• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall TeleSentry, Multiparameter Ambulatory Telemetry

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall TeleSentry, Multiparameter Ambulatory Telemetrysee related information
Date Initiated by FirmJanuary 19, 2012
Date PostedApril 13, 2012
Recall Status1 Terminated 3 on September 27, 2013
Recall NumberZ-1400-2012
Recall Event ID 61370
510(K)NumberK092947 
Product Classification Detector and alarm, arrhythmia - Product Code DSI
ProductTeleSentry, Multi-parameter Ambulatory Telemetry, Model #TS01. Dist. By: ScottCare Corporation, Cleveland, OH. The TeleSentry device is intended for diagnostic evaluation of patients who experience transient symptoms or asymptomatic events that may suggest cardiac arrhythmia. The device continuously monitors and records the data, automatically records events triggered by an arrhythmia detection algorithm or manually by the patient, and automatically transmits the recorded event activity associated with these symptoms for review by a licensed physician.
Code Information Model #TS01, Part #101086, Serial #s: DV10Y01B1987, DV10Y01B1624, DV10Y01B1686, DV10Y01B1747, DV10Y01B1780, DV10Y01B1967, DV10Y01B1268, DV10Y01B1268, DV10Y01B0032, DV10Y01B0119, DV10Y01B0335, DV10Y01B0502

DV10Y01B0505, DV10Y01B0692, DV10Y01B0780, DV10Y01B0825, DV10Y01B0880, DV10Y01B0989, DV10Y01B0992, DV10Y01B1157, DV10Y01B1408, DV10Y01B1497, DV10Y01B1499, DV10Y01B1500, DV10Y01B1531, DV10Y01B1559, DV10Y01B1583, DV10Y01B1594, DV10Y01B1597, DV10Y01B1602, DV10Y01B1605, DV10Y01B1660, DV10Y01B1721, DV10Y01B1739, DV10Y01B1800, DV10Y01B1802, DV10Y01B1817, DV10Y01B1827, DV10Y01B1847, DV10Y01B1874, DV10Y01B1886, DV10Y01B1938, DV10Y01B1954, DV10Y01B1965, DV10Y01B1968, DV10Y01B1972, DV10Y01B1980, DV10Y01B1999, DV10Y01B2000, DV10Y01B0086, DV10Y01B1513, DV10Y01B1551, DV10Y01B1568, DV10Y01B1579, DV10Y01B1695, DV10Y01B1723, DV10Y01B1745, DV10Y01B1791, DV10Y01B1825, DV10Y01B1826, DV10Y01B1868, DV10Y01B1873, DV10Y01B1879, DV10Y01B1884, DV10Y01B1903, DV10Y01B1926, DV10Y01B1931, DV10Y01B1937, DV10Y01B1956, DV10Y01B1960, DV10Y01B1970, DV10Y01B1977, DV10Y01B1979, DV10Y01B1996, DV10Y01B0003, DV10Y01B0715, DV10Y01B0957, DV10Y01B0988, DV10Y01B1502, DV10Y01B1520, DV10Y01B1556, DV10Y01B1563, DV10Y01B1566, DV10Y01B1585, DV10Y01B1592, DV10Y01B1630, DV10Y01B1647, DV10Y01B1674, DV10Y01B1694, DV10Y01B1733, DV10Y01B1765, DV10Y01B1787, DV10Y01B1799, DV10Y01B1803, DV10Y01B1823, DV10Y01B1835, DV10Y01B1838, DV10Y01B1843, DV10Y01B1865, DV10Y01B1893, DV10Y01B1894, DV10Y01B1902, DV10Y01B1955, DV10Y01B1994, DV10Y01B1852, DV10Y01B1880, DV10Y01B1885, DV10Y01B1890, DV10Y01B1636, DV10Y01B1684, DV10Y01B1705, DV10Y01B1860, DV10Y01B1863, DV10Y01B1729, DV10Y01B1925, DV10Y01B0746, DV10Y01B1572, DV10Y01B1575, DV10Y01B1608, DV10Y01B1637, DV10Y01B1655, DV10Y01B1735, DV10Y01B1737, DV10Y01B1755, DV10Y01B1775, DV10Y01B1830, DV10Y01B0450, DV10Y01B0600, DV10Y01B0631, DV10Y01B0662, DV10Y01B0685, DV10Y01B0748, DV10Y01B0845, DV10Y01B0914, DV10Y01B0915, DV10Y01B0949, DV10Y01B0950, DV10Y01B0963, DV10Y01B0990, DV10Y01B0996, DV10Y01B1009, DV10Y01B1498, DV10Y01B1501, DV10Y01B1503, DV10Y01B1506, DV10Y01B1516, DV10Y01B1517, DV10Y01B1519, DV10Y01B1521, DV10Y01B1523, DV10Y01B1524, DV10Y01B1525, DV10Y01B1527, DV10Y01B1530, DV10Y01B1532, DV10Y01B1533, DV10Y01B1534, DV10Y01B1540, DV10Y01B1541, DV10Y01B1542, DV10Y01B1543, DV10Y01B1545, DV10Y01B1547, DV10Y01B1548, DV10Y01B1550, DV10Y01B1552, DV10Y01B1555, DV10Y01B1557, DV10Y01B1558, DV10Y01B1560, DV10Y01B1561, DV10Y01B1562, DV10Y01B1564, DV10Y01B1565, DV10Y01B1567, DV10Y01B1570, DV10Y01B1571, DV10Y01B1574, DV10Y01B1588, DV10Y01B1595, DV10Y01B1596, DV10Y01B1598, DV10Y01B1599, DV10Y01B1604, DV10Y01B1607, DV10Y01B1609, DV10Y01B1610, DV10Y01B1611, DV10Y01B1615, DV10Y01B1619, DV10Y01B1621, DV10Y01B1622, DV10Y01B1623, DV10Y01B1625, DV10Y01B1626, DV10Y01B1627, DV10Y01B1633, DV10Y01B1635, DV10Y01B1642, DV10Y01B1646, DV10Y01B1648, DV10Y01B1649, DV10Y01B1652, DV10Y01B1658, DV10Y01B1663, DV10Y01B1664, DV10Y01B1665, DV10Y01B1670, DV10Y01B1676, DV10Y01B1678, DV10Y01B1681, DV10Y01B1683, DV10Y01B1687, DV10Y01B1688, DV10Y01B1690, DV10Y01B1691, DV10Y01B1693, DV10Y01B1696, DV10Y01B1697, DV10Y01B1698, DV10Y01B1701, DV10Y01B1703, DV10Y01B1707, DV10Y01B1708, DV10Y01B1710, DV10Y01B1711, DV10Y01B1712, DV10Y01B1714, DV10Y01B1715, DV10Y01B1717, DV10Y01B1719, DV10Y01B1726, DV10Y01B1727, DV10Y01B1728, DV10Y01B1730, DV10Y01B1734, DV10Y01B1738, DV10Y01B1740, DV10Y01B1752, DV10Y01B1754, DV10Y01B1757, DV10Y01B1760, DV10Y01B1763, DV10Y01B1764, DV10Y01B1767, DV10Y01B1769, DV10Y01B1770,. DV10Y01B1771, DV10Y01B1772, DV10Y01B1773, DV10Y01B1776, DV10Y01B1777, DV10Y01B1778, DV10Y01B1779, DV10Y01B1782, DV10Y01B1785, DV10Y01B1786, DV10Y01B1793, DV10Y01B1806, DV10Y01B1809, DV10Y01B1811, DV10Y01B1812, DV10Y01B1816, DV10Y01B1818, DV10Y01B1821, DV10Y01B1822, DV10Y01B1829, DV10Y01B1831, DV10Y01B1832, DV10Y01B1836, DV10Y01B1840, DV10Y01B1844, DV10Y01B1846, DV10Y01B1849, DV10Y01B1851, DV10Y01B1853, DV10Y01B1855, DV10Y01B1857, DV10Y01B1859, DV10Y01B1861, DV10Y01B1862, DV10Y01B1864, DV10Y01B1867, DV10Y01B1869, DV10Y01B1878, DV10Y01B1882, DV10Y01B1888, DV10Y01B1896, DV10Y01B1898, DV10Y01B1900, DV10Y01B1901, DV10Y01B1904, DV10Y01B1906, DV10Y01B1908, DV10Y01B1911, DV10Y01B1912, DV10Y01B1913, DV10Y01B1914, DV10Y01B1917, DV10Y01B1918, DV10Y01B1919, DV10Y01B1920, DV10Y01B1922, DV10Y01B1927, DV10Y01B1928, DV10Y01B1930, DV10Y01B1932, DV10Y01B1933, DV10Y01B1935, DV10Y01B1939, DV10Y01B1940, DV10Y01B1941, DV10Y01B1942, DV10Y01B1943, DV10Y01B1944, DV10Y01B1945, DV10Y01B1947, DV10Y01B1949, DV10Y01B1951, DV10Y01B1952, DV10Y01B1957, DV10Y01B1958, DV10Y01B1959, DV10Y01B1961, DV10Y01B1962, DV10Y01B1963, DV10Y01B1964, DV10Y01B1969, DV10Y01B1976, DV10Y01B1981, DV10Y01B1982, DV10Y01B1983, DV10Y01B1985, DV10Y01B1986, DV10Y01B1988, DV10Y01B1989, DV10Y01B1990, DV10Y01B1992, DV10Y01B1993, DV10Y01B1995, DV10Y01B1998, DV10Y01B1515, DV10Y01B1554, DV10Y01B1692, DV10Y01B1720, DV10Y01B1864, DV10Y01B1616, DV10Y01B1732, DV10Y01B1916, DV10Y01B0982, DV10Y01B1549, DV10Y01B1569, DV10Y01B1724, DV10Y01B1731, DV10Y01B1751, DV10Y01B1948, DV10Y01B1553, DV10Y01B1871, DV10Y01B1514, DV10Y01B1529, DV10Y01B1704, DV10Y01B1866, DV10Y01B1883, DV10Y01B1950, DV10Y01B1953, DV10Y01B0048, DV10Y01B0193, DV10Y01B0620, DV10Y01B0991, DV10Y01B0995, DV10Y01B1273, DV10Y01B1504, DV10Y01B1507, DV10Y01B1522, DV10Y01B1539, DV10Y01B1546, DV10Y01B1606, DV10Y01B1634, DV10Y01B1644, DV10Y01B1651, DV10Y01B1657, DV10Y01B1673, DV10Y01B1699, DV10Y01B1709, DV10Y01B1716, DV10Y01B1743, DV10Y01B1762, DV10Y01B1766, DV10Y01B1784, DV10Y01B1792, DV10Y01B1805, DV10Y01B1807, DV10Y01B1808, DV10Y01B1828, DV10Y01B1833, DV10Y01B1842, DV10Y01B1845, DV10Y01B1870, DV10Y01B1875, DV10Y01B1881, DV10Y01B1887, DV10Y01B1891, DV10Y01B1923, DV10Y01B1978, DV10Y01B1984, DV10Y01B0255, DV10Y01B1510, DV10Y01B1512, DV10Y01B1518, DV10Y01B1544, DV10Y01B1576, DV10Y01B1578, DV10Y01B1586, DV10Y01B1614, DV10Y01B1628, DV10Y01B1638, DV10Y01B1643, DV10Y01B1645, DV10Y01B1685, DV10Y01B1689, DV10Y01B1748, DV10Y01B1750, DV10Y01B1781, DV10Y01B1795, DV10Y01B1856, DV10Y01B0565, DV10Y01B1535, DV10Y01B1537, DV10Y01B1584, DV10Y01B1629, DV10Y01B1659, DV10Y01B1662, DV10Y01B1666, DV10Y01B1669, DV10Y01B1768, DV10Y01B1801, DV10Y01B1804, DV10Y01B1573, DV10Y01B1587, DV10Y01B1653, DV10Y01B1858, DV10Y01B1872, DV10Y01B1877, DV10Y01B1921, DV10Y01B1929, DV10Y01B1934, DV10Y01B1936, DV10Y01B1975, DV10Y01B1577, DV10Y01B1617, DV10Y01B1971, DV10Y01B1850, DV10Y01B1834, DV10Y01B0036, DV10Y01B0117, DV10Y01B0160, DV10Y01B0531, DV10Y01B0593, DV10Y01B0994, DV10Y01B1508, DV10Y01B1526, DV10Y01B1581, DV10Y01B1590, DV10Y01B1593, DV10Y01B1600, DV10Y01B1618, DV10Y01B1679, DV10Y01B1706, DV10Y01B1744, DV10Y01B1753, DV10Y01B1783, DV10Y01B1788, DV10Y01B1798, DV10Y01B1820, DV10Y01B1841, DV10Y01B1974, DV10Y01B0533, DV10Y01B1997, DV10Y01B0019, DV10Y01B0941, DV10Y01B1115, DV10Y01B1224, DV10Y01B1528, DV10Y01B1603, DV10Y01B1656, DV10Y01B1667, DV10Y01B1713, DV10Y01B1722, DV10Y01B1725, DV10Y01B1854, DV10Y01B1876, DV10Y01B1889, DV10Y01B1897, DV10Y01B1899, DV10Y01B1905, DV10Y01B1907, DV10Y01B1910, DV10Y01B1915, DV10Y01B1924, DV10Y01B1946, DV10Y01B1966, DV10Y01B0030, DV10Y01B1505, DV10Y01B1509, DV10Y01B1511, DV10Y01B1536, DV10Y01B1538, DV10Y01B1580, DV10Y01B1582, DV10Y01B1589, DV10Y01B1591, DV10Y01B1601, DV10Y01B1612, DV10Y01B1613, DV10Y01B1620, DV10Y01B1631, DV10Y01B1632, DV10Y01B1639, DV10Y01B1640, DV10Y01B1650, DV10Y01B1654, DV10Y01B1661, DV10Y01B1668, DV10Y01B1671, DV10Y01B1672, DV10Y01B1675, DV10Y01B1677, DV10Y01B1680, DV10Y01B1682, DV10Y01B1700, DV10Y01B1702, DV10Y01B1718, DV10Y01B1736, DV10Y01B1741, DV10Y01B1742, DV10Y01B1746, DV10Y01B1749, DV10Y01B1756, DV10Y01B1758, DV10Y01B1759, DV10Y01B1774, DV10Y01B1789, DV10Y01B1790, DV10Y01B1794, DV10Y01B1796, DV10Y01B1797, DV10Y01B1810, DV10Y01B1813, DV10Y01B1814, DV10Y01B1815, DV10Y01B1819, DV10Y01B1824, DV10Y01B1837, DV10Y01B1839, DV10Y01B1848, DV10Y01B1895, DV10Y01B1909, DV10Y01B1973, DV10Y01B1991, DV10Y01B1641, & DV10Y01B1761.
FEI Number 1527715
Recalling Firm/
Manufacturer
Scottcare Corporation
4791 W 150th St
Cleveland OH 44135-3301
For Additional Information ContactTimothy J. Leyva
216-362-0550 Ext. 113
Manufacturer Reason
for Recall
Certain TeleSentry Devices contain a Bad SD Card. When the SD Card goes bad, the TeleSentry Device cannot read or write data to the SD Card. The TeleSentry Device then stops functioning.
FDA Determined
Cause 2
Component design/selection
ActionScottCare sent an "URGENT MEDICAL DEVICE RECALL" letter dated January 19, 2012 to all affected customers. The letter identifies the product, problem, and the actions to be taken by the customers. The letter instructs customers to immediately discontinue use and quarantine any affected products. A Recall Acknowledgement Form was enclosed for customers to complete and return via fax to (216) 257-6129. Contact the firm at (216) 362-0550 ext. 113 for questions regarding this recall.
Quantity in Commerce513 units
DistributionWorldwide Distribution-USA (nationwide) including the states of CA, DE, FL, MI, NJ, NV, NY, OH, PA and TX, and the country of Canada.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
510(K) Database510(K)s with Product Code = DSI
-
-