| Class 2 Device Recall Philips Healthcare PCR Eleva Radiological Image Processing System | |
Date Initiated by Firm | July 27, 2011 |
Date Posted | September 14, 2011 |
Recall Status1 |
Terminated 3 on January 12, 2017 |
Recall Number | Z-3205-2011 |
Recall Event ID |
59473 |
510(K)Number | K964124 |
Product Classification |
Computed Radiography Image Processing System - Product Code LLZ
|
Product | PCR Eleva Radiological Image Processing System with software R.1.2
Philips Healthcare, Inc.
A computed radiography image processing system. |
Code Information |
Software version 1.2 Models: 732040, 732042, 732044, 732048, 732050, 732052, 732080, 732082 |
Recalling Firm/ Manufacturer |
Philips Healthcare Inc. 3000 Minuteman Road Andover MA 01810
|
For Additional Information Contact | SAME 978-687-1501 |
Manufacturer Reason for Recall | The mirror icon on the bottom of each image could be interpreted as a RIGHT marker for side indication if the image is mirrored again within a PACS system. This could potentially lead to a misdiagnosis |
FDA Determined Cause 2 | Software design |
Action | Philips Healthcare Inc. sent out "Urgent --Medical Device Correction" notices dated June 10, 2011 to all affected customers.
The letter described the product and problem. Customers were informed the "burned-in Mirror-Icon" may be misinterpreted as a Right marker. Customers are advised to preferably use lead markers during each X-ray acquisition or alternatively electronic annotations on each image for side indication. The presentation of these markers within each image will clearly show the difference between the intended Left/Right indication and the little mirror ICON. Philips will install improved software that prevents the misinterpretation of the symbol and will schedule the installation of the updated software.
For questions on this recall call Phillips Healthcare Inc. at 978-687-1501. |
Quantity in Commerce | 197 units |
Distribution | Worldwide Distribution: Nationwide distribution; and including the countries of Albania, Angola, Argentina, Australia, Austria, Azerbaijan, Belgium, Bolivia, Brazil, Bulgaria, Canada, Chile, China, Croatia, Denmark, Egypt, France, Germany, Ghana, Greece, Greece, Haiti, Hungary, Indonesia, Iraq, Israel, Italy, Kazakhstan, Kenya, Latvia, Libya, Liechtenstein, Lithuania, Malawi, Malta, Martinique, Mexico, Mozambique, Netherlands, New Zealand, Nigeria, Norway, Oman, Philippines, Portugal, Puerto Rico, Romania, Russia, Saudi-Arabia, Serbia, Slovakia, South Africa, Spain, Sudan, Sweden, Switzerland, Tanzania, Thailand, Tschech. Rep, Turkey, Turkmenistan, UAE, Uganda, and United Kingdom. |
Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database | 510(K)s with Product Code = LLZ
|
|
|
|