| Class 2 Device Recall Trilogy F/M Acetabular Shell with UniHole | |
Date Initiated by Firm | September 14, 2012 |
Date Posted | February 04, 2013 |
Recall Status1 |
Terminated 3 on June 20, 2014 |
Recall Number | Z-0771-2013 |
Recall Event ID |
63531 |
510(K)Number | K934765 |
Product Classification |
Prosthesis, hip, semi-constrained, metal/polymer, porous uncemented - Product Code LPH
|
Product | Trilogy F/M Acetabular Shell with Uni-Hole.
The Trilogy Acetabular System is indicated for either cemented or uncemented use in individuals undergoing surgery for rehabilitating hip damage. |
Code Information |
61282267 61633887 61260645 61237937 61500763 61397941 61584769 61199849 61429445 61412089 00369745 |
Recalling Firm/ Manufacturer |
Zimmer, Inc. 345 E Main St Warsaw IN 46580-2746
|
For Additional Information Contact | Jaime L. Weeks 574-372-4807 |
Manufacturer Reason for Recall | Some units of Trilogy shells manufactured since March 2009 may be below the lower range of the pore size specification (as low as 28% vs. 30% porosity specification). On 10/31/2012, the firm expanded the recall action to include lots of Uni-Hole Trilogy Acetabular System shells. |
FDA Determined Cause 2 | Process control |
Action | Zimmer sent an "URGENT MEDICAL DEVICE NOTIFICATION" letter dated September 20, 2012 to all affected customers. The letter identified the product, problem, and actions to be taken by the customers.The firm expanded the scope of the recall in January 2013 to include devices that use a secondary mean of fixation. New recall letters will be submitted to distributors and surgeons. |
Quantity in Commerce | 275,952 units |
Distribution | Worldwide Distribution-USA (nationwide) and the countries of Denmark, Germany, Greece, Hungary, Ireland, Italy, Netherlands, Spain, Sweden, UK, Poland, France, Australia, Chile, Canada, India, Japan, New Zealand, Peru, South Africa, Colombia, Ukraine, and UAE. |
Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database | 510(K)s with Product Code = LPH
|
|
|
|