• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 1 Device Recall EZ Breathe Medication Cup

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
  Class 1 Device Recall EZ Breathe Medication Cup see related information
Date Initiated by Firm April 30, 2013
Date Posted May 29, 2013
Recall Status1 Terminated 3 on June 28, 2016
Recall Number Z-1373-2013
Recall Event ID 65070
Product Classification Nebulizer, medicinal, non-ventilatory (atomizer) - Product Code CCQ
Product EZ Breathe Medication Cup.

For the delivery of liquid medications for respiratory usage.
Code Information Lots #120902, Lot # 120903 and Lot # 120904
Recalling Firm/
Manufacturer
Nephron Pharmaceuticals Corp.
4121 SW 34th St
Orlando FL 32811-6475
For Additional Information Contact Jerry Webb
800-443-4313 Ext. 2292
Manufacturer Reason
for Recall
Possible dislodgement of the "Plate A", if this instance occurs, it presents a potential choking hazard.
FDA Determined
Cause 2
Nonconforming Material/Component
Action The firm, Nephron Pharmaceuticals Corporation, notified its consignees on April 30, 2013, and May 3, 2013, of the problem via a recall notice and a Press Release, respectively. The firm directed consignees to discontinue use of any EZ Breathe atomizer units contained in the Asthmanefrin Starter Kits with the affected lot numbers and also those that were sold individually. Customers were directed to return all affected product.
Quantity in Commerce 750 Medications Cups
Distribution Worldwide Distribution-USA (nationwide) including DC and Puerto Rico and the states of AL, AZ, CA, CO, CT, FL, IL, IN, IO, GA, KS, KY, LA, MD, MN, MS, MO, NK, NV, NJ, NM, NY, OH, OK, OR, PA, RI, SC, TN , TX, VT, VA, WA, and WV, and the countries of Guatemala, Poland, Panama, and United Kingdom.
Total Product Life Cycle TPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
-
-