| | Class 2 Device Recall Ultraview SL Capnography Module |  |
| Date Initiated by Firm | May 12, 2011 |
| Date Posted | June 20, 2011 |
| Recall Status1 |
Terminated 3 on December 02, 2011 |
| Recall Number | Z-2573-2011 |
| Recall Event ID |
58811 |
| 510(K)Number | K031124 |
| Product Classification |
Monitor, physiological, patient (with arrhythmia detection or alarms) - Product Code MHX
|
| Product | The Ultraview SL 91517 Capnography module is a mainstream and sidestream gas analyzer designed to measure the concentration of carbon dioxide in a gas mixture, to aid in determining the patient's ventilatory, circulatory, and metabolic status.
Mainstream monitoring uses a self-calibrating CO2 sensor to ensure continuous monitoring without interruption of connection. Reusable, lightweight (<14 grams), or single-use airway adapters are available for both adult and neonatal (low dead space) patients.
Respiration can be monitored via an airway adapter and endotracheal or tracheostomy tubes using the mainstream function of the module.
The intended use statement that apply to the 91517 are continuous monitoring of carbon dioxide and oxygen. The Model 95000-A Bedside/Transport Monitor is used at the patient's bedside to actively acquire, analyze, monitor and display a variety of clinical parameters (in clinically relevant formats) for adults, pediatric or neonatal patients. The Model 95000-A Bedside/Transport Monitor is designed for use by a healthcare professional in a hospital, healthcare facility (i.e., clinic), or intra-hospital transport. This product is not designed for home use or for use in transport vehicles. |
| Code Information |
Serial numbers: 1517-000001, 1517-000002, 1517-000003, 1517-000004, 1517-000005, 1517-000006, 1517-000007, 1517-000008, 1517-000009, 1517-000010, 1517-000011, 1517-000012, 1517-000013, 1517-000014, 1517-000015, 1517-000016, 1517-000017, 1517-000018, 1517-000019, 1517-000020, 1517-000021, 1517-000022, 1517-000023, 1517-000024, 1517-000025, 1517-000026, 1517-000027, 1517-000028, 1517-000029, 1517-000030, 1517-000031, 1517-000032, 1517-000033, 1517-000034, 1517-000035, 1517-000036, 1517-000037, 1517-000038, 1517-000039, 1517-000040, 1517-000041, 1517-000042, 1517-000043, 1517-000044, 1517-000045, 1517-000047, 1517-000049, 1517-000050, 1517-000051, 1517-000052, 1517-000053, 1517-000054, 1517-000055, 1517-000056, 1517-000057, 1517-000058, 1517-000059, 1517-000060, 1517-000061, 1517-000062, 1517-000063, 1517-000064, 1517-000065, 1517-000066, 1517-000067, 1517-000068, 1517-000069, 1517-000070, 1517-000071, 1517-000072, 1517-000073, 1517-000074, 1517-000075, 1517-000076, 1517-000077, 1517-000078, 1517-000079, 1517-000080, 1517-000081, 1517-000082, 1517-000083, 1517-000084, 1517-000085, 1517-000086, 1517-000087, 1517-000088, 1517-000089, 1517-000090, 1517-000091, 1517-000092, 1517-000093, 1517-000094, 1517-000095, 1517-000096, 1517-000097, 1517-000098, 1517-000099, 1517-000104, 1517-000105, 1517-000106, 1517-000107, 1517-000108, 1517-000109, 1517-000110, 1517-000111, 1517-000112, 1517-000113, 1517-000114, 1517-000115, 1517-000116, 1517-000117, 1517-000118, 1517-000119, 1517-000120, 1517-000121, 1517-000122, 1517-000123, 1517-000124, 1517-000125, 1517-000126, 1517-000127, 1517-000128, 1517-000129, 1517-000130, 1517-000131, 1517-000132, 1517-000133, 1517-000134, 1517-000135, 1517-000136, 1517-000137, 1517-000138, 1517-000139, 1517-000140, 1517-000141, 1517-000142, 1517-000143, 1517-000144, 1517-000145, 1517-000146, 1517-000147, 1517-000148, 1517-000149, 1517-000150, 1517-000151, 1517-000152, 1517-000153, 1517-000154, 1517-000155, 1517-000156, 1517-000157, 1517-000158, 1517-000159, 1517-000160, 1517-000161, 1517-000162, 1517-000163, 1517-000164, 1517-000165, 1517-000166, 1517-000167, 1517-000168, 1517-000169, 1517-000170, 1517-000171, 1517-000172, 1517-000173, 1517-000174, 1517-000175, 1517-000176, 1517-000177, 1517-000178, 1517-000179, 1517-000180, 1517-000181, 1517-000182, 1517-000183, 1517-000184, 1517-000185, 1517-000186, 1517-000187, 1517-000188, 1517-000189, 1517-000190, 1517-000191, 1517-000192, 1517-000193, 1517-000194, 1517-000195, 1517-000196, 1517-000197, 1517-000198, 1517-000199, 1517-000200, 1517-000201, 1517-000202, 1517-000203, 1517-000204, 1517-000205, 1517-000206, 1517-000207, 1517-000208, 1517-000209, 1517-000210, 1517-000211, 1517-000212, 1517-000213, 1517-000214, 1517-000215, 1517-000216, 1517-000217, 1517-000218, 1517-000219, 1517-000220, 1517-000221, 1517-000222, 1517-000223, 1517-000224, 1517-000225, 1517-000226, 1517-000227, 1517-000228, 1517-000229, 1517-000230, 1517-000231, 1517-000232, 1517-000233, 1517-000234, 1517-000235, 1517-000236, 1517-000237, 1517-000238, 1517-000239, 1517-000240, 1517-000241, 1517-000242, 1517-000243, 1517-000244, 1517-000245, 1517-000246, 1517-000247, 1517-000248, 1517-000249, 1517-000250, 1517-000251, 1517-000252, 1517-000253, 1517-000254, 1517-000255, 1517-000256, 1517-000257, 1517-000258, 1517-000259, 1517-000260, 1517-000261, 1517-000262, 1517-000263, 1517-000264, 1517-000265, 1517-000266, 1517-000267, 1517-000268, 1517-000269, 1517-000270, 1517-000271, 1517-000272, 1517-000273, 1517-000274, 1517-000275, 1517-000276, 1517-000277, 1517-000278, 1517-000279, 1517-0002797, 1517-000280, 1517-0002801, 1517-000281, 1517-000282, 1517-000283, 1517-000284, 1517-000285, 1517-000286, 1517-000287, 1517-000288, 1517-000289, 1517-000290, 1517-000291, 1517-000292, 1517-000293, 1517-000294, 1517-000295, 1517-000296, 1517-000297, 1517-000298, 1517-000299, 1517-000300, 1517-000301, 1517-000302, 1517-000303, 1517-000304, 1517-000305, 1517-000306, 1517-000307, 1517-000308, 1517-000309, 1517-000310, 1517-000311, 1517-000312, 1517-000313, 1517-000314, 1517-000315, 1517-000316, 1517-000317, 1517-000318, 1517-000319, 1517-000320, 1517-000321, 1517-000322, 1517-000323, 1517-000324, 1517-000325, 1517-000326, 1517-000327, 1517-000328, 1517-000329, 1517-000330, 1517-000331, 1517-000332, 1517-000333, 1517-000334, 1517-000335, 1517-000336, 1517-000337, 1517-000338, 1517-000339, 1517-000340, 1517-000341, 1517-000342, 1517-000343, 1517-000344, 1517-000345, 1517-000346, 1517-000347, 1517-000348, 1517-000349, 1517-000350, 1517-000351, 1517-000352, 1517-000353, 1517-000354, 1517-000355, 1517-000356, 1517-000357, 1517-000358, 1517-000359, 1517-000360, 1517-000361, 1517-000362, 1517-000363, 1517-000364, 1517-000365, 1517-000366, 1517-000367, 1517-000368, 1517-000369, 1517-000370, 1517-000371, 1517-000372, 1517-000373, 1517-000374, 1517-000375, 1517-000376, 1517-000377, 1517-000378, 1517-000379, 1517-000380, 1517-000381, 1517-000382, 1517-000383, 1517-000384, 1517-000385, 1517-000386, 1517-000387, 1517-000388, 1517-000389, 1517-000390, 1517-000391, 1517-000392, 1517-000393, 1517-000394, 1517-000395, 1517-000396, 1517-000397, 1517-000398, 1517-000399, 1517-000400, 1517-000401, 1517-000402, 1517-000403, 1517-000404, 1517-000405, 1517-000406, 1517-000407, 1517-000408, 1517-000409, 1517-000410, 1517-000411, 1517-000412, 1517-000413, 1517-000414, 1517-000415, 1517-000416, 1517-000417, 1517-000418, 1517-000419, 1517-000420, 1517-000421, 1517-000422, 1517-000423, 1517-000424, 1517-000425, 1517-000426, 1517-000427, 1517-000428, 1517-000429, 1517-000430, 1517-000431, 1517-000432, 1517-000433, 1517-000434, 1517-000435, 1517-000436, 1517-000437, 1517-000438, 1517-000439, 1517-000440, 1517-000441, 1517-000442, 1517-000443, 1517-000444, 1517-000445, 1517-000446, 1517-000447, 1517-000448, 1517-000449, 1517-000450, 1517-000451, 1517-000452, 1517-000453, 1517-000454, 1517-000455, 1517-000456, 1517-000457, 1517-000458, 1517-000459, 1517-000460, 1517-000461, 1517-000462, 1517-000463, 1517-000464, 1517-000465, 1517-000468, 1517-000469, 1517-000470, 1517-000471, 1517-000472, 1517-000473, 1517-000474, 1517-000475, 1517-000476, 1517-000477, 1517-000478, 1517-000479, 1517-000480, 1517-000481, 1517-000482, 1517-000483, 1517-000484, 1517-000485, 1517-000486, 1517-000487, 1517-000488, 1517-000489, 1517-000490, 1517-000491, 1517-000492, 1517-000493, 1517-000494, 1517-000495, 1517-000496, 1517-000497, 1517-000498, 1517-000499, 1517-000500, 1517-000501, 1517-000502, 1517-000503, 1517-000504, 1517-000505, 1517-000506, 1517-000507, 1517-000508, 1517-000509, 1517-000510, 1517-000511, 1517-000512, 1517-000513, 1517-000514, 1517-000515, 1517-000516, 1517-000517, 1517-000518, 1517-000519, 1517-000520, 1517-000521, 1517-000522, 1517-000523, 1517-000524, 1517-000525, 1517-000526, 1517-000527, 1517-000528, 1517-000529, 1517-000530, 1517-000531, 1517-000532, 1517-000533, 1517-000534, 1517-000535, 1517-000536, 1517-000537, 1517-000538, 1517-000539, 1517-000540, 1517-000541, 1517-000542, 1517-000543, 1517-000544, 1517-000545, 1517-000546, 1517-000547, 1517-000548, 1517-000549, 1517-000550, 1517-000551, 1517-000552, 1517-000553, 1517-000554, 1517-000555, 1517-000556, 1517-000557, 1517-000558, 1517-000559, 1517-000560, 1517-000561, 1517-000562, 1517-000563, 1517-000564, 1517-000565, 1517-000566, 1517-000567, 1517-000568, 1517-000569, 1517-000570, 1517-000571, 1517-000573, 1517-000574, 1517-000575, 1517-000576, 1517-000577, 1517-000578, 1517-000579, 1517-000580, 1517-000581, 1517-000582, 1517-000583, 1517-000584, 1517-000585, 1517-000586, 1517-000587, 1517-000588, 1517-000589, 1517-000590, 1517-000591, 1517-000592, 1517-000593, 1517-000594, 1517-000595, 1517-000596, 1517-000597, 1517-000598, 1517-000599, 1517-000600, 1517-000601, 1517-000602, 1517-000603, 1517-000604, 1517-000605, 1517-000606, 1517-000607, 1517-000608, 1517-000609, 1517-000610, 1517-000611, 1517-000612, 1517-000613, 1517-000614, 1517-000615, 1517-000616, 1517-000617, 1517-000618, 1517-000619, 1517-000620, 1517-000621, 1517-000622, 1517-000623, 1517-000624, 1517-000625, 1517-000626, 1517-000627, 1517-000628, 1517-000629, 1517-000630, 1517-000631, 1517-000632, 1517-000633, 1517-000634, 1517-000635, 1517-000636, 1517-000637, 1517-000638, 1517-000639, 1517-000640, 1517-000641, 1517-000642, 1517-000643, 1517-000644, 1517-000645, 1517-000646, 1517-000647, 1517-000648, 1517-000649, 1517-000650, 1517-000651, 1517-000652, 1517-000653, 1517-000654, 1517-000655, 1517-000656, 1517-000657, 1517-000658, 1517-000659, 1517-000660, 1517-000661, 1517-000662, 1517-000663, 1517-000664, 1517-000665, 1517-000666, 1517-000667, 1517-000668, 1517-000669, 1517-000670, 1517-000671, 1517-000672, 1517-000673, 1517-000674, 1517-000675, 1517-000676, 1517-000677, 1517-000678, 1517-000679, 1517-000680, 1517-000681, 1517-000682, 1517-000683, 1517-000684, 1517-000685, 1517-000686, 1517-000687, 1517-000688, 1517-000689, 1517-000690, 1517-000691, 1517-000692, 1517-000693, 1517-000694, 1517-000695, 1517-000696, 1517-000697, 1517-000698, 1517-000699, 1517-000700, 1517-000701, 1517-000702, 1517-000703, 1517-000704, 1517-000705, 1517-000706, 1517-000707, 1517-000708, 1517-000709, 1517-000710, 1517-000711, 1517-000712, 1517-000713, 1517-000714, 1517-000715, 1517-000716, 1517-000717, 1517-000718, 1517-000719, 1517-000720, 1517-000721, 1517-000722, 1517-000723, 1517-000724, 1517-000725, 1517-000726, 1517-000727, 1517-000728, 1517-000729, 1517-000730, 1517-000731, 1517-000732, 1517-000733, 1517-000734, 1517-000735, 1517-000736, 1517-000737, 1517-000738, 1517-000739, 1517-000740, 1517-000741, 1517-000742, 1517-000743, 1517-000744, 1517-000745, 1517-000746, 1517-000747, 1517-000748, 1517-000749, 1517-000750, 1517-000751, 1517-000752, 1517-000753, 1517-000754, 1517-000755, 1517-000756, 1517-000757, 1517-000758, 1517-000759, 1517-000760, 1517-000761, 1517-000762, 1517-000763, 1517-000764, 1517-000765, 1517-000766, 1517-000767, 1517-000768, 1517-000769, 1517-000770, 1517-000771, 1517-000772, 1517-000773, 1517-000774, 1517-000775, 1517-000776, 1517-000777, 1517-000778, 1517-000779, 1517-000780, 1517-000781, 1517-000782, 1517-000783, 1517-000784, 1517-000785, 1517-000786, 1517-000787, 1517-000788, 1517-000789, 1517-000790, 1517-000791, 1517-000792, 1517-000793, 1517-000794, 1517-000795, 1517-000796, 1517-000797, 1517-000798, 1517-000799, 1517-000800, 1517-000801, 1517-000802, 1517-000803, 1517-000804, 1517-000805, 1517-000806, 1517-000807, 1517-000808, 1517-000809, 1517-000810, 1517-000811, 1517-000812, 1517-000813, 1517-000814, 1517-000815, 1517-000816, 1517-000817, 1517-000818, 1517-000819, 1517-000820, 1517-000821, 1517-000822, 1517-000823, 1517-000824, 1517-000825, 1517-000826, 1517-000827, 1517-000828, 1517-000829, 1517-000830, 1517-000831, 1517-000832, 1517-000833, 1517-000834, 1517-000835, 1517-000836, 1517-000837, 1517-000838, 1517-000839, 1517-000840, 1517-000841, 1517-000842, 1517-000843, 1517-000844, 1517-000845, 1517-000846, 1517-000847, 1517-000848, 1517-000849, 1517-000850, 1517-000851, 1517-000852, 1517-000853, 1517-000854, 1517-000855, 1517-000856, 1517-000857, 1517-000858, 1517-000859, 1517-000860, 1517-000861, 1517-000862, 1517-000863, 1517-000864, 1517-000865, 1517-000866, 1517-000867, 1517-000868, 1517-000869, 1517-000870, 1517-000871, 1517-000872, 1517-000873, 1517-000874, 1517-000875, 1517-000876, 1517-000877, 1517-000878, 1517-000879, 1517-000880, 1517-000881, 1517-000882, 1517-000883, 1517-000884, 1517-000885, 1517-000886, 1517-000887, 1517-000888, 1517-000889, 1517-000890, 1517-000891, 1517-000892, 1517-000893, 1517-000894, 1517-000895, 1517-000896, 1517-000897, 1517-000898, 1517-000899, 1517-000900, 1517-000901, 1517-000902, 1517-000903, 1517-000904, 1517-000905, 1517-000906, 1517-000907, 1517-000908, 1517-000909, 1517-000910, 1517-000911, 1517-000912, 1517-000913, 1517-000914, 1517-000915, 1517-000916, 1517-000917, 1517-000918, 1517-000919, 1517-000920, 1517-000921, 1517-000922, 1517-000923, 1517-000924, 1517-000925, 1517-000926, 1517-000927, 1517-000928, 1517-000929, 1517-000930, 1517-000931, 1517-000932, 1517-000933, 1517-000934, 1517-000935, 1517-000936, 1517-000937, 1517-000938, 1517-000939, 1517-000940, 1517-000941, 1517-000942, 1517-000943, 1517-000944, 1517-000945, 1517-000946, 1517-000947, 1517-000948, 1517-000949, 1517-000950, 1517-000951, 1517-000952, 1517-000953, 1517-000954, 1517-000955, 1517-000956, 1517-000957, 1517-000958, 1517-000959, 1517-000960, 1517-000961, 1517-000962, 1517-000963, 1517-000964, 1517-000965, 1517-000966, 1517-000967, 1517-000968, 1517-000969, 1517-000970, 1517-000971, 1517-000972, 1517-000973, 1517-000974, 1517-000975, 1517-000976, 1517-000977, 1517-000978, 1517-000979, 1517-000980, 1517-000981, 1517-000982, 1517-000983, 1517-000984, 1517-000985, 1517-000986, 1517-000987, 1517-000988, 1517-000989, 1517-000990, 1517-000991, 1517-000992, 1517-000993, 1517-000994, 1517-000995, 1517-000996, 1517-000997, 1517-000998, 1517-000999, 1517-001000, 1517-001001, 1517-001002, 1517-001003, 1517-001004, 1517-001005, 1517-001006, 1517-001007, 1517-001008, 1517-001009, 1517-001010, 1517-001011, 1517-001012, 1517-001013, 1517-001014, 1517-001015, 1517-001016, 1517-001017, 1517-001018, 1517-001019, 1517-001020, 1517-001021, 1517-001022, 1517-001023, 1517-001024, 1517-001025, 1517-001026, 1517-001027, 1517-001028, 1517-001029, 1517-001030, 1517-001031, 1517-001032, 1517-001033, 1517-001034, 1517-001035, 1517-001036, 1517-001037, 1517-001038, 1517-001039, 1517-001042, 1517-001043, 1517-001044, 1517-001045, 1517-001046, 1517-001047, 1517-001048, 1517-001049, 1517-001050, 1517-001051, 1517-001052, 1517-001053, 1517-001054, 1517-001055, 1517-001056, 1517-001057, 1517-001058, 1517-001059, 1517-001060, 1517-001061, 1517-001062, 1517-001063, 1517-001064, 1517-001065, 1517-001066, 1517-001067, 1517-001068, 1517-001069, 1517-001070, 1517-001071, 1517-001072, 1517-001073, 1517-001074, 1517-001075, 1517-001076, 1517-001077, 1517-001078, 1517-001079, 1517-001080, 1517-001081, 1517-001082, 1517-001083, 1517-001084, 1517-001085, 1517-001086, 1517-001087, 1517-001088, 1517-001089, 1517-001090, 1517-001091, 1517-001092, 1517-001093, 1517-001094, 1517-001095, 1517-001096, 1517-001097, 1517-001098, 1517-001099, 1517-001100, 1517-001101, 1517-001102, 1517-001103, 1517-001104, 1517-001105, 1517-001106, 1517-001107, 1517-001108, 1517-001109, 1517-001110, 1517-001111, 1517-001112, 1517-001113, 1517-001114, 1517-001115, 1517-001116, 1517-001117, 1517-001118, 1517-001119, 1517-001120, 1517-001121, 1517-001122, 1517-001123, 1517-001124, 1517-001125, 1517-001126, 1517-001127, 1517-001128, 1517-001129, 1517-001130, 1517-001131, 1517-001132, 1517-001133, 1517-001134, 1517-001135, 1517-001136, 1517-001137, 1517-001138, 1517-001139, 1517-001140, 1517-001141, 1517-001142, 1517-001143, 1517-001144, 1517-001145, 1517-001146, 1517-001147, 1517-001148, 1517-001149, 1517-001150, 1517-001151, 1517-001152, 1517-001153, 1517-001154, 1517-001155, 1517-001156, 1517-001157, 1517-001158, 1517-001159, 1517-001160, 1517-001161, 1517-001162, 1517-001163, 1517-001164, 1517-001165, 1517-001166, 1517-001168, 1517-001169, 1517-001170, 1517-001171, 1517-001172, 1517-001173, 1517-001174, 1517-001175, 1517-001176, 1517-001177, 1517-001178, 1517-001179, 1517-001180, 1517-001181, 1517-001182, 1517-001183, 1517-001184, 1517-001185, 1517-001186, 1517-001187, 1517-001188, 1517-001189, 1517-001190, 1517-001191, 1517-001192, 1517-001193, 1517-001194, 1517-001195, 1517-001196, 1517-001197, 1517-001198, 1517-001199, 1517-001200, 1517-001201, 1517-001202, 1517-001203, 1517-001204, 1517-001205, 1517-001206, 1517-001207, 1517-001208, 1517-001209, 1517-001210, 1517-001211, 1517-001212, 1517-001213, 1517-001214, 1517-001215, 1517-001216, 1517-001217, 1517-001218, 1517-001219, 1517-001220, 1517-001221, 1517-001222, 1517-001223, 1517-001224, 1517-001225, 1517-001226, 1517-001227, 1517-001228, 1517-001229, 1517-001230, 1517-001231, 1517-001232, 1517-001233, 1517-001234, 1517-001235, 1517-001236, 1517-001237, 1517-001238, 1517-001239, 1517-001240, 1517-001241, 1517-001242, 1517-001243, 1517-001244, 1517-001245, 1517-001246, 1517-001247, 1517-001248, 1517-001249, 1517-001250, 1517-001251, 1517-001252, 1517-001253, 1517-001254, 1517-001255, 1517-001256, 1517-001257, 1517-001258, 1517-001259, 1517-001260, 1517-001261, 1517-001262, 1517-001263, 1517-001264, 1517-001265, 1517-001266, 1517-001267, 1517-001268, 1517-001269, 1517-001270, 1517-001271, 1517-001272, 1517-001273, 1517-001274, 1517-001275, 1517-001276, 1517-001277, 1517-001278, 1517-001279, 1517-001280, 1517-001281, 1517-001282, 1517-001283, 1517-001284, 1517-001285, 1517-001286, 1517-001287, 1517-001288, 1517-001289, 1517-001290, 1517-001291, 1517-001292, 1517-001293, 1517-001294, 1517-001295, 1517-001296, 1517-001297, 1517-001298, 1517-001299, 1517-001300, 1517-001301, 1517-001302, 1517-001303, 1517-001304, 1517-001305, 1517-001306, 1517-001307, 1517-001308, 1517-001309, 1517-001310, 1517-001311, 1517-001312, 1517-001313, 1517-001314, 1517-001315, 1517-001316, 1517-001317, 1517-001318, 1517-001319, 1517-001320, 1517-001321, 1517-001322, 1517-001323, 1517-001324, 1517-001325, 1517-001326, 1517-001327, 1517-001328, 1517-001329, 1517-001330, 1517-001331, 1517-001332, 1517-001333, 1517-001334, 1517-001335, 1517-001336, 1517-001337, 1517-001338, 1517-001339, 1517-001340, 1517-001341, 1517-001342, 1517-001343, 1517-001344, 1517-001345, 1517-001346, 1517-001347, 1517-001348, 1517-001349, 1517-001350, 1517-001351, 1517-001352, 1517-001353, 1517-001354, 1517-001355, 1517-001356, 1517-001357, 1517-001358, 1517-001359, 1517-001360, 1517-001361, 1517-001362, 1517-001363, 1517-001364, 1517-001365, 1517-001366, 1517-001367, 1517-001368, 1517-001369, 1517-001370, 1517-001371, 1517-001372, 1517-001373, 1517-001374, 1517-001375, 1517-001376, 1517-001377, 1517-001378, 1517-001379, 1517-001380, 1517-001381, 1517-001382, 1517-001383, 1517-001384, 1517-001385, 1517-001386, 1517-001387, 1517-001388, 1517-001389, 1517-001390, 1517-001391, 1517-001392, 1517-001393, 1517-001394, 1517-001395, 1517-001396, 1517-001397, 1517-001398, 1517-001399, 1517-001400, 1517-001401, 1517-001402, 1517-001403, 1517-001404, 1517-001405, 1517-001406, 1517-001407, 1517-001408, 1517-001409, 1517-001410, 1517-001411, 1517-001412, 1517-001413, 1517-001414, 1517-001415, 1517-001416, 1517-001417, 1517-001418, 1517-001419, 1517-001420, 1517-001421, 1517-001422, 1517-001423, 1517-001424, 1517-001425, 1517-001426, 1517-001427, 1517-001428, 1517-001429, 1517-001430, 1517-001431, 1517-001432, 1517-001433, 1517-001434, 1517-001435, 1517-001436, 1517-001437, 1517-001438, 1517-001439, 1517-001440, 1517-001441, 1517-001442, 1517-001443, 1517-001444, 1517-001445, 1517-001446, 1517-001447, 1517-001448, 1517-001449, 1517-001450, 1517-001451, 1517-001452, 1517-001453, 1517-001454, 1517-001455, 1517-001456, 1517-001457, 1517-001458, 1517-001459, 1517-001460, 1517-001461, 1517-001462, 1517-001463, 1517-001464, 1517-001465, 1517-001466, 1517-001467, 1517-001468, 1517-001469, 1517-001470, 1517-001471, 1517-001472, 1517-001473, 1517-001474, 1517-001475, 1517-001476, 1517-001477, 1517-001478, 1517-001479, 1517-001480, 1517-001481, 1517-001482, 1517-001483, 1517-001484, 1517-001485, 1517-001486, 1517-001487, 1517-001488, 1517-001489, 1517-001490, 1517-001491, 1517-001492, 1517-001493, 1517-001494, 1517-001495, 1517-001496, 1517-001497, 1517-001498, 1517-001499, 1517-001500, 1517-001501, 1517-001502, 1517-001503, 1517-001504, 1517-001505, 1517-001506, 1517-001507, 1517-001508, 1517-001509, 1517-001510, 1517-001511, 1517-001512, 1517-001513, 1517-001514, 1517-001515, 1517-001516, 1517-001517, 1517-001518, 1517-001519, 1517-001520, 1517-001521, 1517-001522, 1517-001523, 1517-001524, 1517-001525, 1517-001526, 1517-001527, 1517-001528, 1517-001529, 1517-001530, 1517-001531, 1517-001532, 1517-001533, 1517-001534, 1517-001535, 1517-001536, 1517-001537, 1517-001538, 1517-001539, 1517-001540, 1517-001541, 1517-001542, 1517-001543, 1517-001544, 1517-001545, 1517-001546, 1517-001547, 1517-001548, 1517-001549, 1517-001550, 1517-001551, 1517-001552, 1517-001553, 1517-001554, 1517-001555, 1517-001556, 1517-001557, 1517-001558, 1517-001559, 1517-001560, 1517-0015605, 1517-0015606, 1517-0015607, 1517-001561, 1517-001562, 1517-001563, 1517-001564, 1517-001565, 1517-001566, 1517-001567, 1517-001568, 1517-001569, 1517-001570, 1517-001571, 1517-001572, 1517-001573, 1517-001574, 1517-001575, 1517-001576, 1517-001577, 1517-001578, 1517-001579, 1517-001580, 1517-001581, 1517-001582, 1517-001583, 1517-001584, 1517-001585, 1517-001586, 1517-001587, 1517-001588, 1517-001589, 1517-001590, 1517-001591, 1517-001592, 1517-001593, 1517-001594, 1517-001595, 1517-001596, 1517-001597, 1517-001598, 1517-001599, 1517-001600, 1517-001601, 1517-001602, 1517-001603, 1517-001604, 1517-001608, 1517-001609, 1517-001610, 1517-001611, 1517-001612, 1517-001613, 1517-001614, 1517-001615, 1517-001616, 1517-001617, 1517-001618, 1517-001619, 1517-001620, 1517-001621, 1517-001622, 1517-001623, 1517-001624, 1517-001625, 1517-001626, 1517-001627, 1517-001628, 1517-001629, 1517-001630, 1517-001631, 1517-001632, 1517-001633, 1517-001634, 1517-001635, 1517-001636, 1517-001637, 1517-001638, 1517-001639, 1517-001640, 1517-001641, 1517-001642, 1517-001643, 1517-001644, 1517-001645, 1517-001646, 1517-001647, 1517-001648, 1517-001649, 1517-001650, 1517-001651, 1517-001652, 1517-001653, 1517-001654, 1517-001655, 1517-001656, 1517-001657, 1517-001658, 1517-001659, 1517-001660, 1517-001661, 1517-001662, 1517-001663, 1517-001664, 1517-001665, 1517-001666, 1517-001667, 1517-001668, 1517-001669, 1517-001670, 1517-001671, 1517-001672, 1517-001673, 1517-001674, 1517-001675, 1517-001676, 1517-001677, 1517-001678, 1517-001679, 1517-001680, 1517-001681, 1517-001682, 1517-001683, 1517-001684, 1517-001685, 1517-001686, 1517-001687, 1517-001688, 1517-001689, 1517-001690, 1517-001691, 1517-001692, 1517-001693, 1517-001694, 1517-001695, 1517-001696, 1517-001697, 1517-001698, 1517-001699, 1517-001700, 1517-001701, 1517-001702, 1517-001703, 1517-001704, 1517-001705, 1517-001706, 1517-001707, 1517-001708, 1517-001709, 1517-001710, 1517-001711, 1517-001712, 1517-001713, 1517-001714, 1517-001715, 1517-001716, 1517-001717, 1517-001718, 1517-001719, 1517-001720, 1517-001721, 1517-001722, 1517-001723, 1517-001724, 1517-001725, 1517-001726, 1517-001727, 1517-001728, 1517-001729, 1517-001730, 1517-001731, 1517-001732, 1517-001733, 1517-001734, 1517-001735, 1517-001736, 1517-001737, 1517-001738, 1517-001739, 1517-001740, 1517-001741, 1517-001742, 1517-001743, 1517-001744, 1517-001745, 1517-001746, 1517-001747, 1517-001748, 1517-001749, 1517-001750, 1517-001751, 1517-001752, 1517-001753, 1517-001754, 1517-001755, 1517-001756, 1517-001757, 1517-001758, 1517-001759, 1517-001760, 1517-001761, 1517-001762, 1517-001763, 1517-001764, 1517-001765, 1517-001766, 1517-001767, 1517-001768, 1517-001769, 1517-001770, 1517-001771, 1517-001772, 1517-001773, 1517-001774, 1517-001775, 1517-001776, 1517-001777, 1517-001778, 1517-001779, 1517-001780, 1517-001781, 1517-001782, 1517-001783, 1517-001784, 1517-001785, 1517-001786, 1517-001787, 1517-001788, 1517-001789, 1517-001790, 1517-001791, 1517-001792, 1517-001793, 1517-001794, 1517-001795, 1517-001796, 1517-001797, 1517-001798, 1517-001799, 1517-001800, 1517-001801, 1517-001802, 1517-001803, 1517-001804, 1517-001805, 1517-001806, 1517-001807, 1517-001808, 1517-001809, 1517-001810, 1517-001811, 1517-001812, 1517-001813, 1517-001814, 1517-001815, 1517-001816, 1517-001817, 1517-001818, 1517-001819, 1517-001820, 1517-001821, 1517-001822, 1517-001823, 1517-001824, 1517-001825, 1517-001826, 1517-001827, 1517-001828, 1517-001829, 1517-001830, 1517-001831, 1517-001832, 1517-001833, 1517-001834, 1517-001835, 1517-001836, 1517-001837, 1517-001838, 1517-001839, 1517-001840, 1517-001841, 1517-001842, 1517-001843, 1517-001844, 1517-001845, 1517-001846, 1517-001847, 1517-001848, 1517-001849, 1517-001850, 1517-001877, 1517-001878, 1517-001879, 1517-001880, 1517-001881, 1517-001885, 1517-001886, 1517-001887, 1517-001888, 1517-001889, 1517-001890, 1517-001891, 1517-001892, 1517-001893, 1517-001894, 1517-001899, 1517-001900, 1517-001901, 1517-001902, 1517-001903, 1517-001904, 1517-001905, 1517-001906, 1517-001907, 1517-001908, 1517-001909, 1517-001910, 1517-001911, 1517-001912, 1517-001913, 1517-001914, 1517-001915, 1517-001916, 1517-001917, 1517-001918, 1517-001919, 1517-001920, 1517-001921, 1517-001922, 1517-001923, 1517-001924, 1517-001925, 1517-001926, 1517-001927, 1517-001928, 1517-001929, 1517-001930, 1517-001931, 1517-001932, 1517-001933, 1517-001934, 1517-001935, 1517-001936, 1517-001937, 1517-001940, 1517-001941, 1517-001942, 1517-001943, 1517-001944, 1517-001945, 1517-001946, 1517-001947, 1517-001948, 1517-001949, 1517-001950, 1517-001951, 1517-001952, 1517-001953, 1517-001954, 1517-001955, 1517-001956, 1517-001957, 1517-001958, 1517-001959, 1517-001960, 1517-001961, 1517-001962, 1517-001963, 1517-001964, 1517-001965, 1517-001966, 1517-001967, 1517-001968, 1517-001969, 1517-001970, 1517-001971, 1517-001972, 1517-001973, 1517-001974, 1517-001975, 1517-001976, 1517-001977, 1517-001978, 1517-001979, 1517-001980, 1517-001981, 1517-001982, 1517-001983, 1517-001984, 1517-001985, 1517-001986, 1517-001987, 1517-001988, 1517-001989, 1517-001990, 1517-001991, 1517-001992, 1517-001993, 1517-001994, 1517-001995, 1517-001996, 1517-001997, 1517-001998, 1517-001999, 1517-002000, 1517-002001, 1517-002002, 1517-002003, 1517-002004, 1517-002005, 1517-002006, 1517-002007, 1517-002008, 1517-002009, 1517-002010, 1517-002011, 1517-002012, 1517-002013, 1517-002014, 1517-002015, 1517-002016, 1517-002017, 1517-002018, 1517-002019, 1517-002020, 1517-002021, 1517-002022, 1517-002023, 1517-002024, 1517-002025, 1517-002026, 1517-002027, 1517-002028, 1517-002029, 1517-002030, 1517-002031, 1517-002032, 1517-002033, 1517-002034, 1517-002035, 1517-002036, 1517-002037, 1517-002038, 1517-002039, 1517-002040, 1517-002041, 1517-002042, 1517-002043, 1517-002044, 1517-002045, 1517-002046, 1517-002047, 1517-002048, 1517-002049, 1517-002050, 1517-002051, 1517-002052, 1517-002053, 1517-002054, 1517-002055, 1517-002056, 1517-002057, 1517-002058, 1517-002059, 1517-002060, 1517-002061, 1517-002062, 1517-002063, 1517-002064, 1517-002065, 1517-002066, 1517-002067, 1517-002068, 1517-002069, 1517-002070, 1517-002071, 1517-002072, 1517-002073, 1517-002074, 1517-002075, 1517-002076, 1517-002077, 1517-002078, 1517-002079, 1517-002080, 1517-002081, 1517-002082, 1517-002083, 1517-002084, 1517-002085, 1517-002086, 1517-002087, 1517-002088, 1517-002089, 1517-002090, 1517-002091, 1517-002092, 1517-002093, 1517-002094, 1517-002095, 1517-002096, 1517-002097, 1517-002098, 1517-002099, 1517-002100, 1517-002101, 1517-002102, 1517-002103, 1517-002104, 1517-002105, 1517-002106, 1517-002107, 1517-002108, 1517-002109, 1517-002110, 1517-002111, 1517-002112, 1517-002113, 1517-002114, 1517-002115, 1517-002116, 1517-002117, 1517-002118, 1517-002119, 1517-002120, 1517-002121, 1517-002122, 1517-002123, 1517-002124, 1517-002125, 1517-002126, 1517-002127, 1517-002128, 1517-002129, 1517-002130, 1517-002131, 1517-002132, 1517-002133, 1517-002134, 1517-002135, 1517-002136, 1517-002137, 1517-002138, 1517-002139, 1517-002140, 1517-002141, 1517-002142, 1517-002143, 1517-002144, 1517-002145, 1517-002146, 1517-002147, 1517-002148, 1517-002149, 1517-002150, 1517-002151, 1517-002152, 1517-002153, 1517-002154, 1517-002155, 1517-002156, 1517-002157, 1517-002158, 1517-002159, 1517-002160, 1517-002161, 1517-002162, 1517-002163, 1517-002164, 1517-002165, 1517-002166, 1517-002167, 1517-002168, 1517-002169, 1517-002170, 1517-002171, 1517-002172, 1517-002173, 1517-002174, 1517-002175, 1517-002176, 1517-002177, 1517-002178, 1517-002179, 1517-002180, 1517-002181, 1517-002182, 1517-002183, 1517-002184, 1517-002185, 1517-002186, 1517-002187, 1517-002188, 1517-002189, 1517-002190, 1517-002191, 1517-002192, 1517-002193, 1517-002194, 1517-002195, 1517-002196, 1517-002197.1517-002198, 1517-002199, 1517-002200, 1517-002201, 1517-002202, 1517-002203, 1517-002204, 1517-002205, 1517-002206, 1517-002207, 1517-002208, 1517-002209, 1517-002210, 1517-002211, 1517-002212, 1517-002213, 1517-002214, 1517-002215, 1517-002216, 1517-002217, 1517-002218, 1517-002219, 1517-002220, 1517-002221, 1517-002222, 1517-002223, 1517-002224, 1517-002225, 1517-002226, 1517-002227, 1517-002228, 1517-002229, 1517-002230, 1517-002231, 1517-002232, 1517-002233, 1517-002234, 1517-002235, 1517-002236, 1517-002237, 1517-002238, 1517-002239, 1517-002240, 1517-002241, 1517-002242, 1517-002243, 1517-002244, 1517-002245, 1517-002246, 1517-002247, 1517-002248, 1517-002249, 1517-002250, 1517-002251, 1517-002252, 1517-002253, 1517-002254, 1517-002255, 1517-002256, 1517-002257, 1517-002258, 1517-002259, 1517-002260, 1517-002261, 1517-002262, 1517-002263, 1517-002264, 1517-002265, 1517-002266, 1517-002267, 1517-002268, 1517-002269, 1517-002270, 1517-002271, 1517-002272, 1517-002273, 1517-002274, 1517-002275, 1517-002276, 1517-002277, 1517-002278, 1517-002279, 1517-002280, 1517-002281, 1517-002282, 1517-002283, 1517-002284, 1517-002285, 1517-002286, 1517-002287, 1517-002288, 1517-002289, 1517-002290, 1517-002291, 1517-002292, 1517-002293, 1517-002294, 1517-002295, 1517-002296, 1517-002297, 1517-002298, 1517-002299, 1517-002300, 1517-002301, 1517-002302, 1517-002303, 1517-002304, 1517-002305, 1517-002306, 1517-002307, 1517-002308, 1517-002309, 1517-002310, 1517-002311, 1517-002312, 1517-002313, 1517-002314, 1517-002315, 1517-002316, 1517-002317, 1517-002318, 1517-002319, 1517-002320, 1517-002321, 1517-002322, 1517-002323, 1517-002324, 1517-002325, 1517-002326, 1517-002327, 1517-002328, 1517-002329, 1517-002330, 1517-002331, 1517-002332, 1517-002333, 1517-002334, 1517-002335, 1517-002336, 1517-002337, 1517-002338, 1517-002339, 1517-002340, 1517-002341, 1517-002342, 1517-002343, 1517-002344, 1517-002345, 1517-002346, 1517-002347, 1517-002348, 1517-002349, 1517-002350, 1517-002351, 1517-002352, 1517-002353, 1517-002354, 1517-002355, 1517-002356, 1517-002357, 1517-002358, 1517-002359, 1517-002360, 1517-002361, 1517-002362, 1517-002363, 1517-002364, 1517-002365, 1517-002366, 1517-002367, 1517-002368, 1517-002369, 1517-002370, 1517-002371, 1517-002372, 1517-002373, 1517-002374, 1517-002375, 1517-002377, 1517-002378, 1517-002379, 1517-002380, 1517-002381, 1517-002382, 1517-002383, 1517-002384, 1517-002385, 1517-002386, 1517-002387, 1517-002388, 1517-002389, 1517-002390, 1517-002391, 1517-002392, 1517-002393, 1517-002394, 1517-002395, 1517-002396, 1517-002397, 1517-002398, 1517-002399, 1517-002400, 1517-002401, 1517-002402, 1517-002403, 1517-002404, 1517-002405, 1517-002406, 1517-002407, 1517-002408, 1517-002409, 1517-002410, 1517-002411, 1517-002412, 1517-002413, 1517-002414, 1517-002415, 1517-002416, 1517-002417, 1517-002418, 1517-002419, 1517-002420, 1517-002421, 1517-002422, 1517-002423, 1517-002424, 1517-002425, 1517-002426, 1517-002427, 1517-002428, 1517-002429, 1517-002430, 1517-002431, 1517-002432, 1517-002433, 1517-002434, 1517-002435, 1517-002436, 1517-002437, 1517-002438, 1517-002439, 1517-002440, 1517-002441, 1517-002442, 1517-002443, 1517-002444, 1517-002445, 1517-002446, 1517-002447, 1517-002448, 1517-002449, 1517-002450, 1517-002451, 1517-002452, 1517-002453, 1517-002454, 1517-002455, 1517-002456, 1517-002457, 1517-002458, 1517-002459, 1517-002460, 1517-002461, 1517-002462, 1517-002463, 1517-002464, 1517-002465, 1517-002466, 1517-002467, 1517-002468, 1517-002469, 1517-002470, 1517-002471, 1517-002472, 1517-002473, 1517-002474, 1517-002475, 1517-002476, 1517-002477, 1517-002478, 1517-002479, 1517-002480, 1517-002481, 1517-002482, 1517-002483, 1517-002484, 1517-002485, 1517-002486, 1517-002487, 1517-002488, 1517-002489, 1517-002490, 1517-002491, 1517-002492, 1517-002493, 1517-002494, 1517-002495, 1517-002496, 1517-002497, 1517-002498, 1517-002499, 1517-002500, 1517-002501, 1517-002502, 1517-002503, 1517-002504, 1517-002505, 1517-002506, 1517-002507, 1517-002508, 1517-002509, 1517-002510, 1517-002511, 1517-002512, 1517-002513, 1517-002514, 1517-002515, 1517-002516, 1517-002517, 1517-002518, 1517-002519, 1517-002520, 1517-002521, 1517-002522, 1517-002523, 1517-002524, 1517-002525, 1517-002526, 1517-002527, 1517-002528, 1517-002529, 1517-002530, 1517-002531, 1517-002532, 1517-002533, 1517-002534, 1517-002535, 1517-002536, 1517-002537, 1517-002538, 1517-002539, 1517-002540, 1517-002541, 1517-002542, 1517-002543, 1517-002544, 1517-002545, 1517-002546, 1517-002547, 1517-002548, 1517-002549, 1517-002550, 1517-002551, 1517-002552, 1517-002553, 1517-002554, 1517-002555, 1517-002556, 1517-002557, 1517-002558, 1517-002559, 1517-002560, 1517-002561, 1517-002562, 1517-002563, 1517-002564, 1517-002565, 1517-002566, 1517-002567, 1517-002568, 1517-002569, 1517-002570, 1517-002571, 1517-002572, 1517-002573, 1517-002574, 1517-002575, 1517-002576, 1517-002577, 1517-002578, 1517-002579, 1517-002580, 1517-002581, 1517-002582, 1517-002583, 1517-002584, 1517-002585, 1517-002586, 1517-002587, 1517-002588, 1517-002589, 1517-002590, 1517-002591, 1517-002592, 1517-002593, 1517-002594, 1517-002595, 1517-002596, 1517-002597, 1517-002598, 1517-002599, 1517-002600, 1517-002601, 1517-002602, 1517-002603, 1517-002604, 1517-002605, 1517-002606, 1517-002607, 1517-002608, 1517-002609, 1517-002610, 1517-002611, 1517-002612, 1517-002613, 1517-002614, 1517-002615, 1517-002616, 1517-002617, 1517-002618, 1517-002619, 1517-002620, 1517-002621, 1517-002622, 1517-002623, 1517-002624, 1517-002625, 1517-002626, 1517-002627, 1517-002628, 1517-002629, 1517-002630, 1517-002631, 1517-002632, 1517-002633, 1517-002634, 1517-002635, 1517-002636, 1517-002637, 1517-002638, 1517-002639, 1517-002640, 1517-002641, 1517-002642, 1517-002643, 1517-002644, 1517-002645, 1517-002646, 1517-002647, 1517-002648, 1517-002649, 1517-002650, 1517-002651, 1517-002652, 1517-002653, 1517-002654, 1517-002655, 1517-002656, 1517-002657, 1517-002658, 1517-002659, 1517-002660, 1517-002661, 1517-002662, 1517-002663, 1517-002664, 1517-002665, 1517-002666, 1517-002667, 1517-002668, 1517-002669, 1517-002670, 1517-002671, 1517-002672, 1517-002673, 1517-002674, 1517-002675, 1517-002676, 1517-002677, 1517-002678, 1517-002679, 1517-002680, 1517-002681, 1517-002682, 1517-002683, 1517-002684, 1517-002685, 1517-002686, 1517-002687, 1517-002688, 1517-002689, 1517-002690, 1517-002691, 1517-002692, 1517-002693, 1517-002694, 1517-002695, 1517-002696, 1517-002697, 1517-002698, 1517-002699, 1517-002700, 1517-002701, 1517-002702, 1517-002703, 1517-002704, 1517-002705, 1517-002706, 1517-002707, 1517-002708, 1517-002709, 1517-002710, 1517-002711, 1517-002712, 1517-002713, 1517-002714, 1517-002715, 1517-002716, 1517-002717, 1517-002718, 1517-002719, 1517-002720, 1517-002721, 1517-002722, 1517-002723, 1517-002724, 1517-002725, 1517-002726, 1517-002727, 1517-002728, 1517-002729, 1517-002730, 1517-002731, 1517-002732, 1517-002733, 1517-002734, 1517-002735, 1517-002736, 1517-002737, 1517-002738, 1517-002739, 1517-002740, 1517-002741, 1517-002742, 1517-002743, 1517-002744, 1517-002745, 1517-002746, 1517-002747, 1517-002748, 1517-002749, 1517-002750, 1517-002751, 1517-002752, 1517-002753, 1517-002754, 1517-002755, 1517-002756, 1517-002757, 1517-002758, 1517-002759, 1517-002760, 1517-002761, 1517-002762, 1517-002763, 1517-002764, 1517-002765, 1517-002766, 1517-002767, 1517-002768, 1517-002769, 1517-002770, 1517-002771, 1517-002772, 1517-002773, 1517-002774, 1517-002775, 1517-002776, 1517-002777, 1517-002778, 1517-002779, 1517-002780, 1517-002781, 1517-002782, 1517-002783, 1517-002784, 1517-002785, 1517-002786, 1517-002787, 1517-002788, 1517-002789, 1517-002790, 1517-002791, 1517-002792, 1517-002793, 1517-002794, 1517-002795, 1517-002796, 1517-002798, 1517-002799, 1517-002800, 1517-002802, 1517-002803, 1517-002804, 1517-002805, 1517-002806, 1517-002807, 1517-002808, 1517-002809, 1517-002810, 1517-002811, 1517-002812, 1517-002813, 1517-002814, 1517-002815, 1517-002816, 1517-002817, 1517-002818, 1517-002819, 1517-002820, 1517-002821, 1517-002822, 1517-002823, 1517-002824, 1517-002825, 1517-002826, 1517-002827, 1517-002828, 1517-002829, 1517-002830, 1517-002831, 1517-002832, 1517-002833, 1517-002834, 1517-002835, 1517-002836, 1517-002837, 1517-002838, 1517-002839, 1517-002840, 1517-002841, 1517-002842, 1517-002843, 1517-002844, 1517-002845, 1517-002846, 1517-002847, 1517-002848, 1517-002849, 1517-002850, 1517-002851, 1517-002852, 1517-002853, 1517-002854, 1517-002855, 1517-002856, 1517-002857, 1517-002858, 1517-002859, 1517-002860, 1517-002861, 1517-002862, 1517-002863, 1517-002864, 1517-002865, 1517-002866, 1517-002867, 1517-002868, 1517-002869, 1517-002870, 1517-002871, 1517-002872, 1517-002873, 1517-002874, 1517-002875, 1517-002876, 1517-002877, 1517-002878, 1517-002879, 1517-002880, 1517-002881, 1517-002882, 1517-002883, 1517-002884, 1517-002885, 1517-002886, 1517-002887, 1517-002888, 1517-002889, 1517-002890, 1517-002891, 1517-002892, 1517-002893, 1517-002894, 1517-002895, 1517-002896, 1517-002897, 1517-002898, 1517-002899, 1517-002900, 1517-002901, 1517-002902, 1517-002903, 1517-002904, 1517-002905, 1517-002906, 1517-002907, 1517-002908, 1517-002909, 1517-002910, 1517-002911, 1517-002912, 1517-002913, 1517-002914, 1517-002915, 1517-002916, 1517-002917, 1517-002918, 1517-002919, 1517-002920, 1517-002921, 1517-002922, 1517-002923, 1517-002924, 1517-002925, 1517-002926, 1517-002927, 1517-002928, 1517-002929, 1517-002930, 1517-002931, 1517-002932, 1517-002933, 1517-002934, 1517-002935, 1517-002936, 1517-002937, 1517-002938, 1517-002939, 1517-002940, 1517-002941, 1517-002942, 1517-002943, 1517-002944, 1517-002945, 1517-002946, 1517-002947, 1517-002948, 1517-002949, 1517-002950, 1517-002951, 1517-002952, 1517-002953, 1517-002954, 1517-002955, 1517-002956, 1517-002957, 1517-002958, 1517-002959, 1517-002960, 1517-002961, 1517-002962, 1517-002963, 1517-002964, 1517-002965, 1517-002966, 1517-002967, 1517-002968, 1517-002969, 1517-002970, 1517-002971, 1517-002972, 1517-002973, 1517-002974, 1517-002975, 1517-002976, 1517-002977, 1517-002978, 1517-002979, 1517-002980, 1517-002981, 1517-002982, 1517-002983, 1517-002984, 1517-002985, 1517-002986, 1517-002987, 1517-002988, 1517-002989, 1517-002990, 1517-002991, 1517-002992, 1517-002993, 1517-002994, 1517-002995, 1517-002996, 1517-002997, 1517-002998, 1517-002999, 1517-003000, 1517-003001, 1517-003002, 1517-003003, 1517-003004, 1517-003005, 1517-003006, 1517-003007, 1517-003008, 1517-003009, 1517-003010, 1517-003011, 1517-003012, 1517-003013, 1517-003014, 1517-003015, 1517-003016, 1517-003017, 1517-003018, 1517-003019, 1517-003020, 1517-003021, 1517-003022, 1517-003023, 1517-003024, 1517-003025, 1517-003026, 1517-003027, 1517-003028, 1517-003029, 1517-003030, 1517-003031, 1517-003032, 1517-003033, 1517-003034, 1517-003035, 1517-003036, 1517-003037, 1517-003038, 1517-003039, 1517-003040, 1517-003041, 1517-003042, 1517-003043, 1517-003044, 1517-003045, 1517-003046, 1517-003047, 1517-003048, 1517-003049, 1517-003050, 1517-003051, 1517-003052, 1517-003053, 1517-003054, 1517-003055, 1517-003056, 1517-003057, 1517-003058, 1517-003059, 1517-003060, 1517-003061, 1517-003062, 1517-003063, 1517-003064, 1517-003065, 1517-003066, 1517-003067, 1517-003068, 1517-003069, 1517-003070, 1517-003071, 1517-003072, 1517-003073, 1517-003074, 1517-003075, 1517-003076, 1517-003077, 1517-003078, 1517-003079, 1517-003080, 1517-003081, 1517-003082, 1517-003083, 1517-003084, 1517-003085, 1517-003086, 1517-003087, 1517-003088, 1517-003089, 1517-003090, 1517-003091, 1517-003092, 1517-003093, 1517-003094, 1517-003095, 1517-003096, 1517-003097, 1517-003098, 1517-003099, 1517-003100, 1517-003101, 1517-003102, 1517-003103, 1517-003104, 1517-003105, 1517-003106, 1517-003107, 1517-003108, 1517-003109, 1517-003110, 1517-003111, 1517-003112, 1517-003113, 1517-003114, 1517-003115, 1517-003116, 1517-003117, 1517-003118, 1517-003119, 1517-003120, 1517-003121, 1517-003122, 1517-003123, 1517-003124, 1517-003125, 1517-003126, 1517-003127, 1517-003128, 1517-003129, 1517-003130, 1517-003131, 1517-003132, 1517-003133, 1517-003134, 1517-003135, 1517-003136, 1517-003137, 1517-003138, 1517-003139, 1517-003140, 1517-003141, 1517-003142, 1517-003143, 1517-003144, 1517-003145, 1517-003146, 1517-003147, 1517-003148, 1517-003149, 1517-003150, 1517-003151, 1517-003152, 1517-003153, 1517-003154, 1517-003155, 1517-003156, 1517-003157, 1517-003158, 1517-003159, 1517-003160, 1517-003161, 1517-003162, 1517-003163, 1517-003164, 1517-003165, 1517-003166, 1517-003167, 1517-003168, 1517-003169, 1517-003170, 1517-003171, 1517-003172, 1517-003173, 1517-003174, 1517-003175, 1517-003176, 1517-003177, 1517-003178, 1517-003179, 1517-003180, 1517-003181, 1517-003182, 1517-003183, 1517-003184, 1517-003185, 1517-003186, 1517-003187, 1517-003188, 1517-003189, 1517-003190, 1517-003191, 1517-003192, 1517-003193, 1517-003194, 1517-003195, 1517-003196, 1517-003197, 1517-003198, 1517-003199, 1517-003200, 1517-003201, 1517-003202, 1517-003203, 1517-003204, 1517-003205, 1517-003206, 1517-003207, 1517-003208, 1517-003209, 1517-003210, 1517-003211, 1517-003212, 1517-003213, 1517-003214, 1517-003215, 1517-003216, 1517-003217, 1517-003218, 1517-003219, 1517-003220, 1517-003221, 1517-003222, 1517-003223, 1517-003224, 1517-003225, 1517-003226, 1517-003227, 1517-003228, 1517-003229, 1517-003230, 1517-003231, 1517-003232, 1517-003233, 1517-003234, 1517-003235, 1517-003236, 1517-003237, 1517-003238, 1517-003239, 1517-003240, 1517-003241, 1517-003242, 1517-003243, 1517-003244, 1517-003245, 1517-003246, 1517-003247, 1517-003248, 1517-003249, 1517-003250, 1517-003251, 1517-003252, 1517-003253, 1517-003254, 1517-003255, 1517-003256, 1517-003257, 1517-003258, 1517-003259, 1517-003260, 1517-003261, 1517-003262, 1517-003263, 1517-003264, 1517-003265, 1517-003266, 1517-003267, 1517-003268, 1517-003269, 1517-003270, 1517-003271, 1517-003272, 1517-003273, 1517-003274, 1517-003275, 1517-003276, 1517-003277, 1517-003278, 1517-003279, 1517-003280, 1517-003281, 1517-003282, 1517-003283, 1517-003284, 1517-003285, 1517-003286, 1517-003287, 1517-003288, 1517-003289, 1517-003290, 1517-003291, 1517-003292, 1517-003293, 1517-003294, 1517-003295, 1517-003296, 1517-003297, 1517-003298, 1517-003299, 1517-003300, 1517-003301, 1517-003302, 1517-003303, 1517-003304, 1517-003305, 1517-003306, 1517-003307, 1517-003308, 1517-003309, 1517-003310, 1517-003311, 1517-003312, 1517-003313, 1517-003314, 1517-003315, 1517-003316, 1517-003317, 1517-003318, 1517-003319, 1517-003320, 1517-003321, 1517-003322, 1517-003323, 1517-003324, 1517-003325, 1517-003326, 1517-003327, 1517-003328, 1517-003329, 1517-003330, 1517-003331, 1517-003332, 1517-003333, 1517-003334, 1517-003335, 1517-003336, 1517-003337, 1517-003338, 1517-003339, 1517-003340, 1517-003341, 1517-003342, 1517-003343, 1517-003344, 1517-003345, 1517-003346, 1517-003347, 1517-003348, 1517-003349, 1517-003350, 1517-003351, 1517-003352, 1517-003353, 1517-003354, 1517-003355, 1517-003356, 1517-003357, 1517-003358, 1517-003359, 1517-003360, 1517-003361, 1517-003362, 1517-003363, 1517-003364, 1517-003365, 1517-003366, 1517-003367, 1517-003368, 1517-003369, 1517-003370, 1517-003371, 1517-003372, 1517-003373, 1517-003374, 1517-003375, 1517-003376, 1517-003377, 1517-003378, 1517-003379, 1517-003380, 1517-003381, 1517-003382, 1517-003383, 1517-003384, 1517-003385, 1517-003386, 1517-003387, 1517-003388, 1517-003389, 1517-003390, 1517-003391, 1517-003392, 1517-003393, 1517-003394, 1517-003395, 1517-003396, 1517-003397, 1517-003398, 1517-003399, 1517-003400, 1517-003401, 1517-003402, 1517-003403, 1517-003404, 1517-003405, 1517-003406, 1517-003407, 1517-003408, 1517-003409, 1517-003410, 1517-003411, 1517-003412, 1517-003413, 1517-003414, 1517-003415, 1517-003416, 1517-003417, 1517-003418, 1517-003419, 1517-003420, 1517-003421, 1517-003422, 1517-003423, 1517-003424, 1517-003425, 1517-003426, 1517-003427, 1517-003428, 1517-003429, 1517-003430, 1517-003431, 1517-003432, 1517-003433, 1517-003434, 1517-003435, 1517-003436, 1517-003437, 1517-003438, 1517-003439, 1517-003440, 1517-003441, 1517-003442, 1517-003443, 1517-003444, 1517-003445, 1517-003446, 1517-003447, 1517-003448, 1517-003449, 1517-003450, 1517-003451, 1517-003452, 1517-003453, 1517-003454, 1517-003455, 1517-003456, 1517-003457, 1517-003458, 1517-003459, 1517-003460, 1517-003461, 1517-003462, 1517-003463, 1517-003464, 1517-003465, 1517-003466, 1517-003467, 1517-003468, 1517-003469, 1517-003470, 1517-003471, 1517-003472, 1517-003473, 1517-003474, 1517-003475, 1517-003476, 1517-003477, 1517-003478, 1517-003479, 1517-003480, 1517-003481, 1517-003482, 1517-003483, 1517-003484, 1517-003485, 1517-003486, 1517-003487, 1517-003488, 1517-003489, 1517-003490, 1517-003491, 1517-003492, 1517-003493, 1517-003494, 1517-003495, 1517-003496, 1517-003497, 1517-003498, 1517-003499, 1517-003500, 1517-003501, 1517-003502, 1517-003503, 1517-003504, 1517-003505, 1517-003506, 1517-003507, 1517-003508, 1517-003509, 1517-003510, 1517-003511, 1517-003512, 1517-003513, 1517-003514, 1517-003515, 1517-003516, 1517-003517, 1517-003518, 1517-003519, 1517-003520, 1517-003521, 1517-003522, 1517-003523, 1517-003524, 1517-003525, 1517-003526, 1517-003527, 1517-003528, 1517-003529, 1517-003530, 1517-003531, 1517-003532, 1517-003533, 1517-003534, 1517-003535, 1517-003536, 1517-003537, 1517-003538, 1517-003539, 1517-003540, 1517-003541, 1517-003542, 1517-003543, 1517-003544, 1517-003545, 1517-003546, 1517-003547, 1517-003548, 1517-003549, 1517-003550, 1517-003551, 1517-003552, 1517-003553, 1517-003554, 1517-003555, 1517-003556, 1517-003557, 1517-003558, 1517-003559, 1517-003560, 1517-003561, 1517-003562, 1517-003563, 1517-003564, 1517-003565, 1517-003566, 1517-003567, 1517-003568, 1517-003569, 1517-003570, 1517-003571, 1517-003572, 1517-003573, 1517-003574, 1517-003575, 1517-003576, 1517-003577, 1517-003578, 1517-003579, 1517-003580, 1517-003581, 1517-003582, 1517-003583, 1517-003584, 1517-003585, 1517-003586, 1517-003587, 1517-003588, 1517-003589, 1517-003590, 1517-003591, 1517-003592, 1517-003593, 1517-003594, 1517-003595, 1517-003596, 1517-003597, 1517-003598, 1517-003599, 1517-003600, 1517-003601, 1517-003602, 1517-003603, 1517-003604, 1517-003605, 1517-003606, 1517-003607, 1517-003608, 1517-003609, 1517-003610, 1517-003611, 1517-003612, 1517-003613, 1517-003614, 1517-003615, 1517-003616, 1517-003617, 1517-003618, 1517-003619, 1517-003620, 1517-003621, 1517-003622, 1517-003623, 1517-003624, 1517-003625, 1517-003626, 1517-003627, 1517-003628, 1517-003629, 1517-003630, 1517-003631, 1517-003632, 1517-003633, 1517-003634, 1517-003635, 1517-003636, 1517-003637, 1517-003638, 1517-003639, 1517-003640, 1517-003641, 1517-003642, 1517-003643, 1517-003644, 1517-003645, 1517-003646, 1517-003647, 1517-003648, 1517-003649, 1517-003650, 1517-003651, 1517-003652, 1517-003653, 1517-003654, 1517-003655, 1517-003656, 1517-003657, 1517-003658, 1517-003659, 1517-003660, 1517-003661, 1517-003662, 1517-003663, 1517-003664, 1517-003665, 1517-003666, 1517-003667, 1517-003668, 1517-003669, 1517-003670, 1517-003671, 1517-003672, 1517-003673, 1517-003674, 1517-003675, 1517-003676, 1517-003677, 1517-003678, 1517-003679, 1517-003680, 1517-003681, 1517-003682, 1517-003683, 1517-003684, 1517-003685, 1517-003686, 1517-003687, 1517-003688, 1517-003689, 1517-003690, 1517-003691, 1517-003692, 1517-003693, 1517-003694, 1517-003695, 1517-003696, 1517-003697, 1517-003698, 1517-003699, 1517-003700, 1517-003701, 1517-003702, 1517-003703, 1517-003704, 1517-003705, 1517-003706, 1517-003707, 1517-003708, 1517-003709, 1517-003710, 1517-003711, 1517-003712, 1517-003713, 1517-003714, 1517-003715, 1517-003716, 1517-003717, 1517-003718, 1517-003719, 1517-003720, 1517-003721, 1517-003722, 1517-003723, 1517-003724, 1517-003725, 1517-003726, 1517-003727, 1517-003728, 1517-003729, 1517-003730, 1517-003731, 1517-003732, 1517-003733, 1517-003734, 1517-003735, 1517-003736, 1517-003737, 1517-003738, 1517-003739, 1517-003740, 1517-003741, 1517-003742, 1517-003743, 1517-003744, 1517-003745, 1517-003746, 1517-003747, 1517-003748, 1517-003749, 1517-003750, 1517-003751, 1517-003752, 1517-003753, 1517-003754, 1517-003755, 1517-003756, 1517-003757, 1517-003758, 1517-003759, 1517-003760, 1517-003761, 1517-003762, 1517-003763, 1517-003764, 1517-003765, 1517-003766, 1517-003767, 1517-003768, 1517-003769, 1517-003770, 1517-003771, 1517-003772, 1517-003773, 1517-003774, 1517-003775, 1517-003776, 1517-003777, 1517-003778, 1517-003779, 1517-003780, 1517-003781, 1517-003782, 1517-003783, 1517-003784, 1517-003785, 1517-003786, 1517-003787, 1517-003788, 1517-003789, 1517-003790, 1517-003791, 1517-003792, 1517-003793, 1517-003794, 1517-003795, 1517-003796, 1517-003797, 1517-003798, 1517-003799, 1517-003800, 1517-003801, 1517-003802, 1517-003803, 1517-003804, 1517-003805, 1517-003806, 1517-003807, 1517-003808, 1517-003809, 1517-003810, 1517-003811, 1517-003812, 1517-003813, 1517-003814, 1517-003815, 1517-003816, 1517-003817, 1517-003818, 1517-003819, 1517-003820, 1517-003821, 1517-003822, 1517-003823, 1517-003824, 1517-003825, 1517-003826, 1517-003827, 1517-003828, 1517-003829, 1517-003830, 1517-003831, 1517-003832, 1517-003833, 1517-003834, 1517-003835, 1517-003836, 1517-003837, 1517-003838, 1517-003839, 1517-003840, 1517-003841, 1517-003842, 1517-003843, 1517-003844, 1517-003845, 1517-003846, 1517-003847, 1517-003848, 1517-003849, 1517-003850, 1517-003851, 1517-003852, 1517-003853, 1517-003854, 1517-003855, 1517-003856, 1517-003857, 1517-003858, 1517-003859, 1517-003860, 1517-003861, 1517-003862, 1517-003863, 1517-003864, 1517-003865, 1517-003866, 1517-003867, 1517-003868, 1517-003869, 1517-003927, 1517-003928, 1517-003929, 1517-003930, 1517-003931, 1517-003932, 1517-003933, 1517-003934, 1517-003935, 1517-003936, 1517-003937, 1517-003938, 1517-003939, 1517-003940, 1517-003941, 1517-003942, 1517-003943, 1517-003944, 1517-003945, 1517-003946, 1517-003947, 1517-003948, 1517-003949, 1517-003950, 1517-003951, 1517-003952, 1517-003953, 1517-003954, 1517-003955, 1517-003956, 1517-003957, 1517-003958, 1517-003959, 1517-003960, 1517-003961, 1517-003962, 1517-003963, 1517-003964, 1517-003965, 1517-003966, 1517-003967, 1517-003968, 1517-003969, 1517-003970, 1517-003971, 1517-003972, 1517-003973, 1517-003974, 1517-003975, 1517-003976, 1517-003977, 1517-003978, 1517-003979, 1517-003980, 1517-003981, 1517-003982, 1517-003983, 1517-003984, 1517-003985, 1517-003986, 1517-003987, 1517-003988, 1517-003989, 1517-003990, 1517-003991, 1517-003992, 1517-003993, 1517-003994, 1517-003995, 1517-003996, 1517-003997, 1517-003998, 1517-003999, 1517-004000, 1517-004001, 1517-004002, 1517-004003, 1517-004004, 1517-004005, 1517-004006, 1517-004007, 1517-004008, 1517-004009, 1517-004010, 1517-004011, 1517-004012, 1517-004013, 1517-004014, 1517-004015, 1517-004016, 1517-004017, 1517-004018, 1517-004019, 1517-004020, 1517-004021, 1517-004022, 1517-004023, 1517-004024, 1517-004025, 1517-004026, 1517-004027, 1517-004028, 1517-004029, 1517-004030, 1517-004031, 1517-004032, 1517-004033, 1517-004034, 1517-004035, 1517-004036, 1517-004037, 1517-004038, 1517-004039, 1517-004040, 1517-004041, 1517-004042, 1517-004043, 1517-004044, 1517-004045, 1517-004046, 1517-004047, 1517-004048, 1517-004049, 1517-004050, 1517-004051, 1517-004052, 1517-004053, 1517-004054, 1517-004055, 1517-004056, 1517-004057, 1517-004058, 1517-004059, 1517-004060, 1517-004061, 1517-004062, 1517-004063, 1517-004064, 1517-004065, 1517-004066, 1517-004067, 1517-004068, 1517-004069, 1517-004070, 1517-004071, 1517-004072, 1517-004073, 1517-004074, 1517-004075, 1517-004076, 1517-004077, 1517-004078, 1517-004079, 1517-004080, 1517-004081, 1517-004082, 1517-004083, 1517-004084, 1517-004085, 1517-004086, 1517-004087, 1517-004088, 1517-004089, 1517-004090, 1517-004091, 1517-004092, 1517-004093, 1517-004094, 1517-004095, 1517-004096, 1517-004097, 1517-004098, 1517-004099, 1517-004100, 1517-004101, 1517-004102, 1517-004103, 1517-004104, 1517-004105, 1517-004106, 1517-004107, 1517-004108, 1517-004109, 1517-004110, 1517-004111, 1517-004112, 1517-004113, 1517-004114, 1517-004115, 1517-004116, 1517-004117, 1517-004118, 1517-004119, 1517-004120, 1517-004121, 1517-004122, 1517-004123, 1517-004124, 1517-004128, 1517-004129, 1517-004130, 1517-004131, 1517-004132, 1517-004133, 1517-004134, 1517-004135, 1517-004136, 1517-004137, 1517-004138, 1517-004139, 1517-004140, 1517-004141, 1517-004142, 1517-004143, 1517-004144, 1517-004145, 1517-004146, 1517-004147, 1517-004148, 1517-004149, 1517-004150, 1517-004151, 1517-004152, 1517-004153, 1517-004154, 1517-004155, 1517-004156, 1517-004157, 1517-004158, 1517-004159, 1517-004160, 1517-004161, 1517-004162, 1517-004163, 1517-004164, 1517-004165, 1517-004166, 1517-004167, 1517-004168, 1517-004169, 1517-004170, 1517-004171, 1517-004172, 1517-004173, 1517-004174, 1517-004175, 1517-004176, 1517-004177, 1517-004178, 1517-004179, 1517-004180, 1517-004181, 1517-004182, 1517-004183, 1517-004184, 1517-004185, 1517-004186, 1517-004187, 1517-004188, 1517-004189, 1517-004190, 1517-004191, 1517-004192, 1517-004193, 1517-004194, 1517-004195, 1517-004196, 1517-004197, 1517-004198, 1517-004199, 1517-004200, 1517-004201, 1517-004202, 1517-004203, 1517-004204, 1517-004205, 1517-004206, 1517-004207, 1517-004208, 1517-004209, 1517-004210, 1517-004211, 1517-004212, 1517-004213, 1517-004214, 1517-004215, 1517-004216, 1517-004217, 1517-004218, 1517-004219, 1517-004220, 1517-004221, 1517-004222, 1517-004223, 1517-004224, 1517-004225, 1517-004226, 1517-004227, 1517-004228, 1517-004229, 1517-004230, 1517-004231, 1517-004232, 1517-004233, 1517-004234, 1517-004235, 1517-004236, 1517-004237, 1517-004238, 1517-004239, 1517-004240, 1517-004241, 1517-004242, 1517-004243, 1517-004244, 1517-004245, 1517-004246, 1517-004247, 1517-004248, 1517-004249, 1517-004250, 1517-004251, 1517-004252, 1517-004253, 1517-004254, 1517-004255, 1517-004256, 1517-004257, 1517-004258, 1517-004259, 1517-004260, 1517-004261, 1517-004262, 1517-004263, 1517-004264, 1517-004265, 1517-004266, 1517-004267, 1517-004268, 1517-004269, 1517-004270, 1517-004271, 1517-004272, 1517-004273, 1517-004274, 1517-004275, 1517-004276, 1517-004277, 1517-004278, 1517-004279, 1517-004280, 1517-004281, 1517-004282, 1517-004283, 1517-004284, 1517-004285, 1517-004286, 1517-004287, 1517-004288, 1517-004289, 1517-004290, 1517-004291, 1517-004292, 1517-004293, 1517-004294, 1517-004295, 1517-004296, 1517-004297, 1517-004298, 1517-004299, 1517-004300, 1517-004301, 1517-004302, 1517-004303, 1517-004304, 1517-004305, 1517-004306, 1517-004307, 1517-004308, 1517-004309, 1517-004310, 1517-004311, 1517-004312, 1517-004313, 1517-004314, 1517-004315, 1517-004316, 1517-004317, 1517-004318, 1517-004319, 1517-004320, 1517-004321, 1517-004322, 1517-004323, 1517-004324, 1517-004325, 1517-004326, 1517-004327, 1517-004328, 1517-004329, 1517-004330, 1517-004331, 1517-004332, 1517-004333, 1517-004334, 1517-004335, 1517-004336, 1517-004337, 1517-004338, 1517-004339, 1517-004340, 1517-004341, 1517-004342, 1517-004343, 1517-004344, 1517-004345, 1517-004346, 1517-004347, 1517-004348, 1517-004349, 1517-004350, 1517-004351, 1517-004352, 1517-004353, 1517-004354, 1517-004355, 1517-004356, 1517-004357, 1517-004358, 1517-004359, 1517-004360, 1517-004361, 1517-004362, 1517-004363, 1517-004364, 1517-004365, 1517-004366, 1517-004367, 1517-004368, 1517-004369, 1517-004370, 1517-004371, 1517-004372, 1517-004373, 1517-004374, 1517-004375, 1517-004376, 1517-004377, 1517-004378, 1517-004379, 1517-004380, 1517-004381, 1517-004382, 1517-004383, 1517-004384, 1517-004385, 1517-004386, 1517-004387, 1517-004388, 1517-004389, 1517-004390, 1517-004391, 1517-004392, 1517-004393, 1517-004394, 1517-004395, 1517-004396, 1517-004397, 1517-004398, 1517-004399, 1517-004400, 1517-004401, 1517-004402, 1517-004403, 1517-004404, 1517-004405, 1517-004406, 1517-004407, 1517-004408, 1517-004409, 1517-004410, 1517-004411, 1517-004412, 1517-004413, 1517-004414, 1517-004415, 1517-004416, 1517-004417, 1517-004418, 1517-004419, 1517-004420, 1517-004421, 1517-004422, 1517-004423, 1517-004425, 1517-004426, 1517-004427, 1517-004428, 1517-004429, 1517-004430, 1517-004431, 1517-004432, 1517-004433, 1517-004434, 1517-004435, 1517-004436, 1517-004437, 1517-004438, 1517-004439, 1517-004440, 1517-004441, 1517-004442, 1517-004443, 1517-004444, 1517-004445, 1517-004446, 1517-004447, 1517-004448, 1517-004449, 1517-004450, 1517-004451, 1517-004452, 1517-004453, 1517-004454, 1517-004455, 1517-004456, 1517-004457, 1517-004458, 1517-004459, 1517-004460, 1517-004461, 1517-004462, 1517-004463, 1517-004464, 1517-004465, 1517-004466, 1517-004467, 1517-004468, 1517-004469, 1517-004470, 1517-004471, 1517-004472, 1517-004473, 1517-004474, 1517-004475, 1517-004476, 1517-004477, 1517-004478, 1517-004479, 1517-004480, 1517-004481, 1517-004482, 1517-004483, 1517-004484, 1517-004485, 1517-004486, 1517-004487, 1517-004488, 1517-004489, 1517-004490, 1517-004491, 1517-004492, 1517-004493, 1517-004494, 1517-004495, 1517-004496, 1517-004497, 1517-004498, 1517-004499, 1517-004500, 1517-004501, 1517-004502, 1517-004503, 1517-004504, 1517-004505, 1517-004506, 1517-004507, 1517-004508, 1517-004509, 1517-004510, 1517-004511, 1517-004512, 1517-004513, 1517-004514, 1517-004515, 1517-004516, 1517-004517, 1517-004518, 1517-004519, 1517-004520, 1517-004521, 1517-004522, 1517-004523, 1517-004524, 1517-004525, 1517-004526, 1517-004527, 1517-004528, 1517-004529, 1517-004530, 1517-004531, 1517-004532, 1517-004533, 1517-004534, 1517-004538, 1517-004539, 1517-004540, 1517-004541, 1517-004542, 1517-004543, 1517-004544, 1517-004545, 1517-004546, 1517-004547, 1517-004548, 1517-004549, 1517-004550, 1517-004551, 1517-004552, 1517-004553, 1517-004554, 1517-004555, 1517-004556, 1517-004557, 1517-004558, 1517-004559, 1517-004560, 1517-004561, 1517-004562, 1517-004563, 1517-004564, 1517-004565, 1517-004566, 1517-004567, 1517-004568, 1517-004569, 1517-004570, 1517-004571, 1517-004572, 1517-004573, 1517-004574, 1517-004575, 1517-004576, 1517-004577, 1517-004578, 1517-004579, 1517-004580, 1517-004581, 1517-004582, 1517-004583, 1517-004584, 1517-004585, 1517-004586, 1517-004587, 1517-004588, 1517-004589, 1517-004590, 1517-004591, 1517-004592, 1517-004593, 1517-004594, 1517-004595, 1517-004596, 1517-004597, 1517-004598, 1517-004599, 1517-004600, 1517-004601, 1517-004602, 1517-004603, 1517-004604, 1517-004605, 1517-004606, 1517-004607, 1517-004608, 1517-004609, 1517-004610, 1517-004611, 1517-004612, 1517-004613, 1517-004614, 1517-004615, 1517-004616, 1517-004617, 1517-004618, 1517-004619, 1517-004620, 1517-004621, 1517-004622, 1517-004623, 1517-004624, 1517-004625, 1517-004626, 1517-004627, 1517-004628, 1517-004629, 1517-004630, 1517-004631, 1517-004632, 1517-004633, 1517-004634, 1517-004635, 1517-004636, 1517-004637, 1517-004638, 1517-004639, 1517-004640, 1517-004641, 1517-004642, 1517-004643, 1517-004644, 1517-004645, 1517-004646, 1517-004647, 1517-004648, 1517-004649, 1517-004650, 1517-004651, 1517-004652, 1517-004653, 1517-004654, 1517-004655, 1517-004656, 1517-004657, 1517-004658, 1517-004659, 1517-004660, 1517-004661, 1517-004662, 1517-004663, 1517-004664, 1517-004665, 1517-004666, 1517-004667, 1517-004668, 1517-004669, 1517-004670, 1517-004671, 1517-004672, 1517-004673, 1517-004674, 1517-004675, 1517-004676, 1517-004677, 1517-004678, 1517-004679, 1517-004680, 1517-004681, 1517-004682, 1517-004683, 1517-004684, 1517-004685, 1517-004686, 1517-004687, 1517-004688, 1517-004689, 1517-004690, 1517-004691, 1517-004692, 1517-004693, 1517-004694, 1517-004695, 1517-004696, 1517-004697, 1517-004698, 1517-004699, 1517-004700, 1517-004701, 1517-004702, 1517-004703, 1517-004704, 1517-004705, 1517-004706, 1517-004707, 1517-004708, 1517-004709, 1517-004710, 1517-004711, 1517-004712, 1517-004713, 1517-004714, 1517-004715, 1517-004716, 1517-004717, 1517-004718, 1517-004719, 1517-004720, 1517-004721, 1517-004722, 1517-004723, 1517-004724, 1517-004725, 1517-004726, 1517-004727, 1517-004728, 1517-004729, 1517-004730, 1517-004731, 1517-004732, 1517-004733, 1517-004734, 1517-004735, 1517-004736, 1517-004737, 1517-004738, 1517-004739, 1517-004740, 1517-004741, 1517-004742, 1517-004743, 1517-004744, 1517-004745, 1517-004746, 1517-004747, 1517-004748, 1517-004749, 1517-004750, 1517-004751, 1517-004752, 1517-004753, 1517-004754, 1517-004755, 1517-004756, 1517-004757, 1517-004758, 1517-004759, 1517-004760, 1517-004761, 1517-004762, 1517-004763, 1517-004766, 1517-004767, 1517-004771, 1517-004773, 1517-004774, 1517-004775, 1517-004776, 1517-004777, 1517-004778, 1517-004779, 1517-004780, 1517-004781, 1517-004782, 1517-004783, 1517-004784, 1517-004785, 1517-004786, 1517-004787, 1517-004788, 1517-004789, 1517-004790, 1517-004791, 1517-004792, 1517-004793, 1517-004794, 1517-004795, 1517-004796, 1517-004797, 1517-004798, 1517-004799, 1517-004800, 1517-004801, 1517-004802, 1517-004803, 1517-004804, 1517-004805, 1517-004806, 1517-004807, 1517-004808, 1517-004809, 1517-004810, 1517-004811, 1517-004812, 1517-004813, 1517-004814, 1517-004815, 1517-004816, 1517-004817, 1517-004818, 1517-004819, 1517-004820, 1517-004821, 1517-004822, 1517-004823, 1517-004824, 1517-004825, 1517-004826, 1517-004827, 1517-004828, 1517-004829, 1517-004830, 1517-004831, 1517-004832, 1517-004833, 1517-004834, 1517-004835, 1517-004836, 1517-004837, 1517-004838, 1517-004839, 1517-004840, 1517-004841, 1517-004842, 1517-004843, 1517-004844, 1517-004845, 1517-004846, 1517-004847, 1517-004848, 1517-004849, 1517-004850, 1517-004851, 1517-004852, 1517-004853, 1517-004854, 1517-004855, 1517-004856, 1517-004857, 1517-004858, 1517-004859, 1517-004860, 1517-004861, 1517-004862, 1517-004863, 1517-004864, 1517-004865, 1517-004866, 1517-004871, 1517-004872, 1517-004873, 1517-004874, 1517-004875, 1517-004876, 1517-004877, 1517-004878, 1517-004879, 1517-004880, 1517-004881, 1517-004882, 1517-004883, 1517-004884, 1517-004885, 1517-004886, 1517-004887, 1517-004888, 1517-004889, 1517-004890, 1517-004891, 1517-004892, 1517-004893, 1517-004894, 1517-004895, 1517-004896, 1517-004897, 1517-004898, 1517-004899, 1517-004900, 1517-004901, 1517-004902, 1517-004903, 1517-004904, 1517-004905, 1517-004906, 1517-004907, 1517-004908, 1517-004909, 1517-004910, 1517-004911, 1517-004912, 1517-004913, 1517-004914, 1517-004915, 1517-004916, 1517-004917, 1517-004918, 1517-004919, 1517-004920, 1517-004921, 1517-004922, 1517-004923, 1517-004924, 1517-004925, 1517-004926, 1517-004927, 1517-004928, 1517-004929, 1517-004930, 1517-004931, 1517-004932, 1517-004933, 1517-004934, 1517-004935, 1517-004936, 1517-004937, 1517-004938, 1517-004939, 1517-004940, 1517-004941, 1517-004942, 1517-004943, 1517-004944, 1517-004945, 1517-004946, 1517-004947, 1517-004948, 1517-004949, 1517-004950, 1517-004951, 1517-004952, 1517-004953, 1517-004954, 1517-004955, 1517-004956, 1517-004957, 1517-004958, 1517-004959, 1517-004960, 1517-004961, 1517-004962, 1517-004963, 1517-004964, 1517-004965, 1517-004966, 1517-004967, 1517-004968, 1517-004969, 1517-004970, 1517-004971, 1517-004972, 1517-004973, 1517-004974, 1517-004975, 1517-004976, 1517-004977, 1517-004978, 1517-004979, 1517-004980, 1517-004981, 1517-004982, 1517-004983, 1517-004984, 1517-004985, 1517-004986, 1517-004987, 1517-004988, 1517-004989, 1517-004990, 1517-004991, 1517-004992, 1517-004993, 1517-004994, 1517-004995, 1517-004996, 1517-004997, 1517-004998, 1517-004999, 1517-005000, 1517-005001, 1517-005002, 1517-005003, 1517-005004, 1517-005005, 1517-005006, 1517-005007, 1517-005008, 1517-005009, 1517-005010, 1517-005011, 1517-005012, 1517-005013, 1517-005014, 1517-005015, 1517-005016, 1517-005017, 1517-005018, 1517-005019, 1517-005020, 1517-005021, 1517-005022, 1517-005023, 1517-005024, 1517-005025, 1517-005026, 1517-005027, 1517-005028, 1517-005029, 1517-005030, 1517-005031, 1517-005032, 1517-005033, 1517-005034, 1517-005035, 1517-005036, 1517-005037, 1517-005038, 1517-005039, 1517-005040, 1517-005041, 1517-005042, 1517-005043, 1517-005044, 1517-005045, 1517-005046, 1517-005047, 1517-005048, 1517-005049, 1517-005050, 1517-005051, 1517-005052, 1517-005053, 1517-005054, 1517-005055, 1517-005056, 1517-005057, 1517-005058, 1517-005059, 1517-005060, 1517-005061, 1517-005062, 1517-005063, 1517-005064, 1517-005065, 1517-005066, 1517-005067, 1517-005068, 1517-005069, 1517-005070, 1517-005071, 1517-005072, 1517-005073, 1517-005074, 1517-005075, 1517-005076, 1517-005077, 1517-005078, 1517-005079, 1517-005080, 1517-005081, 1517-005082, 1517-005083, 1517-005084, 1517-005085, 1517-005086, 1517-005087, 1517-005088, 1517-005089, 1517-005090, 1517-005091, 1517-005092, 1517-005093, 1517-005094, 1517-005095, 1517-005096, 1517-005097, 1517-005098, 1517-005099, 1517-005100, 1517-005101, 1517-005102, 1517-005103, 1517-005104, 1517-005105, 1517-005106, 1517-005107, 1517-005108, 1517-005109, 1517-005110, 1517-005111, 1517-005112, 1517-005113, 1517-005114, 1517-005115, 1517-005116, 1517-005117, 1517-005118, 1517-005119, 1517-005120, 1517-005121, 1517-005122, 1517-005123, 1517-005124, 1517-005125, 1517-005126, 1517-005127, 1517-005128, 1517-005129, 1517-005130, 1517-005131, 1517-005132, 1517-005133, 1517-005134, 1517-005135, 1517-005136, 1517-005137, 1517-005138, 1517-005139, 1517-005140, 1517-005141, 1517-005142, 1517-005143, 1517-005144, 1517-005145, 1517-005146, 1517-005147, 1517-005148, 1517-005149, 1517-005150, 1517-005151, 1517-005152, 1517-005153, 1517-005154, 1517-005155, 1517-005156, 1517-005157, 1517-005158, 1517-005159, 1517-005160, 1517-005161, 1517-005162, 1517-005163, 1517-005164, 1517-005165, 1517-005166, 1517-005167, 1517-005168, 1517-005169, 1517-005170, 1517-005171, 1517-005172, 1517-005173, 1517-005174, 1517-005175, 1517-005176, 1517-005177, 1517-005178, 1517-005179, 1517-005180, 1517-005181, 1517-005182, 1517-005183, 1517-005184, 1517-005185, 1517-005186, 1517-005187, 1517-005188, 1517-005189, 1517-005190, 1517-005191, 1517-005192, 1517-005193, 1517-005194, 1517-005195, 1517-005196, 1517-005197, 1517-005198, 1517-005199, 1517-005200, 1517-005201, 1517-005202, 1517-005203, 1517-005204, 1517-005205, 1517-005206, 1517-005207, 1517-005208, 1517-005209, 1517-005210, 1517-005211, 1517-005212, 1517-005213, 1517-005214, 1517-005215, 1517-005216, 1517-005217, 1517-005218, 1517-005219, 1517-005220, 1517-005221, 1517-005222, 1517-005223, 1517-005224, 1517-005225, 1517-005226, 1517-005227, 1517-005228, 1517-005229, 1517-005230, 1517-005231, 1517-005232, 1517-005233, 1517-005234, 1517-005235, 1517-005236, 1517-005237, 1517-005238, 1517-005239, 1517-005240, 1517-005241, 1517-005242, 1517-005243, 1517-005244, 1517-005245, 1517-005246, 1517-005247, 1517-005248, 1517-005249, 1517-005250, 1517-005251, 1517-005252, 1517-005253, 1517-005254, 1517-005255, 1517-005256, 1517-005257, 1517-005258, 1517-005259, 1517-005260, 1517-005261, 1517-005262, 1517-005263, 1517-005264, 1517-005265, 1517-005266, 1517-005267, 1517-005268, 1517-005269, 1517-005270, 1517-005271, 1517-005272, 1517-005273, 1517-005274, 1517-005275, 1517-005276, 1517-005277, 1517-005278, 1517-005279, 1517-005280, 1517-005281, 1517-005282, 1517-005283, 1517-005284, 1517-005285, 1517-005286, 1517-005287, 1517-005288, 1517-005289, 1517-005290, 1517-005291, 1517-005292, 1517-005293, 1517-005294, 1517-005295, 1517-005346, 1517-005347, 1517-005348, 1517-005349, 1517-005350, 1517-005351, 1517-005356, 1517-005357, 1517-005358, 1517-005359, 1517-005360, 1517-005361, 1517-005362, 1517-005363, 1517-005364, 1517-005365, 1517-005366, 1517-005367, 1517-005368, 1517-005369, 1517-005370, 1517-005371, 1517-005372, 1517-005373, 1517-005374, 1517-005376, 1517-005377, 1517-005378, 1517-005379, 1517-005386, 1517-005387, 1517-005388, 1517-005389, 1517-005390, 1517-005391, 1517-005392, 1517-005393, 1517-005394, 1517-005395, 1517-005396, 1517-005397, 1517-005398, 1517-005399, 1517-005400, 1517-005401, 1517-005402, 1517-005403, 1517-005404, 1517-005405, 1517-005406, 1517-005407, 1517-005408, 1517-005409, 1517-005410, 1517-005411, 1517-005412, 1517-005413, 1517-005414, 1517-005415, 1517-005416, 1517-005417, 1517-005418, 1517-005419, 1517-005420, 1517-005421, 1517-005422, 1517-005423, 1517-005424, 1517-005425, 1517-005426, 1517-005427, 1517-005428, 1517-005429, 1517-005430, 1517-005431, 1517-005432, 1517-005433, 1517-005434, 1517-005435, 1517-005436, 1517-005437, 1517-005438, 1517-005439, 1517-005440, 1517-005441, 1517-005442, 1517-005443, 1517-005444, 1517-005445, 1517-005446, 1517-005447, 1517-005448, 1517-005449, 1517-005450, 1517-005451, 1517-005452, 1517-005453, 1517-005454, 1517-005455, 1517-005456, 1517-005457, 1517-005458, 1517-005459, 1517-005460, 1517-005461, 1517-005462, 1517-005463, 1517-005464, 1517-005465, 1517-005466, 1517-005467, 1517-005468, 1517-005469, 1517-005470, 1517-005471, 1517-005472, 1517-005473, 1517-005474, 1517-005475, 1517-005476, 1517-005477, 1517-005478, 1517-005479, 1517-005480, 1517-005481, 1517-005482, 1517-005483, 1517-005484, 1517-005485, 1517-005486, 1517-005487, 1517-005488, 1517-005489, 1517-005490, 1517-005491, 1517-005493, 1517-005494, 1517-005495, 1517-005496, 1517-005497, 1517-005498, 1517-005499, 1517-005500, 1517-005501, 1517-005502, 1517-005503, 1517-005504, 1517-005505, 1517-005506, 1517-005507, 1517-005508, 1517-005509, 1517-005510, 1517-005511, 1517-005512, 1517-005513, 1517-005514, 1517-005515, 1517-005516, 1517-005517, 1517-005518, 1517-005519, 1517-005520, 1517-005521, 1517-005522, 1517-005523, 1517-005524, 1517-005525, 1517-005526, 1517-005527, 1517-005528, 1517-005529, 1517-005530, 1517-005531, 1517-005532, 1517-005533, 1517-005534, 1517-005535, 1517-005536, 1517-005537, 1517-005538, 1517-005539, 1517-005540, 1517-005541, 1517-005542, 1517-005543, 1517-005544, 1517-005545, 1517-005546, 1517-005547, 1517-005548, 1517-005549, 1517-005550, 1517-005551, 1517-005552, 1517-005553, 1517-005554, 1517-005555, 1517-005556, 1517-005557, 1517-005558, 1517-005559, 1517-005560, 1517-005561, 1517-005562, 1517-005563, 1517-005564, 1517-005565, 1517-005566, 1517-005567, 1517-005568, 1517-005569, 1517-005570, 1517-005571, 1517-005572, 1517-005573, 1517-005574, 1517-005575, 1517-005576, 1517-005577, 1517-005578, 1517-005579, 1517-005580, 1517-005581, 1517-005582, 1517-005583, 1517-005584, 1517-005585, 1517-005586, 1517-005587, 1517-005588, 1517-005589, 1517-005590, 1517-005591, 1517-005592, 1517-005593, 1517-005594, 1517-005595, 1517-005596, 1517-005597, 1517-005601, 1517-005602, 1517-005603, 1517-005604, 1517-005605, 1517-005606, 1517-005607, 1517-005608, 1517-005609, 1517-005610, 1517-005611, 1517-005612, 1517-005613, 1517-005614, 1517-005615, 1517-005616, 1517-005617, 1517-005618, 1517-005619, 1517-005620, 1517-005621, 1517-005622, 1517-005623, 1517-005624, 1517-005625, 1517-005626, 1517-005627, 1517-005628, 1517-005629, 1517-005630, 1517-005631, 1517-005632, 1517-005633, 1517-005634, 1517-005635, 1517-005636, 1517-005637, 1517-005638, 1517-005639, 1517-005640, 1517-005641, 1517-005642, 1517-005643, 1517-005644, 1517-005645, 1517-005646, 1517-005647, 1517-005648, 1517-005649, 1517-005650, 1517-005651, 1517-005652, 1517-005653, 1517-005654, 1517-005655, 1517-005656, 1517-005657, 1517-005658, 1517-005659, 1517-005660, 1517-005661, 1517-005662, 1517-005663, 1517-005664, 1517-005665, 1517-005666, 1517-005667, 1517-005668, 1517-005669, 1517-005670, 1517-005671, 1517-005672, 1517-005673, 1517-005674, 1517-005675, 1517-005676, 1517-005677, 1517-005678, 1517-005679, 1517-005680, 1517-005681, 1517-005682, 1517-005683, 1517-005684, 1517-005685, 1517-005686, 1517-005687, 1517-005688, 1517-005689, 1517-005690, 1517-005691, 1517-005692, 1517-005693, 1517-005694, 1517-005695, 1517-005696, 1517-005697, 1517-005698, 1517-005699, 1517-005700, 1517-005701, 1517-005702, 1517-005703, 1517-005704, 1517-005705, 1517-005706, 1517-005707, 1517-005709, 1517-005710, 1517-005711, 1517-005712, 1517-005713, 1517-005714, 1517-005715, 1517-005716, 1517-005717, 1517-005718, 1517-005719, 1517-005720, 1517-005721, 1517-005722, 1517-005723, 1517-005724, 1517-005725, 1517-005726, 1517-005727, 1517-005728, 1517-005729, 1517-005730, 1517-005731, 1517-005732, 1517-005733, 1517-005734, 1517-005735, 1517-005736, 1517-005737, 1517-005738, 1517-005739, 1517-005740, 1517-005741, 1517-005742, 1517-005743, 1517-005744, 1517-005745, 1517-005746, 1517-005747, 1517-005748, 1517-005749, 1517-005750, 1517-005751, 1517-005752, 1517-005753, 1517-005754, 1517-005755, 1517-005756, 1517-005757, 1517-005758, 1517-005759, 1517-005760, 1517-005761, 1517-005762, 1517-005763, 1517-005764, 1517-005765, 1517-005766, 1517-005767, 1517-005768, 1517-005769, 1517-005770, 1517-005771, 1517-005772, 1517-005773, 1517-005774, 1517-005775, 1517-005776, 1517-005777, 1517-005778, 1517-005779, 1517-005780, 1517-005781, 1517-005782, 1517-005783, 1517-005784, 1517-005785, 1517-005786, 1517-005787, 1517-005788, 1517-005789, 1517-005790, 1517-005791, 1517-005792, 1517-005793, 1517-005794, 1517-005795, 1517-005796, 1517-005797, 1517-005798, 1517-005799, 1517-005800, 1517-005852, 1517-005853, 1517-005854, 1517-005855, 1517-005856, 1517-005857, 1517-005858, 1517-005859, 1517-005860, 1517-005861, 1517-005862, 1517-005863, 1517-005864, 1517-005865, 1517-005866, 1517-005867, 1517-005868, 1517-005869, 1517-005870, 1517-005871, 1517-005872, 1517-005873, 1517-005874, 1517-005875, 1517-005876, 1517-005877, 1517-005878, 1517-005879, 1517-005880, 1517-005881, 1517-005882, 1517-005883, 1517-005884, 1517-005885, 1517-005886, 1517-005887, 1517-005888, 1517-005889, 1517-005890, 1517-005891, 1517-005892, 1517-005893, 1517-005894, 1517-005895, 1517-005896, 1517-005897, 1517-005898, 1517-005899, 1517-005900, 1517-005901, 1517-005902, 1517-005903, 1517-005904, 1517-005905, 1517-005906, 1517-005907, 1517-005908, 1517-005909, 1517-005910, 1517-005911, 1517-005912, 1517-005913, 1517-005914, 1517-005915, 1517-005916, 1517-005917, 1517-005918, 1517-005919, 1517-005920, 1517-005921, 1517-005972, 1517-005973, 1517-005974, 1517-005975, 1517-005976, 1517-005977, 1517-005978, 1517-005979, 1517-005980, 1517-005981, 1517-005982, 1517-005983, 1517-005984, 1517-005985, 1517-005986, 1517-005987, 1517-005988, 1517-005989, 1517-005990, 1517-005991, 1517-005992, 1517-005993, 1517-005994, 1517-005995, 1517-005996, 1517-005997, 1517-005998, 1517-005999, 1517-006000, 1517-006001, 1517-006002, 1517-006003, 1517-006004, 1517-006005, 1517-006006, 1517-006007, 1517-006008, 1517-006009, 1517-006010, 1517-006011, 1517-006012, 1517-006013, 1517-006014, 1517-006015, 1517-006016, 1517-006017, 1517-006018, 1517-006019, 1517-006020, 1517-006021, 1517-006022, 1517-006023, 1517-006024, 1517-006025, 1517-006026, 1517-006027, 1517-006028, 1517-006029, 1517-006030, 1517-006031, 1517-006032, 1517-006033, 1517-006034, 1517-006035, 1517-006036, 1517-006037, 1517-006038, 1517-006039, 1517-006040, 1517-006041, 1517-006042, 1517-006043, 1517-006044, 1517-006045, 1517-006046, 1517-006047, 1517-006050, 1517-006051, 1517-006052, 1517-006053, 1517-006054, 1517-006055, 1517-006056, 1517-006057, 1517-006058, 1517-006059, 1517-006060, 1517-006061, 1517-006062, 1517-006067, 1517-006068, 1517-006069, 1517-006070, 1517-006075, 1517-006076, 1517-006077, 1517-006078, 1517-006079, 1517-006080, 1517-006081, 1517-006082, 1517-006083, 1517-006084, 1517-006085, 1517-006086, 1517-006093, 1517-006094, 1517-006095, 1517-006096, 1517-006097, 1517-006098, 1517-006099, 1517-006100, 1517-006101, 1517-006102, 1517-006103, 1517-006104, 1517-006105, 1517-006106, 1517-006107, 1517-006108, 1517-006109, 1517-006110, 1517-006111, 1517-006112, 1517-006113, 1517-006114, 1517-006115, 1517-006116, 1517-006117, 1517-006118, 1517-006119, 1517-006120, 1517-006121, 1517-006122, 1517-006123, 1517-006124, 1517-006125, 1517-006126, 1517-006127, 1517-006128, 1517-006129, 1517-006130, 1517-006131, 1517-006132, 1517-006133, 1517-006134, 1517-006135, 1517-006136, 1517-006137, 1517-006138, 1517-006139, 1517-006140, 1517-006141, 1517-006142, 1517-006143, 1517-006144, 1517-006145, 1517-006146, 1517-006147, 1517-006148, 1517-006149, 1517-006155, 1517-05492, 1517-06063, 1517-06064, 1517-06065, 1517-06066, 91517-001851, 91517-001852, 91517-001853, 91517-001854, 91517-001855, 91517-001856, 91517-001857, 91517-001858, 91517-001859, 91517-001860, 91517-001861, 91517-001862, 91517-001863, 91517-001864, 91517-001865, 91517-001866, 91517-001867, 91517-001868, 91517-001869, 91517-001870, 91517-001871, 91517-001872, 91517-001873, 91517-001874, 91517-001875, 91517-001876, 91517-001882, 91517-001883, 91517-001884, 91517-001895, 91517-001896, 91517-001897, and 91517-001898. |
Recalling Firm/ Manufacturer |
Spacelabs Healthcare, Llc 5150 220th Ave Se Issaquah WA 98029-6834
|
| For Additional Information Contact | Spacelabs Healthcare 800-522-7025 |
Manufacturer Reason for Recall | Capnography Module, Model 91517, resets 20 seconds after power failure, instead of after three minutes as specified. This reset clears the RAM and the module returns to default settings. |
FDA Determined Cause 2 | Software design |
| Action | Spacelabs Medical sent a "URGENT-MEDICAL DEVICE CORRECTION" letter dated May 12, 2011 to all affected customers. The letter identifies the product, problem, and actions to be taken.
Customers were recommended to set all their Capnography Module Configuration Manager alarm settings to "ON". The letter also recommended that customers brief their staff about the possibility of losing changes to the default settings and to restore them after a power interruption.
A field service engineer will be contacting customers to set up a date and time when affected modules can be upgraded at no cost to the customer.
For questions about this corrective action program, call Spacelabs Medical at 1-800-522-7025, Select 2 for Technical Support. For outside the US, customers can call 1-425-657-7200, x5089. |
| Quantity in Commerce | 5883 units |
| Distribution | Worldwide Distribution--USA (nationwide) and countries of ALGERIA, ARGENTINA, AUSTRALIA, BAHRAIN, BANGLADESH, BELGIUM,
BOLIVIA, BRAZIL, CANADA, CHILE, CHINA,COLOMBIA, CYPRUS, CZECH REPUBLIC, DOMINICAN REPUBLIC, ECUADOR, EGYPT, ESTONIA, FRANCE,
GERMANY, GREECE, GUATEMALA, HONG KONG, INDIA, ISRAEL, ITALY,
JORDAN, KUWAIT, LEBANON, MALAYSIA, MEXICO, MOROCCO, NEPAL,
NETHERLANDS, NEW ZEALAND, NORWAY, PAKISTAN, PALESTINE, PANAMA,
PERU, POLAND, PORTUGAL, ROMANIA, SAUDI ARABIA, SINGAPORE, SLOVAKIA, SOUTH AFRICA, SPAIN, SRI LANKA, SWEDEN, SWITZERLAND,
SYRIAN ARAB REPUBLIC, TAIWAN, REPUBLIC OF CHINA, TRINIDAD AND TOBAGO, TUNISIA, TURKEY, UNITED ARAB EMIRATES, UNITED KINGDOM,
VENEZUELA, and VIET NAM. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
| 510(K) Database | 510(K)s with Product Code = MHX
|
|
|
|