|
Class 2 Device Recall Qube Compact Monitor Model 91390 |
 |
Date Initiated by Firm |
October 19, 2012 |
Date Posted |
November 20, 2012 |
Recall Status1 |
Terminated 3 on May 22, 2013 |
Recall Number |
Z-0403-2013 |
Recall Event ID |
63504 |
510(K)Number |
K120616
|
Product Classification |
Monitor, physiological, patient(with arrhythmia detection or alarms) - Product Code MHX
|
Product |
qube Compact Monitor (Patient Monitor) 91390. The monitor is 26.2 cm x 31.5 cm x 13.2 cm and weighs 4.1 kg (10.3 inches x 12.4 inches x 5.2 inches - 9 lbs). It has a 26.2 cm (12.1 inches) LCD display with 1024 x 768 resolution.
The Spacelabs Healthcare Qube Compact Monitor (91390), functioning as either bedside or central monitors; passively displays data generated by 2 Spacelabs Healthcare parameter modules, Flexports interfaces, and other SDLC based products in the form of waveform and numeric displays, trends and alarms. Key monitored parameters available on the Model 91390 when employing the Spacelabs Command Module, consist of ECU, respiration, invasive and noninvasive blood pressure, Sp02, temperature and cardiac output. Additional parameters and interfaces to other systems are also available depending on the parameter modules employed. The Qube is intended to alert the user to alarm conditions that are reported by Spacelabs Healthcare parameter modules and/or other physiologic monitors via Flexport interfaces. These devices determine a) when an alarm condition is violated; b) the alarm priority (i.e. high, medium or low); c) alarm limits; and d) when to initiate and terminate alarm notifications. The patient monitors are also capable of displaying alarm conditions on other monitors that are on the network through the Alarm Watch feature. The Qube may also function as a generic display or computer terminal. As a generic display or terminal, the patient monitors allow network-based applications to open windows and display information on other networked monitors. The Qube is also designed to communicate with a variety of external devices such as displays, network devices, serial devices, user input devices, audio systems, and local/remote recorders. The Qube is intended for use under the direct supervision of a licensed healthcare practitioner, or by personnel trained in proper use of the equipment in a hospital environment. |
Code Information |
Serial numbers for devices distributed in the US: 1390-000102, 1390-000103, 1390-000104, 1390-000105, 1390-000106, 1390-000107, 1390-000108, 1390-000109, 1390-000110, 1390-000111, 1390-000112, 1390-000113, 1390-000114, 1390-000115, 1390-000116, 1390-000119, 1390-000120, 1390-000150, 1390-000151, 1390-000152, 1390-000153, 1390-000154, 1390-000155, 1390-000156, 1390-000157, 1390-000158, 1390-000159, 1390-000160, 1390-000161, 1390-000162, 1390-000163, 1390-000164, 1390-000165, 1390-000166, 1390-000167, 1390-000168, 1390-000169, 1390-000170, 1390-000171, 1390-000172, 1390-000173, 1390-000174, 1390-000175, 1390-000176, 1390-000177, 1390-000178, 1390-000179, 1390-000180, 1390-000181, 1390-000182, 1390-000183, 1390-000184, 1390-000185, 1390-000186, 1390-000187, 1390-000188, 1390-000189, 1390-000190, 1390-000191, 1390-000192, 1390-000193, 1390-000195, 1390-000196, 1390-000197, 1390-000198, 1390-000199, 1390-000200, 1390-000201, 1390-000202, 1390-000203, 1390-000204, 1390-000205, 1390-000206, 1390-000207, 1390-000208, 1390-000209, 1390-000210, 1390-000211, 1390-000212, 1390-000213, 1390-000214, 1390-000215, 1390-000216, 1390-000217, 1390-000218, 1390-000219, 1390-000220, 1390-000221, 1390-000222, 1390-000223, 1390-000224, 1390-000225, 1390-000226, 1390-000227, 1390-000228, 1390-000229, 1390-000230, 1390-000231, 1390-000232, 1390-000233, 1390-000234, 1390-000235, 1390-000236, 1390-000237, 1390-000238, 1390-000239, 1390-000240, 1390-000241, 1390-000242, 1390-000243, 1390-000244, 1390-000245, 1390-000246, 1390-000247, 1390-000248, 1390-000249, 1390-000250, 1390-000251, 1390-000252, 1390-000253, 1390-000254, 1390-000255, 1390-000256, 1390-000257, 1390-000258, 1390-000259, 1390-000260, 1390-000261, 1390-000262, 1390-000263, 1390-000264, 1390-000266, 1390-000267, 1390-000268, 1390-000269, 1390-000270, 1390-000271, 1390-000272, 1390-000273, 1390-000274, 1390-000275, 1390-000276, 1390-000277, 1390-000278, 1390-000279, 1390-000280, 1390-000281, 1390-000282, 1390-000283, 1390-000284, 1390-000285, 1390-000286, 1390-000287, 1390-000288, 1390-000289, 1390-000296, 1390-000297, 1390-000298, 1390-000299, 1390-000300, 1390-000301, 1390-000302, 1390-000304, 1390-000305, 1390-000307, 1390-000308, 1390-000310, 1390-000311, 1390-000312, 1390-000314, 1390-000315, 1390-000316, 1390-000317, 1390-000318, 1390-000319, 1390-000321, 1390-000322, 1390-000323, 1390-000325, 1390-000341, 1390-000342, 1390-000343, 1390-000344, 1390-000345, 1390-000346, 1390-000347, 1390-000348, 1390-000349, 1390-000350, 1390-000351, 1390-000352, 1390-000353, 1390-000354, 1390-000355, 1390-000356, 1390-000357, 1390-000358, 1390-000359, 1390-000360, 1390-000361, 1390-000362, 1390-000363, 1390-000364, 1390-000365, 1390-000366, 1390-000367, 1390-000368, 1390-000369, 1390-000370, 1390-000371, 1390-000372, 1390-000373, 1390-000374, 1390-000375, 1390-000376, 1390-000377, 1390-000378, 1390-000379, 1390-000380, 1390-000381, 1390-000382, 1390-000383, 1390-000384, 1390-000385, 1390-000386, 1390-000387, 1390-000388, 1390-000389, 1390-000390, 1390-000391, 1390-000392, 1390-000393, 1390-000394, 1390-000420, 1390-000421, 1390-000422, 1390-000423, 1390-000424, 1390-000425, 1390-000426, 1390-000427, 1390-000428, 1390-000429, 1390-000430, 1390-000431, 1390-000432, 1390-000433, 1390-000434, 1390-000435, 1390-000436, 1390-000437, 1390-000438, 1390-000439, 1390-000440, 1390-000441, 1390-000442, 1390-000443, 1390-000444, 1390-000445, 1390-000446, 1390-000447, 1390-000448, 1390-000449, 1390-000450, 1390-000451, 1390-000452, 1390-000453, 1390-000454, 1390-000455, 1390-000456, 1390-000457, 1390-000458, 1390-000459, 1390-000460, 1390-000461, 1390-000462, 1390-000463, 1390-000464, 1390-000465, 1390-000467, 1390-000468, 1390-000469, 1390-000470, 1390-000483, 1390-000564, 1390-000582, 1390-000583, 1390-000584, 1390-000585, 1390-000586, 1390-000587, 1390-000588, 1390-000589, 1390-000590, 1390-000591, 1390-000592, 1390-000593, 1390-000594, 1390-000595, 1390-000597, 1390-000598, 1390-000599, 1390-000600, 1390-000601, 1390-000602, 1390-000603, 1390-000606, 1390-000614, 1390-000615, 1390-000616, 1390-000623, 1390-000624, 1390-000625, 1390-000626, 1390-000627, 1390-000628, 1390-000629, 1390-000630, 1390-000631, 1390-000632, 1390-000633, 1390-000634, 1390-000635, 1390-000636, 1390-000637, 1390-000638, 1390-000639, 1390-000640, 1390-000641, 1390-000642, 1390-000643, 1390-000644, 1390-000645, 1390-000646, 1390-000647, 1390-000648, 1390-000649, 1390-000650, 1390-000651, 1390-000652, 1390-000653, 1390-000654, 1390-000655, 1390-000656, 1390-000657, 1390-000658, 1390-000659, 1390-000660, 1390-000661, 1390-000662, 1390-000663, 1390-000664, 1390-000665, 1390-000666, 1390-000667, 1390-000668, 1390-000669, 1390-000670, 1390-000671, 1390-000672, 1390-000673, 1390-000674, 1390-000675, 1390-000676, 1390-000677, 1390-000678, 1390-000679, 1390-000680, 1390-000681, 1390-000682, 1390-000683, 1390-000684, 1390-000685, 1390-000686, 1390-000687, 1390-000688, 1390-000698, 1390-000699, 1390-000700, 1390-000701, 1390-000702, 1390-000703, 1390-000704, 1390-000705, 1390-000706, 1390-000707, 1390-000708, 1390-000710, 1390-000712, 1390-000713, 1390-000714, 1390-000720, 1390-000721, 1390-000728, 1390-000729, 1390-000730, 1390-000731, 1390-000732, 1390-000733, 1390-000734, 1390-000735, 1390-000736, 1390-000737, 1390-000738, 1390-000742, 1390-000743, 1390-000744, 1390-000745, 1390-000746, 1390-000747, 1390-000748, 1390-000749, 1390-000750, 1390-000751, 1390-000752, 1390-000753, 1390-000754, 1390-000755, 1390-000756, 1390-000757, 1390-000758, 1390-000759, 1390-000760, 1390-000761, 1390-000762, 1390-000763, 1390-000764, 1390-000765, 1390-000766, 1390-000767, 1390-000768, 1390-000769, 1390-000775, 1390-000776, 1390-000777, 1390-000778, 1390-000779, 1390-000780, 1390-000781, 1390-000782, 1390-000792, 1390-000793, 1390-000794, 1390-000795, 1390-000796, 1390-000800, 1390-000801, 1390-000802, 1390-000803, 1390-000804, 1390-000805, 1390-000806, 1390-000807, 1390-000808, 1390-000809, 1390-000810, 1390-000811, 1390-000812, 1390-000813, 1390-000814, 1390-000815, 1390-000816, 1390-000817, 1390-000818, 1390-000819, 1390-000820, 1390-000821, 1390-000822, 1390-000823, 1390-000824, 1390-000825, 1390-000826, 1390-000827, 1390-000828, 1390-000829, 1390-000830, 1390-000831, 1390-000832, 1390-000833, 1390-000834, 1390-000835, 1390-000836, 1390-000837, 1390-000838, 1390-000839, 1390-000840, 1390-000841, 1390-000842, 1390-000843, 1390-000844, 1390-000845, 1390-000846, 1390-000847, 1390-000848, 1390-000849, 1390-000850, 1390-000851, 1390-000852, 1390-000853, 1390-000854, 1390-000855, 1390-000856, 1390-000857, 1390-000858, 1390-000859, 1390-000860, 1390-000861, 1390-000862, 1390-000863, 1390-000870, 1390-000871, 1390-000872, 1390-000887, 1390-000888, 1390-000889, 1390-000890, 1390-000891, 1390-000892, 1390-000893, 1390-000894, 1390-000895, 1390-000896, 1390-000897, 1390-000898, 1390-000899, 1390-000900, 1390-000901, 1390-000902, 1390-000903, and 1390-000904. Serial numbers for devices distributed INTERNATIONAL: On 4/16/13 the firm advised FDA that they added serial number 1390-000055 and 1390-000913. 1390-000290, 1390-000291, 1390-000337, 1390-000338, 1390-000339, 1390-000340, 1390-000607, 1390-000608, 1390-000609, 1390-000610, 1390-000689, 1390-000690, 1390-000691, 1390-000692, 1390-000693, 1390-000695, 1390-000536, 1390-000537, 1390-000709, 1390-000722, 1390-000723, 1390-000724, 1390-000725, 1390-000879, 1390-000880, 1390-000881, 1390-000905, 1390-000906, 1390-000907, 1390-000908, 1390-000909, 1390-000910
1390-000009, 1390-000010, 1390-000011, 1390-000012, 1390-000013, 1390-000014, 1390-000015, 1390-000016, 1390-000017, 1390-000018, 1390-000046, 1390-000048, 1390-000049, 1390-000051, 1390-000052, 1390-000053, 1390-000056, 1390-000057, 1390-000059, 1390-000060, 1390-000061, 1390-000062, 1390-000063, 1390-000101, 1390-000117, 1390-000118, 1390-000125, 1390-000127, 1390-000128, 1390-000138, 1390-000139, 1390-000140, 1390-000141, 1390-000142, 1390-000143, 1390-000144, 1390-000145, 1390-000146, 1390-000147, 1390-000148, 1390-000149, 1390-000326, 1390-000327, 1390-000328, 1390-000329, 1390-000330, 1390-000331, 1390-000332, 1390-000333, 1390-000334, 1390-000335, 1390-000336, 1390-000466, 1390-000535, 1390-000552, 1390-000568, 1390-000605, 1390-000611, 1390-000612, 1390-000613, 1390-000715, 1390-000716, 1390-000739, 1390-000740, 1390-000741, 1390-000770, 1390-000787, 1390-000788, 1390-000789, 1390-000790, 1390-000791, 1390-000797, 1390-000798, 1390-000876, 1390-000877, 1390-000878, 1390-1390-000047, 1390-00002, 1390-000618, 1390-000619, 1390-000620, 1390-000621, 1390-000622, 1390-000509, 1390-000510, 1390-000511, 1390-000512, 1390-000513, 1390-000514, 1390-000553, 1390-000554, 1390-000121, 1390-000524, 1390-000771, 1390-000772, 1390-000292, 1390-000293, 1390-000294, 1390-000295, 1390-000911, 1390-000555, 1390-000556, 1390-000557, 1390-000558, 1390-000559, 1390-000560, 1390-000561, 1390-000562, 1390-000563, 1390-000565, 1390-000566, 1390-000567, 1390-000569, 1390-000570, 1390-000571, 1390-000572, 1390-000573, 1390-000574, 1390-000575, 1390-000576, 1390-000577, 1390-000578, 1390-000579, 1390-000580, 1390-000581
1390-000395, 1390-000396, 1390-000397, 1390-000398, 1390-000399, 1390-000400, 1390-000401, 1390-000402, 1390-000403, 1390-000404, 1390-000405, 1390-000406, 1390-000407, 1390-000408, 1390-000409, 1390-000410, 1390-000411, 1390-000412, 1390-000413, 1390-000414, 1390-000415, 1390-000416, 1390-000417, 1390-000418, 1390-000419, 1390-000604, 1390-000130, 1390-000131, 1390-000132, 1390-000133, 1390-000135, 1390-000136, 1390-000137, 1390-000493, 1390-000496, 1390-000497, 1390-000498, 1390-000499, 1390-000500, 1390-000501, 1390-000502, 1390-000525, 1390-000526, 1390-000527, 1390-000528, 1390-000529, 1390-000530, 1390-000531, 1390-000532, 1390-000533, 1390-000534, 1390-000596, 1390-000617, 1390-000799, and 1390-000696. |
Recalling Firm/ Manufacturer |
Spacelabs Healthcare, Llc 5150 220th Ave Se Issaquah WA 98029-6834
|
For Additional Information Contact |
Global Tech Support 800-522-7025
|
Manufacturer Reason for Recall |
Spacelabs Healthcare has learned through several
reports that the input circuit may fail and the monitor
will go on battery mode. The monitor will function
normally until the batteries become depleted and then
the monitor will shut down. In addition the monitor may
fail to turn on due to an internal short.
|
FDA Determined Cause 2 |
Nonconforming Material/Component |
Action |
Spacelabs Healthcare sent an "URGENT-MEDICAL DEVICE CORRECTION" letter dated October 19, 2012. The letter identified the product, problem, and actions to be taken by the customers. Contact Spacelabs Healthcare at 1-800522-7025 option 2 for questions regarding this notice. |
Quantity in Commerce |
742 total devices distributed; 530 with the U.S. and 212 Internationally. |
Distribution |
Worldwide Distribution.On 4/16/13 the firm advised FDA that they added serial number 1390-000055 and 1390-000913. The two additional serial numbers were added because they were provided as replacements to a French customer whose devices had been returned to support the firm's investigation. The customer had already received the Urgent Medical Device correction, customer letter and required Spacelabs to provide these devices with the understanding that they would require the field corrective action |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database |
510(K)s with Product Code = MHX and Original Applicant = SPACELABS MEDICAL, INC.
|
|
|
|