• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 1 Device Recall HeartWare HVAD System Battery Charger AC Adapter

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 1 Device Recall HeartWare HVAD System Battery Charger AC Adaptersee related information
Date Initiated by FirmJanuary 23, 2020
Date PostedMarch 09, 2020
Recall Status1 Open3, Classified
Recall NumberZ-1427-2020
Recall Event ID 84829
PMA NumberP100047 
Product Classification Ventricular (assist) bypass - Product Code DSQ
ProductHeartWare HVAD System Battery Charger The HeartWare HVAD System is indicated for hemodynamic support in patients with advanced, refractory left ventricular heart failure; either as a Bridge to Cardiac Transplantation (BTT), myocardial recovery, or as Destination Therapy (DT) in patients for whom subsequent transplantation is not planned.
Code Information Model Number: 1610, all Serial Numbers.   Known Model Numbers (Serial Numbers) GTIN Numbers: 1610GB (BCH000001) 00888707002219,  1610GB (BCH000002) 00888707002219,  1610DE (BCH000003) 00888707002202,  1610DE (BCH000004) 00888707002202,  1610IT (BCH000005) 00888707002240,  1610IT (BCH000006) 00888707002240,  1610DE (BCH000007) 00888707002202,  1610DE (BCH000008) 00888707002202,  1610CH (BCH000010) 00888707002196,  1610DE (BCH000011) 00888707002202,  1610DE (BCH000012) 00888707002202,  1610CH (BCH000013) 00888707002196,  1610IT (BCH000014) 00888707002240,  1610DE (BCH000015) 00888707002202,  1610DE (BCH000017) 00888707002202,  1610DE (BCH000018) 00888707002202,  1610DE (BCH000019) 00888707002202,  1610IT (BCH000020) 00888707002240,  1610DE (BCH000021) 00888707002202,  1610GB (BCH000022) 00888707002219,  1610GB (BCH000023) 00888707002219,  1610DE (BCH000024) 00888707002202,  1610IT (BCH000025) 00888707002240,  1610IT (BCH000026) 00888707002240,  1610IT (BCH000027) 00888707002240,  1610DE (BCH000028) 00888707002202,  1610CH (BCH000029) 00888707002196,  1610CH (BCH000030) 00888707002196,  1610DE (BCH000031) 00888707002202,  1610DE (BCH000032) 00888707002202,  1610DE (BCH000033) 00888707002202,  1610DE (BCH000034) 00888707002202,  1610GB (BCH000035) 00888707002219,  1610DE (BCH000036) 00888707002202,  1610GB (BCH000037) 00888707002219,  1610CH (BCH000038) 00888707002196,  1610CH (BCH000039) 00888707002196,  1610DE (BCH000040) 00888707002202,  1610GB (BCH000041) 00888707002219,  1610CH (BCH000042) 00888707002196,  1610GB (BCH000043) 00888707002219,  1610GB (BCH000044) 00888707002219,  1610GB (BCH000045) 00888707002219,  1610IT (BCH000047) 00888707002240,  1610DE (BCH000048) 00888707002202,  1610GB (BCH000050) 00888707002219,  1610DE (BCH000051) 00888707002202,  1610DE (BCH000052) 00888707002202,  1610DE (BCH000053) 00888707002202,  1610GB (BCH000054) 00888707002219,  1610GB (BCH000055) 00888707002219,  1610GB (BCH000056) 00888707002219,  1610DE (BCH000057) 00888707002202,  1610IT (BCH000058) 00888707002240,  1610DE (BCH000059) 00888707002202,  1610DE (BCH000061) 00888707002202,  1610CH (BCH000063) 00888707002196,  1610DE (BCH000064) 00888707002202,  1610DE (BCH000065) 00888707002202,  1610GB (BCH000066) 00888707002219,  1610IT (BCH000067) 00888707002240,  1610CH (BCH000068) 00888707002196,  1610AU (BCH000069) 00888707002165,  1610DE (BCH000071) 00888707002202,  1610GB (BCH000072) 00888707002219,  1610DE (BCH000073) 00888707002202,  1610GB (BCH000074) 00888707002219,  1610GB (BCH000076) 00888707002219,  1610DE (BCH000077) 00888707002202,  1610GB (BCH000078) 00888707002219,  1610DE (BCH000079) 00888707002202,  1610GB (BCH000080) 00888707002219,  1610IT (BCH000081) 00888707002240,  1610CH (BCH000082) 00888707002196,  1610DE (BCH000083) 00888707002202,  1610GB (BCH000084) 00888707002219,  1610DE (BCH000085) 00888707002202,  1610DE (BCH000086) 00888707002202,  1610DE (BCH000087) 00888707002202,  1610DE (BCH000088) 00888707002202,  1610DE (BCH000089) 00888707002202,  1610IT (BCH000090) 00888707002981,  1610CH (BCH000091) 00888707002196,  1610JP (BCH000092) 00888707002257,  1610DE (BCH000093) 00888707002202,  1610CH (BCH000094) 00888707002196,  1610DE (BCH000095) 00888707002202,  1610DE (BCH000096) 00888707002202,  1610CH (BCH000097) 00888707002196,  1610GB (BCH000098) 00888707002219,  1610GB (BCH000099) 00888707002219,  1610CH (BCH000100) 00888707002196,  1610DE (BCH000101) 00888707002202,  1610DE (BCH000102) 00888707002202,  1610DE (BCH000103) 00888707002202,  1610DE (BCH000104) 00888707002202,  1610DE (BCH000105) 00888707002202,  1610DE (BCH000106) 00888707002202,  1610DE (BCH000107) 00888707002202,  1610DE (BCH000108) 00888707002202,  1610DE (BCH000109) 00888707002202,  1610DE (BCH000110) 00888707002202,  1610DE (BCH000111) 00888707002202,  1610DE (BCH000112) 00888707002202,  1610DE (BCH000113) 00888707002202,  1610DE (BCH000114) 00888707002202,  1610DE (BCH000115) 00888707002202,  1610DE (BCH000116) 00888707002202,  1610DE (BCH000117) 00888707002202,  1610DE (BCH000118) 00888707002202,  1610DE (BCH000119) 00888707002202,  1610DE (BCH000120) 00888707002202,  1610DE (BCH000121) 00888707002202,  1610DE (BCH000122) 00888707002202,  1610DE (BCH000123) 00888707002202,  1610DE (BCH000124) 00888707002202,  1610DE (BCH000125) 00888707002202,  1610DE (BCH000126) 00888707002202,  1610DE (BCH000127) 00888707002202,  1610DE (BCH000128) 00888707002202,  1610DE (BCH000129) 00888707002202,  1610DE (BCH000130) 00888707002202,  1610DE (BCH000131) 00888707002202,  1610DE (BCH000132) 00888707002202,  1610DE (BCH000134) 00888707002202,  1610DE (BCH000135) 00888707002202,  1610DE (BCH000136) 00888707002202,  1610DE (BCH000138) 00888707002202,  1610DE (BCH000139) 00888707002202,  1610DE (BCH000140) 00888707002202,  1610DE (BCH000141) 00888707002202,  1610DE (BCH000142) 00888707002202,  1610DE (BCH000143) 00888707002202,  1610DE (BCH000144) 00888707002202,  1610DE (BCH000145) 00888707002202,  1610DE (BCH000146) 00888707002202,  1610DE (BCH000147) 00888707002202,  1610DE (BCH000148) 00888707002202,  1610DE (BCH000149) 00888707002202,  1610DE (BCH000150) 00888707002202,  1610DE (BCH000151) 00888707002202,  1610DE (BCH000152) 00888707002202,  1610DE (BCH000153) 00888707002202,  1610DE (BCH000154) 00888707002202,  1610DE (BCH000155) 00888707002202,  1610DE (BCH000156) 00888707002202,  1610DE (BCH000157) 00888707002202,  1610DE (BCH000158) 00888707002202,  1610DE (BCH000159) 00888707002202,  1610DE (BCH000160) 00888707002202,  1610DE (BCH000161) 00888707002202,  1610DE (BCH000162) 00888707002202,  1610DE (BCH000163) 00888707002202,  1610DE (BCH000164) 00888707002202,  1610DE (BCH000165) 00888707002202,  1610DE (BCH000166) 00888707002202,  1610DE (BCH000168) 00888707002202,  1610DE (BCH000169) 00888707002202,  1610DE (BCH000170) 00888707002202,  1610DE (BCH000171) 00888707002202,  1610DE (BCH000172) 00888707002202,  1610DE (BCH000173) 00888707002202,  1610DE (BCH000174) 00888707002202,  1610DE (BCH000175) 00888707002202,  1610DE (BCH000176) 00888707002202,  1610DE (BCH000177) 00888707002202,  1610DE (BCH000178) 00888707002202,  1610DE (BCH000179) 00888707002202,  1610DE (BCH000180) 00888707002202,  1610DE (BCH000181) 00888707002202,  1610DE (BCH000182) 00888707002202,  1610DE (BCH000183) 00888707002202,  1610DE (BCH000184) 00888707002202,  1610DE (BCH000185) 00888707002202,  1610DE (BCH000186) 00888707002202,  1610DE (BCH000187) 00888707002202,  1610DE (BCH000188) 00888707002202,  1610DE (BCH000189) 00888707002202,  1610DE (BCH000190) 00888707002202,  1610DE (BCH000191) 00888707002202,  1610DE (BCH000192) 00888707002202,  1610DE (BCH000193) 00888707002202,  1610DE (BCH000194) 00888707002202,  1610DE (BCH000195) 00888707002202,  1610DE (BCH000196) 00888707002202,  1610DE (BCH000197) 00888707002202,  1610DE (BCH000198) 00888707002202,  1610DE (BCH000199) 00888707002202,  1610DE (BCH000200) 00888707002202,  1610DE (BCH000203) 00888707002202,  1610DE (BCH000204) 00888707002202,  1610DE (BCH000205) 00888707002202,  1610GB (BCH000206) 00888707002219,  1610DE (BCH000207) 00888707002202,  1610DE (BCH000208) 00888707002202,  1610DE (BCH000209) 00888707002202,  1610DE (BCH000210) 00888707002202,  1610DE (BCH000212) 00888707002202,  1610GB (BCH000213) 00888707002219,  1610GB (BCH000214) 00888707002219,  1610IT (BCH000216) 00888707002240,  1610DE (BCH000217) 00888707002202,  1610DE (BCH000218) 00888707002202,  1610IN (BCH000219) 00888707002998,  1610DE (BCH000220) 00888707002202,  1610DE (BCH000222) 00888707002202,  1610DE (BCH000224) 00888707002202,  1610DE (BCH000225) 00888707002202,  1610IT (BCH000227) 00888707002240,  1610DE (BCH000229) 00888707002202,  1610DE (BCH000230) 00888707002202,  1610DE (BCH000231) 00888707002202,  1610DE (BCH000232) 00888707002202,  1610AU (BCH000233) 00888707002165,  1610DE (BCH000234) 00888707002202,  1610DE (BCH000235) 00888707002202,  1610DE (BCH000236) 00888707002202,  1610DE (BCH000237) 00888707002202,  1610IT (BCH000238) 00888707002240,  1610DE (BCH000240) 00888707002202,  1610DE (BCH000241) 00888707002202,  1610DE (BCH000242) 00888707002202,  1610AU (BCH000243) 00888707002165,  1610DE (BCH000244) 00888707002202,  1610GB (BCH000245) 00888707002219,  1610DE (BCH000246) 00888707002202,  1610DE (BCH000247) 00888707002202,  1610IT (BCH000248) 00888707002240,  1610DE (BCH000249) 00888707002202,  1610DE (BCH000250) 00888707002202,  1610IT (BCH000251) 00888707002240,  1610IT (BCH000252) 00888707002240,  1610DE (BCH000253) 00888707002202,  1610DE (BCH000255) 00888707002202,  1610DE (BCH000256) 00888707002202,  1610DE (BCH000257) 00888707002202,  1610DE (BCH000259) 00888707002202,  1610DE (BCH000261) 00888707002202,  1610DE (BCH000262) 00888707002202,  1610DE (BCH000266) 00888707002202,  1610DE (BCH000267) 00888707002202,  1610DE (BCH000268) 00888707002202,  1610IT (BCH000269) 00888707002240,  1610DE (BCH000270) 00888707002202,  1610AU (BCH000271) 00888707002165,  1610DE (BCH000272) 00888707002202,  1610DE (BCH000274) 00888707002202,  1610DE (BCH000275) 00888707002202,  1610AU (BCH000276) 00888707002165,  1610DE (BCH000277) 00888707002202,  1610DE (BCH000278) 00888707002202,  1610DE (BCH000279) 00888707002202,  1610AU (BCH000280) 00888707002165,  1610DE (BCH000281) 00888707002202,  1610GB (BCH000282) 00888707002219,  1610GB (BCH000283) 00888707002219,  1610DE (BCH000284) 00888707002202,  1610DE (BCH000285) 00888707002202,  1610DE (BCH000286) 00888707002202,  1610DE (BCH000287) 00888707002202,  1610IL (BCH000288) 00888707002226,  1610AU (BCH000290) 00888707002165,  1610DE (BCH000292) 00888707002202,  1610GB (BCH000293) 00888707002219,  1610GB (BCH000294) 00888707002219,  1610GB (BCH000295) 00888707002219,  1610IT (BCH000296) 00888707002240,  1610DE (BCH000297) 00888707002202,  1610DE (BCH000299) 00888707002202,  1610DE (BCH000300) 00888707002202,  1610IT (bch000301) 00888707002240,  1610DE (BCH000302) 00888707002202,  1610GB (BCH000303) 00888707002219,  1610DE (bch000304) 00888707002202,  1610DE (bch000305) 00888707002202,  1610DE (bch000306) 00888707002202,  1610DE (BCH000307) 00888707002202,  1610AU (bch000308) 00888707002165,  1610DE (BCH000309) 00888707002202,  1610DE (BCH000310) 00888707002202,  1610DE (BCH000311) 00888707002202,  1610GB (BCH000312) 00888707002219,  1610DE (bch000313) 00888707002202,  1610DE (bch000314) 00888707002202,  1610DE (bch000315) 00888707002202,  1610DE (bch000316) 00888707002202,  1610DE (BCH000317) 00888707002202,  1610DE (BCH000318) 00888707002202,  1610DE (BCH000319) 00888707002202,  1610DE (BCH000320) 00888707002202,  1610DE (BCH000321) 00888707002202,  1610GB (BCH000322) 00888707002219,  1610GB (BCH000323) 00888707002219,  1610DE (bch000324) 00888707002202,  1610GB (BCH000325) 00888707002219,  1610DE (BCH000326) 00888707002202,  1610DE (BCH000327) 00888707002202,  1610GB (BCH000328) 00888707002219,  1610DE (bch000329) 00888707002202,  1610DE (BCH000330) 00888707002202,  1610DE (BCH000331) 00888707002202,  1610DE (BCH000332) 00888707002202,  1610DE (BCH000333) 00888707002202,  1610CH (BCH000334) 00888707002196,  1610DE (bch000335) 00888707002202,  1610IN (bch000336) 00888707002233,  1610AU (BCH000337) 00888707002165,  1610AU (BCH000338) 00888707002165,  1610CA (BCH000339) 00888707002189,  1610CA (BCH000340) 00888707002189,  1610AU (BCH000341) 00888707002165,  1610DE (bch000342) 00888707002202,  1610DE (bch000343) 00888707002202,  1610DE (BCH000344) 00888707002202,  1610DE (bch000345) 00888707002202,  1610DE (BCH000346) 00888707002202,  1610DE (BCH000347) 00888707002202,  1610GB (BCH000348) 00888707002219,  1610AU (bch000349) 00888707002165,  1610DE (BCH000350) 00888707002202,  1610DE (BCH000351) 00888707002202,  1610DE (BCH000352) 00888707002202,  1610DE (BCH000353) 00888707002202,  1610DE (BCH000354) 00888707002202,  1610DE (BCH000355) 00888707002202,  1610IT (bch000356) 00888707002240,  1610DE (BCH000357) 00888707002202,  1610DE (bch000358) 00888707002202,  1610AU (bch000359) 00888707002165,  1610AU (bch000360) 00888707002165,  1610AU (BCH000361) 00888707002165,  1610DE (bch000362) 00888707002202,  1610DE (BCH000363) 00888707002202,  1610DE (BCH000364) 00888707002202,  1610DE (BCH000365) 00888707002202,  1610DE (BCH000366) 00888707002202,  1610DE (BCH000367) 00888707002202,  1610DE (BCH000368) 00888707002202,  1610DE (BCH000369) 00888707002202,  1610IN (bch000370) 00888707002233,  1610GB (BCH000371) 00888707002219,  1610DE (bch000372) 00888707002202,  1610DE (BCH000373) 00888707002202,  1610DE (BCH000374) 00888707002202,  1610AU (BCH000375) 00888707002165,  1610DE (BCH000376) 00888707002202,  1610DE (BCH000377) 00888707002202,  1610DE (BCH000378) 00888707002202,  1610DE (BCH000379) 00888707002202,  1610DE (BCH000380) 00888707002202,  1610GB (BCH000381) 00888707002219,  1610GB (BCH000382) 00888707002219,  1610GB (BCH000383) 00888707002219,  1610CH (BCH000384) 00888707002196,  1610DE (bch000385) 00888707002202,  1610DE (bch000386) 00888707002202,  1610DE (BCH000387) 00888707002202,  1610DE (bch000388) 00888707002202,  1610DE (bch000389) 00888707002202,  1610DE (bch000390) 00888707002202,  1610DE (bch000391) 00888707002202,  1610DE (bch000392) 00888707002202,  1610GB (BCH000393) 00888707002219,  1610DE (bch000394) 00888707002202,  1610IN (bch000395) 00888707002233,  1610IN (bch000396) 00888707002233,  1610DE (bch000397) 00888707002202,  1610DE (bch000398) 00888707002202,  1610DE (bch000399) 00888707002202,  1610DE (bch000400) 00888707002202,  1610DE (BCH000401) 00888707002202,  1610DE (BCH000402) 00888707002202,  1610DE (BCH000403) 00888707002202,  1610DE (BCH000404) 00888707002202,  1610AU (BCH000405) 00888707002165,  1610AU (BCH000406) 00888707002165,  1610DE (BCH000407) 00888707002202,  1610DE (BCH000408) 00888707002202,  1610DE (BCH000409) 00888707002202,  1610DE (BCH000410) 00888707002202,  1610DE (BCH000411) 00888707002202,  1610DE (BCH000412) 00888707002202,  1610GB (BCH000413) 00888707002219,  1610DE (BCH000414) 00888707002202,  1610DE (BCH000415) 00888707002202,  1610DE (BCH000416) 00888707002202,  1610DE (BCH000417) 00888707002202,  1610DE (BCH000418) 00888707002202,  1610DE (BCH000419) 00888707002202,  1610DE (BCH000420) 00888707002202,  1610DE (BCH000421) 00888707002202,  1610AU (BCH000422) 00888707002165,  1610DE (BCH000423) 00888707002202,  1610DE (BCH000424) 00888707002202,  1610DE (BCH000425) 00888707005784,  1610AU (BCH000426) 00888707002165,  1610DE (BCH000427) 00888707002202,  1610AU (BCH000428) 00888707002165,  1610DE (BCH000429) 00888707002202,  1610DE (BCH000430) 00888707005784,  1610DE (BCH000431) 00888707002202,  1610DE (BCH000432) 00888707002202,  1610DE (BCH000433) 00888707002202,  1610DE (BCH000434) 00888707002202,  1610DE (BCH000435) 00888707002202,  1610DE (BCH000436) 00888707002202,  1610DE (BCH000437) 00888707002202,  1610GB (BCH000438) 00888707002219,  1610DE (BCH000439) 00888707002202,  1610AU (BCH000440) 00888707002165,  1610DE (BCH000441) 00888707002202,  1610DE (BCH000442) 00888707002202,  1610DE (BCH000443) 00888707002202,  1610DE (BCH000444) 00888707002202,  1610DE (BCH000445) 00888707002202,  1610DE (BCH000446) 00888707002202,  1610DE (BCH000447) 00888707002202,  1610DE (BCH000448) 00888707002202,  1610DE (BCH000449) 00888707002202,  1610DE (BCH000450) 00888707002202,  1610AU (BCH000451) 00888707002165,  1610DE (BCH000452) 00888707002202,  1610DE (BCH000453) 00888707002202,  1610DE (BCH000454) 00888707002202,  1610DE (BCH000455) 00888707002202,  1610DE (BCH000456) 00888707002202,  1610GB (BCH000457) 00888707002219,  1610GB (BCH000458) 00888707002219,  1610DE (BCH000459) 00888707002202,  1610DE (BCH000460) 00888707002202,  1610DE (BCH000461) 00888707002202,  1610DE (BCH000462) 00888707002202,  1610DE (BCH000463) 00888707002202,  1610DE (BCH000464) 00888707002202,  1610DE (BCH000465) 00888707002202,  1610DE (BCH000466) 00888707002202,  1610DE (BCH000467) 00888707002202,  1610DE (BCH000468) 00888707002202,  1610DE (BCH000469) 00888707002202,  1610DE (BCH000470) 00888707002202,  1610DE (BCH000471) 00888707002202,  1610DE (BCH000472) 00888707002202,  1610DE (BCH000473) 00888707002202,  1610DE (BCH000474) 00888707002202,  1610DE (BCH000475) 00888707002202,  1610DE (BCH000476) 00888707002202,  1610DE (BCH000477) 00888707002202,  1610GB (BCH000478) 00888707002219,  1610GB (BCH000479) 00888707002219,  1610GB (BCH000480) 00888707002219,  1610DE (BCH000481) 00888707002202,  1610DE (BCH000482) 00888707002202,  1610DE (BCH000483) 00888707002202,  1610DE (BCH000484) 00888707002202,  1610DE (BCH000485) 00888707002202,  1610AU (BCH000486) 00888707002165,  1610DE (BCH000487) 00888707002202,  1610DE (BCH000488) 00888707002202,  1610DE (BCH000489) 00888707002202,  1610DE (BCH000490) 00888707002202,  1610DE (BCH000491) 00888707002202,  1610DE (BCH000492) 00888707002202,  1610AU (BCH000493) 00888707002165,  1610DE (BCH000494) 00888707002202,  1610AU (BCH000495) 00888707002165,  1610DE (BCH000496) 00888707002202,  1610DE (BCH000497) 00888707002202,  1610DE (BCH000498) 00888707002202,  1610IN (BCH000499) 00888707002233,  1610DE (BCH000500) 00888707002202,  1610DE (BCH000501) 00888707002202,  1610DE (BCH000502) 00888707002202,  1610CA (BCH000503) 00888707005760,  1610DE (BCH000504) 00888707002202,  1610DE (BCH000505) 00888707002202,  1610DE (BCH000506) 00888707002202,  1610AU (BCH000507) 00888707002165,  1610JP (BCH000508) 00888707002257,  1610DE (BCH000509) 00888707002202,  1610DE (BCH000510) 00888707002202,  1610DE (BCH000511) 00888707002202,  1610DE (BCH000512) 00888707002202,  1610CA (BCH000513) 00888707005760,  1610IT (BCH000514) 00888707002240,  1610DE (BCH000515) 00888707002202,  1610DE (BCH000516) 00888707002202,  1610DE (BCH000517) 00888707002202,  1610DE (BCH000518) 00888707002202,  1610IT (BCH000519) 00888707002240,  1610DE (BCH000520) 00888707002202,  1610DE (BCH000521) 00888707002202,  1610DE (BCH000522) 00888707002202,  1610DE (BCH000523) 00888707002202,  1610DE (BCH000524) 00888707002202,  1610AU (BCH000525) 00888707002165,  1610DE (BCH000526) 00888707005784,  1610AU (BCH000527) 00888707002165,  1610DE (BCH000528) 00888707002202,  1610DE (BCH000529) 00888707002202,  1610JP (BCH000530) 00888707002257,  1610DE (BCH000531) 00888707005784,  1610DE (BCH000532) 00888707002202,  1610DE (BCH000533) 00888707002202,  1610DE (BCH000534) 00888707002202,  1610DE (BCH000535) 00888707002202,  1610DE (BCH000537) 00888707002202,  1610DE (BCH000538) 00888707002202,  1610DE (BCH000539) 00888707002202,  1610DE (BCH000540) 00888707002202,  1610DE (BCH000541) 00888707002202,  1610DE (BCH000542) 00888707002202,  1610DE (BCH000543) 00888707002202,  1610CA (BCH000544) 00888707005760,  1610CH (BCH000546) 00888707002196,  1610DE (BCH000547) 00888707002202,  1610DE (BCH000548) 00888707002202,  1610DE (BCH000549) 00888707002202,  1610CA (BCH000550) 00888707002189,  1610AU (BCH000551) 00888707002165,  1610CA (BCH000552) 00888707002189,  1610AU (BCH000553) 00888707002165,  1610JP (BCH000554) 00888707002257,  1610CA (BCH000555) 00888707002189,  1610DE (BCH000556) 00888707002202,  1610CA (BCH000557) 00888707002189,  1610DE (BCH000558) 00888707005784,  1610DE (BCH000559) 00888707002202,  1610DE (BCH000560) 00888707005784,  1610DE (BCH000561) 00888707002202,  1610DE (BCH000562) 00888707002202,  1610DE (BCH000563) 00888707002202,  1610DE (BCH000564) 00888707002202,  1610DE (BCH000565) 00888707002202,  1610DE (BCH000566) 00888707002202,  1610DE (BCH000567) 00888707002202,  1610GB (BCH000568) 00888707002219,  1610DE (BCH000569) 00888707002202,  1610DE (BCH000570) 00888707002202. 1610DE (BCH000571) 00888707002202,  1610DE (BCH000572) 00888707002202,  1610DE (BCH000573) 00888707002202,  1610DE (BCH000574) 00888707002202,  1610DE (BCH000575) 00888707002202,  1610DE (BCH000576) 00888707002202,  1610DE (BCH000577) 00888707002202,  1610DE (BCH000578) 00888707002202,  1610AU (BCH000579) 00888707002165,  1610DE (BCH000581) 00888707002202,  1610GB (BCH000582) 00888707002219,  1610DE (BCH000583) 00888707002202,  1610DE (BCH000584) 00888707002202,  1610DE (BCH000585) 00888707002202,  1610DE (BCH000586) 00888707002202,  1610DE (BCH000587) 00888707002202,  1610DE (BCH000588) 00888707002202,  1610DE (BCH000589) 00888707002202,  1610DE (BCH000590) 00888707002202,  1610DE (BCH000591) 00888707002202,  1610DE (BCH000592) 00888707002202,  1610DE (BCH000593) 00888707002202,  1610CA (BCH000594) 00888707005760,  1610DE (BCH000596) 00888707002202,  1610DE (BCH000597) 00888707002202,  1610CA (BCH000598) 00888707005760,  1610DE (BCH000599) 00888707002202,  1610DE (BCH000600) 00888707002202,  1610DE (BCH000602) 00888707002202,  1610DE (BCH000603) 00888707002202,  1610DE (BCH000604) 00888707002202,  1610DE (BCH000605) 00888707002202,  1610DE (BCH000606) 00888707002202,  1610DE (BCH000607) 00888707002202,  1610DE (bch000608) 00888707002202,  1610IT (BCH000609) 00888707002240,  1610DE (BCH000610) 00888707002202,  1610DE (BCH000612) 00888707002202,  1610IN (BCH000613) 00888707002233,  1610IN (BCH000614) 00888707002233,  1610CH (BCH000615) 00888707002196,  1610IT (BCH000616) 00888707002240,  1610DE (BCH000617) 00888707002202,  1610DE (BCH000618) 00888707002202,  1610DE (BCH000619) 00888707002202,  1610DE (BCH000620) 00888707002202,  1610DE (BCH000621) 00888707002202,  1610GB (BCH000622) 00888707002219,  1610DE (BCH000623) 00888707002202,  1610DE (BCH000624) 00888707002202,  1610DE (BCH000625) 00888707002202,  1610GB (BCH000626) 00888707002219,  1610DE (BCH000627) 00888707002202,  1610DE (BCH000628) 00888707002202,  1610DE (BCH000629) 00888707002202,  1610IT (BCH000630) 00888707002240,  1610DE (BCH000631) 00888707002202,  1610DE (BCH000632) 00888707002202,  1610GB (BCH000634) 00888707002219,  1610DE (BCH000635) 00888707002202,  1610GB (BCH000636) 00888707002219,  1610DE (BCH000638) 00888707002202,  1610DE (BCH000639) 00888707002202,  1610DE (BCH000640) 00888707002202,  1610DE (BCH000641) 00888707002202,  1610DE (BCH000642) 00888707002202,  1610DE (BCH000643) 00888707002202,  1610DE (BCH000646) 00888707002202,  1610IT (BCH000647) 00888707002240,  1610DE (BCH000648) 00888707002202,  1610DE (bch000649) 00888707002202,  1610DE (BCH000650) 00888707002202,  1610DE (BCH000651) 00888707002202,  1610DE (BCH000652) 00888707002202,  1610DE (BCH000653) 00888707002202,  1610DE (BCH000654) 00888707002202,  1610DE (BCH000655) 00888707002202,  1610DE (BCH000656) 00888707002202,  1610DE (BCH000657) 00888707002202,  1610DE (BCH000658) 00888707002202,  1610DE (BCH000659) 00888707002202,  1610DE (BCH000660) 00888707002202,  1610DE (BCH000661) 00888707002202,  1610JP (BCH000662) 00888707002257,  1610DE (BCH000663) 00888707002202,  1610DE (BCH000664) 00888707002202,  1610DE (BCH000665) 00888707002202,  1610DE (BCH000666) 00888707002202,  1610DE (BCH000668) 00888707002202,  1610DE (BCH000669) 00888707005784,  1610DE (BCH000670) 00888707002202,  1610DE (BCH000671) 00888707002202,  1610DE (BCH000672) 00888707005784,  1610IT (BCH000673) 00888707002240,  1610IT (BCH000674) 00888707002240,  1610DE (BCH000675) 00888707002202,  1610DE (BCH000677) 00888707002202,  1610DE (BCH000678) 00888707002202,  1610DE (bch000679) 00888707002202,  1610AU (BCH000680) 00888707002165,  1610DE (bch000681) 00888707002202,  1610DE (bch000682) 00888707002202,  1610DE (bch000683) 00888707002202,  1610DE (BCH000684) 00888707002202,  1610DE (BCH000685) 00888707002202,  1610DE (BCH000686) 00888707002202,  1610DE (BCH000687) 00888707002202,  1610DE (bch000689) 00888707002202,  1610AU (BCH000690) 00888707002165,  1610DE (bch000692) 00888707002202,  1610DE (BCH000695) 00888707002202,  1610DE (BCH000697) 00888707002202,  1610GB (BCH000699) 00888707002219,  1610GB (BCH000700) 00888707002219,  1610GB (BCH000701) 00888707002219,  1610IT (BCH000702) 00888707002240,  1610AU (BCH000703) 00888707002165,  1610DE (BCH000704) 00888707002202,  1610IT (BCH000705) 00888707002240,  1610DE (bch000707) 00888707002202,  1610DE (BCH000708) 00888707002202,  1610DE (BCH000709) 00888707002202,  1610DE (BCH000710) 00888707005784,  1610DE (bch000711) 00888707002202,  1610GB (BCH000712) 00888707002219,  1610DE (bch000713) 00888707002202,  1610GB (BCH000714) 00888707002219,  1610DE (BCH000715) 00888707005784,  1610DE (BCH000716) 00888707002202,  1610DE (BCH000717) 00888707002202,  1610GB (BCH000718) 00888707002219,  1610DE (bch000719) 00888707002202,  1610DE (BCH000720) 00888707005784,  1610DE (BCH000721) 00888707002202,  1610AU (BCH000722) 00888707002165,  1610IN (BCH000723) 00888707002233,  1610GB (BCH000724) 00888707002219,  1610DE (BCH000725) 00888707002202,  1610IT (BCH000726) 00888707002240,  1610DE (BCH000727) 00888707005784,  1610DE (BCH000728) 00888707002202,  1610DE (BCH000729) 00888707002202,  1610DE (BCH000730) 00888707002202,  1610DE (BCH000731) 00888707002202,  1610DE (BCH000732) 00888707002202,  1610DE (BCH000733) 00888707002202,  1610DE (BCH000734) 00888707002202,  1610DE (BCH000735) 00888707002202,  1610DE (BCH000736) 00888707002202,  1610DE (BCH000737) 00888707002202,  1610DE (BCH000738) 00888707002202,  1610DE (BCH000739) 00888707002202,  1610DE (bch000740) 00888707002202,  1610DE (bch000742) 00888707002202,  1610DE (BCH000743) 00888707002202,  1610DE (BCH000744) 00888707002202,  1610DE (bch000746) 00888707002202,  1610DE (bch000747) 00888707002202,  1610DE (bch000749) 00888707002202,  1610DE (BCH000750) 00888707002202,  1610GB (BCH000751) 00888707002219,  1610GB (BCH000752) 00888707002219,  1610DE (BCH000753) 00888707005784,  1610DE (BCH000754) 00888707002202,  1610DE (BCH000755) 00888707002202,  1610DE (BCH000756) 00888707002202,  1610DE (bch000757) 00888707002202,  1610IT (BCH000758) 00888707002240,  1610DE (bch000759) 00888707002202,  1610DE (BCH000760) 00888707002202,  1610DE (BCH000765) 00888707002202,  1610DE (BCH000766) 00888707002202,  1610DE (BCH000767) 00888707002202,  1610DE (BCH000768) 00888707002202,  1610DE (BCH000770) 00888707002202,  1610DE (bch000771) 00888707002202,  1610DE (bch000772) 00888707002202,  1610GB (BCH000773) 00888707002219,  1610DE (bch000774) 00888707002202,  1610DE (bch000776) 00888707002202,  1610GB (BCH000780) 00888707002219,  1610DE (BCH000781) 00888707002202,  1610DE (BCH000782) 00888707002202,  1610DE (BCH000783) 00888707002202,  1610DE (BCH000829) 00888707002967,  1610DE (BCH000830) 00888707002967,  1610DE (BCH000831) 00888707002967,  1610DE (BCH000832) 00888707002967,  1610DE (BCH000833) 00888707002967,  1610GB (BCH000834) 00888707002974,  1610DE (BCH000835) 00888707002967,  1610DE (BCH000836) 00888707002967,  1610GB (BCH000837) 00888707002974,  1610DE (BCH000838) 00888707002967,  1610DE (BCH000839) 00888707002967,  1610DE (BCH000840) 00888707002967,  1610GB (BCH000841) 00888707002974,  1610DE (BCH000842) 00888707002967,  1610DE (BCH000843) 00888707002967,  1610AU (BCH000844) 00888707002165,  1610DE (BCH000845) 00888707002967,  1610DE (BCH000846) 00888707002967,  1610DE (BCH000847) 00888707002967,  1610DE (BCH000849) 00888707002967,  1610DE (BCH000851) 00888707002967,  1610CH (BCH000852) 00888707002196,  1610DE (BCH000853) 00888707002967,  1610CH (BCH000854) 00888707002196,  1610DE (BCH000855) 00888707002967,  1610DE (BCH000856) 00888707002967,  1610DE (BCH000857) 00888707002967,  1610DE (BCH000858) 00888707002967,  1610DE (BCH000859) 00888707002967,  1610DE (BCH000860) 00888707002967,  1610DE (BCH000861) 00888707002967,  1610DE (BCH000862) 00888707002967,  1610DE (BCH000863) 00888707002967,  1610DE (BCH000865) 00888707002967,  1610DE (BCH000866) 00888707002967,  1610DE (BCH000867) 00888707002967,  1610DE (BCH000868) 00888707002967,  1610DE (bch000869) 00888707002202,  1610DE (bch000870) 00888707002202,  1610IN (BCH000871) 00888707002233,  1610DE (BCH000872) 00888707002202,  1610DE (bch000873) 00888707002202,  1610DE (BCH000874) 00888707005784,  1610DE (BCH000875) 00888707002202,  1610DE (BCH000876) 00888707002202,  1610DE (BCH000877) 00888707002202,  1610DE (BCH000878) 00888707002202,  1610DE (BCH000879) 00888707002202,  1610JP (BCH000880) 00888707002257,  1610DE (BCH000881) 00888707002202,  1610DE (BCH000882) 00888707002967,  1610JP (BCH000883) 00888707002257,  1610DE (bch000884) 00888707002202,  1610DE (bch000885) 00888707002202,  1610DE (BCH000886) 00888707002202,  1610DE (BCH000887) 00888707002202,  1610DE (bch000888) 00888707002202,  1610DE (bch000889) 00888707002202,  1610DE (BCH000890) 00888707002202,  1610GB (BCH000891) 00888707002219,  1610GB (BCH000892) 00888707002219,  1610GB (BCH000893) 00888707002219,  1610DE (BCH000894) 00888707002202,  1610DE (BCH000896) 00888707002202,  1610GB (BCH000897) 00888707002219,  1610DE (BCH000898) 00888707002202,  1610DE (BCH000899) 00888707002202,  1610DE (BCH000900) 00888707002202,  1610DE (BCH000901) 00888707002202,  1610DE (BCH000902) 00888707002202,  1610DE (BCH000903) 00888707002202,  1610DE (BCH000904) 00888707002202,  1610DE (BCH000905) 00888707002202,  1610DE (BCH000906) 00888707002967,  1610DE (BCH000907) 00888707002202,  1610GB (BCH000908) 00888707002219,  1610GB (BCH000909) 00888707002219,  1610JP (BCH000910) 00888707002257,  1610GB (BCH000911) 00888707002219,  1610DE (BCH000912) 00888707002202,  1610DE (BCH000913) 00888707002202,  1610AU (BCH000914) 00888707002165,  1610DE (BCH000915) 00888707005784,  1610DE (BCH000916) 00888707002202,  1610DE (bch000917) 00888707002202,  1610DE (bch000918) 00888707002202,  1610DE (bch000919) 00888707002202,  1610DE (BCH000920) 00888707002202,  1610DE (BCH000921) 00888707002202,  1610DE (BCH000922) 00888707002202,  1610DE (BCH000923) 00888707002202,  1610DE (BCH000924) 00888707002202,  1610DE (BCH000925) 00888707002202,  1610DE (BCH000926) 00888707002202,  1610DE (BCH000928) 00888707002202,  1610DE (BCH000929) 00888707002202,  1610DE (BCH000930) 00888707002202,  1610DE (BCH000931) 00888707002202,  1610CA (BCH000932) 00888707005760,  1610JP (BCH000933) 00888707002257,  1610DE (BCH000934) 00888707005784,  1610CA (BCH000935) 00888707005760,  1610GB (BCH000936) 00888707002219,  1610DE (BCH000937) 00888707005784,  1610DE (BCH000938) 00888707002202,  1610DE (BCH000939) 00888707002202,  1610DE (BCH000940) 00888707002202,  1610DE (BCH000941) 00888707002202,  1610CA (BCH000942) 00888707005760,  1610DE (BCH000943) 00888707002202,  1610DE (BCH000944) 00888707002202,  1610DE (BCH000945) 00888707002202,  1610DE (BCH000946) 00888707002202,  1610DE (BCH000947) 00888707002202,  1610DE (BCH000948) 00888707002202,  1610DE (BCH000949) 00888707002202,  1610DE (BCH000950) 00888707002202,  1610GB (BCH000951) 00888707002219,  1610JP (BCH000952) 00888707002257,  1610DE (BCH000953) 00888707002202,  1610DE (BCH000954) 00888707002202,  1610CA (BCH000955) 00888707005760,  1610DE (BCH000956) 00888707002202,  1610DE (BCH000957) 00888707002202,  1610DE (BCH000958) 00888707002202,  1610DE (BCH000959) 00888707002202,  1610DE (BCH000960) 00888707005784,  1610CA (BCH000961) 00888707005760,  1610CA (BCH000962) 00888707005760,  1610CA (BCH000963) 00888707005760,  1610JP (BCH000964) 00888707002257,  1610DE (BCH000965) 00888707002202,  1610DE (BCH000966) 00888707002202,  1610DE (BCH000967) 00888707005784,  1610DE (BCH000968) 00888707002202,  1610DE (BCH000969) 00888707002202,  1610DE (BCH000970) 00888707005784,  1610CA (BCH000971) 00888707005760,  1610CA (BCH000972) 00888707005760,  1610DE (BCH000973) 00888707002202,  1610DE (BCH000974) 00888707002202,  1610DE (BCH000975) 00888707002202,  1610DE (BCH000976) 00888707002202,  1610DE (BCH000977) 00888707002202,  1610DE (BCH000978) 00888707002202,  1610DE (BCH000979) 00888707002202,  1610CA (BCH000980) 00888707005760,  1610CA (BCH000981) 00888707002189,  1610DE (BCH000982) 00888707005784,  1610IT (BCH000983) 00888707002240,  1610CA (BCH000984) 00888707005760,  1610DE (BCH000985) 00888707002202,  1610DE (BCH000986) 00888707002202,  1610DE (BCH000987) 00888707002202,  1610DE (BCH000988) 00888707002202,  1610DE (BCH000989) 00888707002202,  1610DE (BCH000990) 00888707002202,  1610DE (BCH000991) 00888707002202,  1610DE (BCH000992) 00888707002202,  1610DE (BCH000993) 00888707002202,  1610DE (BCH000994) 00888707002202,  1610DE (BCH000995) 00888707002202,  1610DE (BCH000996) 00888707005784,  1610DE (BCH000997) 00888707005784,  1610DE (BCH000998) 00888707005784,  1610DE (BCH000999) 00888707002202,  1610DE (BCH001000) 00888707002202,  1610DE (BCH001001) 00888707005784,  1610CA (BCH001002) 00888707005760,  1610DE (BCH001003) 00888707002202,  1610DE (BCH001004) 00888707002202,  1610DE (BCH001005) 00888707002202,  1610CA (BCH001006) 00888707005760,  1610DE (BCH001007) 00888707002202,  1610DE (BCH001008) 00888707002202,  1610DE (BCH001009) 00888707002202,  1610DE (BCH001010) 00888707002202,  1610DE (BCH001011) 00888707002202,  1610DE (BCH001012) 00888707002202,  1610DE (BCH001013) 00888707002202,  1610DE (BCH001014) 00888707002202,  1610DE (bch001015) 00888707002202,  1610DE (BCH001016) 00888707002202,  1610DE (bch001017) 00888707002202,  1610DE (BCH001018) 00888707002202,  1610DE (BCH001019) 00888707002967,  1610DE (BCH001020) 00888707002202,  1610DE (BCH001021) 00888707002967,  1610DE (BCH001022) 00888707002967,  1610DE (BCH001023) 00888707002202,  1610DE (BCH001024) 00888707002967,  1610DE (BCH001025) 00888707002967,  1610DE (BCH001026) 00888707002967,  1610DE (BCH001027) 00888707002967,  1610DE (BCH001028) 00888707002202,  1610DE (BCH001029) 00888707002202,  1610DE (BCH001030) 00888707002967,  1610DE (BCH001031) 00888707002967,  1610DE (BCH001032) 00888707002967,  1610DE (BCH001033) 00888707002967,  1610DE (BCH001034) 00888707002967,  1610DE (BCH001035) 00888707002967,  1610DE (BCH001036) 00888707002202,  1610DE (BCH001037) 00888707002967,  1610DE (BCH001038) 00888707002967,  1610DE (BCH001040) 00888707002967,  1610DE (BCH001041) 00888707002967,  1610DE (BCH001042) 00888707002202,  1610AU (BCH001043) 00888707002165,  1610AU (BCH001044) 00888707002165,  1610DE (BCH001045) 00888707002202,  1610DE (BCH001046) 00888707002202,  1610DE (BCH001047) 00888707002202,  1610DE (BCH001048) 00888707002967,  1610DE (BCH001049) 00888707002967,  1610DE (BCH001050) 00888707002967,  1610DE (BCH001051) 00888707002967,  1610DE (BCH001052) 00888707002967,  1610CA (BCH001053) 00888707003001,  1610DE (BCH001054) 00888707002202,  1610DE (BCH001056) 00888707002967,  1610DE (BCH001057) 00888707002202,  1610DE (BCH001058) 00888707002967,  1610DE (BCH001059) 00888707002967,  1610DE (BCH001060) 00888707002967,  1610DE (BCH001061) 00888707002967,  1610DE (BCH001062) 00888707002967,  1610DE (BCH001063) 00888707002202,  1610DE (BCH001064) 00888707002967,  1610DE (BCH001065) 00888707002202,  1610DE (BCH001066) 00888707002202,  1610DE (BCH001067) 00888707002967,  1610DE (BCH001068) 00888707002967,  1610DE (BCH001069) 00888707002967,  1610DE (BCH001070) 00888707002967,  1610CA (BCH001071) 00888707003001,  1610DE (BCH001072) 00888707002967,  1610DE (BCH001073) 00888707002967,  1610DE (BCH001074) 00888707002967,  1610DE (BCH001075) 00888707002967,  1610DE (BCH001076) 00888707002967,  1610DE (BCH001077) 00888707002967,  1610DE (BCH001078) 00888707002967,  1610DE (BCH001079) 00888707002967,  1610DE (BCH001080) 00888707002967,  1610CA (BCH001081) 00888707003001,  1610DE (BCH001082) 00888707002967,  1610DE (BCH001083) 00888707002967,  1610DE (BCH001084) 00888707002202,  1610DE (BCH001085) 00888707002202,  1610DE (BCH001086) 00888707002202,  1610DE (BCH001087) 00888707002202,  1610DE (BCH001088) 00888707002202,  1610DE (BCH001089) 00888707002202,  1610DE (BCH001090) 00888707002967,  1610DE (BCH001091) 00888707002202,  1610DE (BCH001092) 00888707002202,  1610DE (BCH001093) 00888707002202,  1610DE (BCH001094) 00888707002202,  1610DE (BCH001095) 00888707002967,  1610DE (BCH001096) 00888707002967,  1610DE (BCH001097) 00888707002967,  1610DE (BCH001098) 00888707002967,  1610DE (BCH001099) 00888707002967,  1610DE (BCH001100) 00888707002202,  1610DE (BCH001101) 00888707002202,  1610DE (BCH001102) 00888707002967,  1610CA (BCH001103) 00888707003001,  1610CA (BCH001104) 00888707003001,  1610DE (BCH001105) 00888707002967,  1610DE (BCH001106) 00888707002967,  1610DE (BCH001107) 00888707002967,  1610DE (BCH001108) 00888707002202,  1610DE (BCH001109) 00888707002202,  1610DE (BCH001110) 00888707002202,  1610DE (BCH001111) 00888707002202,  1610DE (BCH001112) 00888707002202,  1610DE (BCH001113) 00888707002202,  1610DE (BCH001114) 00888707002967,  1610DE (BCH001115) 00888707002967,  1610DE (BCH001116) 00888707002967,  1610DE (BCH001117) 00888707002967,  1610DE (BCH001118) 00888707002967,  1610DE (BCH001119) 00888707002967,  1610DE (BCH001120) 00888707002967,  1610DE (BCH001121) 00888707002967,  1610DE (BCH001122) 00888707005784,  1610CA (BCH001123) 00888707003001,  1610DE (BCH001124) 00888707005784,  1610DE (BCH001125) 00888707005784,  1610DE (BCH001126) 00888707005784,  1610DE (BCH001127) 00888707002967,  1610DE (BCH001128) 00888707005784,  1610DE (BCH001129) 00888707002967,  1610DE (BCH001130) 00888707002202,  1610DE (BCH001131) 00888707002202,  1610DE (BCH001132) 00888707005784,  1610CA (BCH001133) 00888707003001,  1610DE (BCH001134) 00888707002967,  1610JP (BCH001135) 00888707002257,  1610DE (BCH001136) 00888707002202,  1610DE (BCH001137) 00888707002202,  1610DE (BCH001138) 00888707002202,  1610JP (BCH001139) 00888707002257,  1610DE (BCH001140) 00888707002967,  1610DE (BCH001141) 00888707005784,  1610DE (BCH001142) 00888707002202,  1610DE (BCH001143) 00888707002202,  1610DE (BCH001145) 00888707002967,  1610DE (BCH001147) 00888707002967,  1610DE (BCH001148) 00888707002967,  1610JP (BCH001149) 00888707002257,  1610DE (BCH001150) 00888707005784,  1610DE (BCH001151) 00888707002202,  1610AU (BCH001152) 00888707002165,  1610DE (BCH001153) 00888707002202,  1610DE (BCH001154) 00888707005784,  1610DE (BCH001155) 00888707005784,  1610DE (BCH001156) 00888707002202,  1610DE (BCH001157) 00888707005784,  1610DE (BCH001158) 00888707005784,  1610DE (BCH001159) 00888707005784,  1610JP (BCH001161) 00888707002257,  1610CA (BCH001162) 00888707003001,  1610DE (BCH001163) 00888707002202,  1610DE (BCH001164) 00888707002202,  1610DE (BCH001165) 00888707002202,  1610DE (BCH001166) 00888707005784,  1610IT (BCH001169) 00888707002981,  1610AU (BCH001170) 00888707002165,  1610DE (BCH001171) 00888707002967,  1610DE (BCH001172) 00888707002967,  1610AU (BCH001173) 00888707002165,  1610DE (BCH001177) 00888707002967,  1610DE (BCH001179) 00888707002967,  1610DE (BCH001180) 00888707002967,  1610DE (BCH001181) 00888707002967,  1610DE (BCH001184) 00888707002967,  1610DE (BCH001185) 00888707002967,  1610GB (BCH001186) 00888707002974,  1610DE (BCH001188) 00888707002967,  1610GB (BCH001189) 00888707002974,  1610DE (BCH001190) 00888707002967,  1610DE (BCH001191) 00888707002967,  1610DE (BCH001192) 00888707002967,  1610GB (BCH001194) 00888707002974,  1610DE (BCH001195) 00888707002967,  1610DE (BCH001196) 00888707002967,  1610AU (BCH001197) 00888707002165,  1610GB (BCH001198) 00888707002974,  1610DE (BCH001199) 00888707002967,  1610AU (BCH001200) 00888707002165,  1610DE (BCH001201) 00888707002967,  1610DE (BCH001202) 00888707002967,  1610DE (BCH001203) 00888707002967,  1610GB (BCH001204) 00888707002974,  1610DE (BCH001206) 00888707002967,  1610DE (BCH001207) 00888707002967,  1610DE (BCH001208) 00888707002967,  1610IT (BCH001209) 00888707002981,  1610DE (BCH001210) 00888707002967,  1610DE (BCH001211) 00888707002967,  1610GB (BCH001212) 00888707002974,  1610DE (BCH001213) 00888707002967,  1610DE (BCH001214) 00888707002967,  1610DE (BCH001215) 00888707002967,  1610GB (BCH001216) 00888707002974,  1610DE (BCH001217) 00888707002967,  1610DE (BCH001218) 00888707002967,  1610DE (BCH001219) 00888707002967,  1610IT (BCH001220) 00888707002981,  1610DE (BCH001222) 00888707002967,  1610DE (BCH001223) 00888707002967,  1610DE (BCH001224) 00888707002967,  1610DE (BCH001225) 00888707002967,  1610DE (BCH001226) 00888707002967,  1610DE (BCH001227) 00888707002967,  1610DE (BCH001228) 00888707002967,  1610IT (BCH001229) 00888707002981,  1610DE (BCH001230) 00888707002967,  1610DE (BCH001231) 00888707002967,  1610DE (BCH001232) 00888707002967,  1610DE (BCH001233) 00888707002967,  1610DE (BCH001234) 00888707002967,  1610DE (BCH001235) 00888707002967,  1610DE (BCH001236) 00888707002967,  1610DE (BCH001237) 00888707002967,  1610DE (BCH001238) 00888707002967,  1610DE (BCH001239) 00888707002967,  1610DE (BCH001240) 00888707002967,  1610DE (BCH001241) 00888707002967,  1610DE (BCH001242) 00888707002967,  1610DE (BCH001243) 00888707002967,  1610DE (BCH001244) 00888707002967,  1610IT (BCH001245) 00888707002981,  1610IT (BCH001246) 00888707002981,  1610DE (BCH001247) 00888707002967,  1610DE (BCH001248) 00888707002967,  1610DE (BCH001249) 00888707002967,  1610GB (BCH001250) 00888707002974,  1610DE (BCH001251) 00888707002967,  1610DE (BCH001252) 00888707002967,  1610DE (BCH001253) 00888707002967,  1610DE (BCH001254) 00888707002967,  1610IT (BCH001255) 00888707002981,  1610CA (BCH001256) 00888707003001,  1610DE (BCH001257) 00888707002967,  1610DE (BCH001258) 00888707002967,  1610DE (BCH001259) 00888707002967,  1610DE (BCH001260) 00888707002967,  1610DE (BCH001261) 00888707002967,  1610DE (BCH001262) 00888707002967,  1610IT (BCH001263) 00888707002981,  1610GB (BCH001264) 00888707002974,  1610DE (BCH001265) 00888707002967,  1610GB (BCH001266) 00888707002974,  1610DE (BCH001267) 00888707002967,  1610DE (BCH001268) 00888707002967,  1610DE (BCH001269) 00888707002967,  1610CH (BCH001270) 00888707002196,  1610IT (BCH001271) 00888707002981,  1610GB (BCH001272) 00888707002974,  1610IT (BCH001273) 00888707002981,  1610CA (BCH001274) 00888707003001,  1610GB (BCH001275) 00888707002974,  1610AU (BCH001276) 00888707002165,  1610CH (BCH001277) 00888707002196,  1610GB (BCH001278) 00888707002974,  1610DE (BCH001279) 00888707002967,  1610CA (BCH001280) 00888707003001,  1610CA (BCH001281) 00888707003001,  1610DE (BCH001282) 00888707002967,  1610DE (BCH001283) 00888707002967,  1610DE (BCH001284) 00888707002967,  1610GB (BCH001285) 00888707002974,  1610DE (BCH001286) 00888707002967,  1610IT (BCH001287) 00888707002981,  1610DE (BCH001288) 00888707002967,  1610IN (BCH001289) 00888707002998,  1610DE (BCH001290) 00888707002967,  1610CH (BCH001291) 00888707002196,  1610DE (BCH001292) 00888707002967,  1610DE (BCH001293) 00888707002967,  1610DE (BCH001294) 00888707002967,  1610IT (BCH001295) 00888707002981,  1610AU (BCH001296) 00888707002165,  1610GB (BCH001297) 00888707002974,  1610DE (BCH001298) 00888707002967,  1610AU (BCH001299) 00888707002165,  1610DE (BCH001300) 00888707002967,  1610DE (BCH001301) 00888707002967,  1610AU (BCH001302) 00888707002165,  1610DE (BCH001303) 00888707002967,  1610DE (BCH001304) 00888707002967,  1610DE (BCH001305) 00888707002967,  1610DE (BCH001306) 00888707002967,  1610DE (BCH001307) 00888707002967,  1610CH (BCH001308) 00888707002196,  1610DE (BCH001309) 00888707002967,  1610DE (BCH001310) 00888707002967,  1610IT (BCH001311) 00888707002981,  1610DE (BCH001312) 00888707002967,  1610DE (BCH001313) 00888707002967,  1610DE (BCH001314) 00888707002967,  1610DE (BCH001315) 00888707002967,  1610DE (BCH001316) 00888707002967,  1610DE (BCH001317) 00888707002967,  1610DE (BCH001318) 00888707002967,  1610DE (BCH001319) 00888707002967,  1610DE (BCH001320) 00888707002967,  1610DE (BCH001321) 00888707002967,  1610DE (BCH001322) 00888707002967,  1610DE (BCH001323) 00888707002967,  1610DE (BCH001324) 00888707002967,  1610DE (BCH001325) 00888707002967,  1610GB (BCH001326) 00888707002974,  1610DE (BCH001327) 00888707002967,  1610DE (BCH001328) 00888707002967,  1610DE (BCH001329) 00888707002967,  1610DE (BCH001330) 00888707002967,  1610DE (BCH001331) 00888707002967,  1610CA (BCH001332) 00888707003001,  1610DE (BCH001333) 00888707002967,  1610DE (BCH001334) 00888707002967,  1610DE (BCH001335) 00888707002967,  1610DE (BCH001336) 00888707002967,  1610DE (BCH001337) 00888707002967,  1610IT (BCH001338) 00888707002981,  1610CH (BCH001339) 00888707002196,  1610GB (BCH001340) 00888707002974,  1610DE (BCH001341) 00888707002967,  1610CA (BCH001342) 00888707003001,  1610DE (BCH001343) 00888707002967,  1610DE (BCH001344) 00888707002967,  1610DE (BCH001345) 00888707002967,  1610DE (BCH001346) 00888707002967,  1610GB (BCH001347) 00888707002974,  1610DE (BCH001348) 00888707002967,  1610CA (BCH001349) 00888707003001,  1610DE (BCH001350) 00888707002967,  1610GB (BCH001351) 00888707002974,  1610GB (BCH001352) 00888707002974,  1610IT (BCH001353) 00888707002981,  1610DE (BCH001354) 00888707002967,  1610DE (BCH001355) 00888707002967,  1610GB (BCH001356) 00888707002974,  1610DE (BCH001357) 00888707002967,  1610DE (BCH001358) 00888707002967,  1610DE (BCH001359) 00888707002967,  1610DE (BCH001360) 00888707002967,  1610IT (BCH001361) 00888707002981,  1610DE (BCH001362) 00888707002967,  1610DE (BCH001363) 00888707002967,  1610DE (BCH001364) 00888707002967,  1610DE (BCH001365) 00888707002967,  1610CH (BCH001366) 00888707002196,  1610DE (BCH001367) 00888707002967,  1610AU (BCH001368) 00888707002165,  1610GB (BCH001369) 00888707002974,  1610GB (BCH001370) 00888707002974,  1610GB (BCH001371) 00888707002974,  1610DE (BCH001372) 00888707002967,  1610DE (BCH001373) 00888707002967,  1610CA (BCH001374) 00888707003001,  1610DE (BCH001375) 00888707002967,  1610DE (BCH001376) 00888707002967,  1610DE (BCH001377) 00888707002967,  1610GB (BCH001378) 00888707002974,  1610GB (BCH001379) 00888707002974,  1610GB (BCH001380) 00888707002974,  1610GB (BCH001381) 00888707002974,  1610GB (BCH001382) 00888707002974,  1610AU (BCH001383) 00888707002165,  1610DE (BCH001384) 00888707002967,  1610DE (BCH001385) 00888707002967,  1610DE (BCH001386) 00888707002967,  1610IT (BCH001387) 00888707002981,  1610DE (BCH001388) 00888707002967,  1610DE (BCH001389) 00888707002967,  1610DE (BCH001391) 00888707002967,  1610DE (BCH001392) 00888707002967,  1610DE (BCH001393) 00888707002967,  1610DE (BCH001394) 00888707002967,  1610DE (BCH001395) 00888707002967,  1610AU (BCH001396) 00888707002165,  1610GB (BCH001397) 00888707002974,  1610GB (BCH001398) 00888707002974,  1610DE (BCH001399) 00888707002967,  1610DE (BCH001400) 00888707002967,  1610DE (BCH001401) 00888707002967,  1610DE (BCH001402) 00888707002967,  1610CA (BCH001403) 00888707003001,  1610CA (BCH001404) 00888707003001,  1610DE (BCH001405) 00888707002967,  1610DE (BCH001406) 00888707002967,  1610DE (BCH001407) 00888707002967,  1610GB (BCH001408) 00888707002974,  1610GB (BCH001409) 00888707002974,  1610GB (BCH001410) 00888707002974,  1610DE (BCH001411) 00888707002967,  1610DE (BCH001412) 00888707002967,  1610DE (BCH001413) 00888707002967,  1610AU (BCH001414) 00888707002165,  1610AU (BCH001415) 00888707002165,  1610DE (BCH001416) 00888707002967,  1610DE (BCH001417) 00888707002967,  1610DE (BCH001418) 00888707002967,  1610GB (BCH001419) 00888707002974,  1610DE (BCH001420) 00888707002967,  1610DE (BCH001421) 00888707002967,  1610DE (BCH001422) 00888707002967,  1610DE (BCH001423) 00888707002967,  1610DE (BCH001424) 00888707002967,  1610DE (BCH001425) 00888707002967,  1610DE (BCH001426) 00888707002967,  1610DE (BCH001427) 00888707002967,  1610DE (BCH001428) 00888707002967,  1610GB (BCH001429) 00888707002974,  1610DE (BCH001430) 00888707002967,  1610GB (BCH001431) 00888707002974,  1610IN (BCH001433) 00888707002998,  1610DE (BCH001434) 00888707002967,  1610DE (BCH001435) 00888707002967,  1610DE (BCH001436) 00888707002967,  1610DE (BCH001437) 00888707002967,  1610IN (BCH001438) 00888707002998,  1610AU (BCH001439) 00888707002165,  1610CH (BCH001440) 00888707002196,  1610AU (BCH001441) 00888707002165,  1610CA (BCH001442) 00888707003001,  1610DE (BCH001443) 00888707002967,  1610DE (BCH001444) 00888707002967,  1610DE (BCH001445) 00888707002967,  1610IN (BCH001446) 00888707002998,  1610DE (BCH001447) 00888707002967,  1610DE (BCH001448) 00888707002967,  1610DE (BCH001449) 00888707002967,  1610IN (BCH001450) 00888707002998,  1610CH (BCH001451) 00888707002196,  1610CH (BCH001452) 00888707002196,  1610DE (BCH001453) 00888707002967,  1610AU (BCH001454) 00888707002165,  1610AU (BCH001455) 00888707002165,  1610DE (BCH001456) 00888707002967,  1610AU (BCH001457) 00888707002165,  1610CA (BCH001458) 00888707003001,  1610GB (BCH001459) 00888707002974,  1610IN (BCH001460) 00888707002998,  1610DE (BCH001461) 00888707002967,  1610AU (BCH001462) 00888707002165,  1610DE (BCH001463) 00888707002967,  1610DE (BCH001464) 00888707002967,  1610DE (BCH001465) 00888707002967,  1610DE (BCH001466) 00888707002967,  1610DE (BCH001467) 00888707002967,  1610GB (BCH001468) 00888707002974,  1610GB (BCH001469) 00888707002974,  1610DE (BCH001473) 00888707002967,  1610DE (BCH001474) 00888707002967,  1610DE (BCH001475) 00888707002967,  1610CA (BCH001476) 00888707003001,  1610GB (BCH001477) 00888707002974,  1610DE (BCH001478) 00888707002967,  1610DE (BCH001479) 00888707002967,  1610DE (BCH001480) 00888707002967,  1610IT (BCH001481) 00888707002981,  1610DE (BCH001482) 00888707002967,  1610DE (BCH001483) 00888707002967,  1610IT (BCH001484) 00888707002981,  1610CA (BCH001485) 00888707003001,  1610IT (BCH001486) 00888707002981,  1610IT (BCH001487) 00888707002981,  1610DE (BCH001488) 00888707002967,  1610DE (BCH001489) 00888707002967,  1610DE (BCH001490) 00888707002967,  1610DE (BCH001491) 00888707002967,  1610GB (BCH001492) 00888707002974,  1610DE (BCH001493) 00888707002967,  1610DE (BCH001494) 00888707002967,  1610DE (BCH001495) 00888707002967,  1610DE (BCH001496) 00888707002967,  1610DE (BCH001497) 00888707002967,  1610DE (BCH001498) 00888707002967,  1610CA (BCH001499) 00888707003001,  1610CA (BCH001500) 00888707003001,  1610IT (BCH001501) 00888707002981,  1610IT (BCH001502) 00888707002981,  1610IT (BCH001503) 00888707002981,  1610DE (BCH001504) 00888707002967,  1610DE (BCH001505) 00888707002967,  1610DE (BCH001506) 00888707002967,  1610DE (BCH001507) 00888707002967,  1610IT (BCH001508) 00888707002981,  1610DE (BCH001509) 00888707002967,  1610CA (BCH001510) 00888707003001,  1610IT (BCH001511) 00888707002981,  1610DE (BCH001512) 00888707002967,  1610DE (BCH001513) 00888707002967,  1610CA (BCH001514) 00888707003001,  1610IN (BCH001515) 00888707002998,  1610DE (BCH001516) 00888707002967,  1610DE (BCH001517) 00888707002967,  1610DE (BCH001518) 00888707002967,  1610DE (BCH001519) 00888707002967,  1610DE (BCH001520) 00888707002967,  1610GB (BCH001521) 00888707002974,  1610CA (BCH001522) 00888707003001,  1610GB (BCH001523) 00888707002974,  1610DE (BCH001524) 00888707002967,  1610DE (BCH001525) 00888707002967,  1610DE (BCH001526) 00888707002967,  1610AU (BCH001527) 00888707002165,  1610IN (BCH001528) 00888707002998,  1610GB (BCH001529) 00888707002974,  1610GB (BCH001530) 00888707002974,  1610AU (BCH001531) 00888707002165,  1610AU (BCH001532) 00888707002165,  1610DE (BCH001533) 00888707002967,  1610CA (BCH001534) 00888707003001,  1610DE (BCH001535) 00888707002967,  1610DE (BCH001536) 00888707002967,  1610CA (BCH001537) 00888707003001,  1610IN (BCH001538) 00888707002998,  1610DE (BCH001539) 00888707002967,  1610DE (BCH001540) 00888707002967,  1610DE (BCH001541) 00888707002967,  1610DE (BCH001542) 00888707002967,  1610DE (BCH001543) 00888707002967,  1610DE (BCH001544) 00888707002967,  1610DE (BCH001545) 00888707002967,  1610CA (BCH001546) 00888707003001,  1610DE (BCH001547) 00888707002967,  1610DE (BCH001548) 00888707002967,  1610DE (BCH001549) 00888707002967,  1610DE (BCH001550) 00888707002967,  1610DE (BCH001551) 00888707002967,  1610DE (BCH001552) 00888707002967,  1610DE (BCH001553) 00888707002967,  1610DE (BCH001554) 00888707002967,  1610DE (BCH001555) 00888707002967,  1610DE (BCH001556) 00888707002967,  1610AU (BCH001557) 00888707002165,  1610DE (BCH001558) 00888707002967,  1610DE (BCH001559) 00888707002967,  1610DE (BCH001560) 00888707002967,  1610DE (BCH001561) 00888707002967,  1610DE (BCH001562) 00888707002967,  1610DE (BCH001563) 00888707002967,  1610DE (BCH001564) 00888707002967,  1610DE (BCH001565) 00888707002967,  1610DE (BCH001566) 00888707002967,  1610CA (BCH001567) 00888707003001,  1610DE (BCH001568) 00888707002967,  1610GB (BCH001569) 00888707002974,  1610DE (BCH001570) 00888707002967,  1610DE (BCH001571) 00888707002967,  1610CA (BCH001572) 00888707003001,  1610DE (BCH001573) 00888707002967,  1610DE (BCH001574) 00888707002967,  1610DE (BCH001575) 00888707002967,  1610DE (BCH001576) 00888707002967,  1610DE (BCH001577) 00888707002967,  1610GB (BCH001578) 00888707002974,  1610DE (BCH001579) 00888707002967,  1610DE (BCH001580) 00888707002967,  1610GB (BCH001581) 00888707002974,  1610DE (BCH001582) 00888707002967,  1610DE (BCH001583) 00888707002967,  1610IT (BCH001584) 00888707002981,  1610GB (BCH001585) 00888707002974,  1610DE (BCH001586) 00888707002967,  1610DE (BCH001587) 00888707002967,  1610DE (BCH001588) 00888707002967,  1610DE (BCH001589) 00888707002967,  1610DE (BCH001590) 00888707002967,  1610DE (BCH001591) 00888707002967,  1610DE (BCH001592) 00888707002967,  1610DE (BCH001593) 00888707002967,  1610DE (BCH001594) 00888707002967,  1610DE (BCH001595) 00888707002967,  1610DE (BCH001596) 00888707002967,  1610CA (BCH001597) 00888707003001,  1610CA (BCH001598) 00888707003001,  1610DE (BCH001599) 00888707002967,  1610DE (BCH001600) 00888707002967,  1610DE (BCH001601) 00888707002967,  1610IN (BCH001602) 00888707002998,  1610CA (BCH001603) 00888707003001,  1610CA (BCH001604) 00888707003001,  1610DE (BCH001605) 00888707002967,  1610DE (BCH001606) 00888707002967,  1610DE (BCH001607) 00888707002967,  1610DE (BCH001608) 00888707002967,  1610JP (BCH001609) 00888707002257,  1610JP (BCH001610) 00888707002257,  1610JP (BCH001611) 00888707002257,  1610DE (BCH001612) 00888707002967,  1610DE (BCH001613) 00888707002967,  1610JP (BCH001614) 00888707002257,  1610JP (BCH001615) 00888707002257,  1610AU (BCH001616) 00888707002165,  1610AU (BCH001617) 00888707002165,  1610DE (BCH001618) 00888707002967,  1610DE (BCH001619) 00888707002967,  1610JP (BCH001620) 00888707002257,  1610DE (BCH001621) 00888707002967,  1610DE (BCH001622) 00888707002967,  1610IT (BCH001623) 00888707002981,  1610JP (BCH001624) 00888707002257,  1610DE (BCH001625) 00888707002967,  1610JP (BCH001626) 00888707002257,  1610IN (BCH001627) 00888707002998,  1610DE (BCH001628) 00888707002967,  1610DE (BCH001629) 00888707002967,  1610DE (BCH001630) 00888707002967,  1610DE (BCH001631) 00888707002967,  1610DE (BCH001632) 00888707002967,  1610DE (BCH001633) 00888707002967,  1610DE (BCH001634) 00888707002967,  1610JP (BCH001635) 00888707002257,  1610IT (BCH001636) 00888707002981,  1610DE (BCH001637) 00888707002967,  1610DE (BCH001638) 00888707002967,  1610DE (BCH001639) 00888707002967,  1610DE (BCH001640) 00888707002967,  1610DE (BCH001641) 00888707002967,  1610DE (BCH001642) 00888707002967,  1610DE (BCH001643) 00888707002967,  1610GB (BCH001644) 00888707002974,  1610DE (BCH001645) 00888707002967,  1610DE (BCH001646) 00888707002967,  1610DE (BCH001647) 00888707002967,  1610DE (BCH001648) 00888707002967,  1610IN (BCH001649) 00888707002998,  1610DE (BCH001650) 00888707002967,  1610IT (BCH001651) 00888707002981,  1610DE (BCH001652) 00888707002967,  1610IT (BCH001653) 00888707002981,  1610DE (BCH001654) 00888707002967,  1610IT (BCH001655) 00888707002981,  1610IN (BCH001656) 00888707002998,  1610DE (BCH001657) 00888707002967,  1610DE (BCH001658) 00888707002967,  1610DE (BCH001659) 00888707002967,  1610IN (BCH001660) 00888707002998,  1610DE (BCH001661) 00888707002967,  1610DE (BCH001662) 00888707002967,  1610IT (BCH001663) 00888707002981,  1610DE (BCH001664) 00888707002967,  1610DE (BCH001665) 00888707002967,  1610DE (BCH001666) 00888707002967,  1610DE (BCH001667) 00888707002967,  1610DE (BCH001668) 00888707002967,  1610DE (BCH001669) 00888707002967,  1610DE (BCH001670) 00888707002967,  1610CA (BCH001671) 00888707003001,  1610DE (BCH001672) 00888707002967,  1610DE (BCH001673) 00888707002967,  1610CH (BCH001674) 00888707002196,  1610CH (BCH001675) 00888707002196,  1610JP (BCH001676) 00888707002257,  1610JP (BCH001677) 00888707002257,  1610DE (BCH001678) 00888707002967,  1610DE (BCH001679) 00888707002967,  1610DE (BCH001680) 00888707002967,  1610IT (BCH001681) 00888707002981,  1610CA (BCH001682) 00888707003001,  1610GB (BCH001683) 00888707002974,  1610DE (BCH001684) 00888707002967,  1610DE (BCH001685) 00888707002967,  1610DE (BCH001686) 00888707002967,  1610DE (BCH001687) 00888707002967,  1610GB (BCH001688) 00888707002974,  1610DE (BCH001689) 00888707002967,  1610DE (BCH001690) 00888707002967,  1610DE (BCH001691) 00888707002967,  1610DE (BCH001692) 00888707002967,  1610DE (BCH001693) 00888707002967,  1610DE (BCH001694) 00888707002967,  1610IT (BCH001695) 00888707002981,  1610DE (BCH001696) 00888707002967,  1610DE (BCH001697) 00888707002967,  1610DE (BCH001698) 00888707002967,  1610JP (BCH001699) 00888707002257,  1610JP (BCH001700) 00888707002257,  1610DE (BCH001701) 00888707002967,  1610JP (BCH001702) 00888707002257,  1610JP (BCH001703) 00888707002257,  1610IT (BCH001704) 00888707002981,  1610DE (BCH001705) 00888707002967,  1610DE (BCH001706) 00888707002967,  1610DE (BCH001707) 00888707002967,  1610DE (BCH001708) 00888707002967,  1610DE (BCH001709) 00888707002967,  1610DE (BCH001710) 00888707002967,  1610DE (BCH001711) 00888707002967,  1610IN (BCH001712) 00888707002998,  1610DE (BCH001713) 00888707002967,  1610IN (BCH001714) 00888707002998,  1610DE (BCH001715) 00888707002967,  1610DE (BCH001716) 00888707002967,  1610GB (BCH001717) 00888707002974,  1610DE (BCH001718) 00888707002967,  1610DE (BCH001719) 00888707002967,  1610DE (BCH001720) 00888707002967,  1610DE (BCH001721) 00888707002967,  1610CH (BCH001722) 00888707002196,  1610CH (BCH001723) 00888707002196,  1610DE (BCH001724) 00888707002967,  1610DE (BCH001725) 00888707002967,  1610DE (BCH001726) 00888707002967,  1610DE (BCH001727) 00888707002967,  1610GB (BCH001728) 00888707002974,  1610DE (BCH001729) 00888707002967,  1610GB (BCH001730) 00888707002974,  1610GB (BCH001731) 00888707002974,  1610IN (BCH001732) 00888707002998,  1610DE (BCH001733) 00888707002967,  1610DE (BCH001734) 00888707002967,  1610GB (BCH001735) 00888707002974,  1610DE (BCH001736) 00888707002967,  1610DE (BCH001737) 00888707002967,  1610CA (BCH001738) 00888707003001,  1610JP (BCH001739) 00888707002257,  1610JP (BCH001740) 00888707002257,  1610JP (BCH001741) 00888707002257,  1610IN (BCH001742) 00888707002998,  1610DE (BCH001743) 00888707002967,  1610DE (BCH001744) 00888707002967,  1610DE (BCH001745) 00888707002967,  1610DE (BCH001746) 00888707002967,  1610JP (BCH001747) 00888707002257,  1610DE (BCH001748) 00888707002967,  1610DE (BCH001749) 00888707002967,  1610DE (BCH001750) 00888707002967,  1610DE (BCH001751) 00888707002967,  1610DE (BCH001752) 00888707002967,  1610DE (BCH001753) 00888707002967,  1610DE (BCH001754) 00888707002967,  1610DE (BCH001755) 00888707002967,  1610DE (BCH001756) 00888707002967,  1610DE (BCH001757) 00888707002967,  1610DE (BCH001758) 00888707002967,  1610DE (BCH001759) 00888707002967,  1610CA (BCH001760) 00888707003001,  1610CA (BCH001761) 00888707003001,  1610CA (BCH001762) 00888707003001,  1610DE (BCH001763) 00888707002967,  1610DE (BCH001764) 00888707002967,  1610DE (BCH001765) 00888707002967,  1610DE (BCH001766) 00888707002967,  1610DE (BCH001767) 00888707002967,  1610DE (BCH001768) 00888707002967,  1610DE (BCH001769) 00888707002967,  1610DE (BCH001770) 00888707002967,  1610DE (BCH001771) 00888707002967,  1610DE (BCH001772) 00888707002967,  1610DE (BCH001773) 00888707002967,  1610DE (BCH001774) 00888707002967,  1610DE (BCH001775) 00888707002967,  1610DE (BCH001776) 00888707002967,  1610DE (BCH001777) 00888707002967,  1610DE (BCH001778) 00888707002967,  1610DE (BCH001779) 00888707002967,  1610DE (BCH001780) 00888707002967,  1610DE (BCH001781) 00888707002967,  1610DE (BCH001782) 00888707002967,  1610DE (BCH001783) 00888707002967,  1610DE (BCH001784) 00888707002967,  1610DE (BCH001785) 00888707002967,  1610DE (BCH001786) 00888707002967,  1610DE (BCH001787) 00888707002967,  1610DE (BCH001788) 00888707002967,  1610DE (BCH001789) 00888707002967,  1610DE (BCH001790) 00888707002967,  1610DE (BCH001791) 00888707002967,  1610DE (BCH001792) 00888707002967,  1610DE (BCH001793) 00888707002967,  1610DE (BCH001794) 00888707002967,  1610DE (BCH001795) 00888707002967,  1610DE (BCH001796) 00888707002967,  1610DE (BCH001797) 00888707002967,  1610DE (BCH001798) 00888707002967,  1610DE (BCH001799) 00888707002967,  1610DE (BCH001800) 00888707002967,  1610DE (BCH001801) 00888707002967,  1610DE (BCH001802) 00888707002967,  1610DE (BCH001803) 00888707002967,  1610DE (BCH001804) 00888707002967,  1610DE (BCH001805) 00888707002967,  1610DE (BCH001806) 00888707002967,  1610DE (BCH001807) 00888707002967,  1610DE (BCH001808) 00888707002967,  1610DE (BCH001809) 00888707002967,  1610IT (BCH001810) 00888707002981,  1610DE (BCH001811) 00888707002967,  1610DE (BCH001812) 00888707002967,  1610DE (BCH001813) 00888707002967,  1610DE (BCH001814) 00888707002967,  1610DE (BCH001815) 00888707002967,  1610DE (BCH001816) 00888707002967,  1610DE (BCH001817) 00888707002967,  1610DE (BCH001818) 00888707002967,  1610DE (BCH001819) 00888707002967,  1610DE (BCH001820) 00888707002967,  1610DE (BCH001821) 00888707002967,  1610DE (BCH001822) 00888707002967,  1610DE (BCH001823) 00888707002967,  1610DE (BCH001824) 00888707002967,  1610DE (BCH001825) 00888707002967,  1610DE (BCH001826) 00888707002967,  1610DE (BCH001827) 00888707002967,  1610DE (BCH001828) 00888707002967,  1610DE (BCH001829) 00888707002967,  1610DE (BCH001830) 00888707002967,  1610DE (BCH001831) 00888707002967,  1610DE (BCH001832) 00888707002967,  1610DE (BCH001833) 00888707002967,  1610DE (BCH001834) 00888707002967,  1610DE (BCH001835) 00888707002967,  1610DE (BCH001836) 00888707002967,  1610GB (BCH001837) 00888707002974,  1610GB (BCH001838) 00888707002974,  1610DE (BCH001839) 00888707002967,  1610DE (BCH001840) 00888707002967,  1610DE (BCH001841) 00888707002967,  1610DE (BCH001842) 00888707002967,  1610DE (BCH001843) 00888707002967,  1610GB (BCH001844) 00888707002974,  1610DE (BCH001845) 00888707002967,  1610GB (BCH001846) 00888707002974,  1610DE (BCH001847) 00888707002967,  1610DE (BCH001848) 00888707002967,  1610DE (BCH001849) 00888707002967,  1610DE (BCH001850) 00888707002967,  1610DE (BCH001851) 00888707002967,  1610DE (BCH001852) 00888707002967,  1610DE (BCH001853) 00888707002967,  1610DE (BCH001854) 00888707002967,  1610DE (BCH001855) 00888707002967,  1610DE (BCH001856) 00888707002967,  1610GB (BCH001857) 00888707002974,  1610IT (BCH001858) 00888707002981,  1610DE (BCH001859) 00888707002967,  1610DE (BCH001860) 00888707002967,  1610DE (BCH001861) 00888707002967,  1610GB (BCH001862) 00888707002219,  1610DE (BCH001863) 00888707002967,  1610GB (BCH001864) 00888707002974,  1610GB (BCH001865) 00888707002219,  1610GB (BCH001866) 00888707002974,  1610DE (BCH001867) 00888707002967,  1610GB (BCH001868) 00888707002974,  1610GB (BCH001869) 00888707002974,  1610DE (BCH001870) 00888707002967,  1610DE (BCH001871) 00888707002967,  1610DE (BCH001872) 00888707002967,  1610DE (BCH001873) 00888707002967,  1610DE (BCH001874) 00888707002967,  1610GB (BCH001875) 00888707002974,  1610DE (BCH001876) 00888707002967,  1610DE (BCH001877) 00888707002967,  1610DE (BCH001878) 00888707002967,  1610GB (BCH001879) 00888707002974,  1610DE (BCH001880) 00888707002967,  1610DE (BCH001881) 00888707002967,  1610GB (BCH001882) 00888707002974,  1610AU (BCH001883) 00888707002165,  1610IT (BCH001884) 00888707002981,  1610DE (BCH001885) 00888707002967,  1610DE (BCH001886) 00888707002967,  1610DE (BCH001887) 00888707002967,  1610DE (BCH001888) 00888707002967,  1610CA (BCH001889) 00888707003001,  1610CA (BCH001890) 00888707003001,  1610DE (BCH001891) 00888707002967,  1610DE (BCH001892) 00888707002967,  1610DE (BCH001893) 00888707002967,  1610DE (BCH001894) 00888707002967,  1610DE (BCH001895) 00888707002967,  1610DE (BCH001896) 00888707002967,  1610DE (BCH001897) 00888707002967,  1610DE (BCH001898) 00888707002967,  1610DE (BCH001899) 00888707002967,  1610DE (BCH001900) 00888707002967,  1610DE (BCH001901) 00888707002967,  1610DE (BCH001902) 00888707002967,  1610DE (BCH001903) 00888707002967,  1610DE (BCH001904) 00888707002967,  1610JP (BCH001905) 00888707002257,  1610JP (BCH001906) 00888707002257,  1610DE (BCH001907) 00888707002967,  1610DE (BCH001908) 00888707002967,  1610DE (BCH001909) 00888707002967,  1610CA (BCH001910) 00888707003001,  1610CA (BCH001911) 00888707003001,  1610DE (BCH001912) 00888707002967,  1610DE (BCH001913) 00888707002967,  1610DE (BCH001914) 00888707002967,  1610DE (BCH001915) 00888707002967,  1610DE (BCH001916) 00888707002967,  1610DE (BCH001917) 00888707002967,  1610DE (BCH001918) 00888707002967,  1610DE (BCH001919) 00888707002967,  1610DE (BCH001920) 00888707002967,  1610DE (BCH001921) 00888707002967,  1610DE (BCH001922) 00888707002967,  1610DE (BCH001923) 00888707002967,  1610DE (BCH001924) 00888707002967,  1610DE (BCH001925) 00888707002967,  1610DE (BCH001926) 00888707002967,  1610DE (BCH001927) 00888707002967,  1610IT (BCH001928) 00888707002981,  1610IT (BCH001929) 00888707002981,  1610DE (BCH001930) 00888707002967,  1610DE (BCH001931) 00888707002967,  1610DE (BCH001932) 00888707002967,  1610DE (BCH001933) 00888707002967,  1610DE (BCH001934) 00888707002967,  1610DE (BCH001935) 00888707002967,  1610DE (BCH001936) 00888707002967,  1610JP (BCH001937) 00888707002257,  1610DE (BCH001938) 00888707002967,  1610DE (BCH001939) 00888707002967,  1610DE (BCH001940) 00888707002967,  1610AU (BCH001941) 00888707002165,  1610AU (BCH001942) 00888707002165,  1610DE (BCH001943) 00888707002967,  1610DE (BCH001944) 00888707002967,  1610DE (BCH001945) 00888707002967,  1610DE (BCH001946) 00888707002967,  1610DE (BCH001947) 00888707002967,  1610DE (BCH001948) 00888707002967,  1610DE (BCH001949) 00888707002967,  1610IT (BCH001950) 00888707002981,  1610DE (BCH001951) 00888707002967,  1610DE (BCH001952) 00888707002967,  1610JP (BCH001953) 00888707002257,  1610DE (BCH001954) 00888707002967,  1610DE (BCH001955) 00888707002967,  1610DE (BCH001956) 00888707002967,  1610DE (BCH001957) 00888707002967,  1610AU (BCH001958) 00888707002165,  1610AU (BCH001959) 00888707002165,  1610DE (BCH001960) 00888707002967,  1610DE (BCH001961) 00888707002967,  1610DE (BCH001962) 00888707002967,  1610JP (BCH001963) 00888707002257,  1610DE (BCH001964) 00888707002967,  1610DE (BCH001965) 00888707002967,  1610AU (BCH001966) 00888707002165,  1610DE (BCH001967) 00888707002967,  1610DE (BCH001968) 00888707002967,  1610DE (BCH001969) 00888707002967,  1610DE (BCH001970) 00888707002967,  1610DE (BCH001971) 00888707002967,  1610DE (BCH001972) 00888707002967,  1610JP (BCH001973) 00888707002257,  1610DE (BCH001974) 00888707002967,  1610DE (BCH001975) 00888707002967,  1610DE (BCH001976) 00888707002967,  1610JP (BCH001977) 00888707002257,  1610DE (BCH001978) 00888707002967,  1610DE (BCH001979) 00888707002967,  1610AU (BCH001980) 00888707002165,  1610AU (BCH001981) 00888707002165,  1610DE (BCH001982) 00888707002967,  1610DE (BCH001983) 00888707002967,  1610AU (BCH001984) 00888707002165,  1610DE (BCH001985) 00888707002967,  1610DE (BCH001986) 00888707002967,  1610IT (BCH001987) 00888707002981,  1610AU (BCH001988) 00888707002165,  1610IT (BCH001989) 00888707002981,  1610AU (BCH001990) 00888707002165,  1610AU (BCH001991) 00888707002165,  1610IL (BCH001992) 00888707002226,  1610AU (BCH001993) 00888707002165,  1610DE (BCH001994) 00888707002967,  1610DE (BCH001995) 00888707002967,  1610DE (BCH001996) 00888707002967,  1610DE (BCH001997) 00888707002967,  1610IL (BCH001998) 00888707002226,  1610IL (BCH001999) 00888707002226,  1610DE (BCH002000) 00888707002967,  1610DE (BCH002001) 00888707002967,  1610DE (BCH002002) 00888707002967,  1610DE (BCH002003) 00888707002967,  1610IL (BCH002004) 00888707002226,  1610DE (BCH002005) 00888707002967,  1610DE (BCH002006) 00888707005784,  1610DE (BCH002007) 00888707005784,  1610DE (BCH002008) 00888707005784,  1610DE (BCH002009) 00888707005784,  1610DE (BCH002010) 00888707005784,  1610AU (BCH002011) 00888707002165,  1610DE (BCH002012) 00888707002967,  1610DE (BCH002013) 00888707005784,  1610DE (BCH002014) 00888707002967,  1610DE (BCH002015) 00888707005784,  1610DE (BCH002016) 00888707005784,  1610DE (BCH002017) 00888707002967,  1610DE (BCH002018) 00888707005784,  1610DE (BCH002019) 00888707005784,  1610DE (BCH002020) 00888707002967,  1610DE (BCH002021) 00888707002967,  1610AU (BCH002022) 00888707002165,  1610DE (BCH002023) 00888707005784,  1610DE (BCH002024) 00888707005784,  1610DE (BCH002025) 00888707005784,  1610DE (BCH002026) 00888707005784,  1610GB (BCH002027) 00888707005791,  1610IT (BCH002028) 00888707005821,  1610IT (BCH002029) 00888707005821,  1610DE (BCH002030) 00888707005784,  1610DE (BCH002031) 00888707005784,  1610DE (BCH002032) 00888707005784,  1610DE (BCH002033) 00888707005784,  1610DE (BCH002034) 00888707005784,  1610DE (BCH002035) 00888707005784,  1610DE (BCH002036) 00888707005784,  1610DE (BCH002037) 00888707005784,  1610DE (BCH002038) 00888707005784,  1610DE (BCH002039) 00888707005784,  1610DE (BCH002040) 00888707005784,  1610DE (BCH002041) 00888707005784,  1610DE (BCH002042) 00888707005784,  1610DE (BCH002043) 00888707005784,  1610GB (BCH002046) 00888707005791,  1610IT (BCH002047) 00888707005821,  1610AU (BCH002048) 00888707002165,  1610AU (BCH002049) 00888707002165,  1610DE (BCH002050) 00888707005784,  1610DE (BCH002051) 00888707005784,  1610DE (BCH002052) 00888707005784,  1610DE (BCH002053) 00888707005784,  1610DE (BCH002054) 00888707005784,  1610DE (BCH002055) 00888707005784,  1610DE (BCH002056) 00888707005784,  1610DE (BCH002057) 00888707005784,  1610DE (BCH002058) 00888707005784,  1610DE (BCH002059) 00888707005784,  1610DE (BCH002060) 00888707005784,  1610DE (BCH002061) 00888707005784,  1610GB (BCH002062) 00888707005791,  1610AU (BCH002063) 00888707002165,  1610GB (BCH002064) 00888707005791,  1610CA (BCH002065) 00888707005760,  1610DE (BCH002066) 00888707005784,  1610DE (BCH002067) 00888707005784,  1610DE (BCH002068) 00888707005784,  1610DE (BCH002069) 00888707005784,  1610DE (BCH002070) 00888707005784,  1610DE (BCH002071) 00888707005784,  1610DE (BCH002072) 00888707005784,  1610GB (BCH002073) 00888707005791,  1610IT (BCH002074) 00888707005821,  1610DE (BCH002075) 00888707005784,  1610IT (BCH002076) 00888707005821,  1610DE (BCH002077) 00888707005784,  1610DE (BCH002078) 00888707005784,  1610DE (BCH002079) 00888707005784,  1610DE (BCH002080) 00888707005784,  1610GB (BCH002081) 00888707005791,  1610GB (BCH002082) 00888707005791,  1610CA (BCH002083) 00888707005760,  1610IT (BCH002084) 00888707005821,  1610DE (BCH002085) 00888707005784,  1610IT (BCH002086) 00888707005821,  1610DE (BCH002090) 00888707005784,  1610DE (BCH002091) 00888707005784,  1610DE (BCH002092) 00888707005784,  1610DE (BCH002093) 00888707005784,  1610DE (BCH002094) 00888707005784,  1610DE (BCH002095) 00888707005784,  1610DE (BCH002096) 00888707005784,  1610DE (BCH002097) 00888707005784,  1610GB (BCH002098) 00888707005791,  1610CA (BCH002099) 00888707005760,  1610DE (BCH002100) 00888707005784,  1610DE (BCH002101) 00888707005784,  1610GB (BCH002102) 00888707005791,  1610DE (BCH002103) 00888707005784,  1610DE (BCH002104) 00888707005784,  1610DE (BCH002105) 00888707005784,  1610DE (BCH002106) 00888707005784,  1610DE (BCH002107) 00888707005784,  1610DE (BCH002108) 00888707005784,  1610CA (BCH002109) 00888707005760,  1610DE (BCH002110) 00888707005784,  1610IT (BCH002111) 00888707005821,  1610GB (BCH002112) 00888707005791,  1610IT (BCH002113) 00888707005821,  1610GB (BCH002114) 00888707005791,  1610GB (BCH002115) 00888707005791,  1610CA (BCH002116) 00888707005760,  1610DE (BCH002117) 00888707005784,  1610GB (BCH002118) 00888707005791,  1610DE (BCH002119) 00888707005784,  1610DE (BCH002120) 00888707005784,  1610CA (BCH002121) 00888707005760,  1610DE (BCH002122) 00888707005784,  1610DE (BCH002123) 00888707005784,  1610DE (BCH002125) 00888707005784,  1610DE (BCH002126) 00888707005784,  1610DE (BCH002127) 00888707005784,  1610DE (BCH002128) 00888707005784,  1610DE (BCH002129) 00888707005784,  1610DE (BCH002130) 00888707005784,  1610JP (BCH002132) 00888707002257,  1610JP (BCH002133) 00888707002257,  1610JP (BCH002134) 00888707002257,  1610JP (BCH002135) 00888707002257,  1610JP (BCH002136) 00888707002257,  1610DE (BCH002137) 00888707005784,  1610DE (BCH002138) 00888707005784,  1610DE (BCH002139) 00888707005784,  1610DE (BCH002140) 00888707005784,  1610DE (BCH002141) 00888707005784,  1610DE (BCH002142) 00888707005784,  1610DE (BCH002143) 00888707005784,  1610DE (BCH002144) 00888707005784,  1610DE (BCH002145) 00888707002967,  1610DE (BCH002146) 00888707005784,  1610DE (BCH002147) 00888707005784,  1610DE (BCH002148) 00888707005784,  1610DE (BCH002149) 00888707005784,  1610DE (BCH002150) 00888707005784,  1610DE (BCH002151) 00888707005784,  1610DE (BCH002152) 00888707005784,  1610DE (BCH002153) 00888707005784,  1610DE (BCH002154) 00888707005784,  1610DE (BCH002155) 00888707005784,  1610DE (BCH002156) 00888707005784,  1610DE (BCH002157) 00888707005784,  1610DE (BCH002158) 00888707005784,  1610DE (BCH002159) 00888707005784,  1610DE (BCH002160) 00888707005784,  1610DE (BCH002161) 00888707005784,  1610DE (BCH002162) 00888707005784,  1610DE (BCH002163) 00888707005784,  1610DE (BCH002164) 00888707005784,  1610JP (BCH002165) 00888707002257,  1610DE (BCH002166) 00888707005784,  1610DE (BCH002167) 00888707005784,  1610DE (BCH002168) 00888707005784,  1610DE (BCH002169) 00888707005784,  1610DE (BCH002170) 00888707005784,  1610DE (BCH002171) 00888707005784,  1610DE (BCH002172) 00888707005784,  1610DE (BCH002173) 00888707005784,  1610DE (BCH002174) 00888707002967,  1610DE (BCH002175) 00888707005784,  1610DE (BCH002176) 00888707002967,  1610DE (BCH002177) 00888707005784,  1610DE (BCH002178) 00888707005784,  1610DE (BCH002179) 00888707002967,  1610DE (BCH002180) 00888707005784,  1610DE (BCH002181) 00888707005784,  1610DE (BCH002182) 00888707005784,  1610DE (BCH002183) 00888707005784,  1610DE (BCH002184) 00888707005784,  1610DE (BCH002185) 00888707005784,  1610DE (BCH002186) 00888707005784,  1610DE (BCH002187) 00888707005784,  1610DE (BCH002188) 00888707005784,  1610DE (BCH002189) 00888707005784,  1610DE (BCH002190) 00888707005784,  1610DE (BCH002191) 00888707005784,  1610DE (BCH002192) 00888707005784,  1610JP (BCH002193) 00888707002257,  1610JP (BCH002194) 00888707002257,  1610JP (BCH002195) 00888707002257,  1610JP (BCH002196) 00888707002257,  1610DE (BCH002197) 00888707005784,  1610DE (BCH002198) 00888707005784,  1610DE (BCH002199) 00888707005784,  1610DE (BCH002200) 00888707005784,  1610DE (BCH002201) 00888707005784,  1610DE (BCH002202) 00888707005784,  1610DE (BCH002204) 00888707005784,  1610DE (BCH002205) 00888707005784,  1610DE (BCH002206) 00888707005784,  1610DE (BCH002207) 00888707005784,  1610DE (BCH002208) 00888707005784,  1610DE (BCH002209) 00888707005784,  1610DE (BCH002210) 00888707005784,  1610DE (BCH002211) 00888707005784,  1610DE (BCH002212) 00888707005784,  1610DE (BCH002213) 00888707005784,  1610DE (BCH002214) 00888707005784,  1610DE (BCH002215) 00888707005784,  1610DE (BCH002216) 00888707005784,  1610DE (BCH002217) 00888707005784,  1610GB (BCH002218) 00888707002974,  1610GB (BCH002219) 00888707005791,  1610GB (BCH002220) 00888707005791,  1610DE (BCH002221) 00888707005784,  1610DE (BCH002222) 00888707005784,  1610DE (BCH002223) 00888707005784,  1610DE (BCH002224) 00888707005784,  1610DE (BCH002225) 00888707005784,  1610DE (BCH002226) 00888707005784,  1610DE (BCH002227) 00888707005784,  1610GB (BCH002228) 00888707005791,  1610DE (BCH002229) 00888707005784,  1610GB (BCH002230) 00888707002974,  1610DE (BCH002231) 00888707005784,  1610DE (BCH002232) 00888707005784,  1610DE (BCH002233) 00888707005784,  1610DE (BCH002235) 00888707005784,  1610DE (BCH002236) 00888707005784,  1610DE (BCH002237) 00888707005784,  1610DE (BCH002238) 00888707005784,  1610DE (BCH002239) 00888707005784,  1610DE (BCH002240) 00888707005784,  1610IN (BCH002241) 00888707005814,  1610DE (BCH002242) 00888707005784,  1610GB (BCH002243) 00888707005791,  1610DE (BCH002244) 00888707005784,  1610DE (BCH002245) 00888707005784,  1610DE (BCH002246) 00888707005784,  1610GB (BCH002247) 00888707002974,  1610DE (BCH002248) 00888707005784,  1610DE (BCH002249) 00888707005784,  1610DE (BCH002250) 00888707005784,  1610DE (BCH002251) 00888707005784,  1610DE (BCH002252) 00888707005784,  1610IL (BCH002253) 00888707005807,  1610DE (BCH002254) 00888707005784,  1610DE (BCH002256) 00888707005784,  1610DE (BCH002257) 00888707005784,  1610DE (BCH002258) 00888707005784,  1610DE (BCH002259) 00888707005784,  1610GB (BCH002260) 00888707005791,  1610DE (BCH002261) 00888707005784,  1610DE (BCH002262) 00888707005784,  1610DE (BCH002263) 00888707005784,  1610DE (BCH002264) 00888707005784,  1610IT (BCH002265) 00888707005821,  1610IT (BCH002266) 00888707005821,  1610DE (BCH002267) 00888707005784,  1610DE (BCH002269) 00888707005784,  1610GB (BCH002270) 00888707005791,  1610JP (BCH002271) 00888707002257,  1610DE (BCH002272) 00888707005784,  1610DE (BCH002273) 00888707005784,  1610GB (BCH002274) 00888707005791,  1610DE (BCH002276) 00888707005784,  1610DE (BCH002277) 00888707005784,  1610IL (BCH002278) 00888707005807,  1610DE (BCH002279) 00888707005784,  1610DE (BCH002280) 00888707005784,  1610DE (BCH002282) 00888707005784,  1610DE (BCH002283) 00888707005784,  1610DE (BCH002284) 00888707005784,  1610DE (BCH002285) 00888707005784,  1610DE (BCH002286) 00888707005784,  1610DE (BCH002287) 00888707005784,  1610DE (BCH002288) 00888707005784,  1610DE (BCH002290) 00888707005784,  1610DE (BCH002291) 00888707005784,  1610DE (BCH002292) 00888707005784,  1610DE (BCH002293) 00888707005784,  1610DE (BCH002295) 00888707005784,  1610DE (BCH002296) 00888707005784,  1610DE (BCH002297) 00888707005784,  1610DE (BCH002298) 00888707005784,  1610DE (BCH002299) 00888707005784,  1610DE (BCH002300) 00888707005784,  1610DE (BCH002301) 00888707005784,  1610DE (BCH002302) 00888707005784,  1610DE (BCH002304) 00888707005784,  1610DE (BCH002305) 00888707005784,  1610DE (BCH002306) 00888707005784,  1610DE (BCH002307) 00888707005784,  1610DE (BCH002308) 00888707005784,  1610DE (BCH002309) 00888707005784,  1610GB (BCH002310) 00888707005791,  1610DE (BCH002311) 00888707005784,  1610DE (BCH002312) 00888707005784,  1610DE (BCH002313) 00888707005784,  1610DE (BCH002314) 00888707005784,  1610DE (BCH002315) 00888707005784,  1610IN (BCH002316) 00888707005814,  1610DE (BCH002318) 00888707005784,  1610DE (BCH002319) 00888707005784,  1610DE (BCH002320) 00888707005784,  1610DE (BCH002321) 00888707005784,  1610DE (BCH002323) 00888707005784,  1610DE (BCH002324) 00888707005784,  1610DE (BCH002325) 00888707005784,  1610JP (BCH002326) 00888707002257,  1610JP (BCH002327) 00888707002257,  1610DE (BCH002328) 00888707005784,  1610GB (BCH002329) 00888707005791,  1610DE (BCH002330) 00888707005784,  1610DE (BCH002331) 00888707005784,  1610DE (BCH002332) 00888707005784,  1610DE (BCH002334) 00888707005784,  1610DE (BCH002335) 00888707005784,  1610GB (BCH002336) 00888707002974,  1610GB (BCH002337) 00888707002974,  1610DE (BCH002338) 00888707005784,  1610DE (BCH002339) 00888707005784,  1610DE (BCH002340) 00888707005784,  1610DE (BCH002341) 00888707005784,  1610DE (BCH002342) 00888707005784,  1610GB (BCH002355) 00888707002974,  1610DE (BCH002356) 00888707005784,  1610DE (BCH002357) 00888707005784,  1610DE (BCH002358) 00888707005784,  1610DE (BCH002359) 00888707005784,  1610DE (BCH002360) 00888707005784,  1610DE (BCH002361) 00888707005784,  1610DE (BCH002362) 00888707005784,  1610DE (BCH002363) 00888707005784,  1610DE (BCH002364) 00888707005784,  1610GB (BCH002365) 00888707002974,  1610GB (BCH002366) 00888707002974,  1610DE (BCH002367) 00888707005784,  1610DE (BCH002368) 00888707005784,  1610DE (BCH002369) 00888707005784,  1610DE (BCH002370) 00888707005784,  1610DE (BCH002371) 00888707005784,  1610DE (BCH002372) 00888707005784,  1610DE (BCH002373) 00888707005784,  1610DE (BCH002374) 00888707005784,  1610AU (BCH002375) 00888707002165,  1610DE (BCH002376) 00888707005784,  1610DE (BCH002377) 00888707005784,  1610DE (BCH002379) 00888707005784,  1610DE (BCH002380) 00888707005784,  1610DE (BCH002381) 00888707005784,  1610DE (BCH002382) 00888707005784,  1610AU (BCH002383) 00888707002165,  1610DE (BCH002384) 00888707005784,  1610JP (BCH002392) 00888707002257,  1610JP (BCH002393) 00888707002257,  1610DE (BCH002394) 00888707005784,  1610JP (BCH002395) 00888707002257,  1610DE (BCH002396) 00888707005784,  1610DE (BCH002397) 00888707005784,  1610JP (BCH002398) 00888707002257,  1610JP (BCH002399) 00888707002257,  1610JP (BCH002400) 00888707002257,  1610JP (BCH002401) 00888707002257,  1610JP (BCH002402) 00888707002257,  1610JP (BCH002403) 00888707002257,  1610DE (BCH002404) 00888707005784,  1610JP (BCH002405) 00888707002257,  1610JP (BCH002406) 00888707002257,  1610JP (BCH002407) 00888707002257,  1610DE (BCH002408) 00888707005784,  1610DE (BCH002409) 00888707005784,  1610JP (BCH002410) 00888707002257,  1610DE (BCH002411) 00888707005784,  1610JP (BCH002412) 00888707002257,  1610JP (BCH002413) 00888707002257,  1610IT (BCH002414) 00888707005821,  1610DE (BCH002415) 00888707005784,  1610GB (BCH002416) 00888707005791,  1610GB (BCH002417) 00888707005791,  1610DE (BCH002418) 00888707005784,  1610DE (BCH002419) 00888707005784,  1610JP (BCH002420) 00888707002257,  1610DE (BCH002421) 00888707005784,  1610JP (BCH002422) 00888707002257,  1610JP (BCH002423) 00888707002257,  1610DE (BCH002424) 00888707005784,  1610JP (BCH002425) 00888707002257,  1610JP (BCH002426) 00888707002257,  1610JP (BCH002427) 00888707002257,  1610JP (BCH002428) 00888707002257,  1610DE (BCH002429) 00888707005784,  1610GB (BCH002430) 00888707005791,  1610DE (BCH002431) 00888707005784,  1610DE (BCH002432) 00888707005784,  1610GB (BCH002433) 00888707005791,  1610GB (BCH002434) 00888707005791,  1610JP (BCH002435) 00888707002257,  1610JP (BCH002436) 00888707002257,  1610JP (BCH002437) 00888707002257,  1610DE (BCH002438) 00888707005784,  1610JP (BCH002439) 00888707002257,  1610JP (BCH002440) 00888707002257,  1610JP (BCH002441) 00888707002257,  1610JP (BCH002442) 00888707002257,  1610DE (BCH002443) 00888707005784,  1610DE (BCH002444) 00888707005784,  1610DE (BCH002445) 00888707005784,  1610JP (BCH002446) 00888707002257,  1610DE (BCH002447) 00888707005784,  1610DE (BCH002448) 00888707005784,  1610DE (BCH002449) 00888707005784,  1610AU (BCH002462) 00888707002165,  1610GB (BCH002463) 00888707005791,  1610AU (BCH002464) 00888707002165,  1610DE (BCH002465) 00888707005784,  1610DE (BCH002466) 00888707005784,  1610DE (BCH002467) 00888707005784,  1610DE (BCH002468) 00888707005784,  1610GB (BCH002469) 00888707005791,  1610DE (BCH002470) 00888707005784,  1610DE (BCH002471) 00888707005784,  1610DE (BCH002472) 00888707005784,  1610GB (BCH002473) 00888707005791,  1610GB (BCH002474) 00888707005791,  1610DE (BCH002475) 00888707005784,  1610GB (BCH002476) 00888707005791,  1610DE (BCH002477) 00888707005784,  1610DE (BCH002478) 00888707005784,  1610GB (BCH002479) 00888707002219,  1610DE (BCH002480) 00888707005784,  1610DE (BCH002481) 00888707005784,  1610GB (BCH002482) 00888707005791,  1610GB (BCH002483) 00888707005791,  1610GB (BCH002484) 00888707005791,  1610DE (BCH002485) 00888707005784,  1610DE (BCH002486) 00888707005784,  1610DE (BCH002487) 00888707005784,  1610DE (BCH002488) 00888707005784,  1610DE (BCH002489) 00888707005784,  1610GB (BCH002490) 00888707005791,  1610CH (BCH002491) 00888707005777,  1610DE (BCH002492) 00888707005784,  1610CH (BCH002493) 00888707005777,  1610DE (BCH002494) 00888707005784,  1610DE (BCH002495) 00888707005784,  1610DE (BCH002496) 00888707005784,  1610GB (BCH002497) 00888707005791,  1610DE (BCH002498) 00888707005784,  1610DE (BCH002499) 00888707005784,  1610DE (BCH002500) 00888707005784,  1610DE (BCH002501) 00888707005784,  1610DE (BCH002502) 00888707005784,  1610IL (BCH002503) 00888707005807,  1610DE (BCH002504) 00888707005784,  1610DE (BCH002505) 00888707005784,  1610DE (BCH002506) 00888707005784,  1610DE (BCH002507) 00888707005784,  1610DE (BCH002508) 00888707005784,  1610GB (BCH002509) 00888707005791,  1610GB (BCH002510) 00888707005791,  1610DE (BCH002511) 00888707005784,  1610IT (BCH002512) 00888707005821,  1610DE (BCH002513) 00888707005784,  1610GB (BCH002514) 00888707005791,  1610JP (BCH002515) 00888707002257,  1610DE (BCH002516) 00888707005784,  1610DE (BCH002517) 00888707005784,  1610DE (BCH002518) 00888707005784,  1610DE (BCH002519) 00888707005784,  1610DE (BCH002520) 00888707005784,  1610DE (BCH002521) 00888707005784,  1610GB (BCH002522) 00888707005791,  1610DE (BCH002523) 00888707005784,  1610DE (BCH002524) 00888707005784,  1610DE (BCH002525) 00888707005784,  1610DE (BCH002526) 00888707005784,  1610DE (BCH002527) 00888707005784,  1610IT (BCH002528) 00888707005821,  1610JP (BCH002530) 00888707002257,  1610DE (BCH002531) 00888707005784,  1610DE (BCH002533) 00888707005784,  1610AU (BCH002534) 00888707002165,  1610DE (BCH002535) 00888707005784,  1610DE (BCH002536) 00888707005784,  1610DE (BCH002537) 00888707005784,  1610DE (BCH002538) 00888707005784,  1610DE (BCH002539) 00888707005784,  1610DE (BCH002540) 00888707005784,  1610DE (BCH002541) 00888707005784,  1610DE (BCH002542) 00888707005784,  1610DE (BCH002543) 00888707005784,  1610DE (BCH002544) 00888707005784,  1610DE (BCH002545) 00888707005784,  1610JP (BCH002546) 00888707002257,  1610IT (BCH002547) 00888707005821,  1610DE (BCH002548) 00888707005784,  1610DE (BCH002549) 00888707005784,  1610IT (BCH002550) 00888707005821,  1610DE (BCH002551) 00888707005784,  1610DE (BCH002552) 00888707005784,  1610DE (BCH002553) 00888707005784,  1610IN (BCH002554) 00888707005814,  1610DE (BCH002555) 00888707005784,  1610DE (BCH002556) 00888707005784,  1610DE (BCH002557) 00888707005784,  1610DE (BCH002558) 00888707005784,  1610DE (BCH002559) 00888707005784,  1610DE (BCH002560) 00888707005784,  1610DE (BCH002561) 00888707005784,  1610DE (BCH002562) 00888707005784,  1610DE (BCH002563) 00888707005784,  1610DE (BCH002564) 00888707005784,  1610GB (BCH002565) 00888707002974,  1610JP (BCH002566) 00888707002257,  1610DE (BCH002567) 00888707005784,  1610DE (BCH002568) 00888707005784,  1610IL (BCH002569) 00888707005807,  1610GB (BCH002570) 00888707005791,  1610DE (BCH002572) 00888707005784,  1610DE (BCH002573) 00888707005784,  1610DE (BCH002574) 00888707005784,  1610DE (BCH002575) 00888707005784,  1610DE (BCH002576) 00888707005784,  1610DE (BCH002577) 00888707005784,  1610IT (BCH002578) 00888707005821,  1610IT (BCH002579) 00888707005821,  1610DE (BCH002580) 00888707005784,  1610JP (BCH002581) 00888707002257,  1610JP (BCH002582) 00888707002257,  1610GB (BCH002583) 00888707005791,  1610JP (BCH002584) 00888707002257,  1610CH (BCH002585) 00888707005777,  1610DE (BCH002586) 00888707005784,  1610IT (BCH002587) 00888707005821,  1610DE (BCH002588) 00888707005784,  1610DE (BCH002589) 00888707005784,  1610DE (BCH002590) 00888707005784,  1610DE (BCH002591) 00888707005784,  1610DE (BCH002592) 00888707005784,  1610DE (BCH002593) 00888707005784,  1610DE (BCH002594) 00888707005784,  1610DE (BCH002595) 00888707005784,  1610DE (BCH002596) 00888707005784,  1610DE (BCH002597) 00888707005784,  1610DE (BCH002598) 00888707005784,  1610DE (BCH002599) 00888707005784,  1610DE (BCH002600) 00888707005784,  1610DE (BCH002601) 00888707005784,  1610DE (BCH002602) 00888707005784,  1610DE (BCH002603) 00888707005784,  1610IT (BCH002604) 00888707005821,  1610DE (BCH002605) 00888707005784,  1610DE (BCH002606) 00888707005784,  1610DE (BCH002607) 00888707005784,  1610JP (BCH002608) 00888707002257,  1610DE (BCH002609) 00888707005784,  1610DE (BCH002610) 00888707005784,  1610DE (BCH002611) 00888707005784,  1610IL (BCH002612) 00888707005807,  1610JP (BCH002613) 00888707002257,  1610DE (BCH002614) 00888707005784,  1610DE (BCH002615) 00888707005784,  1610JP (BCH002616) 00888707002257,  1610DE (BCH002617) 00888707005784,  1610DE (BCH002618) 00888707005784,  1610DE (BCH002619) 00888707005784,  1610GB (BCH002620) 00888707005791,  1610DE (BCH002621) 00888707005784,  1610JP (BCH002622) 00888707002257,  1610GB (BCH002623) 00888707005791,  1610DE (BCH002624) 00888707005784,  1610DE (BCH002625) 00888707005784,  1610DE (BCH002626) 00888707005784,  1610DE (BCH002627) 00888707005784,  1610DE (BCH002628) 00888707005784,  1610IT (BCH002629) 00888707005821,  1610DE (BCH002630) 00888707005784,  1610IN (BCH002631) 00888707005814,  1610DE (BCH002633) 00888707005784,  1610AU (BCH002634) 00888707002165,  1610DE (BCH002635) 00888707005784,  1610DE (BCH002636) 00888707005784,  1610DE (BCH002637) 00888707005784,  1610CH (BCH002638) 00888707005777,  1610DE (BCH002639) 00888707005784,  1610IT (BCH002640) 00888707005821,  1610JP (BCH002641) 00888707002257,  1610IT (BCH002642) 00888707005821,  1610DE (BCH002643) 00888707005784,  1610DE (BCH002644) 00888707005784,  1610IT (BCH002645) 00888707005821,  1610DE (BCH002646) 00888707005784,  1610DE (BCH002647) 00888707005784,  1610JP (BCH002648) 00888707002257,  1610DE (BCH002649) 00888707005784,  1610DE (BCH002650) 00888707005784,  1610DE (BCH002651) 00888707005784,  1610GB (BCH002652) 00888707005791,  1610DE (BCH002653) 00888707005784,  1610DE (BCH002654) 00888707005784,  1610DE (BCH002655) 00888707005784,  1610DE (BCH002656) 00888707005784,  1610DE (BCH002657) 00888707005784,  1610DE (BCH002658) 00888707005784,  1610DE (BCH002661) 00888707005784,  1610CH (BCH002662) 00888707005777,  1610DE (BCH002663) 00888707005784,  1610DE (BCH002664) 00888707005784,  1610DE (BCH002665) 00888707005784,  1610DE (BCH002666) 00888707005784,  1610DE (BCH002667) 00888707005784,  1610DE (BCH002668) 00888707005784,  1610DE (BCH002669) 00888707005784,  1610DE (BCH002670) 00888707005784,  1610DE (BCH002671) 00888707005784,  1610DE (BCH002672) 00888707005784,  1610DE (BCH002675) 00888707005784,  1610DE (BCH002676) 00888707005784,  1610JP (BCH002677) 00888707002257,  1610CH (BCH002678) 00888707005777,  1610DE (BCH002679) 00888707005784,  1610DE (BCH002680) 00888707005784,  1610DE (BCH002681) 00888707005784,  1610DE (BCH002682) 00888707005784,  1610DE (BCH002683) 00888707005784,  1610GB (BCH002684) 00888707002974,  1610DE (BCH002685) 00888707005784,  1610JP (BCH002686) 00888707002257,  1610DE (BCH002688) 00888707005784,  1610DE (BCH002689) 00888707005784,  1610DE (BCH002690) 00888707005784,  1610DE (BCH002691) 00888707005784,  1610GB (BCH002692) 00888707005791,  1610JP (BCH002693) 00888707002257,  1610DE (BCH002694) 00888707005784,  1610IN (BCH002695) 00888707005814,  1610DE (BCH002696) 00888707005784,  1610JP (BCH002697) 00888707002257,  1610DE (BCH002698) 00888707005784,  1610JP (BCH002699) 00888707002257,  1610DE (BCH002702) 00888707005784,  1610DE (BCH002703) 00888707005784,  1610DE (BCH002704) 00888707005784,  1610DE (BCH002705) 00888707005784,  1610DE (BCH002706) 00888707005784,  1610AU (BCH002707) 00888707005753,  1610GB (BCH002708) 00888707005791,  1610DE (BCH002709) 00888707005784,  1610DE (BCH002710) 00888707005784,  1610AU (BCH002711) 00888707005753,  1610DE (BCH002712) 00888707005784,  1610DE (BCH002713) 00888707005784,  1610DE (BCH002714) 00888707005784,  1610DE (BCH002715) 00888707005784,  1610DE (BCH002716) 00888707005784,  1610DE (BCH002717) 00888707005784,  1610DE (BCH002718) 00888707005784,  1610DE (BCH002719) 00888707005784,  1610JP (BCH002720) 00888707002257,  1610DE (BCH002721) 00888707005784,  1610DE (BCH002722) 00888707005784,  1610IT (BCH002723) 00888707005821,  1610IT (BCH002724) 00888707005821,  1610JP (BCH002725) 00888707002257,  1610DE (BCH002726) 00888707005784,  1610DE (BCH002727) 00888707005784,  1610DE (BCH002728) 00888707005784,  1610IT (BCH002729) 00888707005821,  1610DE (BCH002730) 00888707005784,  1610AU (BCH002731) 00888707005753,  1610DE (BCH002732) 00888707005784,  1610AU (BCH002733) 00888707005753,  1610DE (BCH002734) 00888707005784,  1610DE (BCH002735) 00888707005784,  1610AU (BCH002736) 00888707005753,  1610DE (BCH002737) 00888707005784,  1610DE (BCH002738) 00888707005784,  1610JP (BCH002739) 00888707002257,  1610DE (BCH002740) 00888707005784,  1610DE (BCH002741) 00888707005784,  1610DE (BCH002742) 00888707005784,  1610DE (BCH002743) 00888707005784,  1610DE (BCH002744) 00888707005784,  1610DE (BCH002745) 00888707005784,  1610DE (BCH002746) 00888707005784,  1610DE (BCH002747) 00888707005784,  1610DE (BCH002748) 00888707005784,  1610DE (BCH002749) 00888707005784,  1610GB (BCH002750) 00888707005791,  1610DE (BCH002751) 00888707005784,  1610DE (BCH002752) 00888707005784,  1610AU (BCH002753) 00888707005753,  1610DE (BCH002754) 00888707005784,  1610DE (BCH002755) 00888707005784,  1610DE (BCH002756) 00888707005784,  1610DE (BCH002757) 00888707005784,  1610DE (BCH002758) 00888707005784,  1610DE (BCH002759) 00888707005784,  1610DE (BCH002760) 00888707005784,  1610DE (BCH002761) 00888707005784,  1610DE (BCH002762) 00888707005784,  1610AU (BCH002763) 00888707005753,  1610DE (BCH002764) 00888707005784,  1610DE (BCH002765) 00888707005784,  1610JP (BCH002766) 00888707002257,  1610DE (BCH002767) 00888707005784,  1610DE (BCH002768) 00888707005784,  1610DE (BCH002769) 00888707005784,  1610DE (BCH002770) 00888707005784,  1610DE (BCH002771) 00888707005784,  1610GB (BCH002772) 00888707005791,  1610DE (BCH002773) 00888707005784,  1610DE (BCH002774) 00888707005784,  1610DE (BCH002775) 00888707005784,  1610DE (BCH002776) 00888707005784,  1610DE (BCH002777) 00888707005784,  1610DE (BCH002778) 00888707005784,  1610DE (BCH002779) 00888707005784,  1610DE (BCH002780) 00888707005784,  1610DE (BCH002783) 00888707005784,  1610DE (BCH002784) 00888707005784,  1610GB (BCH002785) 00888707005791,  1610DE (BCH002786) 00888707005784,  1610IT (BCH002787) 00888707005821,  1610IT (BCH002788) 00888707005821,  1610IT (BCH002789) 00888707005821,  1610DE (BCH002790) 00888707005784,  1610DE (BCH002791) 00888707005784,  1610DE (BCH002792) 00888707005784,  1610IT (BCH002793) 00888707005821,  1610AU (BCH002794) 00888707005753,  1610AU (BCH002795) 00888707005753,  1610DE (BCH002796) 00888707005784,  1610DE (BCH002797) 00888707005784,  1610DE (BCH002798) 00888707005784,  1610DE (BCH002799) 00888707005784,  1610DE (BCH002800) 00888707005784,  1610GB (BCH002801) 00888707005791,  1610DE (BCH002802) 00888707005784,  1610DE (BCH002803) 00888707005784,  1610DE (BCH002804) 00888707005784,  1610DE (BCH002805) 00888707005784,  1610DE (BCH002806) 00888707005784,  1610DE (BCH002807) 00888707005784,  1610DE (BCH002808) 00888707005784,  1610DE (BCH002809) 00888707005784,  1610DE (BCH002810) 00888707005784,  1610DE (BCH002811) 00888707005784,  1610DE (BCH002812) 00888707005784,  1610GB (BCH002813) 00888707005791,  1610GB (BCH002814) 00888707005791,  1610DE (BCH002815) 00888707005784,  1610GB (BCH002816) 00888707005791,  1610DE (BCH002817) 00888707005784,  1610DE (BCH002818) 00888707005784,  1610DE (BCH002819) 00888707005784,  1610DE (BCH002820) 00888707005784,  1610GB (BCH002821) 00888707005791,  1610DE (BCH002822) 00888707005784,  1610DE (BCH002823) 00888707005784,  1610DE (BCH002824) 00888707005784,  1610DE (BCH002825) 00888707005784,  1610DE (BCH002826) 00888707005784,  1610DE (BCH002827) 00888707005784,  1610DE (BCH002828) 00888707005784,  1610DE (BCH002829) 00888707005784,  1610JP (BCH002830) 00888707002257,  1610DE (BCH002831) 00888707005784,  1610DE (BCH002832) 00888707005784,  1610DE (BCH002833) 00888707005784,  1610AU (BCH002834) 00888707005753,  1610JP (BCH002862) 00888707002257,  1610JP (BCH002863) 00888707002257,  1610DE (BCH002864) 00888707005784,  1610DE (BCH002867) 00888707005784,  1610DE (BCH002869) 00888707005784,  1610GB (BCH002870) 00888707005791,  1610DE (BCH002871) 00888707005784,  1610JP (BCH002873) 00888707002257,  1610GB (BCH002874) 00888707005791,  1610DE (BCH002875) 00888707005784,  1610JP (BCH002876) 00888707002257,  1610DE (BCH002877) 00888707005784,  1610JP (BCH002878) 00888707002257,  1610DE (BCH002879) 00888707005784,  1610JP (BCH002880) 00888707002257,  1610DE (BCH002881) 00888707005784,  1610DE (BCH002883) 00888707005784,  1610DE (BCH002885) 00888707005784,  1610DE (BCH002886) 00888707005784,  1610IT (BCH002888) 00888707005821,  1610IN (BCH002889) 00888707005814,  1610DE (BCH002890) 00888707005784,  1610DE (BCH002891) 00888707005784,  1610DE (BCH002894) 00888707005784,  1610IT (BCH002895) 00888707005821,  1610DE (BCH002896) 00888707005784,  1610DE (BCH002897) 00888707005784,  1610DE (BCH002898) 00888707005784,  1610JP (BCH002901) 00888707002257,  1610DE (BCH002902) 00888707005784,  1610DE (BCH002905) 00888707005784,  1610DE (BCH002906) 00888707005784,  1610DE (BCH002907) 00888707005784,  1610DE (BCH002908) 00888707005784,  1610JP (BCH002910) 00888707002257,  1610DE (BCH002911) 00888707005784,  1610DE (BCH002915) 00888707005784,  1610DE (BCH002916) 00888707005784,  1610GB (BCH002917) 00888707002974,  1610DE (BCH002918) 00888707005784,  1610DE (BCH002919) 00888707005784,  1610DE (BCH002920) 00888707005784,  1610DE (BCH002921) 00888707005784,  1610DE (BCH002923) 00888707005784,  1610DE (BCH002924) 00888707005784,  1610DE (BCH002925) 00888707005784,  1610DE (BCH002926) 00888707005784,  1610GB (BCH002927) 00888707005791,  1610DE (BCH002928) 00888707005784,  1610GB (BCH002929) 00888707005791,  1610DE (BCH002930) 00888707005784,  1610JP (BCH002931) 00888707002257,  1610DE (BCH002932) 00888707005784,  1610DE (BCH002933) 00888707005784,  1610AU (BCH002934) 00888707005753,  1610DE (BCH002935) 00888707005784,  1610AU (BCH002936) 00888707005753,  1610GB (BCH002937) 00888707002974,  1610DE (BCH002938) 00888707005784,  1610AU (BCH002939) 00888707005753,  1610DE (BCH002940) 00888707005784,  1610AU (BCH002941) 00888707005753,  1610DE (BCH002942) 00888707005784,  1610DE (BCH002943) 00888707005784,  1610DE (BCH002944) 00888707005784,  1610DE (BCH002945) 00888707005784,  1610DE (BCH002946) 00888707005784,  1610DE (BCH002947) 00888707005784,  1610DE (BCH002948) 00888707005784,  1610DE (BCH002949) 00888707005784,  1610AU (BCH002950) 00888707005753,  1610DE (BCH002951) 00888707005784,  1610DE (BCH002952) 00888707005784,  1610DE (BCH002953) 00888707005784,  1610CA (BCH002954) 00888707005760,  1610DE (BCH002955) 00888707005784,  1610DE (BCH002956) 00888707005784,  1610GB (BCH002957) 00888707005791,  1610DE (BCH002958) 00888707005784,  1610DE (BCH002959) 00888707005784,  1610CA (BCH002960) 00888707005760,  1610DE (BCH002961) 00888707005784,  1610DE (BCH002962) 00888707005784,  1610IT (BCH002963) 00888707005821,  1610DE (BCH002964) 00888707005784,  1610DE (BCH002965) 00888707005784,  1610DE (BCH002966) 00888707005784,  1610DE (BCH002967) 00888707005784,  1610DE (BCH002968) 00888707005784,  1610DE (BCH002969) 00888707005784,  1610DE (BCH002970) 00888707005784,  1610DE (BCH002971) 00888707005784,  1610DE (BCH002972) 00888707005784,  1610JP (BCH002973) 00888707002257,  1610IT (BCH002974) 00888707005821,  1610DE (BCH002975) 00888707005784,  1610JP (BCH002976) 00888707002257,  1610DE (BCH002977) 00888707005784,  1610DE (BCH002978) 00888707005784,  1610DE (BCH002979) 00888707005784,  1610DE (BCH002980) 00888707005784,  1610DE (BCH002981) 00888707005784,  1610DE (BCH002982) 00888707005784,  1610JP (BCH002983) 00888707002257,  1610DE (BCH002985) 00888707005784,  1610AU (BCH002986) 00888707005753,  1610GB (BCH002987) 00888707005791,  1610DE (BCH002988) 00888707005784,  1610DE (BCH002989) 00888707005784,  1610GB (BCH002990) 00888707005791,  1610DE (BCH002991) 00888707005784,  1610AU (BCH002992) 00888707005753,  1610DE (BCH002993) 00888707005784,  1610DE (BCH002994) 00888707005784,  1610DE (BCH002995) 00888707005784,  1610DE (BCH002996) 00888707005784,  1610DE (BCH002997) 00888707005784,  1610DE (BCH002998) 00888707005784,  1610DE (BCH002999) 00888707005784,  1610JP (BCH003000) 00888707002257,  1610AU (BCH003001) 00888707005753,  1610DE (BCH003002) 00888707005784,  1610DE (BCH003003) 00888707005784,  1610DE (BCH003004) 00888707005784,  1610JP (BCH003005) 00888707002257,  1610DE (BCH003006) 00888707005784,  1610DE (BCH003007) 00888707005784,  1610DE (BCH003008) 00888707005784,  1610DE (BCH003009) 00888707005784,  1610DE (BCH003010) 00888707005784,  1610DE (BCH003011) 00888707005784,  1610DE (BCH003012) 00888707005784,  1610DE (BCH003013) 00888707005784,  1610GB (BCH003014) 00888707005791,  1610DE (BCH003015) 00888707005784,  1610DE (BCH003016) 00888707005784,  1610DE (BCH003017) 00888707005784,  1610GB (BCH003019) 00888707005791,  1610GB (BCH003020) 00888707005791,  1610DE (BCH003022) 00888707005784,  1610DE (BCH003023) 00888707005784,  1610IT (BCH003024) 00888707005821,  1610DE (BCH003025) 00888707005784,  1610DE (BCH003026) 00888707005784,  1610DE (BCH003027) 00888707005784,  1610DE (BCH003028) 00888707005784,  1610DE (BCH003029) 00888707005784,  1610GB (BCH003030) 00888707005791,  1610DE (BCH003031) 00888707005784,  1610IT (BCH003032) 00888707005821,  1610CA (BCH003033) 00888707005760,  1610DE (BCH003034) 00888707005784,  1610DE (BCH003035) 00888707005784,  1610DE (BCH003036) 00888707005784,  1610DE (BCH003037) 00888707005784,  1610AU (BCH003038) 00888707005753,  1610DE (BCH003039) 00888707005784,  1610GB (BCH003040) 00888707005791,  1610IT (BCH003041) 00888707005821,  1610CA (BCH003042) 00888707005760,  1610DE (BCH003043) 00888707005784,  1610DE (BCH003044) 00888707005784,  1610DE (BCH003045) 00888707005784,  1610GB (BCH003046) 00888707005791,  1610DE (BCH003047) 00888707005784,  1610DE (BCH003048) 00888707005784,  1610DE (BCH003049) 00888707005784,  1610DE (BCH003050) 00888707005784,  1610DE (BCH003051) 00888707005784,  1610DE (BCH003052) 00888707005784,  1610IT (BCH003053) 00888707005821,  1610DE (BCH003054) 00888707005784,  1610DE (BCH003055) 00888707005784,  1610DE (BCH003056) 00888707005784,  1610AU (BCH003057) 00888707005753,  1610DE (BCH003058) 00888707005784,  1610DE (BCH003059) 00888707005784,  1610GB (BCH003060) 00888707005791,  1610DE (BCH003061) 00888707005784,  1610AU (BCH003062) 00888707005753,  1610AU (BCH003063) 00888707005753,  1610DE (BCH003065) 00888707005784,  1610DE (BCH003066) 00888707005784,  1610DE (BCH003067) 00888707005784,  1610AU (BCH003068) 00888707005753,  1610DE (BCH003069) 00888707005784,  1610AU (BCH003070) 00888707005753,  1610DE (BCH003071) 00888707005784,  1610DE (BCH003072) 00888707005784,  1610DE (BCH003073) 00888707005784,  1610GB (BCH003074) 00888707005791,  1610DE (BCH003075) 00888707005784,  1610DE (BCH003078) 00888707005784,  1610IT (BCH003079) 00888707005821,  1610DE (BCH003080) 00888707005784,  1610DE (BCH003081) 00888707005784,  1610GB (BCH003082) 00888707002974,  1610DE (BCH003083) 00888707005784,  1610DE (BCH003084) 00888707005784,  1610AU (BCH003085) 00888707005753,  1610DE (BCH003086) 00888707005784,  1610DE (BCH003087) 00888707005784,  1610GB (BCH003089) 00888707005791,  1610GB (BCH003090) 00888707005791,  1610DE (BCH003094) 00888707005784,  1610DE (BCH003095) 00888707005784,  1610DE (BCH003096) 00888707005784,  1610DE (BCH003097) 00888707005784,  1610GB (BCH003099) 00888707005791,  1610GB (BCH003100) 00888707005791,  1610DE (BCH003101) 00888707005784,  1610DE (BCH003102) 00888707005784,  1610DE (BCH003103) 00888707005784,  1610DE (BCH003105) 00888707005784,  1610AU (BCH003106) 00888707005753,  1610DE (BCH003107) 00888707005784,  1610DE (BCH003108) 00888707005784,  1610DE (BCH003172) 00888707001182,  1610DE (BCH003512) 00888707001182,  1610DE (BCH003896) 00888707001182,  1610DE (BCH004236) 00888707001182,  1610AU (BCH004527) 00888707001144,  1610CA (BCH004727) 00888707001175,  1610DE (BCH004728) 00888707001182,  1610DE (BCH004729) 00888707001182,  1610DE (BCH004730) 00888707001182,  1610CA (BCH004732) 00888707001175,  1610DE (BCH004762) 00888707001182,  1610DE (BCH004763) 00888707001182,  1610DE (BCH004764) 00888707001182,  1610DE (BCH004765) 00888707001182,  1610DE (BCH004766) 00888707001182,  1610DE (BCH004767) 00888707001182,  1610DE (BCH004791) 00888707001182,  1610DE (BCH004796) 00888707001182,  1610DE (BCH004797) 00888707001182,  1610GB (BCH004798) 00888707001205,  1610GB (BCH004799) 00888707001205,  1610DE (BCH004800) 00888707001182,  1610IL (BCH004884) 00888707001229,  1610IL (BCH004899) 00888707001229,  1610DE (BCH005045) 00888707001182,  1610DE (BCH005108) 00888707001182,  1610DE (BCH005243) 00888707001182,  1610GB (BCH005324) 00888707001205,  1610GB (BCH005332) 00888707001205,  1610IL (BCH005471) 00888707001229,  1610DE (BCH005524) 00888707001182,  1610DE (BCH005766) 00888707001182,  1610DE (BCH005767) 00888707001182,  1610DE (BCH005768) 00888707001182,  1610DE (BCH005773) 00888707001182,  1610GB (BCH005779) 00888707001205,  1610DE (BCH005781) 00888707001182,  1610DE (BCH005782) 00888707001182,  1610DE (BCH005783) 00888707001182,  1610IT (BCH005785) 00888707001212,  1610DE (BCH005787) 00888707001182,  1610GB (BCH005790) 00888707001205,  1610DE (BCH005794) 00888707001182,  1610GB (BCH005799) 00888707001205,  1610GB (BCH005801) 00888707001205,  1610GB (BCH005811) 00888707001205,  1610DE (BCH005815) 00888707001182,  1610DE (BCH005817) 00888707001182,  1610DE (BCH005819) 00888707001182,  1610IT (BCH005820) 00888707001212,  1610DE (BCH005821) 00888707001182,  1610DE (BCH005822) 00888707001182,  1610DE (BCH005825) 00888707001182,  1610DE (BCH005842) 00888707001182,  1610AU (BCH005868) 00888707001144,  1610AU (BCH005869) 00888707001144,  1610GB (BCH005870) 00888707001205,  1610AU (BCH005871) 00888707001144,  1610BR (BCH005872) 00888707001168,  1610GB (BCH005873) 00888707001205,  1610CA (BCH005875) 00888707001175,  1610GB (BCH005877) 00888707001205,  1610DE (BCH005878) 00888707001182,  1610DE (BCH005881) 00888707001182,  1610DE (BCH005882) 00888707001182,  1610CA (BCH005883) 00888707001175,  1610BR (BCH005884) 00888707001168,  1610BR (BCH005885) 00888707001168,  1610GB (BCH005886) 00888707001205,  1610GB (BCH005890) 00888707001205,  1610GB (BCH005891) 00888707001205,  1610GB (BCH005892) 00888707001205,  1610CA (BCH005893) 00888707001175,  1610GB (BCH005894) 00888707001205,  1610DE (BCH005895) 00888707001182,  1610DE (BCH005896) 00888707001182,  1610DE (BCH005897) 00888707001182,  1610CA (BCH005899) 00888707001175,  1610CA (BCH005900) 00888707001175,  1610GB (BCH005901) 00888707001205,  1610GB (BCH005902) 00888707001205,  1610DE (BCH005903) 00888707001182,  1610AU (BCH005904) 00888707001144,  1610AU (BCH005905) 00888707001144,  1610GB (BCH005906) 00888707001205,  1610DE (BCH005910) 00888707001182,  1610DE (BCH005911) 00888707001182,  1610DE (BCH005912) 00888707001182,  1610DE (BCH005913) 00888707001182,  1610AU (BCH005914) 00888707001144,  1610AU (BCH005915) 00888707001144,  1610DE (BCH005917) 00888707001182,  1610CA (BCH005920) 00888707001175,  1610IT (BCH005921) 00888707001212,  1610IT (BCH005922) 00888707001212,  1610DE (BCH005923) 00888707001182,  1610DE (BCH005924) 00888707001182,  1610CA (BCH005925) 00888707001175,  1610DE (BCH005926) 00888707001182,  1610DE (BCH005927) 00888707001182,  1610DE (BCH005928) 00888707001182,  1610IL (BCH005930) 00888707001229,  1610IL (BCH005931) 00888707001229,  1610IT (BCH005932) 00888707001212,  1610DE (BCH005935) 00888707001182,  1610BR (BCH005939) 00888707001168,  1610IT (BCH005940) 00888707001212,  1610DE (BCH005941) 00888707001182,  1610CA (BCH005942) 00888707001175,  1610CA (BCH005943) 00888707001175,  1610CA (BCH005944) 00888707001175,  1610IT (BCH005945) 00888707001212,  1610DE (BCH005947) 00888707001182,  1610CA (BCH005951) 00888707001175,  1610CA (BCH005952) 00888707001175,  1610GB (BCH005953) 00888707001205,  1610CH (BCH005954) 00888707001199,  1610DE (BCH005955) 00888707001182,  1610IN (BCH005956) 00888707000536,  1610DE (BCH005957) 00888707001182,  1610CH (BCH005961) 00888707001199,  1610DE (BCH005962) 00888707001182,  1610IN (BCH005963) 00888707000536,  1610CA (BCH005964) 00888707001175,  1610DE (BCH005965) 00888707001182,  1610DE (BCH005966) 00888707001182,  1610DE (BCH005967) 00888707001182,  1610DE (BCH005969) 00888707001182,  1610AU (BCH005970) 00888707001144,  1610DE (BCH005972) 00888707001182,  1610DE (BCH005973) 00888707001182,  1610DE (BCH005974) 00888707001182,  1610DE (BCH005975) 00888707001182,  1610DE (BCH005976) 00888707001182,  1610IT (BCH005977) 00888707001212,  1610DE (BCH005979) 00888707001182,  1610GB (BCH005980) 00888707001205,  1610IN (BCH005981) 00888707000536,  1610DE (BCH005983) 00888707001182,  1610DE (BCH005984) 00888707001182,  1610GB (BCH005987) 00888707001205,  1610GB (BCH005988) 00888707001205,  1610DE (BCH005990) 00888707001182,  1610DE (BCH005992) 00888707001182,  1610CA (BCH005993) 00888707001175,  1610DE (BCH005994) 00888707001182,  1610IN (BCH005995) 00888707000536,  1610DE (BCH005996) 00888707001182,  1610CA (BCH005997) 00888707001175,  1610DE (BCH005998) 00888707001182,  1610DE (BCH006000) 00888707001182,  1610IT (BCH006001) 00888707001212,  1610IN (BCH006003) 00888707000536,  1610IN (BCH006004) 00888707000536,  1610DE (BCH006007) 00888707001182,  1610AU (BCH006008) 00888707001144,  1610DE (bch006009) 00888707001182,  1610GB (BCH006010) 00888707001205,  1610DE (BCH006011) 00888707001182,  1610DE (BCH006016) 00888707001182,  1610DE (BCH006017) 00888707001182,  1610DE (BCH006018) 00888707001182,  1610GB (BCH006019) 00888707001205,  1610IT (BCH006020) 00888707001212,  1610IT (BCH006021) 00888707001212,  1610IT (BCH006022) 00888707001212,  1610AU (BCH006023) 00888707001144,  1610AU (BCH006024) 00888707001144,  1610DE (BCH006026) 00888707001182,  1610DE (BCH006027) 00888707001182,  1610IN (BCH006028) 00888707000536,  1610IN (BCH006029) 00888707000536,  1610IN (BCH006030) 00888707000536,  1610IN (BCH006033) 00888707000536,  1610DE (BCH006034) 00888707001182,  1610GB (BCH006035) 00888707001205,  1610GB (BCH006036) 00888707001205,  1610GB (BCH006037) 00888707001205,  1610GB (BCH006038) 00888707001205,  1610IT (BCH006040) 00888707001212,  1610IT (BCH006041) 00888707001212,  1610IT (BCH006042) 00888707001212,  1610DE (BCH006044) 00888707001182,  1610DE (BCH006045) 00888707001182,  1610DE (BCH006046) 00888707001182,  1610DE (BCH006047) 00888707001182,  1610DE (BCH006049) 00888707001182,  1610DE (BCH006051) 00888707001182,  1610DE (BCH006053) 00888707001182,  1610DE (BCH006054) 00888707001182,  1610DE (BCH006062) 00888707001182,  1610DE (BCH006069) 00888707001182,  1610DE (BCH006070) 00888707001182,  1610DE (BCH006071) 00888707001182,  1610DE (BCH006093) 00888707001182,  1610BR (BCH006102) 00888707001168,  1610CA (BCH006104) 00888707001175,  1610DE (BCH006109) 00888707001182,  1610DE (BCH006110) 00888707001182,  1610DE (BCH006114) 00888707001182,  1610DE (BCH006115) 00888707001182,  1610DE (BCH006116) 00888707001182,  1610DE (BCH006119) 00888707001182,  1610DE (BCH006120) 00888707001182,  1610DE (BCH006121) 00888707001182,  1610DE (BCH006122) 00888707001182,  1610DE (BCH006138) 00888707001182,  1610CA (BCH006139) 00888707001175,  1610CA (BCH006140) 00888707001175,  1610CA (BCH006146) 00888707001175,  1610DE (BCH006147) 00888707001182,  1610AU (bch006148) 00888707001144,  1610CA (BCH006149) 00888707001175,  1610DE (BCH006156) 00888707001182,  1610DE (BCH006157) 00888707001182,  1610CA (BCH006158) 00888707001175,  1610CA (BCH006162) 00888707001175,  1610CA (BCH006165) 00888707001175,  1610CA (BCH006171) 00888707001175,  1610CA (BCH006179) 00888707001175,  1610AU (BCH006248) 00888707001144,  1610IN (BCH006249) 00888707000536,  1610DE (BCH006251) 00888707001182,  1610DE (BCH006252) 00888707001182,  1610DE (BCH006253) 00888707001182,  1610AU (BCH006255) 00888707001144,  1610DE (BCH006257) 00888707001182,  1610AU (BCH006258) 00888707001144,  1610AU (BCH006259) 00888707001144,  1610AU (BCH006260) 00888707001144,  1610GB (BCH006262) 00888707001205,  1610DE (BCH006263) 00888707001182,  1610DE (BCH006264) 00888707001182,  1610DE (BCH006265) 00888707001182,  1610DE (BCH006266) 00888707001182,  1610DE (BCH006267) 00888707001182,  1610DE (BCH006268) 00888707001182,  1610DE (BCH006269) 00888707001182,  1610DE (BCH006270) 00888707001182,  1610DE (BCH006271) 00888707001182,  1610AU (BCH006273) 00888707001144,  1610CA (BCH006275) 00888707001175,  1610AU (BCH006276) 00888707001144,  1610AU (BCH006277) 00888707001144,  1610DE (BCH006278) 00888707001182,  1610AU (BCH006279) 00888707001144,  1610DE (BCH006280) 00888707001182,  1610DE (BCH006281) 00888707001182,  1610DE (BCH006282) 00888707001182,  1610AR (BCH006283) 00888707001151,  1610IN (BCH006285) 00888707000536,  1610IN (BCH006286) 00888707000536,  1610AU (BCH006287) 00888707001144,  1610AU (BCH006288) 00888707001144,  1610AU (BCH006289) 00888707001144,  1610DE (BCH006290) 00888707001182,  1610AU (BCH006291) 00888707001144,  1610AU (BCH006292) 00888707001144,  1610CA (BCH006293) 00888707001175,  1610DE (BCH006295) 00888707001182,  1610DE (BCH102179) 00888707005784,  1610DE (BCH102180) 00888707005784,  1610DE (BCH102181) 00888707005784,  1610DE (BCH102182) 00888707005784,  1610DE (BCH102184) 00888707005784,  1610DE (BCH102185) 00888707005784,  1610DE (BCH102188) 00888707005784,  1610DE (BCH102190) 00888707005784,  1610DE (BCH102192) 00888707005784,  1610DE (BCH102196) 00888707005784,  1610IT (BCH102197) 00888707005821,  1610IT (BCH102198) 00888707005821,  1610DE (BCH102206) 00888707005784,  1610DE (BCH102207) 00888707005784,  1610DE (BCH102218) 00888707005784,  1610DE (BCH102219) 00888707005784,  1610DE (BCH102224) 00888707005784,  1610AU (BCH102227) 00888707005753,  1610DE (BCH102229) 00888707005784,  1610DE (BCH102230) 00888707005784,  1610DE (BCH102231) 00888707005784,  1610DE (BCH102232) 00888707005784,  1610DE (BCH102236) 00888707005784,  1610GB (BCH102281) 00888707002974,  1610DE (BCH000808) 00888707001182,  1610DE (BCH002124) 00888707001182,  1610DE (BCH002131) 00888707001182,  1610DE (BCH002203) 00888707001182,  1610CH (BCH000580) 00888707002196,  1610GB (BCH001144) 00888707002974,  1610CA (BCH001146) 00888707003001,  1610GB (BCH001167) 00888707001205,  1610GB (BCH001183) 00888707001205,  1610GB (BCH001193) 00888707001205,  1610GB (BCH003596) 00888707001205,  1610BR (BCH003655) 00888707001168,  1610BR (BCH003656) 00888707001168,  1610BR (BCH003663) 00888707001168,  1610BR (BCH003670) 00888707001168,  1610BR (BCH003672) 00888707001168,  1610BR (BCH003673) 00888707001168,  1610BR (BCH003677) 00888707001168,  1610BR (BCH003698) 00888707001168,  1610BR (BCH003701) 00888707001168,  1610BR (BCH003719) 00888707001168,  1610BR (BCH003732) 00888707001168,  1610BR (BCH003735) 00888707001168,  1610AU (BCH004067) 00888707001144,  1610AU (BCH004295) 00888707001144,  1610JP (BCH004300) 00888707001236,  1610DE (BCH004303) 00888707001182,  1610AU (BCH004317) 00888707001144,  1610IT (BCH004377) 00888707001212,  1610CH (BCH004398) 00888707001199,  1610JP (BCH004422) 00888707001236,  1610JP (BCH004443) 00888707001236,  1610CA-CLIN (BCH004506) 00888707001564,  1610CA-CLIN (BCH004528) 00888707001564,  1610GB (BCH005982) 00888707001205,  1610GB (BCH006012) 00888707001205,  1610BR (BCH006013) 00888707001168,  1610DE (BCH006084) 00888707001182,  1610AU (BCH006261) 00888707001144,  1610AU (BCH006274) 00888707001144,  1610CA (BCH006294) 00888707001175,  1610CA (BCH006103) 00888707001175,  1610CA (BCH005929) 00888707001175,  1610CA (BCH005919) 00888707001175,  1610CA (BCH005889) 00888707001175,  1610DE (BCH005392) 00888707001182,  1610DE (BCH005971) 00888707001182,  1610DE (BCH000611) 00888707002202,  1610GB (BCH005231) 00888707001205,  1610GB (BCH006272) 00888707001205,  1610GB (BCH006284) 00888707001205,  1610CA (BCH006250) 00888707001175,  1610CA (BCH005876) 00888707001175,  1610CA (BCH005887) 00888707001175,  1610DE (BCH000075) 00888707002202,  1610DE (BCH000049) 00888707002202,  1610DE (BCH005999) 00888707001182,  1610DE (BCH006025) 00888707001182,  1610DE (BCH005880) 00888707001182,  1610DE (BCH005946) 00888707001182,  1610IN (BCH006032) 00888707000536,  1610DE (BCH006005) 00888707001182,  1610DE (BCH006006) 00888707001182,  1610DE (BCH006043) 00888707001182,  1610CA (BCH005537) 00888707001175,  1610DE (BCH005916) 00888707001182,  1610DE (BCH005918) 00888707001182,  1610IT (BCH005834) 00888707001212,  1610DE (BCH006052) 00888707001182,  1610DE (BCH006098) 00888707001182,  1610IT (BCH006113) 00888707001212,  1610CA (BCH006254) 00888707001175,  1610CH (BCH005279) 00888707001199,  1610DE (BCH005789) 00888707001182,  1610DE (BCH005796) 00888707001182,  1610DE (BCH006075) 00888707001182,  1610IT (BCH006080) 00888707001212,  1610IT (BCH006095) 00888707001212,  1610DE (BCH006100) 00888707001182,  1610DE (BCH006111) 00888707001182,  1610DE (BCH006112) 00888707001182,  1610CA (BCH005833) 00888707001175,  1610GB (BCH005303) 00888707001205,  1610DE (BCH005713) 00888707001182,  1610CA (BCH005802) 00888707001175,  1610CA (BCH005828) 00888707001175,  1610DE (BCH006091) 00888707001182,  1610DE (BCH006117) 00888707001182,  1610AU (BCH005607) 00888707001144,  1610DE (BCH005728) 00888707001182,  1610DE (BCH006073) 00888707001182,  1610AU (BCH005308) 00888707001144,  1610DE (BCH005673) 00888707001182,  1610DE (BCH005793) 00888707001182,  1610DE (BCH005804) 00888707001182,  1610DE (BCH006078) 00888707001182,  1610DE (BCH006048) 00888707001182,  1610DE (BCH006061) 00888707001182,  1610DE (BCH006074) 00888707001182,  1610GB (BCH005326) 00888707001205,  1610DE (BCH005774) 00888707001182,  1610IN (BCH005791) 00888707000536,  1610DE (BCH005807) 00888707001182,  1610IN (BCH005818) 00888707000536,  1610DE (BCH006060) 00888707001182,  1610DE (BCH006079) 00888707001182,  1610DE (BCH006097) 00888707001182,  1610DE (BCH006108) 00888707001182,  1610DE (BCH005763) 00888707001182,  1610DE (BCH006055) 00888707001182,  1610DE (BCH006059) 00888707001182,  1610DE (BCH006065) 00888707001182,  1610AU (BCH005321) 00888707001144,  1610IT (BCH005689) 00888707001212,  1610IT (BCH005827) 00888707001212,  1610DE (BCH006050) 00888707001182,  1610DE (BCH006057) 00888707001182,  1610DE (BCH006105) 00888707001182,  1610DE (BCH006107) 00888707001182,  1610DE (BCH005697) 00888707001182,  1610DE (BCH005784) 00888707001182,  1610DE (BCH005795) 00888707001182,  1610DE (BCH005800) 00888707001182,  1610DE (BCH005810) 00888707001182,  1610DE (BCH006058) 00888707001182,  1610DE (BCH006068) 00888707001182,  1610DE (BCH006072) 00888707001182,  1610DE (BCH006090) 00888707001182,  1610DE (BCH006094) 00888707001182,  1610DE (BCH006096) 00888707001182,  1610DE (BCH006101) 00888707001182,  1610DE (BCH005755) 00888707001182,  1610DE (BCH005757) 00888707001182,  1610DE (BCH005770) 00888707001182,  1610DE (BCH006092) 00888707001182,  1610DE (BCH006099) 00888707001182,  1610DE (BCH006106) 00888707001182,  1610DE (BCH005780) 00888707001182,  1610DE (BCH005809) 00888707001182,  1610DE (BCH005814) 00888707001182,  1610DE (BCH005816) 00888707001182,  1610DE (BCH005806) 00888707001182,  1610DE (BCH005824) 00888707001182,  1610DE (BCH005841) 00888707001182,  1610AU (BCH005323) 00888707001144,  1610AU (BCH005628) 00888707001144,  1610DE (BCH005772) 00888707001182,  1610DE (BCH005788) 00888707001182,  1610DE (BCH005798) 00888707001182,  1610DE (BCH005813) 00888707001182,  1610DE (BCH005835) 00888707001182,  1610DE (BCH005840) 00888707001182,  1610GB (BCH005445) 00888707001205,  1610DE (BCH005716) 00888707001182,  1610CA (BCH005748) 00888707001175,  1610DE (BCH005805) 00888707001182,  1610GB (BCH005832) 00888707001205,  1610DE (BCH006076) 00888707001182,  1610CA (BCH006118) 00888707001175,  1610DE (BCH005725) 00888707001182,  1610DE (BCH005829) 00888707001182,  1610DE (BCH005645) 00888707001182,  1610DE (BCH005714) 00888707001182,  1610DE (BCH005641) 00888707001182,  1610DE (BCH005734) 00888707001182,  1610IN (BCH005267) 00888707000536,  1610AU (BCH005544) 00888707001144,  1610AU (BCH005627) 00888707001144,  1610DE (BCH005674) 00888707001182,  1610DE (BCH005699) 00888707001182,  1610DE (BCH005709) 00888707001182,  1610DE (BCH005721) 00888707001182,  1610DE (BCH005736) 00888707001182,  1610DE (BCH005644) 00888707001182,  1610DE (BCH005647) 00888707001182,  1610DE (BCH005701) 00888707001182,  1610DE (BCH005738) 00888707001182,  1610DE (BCH005730) 00888707001182,  1610DE (BCH005731) 00888707001182,  1610DE (BCH005732) 00888707001182,  1610DE (BCH005741) 00888707001182,  1610CA (BCH005749) 00888707001175,  1610DE (BCH005797) 00888707001182,  1610DE (BCH005831) 00888707001182,  1610GB (BCH005531) 00888707001205,  1610DE (BCH005685) 00888707001182,  1610DE (BCH005686) 00888707001182,  1610DE (BCH005687) 00888707001182,  1610DE (BCH005729) 00888707001182,  1610DE (BCH005740) 00888707001182,  1610DE (BCH005754) 00888707001182,  1610DE (BCH005808) 00888707001182,  1610GB (BCH005839) 00888707001205,  1610DE (BCH006064) 00888707001182,  1610GB (BCH005331) 00888707001205,  1610DE (BCH005756) 00888707001182,  1610DE (BCH005776) 00888707001182,  1610DE (BCH005750) 00888707001182,  1610DE (BCH005639) 00888707001182,  1610CH (BCH005717) 00888707001199,  1610DE (BCH006063) 00888707001182,  1610DE (BCH005680) 00888707001182,  1610DE (BCH005771) 00888707001182,  1610DE (BCH005803) 00888707001182,  1610DE (BCH005838) 00888707001182,  1610AU (BCH005447) 00888707001144,  1610AU (BCH005461) 00888707001144,  1610AU (BCH005506) 00888707001144,  1610DE (BCH005778) 00888707001182,  1610DE (BCH006056) 00888707001182,  1610IN (BCH005336) 00888707000536,  1610DE (BCH005760) 00888707001182,  1610DE (BCH005762) 00888707001182,  1610DE (BCH005636) 00888707001182,  1610DE (BCH005640) 00888707001182,  1610DE (BCH005642) 00888707001182,  1610DE (BCH005318) 00888707001182,  1610GB (BCH005329) 00888707001205,  1610DE (BCH005753) 00888707001182,  1610DE (BCH005764) 00888707001182,  1610DE (BCH005769) 00888707001182,  1610DE (BCH005777) 00888707001182,  1610DE (BCH005823) 00888707001182,  1610DE (BCH005830) 00888707001182,  1610CA (BCH005733) 00888707001175,  1610DE (BCH005759) 00888707001182,  1610GB (BCH004357) 00888707001205,  1610IT (BCH005606) 00888707001212,  1610DE (BCH005646) 00888707001182,  1610DE (BCH005696) 00888707001182,  1610DE (BCH005707) 00888707001182,  1610DE (BCH005718) 00888707001182,  1610IT (BCH005726) 00888707001212,  1610DE (BCH005727) 00888707001182,  1610IT (BCH005735) 00888707001212,  1610IT (BCH005739) 00888707001212,  1610DE (BCH005712) 00888707001182,  1610DE (BCH005675) 00888707001182,  1610DE (BCH005679) 00888707001182,  1610CH (BCH005690) 00888707001199,  1610DE (BCH005700) 00888707001182,  1610DE (BCH005722) 00888707001182,  1610DE (BCH005737) 00888707001182,  1610DE (BCH005775) 00888707001182,  1610DE (BCH005698) 00888707001182,  1610DE (BCH005705) 00888707001182,  1610DE (BCH005715) 00888707001182,  1610GB (BCH004432) 00888707001205,  1610DE (BCH005637) 00888707001182,  1610DE (BCH005710) 00888707001182,  1610DE (BCH005724) 00888707001182,  1610DE (BCH005751) 00888707001182,  1610DE (BCH005614) 00888707001182,  1610DE (BCH005681) 00888707001182,  1610GB (BCH005837) 00888707001205,  1610DE (BCH005266) 00888707001182,  1610GB (BCH005327) 00888707001205,  1610AU (BCH005435) 00888707001144,  1610AU (BCH005505) 00888707001144,  1610DE (BCH005634) 00888707001182,  1610DE (BCH005648) 00888707001182,  1610DE (BCH005676) 00888707001182,  1610DE (BCH005677) 00888707001182,  1610DE (BCH005719) 00888707001182,  1610DE (BCH005761) 00888707001182,  1610DE (BCH005635) 00888707001182,  1610DE (BCH005649) 00888707001182,  1610DE (BCH005706) 00888707001182,  1610DE (BCH005742) 00888707001182,  1610DE (BCH004439) 00888707001182,  1610DE (BCH005328) 00888707001182,  1610DE (BCH005591) 00888707001182,  1610DE (BCH005618) 00888707001182,  1610DE (BCH005683) 00888707001182,  1610DE (BCH005684) 00888707001182,  1610DE (BCH005692) 00888707001182,  1610DE (BCH005694) 00888707001182,  1610DE (BCH005695) 00888707001182,  1610DE (BCH005704) 00888707001182,  1610DE (BCH005743) 00888707001182,  1610DE (BCH005747) 00888707001182,  1610GB (BCH005314) 00888707001205,  1610DE (BCH005638) 00888707001182,  1610DE (BCH005643) 00888707001182,  1610DE (BCH005292) 00888707001182,  1610DE (BCH005315) 00888707001182,  1610AU (BCH005434) 00888707001144,  1610DE (BCH005678) 00888707001182,  1610DE (BCH005688) 00888707001182,  1610DE (BCH005693) 00888707001182,  1610DE (BCH005702) 00888707001182,  1610DE (BCH005723) 00888707001182,  1610DE (BCH005604) 00888707001182,  1610DE (BCH004979) 00888707001182,  1610DE (BCH005305) 00888707001182,  1610GB (BCH005306) 00888707001205,  1610DE (BCH005313) 00888707001182,  1610DE (BCH005350) 00888707001182,  1610DE (BCH005317) 00888707001182,  1610DE (BCH005711) 00888707001182,  1610DE (BCH004314) 00888707001182,  1610IT (BCH004421) 00888707001212,  1610DE (BCH005333) 00888707001182,  1610IT (BCH005543) 00888707001212,  1610IT (BCH005596) 00888707001212,  1610IT (BCH005602) 00888707001212,  1610DE (BCH005703) 00888707001182,  1610DE (BCH004415) 00888707001182,  1610DE (BCH005274) 00888707001182,  1610DE (BCH005334) 00888707001182,  1610DE (BCH005548) 00888707001182,  1610DE (BCH005633) 00888707001182,  1610DE (BCH005682) 00888707001182,  1610DE (BCH004119) 00888707001182,  1610DE (BCH004019) 00888707001182,  1610DE (BCH004174) 00888707001182,  1610DE (BCH004320) 00888707001182,  1610DE (BCH004410) 00888707001182,  1610DE (BCH005276) 00888707001182,  1610DE (BCH005302) 00888707001182,  1610DE (BCH005322) 00888707001182,  1610DE (BCH005547) 00888707001182,  1610DE (BCH005307) 00888707001182,  1610DE (BCH004321) 00888707001182,  1610DE (BCH004438) 00888707001182,  1610DE (BCH005261) 00888707001182,  1610DE (BCH005310) 00888707001182,  1610DE (BCH005335) 00888707001182,  1610DE (BCH005744) 00888707001182,  1610DE (BCH005110) 00888707001182,  1610DE (BCH005278) 00888707001182,  1610DE (BCH005296) 00888707001182,  1610CH (BCH005338) 00888707001199,  1610GB (BCH005826) 00888707001205,  1610DE (BCH004293) 00888707001182,  1610DE (BCH005268) 00888707001182,  1610DE (BCH005291) 00888707001182,  1610DE (BCH005319) 00888707001182,  1610DE (BCH004319) 00888707001182,  1610DE (BCH004362) 00888707001182,  1610DE (BCH004434) 00888707001182,  1610DE (BCH005259) 00888707001182,  1610DE (BCH005269) 00888707001182,  1610DE (BCH005270) 00888707001182,  1610GB (BCH005557) 00888707001205,  1610DE (BCH004301) 00888707001182,  1610DE (BCH005320) 00888707001182,  1610CA (BCH005745) 00888707001175,  1610CH (BCH005277) 00888707001199,  1610IT (BCH005691) 00888707001212,  1610IT (BCH005708) 00888707001212,  1610DE (BCH005271) 00888707001182,  1610CH (BCH005286) 00888707001199,  1610DE (BCH005297) 00888707001182,  1610GB (BCH005551) 00888707001205,  1610CH (BCH004315) 00888707001199,  1610DE (BCH005260) 00888707001182,  1610DE (BCH005273) 00888707001182,  1610DE (BCH005720) 00888707001182,  1610DE (BCH005257) 00888707001182,  1610DE (BCH005289) 00888707001182,  1610DE (BCH005312) 00888707001182,  1610DE (BCH004309) 00888707001182,  1610DE (BCH004354) 00888707001182,  1610DE (BCH004390) 00888707001182,  1610DE (BCH005264) 00888707001182,  1610GB (BCH005567) 00888707001205,  1610DE (BCH004353) 00888707001182,  1610DE (BCH004391) 00888707001182,  1610DE (BCH005256) 00888707001182,  1610DE (BCH005262) 00888707001182,  1610DE (BCH005619) 00888707001182,  1610DE (BCH005300) 00888707001182,  1610DE (BCH005311) 00888707001182,  1610CH (BCH005416) 00888707001199,  1610GB (BCH005558) 00888707001205,  1610DE (BCH005275) 00888707001182,  1610DE (BCH005282) 00888707001182,  1610DE (BCH005284) 00888707001182,  1610DE (BCH005290) 00888707001182,  1610DE (BCH005293) 00888707001182,  1610DE (BCH005479) 00888707001182,  1610DE (BCH005545) 00888707001182,  1610DE (BCH005589) 00888707001182,  1610DE (BCH005594) 00888707001182,  1610DE (BCH005595) 00888707001182,  1610DE (BCH005624) 00888707001182,  1610DE (BCH005625) 00888707001182,  1610JP (BCH004194) 00888707001236,  1610DE (BCH005283) 00888707001182,  1610DE (BCH005613) 00888707001182,  1610IT (BCH005621) 00888707001212,  1610CA (BCH005288) 00888707001175,  1610CA (BCH005294) 00888707001175,  1610CA (BCH005295) 00888707001175,  1610CA (BCH005304) 00888707001175,  1610DE (BCH005460) 00888707001182,  1610DE (BCH005462) 00888707001182,  1610DE (BCH005495) 00888707001182,  1610DE (BCH005522) 00888707001182,  1610IT (BCH005532) 00888707001212,  1610DE (BCH005546) 00888707001182,  1610DE (BCH005599) 00888707001182,  1610IT (BCH004209) 00888707001212,  1610IT (BCH004307) 00888707001212,  1610IT (BCH004442) 00888707001212,  1610DE (BCH005483) 00888707001182,  1610DE (BCH005597) 00888707001182,  1610DE (BCH005280) 00888707001182,  1610DE (BCH005258) 00888707001182,  1610DE (BCH005455) 00888707001182,  1610DE (BCH005565) 00888707001182,  1610DE (BCH005629) 00888707001182,  1610DE (BCH005337) 00888707001182,  1610IT (BCH005525) 00888707001212,  1610IT (BCH005529) 00888707001212,  1610IT (BCH005542) 00888707001212,  1610DE (BCH005556) 00888707001182,  1610DE (BCH005569) 00888707001182,  1610DE (BCH005577) 00888707001182,  1610IT (BCH005588) 00888707001212,  1610DE (BCH005608) 00888707001182,  1610DE (BCH005610) 00888707001182,  1610DE (BCH004219) 00888707001182,  1610DE (BCH005573) 00888707001182,  1610DE (BCH005593) 00888707001182,  1610DE (BCH005281) 00888707001182,  1610GB (BCH005468) 00888707001205,  1610DE (BCH005555) 00888707001182,  1610DE (BCH005605) 00888707001182,  1610DE (BCH005299) 00888707001182,  1610DE (BCH005570) 00888707001182,  1610IN (BCH005459) 00888707000536,  1610AU (BCH005521) 00888707001144,  1610DE (BCH005572) 00888707001182,  1610DE (BCH005575) 00888707001182,  1610CA (BCH005586) 00888707001175,  1610DE (BCH005534) 00888707001182,  1610DE (BCH005578) 00888707001182,  1610DE (BCH005581) 00888707001182,  1610DE (BCH005582) 00888707001182,  1610DE (BCH005611) 00888707001182,  1610DE (BCH005612) 00888707001182,  1610DE (BCH004382) 00888707001182,  1610DE (BCH005265) 00888707001182,  1610DE (BCH005325) 00888707001182,  1610DE (BCH005533) 00888707001182,  1610DE (BCH005535) 00888707001182,  1610DE (BCH005549) 00888707001182,  1610DE (BCH005559) 00888707001182,  1610DE (BCH005592) 00888707001182,  1610DE (BCH005600) 00888707001182,  1610DE (BCH005601) 00888707001182,  1610DE (BCH005603) 00888707001182,  1610DE (BCH005609) 00888707001182,  1610DE (BCH005615) 00888707001182,  1610IN (BCH005623) 00888707000536,  1610DE (BCH005263) 00888707001182,  1610DE (BCH005360) 00888707001182,  1610DE (BCH005518) 00888707001182,  1610DE (BCH005626) 00888707001182,  1610DE (BCH005285) 00888707001182,  1610DE (BCH005298) 00888707001182,  1610DE (BCH005330) 00888707001182,  1610CA (BCH005478) 00888707001175,  1610DE (BCH005497) 00888707001182,  1610CH (BCH005520) 00888707001199,  1610DE (BCH005539) 00888707001182,  1610DE (BCH005553) 00888707001182,  1610DE (BCH005554) 00888707001182,  1610DE (BCH005620) 00888707001182,  1610IT (BCH005137) 00888707001212,  1610DE (BCH005272) 00888707001182,  1610DE (BCH005316) 00888707001182,  1610IT (BCH005387) 00888707001212,  1610IT (BCH005451) 00888707001212,  1610GB (BCH005104) 00888707001205,  1610DE (BCH005301) 00888707001182,  1610DE (BCH005583) 00888707001182,  1610CH (BCH005432) 00888707001199,  1610CH (BCH005474) 00888707001199,  1610DE (BCH005516) 00888707001182,  1610DE (BCH005580) 00888707001182,  1610DE (BCH005617) 00888707001182,  1610DE (BCH005562) 00888707001182,  1610DE (BCH005574) 00888707001182,  1610DE (BCH005616) 00888707001182,  1610DE (BCH005563) 00888707001182,  1610DE (BCH005488) 00888707001182,  1610DE (BCH005496) 00888707001182,  1610DE (BCH005514) 00888707001182,  1610DE (BCH005560) 00888707001182,  1610DE (BCH005576) 00888707001182,  1610CH (BCH005415) 00888707001199,  1610DE (BCH005441) 00888707001182,  1610DE (BCH005463) 00888707001182,  1610DE (BCH005464) 00888707001182,  1610DE (BCH005486) 00888707001182,  1610DE (BCH005492) 00888707001182,  1610DE (BCH005494) 00888707001182,  1610DE (BCH005501) 00888707001182,  1610DE (BCH005503) 00888707001182,  1610DE (BCH005515) 00888707001182,  1610DE (BCH005566) 00888707001182,  1610DE (BCH005517) 00888707001182,  1610GB (BCH005087) 00888707001205,  1610DE (BCH005122) 00888707001182,  1610DE (BCH005436) 00888707001182,  1610DE (BCH005437) 00888707001182,  1610DE (BCH005446) 00888707001182,  1610DE (BCH005472) 00888707001182,  1610DE (BCH005498) 00888707001182,  1610DE (BCH005590) 00888707001182,  1610DE (BCH005452) 00888707001182,  1610DE (BCH005485) 00888707001182,  1610DE (BCH005487) 00888707001182,  1610DE (BCH005561) 00888707001182,  1610DE (BCH005469) 00888707001182,  1610DE (BCH005473) 00888707001182,  1610IT (BCH005128) 00888707001212,  1610IT (BCH005422) 00888707001212,  1610GB (BCH005444) 00888707001205,  1610DE (BCH005512) 00888707001182,  1610DE (BCH005440) 00888707001182,  1610GB (BCH005069) 00888707001205,  1610DE (BCH005442) 00888707001182,  1610DE (BCH005530) 00888707001182,  1610DE (BCH005584) 00888707001182,  1610DE (BCH005622) 00888707001182,  1610GB (BCH005101) 00888707001205,  1610DE (BCH005119) 00888707001182,  1610DE (BCH005467) 00888707001182,  1610DE (BCH005457) 00888707001182,  1610DE (BCH005510) 00888707001182,  1610DE (BCH005519) 00888707001182,  1610DE (BCH005571) 00888707001182,  1610DE (BCH005372) 00888707001182,  1610DE (BCH005465) 00888707001182,  1610DE (BCH005470) 00888707001182,  1610DE (BCH005526) 00888707001182,  1610DE (BCH005086) 00888707001182,  1610DE (BCH005448) 00888707001182,  1610DE (BCH005456) 00888707001182,  1610DE (BCH005475) 00888707001182,  1610DE (BCH005490) 00888707001182,  1610DE (BCH005491) 00888707001182,  1610DE (BCH005504) 00888707001182,  1610DE (BCH005527) 00888707001182,  1610DE (BCH005528) 00888707001182,  1610DE (BCH005587) 00888707001182,  1610DE (BCH005438) 00888707001182,  1610DE (BCH005449) 00888707001182,  1610DE (BCH005454) 00888707001182,  1610GB (BCH005538) 00888707001205,  1610GB (BCH005540) 00888707001205,  1610GB (BCH005541) 00888707001205,  1610DE (BCH005466) 00888707001182,  1610CH (BCH005236) 00888707001199,  1610DE (BCH005120) 00888707001182,  1610DE (BCH005443) 00888707001182,  1610DE (BCH005500) 00888707001182,  1610DE (BCH005502) 00888707001182,  1610DE (BCH005096) 00888707001182,  1610CH (BCH005251) 00888707001199,  1610DE (BCH005354) 00888707001182,  1610DE (BCH005396) 00888707001182,  1610DE (BCH005385) 00888707001182,  1610DE (BCH005403) 00888707001182,  1610GB (BCH005131) 00888707001205,  1610GB (BCH005349) 00888707001205,  1610DE (BCH005355) 00888707001182,  1610IT (BCH005439) 00888707001212,  1610GB (BCH005450) 00888707001205,  1610DE (BCH005476) 00888707001182,  1610IT (BCH005477) 00888707001212,  1610DE (BCH005023) 00888707001182,  1610DE (BCH005552) 00888707001182,  1610GB (BCH005568) 00888707001205,  1610CA (BCH005345) 00888707001175,  1610DE (BCH005374) 00888707001182,  1610DE (BCH005480) 00888707001182,  1610DE (BCH005489) 00888707001182,  1610DE (BCH005630) 00888707001182,  1610DE (BCH005523) 00888707001182,  1610DE (BCH005499) 00888707001182,  1610DE (BCH005424) 00888707001182,  1610DE (BCH005401) 00888707001182,  1610DE (BCH005433) 00888707001182,  1610DE (BCH005458) 00888707001182,  1610DE (BCH005481) 00888707001182,  1610DE (BCH005482) 00888707001182,  1610DE (BCH005484) 00888707001182,  1610DE (BCH005507) 00888707001182,  1610DE (BCH005585) 00888707001182,  1610CH (BCH005348) 00888707001199,  1610DE (BCH005493) 00888707001182,  1610DE (BCH005509) 00888707001182,  1610DE (BCH005513) 00888707001182,  1610DE (BCH005536) 00888707001182,  1610GB (BCH005089) 00888707001205,  1610DE (BCH005511) 00888707001182,  1610DE (BCH005579) 00888707001182,  1610IT (BCH005037) 00888707001212,  1610IN (BCH005426) 00888707000536,  1610DE (BCH005393) 00888707001182,  1610DE (BCH005117) 00888707001182,  1610DE (BCH005129) 00888707001182,  1610DE (BCH005368) 00888707001182,  1610AU (BCH005357) 00888707001144,  1610DE (BCH005362) 00888707001182,  1610AU (BCH005381) 00888707001144,  1610AU (BCH005417) 00888707001144,  1610IT (BCH005564) 00888707001212,  1610IT (BCH004923) 00888707001212,  1610DE (BCH005081) 00888707001182,  1610DE (BCH005097) 00888707001182,  1610DE (BCH005098) 00888707001182,  1610DE (BCH005107) 00888707001182,  1610AU (BCH005116) 00888707001144,  1610DE (BCH005118) 00888707001182,  1610DE (BCH005121) 00888707001182,  1610AU (BCH005351) 00888707001144,  1610AU (BCH005352) 00888707001144,  1610IT (BCH005358) 00888707001212,  1610DE (BCH005373) 00888707001182,  1610DE (BCH005375) 00888707001182,  1610DE (BCH005376) 00888707001182,  1610DE (BCH005378) 00888707001182,  1610DE (BCH005388) 00888707001182,  1610DE (BCH005389) 00888707001182,  1610DE (BCH005425) 00888707001182,  1610CH (BCH004879) 00888707001199,  1610DE (BCH005342) 00888707001182,  1610DE (BCH005343) 00888707001182,  1610DE (BCH005346) 00888707001182,  1610DE (BCH005365) 00888707001182,  1610DE (BCH005366) 00888707001182,  1610DE (BCH005394) 00888707001182,  1610DE (BCH005407) 00888707001182,  1610CA (BCH005550) 00888707001175,  1610DE (BCH005079) 00888707001182,  1610DE (BCH005344) 00888707001182,  1610DE (BCH005078) 00888707001182,  1610DE (BCH005093) 00888707001182,  1610DE (BCH005429) 00888707001182,  1610CA (BCH005041) 00888707001175,  1610DE (BCH005066) 00888707001182,  1610DE (BCH005083) 00888707001182,  1610GB (BCH005085) 00888707001205,  1610CA (BCH005453) 00888707001175,  1610DE (BCH005072) 00888707001182,  1610DE (BCH005115) 00888707001182,  1610DE (BCH005130) 00888707001182,  1610DE (BCH005356) 00888707001182,  1610DE (BCH005367) 00888707001182,  1610DE (BCH005413) 00888707001182,  1610DE (BCH005430) 00888707001182,  1610IT (BCH004889) 00888707001212,  1610AU (BCH005068) 00888707001144,  1610IT (BCH005112) 00888707001212,  1610DE (BCH005364) 00888707001182,  1610DE (BCH005382) 00888707001182,  1610DE (BCH005406) 00888707001182,  1610CA (BCH005418) 00888707001175,  1610DE (BCH005423) 00888707001182,  1610DE (BCH005508) 00888707001182,  1610DE (BCH005082) 00888707001182,  1610GB (BCH005084) 00888707001205,  1610DE (BCH005384) 00888707001182,  1610DE (BCH005404) 00888707001182,  1610DE (BCH005412) 00888707001182,  1610GB (BCH005248) 00888707001205,  1610DE (BCH005370) 00888707001182,  1610DE (BCH005400) 00888707001182,  1610DE (BCH005410) 00888707001182,  1610CH (BCH005040) 00888707001199,  1610DE (BCH005369) 00888707001182,  1610CH (BCH005033) 00888707001199,  1610DE (BCH005095) 00888707001182,  1610DE (BCH005141) 00888707001182,  1610GB (BCH005252) 00888707001205,  1610DE (BCH005390) 00888707001182,  1610DE (BCH005402) 00888707001182,  1610GB (BCH005088) 00888707001205,  1610DE (BCH005353) 00888707001182,  1610DE (BCH005386) 00888707001182,  1610GB (BCH004815) 00888707001205,  1610GB (BCH005230) 00888707001205,  1610DE (BCH005241) 00888707001182,  1610DE (BCH005363) 00888707001182,  1610DE (BCH005391) 00888707001182,  1610DE (BCH005409) 00888707001182,  1610DE (BCH005411) 00888707001182,  1610DE (BCH005428) 00888707001182,  1610DE (BCH005431) 00888707001182,  1610DE (BCH005092) 00888707001182,  1610DE (BCH005124) 00888707001182,  1610DE (BCH005380) 00888707001182,  1610DE (BCH005395) 00888707001182,  1610DE (BCH005427) 00888707001182,  1610IT (BCH004945) 00888707001212,  1610IT (BCH005028) 00888707001212,  1610DE (BCH005132) 00888707001182,  1610DE (BCH005240) 00888707001182,  1610GB (BCH005244) 00888707001205,  1610DE (BCH005090) 00888707001182,  1610DE (BCH005091) 00888707001182,  1610CH (BCH005208) 00888707001199,  1610GB (BCH005029) 00888707001205,  1610GB (BCH005051) 00888707001205,  1610DE (BCH005075) 00888707001182,  1610DE (BCH005109) 00888707001182,  1610DE (BCH005399) 00888707001182,  1610IN (BCH004924) 00888707000536,  1610CA (BCH004946) 00888707001175,  1610DE (BCH005125) 00888707001182,  1610IL (BCH005227) 00888707001229,  1610IT (BCH004913) 00888707001212,  1610IT (BCH005015) 00888707001212,  1610DE (BCH005076) 00888707001182,  1610DE (BCH005106) 00888707001182,  1610DE (BCH005379) 00888707001182,  1610DE (BCH005105) 00888707001182,  1610AU (BCH005136) 00888707001144,  1610CA (BCH005347) 00888707001175,  1610DE (BCH005073) 00888707001182,  1610CA (BCH005359) 00888707001175,  1610DE (BCH005250) 00888707001182,  1610DE (BCH005408) 00888707001182,  1610DE (BCH005070) 00888707001182,  1610DE (BCH005074) 00888707001182,  1610DE (BCH005094) 00888707001182,  1610DE (BCH005099) 00888707001182,  1610DE (BCH005133) 00888707001182,  1610DE (BCH005134) 00888707001182,  1610DE (BCH005144) 00888707001182,  1610CA (BCH005233) 00888707001175,  1610AU (BCH005067) 00888707001144,  1610DE (BCH005111) 00888707001182,  1610DE (BCH005232) 00888707001182,  1610DE (BCH005223) 00888707001182,  1610DE (BCH005371) 00888707001182,  1610DE (BCH005377) 00888707001182,  1610DE (BCH004310) 00888707001182,  1610DE (BCH005071) 00888707001182,  1610DE (BCH005100) 00888707001182,  1610DE (BCH005102) 00888707001182,  1610DE (BCH005103) 00888707001182,  1610DE (BCH005148) 00888707001182,  1610DE (BCH005361) 00888707001182,  1610DE (BCH005145) 00888707001182,  1610DE (BCH005221) 00888707001182,  1610DE (BCH004425) 00888707001182,  1610DE (BCH005114) 00888707001182,  1610DE (BCH005123) 00888707001182,  1610DE (BCH005113) 00888707001182,  1610DE (BCH005135) 00888707001182,  1610DE (BCH005179) 00888707001182,  1610DE (BCH005182) 00888707001182,  1610DE (BCH005199) 00888707001182,  1610DE (BCH005219) 00888707001182,  1610DE (BCH005253) 00888707001182,  1610DE (BCH004339) 00888707001182,  1610AU (BCH005077) 00888707001144,  1610DE (BCH005191) 00888707001182,  1610DE (BCH005198) 00888707001182,  1610AU (BCH005225) 00888707001144,  1610AU (BCH005226) 00888707001144,  1610DE (BCH005237) 00888707001182,  1610DE (BCH004389) 00888707001182,  1610IT (BCH004881) 00888707001212,  1610DE (BCH005196) 00888707001182,  1610DE (BCH005197) 00888707001182,  1610DE (BCH004304) 00888707001182,  1610DE (BCH004327) 00888707001182,  1610DE (BCH005217) 00888707001182,  1610CA (BCH005218) 00888707001175,  1610GB (BCH005229) 00888707001205,  1610GB (BCH004960) 00888707001205,  1610DE (BCH004274) 00888707001182,  1610DE (BCH004306) 00888707001182,  1610DE (BCH004393) 00888707001182,  1610DE (BCH004406) 00888707001182,  1610GB (BCH005059) 00888707001205,  1610DE (BCH005149) 00888707001182,  1610DE (BCH005222) 00888707001182,  1610DE (BCH005247) 00888707001182,  1610DE (BCH005176) 00888707001182,  1610DE (BCH005192) 00888707001182,  1610DE (BCH005204) 00888707001182,  1610DE (BCH005205) 00888707001182,  1610DE (BCH005238) 00888707001182,  1610DE (BCH005126) 00888707001182,  1610DE (BCH005127) 00888707001182,  1610DE (BCH005157) 00888707001182,  1610DE (BCH005159) 00888707001182,  1610DE (BCH005180) 00888707001182,  1610DE (BCH005185) 00888707001182,  1610GB (BCH004942) 00888707001205,  1610DE (BCH005215) 00888707001182,  1610AU (BCH005216) 00888707001144,  1610AU (BCH005245) 00888707001144,  1610DE (BCH005151) 00888707001182,  1610DE (BCH005168) 00888707001182,  1610DE (BCH005181) 00888707001182,  1610DE (BCH005183) 00888707001182,  1610DE (BCH005195) 00888707001182,  1610DE (BCH005201) 00888707001182,  1610DE (BCH005211) 00888707001182,  1610DE (BCH005212) 00888707001182,  1610DE (BCH005214) 00888707001182,  1610DE (BCH005405) 00888707001182,  1610DE (BCH005152) 00888707001182,  1610DE (BCH005175) 00888707001182,  1610DE (BCH005177) 00888707001182,  1610DE (BCH005178) 00888707001182,  1610DE (BCH005189) 00888707001182,  1610DE (BCH005190) 00888707001182,  1610DE (BCH005213) 00888707001182,  1610GB (BCH004888) 00888707001205,  1610DE (BCH005019) 00888707001182,  1610CH (BCH005140) 00888707001199,  1610DE (BCH005234) 00888707001182,  1610DE (BCH005239) 00888707001182,  1610DE (BCH005080) 00888707001182,  1610DE (BCH005202) 00888707001182,  1610IT (BCH004909) 00888707001212,  1610DE (BCH004915) 00888707001182,  1610IT (BCH004934) 00888707001212,  1610DE (BCH005150) 00888707001182,  1610AU (BCH005160) 00888707001144,  1610DE (BCH005209) 00888707001182,  1610AU (BCH005246) 00888707001144,  1610DE (BCH005038) 00888707001182,  1610DE (BCH005052) 00888707001182,  1610DE (BCH005162) 00888707001182,  1610DE (BCH005163) 00888707001182,  1610GB (BCH004984) 00888707001205,  1610DE (BCH005046) 00888707001182,  1610DE (BCH005060) 00888707001182,  1610DE (BCH005146) 00888707001182,  1610DE (BCH005153) 00888707001182,  1610DE (BCH005154) 00888707001182,  1610DE (BCH005164) 00888707001182,  1610DE (BCH005169) 00888707001182,  1610DE (BCH005186) 00888707001182,  1610CA (BCH005220) 00888707001175,  1610IT (BCH004920) 00888707001212,  1610IT (BCH004958) 00888707001212,  1610DE (BCH005138) 00888707001182,  1610DE (BCH005184) 00888707001182,  1610DE (BCH005165) 00888707001182,  1610DE (BCH005167) 00888707001182,  1610DE (BCH005170) 00888707001182,  1610DE (BCH005172) 00888707001182,  1610DE (BCH005173) 00888707001182,  1610DE (BCH004927) 00888707001182,  1610GB (BCH004982) 00888707001205,  1610AU (BCH005020) 00888707001144,  1610AU (BCH005056) 00888707001144,  1610DE (BCH005053) 00888707001182,  1610DE (BCH005142) 00888707001182,  1610DE (BCH005143) 00888707001182,  1610DE (BCH005166) 00888707001182,  1610DE (BCH005174) 00888707001182,  1610DE (BCH005187) 00888707001182,  1610AU (BCH005188) 00888707001144,  1610DE (BCH005200) 00888707001182,  1610DE (BCH005207) 00888707001182,  1610DE (BCH004802) 00888707001182,  1610DE (BCH004804) 00888707001182,  1610DE (BCH004808) 00888707001182,  1610DE (BCH004843) 00888707001182,  1610DE (BCH004848) 00888707001182,  1610DE (BCH004852) 00888707001182,  1610DE (BCH004869) 00888707001182,  1610IT (BCH004921) 00888707001212,  1610DE (BCH004931) 00888707001182,  1610IT (BCH004937) 00888707001212,  1610DE (BCH005036) 00888707001182,  1610DE (BCH005047) 00888707001182,  1610DE (BCH005063) 00888707001182,  1610DE (BCH005158) 00888707001182,  1610DE (BCH005193) 00888707001182,  1610DE (BCH005203) 00888707001182,  1610DE (BCH005206) 00888707001182,  1610DE (BCH005031) 00888707001182,  1610DE (BCH005042) 00888707001182,  1610DE (BCH005049) 00888707001182,  1610DE (BCH005050) 00888707001182,  1610GB (BCH004976) 00888707001205,  1610DE (BCH005061) 00888707001182,  1610DE (BCH004806) 00888707001182,  1610DE (BCH005057) 00888707001182,  1610DE (BCH005147) 00888707001182,  1610DE (BCH005156) 00888707001182,  1610GB (BCH004231) 00888707001205,  1610GB (BCH004545) 00888707001205,  1610DE (BCH004819) 00888707001182,  1610DE (BCH005021) 00888707001182,  1610DE (BCH005022) 00888707001182,  1610DE (BCH005048) 00888707001182,  1610DE (BCH005210) 00888707001182,  1610CH (BCH004981) 00888707001199,  1610AU (BCH005155) 00888707001144,  1610DE (BCH005058) 00888707001182,  1610DE (BCH005062) 00888707001182,  1610DE (BCH004463) 00888707001182,  1610DE (BCH004828) 00888707001182,  1610GB (BCH004922) 00888707001205,  1610DE (BCH005016) 00888707001182,  1610DE (BCH005018) 00888707001182,  1610DE (BCH005024) 00888707001182,  1610DE (BCH005055) 00888707001182,  1610DE (BCH005139) 00888707001182,  1610DE (BCH005025) 00888707001182,  1610DE (BCH005035) 00888707001182,  1610DE (BCH005194) 00888707001182,  1610CA (BCH005235) 00888707001175,  1610DE (BCH004867) 00888707001182,  1610DE (BCH004936) 00888707001182,  1610DE (BCH005006) 00888707001182,  1610DE (BCH005054) 00888707001182,  1610DE (BCH004383) 00888707001182,  1610DE (BCH005030) 00888707001182,  1610AU (BCH004629) 00888707001144,  1610AU (BCH004657) 00888707001144,  1610DE (BCH004821) 00888707001182,  1610DE (BCH004834) 00888707001182,  1610IN (BCH005026) 00888707000536,  1610DE (BCH005171) 00888707001182,  1610CA (BCH005242) 00888707001175,  1610CA (BCH005249) 00888707001175,  1610DE (BCH004823) 00888707001182,  1610DE (BCH004829) 00888707001182,  1610DE (BCH004849) 00888707001182,  1610DE (BCH005027) 00888707001182,  1610DE (BCH005044) 00888707001182,  1610DE (BCH004820) 00888707001182,  1610DE (BCH005009) 00888707001182,  1610CH (BCH005034) 00888707001199,  1610DE (BCH004218) 00888707001182,  1610AU (BCH004308) 00888707001144,  1610AU (BCH004547) 00888707001144,  1610DE (BCH004830) 00888707001182,  1610DE (BCH004926) 00888707001182,  1610DE (BCH004983) 00888707001182,  1610CA (BCH005032) 00888707001175,  1610CA (BCH005161) 00888707001175,  1610DE (BCH004933) 00888707001182,  1610DE (BCH004947) 00888707001182,  1610DE (BCH004818) 00888707001182,  1610GB (BCH004895) 00888707001205,  1610CH (BCH004939) 00888707001199,  1610DE (BCH005043) 00888707001182,  1610DE (BCH004807) 00888707001182,  1610DE (BCH004894) 00888707001182,  1610CH (BCH004971) 00888707001199,  1610GB (BCH004980) 00888707001205,  1610DE (BCH004824) 00888707001182,  1610DE (BCH004835) 00888707001182,  1610IN (BCH005004) 00888707000536,  1610DE (BCH005017) 00888707001182,  1610DE (BCH004851) 00888707001182,  1610CA (BCH004898) 00888707001175,  1610DE (BCH004827) 00888707001182,  1610GB (BCH004970) 00888707001205,  1610DE (BCH005011) 00888707001182,  1610DE (BCH005039) 00888707001182,  1610DE (BCH004832) 00888707001182,  1610DE (BCH004880) 00888707001182,  1610DE (BCH004882) 00888707001182,  1610DE (BCH004883) 00888707001182,  1610DE (BCH004893) 00888707001182,  1610DE (BCH004905) 00888707001182,  1610DE (BCH004910) 00888707001182,  1610DE (BCH004918) 00888707001182,  1610DE (BCH004932) 00888707001182,  1610DE (BCH004938) 00888707001182,  1610DE (BCH004940) 00888707001182,  1610DE (BCH004941) 00888707001182,  1610DE (BCH004949) 00888707001182,  1610DE (BCH004951) 00888707001182,  1610DE (BCH004953) 00888707001182,  1610DE (BCH004955) 00888707001182,  1610DE (BCH004959) 00888707001182,  1610DE (BCH004968) 00888707001182,  1610DE (BCH004977) 00888707001182,  1610DE (BCH004985) 00888707001182,  1610DE (BCH004999) 00888707001182,  1610DE (BCH005000) 00888707001182,  1610DE (BCH005002) 00888707001182,  1610DE (BCH005005) 00888707001182,  1610DE (BCH005007) 00888707001182,  1610DE (BCH004902) 00888707001182,  1610DE (BCH004917) 00888707001182,  1610GB (BCH005008) 00888707001205,  1610DE (BCH004916) 00888707001182,  1610DE (BCH004956) 00888707001182,  1610DE (BCH004822) 00888707001182,  1610DE (BCH004886) 00888707001182,  1610DE (BCH004907) 00888707001182,  1610DE (BCH004962) 00888707001182,  1610DE (BCH004963) 00888707001182,  1610DE (BCH004836) 00888707001182,  1610DE (BCH004965) 00888707001182,  1610DE (BCH005014) 00888707001182,  1610GB (BCH004801) 00888707001205,  1610DE (BCH004814) 00888707001182,  1610DE (BCH004833) 00888707001182,  1610DE (BCH004857) 00888707001182,  1610DE (BCH004900) 00888707001182,  1610DE (BCH004908) 00888707001182,  1610DE (BCH004911) 00888707001182,  1610DE (BCH004914) 00888707001182,  1610DE (BCH004929) 00888707001182,  1610DE (BCH004930) 00888707001182,  1610DE (BCH004943) 00888707001182,  1610DE (BCH004948) 00888707001182,  1610DE (BCH004969) 00888707001182,  1610DE (BCH004973) 00888707001182,  1610AU (BCH004632) 00888707001144,  1610DE (BCH004887) 00888707001182,  1610CH (BCH004897) 00888707001199,  1610DE (BCH004901) 00888707001182,  1610DE (BCH005010) 00888707001182,  1610DE (BCH004944) 00888707001182,  1610DE (BCH004954) 00888707001182,  1610DE (BCH004975) 00888707001182,  1610DE (BCH005012) 00888707001182,  1610DE (BCH005013) 00888707001182,  1610DE (BCH004846) 00888707001182,  1610CH (BCH004961) 00888707001199,  1610DE (BCH004356) 00888707001182,  1610DE (BCH004896) 00888707001182,  1610DE (BCH004842) 00888707001182,  1610DE (BCH004906) 00888707001182,  1610DE (BCH004964) 00888707001182,  1610DE (BCH004967) 00888707001182,  1610DE (BCH004811) 00888707001182,  1610DE (BCH004826) 00888707001182,  1610DE (BCH004831) 00888707001182,  1610DE (BCH004861) 00888707001182,  1610DE (BCH005001) 00888707001182,  1610DE (BCH004206) 00888707001182,  1610DE (BCH004825) 00888707001182,  1610DE (BCH004847) 00888707001182,  1610CH (BCH004904) 00888707001199,  1610DE (BCH004837) 00888707001182,  1610DE (BCH004839) 00888707001182,  1610DE (BCH004871) 00888707001182,  1610DE (BCH004844) 00888707001182,  1610DE (BCH004845) 00888707001182,  1610DE (BCH004854) 00888707001182,  1610DE (BCH004860) 00888707001182,  1610DE (BCH004872) 00888707001182,  1610DE (BCH004817) 00888707001182,  1610DE (BCH004865) 00888707001182,  1610DE (BCH004866) 00888707001182,  1610DE (BCH004870) 00888707001182,  1610DE (BCH004885) 00888707001182,  1610DE (BCH004858) 00888707001182,  1610IT (BCH004890) 00888707001212,  1610IT (BCH004903) 00888707001212,  1610IT (BCH004925) 00888707001212,  1610DE (BCH004862) 00888707001182,  1610DE (BCH004863) 00888707001182,  1610DE (BCH004874) 00888707001182,  1610DE (BCH004891) 00888707001182,  1610DE (BCH004912) 00888707001182,  1610DE (BCH004974) 00888707001182,  1610DE (BCH004838) 00888707001182,  1610DE (BCH004856) 00888707001182,  1610DE (BCH004864) 00888707001182,  1610DE (BCH003836) 00888707001182,  1610DE (BCH004189) 00888707001182,  1610DE (BCH004199) 00888707001182,  1610DE (BCH004208) 00888707001182,  1610DE (BCH004210) 00888707001182,  1610DE (BCH004338) 00888707001182,  1610DE (BCH004366) 00888707001182,  1610DE (BCH004429) 00888707001182,  1610DE (BCH004477) 00888707001182,  1610DE (BCH004499) 00888707001182,  1610DE (BCH004803) 00888707001182,  1610DE (BCH004805) 00888707001182,  1610DE (BCH004859) 00888707001182,  1610GB (BCH004972) 00888707001205,  1610GB (BCH004503) 00888707001205,  1610DE (BCH004812) 00888707001182,  1610DE (BCH004892) 00888707001182,  1610DE (BCH004810) 00888707001182,  1610DE (BCH004873) 00888707001182,  1610DE (BCH004950) 00888707001182,  1610DE (BCH004176) 00888707001182,  1610CH (BCH004187) 00888707001199,  1610DE (BCH004809) 00888707001182,  1610DE (BCH004853) 00888707001182,  1610DE (BCH004875) 00888707001182,  1610DE (BCH004841) 00888707001182,  1610DE (BCH004850) 00888707001182,  1610DE (BCH004855) 00888707001182,  1610CH (BCH004215) 00888707001199,  1610DE (BCH004816) 00888707001182,  1610DE (BCH004840) 00888707001182,  1610DE (BCH003884) 00888707001182,  1610DE (BCH004223) 00888707001182,  1610DE (BCH004400) 00888707001182,  1610DE (BCH004146) 00888707001182,  1610DE (BCH004216) 00888707001182,  1610GB (BCH004229) 00888707001205,  1610DE (BCH004238) 00888707001182,  1610DE (BCH004299) 00888707001182,  1610DE (BCH004371) 00888707001182,  1610DE (BCH004414) 00888707001182,  1610DE (BCH004419) 00888707001182,  1610DE (BCH004420) 00888707001182,  1610DE (BCH004104) 00888707001182,  1610DE (BCH004184) 00888707001182,  1610DE (BCH004234) 00888707001182,  1610DE (BCH004297) 00888707001182,  1610DE (BCH004196) 00888707001182,  1610DE (BCH004235) 00888707001182,  1610DE (BCH004286) 00888707001182,  1610DE (BCH004324) 00888707001182,  1610DE (BCH004355) 00888707001182,  1610CA (BCH003987) 00888707001175,  1610CA (BCH003997) 00888707001175,  1610DE (BCH004121) 00888707001182,  1610DE (BCH004179) 00888707001182,  1610DE (BCH004349) 00888707001182,  1610DE (BCH004106) 00888707001182,  1610DE (BCH004180) 00888707001182,  1610DE (BCH004203) 00888707001182,  1610GB (BCH004221) 00888707001205,  1610CH (BCH004224) 00888707001199,  1610DE (BCH004279) 00888707001182,  1610DE (BCH004342) 00888707001182,  1610DE (BCH004475) 00888707001182,  1610DE (BCH004173) 00888707001182,  1610DE (BCH004352) 00888707001182,  1610DE (BCH004298) 00888707001182,  1610DE (BCH004333) 00888707001182,  1610DE (BCH004346) 00888707001182,  1610DE (BCH004335) 00888707001182,  1610DE (BCH004120) 00888707001182,  1610DE (BCH004345) 00888707001182,  1610GB (BCH004365) 00888707001205,  1610DE (BCH004471) 00888707001182,  1610GB (BCH004658) 00888707001205,  1610DE (BCH004172) 00888707001182,  1610DE (BCH004233) 00888707001182,  1610DE (BCH004222) 00888707001182,  1610DE (BCH004289) 00888707001182,  1610DE (BCH004401) 00888707001182,  1610DE (BCH004192) 00888707001182,  1610CH (BCH004217) 00888707001199,  1610DE (BCH004232) 00888707001182,  1610DE (BCH004239) 00888707001182,  1610DE (BCH004322) 00888707001182,  1610DE (BCH004325) 00888707001182,  1610AU (BCH004402) 00888707001144,  1610DE (BCH004110) 00888707001182,  1610AU (BCH004123) 00888707001144,  1610DE (BCH004276) 00888707001182,  1610GB (BCH004369) 00888707001205,  1610DE (BCH004100) 00888707001182,  1610DE (BCH004116) 00888707001182,  1610DE (BCH004181) 00888707001182,  1610GB (BCH004413) 00888707001205,  1610IN (BCH004508) 00888707000536,  1610DE (BCH004200) 00888707001182,  1610DE (BCH004109) 00888707001182,  1610AU (BCH004191) 00888707001144,  1610AU (BCH004240) 00888707001144,  1610DE (BCH004267) 00888707001182,  1610DE (BCH004273) 00888707001182,  1610AU (BCH004277) 00888707001144,  1610DE (BCH004287) 00888707001182,  1610DE (BCH004101) 00888707001182,  1610AU (BCH004198) 00888707001144,  1610DE (BCH004220) 00888707001182,  1610DE (BCH004403) 00888707001182,  1610DE (BCH004102) 00888707001182,  1610DE (BCH004108) 00888707001182,  1610DE (BCH004237) 00888707001182,  1610DE (BCH004275) 00888707001182,  1610DE (BCH004311) 00888707001182,  1610DE (BCH004326) 00888707001182,  1610DE (BCH004334) 00888707001182,  1610DE (BCH004344) 00888707001182,  1610DE (BCH004347) 00888707001182,  1610DE (BCH004381) 00888707001182,  1610DE (BCH004175) 00888707001182,  1610GB (BCH004212) 00888707001205,  1610DE (BCH004205) 00888707001182,  1610IN (BCH003852) 00888707000536,  1610DE (BCH004202) 00888707001182,  1610DE (BCH004378) 00888707001182,  1610DE (BCH004380) 00888707001182,  1610DE (BCH004099) 00888707001182,  1610DE (BCH004118) 00888707001182,  1610DE (BCH004183) 00888707001182,  1610DE (BCH004207) 00888707001182,  1610DE (BCH004328) 00888707001182,  1610GB (BCH004098) 00888707001205,  1610DE (BCH004440) 00888707001182,  1610GB (BCH004061) 00888707001205,  1610DE (BCH004178) 00888707001182,  1610DE (BCH004340) 00888707001182,  1610DE (BCH004350) 00888707001182,  1610DE (BCH004361) 00888707001182,  1610DE (BCH004363) 00888707001182,  1610DE (BCH004182) 00888707001182,  1610DE (BCH004296) 00888707001182,  1610DE (BCH004341) 00888707001182,  1610DE (BCH004405) 00888707001182,  1610DE (BCH004437) 00888707001182,  1610DE (BCH004318) 00888707001182,  1610DE (BCH003860) 00888707001182,  1610DE (BCH004103) 00888707001182,  1610DE (BCH004226) 00888707001182,  1610DE (BCH004430) 00888707001182,  1610DE (BCH004448) 00888707001182,  1610DE (BCH003871) 00888707001182,  1610IT (BCH004188) 00888707001212,  1610IT (BCH004190) 00888707001212,  1610IT (BCH004193) 00888707001212,  1610IT (BCH004201) 00888707001212,  1610GB (BCH004083) 00888707001205,  1610DE (BCH004117) 00888707001182,  1610DE (BCH004230) 00888707001182,  1610DE (BCH004480) 00888707001182,  1610DE (BCH004482) 00888707001182,  1610AU (BCH004653) 00888707001144,  1610DE (BCH004105) 00888707001182,  1610DE (BCH004214) 00888707001182,  1610DE (BCH004364) 00888707001182,  1610DE (BCH004423) 00888707001182,  1610DE (BCH004241) 00888707001182,  1610DE (BCH003845) 00888707001182,  1610DE (BCH004122) 00888707001182,  1610DE (BCH004195) 00888707001182,  1610DE (BCH004197) 00888707001182,  1610DE (BCH004368) 00888707001182,  1610DE (BCH004370) 00888707001182,  1610DE (BCH004392) 00888707001182,  1610DE (BCH004411) 00888707001182,  1610DE (BCH004418) 00888707001182,  1610DE (BCH004441) 00888707001182,  1610DE (BCH004465) 00888707001182,  1610DE (BCH004488) 00888707001182,  1610DE (BCH004489) 00888707001182,  1610GB (BCH003609) 00888707001205,  1610DE (BCH004185) 00888707001182,  1610DE (BCH004213) 00888707001182,  1610DE (BCH004331) 00888707001182,  1610DE (BCH004336) 00888707001182,  1610DE (BCH004375) 00888707001182,  1610DE (BCH004379) 00888707001182,  1610DE (BCH004227) 00888707001182,  1610DE (BCH004376) 00888707001182,  1610DE (BCH004394) 00888707001182,  1610DE (BCH004435) 00888707001182,  1610DE (BCH003224) 00888707001182,  1610DE (BCH004186) 00888707001182,  1610DE (BCH004330) 00888707001182,  1610DE (BCH004409) 00888707001182,  1610DE (BCH004204) 00888707001182,  1610DE (BCH004323) 00888707001182,  1610DE (BCH004373) 00888707001182,  1610DE (BCH004384) 00888707001182,  1610DE (BCH004426) 00888707001182,  1610DE (BCH004457) 00888707001182,  1610CH (BCH004460) 00888707001199,  1610DE (BCH003850) 00888707001182,  1610DE (BCH004329) 00888707001182,  1610DE (BCH004360) 00888707001182,  1610GB (BCH004433) 00888707001205,  1610DE (BCH004228) 00888707001182,  1610GB (BCH004536) 00888707001205,  1610DE (BCH003851) 00888707001182,  1610DE (BCH004124) 00888707001182,  1610DE (BCH004294) 00888707001182,  1610DE (BCH004332) 00888707001182,  1610DE (BCH004359) 00888707001182,  1610DE (BCH004424) 00888707001182,  1610DE (BCH004461) 00888707001182,  1610DE (BCH004476) 00888707001182,  1610GB (BCH004071) 00888707001205,  1610GB (BCH004091) 00888707001205,  1610CA (BCH004092) 00888707001175,  1610DE (BCH004337) 00888707001182,  1610DE (BCH004372) 00888707001182,  1610DE (BCH004404) 00888707001182,  1610GB (BCH004539) 00888707001205,  1610GB (BCH004606) 00888707001205,  1610GB (BCH003998) 00888707001205,  1610DE (BCH004358) 00888707001182,  1610DE (BCH004450) 00888707001182,  1610DE (BCH004484) 00888707001182,  1610DE (BCH003750) 00888707001182,  1610DE (BCH003867) 00888707001182,  1610DE (BCH004386) 00888707001182,  1610IT (BCH004436) 00888707001212,  1610IN (BCH004561) 00888707000536,  1610DE (BCH004626) 00888707001182,  1610DE (BCH003869) 00888707001182,  1610DE (BCH004302) 00888707001182,  1610DE (BCH004485) 00888707001182,  1610CH (BCH003810) 00888707001199,  1610AU (BCH004343) 00888707001144,  1610DE (BCH004367) 00888707001182,  1610DE (BCH004447) 00888707001182,  1610DE (BCH004492) 00888707001182,  1610DE (BCH003872) 00888707001182,  1610DE (BCH004348) 00888707001182,  1610DE (BCH004388) 00888707001182,  1610CH (BCH004407) 00888707001199,  1610DE (BCH004416) 00888707001182,  1610DE (BCH004481) 00888707001182,  1610DE (BCH003692) 00888707001182,  1610IT (BCH003726) 00888707001212,  1610DE (BCH003844) 00888707001182,  1610IT (BCH004351) 00888707001212,  1610IT (BCH004452) 00888707001212,  1610DE (BCH004455) 00888707001182,  1610DE (BCH004469) 00888707001182,  1610GB (BCH004654) 00888707001205,  1610DE (BCH004493) 00888707001182,  1610DE (BCH004458) 00888707001182,  1610DE (BCH004490) 00888707001182,  1610DE (BCH004497) 00888707001182,  1610CA (BCH003975) 00888707001175,  1610IT (BCH004168) 00888707001212,  1610DE (BCH004449) 00888707001182,  1610GB (BCH004676) 00888707001205,  1610GB (BCH003968) 00888707001205,  1610DE (BCH004396) 00888707001182,  1610DE (BCH004397) 00888707001182,  1610CH (BCH004459) 00888707001199,  1610DE (BCH004483) 00888707001182,  1610DE (BCH003607) 00888707001182,  1610DE (BCH003809) 00888707001182,  1610DE (BCH004395) 00888707001182,  1610DE (BCH004408) 00888707001182,  1610DE (BCH004495) 00888707001182,  1610DE (BCH003828) 00888707001182,  1610DE (BCH003831) 00888707001182,  1610DE (BCH003843) 00888707001182,  1610IT (BCH004305) 00888707001212,  1610IT (BCH004399) 00888707001212,  1610DE (BCH004451) 00888707001182,  1610DE (BCH003691) 00888707001182,  1610DE (BCH003870) 00888707001182,  1610DE (BCH004242) 00888707001182,  1610DE (BCH004246) 00888707001182,  1610DE (BCH004385) 00888707001182,  1610DE (BCH004412) 00888707001182,  1610DE (BCH004456) 00888707001182,  1610DE (BCH004478) 00888707001182,  1610DE (BCH004479) 00888707001182,  1610DE (BCH004487) 00888707001182,  1610DE (BCH003644) 00888707001182,  1610DE (BCH003847) 00888707001182,  1610DE (BCH003865) 00888707001182,  1610CH (BCH003964) 00888707001199,  1610DE (BCH004431) 00888707001182,  1610DE (BCH003811) 00888707001182,  1610JP (BCH004427) 00888707001236,  1610DE (BCH003822) 00888707001182,  1610DE (BCH003832) 00888707001182,  1610DE (BCH003862) 00888707001182,  1610DE (BCH003866) 00888707001182,  1610AU (BCH003945) 00888707001144,  1610DE (BCH003814) 00888707001182,  1610DE (BCH003827) 00888707001182,  1610DE (BCH003859) 00888707001182,  1610DE (BCH003887) 00888707001182,  1610IT (BCH004446) 00888707001212,  1610DE (BCH003690) 00888707001182,  1610DE (BCH003738) 00888707001182,  1610DE (BCH003858) 00888707001182,  1610GB (BCH004060) 00888707001205,  1610DE (BCH003603) 00888707001182,  1610DE (BCH003823) 00888707001182,  1610AU (BCH004068) 00888707001144,  1610DE (BCH004428) 00888707001182,  1610IT (BCH004445) 00888707001212,  1610IT (BCH004486) 00888707001212,  1610DE (BCH003619) 00888707001182,  1610DE (BCH003622) 00888707001182,  1610DE (BCH003816) 00888707001182,  1610DE (BCH003890) 00888707001182,  1610DE (BCH004466) 00888707001182,  1610DE (BCH003830) 00888707001182,  1610DE (BCH003846) 00888707001182,  1610DE (BCH003861) 00888707001182,  1610DE (BCH003864) 00888707001182,  1610IT (BCH003879) 00888707001212,  1610DE (BCH003888) 00888707001182,  1610DE (BCH003889) 00888707001182,  1610IT (BCH004444) 00888707001212,  1610IT (BCH004467) 00888707001212,  1610IT (BCH003868) 00888707001212,  1610DE (BCH003876) 00888707001182,  1610IT (BCH004472) 00888707001212,  1610IT (BCH004474) 00888707001212,  1610GB (BCH004677) 00888707001205,  1610DE (BCH003643) 00888707001182,  1610DE (BCH003817) 00888707001182,  1610DE (BCH003824) 00888707001182,  1610DE (BCH003829) 00888707001182,  1610DE (BCH003863) 00888707001182,  1610DE (BCH004652) 00888707001182,  1610DE (BCH004468) 00888707001182,  1610IT (BCH003638) 00888707001212,  1610IT (BCH003688) 00888707001212,  1610DE (BCH003848) 00888707001182,  1610DE (BCH004453) 00888707001182,  1610DE (BCH004462) 00888707001182,  1610DE (BCH003601) 00888707001182,  1610AU (BCH004013) 00888707001144,  1610IT (BCH004387) 00888707001212,  1610IT (BCH004417) 00888707001212,  1610DE (BCH004454) 00888707001182,  1610DE (BCH004494) 00888707001182,  1610DE (BCH003642) 00888707001182,  1610DE (BCH003749) 00888707001182,  1610GB (BCH003984) 00888707001205,  1610GB (BCH003999) 00888707001205,  1610GB (BCH004090) 00888707001205,  1610DE (BCH004464) 00888707001182,  1610CA (BCH004689) 00888707001175,  1610DE (BCH003584) 00888707001182,  1610DE (BCH003593) 00888707001182,  1610DE (BCH003600) 00888707001182,  1610DE (BCH003602) 00888707001182,  1610DE (BCH003628) 00888707001182,  1610DE (BCH003632) 00888707001182,  1610DE (BCH003647) 00888707001182,  1610DE (BCH003747) 00888707001182,  1610DE (BCH003803) 00888707001182,  1610DE (BCH003612) 00888707001182,  1610GB (BCH004070) 00888707001205,  1610DE (BCH004281) 00888707001182,  1610DE (BCH003592) 00888707001182,  1610GB (BCH003608) 00888707001205,  1610DE (BCH003613) 00888707001182,  1610DE (BCH003727) 00888707001182,  1610AU (BCH003927) 00888707001144,  1610CH (BCH003928) 00888707001199,  1610GB (BCH004051) 00888707001205,  1610GB (BCH004656) 00888707001205,  1610DE (BCH003825) 00888707001182,  1610DE (BCH003925) 00888707001182,  1610CA (BCH003985) 00888707001175,  1610DE (BCH003588) 00888707001182,  1610DE (BCH003645) 00888707001182,  1610CH (BCH003936) 00888707001199,  1610DE (BCH003637) 00888707001182,  1610DE (BCH003587) 00888707001182,  1610DE (BCH003604) 00888707001182,  1610DE (BCH003605) 00888707001182,  1610DE (BCH003610) 00888707001182,  1610DE (BCH003629) 00888707001182,  1610DE (BCH003636) 00888707001182,  1610DE (BCH003646) 00888707001182,  1610DE (BCH003674) 00888707001182,  1610DE (BCH003683) 00888707001182,  1610DE (BCH003687) 00888707001182,  1610DE (BCH003807) 00888707001182,  1610DE (BCH003820) 00888707001182,  1610DE (BCH003821) 00888707001182,  1610DE (BCH003877) 00888707001182,  1610CA (BCH004590) 00888707001175,  1610DE (BCH003618) 00888707001182,  1610DE (BCH003685) 00888707001182,  1610DE (BCH003835) 00888707001182,  1610JP (BCH003849) 00888707001236,  1610DE (BCH003883) 00888707001182,  1610CA-CLIN (BCH004538) 00888707001564,  1610DE (BCH003724) 00888707001182,  1610DE (BCH003725) 00888707001182,  1610GB (BCH004078) 00888707001205,  1610DE (BCH003662) 00888707001182,  1610DE (BCH003741) 00888707001182,  1610DE (BCH003630) 00888707001182,  1610DE (BCH003720) 00888707001182,  1610DE (BCH003894) 00888707001182,  1610DE (BCH003949) 00888707001182,  1610AU (BCH004023) 00888707001144,  1610CA (BCH004633) 00888707001175,  1610DE (BCH004065) 00888707001182,  1610IT (BCH004470) 00888707001212,  1610JP (BCH003120) 00888707001236,  1610DE (BCH003595) 00888707001182,  1610DE (BCH003597) 00888707001182,  1610DE (BCH003598) 00888707001182,  1610DE (BCH003633) 00888707001182,  1610DE (BCH003661) 00888707001182,  1610DE (BCH003742) 00888707001182,  1610DE (BCH003804) 00888707001182,  1610DE (BCH003722) 00888707001182,  1610AU (BCH004001) 00888707001144,  1610AU (BCH004038) 00888707001144,  1610GB (BCH003419) 00888707001205,  1610IT (BCH003583) 00888707001212,  1610DE (BCH003606) 00888707001182,  1610IT (BCH003681) 00888707001212,  1610DE (BCH003713) 00888707001182,  1610DE (BCH003721) 00888707001182,  1610IT (BCH003800) 00888707001212,  1610CH (BCH004261) 00888707001199,  1610DE (BCH003671) 00888707001182,  1610DE (BCH003631) 00888707001182,  1610DE (BCH003729) 00888707001182,  1610DE (BCH003838) 00888707001182,  1610AU (BCH004037) 00888707001144,  1610DE (BCH003743) 00888707001182,  1610DE (BCH003611) 00888707001182,  1610DE (BCH003660) 00888707001182,  1610DE (BCH003682) 00888707001182,  1610DE (BCH003635) 00888707001182,  1610IT (BCH003737) 00888707001212,  1610IT (BCH003740) 00888707001212,  1610DE (BCH003788) 00888707001182,  1610DE (BCH003841) 00888707001182,  1610DE (BCH003931) 00888707001182,  1610DE (BCH003590) 00888707001182,  1610DE (BCH003723) 00888707001182,  1610DE (BCH003893) 00888707001182,  1610DE (BCH003882) 00888707001182,  1610DE (BCH003591) 00888707001182,  1610DE (BCH003599) 00888707001182,  1610DE (BCH003614) 00888707001182,  1610DE (BCH003676) 00888707001182,  1610DE (BCH003679) 00888707001182,  1610DE (BCH003693) 00888707001182,  1610DE (BCH003695) 00888707001182,  1610DE (BCH003739) 00888707001182,  1610DE (BCH003799) 00888707001182,  1610DE (BCH003806) 00888707001182,  1610DE (BCH003826) 00888707001182,  1610DE (BCH003897) 00888707001182,  1610IT (BCH004555) 00888707001212,  1610IT (BCH004625) 00888707001212,  1610DE (BCH003748) 00888707001182,  1610GB (BCH004603) 00888707001205,  1610DE (BCH003808) 00888707001182,  1610DE (BCH003881) 00888707001182,  1610DE (BCH003818) 00888707001182,  1610DE (BCH003878) 00888707001182,  1610DE (BCH003880) 00888707001182,  1610CH (BCH004260) 00888707001199,  1610DE (BCH003689) 00888707001182,  1610DE (BCH003819) 00888707001182,  1610DE (BCH003842) 00888707001182,  1610DE (BCH004530) 00888707001182,  1610DE (BCH003651) 00888707001182,  1610DE (BCH003686) 00888707001182,  1610IT (BCH003798) 00888707001212,  1610AU (BCH003947) 00888707001144,  1610IT (BCH004135) 00888707001212,  1610GB (BCH004648) 00888707001205,  1610DE (BCH003805) 00888707001182,  1610DE (BCH003617) 00888707001182,  1610DE (BCH003654) 00888707001182,  1610IT (BCH003913) 00888707001212,  1610IT (BCH004548) 00888707001212,  1610DE (BCH003634) 00888707001182,  1610DE (BCH003714) 00888707001182,  1610DE (BCH003837) 00888707001182,  1610DE (BCH003891) 00888707001182,  1610DE (BCH003833) 00888707001182,  1610DE (BCH003703) 00888707001182,  1610DE (BCH003733) 00888707001182,  1610DE (BCH003734) 00888707001182,  1610DE (BCH003795) 00888707001182,  1610DE (BCH003664) 00888707001182,  1610DE (BCH003886) 00888707001182,  1610GB (BCH004075) 00888707001205,  1610AU (BCH004627) 00888707001144,  1610DE (BCH003667) 00888707001182,  1610DE (BCH003668) 00888707001182,  1610DE (BCH003702) 00888707001182,  1610DE (BCH003731) 00888707001182,  1610DE (BCH003802) 00888707001182,  1610DE (BCH003854) 00888707001182,  1610DE (BCH003857) 00888707001182,  1610DE (BCH003885) 00888707001182,  1610IL (BCH003840) 00888707001229,  1610DE (BCH003856) 00888707001182,  1610DE (BCH003895) 00888707001182,  1610IL (BCH004031) 00888707001229,  1610GB (BCH004507) 00888707001205,  1610DE (BCH003653) 00888707001182,  1610DE (BCH003705) 00888707001182,  1610DE (BCH003796) 00888707001182,  1610DE (BCH003834) 00888707001182,  1610DE (BCH004020) 00888707001182,  1610DE (BCH004021) 00888707001182,  1610DE (BCH003658) 00888707001182,  1610DE (BCH003666) 00888707001182,  1610DE (BCH003680) 00888707001182,  1610DE (BCH003684) 00888707001182,  1610DE (BCH003709) 00888707001182,  1610DE (BCH003853) 00888707001182,  1610DE (BCH003794) 00888707001182,  1610DE (BCH003812) 00888707001182,  1610DE (BCH003874) 00888707001182,  1610DE (BCH003797) 00888707001182,  1610DE (BCH003892) 00888707001182,  1610DE (BCH004062) 00888707001182,  1610DE (BCH003694) 00888707001182,  1610DE (BCH003659) 00888707001182,  1610DE (BCH003675) 00888707001182,  1610DE (BCH003652) 00888707001182,  1610DE (BCH003657) 00888707001182,  1610DE (BCH003730) 00888707001182,  1610DE (BCH003736) 00888707001182,  1610DE (BCH004063) 00888707001182,  1610DE (BCH003533) 00888707001182,  1610DE (BCH003712) 00888707001182,  1610DE (BCH003813) 00888707001182,  1610DE (BCH003815) 00888707001182,  1610DE (BCH003855) 00888707001182,  1610DE (BCH003665) 00888707001182,  1610DE (BCH003696) 00888707001182,  1610DE (BCH003704) 00888707001182,  1610DE (BCH003710) 00888707001182,  1610DE (BCH003875) 00888707001182,  1610GB (BCH003982) 00888707001205,  1610DE (BCH004077) 00888707001182,  1610DE (BCH004529) 00888707001182,  1610CA (BCH003503) 00888707001175,  1610DE (BCH003708) 00888707001182,  1610DE (BCH003717) 00888707001182,  1610DE (BCH003718) 00888707001182,  1610GB (BCH003942) 00888707001205,  1610GB (BCH004563) 00888707001205,  1610CA-CLIN (BCH004607) 00888707001564,  1610DE (BCH003582) 00888707001182,  1610DE (BCH003616) 00888707001182,  1610DE (BCH004000) 00888707001182,  1610DE (BCH004006) 00888707001182,  1610CA (BCH004663) 00888707001175,  1610DE (BCH003458) 00888707001182,  1610DE (BCH004087) 00888707001182,  1610DE (BCH003621) 00888707001182,  1610DE (BCH003640) 00888707001182,  1610GB (BCH004532) 00888707001205,  1610CA (BCH004635) 00888707001175,  1610IT (BCH003711) 00888707001212,  1610DE (BCH003801) 00888707001182,  1610DE (BCH004086) 00888707001182,  1610DE (BCH004088) 00888707001182,  1610IT (BCH004253) 00888707001212,  1610IT (BCH004558) 00888707001212,  1610DE (BCH004017) 00888707001182,  1610DE (BCH004128) 00888707001182,  1610GB (BCH004553) 00888707001205,  1610DE (BCH003457) 00888707001182,  1610CH (BCH003480) 00888707001199,  1610DE (BCH003594) 00888707001182,  1610DE (BCH003951) 00888707001182,  1610DE (BCH003623) 00888707001182,  1610DE (BCH004139) 00888707001182,  1610DE (BCH004007) 00888707001182,  1610CA (BCH004572) 00888707001175,  1610DE (BCH003585) 00888707001182,  1610DE (BCH004042) 00888707001182,  1610GB (BCH004059) 00888707001205,  1610DE (BCH004129) 00888707001182,  1610GB (BCH004540) 00888707001205,  1610CH (BCH004057) 00888707001199,  1610GB (BCH004534) 00888707001205,  1610GB (BCH004546) 00888707001205,  1610GB (BCH004604) 00888707001205,  1610DE (BCH004620) 00888707001182,  1610DE (BCH003648) 00888707001182,  1610CH (BCH004058) 00888707001199,  1610DE (BCH003627) 00888707001182,  1610DE (BCH003873) 00888707001182,  1610IT (BCH004518) 00888707001212,  1610IT (BCH004519) 00888707001212,  1610DE (BCH004018) 00888707001182,  1610DE (BCH004080) 00888707001182,  1610DE (BCH004089) 00888707001182,  1610IN (BCH004562) 00888707000536,  1610AU (BCH004682) 00888707001144,  1610DE (BCH003580) 00888707001182,  1610DE (BCH003581) 00888707001182,  1610DE (BCH003649) 00888707001182,  1610DE (BCH004085) 00888707001182,  1610DE (BCH004143) 00888707001182,  1610DE (BCH004266) 00888707001182,  1610GB (BCH003568) 00888707001205,  1610DE (BCH003624) 00888707001182,  1610GB (BCH003967) 00888707001205,  1610DE (BCH004016) 00888707001182,  1610DE (BCH004030) 00888707001182,  1610DE (BCH003625) 00888707001182,  1610DE (BCH003626) 00888707001182,  1610DE (BCH003641) 00888707001182,  1610DE (BCH003906) 00888707001182,  1610DE (BCH004002) 00888707001182,  1610GB (BCH004010) 00888707001205,  1610DE (BCH004054) 00888707001182,  1610CA (BCH004628) 00888707001175,  1610GB (BCH003553) 00888707001205,  1610DE (BCH004005) 00888707001182,  1610GB (BCH004605) 00888707001205,  1610GB (BCH004662) 00888707001205,  1610DE (BCH004064) 00888707001182,  1610DE (BCH003579) 00888707001182,  1610DE (BCH003615) 00888707001182,  1610DE (BCH003650) 00888707001182,  1610DE (BCH003678) 00888707001182,  1610DE (BCH003933) 00888707001182,  1610DE (BCH003944) 00888707001182,  1610DE (BCH003950) 00888707001182,  1610DE (BCH003586) 00888707001182,  1610DE (BCH003922) 00888707001182,  1610DE (BCH003958) 00888707001182,  1610DE (BCH003989) 00888707001182,  1610DE (BCH004066) 00888707001182,  1610IT (BCH004544) 00888707001212,  1610IT (BCH004559) 00888707001212,  1610IT (BCH004668) 00888707001212,  1610DE (BCH004079) 00888707001182,  1610AU (BCH004683) 00888707001144,  1610GB (BCH004050) 00888707001205,  1610GB (BCH004082) 00888707001205,  1610DE (BCH003900) 00888707001182,  1610GB (BCH003904) 00888707001205,  1610DE (BCH003916) 00888707001182,  1610DE (BCH003935) 00888707001182,  1610GB (BCH003938) 00888707001205,  1610CA (BCH004501) 00888707001175,  1610DE (BCH003620) 00888707001182,  1610DE (BCH003639) 00888707001182,  1610DE (BCH003901) 00888707001182,  1610DE (BCH003912) 00888707001182,  1610DE (BCH003926) 00888707001182,  1610DE (BCH003932) 00888707001182,  1610DE (BCH003948) 00888707001182,  1610DE (BCH003988) 00888707001182,  1610DE (BCH004055) 00888707001182,  1610DE (BCH004673) 00888707001182,  1610DE (BCH003898) 00888707001182,  1610DE (BCH004053) 00888707001182,  1610GB (BCH004608) 00888707001205,  1610DE (BCH003899) 00888707001182,  1610DE (BCH004093) 00888707001182,  1610IT (BCH004257) 00888707001212,  1610DE (BCH003934) 00888707001182,  1610DE (BCH003513) 00888707001182,  1610DE (BCH003902) 00888707001182,  1610DE (BCH003920) 00888707001182,  1610DE (BCH003954) 00888707001182,  1610DE (BCH004097) 00888707001182,  1610AU (BCH003500) 00888707001144,  1610DE (BCH003907) 00888707001182,  1610DE (BCH003908) 00888707001182,  1610DE (BCH003909) 00888707001182,  1610DE (BCH004026) 00888707001182,  1610IN (BCH004619) 00888707000536,  1610CA (BCH003569) 00888707001175,  1610DE (BCH003911) 00888707001182,  1610DE (BCH003921) 00888707001182,  1610DE (BCH003930) 00888707001182,  1610DE (BCH003939) 00888707001182,  1610DE (BCH003953) 00888707001182,  1610DE (BCH003956) 00888707001182,  1610DE (BCH004011) 00888707001182,  1610DE (BCH004028) 00888707001182,  1610DE (BCH004052) 00888707001182,  1610IN (BCH004250) 00888707000536,  1610GB (BCH004651) 00888707001205,  1610DE (BCH003923) 00888707001182,  1610DE (BCH003941) 00888707001182,  1610DE (BCH004014) 00888707001182,  1610DE (BCH004022) 00888707001182,  1610DE (BCH003914) 00888707001182,  1610DE (BCH003918) 00888707001182,  1610DE (BCH003955) 00888707001182,  1610DE (BCH003969) 00888707001182,  1610DE (BCH003983) 00888707001182,  1610DE (BCH004029) 00888707001182,  1610DE (BCH003963) 00888707001182,  1610DE (BCH004034) 00888707001182,  1610DE (BCH003915) 00888707001182,  1610DE (BCH003946) 00888707001182,  1610DE (BCH003962) 00888707001182,  1610DE (BCH003924) 00888707001182,  1610GB (BCH003937) 00888707001205,  1610DE (BCH003957) 00888707001182,  1610DE (BCH003959) 00888707001182,  1610DE (BCH003960) 00888707001182,  1610DE (BCH003965) 00888707001182,  1610DE (BCH003993) 00888707001182,  1610GB (BCH004056) 00888707001205,  1610DE (BCH003910) 00888707001182,  1610DE (BCH003952) 00888707001182,  1610IN (BCH004169) 00888707000536,  1610IN (BCH003387) 00888707000536,  1610DE (BCH003905) 00888707001182,  1610DE (BCH003917) 00888707001182,  1610DE (BCH003940) 00888707001182,  1610DE (BCH004035) 00888707001182,  1610DE (BCH004049) 00888707001182,  1610DE (BCH003929) 00888707001182,  1610DE (BCH003972) 00888707001182,  1610DE (BCH004027) 00888707001182,  1610DE (BCH004043) 00888707001182,  1610DE (BCH003525) 00888707001182,  1610DE (BCH004008) 00888707001182,  1610DE (BCH004032) 00888707001182,  1610DE (BCH004069) 00888707001182,  1610DE (BCH004509) 00888707001182,  1610DE (BCH004639) 00888707001182,  1610DE (BCH003535) 00888707001182,  1610DE (BCH003543) 00888707001182,  1610DE (BCH003903) 00888707001182,  1610DE (BCH003517) 00888707001182,  1610DE (BCH003961) 00888707001182,  1610DE (BCH004554) 00888707001182,  1610DE (BCH004570) 00888707001182,  1610DE (BCH003554) 00888707001182,  1610DE (BCH003994) 00888707001182,  1610DE (BCH004048) 00888707001182,  1610DE (BCH004096) 00888707001182,  1610DE (BCH003970) 00888707001182,  1610DE (BCH004040) 00888707001182,  1610IL (BCH004081) 00888707001229,  1610DE (BCH004610) 00888707001182,  1610DE (BCH004646) 00888707001182,  1610DE (BCH003508) 00888707001182,  1610DE (BCH003511) 00888707001182,  1610DE (BCH003520) 00888707001182,  1610DE (BCH003523) 00888707001182,  1610DE (BCH003542) 00888707001182,  1610DE (BCH003556) 00888707001182,  1610DE (BCH003558) 00888707001182,  1610DE (BCH004537) 00888707001182,  1610DE (BCH004564) 00888707001182,  1610DE (BCH004671) 00888707001182,  1610AU (BCH004681) 00888707001144,  1610DE (BCH003510) 00888707001182,  1610DE (BCH003522) 00888707001182,  1610DE (BCH003527) 00888707001182,  1610DE (BCH003532) 00888707001182,  1610DE (BCH003546) 00888707001182,  1610DE (BCH003578) 00888707001182,  1610DE (BCH003509) 00888707001182,  1610DE (BCH003526) 00888707001182,  1610DE (BCH003552) 00888707001182,  1610DE (BCH003979) 00888707001182,  1610DE (BCH004047) 00888707001182,  1610DE (BCH004072) 00888707001182,  1610DE (BCH004074) 00888707001182,  1610DE (BCH004551) 00888707001182,  1610DE (BCH003971) 00888707001182,  1610DE (BCH003980) 00888707001182,  1610DE (BCH003981) 00888707001182,  1610DE (BCH003992) 00888707001182,  1610DE (BCH004025) 00888707001182,  1610DE (BCH003515) 00888707001182,  1610GB (BCH003990) 00888707001205,  1610GB (BCH004033) 00888707001205,  1610GB (BCH004036) 00888707001205,  1610DE (BCH003537) 00888707001182,  1610DE (BCH003538) 00888707001182,  1610DE (BCH003539) 00888707001182,  1610DE (BCH003560) 00888707001182,  1610DE (BCH003531) 00888707001182,  1610DE (BCH003557) 00888707001182,  1610DE (BCH003559) 00888707001182,  1610DE (BCH003561) 00888707001182,  1610DE (BCH003974) 00888707001182,  1610DE (BCH003976) 00888707001182,  1610DE (BCH004094) 00888707001182,  1610DE (BCH003572) 00888707001182,  1610DE (BCH003986) 00888707001182,  1610DE (BCH004045) 00888707001182,  1610DE (BCH003562) 00888707001182,  1610DE (BCH003995) 00888707001182,  1610DE (BCH003996) 00888707001182,  1610DE (BCH004024) 00888707001182,  1610DE (BCH004044) 00888707001182,  1610DE (BCH004012) 00888707001182,  1610DE (BCH004015) 00888707001182,  1610DE (BCH004046) 00888707001182,  1610DE (BCH003575) 00888707001182,  1610DE (BCH003577) 00888707001182,  1610DE (BCH003991) 00888707001182,  1610GB (BCH004039) 00888707001205,  1610AU (BCH003493) 00888707001144,  1610AU (BCH003502) 00888707001144,  1610DE (BCH003973) 00888707001182,  1610DE (BCH003977) 00888707001182,  1610AU (BCH004152) 00888707001144,  1610DE (BCH003514) 00888707001182,  1610DE (BCH004003) 00888707001182,  1610DE (BCH004004) 00888707001182,  1610CH (BCH004284) 00888707001199,  1610DE (BCH003534) 00888707001182,  1610DE (BCH003571) 00888707001182,  1610DE (BCH003978) 00888707001182,  1610DE (BCH004073) 00888707001182,  1610CA (BCH003282) 00888707001175,  1610CA (BCH003283) 00888707001175,  1610DE (BCH003566) 00888707001182,  1610DE (BCH004084) 00888707001182,  1610CH (BCH004285) 00888707001199,  1610DE (BCH003548) 00888707001182,  1610DE (BCH003550) 00888707001182,  1610DE (BCH003567) 00888707001182,  1610DE (BCH003574) 00888707001182,  1610IT (BCH004543) 00888707001212,  1610IT (BCH004666) 00888707001212,  1610DE (BCH004669) 00888707001182,  1610IN (BCH003385) 00888707000536,  1610DE (BCH003529) 00888707001182,  1610DE (BCH004258) 00888707001182,  1610DE (BCH004510) 00888707001182,  1610DE (BCH004565) 00888707001182,  1610IT (BCH004612) 00888707001212,  1610AU (BCH003499) 00888707001144,  1610DE (BCH003524) 00888707001182,  1610DE (BCH003540) 00888707001182,  1610DE (BCH003541) 00888707001182,  1610DE (BCH003547) 00888707001182,  1610DE (BCH003555) 00888707001182,  1610DE (BCH004095) 00888707001182,  1610CH (BCH004140) 00888707001199,  1610GB (BCH004504) 00888707001205,  1610DE (BCH004531) 00888707001182,  1610DE (BCH004552) 00888707001182. 1610DE (BCH004566) 00888707001182,  1610CA (BCH004571) 00888707001175,  1610CA (BCH004634) 00888707001175,  1610DE (BCH003506) 00888707001182,  1610DE (BCH003507) 00888707001182,  1610DE (BCH003516) 00888707001182,  1610DE (BCH003521) 00888707001182,  1610DE (BCH003530) 00888707001182,  1610DE (BCH003536) 00888707001182,  1610DE (BCH003544) 00888707001182,  1610DE (BCH003545) 00888707001182,  1610DE (BCH003549) 00888707001182,  1610DE (BCH003551) 00888707001182,  1610DE (BCH003563) 00888707001182,  1610DE (BCH003564) 00888707001182,  1610DE (BCH003565) 00888707001182,  1610DE (BCH003570) 00888707001182,  1610DE (BCH003573) 00888707001182,  1610DE (BCH003576) 00888707001182,  1610DE (BCH004526) 00888707001182,  1610DE (BCH004609) 00888707001182,  1610DE (BCH004611) 00888707001182,  1610DE (BCH004659) 00888707001182,  1610DE (BCH004670) 00888707001182,  1610CH (BCH003468) 00888707001199,  1610DE (BCH003504) 00888707001182,  1610DE (BCH003505) 00888707001182,  1610DE (BCH004602) 00888707001182,  1610GB (BCH003498) 00888707001205,  1610DE (BCH003501) 00888707001182,  1610DE (BCH003518) 00888707001182,  1610DE (BCH004156) 00888707001182,  1610DE (BCH004158) 00888707001182,  1610DE (BCH004159) 00888707001182,  1610GB (BCH004165) 00888707001205,  1610DE (BCH004600) 00888707001182,  1610DE (BCH004645) 00888707001182,  1610DE (BCH004136) 00888707001182,  1610DE (BCH004516) 00888707001182,  1610IT (BCH003218) 00888707001212,  1610DE (BCH004107) 00888707001182,  1610DE (BCH004515) 00888707001182,  1610DE (BCH004517) 00888707001182,  1610IT (BCH004568) 00888707001212,  1610DE (BCH004655) 00888707001182,  1610DE (BCH004664) 00888707001182,  1610DE (BCH004672) 00888707001182,  1610DE (BCH004693) 00888707001182,  1610DE (BCH003519) 00888707001182,  1610DE (BCH004160) 00888707001182,  1610DE (BCH004505) 00888707001182,  1610DE (BCH004125) 00888707001182,  1610DE (BCH004265) 00888707001182,  1610DE (BCH004613) 00888707001182,  1610DE (BCH004615) 00888707001182,  1610DE (BCH003528) 00888707001182,  1610DE (BCH004161) 00888707001182,  1610DE (BCH004282) 00888707001182,  1610DE (BCH004520) 00888707001182,  1610DE (BCH004569) 00888707001182,  1610DE (BCH004599) 00888707001182,  1610DE (BCH004623) 00888707001182,  1610DE (BCH004674) 00888707001182,  1610AU (BCH003392) 00888707001144,  1610DE (BCH004601) 00888707001182,  1610DE (BCH004667) 00888707001182,  1610GB (BCH004164) 00888707001205,  1610GB (BCH004280) 00888707001205,  1610DE (BCH004557) 00888707001182,  1610DE (BCH004560) 00888707001182,  1610DE (BCH004567) 00888707001182,  1610GB (BCH004167) 00888707001205,  1610DE (BCH004511) 00888707001182,  1610CH (BCH004290) 00888707001199,  1610DE (BCH004630) 00888707001182,  1610IT (BCH003213) 00888707001212,  1610IT (BCH003285) 00888707001212,  1610IT (BCH003451) 00888707001212,  1610DE (BCH004500) 00888707001182,  1610DE (BCH004522) 00888707001182,  1610GB (BCH004130) 00888707001205,  1610DE (BCH004640) 00888707001182,  1610CA-CLIN (BCH003432) 00888707001564,  1610AU (BCH004262) 00888707001144,  1610DE (BCH004498) 00888707001182,  1610DE (BCH004512) 00888707001182,  1610DE (BCH004513) 00888707001182,  1610DE (BCH004521) 00888707001182,  1610DE (BCH004577) 00888707001182,  1610DE (BCH004579) 00888707001182,  1610DE (BCH004586) 00888707001182,  1610DE (BCH004587) 00888707001182,  1610DE (BCH004616) 00888707001182,  1610DE (BCH004617) 00888707001182,  1610DE (BCH004624) 00888707001182,  1610DE (BCH004636) 00888707001182,  1610DE (BCH004637) 00888707001182,  1610DE (BCH004660) 00888707001182,  1610DE (BCH004695) 00888707001182,  1610AU (BCH004131) 00888707001144,  1610DE (BCH004269) 00888707001182,  1610DE (BCH004523) 00888707001182,  1610DE (BCH004268) 00888707001182,  1610DE (BCH004687) 00888707001182,  1610DE (BCH004574) 00888707001182,  1610DE (BCH004593) 00888707001182,  1610DE (BCH004111) 00888707001182,  1610DE (BCH004154) 00888707001182,  1610DE (BCH004597) 00888707001182,  1610CA (BCH004618) 00888707001175,  1610CA (BCH004650) 00888707001175,  1610DE (BCH004127) 00888707001182,  1610DE (BCH004155) 00888707001182,  1610DE (BCH004502) 00888707001182,  1610DE (BCH004556) 00888707001182,  1610IT (BCH003196) 00888707001212,  1610IT (BCH003217) 00888707001212,  1610IT (BCH003288) 00888707001212,  1610AU (BCH003477) 00888707001144,  1610CA (BCH004541) 00888707001175,  1610DE (BCH004573) 00888707001182,  1610DE (BCH004585) 00888707001182,  1610DE (BCH004589) 00888707001182,  1610DE (BCH004594) 00888707001182,  1610CA (BCH004524) 00888707001175,  1610DE (BCH004576) 00888707001182,  1610DE (BCH004578) 00888707001182,  1610DE (BCH004592) 00888707001182,  1610DE (BCH004621) 00888707001182,  1610DE (BCH004661) 00888707001182,  1610GB (BCH004126) 00888707001205,  1610DE (BCH004153) 00888707001182,  1610DE (BCH004581) 00888707001182,  1610DE (BCH004622) 00888707001182,  1610CA (BCH004647) 00888707001175,  1610DE (BCH004694) 00888707001182,  1610CH (BCH003415) 00888707001199,  1610GB (BCH003459) 00888707001205,  1610CA-CLIN (BCH003481) 00888707001564,  1610DE (BCH004112) 00888707001182,  1610DE (BCH004580) 00888707001182,  1610DE (BCH004582) 00888707001182,  1610DE (BCH004591) 00888707001182,  1610DE (BCH004595) 00888707001182,  1610DE (BCH004642) 00888707001182,  1610DE (BCH004644) 00888707001182,  1610DE (BCH004696) 00888707001182,  1610DE (BCH004244) 00888707001182,  1610CA (BCH004525) 00888707001175,  1610DE (BCH004584) 00888707001182,  1610DE (BCH004134) 00888707001182,  1610DE (BCH004145) 00888707001182,  1610DE (BCH004151) 00888707001182,  1610DE (BCH004166) 00888707001182,  1610DE (BCH004248) 00888707001182,  1610DE (BCH004264) 00888707001182,  1610DE (BCH004270) 00888707001182,  1610DE (BCH004288) 00888707001182,  1610DE (BCH004575) 00888707001182,  1610DE (BCH004583) 00888707001182,  1610DE (BCH004596) 00888707001182,  1610DE (BCH004598) 00888707001182,  1610DE (BCH004137) 00888707001182,  1610DE (BCH004144) 00888707001182,  1610DE (BCH004247) 00888707001182,  1610CA (BCH004550) 00888707001175,  1610DE (BCH004638) 00888707001182,  1610GB (BCH003156) 00888707001205,  1610DE (BCH004133) 00888707001182,  1610DE (BCH004147) 00888707001182,  1610DE (BCH004157) 00888707001182,  1610DE (BCH004255) 00888707001182,  1610DE (BCH004272) 00888707001182,  1610CA (BCH004549) 00888707001175,  1610DE (BCH004641) 00888707001182,  1610CA-CLIN (BCH003483) 00888707001564,  1610CA-CLIN (BCH003793) 00888707001564,  1610DE (BCH004278) 00888707001182,  1610CA (BCH004631) 00888707001175,  1610DE (BCH004688) 00888707001182,  1610IN (BCH003434) 00888707000536,  1610DE (BCH004132) 00888707001182,  1610DE (BCH004163) 00888707001182,  1610DE (BCH004291) 00888707001182,  1610IT (BCH004614) 00888707001212,  1610DE (BCH004150) 00888707001182,  1610DE (BCH004162) 00888707001182,  1610DE (BCH003440) 00888707001182,  1610GB (BCH004243) 00888707001205,  1610DE (BCH004251) 00888707001182,  1610DE (BCH004292) 00888707001182,  1610DE (BCH003398) 00888707001182,  1610AU (BCH003428) 00888707001144,  1610GB (BCH004149) 00888707001205,  1610DE (BCH004283) 00888707001182,  1610DE (BCH004643) 00888707001182,  1610DE (BCH004649) 00888707001182,  1610DE (BCH003439) 00888707001182,  1610DE (BCH004138) 00888707001182,  1610DE (BCH004249) 00888707001182,  1610DE (BCH004271) 00888707001182,  1610DE (BCH003446) 00888707001182,  1610DE (BCH003472) 00888707001182,  1610DE (BCH004245) 00888707001182,  1610DE (BCH004259) 00888707001182,  1610DE (BCH004263) 00888707001182,  1610DE (BCH003388) 00888707001182,  1610DE (BCH004254) 00888707001182,  1610CA (BCH004665) 00888707001175,  1610DE (BCH003244) 00888707001182,  1610CA (BCH004588) 00888707001175,  1610DE (BCH003142) 00888707001182,  1610CH (BCH003238) 00888707001199,  1610DE (BCH003394) 00888707001182,  1610DE (BCH003416) 00888707001182,  1610DE (BCH003143) 00888707001182,  1610CA (BCH003397) 00888707001175,  1610DE (BCH003421) 00888707001182,  1610GB (BCH003453) 00888707001205,  1610CH (BCH003223) 00888707001199,  1610DE (BCH003401) 00888707001182,  1610DE (BCH003438) 00888707001182,  1610DE (BCH003135) 00888707001182,  1610DE (BCH003393) 00888707001182,  1610DE (BCH003417) 00888707001182,  1610DE (BCH003159) 00888707001182,  1610DE (BCH003405) 00888707001182,  1610DE (BCH003406) 00888707001182,  1610DE (BCH003447) 00888707001182,  1610DE (BCH004141) 00888707001182,  1610DE (BCH003395) 00888707001182,  1610GB (BCH003399) 00888707001205,  1610DE (BCH004142) 00888707001182,  1610GB (BCH003418) 00888707001205,  1610DE (BCH004252) 00888707001182,  1610DE (BCH003382) 00888707001182,  1610DE (BCH003396) 00888707001182,  1610DE (BCH003427) 00888707001182,  1610DE (BCH003441) 00888707001182,  1610DE (BCH003452) 00888707001182,  1610DE (BCH003478) 00888707001182,  1610DE (BCH003144) 00888707001182,  1610DE (BCH003390) 00888707001182,  1610IT (BCH003404) 00888707001212,  1610DE (BCH003425) 00888707001182,  1610DE (BCH003426) 00888707001182,  1610DE (BCH003445) 00888707001182,  1610CA (BCH003479) 00888707001175,  1610DE (BCH003391) 00888707001182,  1610GB (BCH003470) 00888707001205,  1610GB (BCH004148) 00888707001205,  1610DE (BCH004256) 00888707001182,  1610DE (BCH003409) 00888707001182,  1610DE (BCH003431) 00888707001182,  1610DE (BCH003442) 00888707001182,  1610DE (BCH003466) 00888707001182,  1610DE (BCH003433) 00888707001182,  1610DE (BCH003437) 00888707001182,  1610DE (BCH003443) 00888707001182,  1610DE (BCH003444) 00888707001182,  1610GB (BCH003207) 00888707001205,  1610DE (BCH003482) 00888707001182,  1610DE (BCH004114) 00888707001182,  1610DE (BCH004115) 00888707001182,  1610DE (BCH004171) 00888707001182,  1610DE (BCH003260) 00888707001182,  1610DE (BCH003298) 00888707001182,  1610DE (BCH003379) 00888707001182,  1610DE (BCH003413) 00888707001182,  1610DE (BCH003465) 00888707001182,  1610DE (BCH003791) 00888707001182,  1610AU (BCH003407) 00888707001144,  1610GB (BCH003420) 00888707001205,  1610DE (BCH003467) 00888707001182,  1610DE (BCH003474) 00888707001182,  1610GB (BCH003183) 00888707001205,  1610GB (BCH003185) 00888707001205,  1610DE (BCH003381) 00888707001182,  1610DE (BCH003402) 00888707001182,  1610DE (BCH003412) 00888707001182,  1610DE (BCH003414) 00888707001182,  1610DE (BCH003436) 00888707001182,  1610DE (BCH003464) 00888707001182,  1610DE (BCH003792) 00888707001182,  1610DE (BCH004170) 00888707001182,  1610DE (BCH003380) 00888707001182,  1610DE (BCH003463) 00888707001182,  1610DE (BCH004113) 00888707001182,  1610DE (BCH003131) 00888707001182,  1610DE (BCH003245) 00888707001182,  1610DE (BCH003258) 00888707001182,  1610DE (BCH003461) 00888707001182,  1610GB (BCH003193) 00888707001205,  1610DE (BCH003389) 00888707001182,  1610DE (BCH003410) 00888707001182,  1610DE (BCH003430) 00888707001182,  1610DE (BCH003454) 00888707001182,  1610DE (BCH003455) 00888707001182,  1610DE (BCH003460) 00888707001182,  1610DE (BCH003775) 00888707001182,  1610GB (BCH003158) 00888707001205,  1610DE (BCH003462) 00888707001182,  1610DE (BCH003177) 00888707001182,  1610GB (BCH003189) 00888707001205,  1610GB (BCH003190) 00888707001205,  1610DE (BCH003191) 00888707001182,  1610IT (BCH003195) 00888707001212,  1610DE (BCH003209) 00888707001182,  1610DE (BCH003214) 00888707001182,  1610DE (BCH003263) 00888707001182,  1610DE (BCH003273) 00888707001182,  1610DE (BCH003275) 00888707001182,  1610CA (BCH003384) 00888707001175,  1610DE (BCH003456) 00888707001182,  1610GB (BCH003147) 00888707001205,  1610DE (BCH003157) 00888707001182,  1610DE (BCH003178) 00888707001182,  1610DE (BCH003243) 00888707001182,  1610DE (BCH003253) 00888707001182,  1610DE (BCH003254) 00888707001182,  1610GB (BCH003429) 00888707001205,  1610IT (BCH003211) 00888707001212,  1610DE (BCH003300) 00888707001182,  1610DE (BCH003424) 00888707001182,  1610DE (BCH003448) 00888707001182,  1610DE (BCH003469) 00888707001182,  1610DE (BCH003475) 00888707001182,  1610CH (BCH003230) 00888707001199,  1610DE (BCH003408) 00888707001182,  1610DE (BCH003423) 00888707001182,  1610DE (BCH003449) 00888707001182,  1610DE (BCH003450) 00888707001182,  1610DE (BCH003495) 00888707001182,  1610DE (BCH003471) 00888707001182,  1610DE (BCH003400) 00888707001182,  1610DE (BCH003241) 00888707001182,  1610DE (BCH003212) 00888707001182,  1610DE (BCH003240) 00888707001182,  1610DE (BCH003473) 00888707001182,  1610DE (BCH003220) 00888707001182,  1610CH (BCH003221) 00888707001199,  1610DE (BCH003293) 00888707001182,  1610DE (BCH003222) 00888707001182,  1610DE (BCH003422) 00888707001182,  1610DE (BCH003476) 00888707001182,  1610DE (BCH003494) 00888707001182,  1610DE (BCH003208) 00888707001182,  1610DE (BCH003252) 00888707001182,  1610DE (BCH003403) 00888707001182,  1610DE (BCH003197) 00888707001182,  1610DE (BCH003216) 00888707001182,  1610DE (BCH003219) 00888707001182,  1610DE (BCH003229) 00888707001182,  1610DE (BCH003231) 00888707001182,  1610DE (BCH003235) 00888707001182,  1610DE (BCH003242) 00888707001182,  1610DE (BCH003249) 00888707001182,  1610DE (BCH003250) 00888707001182,  1610DE (BCH003251) 00888707001182,  1610CA (BCH003383) 00888707001175,  1610IN (BCH003435) 00888707000536,  1610DE (BCH003781) 00888707001182,  1610DE (BCH003154) 00888707001182,  1610DE (BCH003164) 00888707001182,  1610DE (BCH003179) 00888707001182,  1610DE (BCH003228) 00888707001182,  1610DE (BCH003289) 00888707001182,  1610DE (BCH003236) 00888707001182,  1610DE (BCH003137) 00888707001182,  1610DE (BCH003141) 00888707001182,  1610DE (BCH003155) 00888707001182,  1610DE (BCH003162) 00888707001182,  1610DE (BCH003200) 00888707001182,  1610DE (BCH003205) 00888707001182,  1610DE (BCH003232) 00888707001182,  1610DE (BCH003259) 00888707001182,  1610DE (BCH003279) 00888707001182,  1610DE (BCH003280) 00888707001182,  1610DE (BCH003754) 00888707001182,  1610DE (BCH003761) 00888707001182,  1610DE (BCH003774) 00888707001182,  1610DE (BCH003780) 00888707001182,  1610DE (BCH003140) 00888707001182,  1610DE (BCH003152) 00888707001182,  1610DE (BCH003153) 00888707001182,  1610CH (BCH003161) 00888707001199,  1610DE (BCH003163) 00888707001182,  1610DE (BCH003237) 00888707001182,  1610CA (BCH003386) 00888707001175,  1610DE (BCH003133) 00888707001182,  1610DE (BCH003136) 00888707001182,  1610GB (BCH003184) 00888707001205,  1610DE (BCH003215) 00888707001182,  1610DE (BCH003239) 00888707001182,  1610IN (BCH003257) 00888707000536,  1610DE (BCH003292) 00888707001182,  1610DE (BCH003299) 00888707001182,  1610DE (BCH003138) 00888707001182,  1610IN (BCH003255) 00888707000536,  1610DE (BCH003276) 00888707001182,  1610DE (BCH003290) 00888707001182,  1610DE (BCH003751) 00888707001182,  1610GB (BCH003160) 00888707001205,  1610DE (BCH003274) 00888707001182,  1610GB (BCH003166) 00888707001205,  1610DE (BCH003225) 00888707001182,  1610DE (BCH003261) 00888707001182,  1610DE (BCH003202) 00888707001182,  1610GB (BCH003264) 00888707001205,  1610DE (BCH003127) 00888707001182,  1610DE (BCH003165) 00888707001182,  1610DE (BCH003173) 00888707001182,  1610DE (BCH003188) 00888707001182,  1610CH (BCH003234) 00888707001199,  1610DE (BCH003265) 00888707001182,  1610DE (BCH003267) 00888707001182,  1610GB (BCH003286) 00888707001205,  1610DE (BCH003295) 00888707001182,  1610DE (BCH003746) 00888707001182,  1610DE (BCH003129) 00888707001182,  1610DE (BCH003176) 00888707001182,  1610DE (BCH003180) 00888707001182,  1610IT (BCH003203) 00888707001212,  1610DE (BCH003488) 00888707001182,  1610IT (BCH003771) 00888707001212,  1610DE (BCH003130) 00888707001182,  1610DE (BCH003175) 00888707001182,  1610CH (BCH003233) 00888707001199,  1610CH (BCH003146) 00888707001199,  1610GB (BCH003192) 00888707001205,  1610DE (BCH003145) 00888707001182,  1610DE (BCH003174) 00888707001182,  1610IT (BCH003194) 00888707001212,  1610GB (BCH003210) 00888707001205,  1610CH (BCH003256) 00888707001199,  1610CA (BCH003284) 00888707001175,  1610IT (BCH003757) 00888707001212,  1610IT (BCH003767) 00888707001212,  1610DE (BCH003132) 00888707001182,  1610GB (BCH003151) 00888707001205,  1610DE (BCH003167) 00888707001182,  1610DE (BCH003168) 00888707001182,  1610GB (BCH003169) 00888707001205,  1610GB (BCH003247) 00888707001205,  1610DE (BCH003262) 00888707001182,  1610DE (BCH003266) 00888707001182,  1610DE (BCH003128) 00888707001182,  1610DE (BCH003170) 00888707001182,  1610GB (BCH003181) 00888707001205,  1610GB (BCH003182) 00888707001205,  1610DE (BCH003199) 00888707001182,  1610DE (BCH003287) 00888707001182,  1610DE (BCH003109) 00888707001182,  1610DE (BCH003148) 00888707001182,  1610DE (BCH003171) 00888707001182,  1610DE (BCH003198) 00888707001182,  1610DE (BCH003204) 00888707001182,  1610DE (BCH003268) 00888707001182,  1610DE (BCH003271) 00888707001182,  1610DE (BCH003291) 00888707001182,  1610DE (BCH003294) 00888707001182,  1610GB (BCH003246) 00888707001205,  1610DE (BCH003187) 00888707001182,  1610DE (BCH003206) 00888707001182,  1610CH (BCH003227) 00888707001199,  1610GB (BCH003134) 00888707001205,  1610DE (BCH003139) 00888707001182,  1610DE (BCH003270) 00888707001182,  1610DE (BCH003278) 00888707001182,  1610DE (BCH003281) 00888707001182,  1610DE (BCH003296) 00888707001182,  1610DE (BCH003759) 00888707001182,  1610DE (BCH003760) 00888707001182,  1610DE (BCH003787) 00888707001182,  1610DE (BCH003752) 00888707001182,  1610DE (BCH003786) 00888707001182,  1610DE (BCH003497) 00888707001182,  1610DE (BCH003785) 00888707001182,  1610DE (BCH003790) 00888707001182,  1610GB (BCH003201) 00888707001205,  1610DE (BCH003762) 00888707001182,  1610DE (BCH003789) 00888707001182,  1610DE (BCH003117) 00888707001182,  1610GB (BCH003248) 00888707001205,  1610CH (BCH003269) 00888707001199,  1610DE (BCH003744) 00888707001182,  1610DE (BCH003745) 00888707001182,  1610DE (BCH003755) 00888707001182,  1610DE (BCH003764) 00888707001182,  1610DE (BCH003765) 00888707001182,  1610DE (BCH003772) 00888707001182,  1610DE (BCH003777) 00888707001182,  1610GB (BCH003186) 00888707001205,  1610GB (BCH003226) 00888707001205,  1610GB (BCH003277) 00888707001205,  1610DE (BCH003753) 00888707001182,  1610DE (BCH003770) 00888707001182,  1610DE (BCH003779) 00888707001182,  1610DE (BCH002984) 00888707001182,  1610DE (BCH003112) 00888707001182,  1610DE (BCH003113) 00888707001182,  1610DE (BCH003116) 00888707001182,  1610GB (BCH003297) 00888707001205,  1610DE (BCH002840) 00888707001182,  1610GB (BCH003149) 00888707001205,  1610GB (BCH003150) 00888707001205,  1610DE (BCH002852) 00888707001182,  1610DE (BCH003486) 00888707001182,  1610DE (BCH003778) 00888707001182,  1610DE (BCH003485) 00888707001182,  1610GB (BCH003758) 00888707001205,  1610CA (BCH003272) 00888707001175,  1610GB (BCH003064) 00888707001205,  1610CA (BCH003756) 00888707001175,  1610CA (BCH003763) 00888707001175,  1610GB (BCH003766) 00888707001205,  1610DE (BCH003115) 00888707001182,  1610DE (BCH003118) 00888707001182,  1610DE (BCH003496) 00888707001182,  1610GB (BCH003110) 00888707001205,  1610GB (BCH003121) 00888707001205,  1610DE (BCH003123) 00888707001182,  1610GB (BCH003124) 00888707001205,  1610IT (BCH003125) 00888707001212,  1610DE (BCH003484) 00888707001182,  1610IT (BCH003487) 00888707001212,  1610IT (BCH003489) 00888707001212,  1610IT (BCH003492) 00888707001212,  1610GB (BCH003782) 00888707001205,  1610GB (BCH003783) 00888707001205,  1610DE (BCH002843) 00888707001182,  1610DE (BCH003122) 00888707001182,  1610DE (BCH003491) 00888707001182,  1610GB (BCH003769) 00888707001205,  1610CA (BCH003768) 00888707001175,  1610CA (BCH003114) 00888707001175,  1610IT (BCH003119) 00888707001212,  1610IT (BCH003773) 00888707001212,  1610IT (BCH003077) 00888707001212,  1610IT (BCH003111) 00888707001212. 1610DE (BCH002854) 00888707001182,  1610CA (BCH003490) 00888707001175,  1610CA (BCH003776) 00888707001175,  1610DE (BCH002846) 00888707001182,  1610DE (BCH002850) 00888707001182,  1610DE (BCH002861) 00888707001182,  1610DE (BCH002837) 00888707001182,  1610DE (BCH002860) 00888707001182,  1610DE (BCH002836) 00888707001182,  1610CA (BCH002856) 00888707001175,  1610CA (BCH003126) 00888707001175,  1610DE (BCH002839) 00888707001182,  1610DE (BCH002847) 00888707001182,  1610DE (BCH002849) 00888707001182,  1610DE (BCH002857) 00888707001182,  1610DE (BCH002844) 00888707001182,  1610IT (BCH002853) 00888707001212,  1610DE (BCH002842) 00888707001182,  1610DE (BCH002841) 00888707001182,  1610GB (BCH002838) 00888707001205,  1610GB (BCH002845) 00888707001205,  1610IT (BCH002858) 00888707001212,  1610DE (BCH002352) 00888707001182,  1610IT (BCH002848) 00888707001212,  1610DE (BCH002346) 00888707001182,  1610DE (BCH002459) 00888707001182,  1610DE (BCH002451) 00888707001182,  1610GB (BCH002700) 00888707001205,  1610CH (BCH002835) 00888707001199,  1610DE (BCH002859) 00888707001182,  1610CA (BCH002851) 00888707001175,  1610DE (BCH002454) 00888707001182,  1610DE (BCH002387) 00888707001182,  1610DE (BCH002455) 00888707001182,  1610DE (BCH002450) 00888707001182,  1610CA (BCH002044) 00888707005760,  1610CA (BCH002855) 00888707001175,  1610DE (BCH002701) 00888707001182,  1610DE (BCH002389) 00888707001182,  1610DE (BCH002457) 00888707001182,  1610DE (BCH002456) 00888707001182,  1610DE (BCH002460) 00888707001182,  1610DE (BCH002453) 00888707001182,  1610DE (BCH002385) 00888707001182,  1610DE (BCH002390) 00888707001182,  1610DE (BCH002452) 00888707001182,  1610DE (BCH002378) 00888707001182,  1610DE (BCH002461) 00888707001182,  1610DE (BCH002386) 00888707001182,  1610DE (BCH002388) 00888707001182,  1610DE (BCH002268) 00888707001182,  1610GB (BCH002458) 00888707001205,  1610DE (BCH002255) 00888707001182,  1610DE (BCH002351) 00888707001182,  1610DE (BCH002529) 00888707001182,  1610DE (BCH002317) 00888707001182,  1610GB (BCH002344) 00888707001205,  1610DE (BCH002234) 00888707001182,  1610DE (BCH002532) 00888707001182,  1610DE (BCH002348) 00888707001182,  1610GB (BCH002322) 00888707001205,  1610DE (BCH002289) 00888707001182,  1610DE (BCH002354) 00888707001182,  1610DE (BCH002303) 00888707001182,  1610DE (BCH002343) 00888707001182,  1610DE (BCH002350) 00888707001182,  1610DE (BCH002333) 00888707001182,  1610DE (BCH002349) 00888707001182,  1610DE (BCH002294) 00888707001182,  1610CA (BCH002353) 00888707001175,  1610DE (BCH002347) 00888707001182,  1610GB (BCH002281) 00888707001205,  1610DE (BCH002275) 00888707001182,  1610DE (BCH001176) 00888707001182,  1610DE (BCH001178) 00888707001182,  1610DE (BCH001390) 00888707001182,  1610DE (BCH001174) 00888707001182,  1610DE (BCH001175) 00888707001182,  1610IT (BCH001205) 00888707001212,  1610DE (BCH000777) 00888707002202,  1610DE (BCH000864) 00888707002967,  1610DE (BCH001168) 00888707001182,  1610DE (BCH002087) 00888707001182,  1610GB (BCH000800) 00888707001205,  1610GB (BCH000825) 00888707001205,  1610DE (BCH001182) 00888707001182,  1610DE (BCH000816) 00888707001182,  1610GB (BCH000810) 00888707001205,  1610IN (BCH000927) 00888707002233,  1610CA (BCH001160) 00888707003001,  1610DE (BCH000775) 00888707002202,  1610GB (BCH000824) 00888707001205,  1610DE (BCH000694) 00888707002202,  1610DE (BCH001472) 00888707001182,  1610GB (BCH000791) 00888707001205,  1610DE (BCH000796) 00888707001182,  1610DE (BCH001471) 00888707001182,  1610DE (BCH000226) 00888707002202,  1610DE (BCH000764) 00888707002202,  1610DE (BCH000706) 00888707002202,  1610DE (BCH000809) 00888707001182,  1610DE (BCH000817) 00888707001182,  1610DE (BCH000823) 00888707001182,  1610DE (BCH001432) 00888707001182,  1610DE (BCH001470) 00888707001182,  1610GB (BCH001055) 00888707002974,  1610GB (BCH000801) 00888707001205,  1610DE (BCH000794) 00888707001182,  1610GB (BCH000799) 00888707001205,  1610DE (BCH000822) 00888707001182,  1610DE (BCH000826) 00888707001182,  1610DE (BCH000676) 00888707002202,  1610GB (BCH000802) 00888707001205,  1610DE (BCH000667) 00888707002202,  1610DE (BCH001221) 00888707001182,  1610DE (BCH000784) 00888707001182,  1610DE (BCH000693) 00888707002202,  1610AU (BCH001039) 00888707002165,  1610DE (BCH000814) 00888707001182,  1610DE (BCH000691) 00888707002202,  1610DE (BCH000818) 00888707001182,  1610DE (BCH000815) 00888707001182,  1610DE (BCH000793) 00888707001182,  1610DE (BCH000795) 00888707001182,  1610DE (BCH000805) 00888707001182,  1610DE (BCH000812) 00888707001182,  1610DE (BCH000792) 00888707001182,  1610DE (BCH000798) 00888707001182,  1610IL (BCH000698) 00888707002226,  1610DE (BCH000811) 00888707001182,  1610DE (BCH000790) 00888707001182,  1610DE (BCH000763) 00888707002202,  1610DE (BCH000850) 00888707001182,  1610DE (BCH000813) 00888707001182,  1610DE (BCH000804) 00888707001182,  1610IT (BCH000895) 00888707001212,  1610BR (BCH000803) 00888707001168,  1610DE (BCH000797) 00888707001182,  1610DE (BCH000769) 00888707002202,  1610DE (BCH000741) 00888707002202,  1610IL (BCH000779) 00888707002226,  1610DE (BCH000785) 00888707001182,  1610DE (BCH000696) 00888707002202,  1610DE (BCH000745) 00888707002202,  1610DE (BCH000786) 00888707001182,  1610DE (BCH000748) 00888707002202,  1610DE (BCH000820) 00888707001182,  1610IL (BCH000778) 00888707002226,  1610DE (BCH000819) 00888707001182,  1610DE (BCH000637) 00888707002202,  1610DE (BCH000644) 00888707002202,  1610DE (BCH000645) 00888707002202,  1610IN (BCH000536) 00888707002233,  1610CH (BCH000595) 00888707002196,  1610DE (BCH000633) 00888707002202,  1610DE (BCH000807) 00888707001182,  1610DE (BCH000761) 00888707002202,  1610DE (BCH000688) 00888707002202,  1610DE (BCH000788) 00888707001182,  1610DE (BCH000827) 00888707001182,  1610DE (BCH000848) 00888707002967,  1610DE (BCH000806) 00888707001182,  1610DE (BCH000789) 00888707001182,  1610DE (BCH000828) 00888707001182,  1610DE (BCH000762) 00888707002202,  1610DE (BCH000787) 00888707001182,  1610DE (BCH000133) 00888707002202,  1610DE (BCH000601) 00888707002202,  1610DE (BCH000298) 00888707002202,  1610CH (BCH000167) 00888707001199,  1610IN (BCH000545) 00888707002233,  1610DE (BCH000201) 00888707002202,  1610DE (BCH000202) 00888707002202,  1610DE (BCH000221) 00888707002202,  1610DE (BCH000289) 00888707002202,  1610DE (BCH000260) 00888707002202,  1610DE (BCH000228) 00888707002202,  1610DE (BCH000258) 00888707002202,  1610DE (BCH000254) 00888707002202,  1610DE (BCH000215) 00888707002202,  1610DE (BCH000137) 00888707002202,  1610DE (BCH000016) 00888707002202,  1610DE (BCH000265) 00888707002202,  1610DE (BCH000273) 00888707002202,  1610DE (BCH000263) 00888707002202,  1610DE (BCH000264) 00888707002202,  1610DE (BCH000239) 00888707002202,  1610DE (BCH000291) 00888707002202,  1610DE (BCH000223) 00888707002202,  1610DE (BCH000211) 00888707002202,  1610DE (BCH000062) 00888707002202,  1610DE (BCH000009) 00888707002202,  1610DE (BCH000046) 00888707002202,  1610DE (BCH000060) 00888707002202,  1610DE (BCH000070) 00888707002202,  
FEI Number 3007042319
Recalling Firm/
Manufacturer
Heartware, Inc.
14400 Nw 60th Ave
Miami Lakes FL 33014-2807
For Additional Information Contact
877-367-4823
Manufacturer Reason
for Recall
Potential for an HVAD System user to mistakenly insert the Battery Charger AC Adapter into an HVAD Controller power port.
FDA Determined
Cause 2
Device Design
ActionHeartware notified customers on about 01/23/2020 via "Urgent Field Safety Notice" letter. Customers were notified of the potential for an HVAD System user to mistakenly insert the Battery Charger AC Adapter into an HVAD Controller power port. When this occurs, communication circuits between the battery and the Controller may be damaged. If circuit damage occurs, it will be necessary to exchange the Controller to restore communication between the batteries and Controller. Instructions included to advise patients to carefully follow the guidance provided in the patient manual related to connecting a power source. Pay attention when connecting and disconnecting power supply cables to ensure users verify the correct cable is being connected to their Controller. Patients were instructed to report any persistent, unexpected audible tones to their Clinician for additional instructions. Additional instructions included to ensure all personnel within your organization or any customers if further distributed are notified of this recall. If you have any questions, please contact the Cardiac and Vascular Group Customer Focused Quality at 763-526-2513 or by email to: carlos.j.alfonzo@medtronic.com.
Quantity in Commerce5,489 devices total
DistributionNo affected devices distributed within the US. International distribution to Australia, Austria, Belgium, Brazil, Canada, Chile, Colombia, Croatia, Czech Republic, Denmark, EGYPT, Estonia, Finland, France, Germany, Greece, Hong Kong, India, Israel, Italy, Japan, Kazakhstan, Korea, Korea, Republic of, Kuwait, Latvia, Lebanon, Lithuania, Luxembourg, Macedonia, Malaysia, Netherlands, New Zealand, Norway, Poland, Romania, Saudi Arabia, Serbia, Singapore, Slovakia, South Africa, Spain, Sweden, Switzerland, Taiwan, Taiwan, Province of China, Turkey, United Arab Emirates, United Kingdom, Vietnam.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
PMA DatabasePMAs with Product Code = DSQ
-
-