| | Class 2 Device Recall Phiips Medical Systems |  |
| Date Initiated by Firm | February 07, 2011 |
| Date Posted | March 08, 2011 |
| Recall Status1 |
Terminated 3 on March 16, 2017 |
| Recall Number | Z-1609-2011 |
| Recall Event ID |
57846 |
| 510(K)Number | K051134 |
| Product Classification |
Automated External Defibrillators (non-wearable) - Product Code MKJ
|
| Product | Philips HeartStart MRx Defibrillator Monitor (M3535A/M3536A) with software (SW) versions F.01.00 or F.01.01 and with all of the following hardware options EtCO2, IP and Temp.
Used for the termination of ventricular tachycardia and ventricular fibrillation. |
| Code Information |
Software versions: F.01.00 or F.01.01 |
Recalling Firm/ Manufacturer |
Philips Healthcare Inc. 3000 Minuteman Road Andover MA 01810
|
| For Additional Information Contact | 978-687-1501 |
Manufacturer Reason for Recall | Software: EtCO2 and Ventilation Rate values may be labeled incorrectly. |
FDA Determined Cause 2 | Software design |
| Action | Philips Healthcare representatives issued an Urgent Medical Device letter on 2/7/11.
Customers are asked to follow the Action to be taken by Customer/User section of the Urgent Medical Device Correction Notification. In this section, customers are given the following instructions:
Philips software (SW) version F.01.03 (for M3535A and M3536A) and R.01.00 (for M3536A, English only) corrects this problem.
If you have affected product a Philips Service Engineer will contact you to install software version F.01.03 or R.01.00.
Until your software is updated your HeartStart MRx monitor/defibrillator can remain in service. If you are using the device in one of the use modes listed under problem description, make sure that operators are aware that even though the EtCO2 and ventilation rate values can display interchangeably (unexpectedly or randomly switch) under either the ventilation rate label or the EtCO2 label, the values themselves can be identified by their color (for example if the ventilation rate label is white, the ventilation value is displayed as white).
Customers are to contact their local Philips representative at 1-800-722-9377, if they need further information or support regarding this issue. |
| Quantity in Commerce | 2382 units |
| Distribution | Worldwide -- USA, Canada, AUSTRALIA, AUSTRIA, BAHRAIN, BELGIUM, BRAZIL, CHINA, COLOMBIA, CZECH REPUBLIC, DENMARK, EGYPT, ESTONIA,
FRANCE, GERMANY, HONG KONG, INDIA, IRELAND, ISRAEL, ITALY, KENYA, LATVIA, MALAYSIA, MEXICO, NEW ZEALAND, NORWAY, PHILIPPINES, REUNION, SAUDI ARABIA, SINGAPORE, SOUTH AFRICA, SPAIN, SWITZERLAND, THAILAND, UNITED ARAB EMIRATES, and UNITED KINGDOM. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
| 510(K) Database | 510(K)s with Product Code = MKJ
|
|
|
|