Class 1 Device Recall Spectrum Central Venous Catheter Tray, Minocydine/Rifampin Antibiotic Impregnated, Double Lumen

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 1 Device Recall Spectrum Central Venous Catheter Tray, Minocydine/Rifampin Antibiotic Impregnated, Double Lumensee related information
Date Initiated by FirmOctober 04, 2010
Date PostedFebruary 23, 2011
Recall Status1 Terminated 3 on June 21, 2011
Recall NumberZ-1098-2011
Recall Event ID 57663
510(K)NumberK081113 
Product Classification Peripherally Inserted, Central Venous Catheter Tray - Product Code FOZ
ProductSpectrum Central Venous Catheter Tray, Minocydine/Rifampin Antibiotic Impregnated, Double Lumen, 7.5Fr/20cm, CUDLMY751JRSCABRMFST, G13482, Sterile, Cook Medical, Cook Incorporated, Bloomington, IN.
Code Information Part Number: C¿UDLMY¿751J-RSC-ABRM-FST; Global Part Number G13482; Size 7.5Fr/20cm;   Lot numbers: F2361610, F2372057, F2379389. 
FEI Number 1820334
Recalling Firm/
Manufacturer
Cook, Inc.
750 Daniels Way
Bloomington IN 47404-9120
For Additional Information ContactRita A. Harden
812-339-2235
Manufacturer Reason
for Recall
Cook, Inc. is recalling Single, Double, Triple, and Five-Lumen Central Venous Catheter Trays AND Single and Double Lumen PICC Peripherally Inserted Central Venous Catheter Trays. These kits and trays contain 0.9% Sodium Chloride Injection USP 5 mL fill in 6 mL flush syringes that have been recalled by Excelsior Medical. During a routine test for syringe sterility, a possible integrity failure wa
FDA Determined
Cause 2
Nonconforming Material/Component
ActionCook, Inc. sent URGENT DEVICE RECALLS letters to its customers dated 10/4/2010. The letters asked customers to examine their inventory and notify Cook for product return. The letter included a response form.
Quantity in Commerce51,473 total kits/trays
DistributionWorldwide distribution: USA, Chile, Colombia, Panama, Barbados, Canada, Guatemala, and Saudi Arabia.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
510(K) Database510(K)s with Product Code = FOZ
-
-