| Class 2 Device Recall BioPlex 2200 EBV IgG Calibrator Set. | |
Date Initiated by Firm | March 11, 2011 |
Date Posted | April 27, 2011 |
Recall Status1 |
Terminated 3 on December 22, 2011 |
Recall Number | Z-2082-2011 |
Recall Event ID |
58291 |
510(K)Number | K062211 |
Product Classification |
Epstein-barr virus, other - Product Code LSE
|
Product | BioPlex 2200 EBV IgG Calibrator Set. Catalog #663-1200.
Product is labeled in part: "BIO-RAD***BioPlex 2200 EBV IgG Calibrator Set***7 x 0.5 mL***A Calibrator set for the BioPlex 2200 EBV IgG Reagent Pack***UNITED STATES, Bio-Rad Laboratories, Inc., Hercules, CA 94547".
BioPlex 2200 EBV IgG Calibrator Set is intended for the calibration of the BioPlex 2200 EBV IgG Reagent Pack. |
Code Information |
Lot #46704. |
Recalling Firm/ Manufacturer |
Bio-Rad Laboratories 6565 185th Ave NE Redmond WA 98052-5039
|
For Additional Information Contact | Bio-Rad Technical Support 510-741-4431 |
Manufacturer Reason for Recall | Bio-Rad Laboratories confirmed a downward shift in control values for EBV NA-1 analyte when using the BioPlex 2200 EBV IgG Calibrator Set Lot #46704. |
FDA Determined Cause 2 | Other |
Action | The firm, Bio-Rad Laboratories, sent an "Voluntary Field Correction: BioPlex 2200 EBV IgG Calibrator Set, Lot# 46704 for EBV NA-1 Analyte" letter dated March 16, 2011 domestic customers via Fed-Ex and and to international subsidiaries via e-mail. The letter described the product, problem and actions to be taken. The consignees/customers were informed that Bio-Rad confirmed downward shift in control values for the EBV NA-1 analyte when using the BioPlex 2200 EBV IgG Calibrator Set, Lot # 46704. If this occurs, the customers were instructed to readjust the positive control mean using the procedure presented in the customer notification "BioPlex 2200 Quality Control Values (L60013402)". If the re-adjusted mean of the EBV NA-1 controls is within the manufacturer's limit, patient results may be reported.
If the new assay positive control mean still does not fall within the manufacturer's limit, the customers were instructed to contact their local Bio-Rad Technical Support and do not report the EBV NA-1 values. The customers were also instructed to complete and return the CUSTOMER RESPONSE FORM via fax to 510-741-5775.
If the BioPlex 2200 EBV VCA IgG and EBV EA-D IgG positive and negative control means are within the acceptable QC range, and subsequent passing QC runs are obtained, patient results may be reported for these analytes.
Bio-Rad will be manufacturing a new lot of the BioPlex 2200 EBV IgG Calibrator Set which will bring the control values within the manufacturer's limit. It may be necessary to readjust the QC ranges after the consignees receive this new calibrator set lot.
Bio-Rad will send their consignees/customers the replacement calibrators and calibrator CDs when available.
The consignees/customers were also instructed to discard the calibrator set lot# 46704 in accordance with national and/or local regulations.
Customers can call Clinical Diagnostics Group at 510-741-4431 or email to BPD-ProductSupport@bio-rad.com for any questions |
Quantity in Commerce | 270 sets |
Distribution | Worldwide distribution: USA including the following states: AL, AZ, CA, CT, FL, GA, HI, IL, IN, LA, MA, MD, MI, MN, NC, NJ, NY, OH, OR, PA, TN, TX, VA, WA, and WI; and countries of: Australia, Czech Republic, France, Germany, Italy and Republic of Korea. |
Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database | 510(K)s with Product Code = LSE
|
|
|
|