Class 2 Device Recall GE Healthcare, Dash 3000/4000/5000.

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
  Class 2 Device Recall GE Healthcare, Dash 3000/4000/5000. see related information
Date Initiated by Firm June 26, 2012
Date Posting Updated December 16, 2012
Recall Status1 Terminated 3 on June 23, 2017
Recall Number Z-0494-2013
Recall Event ID 63650
510(K)Number K073462  
Product Classification Monitor, physiological, patient(with arrhythmia detection or alarms) - Product Code MHX
Product GE Healthcare, Dash 3000/4000/5000.

The Dash 3000/4000/5000 patient monitor is intended for use under the direct supervision of a licensed healthcare practitioner. The intended use of the system is to monitor physiologic parameter data on adult, pediatric and neonatal patients. The Dash 3000/4000/5000 patient monitor is designed as a bedside, portable, and intra-hospital transport monitor that can operate in all professional medical facilities including but not limited to: emergency department, operating room, post anesthesia recovery, critical care, surgical intensive care, respiratory intensive care, coronary care, medical intensive care, pediatric intensive care, or neonatal intensive care areas located in hospitals, outpatient clinics, free-standing surgical centers, and other alternate care facilities. Physiologic data includes, but is not limited to: electrocardiogram, invasive blood pressure, noninvasive blood pressure (NBP), heart rate, temperature, cardiac output, respiration, pulse oximetry, carbon dioxide, bi-spectral index, impedance cardiography, oxygen, and anesthetic agents as summarized in the operator's manual. The Dash 3000/4000/5000 patient monitor is also intended to provide physiologic data over the UNITY NETWORKTM indirectly to clinical information systems (via our Enterprise Gateway) and allow the user to access hospital data at the point-of-care. The information can be displayed, trended, stored, and printed. The Dash 3000/4000/5000 patient monitor was developed to interface with nonproprietary third party peripheral devices that support serial data outputs.
Code Information 58SA, SHQ12217859SA, SHQ12217860SA, SHQ12217861SA, SHQ12217863SA, SHQ12217864SA, SHQ12217867SA, SHQ12217868SA, SHQ12217869SA, SHQ12217884SA, SHQ12217913SA, SHQ12218018SA, SHQ12218019SA, SHQ12218020SA, SHQ12218021SA, SHQ12218058SA, SHQ12218059SA, SHQ12218060SA, SHQ12218062SA, SHQ12218063SA, SHQ12218064SA, SHQ12218065SA, SHQ12218066SA, SHQ12218067SA, SHQ12218071SA, SHQ12218072SA, SHQ12218073SA, SHQ12218075SA, SHQ12218077SA, SHQ12218080SA, SHQ12218082SA, SHQ12218084SA, SHQ12218086SA, SHQ12218087SA, SHQ12218089SA, SHQ12218090SA, SHQ12218094SA, SHQ12218095SA, SHQ12218096SA, SHQ12218097SA, SHQ12228118SA, SHQ12228119SA, SHQ12228120SA, SHQ12228121SA, SHQ12228122SA, SHQ12228123SA, SHQ12228124SA, SHQ12228125SA, SHQ12228126SA, SHQ12228127SA, SHQ12228173SA, SHQ12228174SA, SHQ12228175SA, SHQ12228253SA, SHQ12228255SA, SHQ12228259SA, SHQ12228264SA, SHQ12228265SA, SHQ12238370SA, SHQ12238371SA, SHQ12238372SA, SHQ12238373SA, SHQ12238374SA, SHQ12238375SA, SHQ12238376SA, SHQ12238377SA, SHQ12238378SA, SHQ12258750SA, SHQ12278970SA, SHQ12278971SA, SHQ12278972SA, SHQ12278973SA, SHQ12278974SA, SHQ12278975SA, SHQ12278976SA, SHQ12278977SA, SHQ12278978SA, SHQ12278979SA, SHQ12278981SA, SHQ12278984SA, SHQ12278988SA, SHQ12278990SA, SHQ12278992SA, SHQ12278994SA, SHQ12278995SA, SHQ12278996SA, SHQ12278997SA, SHQ12278998SA, SHQ12278999SA, SHQ12279000SA, SHQ12289055SA, SHQ12289057SA, SHQ12289058SA, SHQ12289059SA, SHQ12289086SA, SHQ12289087SA, SHQ12289090SA, SHQ12289091SA, SHQ12289092SA, SHQ12289097SA, SHQ12289098SA, SHQ12289100SA, SHQ12289104SA, SHQ12289105SA, SHQ12289106SA, SHQ12289108SA, SHQ12289109SA, SHQ12289138SA, SHQ12289141SA, SHQ12289142SA, SHQ12289143SA, SHQ12289145SA, SHQ12289148SA, SHQ12289149SA, SHQ12289150SA, SHQ12289235SA, SHQ12289236SA, SHQ12289237SA, SHQ12289238SA, SHQ12289239SA, SHQ12289241SA, SHQ12289242SA, SHQ12289246SA, SHQ12289249SA, SHQ12299406SA, SHQ12299407SA, SHQ12299408SA, SHQ12299410SA, SHQ12299411SA, SHQ12299412SA, SHQ12299413SA, SHQ12299414SA, SHQ12299415SA, SHQ12299416SA, SHQ12299417SA, SHQ12299418SA, SHQ12299419SA, SHQ12299421SA, SHQ12299423SA, SHQ12299426SA, SHQ12299429SA, SHQ12299431SA, SHQ12309600SA, SHQ12309601SA, SHQ12309602SA, SHQ12309603SA, SHQ12309605SA, SHQ12309606SA, SHQ12309607SA, SHQ12309608SA, SHQ12309609SA, SHQ12309610SA, SHQ12309611SA, SHQ12309612SA, SHQ12309613SA, SHQ12309614SA, SHQ12309615SA, SHQ12309616SA, SHQ12309617SA, SHQ12309619SA, SHQ12309621SA, SHQ12309623SA, SHQ12309624SA, SHQ12309625SA, SHQ12309626SA, SHQ12309627SA, SHQ12309628SA, SHQ12309629SA, SHQ12309630SA, SHQ12309631SA, SHQ12309632SA, SHQ12309633SA, SHQ12309634SA, SHQ12309635SA, SHQ12309636SA, SHQ12309637SA, SHQ12309638SA, SHQ12309639SA, SHQ12309640SA, SHQ12309641SA, SHQ12309642SA, SHQ12309643SA, SHQ12309644SA, SHQ12309645SA, SHQ12309646SA, SHQ12309647SA, SHQ12309648SA, SHQ12309649SA, SHQ12309650SA, SHQ12309651SA, SHQ12309652SA, SHQ12309656SA, SHQ12309657SA, SHQ12309658SA, SHQ12309659SA, SHQ12309660SA, SHQ12309661SA, SHQ12309662SA, SHQ12309663SA, SHQ12309664SA, SHQ12309665SA, SHQ12309666SA, SHQ12309667SA, SHQ12309668SA, SHQ12309670SA, SHQ12309673SA, SHQ12309674SA, SHQ12309675SA, SHQ12309677SA, SHQ12309678SA, SHQ12309679SA, SHQ12309680SA, SHQ12309682SA, SHQ12309684SA, SHQ12309685SA, SHQ12309687SA, SHQ12309690SA, SHQ12309692SA, SHQ12309693SA, SHQ12309694SA, SHQ12309695SA, SHQ12309696SA, SHQ12309697SA, SHQ12309699SA, SHQ12309700SA, SHQ12309701SA, SHQ12309702SA, SHQ12309703SA, SHQ12309704SA, SHQ12309705SA, SHQ12309706SA, SHQ12309707SA, SHQ12309708SA, SHQ12309709SA, SHQ12309710SA, SHQ12309711SA, SHQ12309712SA, SHQ12309713SA, SHQ12309714SA, SHQ12309715SA, SHQ12309717SA, SHQ12309718SA, SHQ12309719SA, SHQ12309720SA, SHQ12309722SA, SHQ12309723SA, SHQ12309724SA, SHQ12320257SA, SHQ12320258SA, SHQ12320259SA, SHQ12320260SA, SHQ12320261SA, SHQ12320262SA, SHQ12320264SA, SHQ12320265SA, SHQ12320266SA, SHQ12320267SA, SHQ12320268SA, SHQ12320270SA, SHQ12320272SA, SHQ12320275SA, SHQ12320278SA, SHQ12320279SA, SHQ12320282SA, SHQ12320283SA, SHQ12320292SA, SHQ12320293SA, SHQ1
Recalling Firm/
Manufacturer
GE Healthcare, LLC
3000 N Grandview Blvd
Waukesha WI 53188-1615
For Additional Information Contact
262-513-4122
Manufacturer Reason
for Recall
GE Healthcare has recently become aware of a potential issue due to an unexpected loss of Custom Automatic NBP Measurement cycling associated with the Dash 3000/4000/5000 patient monitors.
FDA Determined
Cause 2
Software design
Action GE Healthcare sent a Important Product Information Letter dated June 21, 2012, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. If your facility is using the Custom Auto NBP Measurement feature in conjunction with Automatic View On Alarm and/or View Other Beds options, GE Healthcare recommends that you stop using the Custom Automatic NBP Measurement feature and use the Auto NBP function instead. All Dash3000/4000/5000 patient monitors with software version 6.0 and later with the networking option enabled may be affected. You can determine the software version of the Dash monitors by selecting MORE MENUS > MONITOR SETUP > SOFTWARE CONFIGURATION. Product Correction GE Healthcare is working on a software correction that will be provided to you at no charge once it is available. We will contact you to schedule the correction. Contact Information If you have any questions regarding this notification, please contact Technical Support at 1-800-558-7044 or your local Service Representative. Please be assured that maintaining a high level of safety and quality is our highest priority. If you have any questions, please contact us immediately per the contact information above. .
Quantity in Commerce 119,260 devices
Distribution Worldwide Distribution -- USA (nationwide) including the countries of AUSTRALIA, ANGOLA,ARGENTINA, AUSTRIA, BAHAMAS, BAHRAIN, BANGLADESH, BELGIUM, BOSNIA HERZEGOVINA, BRAZIL, BRUNEI, BULGARIA, CAMBODIA, CANADA, CHILE, CHINA, COLOMBIA, COSTA RICA, CROATIA, CUBA, CYPRUS, CZECH REPUBLIC, DENMARK, ECUADOR, EGYPT, EL SALVADOR, ESTONIA, FINLAND, FRANCE, GEORGIA, GERMANY, GHANA, GREECE, GUATEMALA, HONDURAS, HONG KONG, HUNG ARIA, ICELAND, INDIA, INDONESIA, IRAQ, IRELAND, ISRAEL, ITALY, JAPAN, JORDAN, KAZAHSTAN, KENYA, KOREA, KUWAIT, LATIVIA, LEBANON, LIBYA, LITHUANIA, MACEDONIA, MALAYSIA, MEXICO, MALTA, MOROCCO, NEPAL, NETHERLAND, NEW ZEALAND, NIGERIA, NORWAY, PAKISTAN, PALESTINIAN, PANAMA, PARAGUAY, PERU, PHILIPPINES, POLAND, PORTUGAL, QATAR, ROMANIA, RUSSIA, SAUDI ARABIA, SINGAPORE, SLOVENIA, SOUTH AFRICA, SPAIN, SRI LANKA, SWEDEN, SWITZERLAND, SYRIA, TAIWAN THAILAND, TRINIDAD AND TOBAGO,K TUNISIA, TURKEY, UGANDA, UKRAINE, UNITED ARB EMIRATES , UNITED KINGDOM and URUGUAY
Total Product Life Cycle TPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
510(K) Database 510(K)s with Product Code = MHX and Original Applicant = GE MEDICAL SYSTEMS INFORMATION TECHNOLOGIES
-
-