• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall Terumo Advanced Perfusion System 1

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Terumo Advanced Perfusion System 1see related information
Date Initiated by FirmJune 25, 2012
Date PostedJuly 09, 2012
Recall Status1 Terminated 3 on March 01, 2013
Recall NumberZ-1971-2012
Recall Event ID 62373
510(K)NumberK022947 
Product Classification Console, heart-lung machine, cardiopulmonary bypass - Product Code DTQ
ProductSystem 1 Base 220/240V. The Terumo Advanced Perfusion System 1 is indicated for use for up to 6 hours on the extracorporeal circulation of blood for arterial perfusion, regional perfusion, and cardiopulmonary bypass procedures, when used by a qualified medical professional who is experienced in the operation of this or similar equipment.
Code Information catalog number: 801764 and serial numbers: 0006-0066, 0100-0311, 0313-0322, 1001-1027, 1100-1434, 1436,1440, and 1441.
FEI Number 1828100
Recalling Firm/
Manufacturer
Terumo Cardiovascular Systems Corporation
6200 Jackson Road
Ann Arbor MI 48103-9586
For Additional Information Contact
734-741-6173
Manufacturer Reason
for Recall
Terumo Cardiovascular Systems (Terumo CVS) has received reports of various roller pump malfunctions of the Terumo Advanced Perfusion System 1 that could result in the inability to use the pump and would then require replacement with back-up equipment. Description of malfunctions: " Unplanned pump stop or pause " Changes in pump speed If it is necessary to replace a roller pump during cardiopu
FDA Determined
Cause 2
Device Design
ActionTerumo CVS sent an "URGENT MEDICAL DEVICE RECALL" letter dated May 4, 2012 to all affected customers. The letter identifies the product, problem, and actions to be taken by the customers. Customers were asked to complete and return a Customer Response Form to the firm via fax at 1-800-292-6551. Contact the firm at 1-800-521-2818 for questions regarding this recall.
Quantity in Commerce648 units
DistributionWorldwide Distribution-USA (nationwide) and the countries of ARGENTINA, AUSTRALIA, BELGIUM, CANADA, CHILE, COLOMBIA, COLOMBIA, Costa Rica, Dominican Republic, Egypt, Germany, Guatamala, Honduras, Hong Kong, India, Indonesia, Israel, Japan, Jordan, Kuwait, Malaysia, Mexico, Pakistan, Philippines, Russia, Saudi Arabia, Singapore, South Africa, South Korea, Taiwan, Thailand, Turkey, UNITED ARAB EMIRATES (UAE), and Vietnam.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
510(K) Database510(K)s with Product Code = DTQ
-
-