| Class 2 Device Recall Cirrus HDOCT Model 4000 instruments |  |
Date Initiated by Firm | March 10, 2011 |
Date Posted | May 02, 2011 |
Recall Status1 |
Terminated 3 on March 29, 2012 |
Recall Number | Z-2127-2011 |
Recall Event ID |
58313 |
510(K)Number | K083291 |
Product Classification |
Tomography, optical coherence - Product Code OBO
|
Product | Cirrus HD-OCT Model 4000 instruments, new and refurbished, distributed with, or upgraded to, version v 1.0 - 4.5, 5.0.0 or 5.1.0
The Cirrus HD-OCT is a non-contact, high resolution tomographic and biomicroscopic imaging device. |
Code Information |
Software kit part number: 2660021136948 ( v.5.0.0) 2660021140099 (v 5.1.0)(US and OUS) part number 2660021136966 - (v 5.0.0 - China) part number 2660021140811 ( v.5.1.0 - China): Serial numers: 4000-5907 4000-5660 4000-2728 4000-4165 4000-5970 4000-5906 4000-5814 4000-5562 4000-5921 4000-5797 4000-5990 4000-6048 4000-5726 4000-6065 4000-5972 4000-5999 4000-1811 4000-5545 4000-6645 4000-5648 4000-6038 4000-5669 4000-6021 4000-5944 4000-6080 4000-5799 4000-5902 4000-1243 4000-5769 4000-1381 4000-2570 4000-5633 4000-4370 4000-5930 4000-5840 4000-5833 4000-3973 4000-4266 4000-2760 4000-2951 4000-4386 4000-5818 4000-6906 4000-5718 4000-6025 4000-5832 4000-6074 4000-3691 4000-1240 4000-6313 4000-4786 4000-1788 4000-5157 4000-3645 4000-5964 4000-1844 4000-1013 4000-2817 4000-3787 4000-5667 4000-1335 4000-2607 4000-5685 4000-3818 4000-7377 4000-5039 4000-5825 4000-4394 4000-1621 4000-6026 4000-2710 4000-4876 4000-4270 4000-3617 4000-2668 4000-3942 4000-4276 4000-2361 4000-2603 4000-5843 4000-6131 4000-3471 4000-1062 4000-5885 4000-1235 4000-1353 4000-3489 4000-3581 4000-5859 4000-6274 4000-1502 4000-7029 4000-2624 4000-3892 4000-6272 4000-3495 4000-4992 4000-3360 4000-5199 4000-4073 4000-4858 4000-1816 4000-5315 4000-4374 4000-5179 4000-3650 4000-4641 4000-3621 4000-4526 4000-5749 4000-7038 4000-1302 4000-7161 4000-5572 4000-2253 4000-2397 4000-2238 4000-2875 4000-6825 4000-5060 4000-2225 4000-5593 4000-1649 4000-5807 4000-6432 4000-5414 4000-3773 4000-3575 4000-2313 4000-3560 4000-3682 4000-3245 4000-3546 4000-3180 4000-6963 4000-1641 4000-1769 4000-5601 4000-4804 4000-1549 4000-5241 4000-3889 4000-5984 4000-3809 4000-1271 4000-4347 4000-6362 4000-5704 4000-6810 4000-4672 4000-5809 4000-6759 4000-3131 4000-6311 4000-4981 4000-1633 4000-1273 4000-1256 4000-4214 4000-4653 4000-2812 4000-1249 4000-6106 4000-4135 4000-6165 4000-1261 4000-4166 4000-4919 4000-2752 4000-3284 4000-3473 4000-5643 4000-2614 4000-2177 4000-5127 4000-3825 4000-1125 4000-5629 4000-2063 4000-3452 4000-2762 4000-5589 4000-3917 4000-2630 4000-5677 4000-1087 4000-1096 4000-6450 4000-6173 4000-7007 4000-6267 4000-3276 4000-1408 4000-1748 4000-3209 4000-3275 4000-6204 4000-2977 4000-2098 4000-2881 4000-3995 4000-2589 4000-5646 4000-3504 4000-2404 4000-3337 4000-4840 4000-3519 4000-3024 4000-1375 4000-4854 4000-6974 4000-3078 4000-3328 4000-4809 4000-5775 4000-5279 4000-7113 4000-3246 4000-2754 4000-6532 4000-6789 4000-4584 4000-1227 4000-6519 4000-5042 4000-1642 4000-5728 4000-5747 4000-3525 4000-2972 4000-4089 4000-5347 4000-4452 4000-3025 4000-3158 4000-7082 4000-4245 4000-4147 4000-6610 4000-4572 4000-6399 4000-2486 4000-3695 4000-5758 4000-6688 4000-7136 4000-3381 4000-1308 4000-1155 4000-6341 4000-2869 4000-6580 4000-5689 4000-2402 4000-5715 4000-5690 4000-1607 4000-1615 4000-4874 4000-5103 4000-5357 4000-4268 4000-6685 4000-4282 4000-1505 4000-5221 4000-6911 4000-3049 4000-7236 4000-4719 4000-6212 4000-7046 4000-4139 4000-7209 4000-5251 4000-3392 4000-6651 4000-4721 4000-1440 4000-3761 4000-2999 4000-2074 4000-6468 4000-3299 4000-4373 4000-5069 4000-2259 4000-3900 4000-4772 4000-4864 4000-6255 4000-3946 4000-3516 4000-6886 4000-5247 4000-2613 4000-5036 4000-5050 4000-3344 4000-2363 4000-4109 4000-6001 4000-6674 4000-3852 4000-4810 4000-3396 4000-4301 4000-5032 4000-2544 4000-2777 4000-3353 4000-2375 4000-7366 4000-5030 4000-4226 4000-6805 4000-3478 4000-4170 4000-6739 4000-4617 4000-5887 4000-2154 4000-4175 4000-3367 4000-2575 4000-2220 4000-3858 4000-6737 4000-3496 4000-5158 4000-4680 4000-2967 4000-7329 4000-3799 4000-6318 4000-4274 4000-3707 4000-6711 4000-5697 4000-4335 4000-2611 4000-2667 4000-2620 4000-3796 4000-2617 4000-4145 4000-4123 4000-3021 4000-3015 4000-1550 4000-7033 4000-2279 4000-1566 4000-6184 4000-5717 4000-3528 4000-2465 4000-6339 4000-6515 4000-4695 4000-1003 4000-5079 4000-3810 4000-4427 4000-4814 4000-4961 4000-7132 4000-4067 4000-4889 4000-2121 4000-4227 4000-3075 4000-1796 4000-6513 4000-3945 4000-6036 4000-3843 4000-4008 4000-2798 4000-4848 4000-2423 4000-6202 4000-6168 4000-3294 4000-3009 4000-2376 4000-3058 4000-1594 4000-6262 4000-1145 4000-5791 4000-6340 4000-6193 4000-6636 4000-5908 4000-5444 4000-4252 4000-6594 4000-6309 4000-3282 4000-2378 4000-4408 4000-4146 4000-6229 4000-7181 4000-4601 4000-2300 4000-4292 4000-4770 4000-5255 4000-6893 4000-2814 4000-3601 4000-3262 4000-3907 4000-6052 4000-6908 4000-4463 4000-3938 4000-3019 4000-1190 4000-5020 4000-4353 4000-6683 4000-3782 4000-1216 4000-3988 4000-3925 4000-2377 4000-2561 4000-5223 4000-1611 4000-3198 4000-7093 4000-5973 4000-4826 4000-3408 4000-4588 4000-6465 4000-5901 4000-5639 4000-2480 4000-6629 4000-3829 4000-3957 4000-2239 4000-2222 4000-1077 4000-3301 4000-6100 4000-2593 4000-6056 4000-6916 4000-1208 4000-6687 4000-4296 4000-4099 4000-4861 4000-7111 4000-6119 4000-6357 4000-3708 4000-1218 4000-6256 4000-1435 4000-2592 4000-3511 4000-6243 4000-4968 4000-4333 4000-6493 4000-3416 4000-7177 4000-7098 4000-7124 4000-7207 4000-5011 4000-4596 4000-2843 4000-3647 4000-3696 4000-6163 4000-1568 4000-2441 4000-2431 4000-6656 4000-5512 4000-4371 4000-6543 4000-5004 4000-3274 4000-2746 4000-7190 4000-6443 4000-3214 4000-2200 4000-2321 4000-4273 4000-1293 4000-4489 4000-1114 4000-6679 4000-2599 4000-3928 4000-6650 4000-7221 4000-6890 4000-2997 4000-3676 4000-1813 4000-5029 4000-3422 4000-1173 4000-5217 4000-4936 4000-4222 4000-1157 4000-5257 4000-6572 4000-2559 4000-5260 4000-3736 4000-1156 4000-5343 4000-4531 4000-4569 4000-6921 4000-1524 4000-6702 4000-6770 4000-6584 4000-4230 4000-3356 4000-1336 4000-2580 4000-6441 4000-6518 4000-2916 4000-3340 4000-7183 4000-2991 4000-4117 4000-7301 4000-3036 4000-4290 4000-5954 4000-3838 4000-3879 4000-2405 4000-4844 4000-2073 4000-5183 4000-2608 4000-7010 4000-6897 4000-2284 4000-3618 4000-3574 4000-1794 4000-5456 4000-2206 4000-5929 4000-2097 4000-2962 4000-2978 4000-3051 4000-6887 4000-5107 4000-3177 4000-1167 4000-2969 4000-3062 4000-6922 4000-2082 4000-4169 4000-4348 4000-1135 4000-6590 4000-4539 4000-4640 4000-5176 4000-5034 4000-3941 4000-3702 4000-4294 4000-3731 4000-1174 4000-5702 4000-5918 4000-7066 4000-6901 4000-3919 4000-2226 4000-4259 4000-2055 4000-3677 4000-1368 4000-4946 4000-6774 4000-2210 4000-1458 4000-5550 4000-5001 4000-3717 4000-1736 4000-1001 4000-2600 4000-2960 4000-3138 4000-2919 4000-2069 4000-3965 4000-6294 4000-5672 4000-3263 4000-2000 4000-4002 4000-3437 4000-2439 4000-2442 4000-2081 4000-3792 4000-6299 4000-4782 4000-5261 4000-4532 4000-4847 4000-3233 4000-2051 4000-4261 4000-2408 4000-3804 4000-3048 4000-6612 4000-4156 4000-2385 4000-5842 4000-7175 4000-3974 4000-6365 4000-4108 4000-4841 4000-4464 4000-1154 4000-4768 4000-6622 4000-2815 4000-3745 4000-4157 4000-7012 4000-3238 4000-1188 4000-4129 4000-5977 4000-4561 4000-7003 4000-5722 4000-5975 4000-3494 4000-4862 4000-1535 4000-1316 4000-5197 4000-3542 4000-4004 4000-6207 4000-1230 4000-4712 4000-4974 4000-1252 4000-6979 4000-4180 4000-2828 4000-6547 4000-6430 4000-1759 4000-3586 4000-3183 4000-5897 4000-5327 4000-1839 4000-5936 4000-3411 4000-5723 4000-3940 4000-6517 4000-6662 4000-3993 4000-5927 4000-3823 4000-2673 4000-4116 4000-3213 4000-3983 4000-4254 4000-1107 4000-2823 4000-5027 4000-2937 4000-6868 4000-6676 4000-1436 4000-6639 4000-3657 4000-3362 4000-2141 4000-5570 4000-3757 4000-6881 4000-3863 4000-3216 4000-5808 4000-3389 4000-3926 4000-1564 4000-5438 4000-3223 4000-5647 4000-5654 4000-5708 4000-7307 4000-3788 4000-2369 4000-1300 4000-5017 4000-3413 4000-6695 4000-6769 4000-2849 4000-3987 4000-4860 4000-3977 4000-6892 4000-6222 4000-4423 4000-3020 4000-6208 4000-2164 4000-5171 4000-5333 4000-3231 4000-5724 4000-7081 4000-5566 4000-7102 4000-6758 4000-2448 4000-4417 4000-3896 4000-3123 4000-4056 4000-2394 4000-7179 4000-6790 4000-3840 4000-6259 4000-3937 4000-6409 4000-3521 4000-5895 4000-5238 4000-3192 4000-4731 4000-6947 4000-6780 4000-3433 4000-3105 4000-1474 4000-2983 4000-6499 4000-2052 4000-5877 4000-6292 4000-2922 4000-5385 4000-7107 4000-4699 4000-4700 4000-5779 4000-6614 4000-3698 4000-6392 4000-3111 4000-1724 4000-4105 4000-3001 4000-3735 4000-4714 4000-2417 4000-5739 4000-7068 4000-2409 4000-5378 4000-3805 4000-6652 4000-6986 4000-4212 4000-4131 4000-3045 4000-5276 4000-6621 4000-3553 4000-3306 4000-3052 4000-3638 4000-5131 4000-1169 4000-6913 4000-5262 4000-6533 4000-2596 4000-6449 4000-5130 4000-7149 4000-7293 4000-2357 4000-3014 4000-2064 4000-7295 4000-6527 4000-1078 4000-6665 4000-2598 4000-6005 4000-6889 4000-3106 4000-3067 4000-3167 4000-6446 4000-6690 4000-6754 4000-6750 4000-6799 4000-4082 4000-6751 4000-4615 4000-4868 4000-4823 4000-3697 4000-5536 4000-3497 4000-7039 4000-2735 4000-2582 4000-4642 4000-4076 4000-6363 4000-3376 4000-1298 4000-7188 4000-4284 4000-6452 4000-5007 4000-5740 4000-6169 4000-5190 4000-5006 4000-6635 4000-4064 4000-7025 4000-4174 4000-6534 4000-7268 4000-4181 4000-5514 4000-4827 4000-2595 4000-3596 4000-6521 4000-6034 4000-2426 4000-6551 4000-5951 4000-4085 4000-6992 4000-5280 4000-3573 4000-4141 4000-4140 4000-4142 4000-3037 4000-1318 4000-3834 4000-3876 4000-3182 4000-5165 4000-6918 4000-7237 4000-6659 4000-5055 4000-6644 4000-5076 4000-6483 4000-4003 4000-4972 4000-3215 4000-2990 4000-1393 4000-5208 4000-3172 4000-5349 4000-4932 4000-5679 4000-3795 4000-4119 4000-6471 4000-5304 4000-5353 4000-2803 4000-5671 4000-2832 4000-7133 4000-5582 4000-1370 4000-6467 4000-4219 4000-6625 4000-6620 4000-2709 4000-4024 4000-6264 4000-3562 4000-2743 4000-2466 4000-2738 4000-5561 4000-5738 4000-5905 4000-3769 4000-2233 4000-1679 4000-4069 4000-3885 4000-6433 4000-7251 4000-6795 4000-7047 4000-1016 4000-3477 4000-2929 4000-4232 4000-6324 4000-5649 4000-3304 4000-3128 4000-3334 4000-1465 4000-2993 4000-7228 4000-6215 4000-5751 4000-4388 4000-4845 4000-5935 4000-6856 4000-5224 4000-2638 4000-5013 4000-7115 4000-3410 4000-2463 4000-4432 4000-3793 4000-3461 4000-3985 4000-4364 4000-1847 4000-2443 4000-3960 4000-3557 4000-3710 4000-6680 4000-2928 4000-3836 4000-2418 4000-4812 4000-3860 4000-3786 4000-3961 4000-7140 4000-5966 4000-5300 4000-2954 4000-3174 4000-6514 4000-2429 4000-5078 4000-4797 4000-3577 4000-3774 4000-2904 4000-6565 4000-3298 4000-2606 4000-6865 4000-2930 4000-7126 4000-1629 4000-3493 4000-5458 4000-2517 4000-3064 4000-5753 4000-4055 4000-3072 4000-5492 4000-2242 4000-2860 4000-5240 4000-4053 4000-3030 4000-4598 4000-7157 4000-5008 4000-3923 4000-4923 4000-1516 4000-6596 4000-5981 4000-1648 4000-2059 4000-5146 4000-3420 4000-4629 4000-6634 4000-4326 4000-7024 4000-6951 4000-2591 4000-5308 4000-2407 4000-6257 4000-4114 4000-4337 4000-3949 4000-6694 4000-7065 4000-7148 4000-3498 4000-6761 4000-5548 4000-3723 4000-3572 4000-1158 4000-4192 4000-3438 4000-7016 4000-3538 4000-1017 4000-2103 4000-5721 4000-6271 4000-1165 4000-5231 4000-6699 4000-1126 4000-2623 4000-6166 4000-2392 4000-7170 4000-6576 4000-7122 4000-2626 4000-2913 4000-4124 4000-6904 4000-5189 4000-1231 4000-1319 4000-3126 4000-4103 4000-6613
4000-1828 4000-4234 4000-3491 4000-3513 4000-5659 4000-2809 4000-4459 4000-2229 4000-4429 4000-3951 4000-2224 4000-1196 4000-3815 4000-6246 4000-3423 4000-6300 4000-3683 4000-5700 4000-4281 4000-3371 4000-3395 4000-3837 4000-5071 4000-2808 4000-6189 4000-6919 4000-1513 4000-2821 4000-3163 4000-5567 4000-5558 4000-3523 4000-2223 4000-5705 4000-6779 4000-5711 4000-6771 4000-5216 4000-3291 4000-2693 4000-3894 4000-4246 4000-3363 4000-2988 4000-2959 4000-1063 4000-2998 4000-5206 4000-4265 4000-2594 4000-4807 4000-2475 4000-4706 4000-3208 4000-3207 4000-6975 4000-5194 4000-3086 4000-1133 4000-4280 4000-6640 4000-5376 4000-2813 4000-5009 4000-5746 4000-4291 4000-3826 4000-6917 4000-4176 4000-4150 4000-2819 4000-3777 4000-5701 4000-1643 4000-2395 4000-3116 4000-2909 4000-6492 4000-3091 4000-1737 4000-4507 4000-6655 4000-5922 4000-4813 4000-7273 4000-3976 4000-6738 4000-3061 4000-6637 4000-4162 4000-3431 4000-5786 4000-4015 4000-3654 4000-5497 4000-1232 4000-7079 4000-4101 4000-6959 4000-2251 4000-1259 4000-3210 4000-6978 4000-5358 4000-5345 4000-7085 4000-2453 4000-7346 4000-6412 4000-1170 4000-4111 4000-2167 4000-5602 4000-3570 4000-4975 4000-3980 4000-6054 4000-6385 4000-2125 4000-7244 4000-4916 4000-1121 4000-2662 4000-1625 4000-1352 4000-5713 4000-5516 4000-1166 4000-3855 4000-3077 4000-4873 4000-1685 4000-3000 4000-5340 4000-2943 4000-2143 4000-1801 4000-1272 4000-1585 4000-2992 4000-1434 4000-3585 4000-5621 4000-3398 4000-6898 4000-1667 4000-5880 4000-6797 4000-5244 4000-5364 4000-4922 4000-5052 4000-1675 4000-6540 4000-2980 4000-4106 4000-6503 4000-5229 4000-6693 4000-3911 4000-5181 4000-2736 4000-4361 4000-3822 4000-1557 4000-3742 4000-1296 4000-7367 4000-3355 4000-3947 4000-3933 4000-5168 4000-3390 4000-5200 4000-3999 4000-5893 4000-3137 4000-6902 4000-2887 4000-7083 4000-2458 4000-1279 4000-2644 4000-6293 4000-4510 4000-2449 4000-3003 4000-3459 4000-4179 4000-2724 4000-6377 4000-5072 4000-3548 4000-6785 4000-3916 4000-4514 4000-1006 4000-3537 4000-3740 4000-6791 4000-2945 4000-1180 4000-1617 4000-4713 4000-7203 4000-4369 4000-3784 4000-6279 4000-5957 4000-7200 4000-6587 4000-6254 4000-5917 4000-5736 4000-1389 4000-4727 4000-6444 4000-3255 4000-6312 4000-2010 4000-2089 4000-5192 4000-3076 4000-3033 4000-5211 4000-4168 4000-7243 4000-2729 4000-6370 4000-6854 4000-3547 4000-6445 4000-5703 4000-4982 4000-6914 4000-6556 4000-2399 4000-5195 4000-1355 4000-7282 4000-3929 4000-1278 4000-4317 4000-6643 4000-6553 4000-3366 4000-1547 4000-6767 4000-4878 4000-6284 4000-3425 4000-2437 4000-4553 4000-2721 4000-6304 4000-3659 4000-2204 4000-1191 4000-6384 4000-2737 4000-4747 4000-4765 4000-5294 4000-1573 4000-3007 4000-3002 4000-2386 4000-7159 4000-6879 4000-4450 4000-2379 4000-1189 4000-7219 4000-4112 4000-3529 4000-3866 4000-2956 4000-7116 4000-7191 4000-7153 4000-3272 4000-6376 4000-4784 4000-3280 4000-3229 4000-3978 4000-2579 4000-1612 4000-2679 4000-3845 4000-6548 4000-7239 4000-6525 4000-3428 4000-3944 4000-6146 4000-4258 4000-2400 4000-5163 4000-5344 4000-2150 4000-5992 4000-6597 4000-6460 4000-5166 4000-1438 4000-1834 4000-2274 4000-2446 4000-5594 4000-2751 4000-2474 4000-3442 4000-6199 4000-6154 4000-1176 4000-6985 4000-6800 4000-6907 4000-6352 4000-1412 4000-5193 4000-2390 4000-3201 4000-6994 4000-7114 4000-3990 4000-1263 4000-6910 4000-2628 4000-4979 4000-7005 4000-6593 4000-7287 4000-4237 4000-6185 4000-1172 4000-7227 4000-4190 4000-4833 4000-1712 4000-7291 4000-1294 4000-2923 4000-4269 4000-4828 4000-3399 4000-6372 4000-5642 4000-3880 4000-5429 4000-6749 4000-5707 4000-2733 4000-5016 4000-6458 4000-4469 4000-4360 4000-6602 4000-3079 4000-5933 4000-5624 4000-7105 4000-4831 4000-1246 4000-7021 4000-4547 4000-3022 4000-7027 4000-6653 4000-5729 4000-5459 4000-7296 4000-7306 4000-7302 4000-2996 4000-6894 4000-5383 4000-4424 4000-4133 4000-4300 4000-2995 4000-1493 4000-4652 4000-5341 4000-6511 4000-3920 4000-7055 4000-6595 4000-6649 4000-3743 4000-1181 4000-1716 4000-2676 4000-4102 4000-1312 4000-6188 4000-3955 4000-3144 4000-2612 4000-6537 4000-5831 4000-1558 4000-6239 4000-2189 4000-1254 4000-4020 4000-5173 4000-5167 4000-3279 4000-5215 4000-1399 4000-2234 4000-2382 4000-4152 4000-4159 4000-4130 4000-5656 4000-5368 4000-1775 4000-6305 4000-5868 4000-4387 4000-5222 4000-7137 4000-7134 4000-7160 4000-3555 4000-6930 4000-2740 4000-1756 4000-4239 4000-6698 4000-1253 4000-3211 4000-1258 4000-4793 4000-6692 4000-4767 4000-6454 4000-4824 4000-3386 4000-3193 4000-6425 4000-1754 4000-2104 4000-5612 4000-6670 4000-4604 4000-2973 4000-3593 4000-6504 4000-4710 4000-7036 4000-3655 4000-5710 4000-1666 4000-2918 4000-7231 4000-7178 4000-4720 4000-4241 4000-6905 4000-5600 4000-5610 4000-1630 4000-4639 4000-6611 4000-4221 4000-2006 4000-4785 4000-2847 4000-2841 4000-2118 4000-4352 4000-5676 4000-2383 4000-7020 4000-3486 4000-1424 4000-6884 4000-6923 4000-2836 4000-3467 4000-6773 4000-3857 4000-5727 4000-3872 4000-4235 4000-5617 4000-1567 4000-3350 4000-4295 4000-4126 4000-1751 4000-2581 4000-1234 4000-3043 4000-4806 4000-3292 4000-6354 4000-4187 4000-5361 4000-5352 4000-7135 4000-5205 4000-3404 4000-2745 4000-3278 4000-3338 4000-6747 4000-6316 4000-2530 4000-6657 4000-6205 4000-3068 4000-4789 4000-4857 4000-2914 4000-6811 4000-5334 4000-6451 4000-7288 4000-5730 4000-1470 4000-2434 4000-5460 4000-2772 4000-4238 4000-4837 4000-3290 4000-5250 4000-1215 4000-2551 4000-6766 4000-4783 4000-6436 4000-3741 4000-5652 4000-4472 4000-4182 4000-4473 4000-1829 4000-7194 4000-2583 4000-7201 4000-3348 4000-5137 4000-3959 4000-3044 4000-3297 4000-7198 4000-5804 4000-5366 4000-4480 4000-7156 4000-2971 4000-6672 4000-4354 4000-4711 4000-4215 4000-4648 4000-5760 4000-1785 4000-3004 4000-5867 4000-3195 4000-6869 4000-3026 4000-2939 4000-3505 4000-3943 4000-4851 4000-4978 4000-4477 4000-6895 4000-4524 4000-4275 4000-5741 4000-6753 4000-2747 4000-5691 4000-6066 4000-3259 4000-3934 4000-6948 4000-6489 4000-1233 4000-3383 4000-5650 4000-4127 4000-4835 4000-3492 4000-6338 4000-3718 4000-4527 4000-1301 4000-3136 4000-5764 4000-6667 4000-6325 4000-5024 4000-6462 4000-4475 4000-5212 4000-1089 4000-3107 4000-4161 4000-2213 4000-1581 4000-4703 4000-5090 4000-1220 4000-2391 4000-6664 4000-3010 4000-6356 4000-4256 4000-4272 4000-6358 4000-5664 4000-3812 4000-5665 4000-3559 4000-5196 4000-7058 4000-3287 4000-5014 4000-4509 4000-4437 4000-4994 4000-1281 4000-6544 4000-2818 4000-7032 4000-3435 4000-3687 4000-6684 4000-3418 4000-1804 4000-7158 4000-5899 4000-2574 4000-2924 4000-2981 4000-6569 4000-4842 4000-2578 4000-5038 4000-7215 4000-7213 4000-6156 4000-1161 4000-4701 4000-4430 4000-4433 4000-4428 4000-2384 4000-4158 4000-6550 4000-5471 4000-1721 4000-3569 4000-7202 4000-3759 4000-2757 4000-7011 4000-6453 4000-2715 4000-2965 4000-6479 4000-2172 4000-1153 4000-3755 4000-4104 4000-5370 4000-6516 4000-6617 4000-6179 4000-6536 4000-7176 4000-1430 4000-2187 4000-2219 4000-4462 4000-4321 4000-6470 4000-6170 4000-2649 4000-2950 4000-6157 4000-2529 4000-4921 4000-5132 4000-2765 4000-6512 4000-7164 4000-3971 4000-6496 4000-7084 4000-5662 4000-6347 4000-7155 4000-3699 4000-4523 4000-1309 4000-4803 4000-1276 4000-1562 4000-7006 4000-4080 4000-4875 4000-6326 4000-6520 4000-3484 4000-6880 4000-4522 4000-4051 4000-2516 4000-2750 4000-2753 4000-2372 4000-3432 4000-7028 4000-6977 4000-1653 4000-3935 4000-2744 4000-6342 4000-7185 4000-4263 4000-3212 4000-5457 4000-6228 4000-7300 4000-6835 4000-1783 4000-4802 4000-3867 4000-1244 4000-6661 4000-4735 4000-3912 4000-3520 4000-4787 4000-4297 4000-7087 4000-1646 4000-5043 4000-2221 4000-4582 4000-3239 4000-3620 4000-5982 4000-5437 4000-1140 4000-1141 4000-3006 4000-5202 4000-6297 4000-3295 4000-5115 4000-4613 4000-6402 4000-5218 4000-4288 4000-3662 4000-5339 4000-2048 4000-6182 4000-5742 4000-3590 4000-5063 4000-7205 4000-7022 4000-6646 4000-3268 4000-2247 4000-4832 4000-3419 4000-3415 4000-3567 4000-2456 4000-3405 4000-4405 4000-2230 4000-2419 4000-6086 4000-6424 4000-2553 4000-5124 4000-4384 4000-5012 4000-3778 4000-1433 4000-2265 4000-1484 4000-4093 4000-3322 4000-3027 4000-3884 4000-6448 4000-5053 4000-5578 4000-6192 4000-6864 4000-4983 4000-3979 4000-1580 4000-1236 4000-7008 4000-2460 4000-5106 4000-3414 4000-4796 4000-1168 4000-5210 4000-4220 4000-3561 4000-5849 4000-2467 4000-6315 4000-7097 4000-7086 4000-3436 4000-2601 4000-4819 4000-5955 4000-1159 4000-1392 4000-4365 4000-4224 4000-7189 4000-5230 4000-3288 4000-3948 4000-2633 4000-6023 4000-2674 4000-4745 4000-2938 4000-6700 4000-5000 4000-2393 4000-1367 4000-2243 4000-4538 4000-6568 4000-4418 4000-7052 4000-5693 4000-5237 4000-2933 4000-2894 4000-2297 4000-5374 4000-1117 4000-6404 4000-7206 4000-3488 4000-5369 4000-7351 4000-2415 4000-5588 4000-6213 4000-1095 4000-1314 4000-3479 4000-5607 4000-1772 4000-6434 4000-6435 4000-3984 4000-4172 4000-6600 4000-1395 4000-6776 4000-3401 4000-4185 4000-1149 4000-4173 4000-1396 4000-2182 4000-1124 4000-4083 4000-3584 4000-4077 4000-3394 4000-4363 4000-1835 4000-6269 4000-4705 4000-1069 4000-3791 4000-2826 4000-2618 4000-1323 4000-2250 4000-4356 4000-5542 4000-6896 4000-7019 4000-6748 4000-4315 4000-4708 4000-3558 4000-1118 4000-3564 4000-3672 4000-4839 4000-1142 4000-4816 4000-4856 4000-1350 4000-7294 4000-4654 4000-4425 4000-5028 4000-1305 4000-2029 4000-2054 4000-6007 4000-3694 4000-7187 4000-6258 4000-7232 4000-6218 4000-2878 4000-3227 4000-5428 4000-2062 4000-1446 4000-6883 4000-3994 4000-2403 4000-1472 4000-4079 4000-2414 4000-5259 4000-3343 4000-4752 4000-1303 4000-4287 4000-4125 4000-3188 4000-2911 4000-6395 4000-6263 4000-3502 4000-3487 4000-1437 4000-6691 4000-1289 4000-4355 4000-7333 4000-3130 4000-5219 4000-4697 4000-1565 4000-4376 4000-3797 4000-2398 4000-4134 4000-4345 4000-4090 4000-3403 4000-1511 4000-2989 4000-1021 4000-5209 4000-5324 4000-3915 4000-2374 4000-7057 4000-2477 4000-4057 4000-6628 4000-3443 4000-1277 4000-3186 4000-7060 4000-2619 4000-2602 4000-4869 4000-5162 4000-5356 4000-6618 4000-7341 4000-4115 4000-5254 4000-2257 4000-3730 4000-5733 4000-6582 4000-5748 4000-5351 4000-1439 4000-6697 4000-5177 4000-4070 4000-4171 4000-1068 4000-1765 4000-4231 4000-1310 4000-7230 4000-4821 4000-3800 4000-1475 4000-3656 4000-5061 4000-3580 4000-7096 4000-3909 4000-2388 4000-4251 4000-3499 4000-7405 4000-4943 4000-1113 4000-6819 4000-1299 4000-3374 4000-4952 4000-6604 4000-4605 4000-4704 4000-2459 4000-3346 4000-4138 4000-6440 4000-5372 4000-6626 4000-5452 4000-6526 4000-1187 4000-2609 4000-3568 4000-4121 4000-5943 4000-1147 4000-6394 4000-2935 4000-3440 4000-4573 4000-7018 4000-5731 4000-5161 4000-3737 4000-5491 4000-3895 4000-2370 4000-3918 4000-1313 4000-6993 4000-6296 4000-6391 4000-1714 4000-4403 4000-7004 4000-5694 4000-3217 4000-2106 4000-1164 4000-3426 4000-4164 4000-6382 4000-1175 4000-3016 4000-1390 4000-5214 4000-5712 4000-5490 4000-5584 4000-4937 4000-1320 4000-4759 4000-1598 4000-6410 4000-3429 4000-4958 4000-5942 4000-7214 4000-4304 4000-6686 4000-4729 4000-4722 4000-2927 4000-6564 4000-5475 4000-5473 4000-7151 4000-2139 4000-4624 4000-2110 4000-1556 4000-2468 4000-6574 4000-3032 4000-6900 4000-2071 4000-3397 4000-2420 4000-3549 4000-3615 4000-6920 4000-3219 4000-5971 4000-6390 4000-5025 4000-3610 4000-5201 4000-4945 4000-5362 4000-3029 4000-2863 4000-5120 4000-3545 4000-4499 4000-2646 4000-3190 4000-1282 4000-6524 4000-7013 4000-5346 4000-2797 4000-4113 4000-6147 4000-1763 4000-2119 4000-5225 4000-6705 4000-6198 4000-6383 4000-6429 4000-3008 4000-6764 4000-1171 4000-6638 4000-7233 4000-7044 4000-3888 4000-6064 4000-6763 4000-6591 4000-2770 4000-6055 4000-5253 4000-7350 4000-5087 4000-4478 4000-2636 4000-3963 4000-6014 4000-6932 4000-1306 4000-5204 4000-4148 4000-6187 4000-3382 4000-3121 4000-4973 4000-6988 4000-7131 4000-4087 4000-4088 4000-4351 4000-4938 4000-1776 4000-2228 4000-1740 4000-2986 4000-1119 4000-5448 4000-1255 4000-5355 4000-4683 4000-3864 4000-4849 4000-7319 4000-1287 4000-7145 4000-1510 4000-6501 4000-2109 4000-4818 4000-2573 4000-4243 4000-1554 4000-4071 4000-6067 4000-3224 4000-3527 4000-5752 4000-6681 4000-7165 4000-3966 4000-3114 4000-2425 4000-3230 4000-5328 4000-5088 4000-4286 4000-6260 4000-6387 4000-3582 4000-2880 4000-6632 4000-4054 4000-3734 4000-7212 4000-3981 4000-5175 4000-2112 4000-6369 4000-3927 4000-6984 4000-2159 4000-2932 4000-3185 4000-6903 4000-4250 4000-4846 4000-3565 4000-7063 4000-1779 4000-1311 4000-5129 4000-5632 4000-3721 4000-3897 4000-1579 4000-3669 4000-4718 4000-2789 4000-4822 4000-3359 4000-3870 4000-3967 4000-1522 4000-3057 4000-5714 4000-1791 4000-6447 4000-7395 4000-3982 4000-1143 4000-3354 4000-2352 4000-1237 4000-5198 4000-7173 4000-5329 4000-2102 4000-2854 4000-6349 4000-5141 4000-6463 4000-2643 4000-3385 4000-6567 4000-7298 4000-3184 4000-4474 4000-4530 4000-3347 4000-3059 4000-5234 4000-2142 4000-6331 4000-6416 4000-4047 4000-6431 4000-6009 4000-4777 4000-6875 4000-6756 4000-1388 4000-1148 4000-5031 4000-4774 4000-2641 4000-7130 4000-2632 4000-6675 4000-1163 4000-4769 4000-2982 4000-5373 4000-2373 4000-5606 4000-6361 4000-5227 4000-3269 4000-1291 4000-4581 4000-3011 4000-1750 4000-2979 4000-4034 4000-4481 4000-6321 4000-4957 4000-3341 4000-1130 4000-1444 4000-6627 4000-2416 4000-2067 4000-1162 4000-5900 4000-1245 4000-4933 4000-6743 4000-1295 4000-4866 4000-1799 4000-5033 4000-5744 4000-2396 4000-5023 4000-1366 4000-4970 4000-4299 4000-3485 4000-6210 4000-3972 4000-2413 4000-5716 4000-3196 4000-5439 4000-4277 4000-4482 4000-7283 4000-7297 4000-3903 4000-2915 4000-7150 4000-4084 4000-5533 4000-1743 4000-5801 4000-3821 4000-5174 4000-5213 4000-3462 4000-6388 4000-2295 4000-2162 4000-3578 4000-4790 4000-6577 4000-6557 4000-4915 4000-1682 4000-6654 4000-6666 4000-5246 4000-5756 4000-2865 4000-3556 4000-3221 4000-5191 4000-4579 4000-4795 4000-4298 4000-7163 4000-3187 4000-6281 4000-6787 4000-2470 4000-4521 4000-1007 4000-4980 4000-5081 4000-5969 4000-5281 4000-3069 4000-1093 4000-6813 4000-5184 4000-3714 4000-7056 4000-3666 4000-3055 4000-1317 4000-1284 4000-5898 4000-3453 4000-3439 4000-6678 4000-1072 4000-1022 4000-2822 4000-1257 4000-5245 4000-1008 4000-7000 4000-1652 4000-2117 4000-3305 4000-1014 4000-1413 4000-5021 4000-5595 4000-7101 4000-2315 4000-2358 4000-2901 4000-6736 4000-5931 4000-3875 4000-6641 4000-4616 4000-1376 4000-1260 4000-1011 4000-1356 4000-4838 4000-1703 4000-4100 4000-4726 4000-3339 4000-4658 4000-1473 4000-7017 4000-6413 4000-7152 4000-3986 4000-2585 4000-3873 4000-5928 4000-4755 4000-4591 4000-5159 4000-3203 4000-5035 4000-6411 4000-6317 4000-3954 4000-2451 4000-4558 4000-6760 4000-4257 4000-5203 4000-6174 4000-6335 4000-4346 4000-3017 4000-1274 4000-5133 4000-4385 4000-4407 4000-2739 4000-2450 4000-2430 4000-5888 4000-1100 4000-5180 4000-1115 4000-2202 4000-2469 4000-3969 4000-2961 4000-1674 4000-2359 4000-6929 4000-6579 4000-7266 4000-2917 4000-5164 4000-2368 4000-3512 4000-6295 4000-6786 4000-2912 4000-7331 4000-3200 4000-3194 4000-6668 4000-5015 4000-1111 4000-3050 4000-6689 4000-1285 4000-6633 4000-6327 4000-6507 4000-5178 4000-6682 4000-2454 4000-6803 4000-6455 4000-6996 4000-6778 4000-3191 4000-6775 4000-3597 4000-7286 4000-2906 4000-2902 4000-2900 4000-6417 4000-6777 4000-2670 4000-1441 4000-2825 4000-2874 4000-3199 4000-2447 4000-2907 4000-5018 4000-2877 4000-5182 4000-3205 4000-3197 4000-5066 4000-1275 4000-4593 4000-2216 4000-6155 4000-5367 4000-2406 4000-3827 4000-2984 4000-1800 4000-6808 4000-6084 4000-5185 4000-1623 4000-2108 4000-4724 4000-2227 4000-4934 4000-4627 4000-2232 4000-4468 4000-3109 4000-6288 4000-6601 4000-1351 4000-1179 4000-5732 4000-6240 4000-2367 4000-6282 4000-2975 4000-3380 4000-3253 4000-3901 4000-4018 4000-4947 4000-3204 4000-4118 4000-2421 4000-3285 4000-6624 4000-5062 4000-3628 4000-5745 4000-5419 4000-5256 4000-3053 4000-3202 4000-3692 4000-1247 4000-5834 4000-2401 4000-2411 4000-4788 4000-5719 4000-3859 4000-4503 4000-6428 4000-2065 4000-3958 4000-4107 4000-3476 4000-1283 4000-6623 4000-1280 4000-2013 4000-4091 4000-1076 4000-1083 4000-5331 4000-3551 4000-5855 4000-4998 4000-3539 4000-4167 4000-4476 4000-7166 4000-5348 4000-1364 4000-4233 4000-1509 4000-5318 4000-6765 4000-1292 4000-6033 4000-6558 4000-6171 4000-2966 4000-4536 4000-2976 4000-5952 4000-1152 4000-5565 4000-4096 4000-4092 4000-2921 4000-6303 4000-4191 4000-3028 4000-2734 4000-5875 4000-7199 4000-4163 4000-4761 4000-7318 4000-3417 4000-5207 4000-2627 4000-6373 4000-1304 4000-1213 4000-7070 4000-3968 4000-4723 4000-5019 4000-6286 4000-4863 4000-3668 4000-6701 4000-6393 4000-5326 4000-3065 4000-7193 4000-3653 4000-6793 4000-4072 4000-4184 4000-6658 4000-2622 4000-2571 4000-3831 4000-4293 4000-3939 4000-6528 4000-3329 4000-3335 4000-6891 4000-2535 4000-3975 4000-3854 4000-4457 4000-6323 4000-4001 4000-5653 4000-2666 4000-2115 4000-4426 4000-2616 4000-2637 4000-4398 4000-6619 4000-6768 4000-4935 4000-3047 4000-5910 4000-5945 4000-3964 4000-5720 4000-5985 4000-5896 4000-6051 4000-6076 4000-5956 4000-5794 4000-6000 4000-6050 4000-5881 4000-6072 4000-5979 4000-6002 4000-4029 4000-5750 4000-6647 4000-5865 4000-5026 4000-5863 4000-6040 4000-5948 4000-5811 4000-5926 4000-2964 4000-4763 4000-3038 4000-4149 4000-6070 4000-5835 4000-5989 4000-6027 4000-3355 4000-6008 4000-6583 4000-5823 4000-5852 4000-6046 4000-5836 4000-5946 4000-3887 4000-3365 4000-6314 4000-4794 4000-2827 4000-5186 4000-6153 4000-5980 4000-2804 4000-1352 4000-2820 4000-6150 4000-1718 4000-6209 4000-7078 4000-6542 4000-2920 4000-6261 4000-7127 4000-6350 4000-5591 4000-4283 4000-5840 4000-4853 4000-2380 4000-6151 4000-6990 4000-2218 4000-5938 4000-6704 4000-7162 4000-4075 4000-5919 4000-6277 4000-1002 4000-7226 4000-4543 4000-2480 4000-3562 4000-6480 4000-7182 4000-5242 4000-7217 4000-6939 4000-6608 4000-5187 4000-4643 4000-1744 4000-7254 4000-2213 4000-2811 4000-7184 4000-6334 4000-6746 4000-3970 4000-2958 4000-3051 4000-3566 4000-2069 4000-5998 4000-7146 4000-5130 4000-3735 4000-2926 4000-5049 4000-6563 4000-3236 4000-4622 4000-7171 4000-1786 4000-7280 4000-6310 4000-3784 4000-5912 4000-5037 4000-1297 4000-5604 4000-4372 4000-4859 4000-5809 4000-6696 4000-3643 4000-6812 4000-5239 4000-4630 4000-3727 4000-4381 4000-4656 4000-4249 4000-6586 4000-4762 4000-3956 4000-5916 4000-4178 4000-4919 4000-3434 4000-3475 4000-5674 4000-7076 4000-6560 4000-5233 4000-4638 4000-1286 4000-6663 4000-5128 4000-2471 4000-3950 4000-6466 4000-2974 4000-2462 4000-6862 4000-6581 4000-7197 4000-5846 4000-4271 4000-2677 4000-3031 4000-4151 4000-3790 4000-7347 4000-6938 4000-3257 4000-4155 4000-4267 4000-6943 4000-6721 4000-6250 4000-3576 4000-5075 4000-5920 4000-5947 4000-3189 4000-5939 4000-5962 4000-5812 4000-6060 4000-6078 4000-5978 4000-5810 4000-5988 4000-5821 4000-5850 4000-6028 4000-6019 4000-6044 4000-4458 4000-5793 4000-7249 4000-5866 4000-5958 4000-5871 4000-5986 4000-5816 4000-5891 4000-4247 4000-6234 4000-3142 4000-6071 4000-5220 4000-5924 4000-6478 4000-4264 4000-6029 4000-6016 4000-1120 4000-3930 4000-1091 4000-6915 4000-4471 4000-5170 4000-6162 4000-5991 4000-6337 4000-5925 4000-5949 4000-1601 4000-5884 4000-5994 4000-5894 4000-5784 4000-5817 4000-6035 4000-6003 4000-5822 4000-6057 4000-6058 4000-6022 4000-6059 4000-4479 4000-7250 4000-5870 4000-5768 4000-5911 4000-4110 4000-2597 4000-5243 4000-6077 4000-5950 4000-2807 4000-5879 4000-5995 4000-5856 4000-6031 4000-6047 4000-6045 4000-5853 4000-5824 4000-6024 4000-6068 4000-4486 4000-7252 4000-5837 4000-6093 4000-5953 4000-3225 4000-5883 4000-6010 4000-4015 4000-6053 4000-5965 4000-6083 4000-6041 4000-5882 4000-5961 4000-3226 4000-6094 4000-6012 4000-5856 4000-6073 4000-5886 4000-5963 4000-5913 4000-6097 4000-6018 4000-5886 4000-6061 4000-6042 4000-3375 4000-6037 4000-3715 4000-5890 4000-6020 4000-3801 4000-6043 4000-3890 4000-4189: International: 4000-5872 4000-2908 4000-3989 4000-3996 4000-4600 4000-4614 4000-5064 4000-5792 4000-6177 4000-6186 4000-6573 4000-6762 4000-1020 4000-1067 4000-1073 4000-1160 4000-1194 4000-1195 4000-1225 4000-1242 4000-1248 4000-1265 4000-1288 4000-1290 4000-1325 4000-1330 4000-1432 4000-1451 4000-1530 4000-1531 4000-1543 4000-1545 4000-1551 4000-1561 4000-1563 4000-1577 4000-1686 4000-1690 4000-1695 4000-1702 4000-1761 4000-1764 4000-1766 4000-1768 4000-1770 4000-1778 4000-1787 4000-2057 4000-2083 4000-2092 4000-2093 4000-2094 4000-2095 4000-2101 4000-2291 4000-2308 4000-2310 4000-2339 4000-2340 4000-2342 4000-2345 4000-2347 4000-2360 4000-2362 4000-2365 4000-2366 4000-2508 4000-2562 4000-2564 4000-2566 4000-2568 4000-2569 4000-2576 4000-2691 4000-2701 4000-2704 4000-2711 4000-2714 4000-2716 4000-2718 4000-2850 4000-2885 4000-2888 4000-2892 4000-2897 4000-2899 4000-2903 4000-2925 4000-2934 4000-2936 4000-2940 4000-2941 4000-2942 4000-2944 4000-2946 4000-2948 4000-2949 4000-2952 4000-2953 4000-2955 4000-2957 4000-2968 4000-3454 4000-3474 4000-3480 4000-3481 4000-3482 4000-3501 4000-3517 4000-3665 4000-3675 4000-3678 4000-3729 4000-3746 4000-3747 4000-3817 4000-3830 4000-3833 4000-3877 4000-3905 4000-3931 4000-3998 4000-4007 4000-4009 4000-4030 4000-4039 4000-4153 4000-4390 4000-4401 4000-4421 4000-4431 4000-4434 4000-4435 4000-4439 4000-4440 4000-4441 4000-4445 4000-4446 4000-4449 4000-4456 4000-4465 4000-4467 4000-4470 4000-4529 4000-4554 4000-4559 4000-4564 4000-4580 4000-4583 4000-4623 4000-4625 4000-4734 4000-4744 4000-4746 4000-4749 4000-4751 4000-4766 4000-4771 4000-4773 4000-4779 4000-4912 4000-4924 4000-4925 4000-4977 4000-4987 4000-4988 4000-5041 4000-5045 4000-5054 4000-5067 4000-5264 4000-5266 4000-5267 4000-5269 4000-5270 4000-5271 4000-5277 4000-5289 4000-5290 4000-5295 4000-5299 4000-5305 4000-5307 4000-5309 4000-5310 4000-5321 4000-5388 4000-5391 4000-5392 4000-5430 4000-5442 4000-5446 4000-5447 4000-5449 4000-5451 4000-5453 4000-5661 4000-5666 4000-5680 4000-5790 4000-5819 4000-5828 4000-5987 4000-6069 4000-6092 4000-6183 4000-6247 4000-6268 4000-6276 4000-6280 4000-6328 4000-6329 4000-6332 4000-6336 4000-6348 4000-6366 4000-6371 4000-6380 4000-6386 4000-6389 4000-6400 4000-6401 4000-6403 4000-6405 4000-6414 4000-6415 4000-6421 4000-6459 4000-6473 4000-6486 4000-6488 4000-6575 4000-6648 4000-6809 4000-6818 4000-6877 4000-6924 4000-6933 4000-6935 4000-6964 4000-6982 4000-6983 4000-6987 4000-6989 4000-6997 4000-7037 4000-7041 4000-7042 4000-7043 4000-7048 4000-7050 4000-7051 4000-7054 4000-7059 4000-7061 4000-7064 4000-7067 4000-7077 4000-7208 4000-3490 4000-4634 4000-5005 4000-5844 4000-3644 4000-4886 4000-7015 4000-1613 4000-1624 4000-1661 4000-1699 4000-1707 4000-1713 4000-1760 4000-1780 4000-2085 4000-2088 4000-2090 4000-2091 4000-2240 4000-2244 4000-2245 4000-2246 4000-2248 4000-3237 4000-3671 4000-3733 4000-3779 4000-4303 4000-4525 4000-4535 4000-4544 4000-4549 4000-4570 4000-4754 4000-4764 4000-5316 4000-5325 4000-5332 4000-5443 4000-5454 4000-5564 4000-5571 4000-5573 4000-5575 4000-5576 4000-5577 4000-5583 4000-5640 4000-5675 4000-5692 4000-5695 4000-5830 4000-5838 4000-5960 4000-6102 4000-6104 4000-6107 4000-6108 4000-6109 4000-6115 4000-6236 4000-6242 4000-6397 4000-6423 4000-6820 4000-6822 4000-6830 4000-6839 4000-6843 4000-6844 4000-6937 4000-6940 4000-6945 4000-6952 4000-6953 4000-6955 4000-7053 4000-7072 4000-7075 4000-7080 4000-7269 4000-7270 4000-7327 4000-7399 4000-1000 4000-1019 4000-1129 4000-1131 4000-1226 4000-1241 4000-1343 4000-1349 4000-1358 4000-1409 4000-1410 4000-1411 4000-1415 4000-1420 4000-1427 4000-1464 4000-1478 4000-1481 4000-1482 4000-1485 4000-1559 4000-1572 4000-1578 4000-1587 4000-1590 4000-1592 4000-1622 4000-1638 4000-1640 4000-1651 4000-1723 4000-1728 4000-1742 4000-1747 4000-1815 4000-1826 4000-1832 4000-1836 4000-1841 4000-1842 4000-1843 4000-1849 4000-2105 4000-2132 4000-2146 4000-2158 4000-2301 4000-2304 4000-2305 4000-2307 4000-2316 4000-2433 4000-2436 4000-2455 4000-2457 4000-2461 4000-2494 4000-2523 4000-2660 4000-2680 4000-2708 4000-2712 4000-2717 4000-2719 4000-2720 4000-2722 4000-2723 4000-2726 4000-2732 4000-2776 4000-2963 4000-2987 4000-3060 4000-3154 4000-3156 4000-3164 4000-3166 4000-3168 4000-3170 4000-3173 4000-3312 4000-3316 4000-3319 4000-3324 4000-3506 4000-3509 4000-3514 4000-3518 4000-3522 4000-3651 4000-3686 4000-3700 4000-3724 4000-3725 4000-3921 4000-3936 4000-3997 4000-4023 4000-4025 4000-4330 4000-4331 4000-4339 4000-4342 4000-4343 4000-4563 4000-4565 4000-4567 4000-4574 4000-4619 4000-4621 4000-4673 4000-4675 4000-4678 4000-4682 4000-4688 4000-4691 4000-4692 4000-4693 4000-4698 4000-4709 4000-4715 4000-4748 4000-4757 4000-4760 4000-4894 4000-4950 4000-4962 4000-4964 4000-4969 4000-4971 4000-5098 4000-5099 4000-5102 4000-5109 4000-5117 4000-5144 4000-5169 4000-5172 4000-5188 4000-5226 4000-5263 4000-5297 4000-5311 4000-5337 4000-5390 4000-5393 4000-5395 4000-5396 4000-5400 4000-5402 4000-5405 4000-5413 4000-5464 4000-5468 4000-5472 4000-5506 4000-5544 4000-5568 4000-5579 4000-5581 4000-5586 4000-5587 4000-5590 4000-5599 4000-5614 4000-5616 4000-5620 4000-5626 4000-5909 4000-5959 4000-5967 4000-5983 4000-6011 4000-6206 4000-6216 4000-6225 4000-6226 4000-6249 4000-6283 4000-6285 4000-6291 4000-6298 4000-6301 4000-6306 4000-6308 4000-6319 4000-6330 4000-6333 4000-6343 4000-6346 4000-6351 4000-6355 4000-6367 4000-6374 4000-6378 4000-6381 4000-6545 4000-6598 4000-6603 4000-6605 4000-6606 4000-6631 4000-6673 4000-6740 4000-6742 4000-6745 4000-6755 4000-6782 4000-6783 4000-6792 4000-6798 4000-6867 4000-6909 4000-6967 4000-6991 4000-6995 4000-7001 4000-7002 4000-7014 4000-7026 4000-7030 4000-7031 4000-7035 4000-7257 4000-7258 4000-7275 4000-7315 4000-7322 4000-7325 4000-7330 4000-7334 4000-3317 4000-3588 4000-3571 4000-3591 4000-3599 4000-3589 4000-3598 4000-3600 4000-3603 4000-3606 4000-1584 4000-1688 4000-1693 4000-2555 4000-2855 4000-3220 4000-3430 4000-4607 4000-4610 4000-4879 4000-4984 4000-5342 4000-5580 4000-6541 4000-7069 4000-7094 4000-7100 4000-1092 4000-1102 4000-1106 4000-2758 4000-2796 4000-3071 4000-3122 4000-3171 4000-3323 4000-3409 4000-3412 4000-3463 4000-3468 4000-3469 4000-3483 4000-3503 4000-3679 4000-3716 4000-3722 4000-3794 4000-3811 4000-3865 4000-3924 4000-3952 4000-4010 4000-4048 4000-4052 4000-4058 4000-4068 4000-4094 4000-4154 4000-4199 4000-4236 4000-4248 4000-4253 4000-4260 4000-4262 4000-4383 4000-4592 4000-4655 4000-4926 4000-4955 4000-4966 4000-4967 4000-4976 4000-5147 4000-5148 4000-5149 4000-5150 4000-5151 4000-5152 4000-5153 4000-5338 4000-5462 4000-5463 4000-5502 4000-5503 4000-5504 4000-5509 4000-5510 4000-5518 4000-5519 4000-5520 4000-5625 4000-5772 4000-5774 4000-6116 4000-6122 4000-6128 4000-6133 4000-6134 4000-6152 4000-6251 4000-6252 4000-6265 4000-6266 4000-6359 4000-6360 4000-6475 4000-6523 4000-6561 4000-6615 4000-6616 4000-6715 4000-6716 4000-6842 4000-6846 4000-6847 4000-6848 4000-6849 4000-6878 4000-6888 4000-7034 4000-7117 4000-7123 4000-7141 4000-7142 4000-7143 4000-7167 4000-7174 4000-7235 4000-7265 4000-7312 4000-7314 4000-7317 4000-7369 4000-1192 4000-2910 4000-4078 4000-4444 4000-5046 4000-5113 4000-5232 4000-5235 4000-5236 4000-6874 4000-6876 4000-6956 4000-6961 4000-6270 4000-1105 4000-1838 4000-2241 4000-2554 4000-3806 4000-3808 4000-4402 4000-4451 4000-4461 4000-5569 4000-6096 4000-1589 4000-6802 4000-1004 4000-1005 4000-1015 4000-1024 4000-1079 4000-1080 4000-1081 4000-1082 4000-1084 4000-1097 4000-1101 4000-1103 4000-1108 4000-1110 4000-1112 4000-1116 4000-1122 4000-1123 4000-1127 4000-1134 4000-1137 4000-1139 4000-1150 4000-1177 4000-1185 4000-1193 4000-1197 4000-1206 4000-1207 4000-1211 4000-1212 4000-1214 4000-1251 4000-1267 4000-1315 4000-1321 4000-1322 4000-1326 4000-1327 4000-1329 4000-1332 4000-1333 4000-1334 4000-1338 4000-1340 4000-1341 4000-1344 4000-1348 4000-1360 4000-1361 4000-1362 4000-1365 4000-1372 4000-1373 4000-1380 4000-1382 4000-1386 4000-1391 4000-1394 4000-1397 4000-1398 4000-1400 4000-1401 4000-1403 4000-1414 4000-1416 4000-1417 4000-1418 4000-1421 4000-1423 4000-1426 4000-1442 4000-1447 4000-1448 4000-1449 4000-1450 4000-1453 4000-1454 4000-1459 4000-1460 4000-1461 4000-1468 4000-1469 4000-1471 4000-1477 4000-1479 4000-1480 4000-1487 4000-1488 4000-1490 4000-1491 4000-1492 4000-1494 4000-1495 4000-1496 4000-1498 4000-1499 4000-1501 4000-1503 4000-1506 4000-1507 4000-1518 4000-1519 4000-1521 4000-1523 4000-1525 4000-1526 4000-1527 4000-1528 4000-1529 4000-1532 4000-1534 4000-1536 4000-1539 4000-1540 4000-1544 4000-1548 4000-1552 4000-1553 4000-1555 4000-1569 4000-1570 4000-1571 4000-1575 4000-1576 4000-1583 4000-1600 4000-1603 4000-1605 4000-1606 4000-1608 4000-1616 4000-1618 4000-1619 4000-1626 4000-1627 4000-1632 4000-1635 4000-1636 4000-1654 4000-1657 4000-1658 4000-1659 4000-1668 4000-1677 4000-1680 4000-1681 4000-1691 4000-1694 4000-1696 4000-1697 4000-1698 4000-1700 4000-1701 4000-1704 4000-1705 4000-1706 4000-1708 4000-1709 4000-1710 4000-1711 4000-1722 4000-1725 4000-1727 4000-1729 4000-1730 4000-1731 4000-1732 4000-1733 4000-1734 4000-1735 4000-1738 4000-1739 4000-1741 4000-1745 4000-1746 4000-1749 4000-1752 4000-1753 4000-1755 4000-1757 4000-1758 4000-1762 4000-1774 4000-1777 4000-1781 4000-1782 4000-1789 4000-1790 4000-1793 4000-1795 4000-1805 4000-1807 4000-1810 4000-1822 4000-1827 4000-1831 4000-1833 4000-1837 4000-1845 4000-1846 4000-1848 4000-2007 4000-2018 4000-2027 4000-2030 4000-2031 4000-2032 4000-2033 4000-2034 4000-2038 4000-2039 4000-2040 4000-2043 4000-2045 4000-2047 4000-2056 4000-2058 4000-2061 4000-2068 4000-2072 4000-2075 4000-2076 4000-2077 4000-2079 4000-2080 4000-2084 4000-2086 4000-2111 4000-2113 4000-2114 4000-2116 4000-2122 4000-2123 4000-2128 4000-2130 4000-2144 4000-2145 4000-2147 4000-2148 4000-2149 4000-2153 4000-2155 4000-2157 4000-2160 4000-2163 4000-2165 4000-2168 4000-2170 4000-2171 4000-2173 4000-2175 4000-2178 4000-2180 4000-2185 4000-2186 4000-2190 4000-2194 4000-2195 4000-2196 4000-2197 4000-2198 4000-2217 4000-2252 4000-2255 4000-2258 4000-2261 4000-2262 4000-2263 4000-2266 4000-2268 4000-2269 4000-2271 4000-2272 4000-2275 4000-2276 4000-2278 4000-2282 4000-2283 4000-2292 4000-2312 4000-2314 4000-2317 4000-2318 4000-2319 4000-2320 4000-2322 4000-2324 4000-2325 4000-2326 4000-2327 4000-2328 4000-2329 4000-2330 4000-2331 4000-2332 4000-2333 4000-2334 4000-2335 4000-2336 4000-2337 4000-2338 4000-2341 4000-2343 4000-2344 4000-2346 4000-2348 4000-2349 4000-2350 4000-2351 4000-2353 4000-2354 4000-2356 4000-2387 4000-2389 4000-2452 4000-2479 4000-2484 4000-2487 4000-2488 4000-2489 4000-2493 4000-2496 4000-2498 4000-2502 4000-2503 4000-2504 4000-2506 4000-2507 4000-2509 4000-2510 4000-2511 4000-2512 4000-2513 4000-2514 4000-2515 4000-2518 4000-2519 4000-2520 4000-2521 4000-2522 4000-2524 4000-2525 4000-2526 4000-2528 4000-2531 4000-2533 4000-2534 4000-2537 4000-2538 4000-2541 4000-2543 4000-2547 4000-2550 4000-2572 4000-2584 4000-2586 4000-2587 4000-2588 4000-2604 4000-2605 4000-2610 4000-2615 4000-2621 4000-2625 4000-2629 4000-2631 4000-2634 4000-2639 4000-2640 4000-2642 4000-2645 4000-2647 4000-2650 4000-2651 4000-2652 4000-2653 4000-2654 4000-2661 4000-2671 4000-2672 4000-2675 4000-2678 4000-2681 4000-2682 4000-2683 4000-2684 4000-2685 4000-2686 4000-2687 4000-2688 4000-2690 4000-2692 4000-2694 4000-2696 4000-2697 4000-2698 4000-2699 4000-2700 4000-2703 4000-2705 4000-2707 4000-2713 4000-2727 4000-2756 4000-2763 4000-2764 4000-2766 4000-2768 4000-2769 4000-2771 4000-2773 4000-2774 4000-2775 4000-2778 4000-2779 4000-2780 4000-2781 4000-2782 4000-2783 4000-2784 4000-2785 4000-2786 4000-2787 4000-2788 4000-2790 4000-2792 4000-2793 4000-2794 4000-2795 4000-2799 4000-2800 4000-2801 4000-2802 4000-2844 4000-2851 4000-2852 4000-2853 4000-2859 4000-2867 4000-2868 4000-2870 4000-2871 4000-2872 4000-2876 4000-2879 4000-2882 4000-2884 4000-2889 4000-2890 4000-2891 4000-2893 4000-2895 4000-2896 4000-2970 4000-2985 4000-2994 4000-3005 4000-3012 4000-3013 4000-3018 4000-3023 4000-3054 4000-3063 4000-3070 4000-3099 4000-3103 4000-3110 4000-3112 4000-3113 4000-3115 4000-3117 4000-3118 4000-3119 4000-3120 4000-3124 4000-3125 4000-3127 4000-3129 4000-3132 4000-3133 4000-3134 4000-3139 4000-3140 4000-3141 4000-3143 4000-3145 4000-3146 4000-3148 4000-3150 4000-3151 4000-3155 4000-3157 4000-3159 4000-3160 4000-3162 4000-3165 4000-3169 4000-3175 4000-3176 4000-3178 4000-3179 4000-3181 4000-3248 4000-3258 4000-3264 4000-3271 4000-3273 4000-3277 4000-3281 4000-3286 4000-3289 4000-3293 4000-3296 4000-3300 4000-3302 4000-3303 4000-3307 4000-3308 4000-3309 4000-3310 4000-3313 4000-3314 4000-3315 4000-3318 4000-3321 4000-3325 4000-3326 4000-3327 4000-3330 4000-3331 4000-3332 4000-3336 4000-3345 4000-3427 4000-3441 4000-3444 4000-3445 4000-3446 4000-3447 4000-3448 4000-3449 4000-3450 4000-3451 4000-3455 4000-3456 4000-3457 4000-3458 4000-3460 4000-3464 4000-3466 4000-3507 4000-3510 4000-3524 4000-3526 4000-3530 4000-3531 4000-3532 4000-3533 4000-3534 4000-3535 4000-3536 4000-3540 4000-3541 4000-3543 4000-3544 4000-3550 4000-3552 4000-3554 4000-3563 4000-3583 4000-3587 4000-3594 4000-3602 4000-3604 4000-3605 4000-3607 4000-3608
4000-3609 4000-3611 4000-3612 4000-3613 4000-3616 4000-3619 4000-3624 4000-3625 4000-3634 4000-3636 4000-3637 4000-3639 4000-3640 4000-3641 4000-3646 4000-3658 4000-3660 4000-3661 4000-3664 4000-3667 4000-3670 4000-3674 4000-3681 4000-3684 4000-3685 4000-3688 4000-3689 4000-3690 4000-3693 4000-3701 4000-3703 4000-3704 4000-3705 4000-3706 4000-3709 4000-3711 4000-3712 4000-3713 4000-3719 4000-3720 4000-3726 4000-3728 4000-3732 4000-3744 4000-3748 4000-3751 4000-3752 4000-3756 4000-3760 4000-3762 4000-3763 4000-3764 4000-3765 4000-3766 4000-3767 4000-3768 4000-3770 4000-3771 4000-3780 4000-3798 4000-3802 4000-3803 4000-3807 4000-3813 4000-3814 4000-3816 4000-3819 4000-3820 4000-3824 4000-3828 4000-3832 4000-3835 4000-3841 4000-3842 4000-3844 4000-3847 4000-3848 4000-3849 4000-3850 4000-3853 4000-3856 4000-3861 4000-3869 4000-3871 4000-3881 4000-3882 4000-3883 4000-3893 4000-3898 4000-3899 4000-3902 4000-3904 4000-3906 4000-3908 4000-3910 4000-3913 4000-3914 4000-4028 4000-4033 4000-4036 4000-4041 4000-4043 4000-4044 4000-4045 4000-4049 4000-4050 4000-4059 4000-4060 4000-4065 4000-4120 4000-4122 4000-4128 4000-4132 4000-4136 4000-4137 4000-4186 4000-4188 4000-4208 4000-4209 4000-4210 4000-4211 4000-4213 4000-4216 4000-4217 4000-4218 4000-4223 4000-4225 4000-4240 4000-4242 4000-4244 4000-4278 4000-4279 4000-4289 4000-4302 4000-4305 4000-4306 4000-4308 4000-4309 4000-4310 4000-4313 4000-4314 4000-4318 4000-4319 4000-4320 4000-4325 4000-4327 4000-4338 4000-4341 4000-4344 4000-4349 4000-4359 4000-4362 4000-4368 4000-4396 4000-4399 4000-4400 4000-4409 4000-4410 4000-4411 4000-4412 4000-4413 4000-4414 4000-4415 4000-4416 4000-4420 4000-4442 4000-4453 4000-4454 4000-4455 4000-4466 4000-4485 4000-4487 4000-4490 4000-4491 4000-4492 4000-4493 4000-4494 4000-4495 4000-4496 4000-4497 4000-4498 4000-4513 4000-4519 4000-4520 4000-4528 4000-4533 4000-4534 4000-4537 4000-4540 4000-4541 4000-4542 4000-4545 4000-4546 4000-4548 4000-4550 4000-4551 4000-4555 4000-4556 4000-4557 4000-4560 4000-4571 4000-4575 4000-4576 4000-4577 4000-4589 4000-4594 4000-4595 4000-4597 4000-4599 4000-4603 4000-4608 4000-4609 4000-4611 4000-4620 4000-4626 4000-4628 4000-4631 4000-4632 4000-4635 4000-4636 4000-4637 4000-4644 4000-4645 4000-4646 4000-4647 4000-4649 4000-4651 4000-4657 4000-4659 4000-4660 4000-4661 4000-4663 4000-4664 4000-4665 4000-4666 4000-4667 4000-4668 4000-4669 4000-4670 4000-4674 4000-4679 4000-4681 4000-4684 4000-4685 4000-4686 4000-4687 4000-4689 4000-4694 4000-4696 4000-4707 4000-4716 4000-4725 4000-4728 4000-4730 4000-4732 4000-4733 4000-4736 4000-4737 4000-4738 4000-4739 4000-4740 4000-4741 4000-4742 4000-4743 4000-4750 4000-4753 4000-4756 4000-4758 4000-4792 4000-4808 4000-4811 4000-4817 4000-4820 4000-4834 4000-4843 4000-4850 4000-4852 4000-4855 4000-4865 4000-4867 4000-4870 4000-4871 4000-4872 4000-4877 4000-4880 4000-4881 4000-4882 4000-4883 4000-4884 4000-4885 4000-4888 4000-4890 4000-4891 4000-4892 4000-4893 4000-4895 4000-4896 4000-4897 4000-4898 4000-4899 4000-4905 4000-4906 4000-4907 4000-4910 4000-4913 4000-4927 4000-4990 4000-4991 4000-4993 4000-4995 4000-4996 4000-4997 4000-4999 4000-5002 4000-5003 4000-5040 4000-5044 4000-5051 4000-5059 4000-5068 4000-5070 4000-5074 4000-5077 4000-5080 4000-5082 4000-5084 4000-5085 4000-5086 4000-5089 4000-5091 4000-5092 4000-5093 4000-5094 4000-5096 4000-5097 4000-5101 4000-5104 4000-5105 4000-5112 4000-5118 4000-5119 4000-5123 4000-5135 4000-5140 4000-5145 4000-5274 4000-5275 4000-5282 4000-5284 4000-5291 4000-5296 4000-5298 4000-5301 4000-5302 4000-5306 4000-5313 4000-5314 4000-5317 4000-5319 4000-5323 4000-5335 4000-5336 4000-5350 4000-5354 4000-5359 4000-5360 4000-5363 4000-5365 4000-5371 4000-5375 4000-5377 4000-5379 4000-5381 4000-5382 4000-5384 4000-5389 4000-5399 4000-5401 4000-5403 4000-5404 4000-5407 4000-5408 4000-5410 4000-5412 4000-5415 4000-5416 4000-5417 4000-5420 4000-5422 4000-5423 4000-5424 4000-5425 4000-5427 4000-5432 4000-5433 4000-5434 4000-5436 4000-5441 4000-5461 4000-5465 4000-5466 4000-5467 4000-5469 4000-5470 4000-5474 4000-5476 4000-5477 4000-5479 4000-5480 4000-5482 4000-5484 4000-5487 4000-5494 4000-5499 4000-5501 4000-5507 4000-5513 4000-5522 4000-5526 4000-5532 4000-5535 4000-5537 4000-5538 4000-5539 4000-5541 4000-5543 4000-5549 4000-5553 4000-5555 4000-5556 4000-5559 4000-5563 4000-5574 4000-5596 4000-5597 4000-5603 4000-5605 4000-5611 4000-5613 4000-5615 4000-5619 4000-5622 4000-5623 4000-5627 4000-5628 4000-5634 4000-5635 4000-5636 4000-5637 4000-5638 4000-5644 4000-5645 4000-5651 4000-5655 4000-5657 4000-5658 4000-5663 4000-5670 4000-5673 4000-5678 4000-5682 4000-5683 4000-5684 4000-5686 4000-5687 4000-5688 4000-5696 4000-5698 4000-5699 4000-5709 4000-5734 4000-5735 4000-5754 4000-5755 4000-5757 4000-5759 4000-5761 4000-5762 4000-5763 4000-5766 4000-5773 4000-5776 4000-5777 4000-5778 4000-5780 4000-5781 4000-5785 4000-5805 4000-5826 4000-5841 4000-5860 4000-5864 4000-5873 4000-5876 4000-5878 4000-5889 4000-5903 4000-5914 4000-5976 4000-6013 4000-6062 4000-6063 4000-6075 4000-6079 4000-6085 4000-6087 4000-6088 4000-6089 4000-6090 4000-6091 4000-6095 4000-6098 4000-6099 4000-6101 4000-6121 4000-6130 4000-6132 4000-6135 4000-6136 4000-6137 4000-6138 4000-6139 4000-6141 4000-6142 4000-6143 4000-6144 4000-6145 4000-6148 4000-6149 4000-6158 4000-6159 4000-6160 4000-6161 4000-6164 4000-6172 4000-6175 4000-6176 4000-6178 4000-6181 4000-6190 4000-6191 4000-6194 4000-6196 4000-6197 4000-6201 4000-6211 4000-6217 4000-6219 4000-6220 4000-6223 4000-6224 4000-6227 4000-6230 4000-6232 4000-6233 4000-6235 4000-6237 4000-6238 4000-6241 4000-6244 4000-6245 4000-6248 4000-6253 4000-6273 4000-6275 4000-6345 4000-6469 4000-6476 4000-6481 4000-6482 4000-6485 4000-6494 4000-6522 4000-6535 4000-6539 4000-6562 4000-6570 4000-6571 4000-6578 4000-6588 4000-6589 4000-6660 4000-6677 4000-6703 4000-6709 4000-6710 4000-6712 4000-6713 4000-6714 4000-6717 4000-6718 4000-6719 4000-6720 4000-6722 4000-6723 4000-6724 4000-6725 4000-6726 4000-6727 4000-6728 4000-6729 4000-6730 4000-6731 4000-6732 4000-6733 4000-6735 4000-6752 4000-6781 4000-6784 4000-6794 4000-6796 4000-6801 4000-6804 4000-6807 4000-6899 4000-6912 4000-6925 4000-6926 4000-6928 4000-6931 4000-6934 4000-6936 4000-6941 4000-6942 4000-6944 4000-6946 4000-6949 4000-6950 4000-6954 4000-6958 4000-6960 4000-6962 4000-6965 4000-6966 4000-6968 4000-6969 4000-6970 4000-6971 4000-6972 4000-6973 4000-6976 4000-6980 4000-6981 4000-7049 4000-7062 4000-7088 4000-7090 4000-7091 4000-7092 4000-7095 4000-7099 4000-7103 4000-7104 4000-7106 4000-7108 4000-7109 4000-7112 4000-7119 4000-7120 4000-7138 4000-7180 4000-7192 4000-7196 4000-7218 4000-7220 4000-7222 4000-7234 4000-7238 4000-7255 4000-7292 4000-7303 4000-7304 4000-7305 4000-7308 4000-7309 4000-7311 4000-7313 4000-7316 4000-7321 4000-7323 4000-7332 4000-7335 4000-7336 4000-7337 4000-7338 4000-7339 4000-7344 4000-7345 4000-7353 4000-7354 4000-7360 4000-7363 4000-7365 4000-7371 4000-7373 4000-7374 4000-7378 4000-7379 4000-4504 4000-2231 4000-3851 4000-1012 4000-1670 4000-1689 4000-1773 4000-1840 4000-2012 4000-2017 4000-2026 4000-2731 4000-2861 4000-4066 4000-4799 4000-4800 4000-6221 4000-6841 4000-6927 4000-6957 4000-7223 4000-7224 4000-7242 4000-7310 4000-7381 4000-1074 4000-1086 4000-1128 4000-1183 4000-1346 4000-1655 4000-1683 4000-1687 4000-1715 4000-1720 4000-1726 4000-2001 4000-2107 4000-2293 4000-2294 4000-2296 4000-2306 4000-2412 4000-2546 4000-2558 4000-2567 4000-2702 4000-2706 4000-2761 4000-2806 4000-2824 4000-2835 4000-2905 4000-3228 4000-3421 4000-3465 4000-3500 4000-3758 4000-3772 4000-3862 4000-3962 4000-4019 4000-4063 4000-4177 4000-4324 4000-4328 4000-4350 4000-4397 4000-4484 4000-4512 4000-4606 4000-4633 4000-4690 4000-4776 4000-4778 4000-4791 4000-4829 4000-4836 4000-4985 4000-5047 4000-5048 4000-5228 4000-5265 4000-5330 4000-5435 4000-5551 4000-5557 4000-5560 4000-5592 4000-5631 4000-5681 4000-5829 4000-6015 4000-6110 4000-6203 4000-6287 4000-6407 4000-6505 4000-6549 4000-6592 4000-6630 4000-6734 4000-6741 4000-7168 4000-7204 4000-7211 4000-7225 4000-7320 4000-7355 4000-7370 4000-1650 4000-6379 4000-1064 4000-1065 4000-1066 4000-1075 4000-1085 4000-1088 4000-1090 4000-1094 4000-1099 4000-1104 4000-1132 4000-1136 4000-1138 4000-1144 4000-1178 4000-1198 4000-1199 4000-1200 4000-1201 4000-1202 4000-1203 4000-1204 4000-1205 4000-1209 4000-1210 4000-1217 4000-1219 4000-1221 4000-1222 4000-1223 4000-1224 4000-1228 4000-1229 4000-1238 4000-1239 4000-1250 4000-1262 4000-1264 4000-1266 4000-1268 4000-1270 4000-1324 4000-1328 4000-1331 4000-1337 4000-1339 4000-1342 4000-1345 4000-1347 4000-1354 4000-1363 4000-1369 4000-1371 4000-1377 4000-1379 4000-1383 4000-1384 4000-1385 4000-1402 4000-1404 4000-1405 4000-1406 4000-1407 4000-1419 4000-1422 4000-1425 4000-1428 4000-1429 4000-1445 4000-1452 4000-1455 4000-1456 4000-1457 4000-1462 4000-1463 4000-1466 4000-1467 4000-1476 4000-1483 4000-1486 4000-1489 4000-1497 4000-1500 4000-1504 4000-1508 4000-1512 4000-1514 4000-1515 4000-1517 4000-1520 4000-1533 4000-1537 4000-1538 4000-1541 4000-1542 4000-1546 4000-1574 4000-1586 4000-1588 4000-1591 4000-1593 4000-1595 4000-1596 4000-1597 4000-1602 4000-1604 4000-1614 4000-1620 4000-1628 4000-1631 4000-1634 4000-1637 4000-1639 4000-1644 4000-1645 4000-1660 4000-1662 4000-1663 4000-1664 4000-1665 4000-1669 4000-1671 4000-1672 4000-1676 4000-1678 4000-1684 4000-1792 4000-1797 4000-1798 4000-1802 4000-1803 4000-1806 4000-1808 4000-1809 4000-1812 4000-1814 4000-1817 4000-1818 4000-1819 4000-1820 4000-1821 4000-1823 4000-1824 4000-1825 4000-1830 4000-2002 4000-2003 4000-2004 4000-2005 4000-2008 4000-2009 4000-2011 4000-2014 4000-2015 4000-2016 4000-2019 4000-2020 4000-2021 4000-2022 4000-2023 4000-2024 4000-2025 4000-2028 4000-2035 4000-2036 4000-2037 4000-2041 4000-2042 4000-2044 4000-2046 4000-2049 4000-2050 4000-2053 4000-2060 4000-2066 4000-2078 4000-2096 4000-2099 4000-2100 4000-2120 4000-2124 4000-2126 4000-2127 4000-2129 4000-2134 4000-2135 4000-2136 4000-2137 4000-2138 4000-2174 4000-2176 4000-2179 4000-2181 4000-2183 4000-2184 4000-2188 4000-2191 4000-2192 4000-2193 4000-2199 4000-2201 4000-2203 4000-2205 4000-2207 4000-2208 4000-2209 4000-2211 4000-2215 4000-2249 4000-2254 4000-2256 4000-2264 4000-2267 4000-2270 4000-2273 4000-2277 4000-2280 4000-2281 4000-2285 4000-2286 4000-2287 4000-2288 4000-2289 4000-2290 4000-2298 4000-2302 4000-2303 4000-2309 4000-2311 4000-2323 4000-2381 4000-2440 4000-2445 4000-2464 4000-2472 4000-2473 4000-2476 4000-2478 4000-2481 4000-2482 4000-2483 4000-2485 4000-2490 4000-2491 4000-2492 4000-2495 4000-2497 4000-2499 4000-2500 4000-2501 4000-2505 4000-2527 4000-2539 4000-2540 4000-2542 4000-2545 4000-2548 4000-2549 4000-2552 4000-2635 4000-2648 4000-2655 4000-2656 4000-2657 4000-2658 4000-2659 4000-2663 4000-2664 4000-2665 4000-2669 4000-2759 4000-2829 4000-2830 4000-2831 4000-2833 4000-2834 4000-2837 4000-2838 4000-2839 4000-2840 4000-2842 4000-2845 4000-2846 4000-2856 4000-2857 4000-2858 4000-2864 4000-3034 4000-3042 4000-3046 4000-3056 4000-3066 4000-3073 4000-3074 4000-3080 4000-3081 4000-3082 4000-3083 4000-3084 4000-3085 4000-3087 4000-3088 4000-3089 4000-3090 4000-3092 4000-3093 4000-3094 4000-3095 4000-3096 4000-3097 4000-3098 4000-3100 4000-3101 4000-3102 4000-3104 4000-3108 4000-3147 4000-3149 4000-3152 4000-3153 4000-3161 4000-3232 4000-3234 4000-3235 4000-3240 4000-3241 4000-3242 4000-3243 4000-3244 4000-3249 4000-3250 4000-3251 4000-3252 4000-3254 4000-3256 4000-3260 4000-3261 4000-3265 4000-3266 4000-3267 4000-3270 4000-3342 4000-3349 4000-3351 4000-3352 4000-3357 4000-3358 4000-3361 4000-3364 4000-3369 4000-3370 4000-3372 4000-3373 4000-3377 4000-3378 4000-3379 4000-3387 4000-3391 4000-3393 4000-3402 4000-3406 4000-3407 4000-3472 4000-3592 4000-3614 4000-3622 4000-3623 4000-3626 4000-3627 4000-3629 4000-3630 4000-3631 4000-3632 4000-3633 4000-3635 4000-3642 4000-3648 4000-3649 4000-3750 4000-3753 4000-3754 4000-3775 4000-3776 4000-3785 4000-3789 4000-3839 4000-3922 4000-3932 4000-3991 4000-3992 4000-4000 4000-4005 4000-4006 4000-4011 4000-4012 4000-4013 4000-4014 4000-4016 4000-4017 4000-4021 4000-4022 4000-4026 4000-4027 4000-4031 4000-4032 4000-4035 4000-4037 4000-4038 4000-4040 4000-4042 4000-4046 4000-4061 4000-4062 4000-4143 4000-4183 4000-4193 4000-4194 4000-4195 4000-4196 4000-4197 4000-4198 4000-4200 4000-4201 4000-4202 4000-4203 4000-4204 4000-4205 4000-4206 4000-4207 4000-4228 4000-4229 4000-4311 4000-4323 4000-4332 4000-4334 4000-4336 4000-4340 4000-4357 4000-4358 4000-4366 4000-4367 4000-4375 4000-4377 4000-4378 4000-4379 4000-4380 4000-4382 4000-4389 4000-4392 4000-4393 4000-4395 4000-4406 4000-4483 4000-4488 4000-4500 4000-4501 4000-4502 4000-4505 4000-4508 4000-4515 4000-4516 4000-4518 4000-4562 4000-4568 4000-4585 4000-4586 4000-4590 4000-4815 4000-4900 4000-4901 4000-4902 4000-4903 4000-4904 4000-4908 4000-4909 4000-4911 4000-4914 4000-4917 4000-4918 4000-4928 4000-4929 4000-4930 4000-4931 4000-4939 4000-4940 4000-4941 4000-4942 4000-4948 4000-4949 4000-4956 4000-4960 4000-4963 4000-5022 4000-5083 4000-5100 4000-5108 4000-5110 4000-5111 4000-5114 4000-5116 4000-5121 4000-5122 4000-5125 4000-5126 4000-5134 4000-5136 4000-5138 4000-5139 4000-5154 4000-5155 4000-5156 4000-5160 4000-5272 4000-5273 4000-5278 4000-5283 4000-5285 4000-5287 4000-5288 4000-5293 4000-5303 4000-5312 4000-5481 4000-5483 4000-5485 4000-5486 4000-5488 4000-5489 4000-5493 4000-5495 4000-5496 4000-5498 4000-5500 4000-5505 4000-5511 4000-5515 4000-5517 4000-5521 4000-5523 4000-5524 4000-5525 4000-5527 4000-5528 4000-5529 4000-5530 4000-5531 4000-5534 4000-5540 4000-5546 4000-5547 4000-5552 4000-5554 4000-5706 4000-5737 4000-5743 4000-5765 4000-5767 4000-5770 4000-5771 4000-5782 4000-5787 4000-5788 4000-5789 4000-5796 4000-5798 4000-5800 4000-5802 4000-5806 4000-5815 4000-5820 4000-5839 4000-5854 4000-5862 4000-5874 4000-5892 4000-5915 4000-5932 4000-5934 4000-5940 4000-5974 4000-5993 4000-5997 4000-6004 4000-6006 4000-6105 4000-6111 4000-6112 4000-6113 4000-6114 4000-6117 4000-6118 4000-6120 4000-6123 4000-6124 4000-6125 4000-6126 4000-6127 4000-6129 4000-6140 4000-6320 4000-6408 4000-6418 4000-6422 4000-6426 4000-6437 4000-6438 4000-6439 4000-6442 4000-6461 4000-6472 4000-6474 4000-6484 4000-6490 4000-6497 4000-6498 4000-6502 4000-6506 4000-6508 4000-6529 4000-6530 4000-6538 4000-6546 4000-6552 4000-6554 4000-6566 4000-6607 4000-6669 4000-6671 4000-6707 4000-6806 4000-6821 4000-6823 4000-6826 4000-6827 4000-6828 4000-6829 4000-6831 4000-6832 4000-6833 4000-6834 4000-6836 4000-6837 4000-6838 4000-6840 4000-6845 4000-6850 4000-6851 4000-6853 4000-6855 4000-6857 4000-6858 4000-6859 4000-6860 4000-6861 4000-6863 4000-6866 4000-6872 4000-6873 4000-7147 4000-7216 4000-7256 4000-7259 4000-7261 4000-7262 4000-7264 4000-7267 4000-7271 4000-7276 4000-7277 4000-7278 4000-7281 4000-7284 4000-7285 4000-7290 4000-2371 4000-1378 4000-2563 4000-2577 4000-2590 4000-3035 4000-3206 4000-4566 4000-4662 4000-4920 4000-6081 4000-6180 4000-6456 4000-6870 4000-2214 4000-2435 4000-2444 4000-2725 4000-2742 4000-2749 4000-2805 4000-2810 4000-2873 4000-2886 4000-3368 4000-3846 4000-4074 4000-4160 4000-4255 4000-4312 4000-4404 4000-4422 4000-4506 4000-4587 4000-4944 4000-4951 4000-4953 4000-4954 4000-4965 4000-5292 4000-5409 4000-5418 4000-5440 4000-5630 4000-5725 4000-5869 4000-5904 4000-6017 4000-6368 4000-6491 4000-6500 4000-6510 4000-6585 4000-6772 4000-7074 4000-7186 4000-7210 4000-7358 4000-2070 4000-7129 4000-1374 4000-1771 4000-1784 4000-2556 4000-2560 4000-2565 4000-2695 4000-2730 4000-2755 4000-3320 4000-4391 4000-4436 4000-4443 4000-4448 4000-4986 4000-5073 4000-5286 4000-5398 4000-5450 4000-5598 4000-6290 4000-6398 4000-6427 4000-6457 4000-6815 4000-6816 4000-1673 4000-2087 4000-3333 4000-3384 4000-3388 4000-5058 4000-5431 4000-6353 4000-6464 4000-1009 4000-1061 4000-1146 4000-1387 4000-1582 4000-2131 4000-2140 4000-2152 4000-2156 4000-2161 4000-2169 4000-2355 4000-2364 4000-2848 4000-2862 4000-3039 4000-3040 4000-3222 4000-3283 4000-3311 4000-3470 4000-4144 4000-4438 4000-4517 4000-4612 4000-4650 4000-4887 4000-5065 4000-5143 4000-5320 4000-5380 4000-5386 4000-5394 4000-5455 4000-5609 4000-5795 4000-5996 4000-6103 4000-6167 4000-6195 4000-6214 4000-6289 4000-6509 4000-6757 4000-6814 4000-6824 4000-7071 4000-7110 4000-7118 4000-7121 4000-7125 4000-7144 4000-1098 4000-1109 4000-1269 4000-1431 4000-1719 4000-2422 4000-2424 4000-2427 4000-2428 4000-2432 4000-2741 4000-2866 4000-2883 4000-2898 4000-3247 4000-3400 4000-3424 4000-3508 4000-3515 4000-3663 4000-3783 4000-3868 4000-3874 4000-3886 4000-4086 4000-4322 4000-4329 4000-4419 4000-4511 4000-4602 4000-4618 4000-4671 4000-4781 4000-4825 4000-4830 4000-4989 4000-5010 4000-5095 4000-5322 4000-5387 4000-5397 4000-5411 4000-5426 4000-5478 4000-5585 4000-5608 4000-5668 4000-5827 4000-6278 4000-6364 4000-6495 4000-6817 4000-6998 4000-6999 4000-7009 4000-7045 4000-7089 4000-7128 4000-7195 4000-7253 4000-7349 4000-1186 4000-3781 4000-4676 4000-4677 4000-4702 4000-4959 4000-5445 4000-5618 4000-6375 4000-1656 4000-2151 4000-2235 4000-2236 4000-2237 4000-2689 4000-3041 4000-3738 4000-4460 4000-4552 4000-4578 4000-4717 4000-5056 4000-5421 4000-6200 4000-6531 4000-7241 4000-7247 4000-7343 4000-7348 4000-7361 4000-2438 4000-2536 4000-2557 4000-2767 4000-2816 4000-4095 4000-4098 4000-4307 4000-4316 4000-4775 4000-5248 4000-5249 4000-5252 4000-5258 4000-6344 4000-7023 4000-7040 |
Recalling Firm/ Manufacturer |
Carl Zeiss Meditec Inc 5160 Hacienda Dr Dublin CA 94568
|
For Additional Information Contact | Judith A. Brimcombe. MA 925-557-4616 |
Manufacturer Reason for Recall | A software defect exists in versions 5.0.0 and 5.1.0 where when following a specific workflow, exam data may be saved under the name of the first patient |
FDA Determined Cause 2 | Software design |
Action | Carl Zeiss Meditec, Inc. sent an Field Correction Action letter dated March 9, 2011, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. The letters requested acknowledgement of the letter and a response and the fact that a follow up packet with revised software will be sent to all customers with V 5.0.0 or 5.1.0 software.
For questions, please call 1-855-853-9461. |
Quantity in Commerce | 3465 US; 3423 OUS |
Distribution | Worldwide Distribution - USA including AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, LA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SC, TN, TX, UT, VA, VT, WA, WI, and WV, and the countries of Algeria, Argentina, Australia, Bangladesh, Belgium, Boliva, Brazil, Canada, France, Chili, China, Columbia, Denmark, Ecuador,Germany, Great Britain, Greece, Guatemala, Hong Kong, India, Japan, Lebanon, Malaysia, Mexico, New Zealand, Pakistan, Peru, Singapore,South Korea, Spain, Sri Lanka, Taiwan, Uruguay and Venezuela |
Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database | 510(K)s with Product Code = OBO
|
|
|
|