|
Class 2 Device Recall Synchron Systems LX20/ UniCel DxC ISE Reference Reagent |
|
Date Initiated by Firm |
September 22, 2010 |
Date Posted |
August 10, 2013 |
Recall Status1 |
Terminated 3 on August 10, 2013 |
Recall Number |
Z-1951-2013 |
Recall Event ID |
65655 |
510(K)Number |
K965240
|
Product Classification |
Electrode, ion specific, potassium - Product Code CEM
|
Product |
ISE Electrolyte Reference reagent (for Synchron LX Systems and UniCel DxC 600/800 Systems), Part No. A28937, A28945, 467935, 467915.
Product Usage: ISE Electrolyte Buffer reagent and ISE Electrolyte Reference reagent, when used in conjunction with SYNCHRON LX System(s), UniCel DxC 600/800 System(s) and SYNCHRON Systems AQUA CAL 1, 2 and 3, are intended for the quantitative determination of potassium concentration in human serum, plasma or urine. |
Code Information |
Lot No. T006042 |
Recalling Firm/ Manufacturer |
Beckman Coulter Inc. 250 S Kraemer Blvd Brea CA 92821-6232
|
For Additional Information Contact |
Clair K. O'Donovan, Ph.D. 714-961-4483
|
Manufacturer Reason for Recall |
Beckman Coulter is recalling the Synchron Systems LX20/UniCel DxC ISE Reference Reagent because it has a potassium (K) response near the instrument calibration flag limits and may exceed the limits on some instruments.
|
FDA Determined Cause 2 |
Process control |
Action |
Beckman Coulter Inc sent an Urgent Product Corrective Action (PCA) letter dated September 22, 2010 to all affected customers. Customers were notified of the the problems identified and the actions to be completed. Customers are instructed to complete and return the enclosed response form within 10 days. Customers with questions are instructed to contact Customer Support Center (Call Center) on the Beckman Coulter website at http://www.beckmancoulter.com/customersupport/support/ or call (800) 854-3633 in the US and in Canada. |
Quantity in Commerce |
718 units total (132 units in US) |
Distribution |
Worldwide Distribution - USA Nationwide and the countries of
Belgium, France, Germany, Netherlands, Poland, Spain, Sweden, Switzerland, and United Kingdom |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database |
510(K)s with Product Code = CEM and Original Applicant = BECKMAN INSTRUMENTS, INC.
|
|
|
|