| Class 1 Device Recall MoniTorr | |
Date Initiated by Firm | April 08, 2019 |
Date Posted | May 20, 2019 |
Recall Status1 |
Terminated 3 on January 27, 2021 |
Recall Number | Z-1310-2019 |
Recall Event ID |
82659 |
510(K)Number | K022554 K062599 K920156 |
Product Classification |
Shunt, central nervous system and components - Product Code JXG
|
Product | MoniTorr 10100 10-100 CSF Drainage System w/Patient Line One Way Valve
MoniTorr 10102 10-102 CSF Drainage System w/Patient Line One Way Valve w
MoniTorr 10110 10-110 CSF Drainage System used with Pole Mount System
MoniTorr 10140 10-140 CSF Drainage System w/Vinyl Measuring Strip
MoniTorr 10150 10-150 CSF Drainage System Simple Bag and Line
MoniTorr 10110A 10-110A CSF Drainage System w/o Manifold Stopcock
MoniTorr SP0017 SP0017 Special EVD 10-110 w/o Y Site Latex Free Sites
MoniTorr SP0042 SP0042 Special EVD 10-100 w/o Y Site
MoniTorr SP0090 SP0090 Special EVD 10-110 w/o Y Site
MoniTorr SP0164 SP0164 Special EVD 10-140 w/Y Site & Stopcock Reverse
MoniTorr SP0236 SP0236 INS 1100 WITH NEEDLE AT Y-ACCESS AND NEEDLELESS SI
Product Usage: The MoniTorr ICP External CSF Drainage and Monitoring Systems provide a simple to use, closed system for the drainage of cerebrospinal fluid (CSF) from ventricles of the brain or the lumbar subarachnoid space to a drainage bag. The system may be used with a pole mounted assembly that allows for simple, quick and accurate alignment with the patient and secure positive or negative pressure level setting. The system also has been designed to provide for ease of patient transport through a compact design and antimicrobial hydrophobic vent feature that resists occlusion.
Indications: The MoniTorr ICP system allows for drainage and monitoring of CSF from the lateral ventricles of the brain and the lumbar subarachnoid space in selected patients to reduce intracranial pressure (ICP), to monitor CSF, to provide temporary drainage of CSF inpatients with infected CSF shunts, and to monitor ICP. |
Code Information |
All lots |
Recalling Firm/ Manufacturer |
Integra LifeSciences Corp. 311 Enterprise Dr Plainsboro NJ 08536-3344
|
For Additional Information Contact | Dr. Patricia Kihn 609-275-0500 |
Manufacturer Reason for Recall | The firm has identified a complaint trend regarding breakage of the LimiTorr Transducer Mount stopcock and MoniTorr Panel Mount stopcock. |
FDA Determined Cause 2 | Device Design |
Action | On April 8, 2019, the firm distributed Urgent Medical Device Correction letters to its affected customers. The communication advised customers of the possible breakage of the stopcock.
Customers were advised to take the following precautions: When connecting a fluid-filled transducer (FFT) to the Panel Mount/Transducer Mount stopcocks:
" Do not apply excessive force while attaching the FFT to the stopcock.
" Take care during the priming process to not torque or bend the syringe and FFT.
" Ensure luer lock is "finger tight;" do not over-tighten connection at the stopcock junction.
When managing an external ventricular or lumbar drain (EVD) with a FFT attached:
" Take care when manipulating the stopcock to zero the FFT and to obtain pressure readings, do
not apply excessive force to the stopcock junction.
" Use caution during patient transportation and during routine ICU care, to avoid direct contact of
the FFT with other devices or equipment.
" If the FFT must be flushed or replaced, do not apply excessive torque or force to the FFT/
stopcock junction.
" If the FFT is connected to the ICU bedside monitor, ensure the interface cable is positioned so
that it does not place excess force on the FFT (e.Q. do not allow cable to touch the floor.
Customers were asked to determine if they had any affected product on hand, review and understand the precautions, and complete and return a completed acknowledgement form to the firm. Distributors were asked to forward the customer letter to any customers who received the affected product.
Should you have any questions regarding these instructions, please contact Customer Service
Monday to Friday 8:00 AM - 8:00 PM EST
USA: 1-800-654-2873: email: custsvcnj@integralife.com
Outside USA: 781-565-1401
***UPDATE***
On October 22, 2019, the firm sent out an updated notification informing customers that a new stopcock has been designed for the LimiTorr and MoniTorr, and is now available for replacement of th |
Quantity in Commerce | 42102 units |
Distribution | Worldwide Distribution - US Nationwide
International distribution to Argentina, American Samoa, Australia, Canada, and Guam. |
Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|