• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall OEC 9800 Plus

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
  Class 2 Device Recall OEC 9800 Plus see related information
Date Initiated by Firm March 27, 2007
Date Posted August 01, 2012
Recall Status1 Terminated 3 on August 02, 2012
Recall Number Z-2125-2012
Recall Event ID 62449
510(K)Number K021049  
Product Classification System, x-ray, fluoroscopic, image-intensified - Product Code IZL
Product OEC 9800 Plus Digital; MDL number A349855.

Product Usage: Is designed to provide fluoroscopic and spot-film imaging of the patient during diagnostic, surgical and interventional procedures. Clinical applications may include cholangiography, endoscopic, urologic, orthopedic, neurologic, vascular, cardiac, critical care and emergency room procedures. The system may be used for other imaging applications at the physician's discretion.
Code Information 62-0212 , 69-0054 , 82-0023-RC , 82-0034 , 82-0072 , 82-0274 , 82-0587, 8S-0668 , 82-1710 , 82-1803 , 82-2147 , 82-2620 , 82-2754 , 82-2815 , 82-3014 , 82-3234 , 82-7010-MH , 86-0254 , 86-0354 , 86-1078 , 89-0505 , 89-0509 , 89-0595 , 89-0605 , 89-0987 , 89-1048 , 89-1776 , 89-1837 , 89-2132 , 89-2303 , 89-2335 , 89-2348 , 89-2350 , 89-2610 , 89-2645 , 8S-0066 , 8S-0079 , 8S-0341 , 8S-0562 , 8S-0615-C , 8S-0847 , 8S-0849-C , 8S-1312 , 8S-1971 , 8S-2015-N , 8S-2116 , 8S-2117 , 8S-2199 , 8S-2204 , 8S-2320 , 8S-2411 , P4-0097-R , P4-0101-R , P4-0325-R , P4-0346-L , P4-0477-L , P6-0432-R , W2-0012 , W2-0580-R , W2-0591-R , 82-0030
Recalling Firm/
Manufacturer
GE OEC Medical Systems, Inc
384 Wright Brothers Dr
Salt Lake City UT 84116-2862
For Additional Information Contact Deena Pease
801-536-4516
Manufacturer Reason
for Recall
Please be aware that this is not a new recall. The firm has taken action; but, due to administrative issues this recall is now being classified by the Agency. GE OEC had discovered an issue with the hard disk drives that were installed on OEC device models: UroView 2800; 9800 Plus Digital; MiniView 6800; and FlexiView 8800 Mobile C-Arm. The devices were manufactured between October 15, 2004 and
FDA Determined
Cause 2
Software design
Action Please be aware that this is not a new recall. The firm has taken action; but, due to administrative issues this recall is now being classified by the Agency. GE OEC sent an Urgent Safety Advisory Notices dated October 11, 2006 and February 21, 2007 via certified mail delivery letter to affected customers. The notices identified the affected products, problem and actions to be taken. Customers were informed that a GEHC OEC field service engineer would be contacting their facility to arrange to have a replacement hard disk drive installed on the affected system(s). Customers were instructed to ensure that images on the hard disk drive were stored via an alternate permanent media as the hard disk drive approaches the 170 image capacity until a GEHC OEC field service engineer had replaced the drive. For questions regardiug this issues contact the service team at 800-874-7378.
Quantity in Commerce Part of a total of 82 units distributed.
Distribution Worldwide Distribution - US (nationwide) and the countries of: Canada, France, Japan, and China.
Total Product Life Cycle TPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
510(K) Database 510(K)s with Product Code = IZL and Original Applicant = GE OEC MEDICAL SYSTEMS
-
-