• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall IDSiSYS Ostase BAP (Bone Alkaline Phosphatase),

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall IDSiSYS Ostase BAP (Bone Alkaline Phosphatase),see related information
Date Initiated by FirmAugust 10, 2012
Date PostedOctober 10, 2012
Recall Status1 Terminated 3 on December 17, 2012
Recall NumberZ-0040-2013
Recall Event ID 62968
Product Classification N/A - Product Code CIN
ProductIDS-iSYS Ostase BAP (Bone Alkaline Phosphatase), Catalog Number IS-2800. Packaged in cardboard Box containing plastic bottles. The IDS-iSYS Ostase BAP Assay (IDS-iSYS BAP) is intended for the quantitative determination of bone-specific alkaline phosphatase (BAP), an indicator of osteoblastic activity, in human serum or plasma on the IDS-iSYS Multi- Discipline Automated Analyser (Analyser). Results are to be used in conjunction with other clinical and laboratory data to aid the clinician in the management of post menopausal osteoporosis and Pagets disease.
Code Information Lot 1184, EXP Aug 2012 
FEI Number 3002485711
Recalling Firm/
Manufacturer
Immunodiagnostics Systems Ltd
10 Didcot Way Bolden Business Park
Boldon Tyne & Wear United Kingdom
Manufacturer Reason
for Recall
It has been noted by staff at IDS Inc. that stock of the affected products held at IDS Inc. are labeled with IVD labels although they are not registered for sale as IVDs in the United States. The products should have been labeled as Research Use Only or Investigational use only dependent on their usage.
FDA Determined
Cause 2
Labeling False and Misleading
ActionImmunodiagnostics Systems sent Customer notification letters via email and notified customers of the recall. The notification included the reason for recall, product information with codes, lot number, manufacturing/distribution date, and exp date. The notification included instructions to customers: Actions to be taken by the customer: Customers should complete the attached Recall and Acknowledgement Form and take the following actions: 1) Check your stock for the products included within the scope of the recall. Cease use and distribution and quarantine all affected product lots immediately. 2) Complete the enclosed Recall & Acknowledgement Form and immediately fax back to Immunodiagnostic Systems. This will allow us to document your receipt of this letter and the amount of product you have on hand for return. 3) Return any affected product freight collect, along with the original completed Recall & Acknowledgement Form to your local IDS office. Contact information was also provided. For questions regarding this recall call +44 (191) 5195212.
Quantity in Commerce3
DistributionNationwide Distribution including NY, CA, ME and CT.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
-
-