| Class 2 Device Recall Mindray DS USA Inc. | |
Date Initiated by Firm | August 11, 2010 |
Date Posted | December 11, 2012 |
Recall Status1 |
Terminated 3 on August 02, 2013 |
Recall Number | Z-0513-2013 |
Recall Event ID |
63614 |
510(K)Number | K092449 |
Product Classification |
Monitor, physiological, patient(with arrhythmia detection or alarms) - Product Code MHX
|
Product | DPM 6/7 Patient Monitor
Manufactured in China for :
Mindray DS USA, Inc.
800 MacArthur Blvd.
Mahwah, NJ 07430 USA.
Intended to be used for monitoring, displaying, reviewing, storing and alarming of multiple physiological parameters including ECG (3-lead or 5-lead or 12-lead selectable). arrhythmia detection, ST Segment analysis, heart rate (HR) , respiration rate (RESP), temperature (TEMP), pulse oxygen saturation (SpO2), pulse rate (PR), non-invasive blood pressure (NIBP), invasive blood pressure (IBP), carbon dioxide (CO2), anesthetic gas (AG),impedance cardiograph (ICG), bispectral index (BIS), cardiac output (C.O.) and respiration mechanics (RM). |
Code Information |
K092449 DPM 6- p/n 6802F-PA00001 DPM 7 - p/n 6800F-PA00001 |
Recalling Firm/ Manufacturer |
Mindray DS USA, Inc. d.b.a. Mindray North America 800 MacArthur Blvd. Mahwah NJ 07430
|
For Additional Information Contact | Ms. Diane Arpino 201-995-8407 |
Manufacturer Reason for Recall | There is an issue with the DPM 6 and DPM 7 monitors where the following features may be unavailable : Full Disclosure, Drug, Hemodynamic Calculations, Renal Calculations, Oxygenation Calculations, and Ventilation Calculations. |
FDA Determined Cause 2 | Device Design |
Action | Mindray sent a "PRODUCT CORRECTIVE ACTION" letter dated August 11, 2010 to all affected customers. The letter identified the product, problem, and actions to be taken by the customers. Customers should contact their Mindray Service Representative at 1-800-288-2121 for assistance with this issue. |
Quantity in Commerce | 124 units |
Distribution | Nationwide Distribution-USA (nationwide) including the states of WA, IL, FL, MN, SC, NV, MI, TN, PA, LA, NE, TX, and IA. |
Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database | 510(K)s with Product Code = MHX
|
|
|
|