Date Initiated by Firm |
March 16, 2011 |
Create Date |
November 04, 2016 |
Recall Status1 |
Terminated 3 on November 28, 2016 |
Recall Number |
Z-0361-2017 |
Recall Event ID |
64005 |
510(K)Number |
K110901 K041550
|
Product Classification |
Pump, infusion - Product Code FRN
|
Product |
Symbiq One Channel Infuser; an Rx medical device infusion pump used to administer I.V. fluids. List number 16026. Note that Hospira retired this pump June 30, 2015. The Hospira Symbiq Infusion System is intended for the delivery of fluids, solutions, drugs, agents, nutritionals, electrolytes, blood, and blood products via parenteral, enteral, intravenous, intra-arterial, subcutaneous, epidural, or irrigation routes of administration. |
Code Information |
14339017 14339063 14339071 14339076 14339078 14339084 14339086 14339087 14339091 14339095 14339097 14339104 14339105 14339108 14339114 14339115 14339119 14339125 14339130 14339140 14339142 14339145 14339146 14339149 14339158 14339171 14339172 14339192 14339195 14339200 14339271 14339283 14339304 14339308 14339322 14339327 14339328 14339344 14339358 14339359 14339362 14339367 14339369 14339370 14339377 14339380 14339384 14339389 14339391 14339392 14339396 14339407 14339411 14339413 14339414 14339415 14339422 14339424 14339426 14339504 14339517 14339519 14339524 14339529 14339534 14339545 14339569 14339572 14339664 14339669 14339681 14339723 14339789 14339849 14339905 14339921 14339932 14339933 14339985 14339987 14339989 14339990 14340000 14340002 14340004 14340006 14340008 14340017 14340027 14340038 14340039 14340041 14340048 14340053 14340058 14340060 14340063 14340065 14340067 14340071 14340072 14340077 14340078 14340080 14340084 14340088 14340089 14340103 14340105 14340106 14340123 14340125 14340126 14340128 14340138 14340150 14340153 14340167 14340170 14340180 14340186 14340192 14340205 14340214 14340217 14340223 14340232 14340255 14340272 14340273 14340277 14340287 14340377 14340627 14340659 14340709 14340722 14340747 14340760 14340770 14340816 14340845 14340865 14340913 14340935 14341101 14341120 14341141 14341158 14341160 14341246 14341290 14341342 14341363 14341402 14341414 14341461 14341469 14341481 14341502 14341521 14341523 14341553 14341579 14341681 14341727 14341746 14341752 14341763 14341825 14341879 14341889 14341924 14341944 14341969 14341986 14342018 14342129 14342146 14342153 14342161 14342177 14342183 14342197 14342253 14342275 14342322 14342375 14342376 14342384 14342400 14342491 14342493 14342535 14342581 14342619 14342657 14342664 14342668 14342678 14342694 14342752 14342857 14342886 14342907 14342911 14342926 14342939 14342958 14342986 14343051 14343068 14343091 14343149 14343168 14343194 14343254 14343258 14343328 14343331 14343359 14343407 14343429 14343441 14343499 14343558 14343585 14343588 14343636 14343643 14343651 14343657 14343703 14343781 14343823 14343825 14343876 14343896 14343919 14343945 14344020 14344045 14344056 14344100 14344256 14344258 14344261 14344296 14344305 14344313 14344326 14344328 14344349 14344360 14344377 14344394 14344396 14344419 14344436 14344520 14344523 14344552 14344563 14344572 14344608 14344633 14344640 14344643 14344645 14344662 14344669 14344701 14344718 14344734 14344738 14344740 14344770 14344867 14344908 14344918 14344920 14344928 14344936 14344993 14345052 14345114 14345149 14345248 14345362 14345363 14345372 14345437 14345468 14345475 14345777 14345812 14345815 14345832 14345841 14345845 14345847 14345949 14346046 14346065 14346105 14346107 14346120 14346146 14346174 14346289 14346299 14346372 14346480 14346608 14346638 14346693 14346741 14346772 14346784 14346851 14346876 14347015 14347038 14347061 14347066 14347067 14347068 14347069 14347070 14347071 14347072 14347074 14347075 14347076 14347077 14347078 14347079 14347080 14347101 14347109 14347112 14347113 14347115 14347116 14347118 14347120 14347122 14347124 14347125 14347126 14347130 14347132 14347134 14347136 14347138 14347139 14347140 14347141 14347142 14347143 14347144 14347145 14347146 14347149 14347150 14347151 14347152 14347153 14347154 14347155 14347156 14347157 14347160 14347161 14347162 14347163 14347164 14347165 14347167 14347168 14347169 14347170 14347171 14347172 14347173 14347175 14347176 14347179 14347180 14347182 14347184 14347185 14347186 14347187 14347188 14347190 14347191 14347192 14347194 14347197 14347198 14347199 14347200 14347201 14347202 14347205 14347206 14347207 14347208 14347210 14347211 14347212 14347213 14347215 14347216 14347218 14347221 14347222 14347223 14347224 14347225 14347229 14347230 14347231 14347232 14347233 14347234 14347238 14347240 14347241 14347242 14347244 14347246 14347247 14347251 14347252 14347258 14347260 14347265 14347266 14347267 14347268 14347270 14347272 14347276 14347278 14347280 14347281 14347283 14347286 14347287 14347288 14347291 14347293 14347295 14347296 14347298 14347301 14347302 14347303 14347304 14347305 14347307 14347311 14347312 14347317 14347318 14347319 14347320 14347321 14347323 14347325 14347326 14347328 14347330 14347331 14347333 14347338 14347339 14347340 14347341 14347342 14347343 14347344 14347345 14347346 14347347 14347351 14347353 14347354 14347355 14347356 14347357 14347358 14347360 14347361 14347362 14347363 14347364 14347365 14347368 14347369 14347370 14347372 14347374 14347376 14347377 14347378 14347379 14347381 14347382 14347384 14347385 14347388 14347389 14347390 14347392 14347395 14347396 14347397 14347398 14347401 14347404 14347405 14347407 14347408 14347409 14347410 14347411 14347412 14347413 14347414 14347418 14347419 14347420 14347422 14347423 14347424 14347427 14347428 14347430 14347431 14347432 14347433 14347434 14347436 14347437 14347442 14347445 14347446 14347447 14347448 14347450 14347451 14347452 14347454 14347456 14347457 14347459 14347460 14347463 14347464 14347465 14347466 14347467 14347468 14347470 14347473 14347474 14347477 14347478 14347479 14347480 14347481 14347482 14347483 14347484 14347485 14347486 14347487 14347489 14347491 14347493 14347497 14347498 14347499 14347501 14347502 14347503 14347505 14347509 14347510 14347515 14347516 14347519 14347523 14347524 14347527 14347532 14347533 14347535 14347537 14347538 14347540 14347541 14347543 14347544 14347545 14347547 14347548 14347551 14347557 14347561 14347563 14347564 14347565 14347566 14347567 14347568 14347569 14347570 14347571 14347572 14347573 14347574 14347576 14347578 14347579 14347580 14347581 14347587 14347588 14347589 14347592 14347594 14347596 14347597 14347599 14347600 14347601 14347603 14347606 14347607 14347609 14347613 14347614 14347617 14347618 14347620 14347624 14347627 14347634 14347636 14347637 14347650 14347653 14347654 14347655 14347662 14347663 14347664 14347668 14347670 14347681 14347683 14347688 14347690 14347698 14347700 14347705 14347707 14347709 14347711 14347712 14347713 14347714 14347715 14347716 14347717 14347718 14347719 14347720 14347721 14347724 14347725 14347726 14347727 14347729 14347730 14347731 14347732 14347733 14347734 14347736 14347737 14347738 14347739 14347740 14347741 14347742 14347743 14347744 14347745 14347746 14347748 14347749 14347750 14347751 14347752 14347753 14347754 14347755 14347756 14347758 14347759 14347760 14347761 14347762 14347763 14347764 14347765 14347767 14347768 14347769 14347770 14347771 14347772 14347773 14347774 14347775 14347776 14347777 14347778 14347779 14347780 14347781 14347783 14347784 14347785 14347787 14347790 14347791 14347793 14347795 14347796 14347797 14347798 14347799 14347801 14347803 14347804 14347805 14347807 14347808 14347809 14347810 14347811 14347812 14347814 14347815 14347816 14347817 14347818 14347820 14347824 14347826 14347827 14347828 14347829 14347831 14347832 14347833 14347834 14347835 14347836 14347837 14347838 14347839 14347840 14347841 14347842 14347844 14347845 14347848 14347851 14347852 14347853 14347854 14347856 14347857 14347858 14347859 14347860 14347861 14347862 14347864 14347865 14347866 14347871 14347872 14347873 14347874 14347875 14347877 14347879 14347880 14347881 14347882 14347884 14347885 14347886 14347887 14347892 14347894 14347895 14347896 14347897 14347898 14347899 14347901 14347902 14347903 14347904 14347905 14347906 14347908 14347910 14347913 14347914 14347915 14347916 14347917 14347919 14347920 14347921 14347924 14347925 14347926 14347927 14347929 14347931 14347933 14347934 14347935 14347937 14347938 14347940 14347942 14347944 14347945 14347946 14347947 14347951 14347952 14347953 14347954 14347955 14347956 14347957 14347958 14347961 14347962 14347963 14347964 14347965 14347968 14347969 14347970 14347971 14347973 14347974 14347975 14347978 14347979 14347981 14347982 14347983 14347988 14347990 14347992 14347993 14347994 14347996 14347997 14347998 14347999 14348000 14348002 14348003 14348004 14348006 14348007 14348008 14348009 14348010 14348011 14348013 14348014 14348015 14348016 14348018 14348019 14348020 14348021 14348022 14348023 14348024 14348025 14348026 14348027 14348028 14348030 14348031 14348032 14348033 14348035 14348036 14348037 14348040 14348041 14348044 14348045 14348046 14348047 14348048 14348049 14348050 14348051 14348052 14348053 14348054 14348055 14348056 14348057 14348058 14348059 14348060 14348061 14348062 14348063 14348064 14348066 14348067 14348069 14348071 14348072 14348073 14348074 14348075 14348076 14348077 14348079 14348080 14348082 14348083 14348085 14348086 14348087 14348088 14348089 14348090 14348091 14348092 14348093 14348094 14348095 14348097 14348098 14348099 14348100 14348101 14348102 14348103 14348104 14348105 14348108 14348109 14348111 14348112 14348113 14348114 14348115 14348116 14348117 14348118 14348119 14348120 14348121 14348122 14348123 14348124 14348125 14348128 14348130 14348131 14348132 14348133 14348134 14348135 14348137 14348142 14348144 14348145 14348146 14348147 14348148 14348149 14348151 14348152 14348153 14348154 14348156 14348157 14348158 14348159 14348160 14348161 14348162 14348163 14348164 14348166 14348169 14348170 14348171 14348172 14348173 14348174 14348175 14348176 14348177 14348179 14348180 14348181 14348184 14348185 14348191 14348192 14348193 14348195 14348197 14348198 14348200 14348201 14348202 14348203 14348204 14348207 14348208 14348209 14348210 14348211 14348212 14348213 14348214 14348215 14348216 14348217 14348219 14348220 14348221 14348222 14348223 14348225 14348226 14348229 14348231 14348232 14348235 14348236 14348237 14348238 14348240 14348241 14348243 14348244 14348245 14348247 14348251 14348252 14348253 14348255 14348257 14348258 14348259 14348260 14348261 14348263 14348265 14348266 14348268 14348269 14348270 14348271 14348272 14348273 14348274 14348275 14348277 14348278 14348279 14348280 14348281 14348282 14348283 14348284 14348285 14348287 14348289 14348291 14348293 14348294 14348295 14348296 14348298 14348299 14348300 14348301 14348303 14348304 14348306 14348307 14348310 14348311 14348312 14348313 14348314 14348315 14348317 14348318 14348319 14348320 14348321 14348322 14348323 14348324 14348325 14348326 14348327 14348328 14348329 14348330 14348332 14348333 14348334 14348335 14348336 14348337 14348338 14348339 14348341 14348343 14348345 14348346 14348347 14348348 14348349 14348350 14348351 14348353 14348354 14348355 14348356 14348357 14348358 14348359 14348364 14348365 14348366 14348367 14348368 14348369 14348370 14348372 14348373 14348374 14348375 14348377 14348378 14348379 14348380 14348381 14348382 14348383 14348384 14348385 14348386 14348387 14348388 14348389 14348390 14348391 14348392 14348393 14348396 14348397 14348398 14348399 14348400 14348401 14348402 14348403 14348404 14348406 14348408 14348409 14348410 14348411 14348412 14348413 14348414 14348415 14348417 14348418 14348421 14348422 14348423 14348426 14348428 14348429 14348430 14348432 14348433 14348435 14348438 14348439 14348440 14348441 14348442 14348443 14348445 14348446 14348447 14348448 14348449 14348450 14348451 14348452 14348453 14348454 14348455 14348456 14348457 14348458 14348459 14348460 14348461 14348462 14348463 14348464 14348466 14348467 14348468 14348470 14348471 14348472 14348473 14348474 14348475 14348476 14348478 14348479 14348480 14348482 14348483 14348484 14348485 14348486 14348487 14348488 14348489 14348490 14348491 14348492 14348494 14348495 14348496 14348497 14348498 14348499 14348500 14348502 14348503 14348504 14348505 14348506 14348507 14348508 14348509 14348510 14348511 14348512 14348514 14348515 14348516 14348517 14348518 14348519 14348520 14348521 14348523 14348524 14348525 14348526 14348527 14348529 14348531 14348532 14348533 14348534 14348535 14348536 14348539 14348540 14348541 14348542 14348545 14348546 14348547 14348548 14348549 14348550 14348551 14348552 14348553 14348554 14348556 14348557 14348558 14348559 14348560 14348561 14348562 14348563 14348568 14348569 14348570 14348572 14348573 14348575 14348576 14348577 14348578 14348580 14348581 14348582 14348584 14348585 14348586 14348587 14348588 14348589 14348590 14348591 14348592 14348593 14348594 14348595 14348596 14348597 14348598 14348599 14348601 14348602 14348604 14348605 14348608 14348609 14348610 14348611 14348614 14348616 14348617 14348618 14348619 14348620 14348621 14348622 14348623 14348624 14348625 14348626 14348627 14348628 14348629 14348630 14348631 14348632 14348633 14348634 14348635 14348636 14348637 14348639 14348640 14348641 14348642 14348643 14348644 14348645 14348646 14348647 14348649 14348651 14348653 14348654 14348655 14348656 14348657 14348660 14348661 14348662 14348664 14348665 14348666 14348667 14348668 14348670 14348672 14348673 14348675 14348676 14348678 14348679 14348682 14348683 14348686 14348687 14348688 14348691 14348692 14348693 14348694 14348695 14348696 14348697 14348698 14348702 14348704 14348705 14348707 14348708 14348709 14348711 14348713 14348714 14348716 14348719 14348720 14348721 14348722 14348723 14348724 14348725 14348726 14348727 14348728 14348731 14348732 14348733 14348735 14348736 14348738 14348739 14348741 14348742 14348744 14348746 14348747 14348748 14348749 14348750 14348751 14348754 14348755 14348758 14348760 14348761 14348762 14348763 14348764 14348765 14348766 14348769 14348770 14348772 14348776 14348778 14348779 14348780 14348781 14348782 14348783 14348786 14348789 14348790 14348791 14348792 14348796 14348799 14348800 14348801 14348802 14348803 14348805 14348807 14348808 14348810 14348811 14348813 14348814 14348817 14348821 14348822 14348823 14348824 14348825 14348826 14348827 14348828 14348829 14348830 14348831 14348832 14348833 14348836 14348837 14348838 14348839 14348840 14348841 14348842 14348843 14348844 14348845 14348846 14348848 14348849 14348850 14348851 14348852 14348854 14348855 14348856 14348857 14348858 14348860 14348862 14348863 14348864 14348865 14348866 14348867 14348868 14348869 14348870 14348872 14348873 14348875 14348876 14348877 14348879 14348880 14348881 14348882 14348884 14348885 14348887 14348888 14348890 14348891 14348893 14348894 14348895 14348896 14348899 14348901 14348902 14348903 14348905 14348907 14348910 14348912 14348913 14348914 14348915 14348917 14348918 14348919 14348920 14348921 14348922 14348923 14348924 14348925 14348926 14348927 14348928 14348929 14348930 14348931 14348932 14348933 14348935 14348936 14348941 14348942 14348943 14348944 14348946 14348948 14348949 14348950 14348951 14348952 14348953 14348954 14348955 14348956 14348958 14348959 14348960 14348961 14348962 14348963 14348965 14348967 14348968 14348969 14348972 14348974 14348977 14348978 14348979 14348980 14348981 14348983 14348984 14348985 14348987 14348988 14348989 14348992 14348993 14348995 14348998 14348999 14349000 14349001 14349002 14349003 14349004 14349005 14349006 14349007 14349008 14349011 14349012 14349013 14349014 14349016 14349018 14349019 14349020 14349021 14349022 14349023 14349025 14349027 14349028 14349031 14349032 14349033 14349034 14349035 14349038 14349039 14349040 14349042 14349043 14349044 14349045 14349046 14349047 14349048 14349049 14349051 14349052 14349053 14349054 14349056 14349058 14349059 14349060 14349065 14349066 14349068 14349069 14349070 14349072 14349073 14349074 14349075 14349076 14349077 14349078 14349081 14349082 14349084 14349085 14349087 14349088 14349089 14349092 14349096 14349097 14349098 14349099 14349100 14349102 14349103 14349104 14349108 14349109 14349113 14349114 14349116 14349117 14349118 14349119 14349120 14349123 14349124 14349125 14349126 14349127 14349128 14349129 14349133 14349135 14349136 14349137 14349138 14349140 14349141 14349142 14349143 14349144 14349145 14349146 14349149 14349150 14349151 14349152 14349153 14349154 14349156 14349157 14349158 14349159 14349160 14349161 14349163 14349164 14349165 14349166 14349167 14349168 14349170 14349172 14349173 14349174 14349175 14349176 14349178 14349179 14349180 14349182 14349183 14349184 14349187 14349188 14349189 14349190 14349192 14349193 14349194 14349195 14349196 14349197 14349199 14349200 14349201 14349202 14349205 14349207 14349208 14349209 14349210 14349211 14349212 14349214 14349216 14349217 14349218 14349219 14349221 14349222 14349223 14349224 14349226 14349227 14349228 14349229 14349231 14349232 14349237 14349242 14349245 14349246 14349247 14349248 14349250 14349251 14349254 14349255 14349258 14349259 14349260 14349263 14349264 14349265 14349266 14349268 14349269 14349270 14349271 14349272 14349273 14349274 14349275 14349276 14349277 14349283 14349284 14349285 14349286 14349287 14349288 14349290 14349291 14349294 14349295 14349296 14349298 14349300 14349301 14349303 14349304 14349305 14349306 14349307 14349308 14349310 14349311 14349312 14349313 14349314 14349315 14349316 14349320 14349323 14349324 14349325 14349326 14349327 14349328 14349329 14349330 14349331 14349332 14349333 14349334 14349336 14349338 14349340 14349342 14349344 14349345 14349347 14349348 14349349 14349351 14349352 14349353 14349355 14349357 14349358 14349360 14349361 14349362 14349367 14349368 14349369 14349370 14349373 14349376 14349379 14349380 14349381 14349382 14349383 14349384 14349385 14349386 14349388 14349389 14349390 14349391 14349392 14349394 14349395 14349396 14349397 14349400 14349401 14349404 14349405 14349407 14349408 14349409 14349410 14349411 14349413 14349414 14349415 14349416 14349419 14349420 14349422 14349423 14349426 14349428 14349429 14349430 14349432 14349433 14349435 14349436 14349437 14349439 14349441 14349442 14349443 14349444 14349445 14349448 14349450 14349451 14349455 14349457 14349458 14349459 14349460 14349461 14349462 14349464 14349465 14349467 14349469 14349470 14349471 14349472 14349473 14349474 14349477 14349478 14349480 14349481 14349484 14349486 14349490 14349492 14349495 14349496 14349498 14349502 14349503 14349506 14349508 14349511 14349513 14349517 14349518 14349519 14349520 14349524 14349525 14349526 14349527 14349529 14349530 14349532 14349534 14349535 14349537 14349541 14349544 14349546 14349547 14349548 14349556 14349695 14349696 14349698 14349700 14349701 14349703 14349704 14349705 14349706 14349707 14349708 14349709 14349711 14349713 14349714 14349715 14349716 14349717 14349718 14349719 14349720 14349721 14349722 14349723 14349724 14349726 14349728 14349729 14349730 14349731 14349733 14349734 14349735 14349737 14349738 14349739 14349740 14349742 14349743 14349744 14349745 14349746 14349747 14349749 14349750 14349751 14349752 14349754 14349755 14349757 14349758 14349759 14349760 14349766 14349768 14349769 14349770 14349771 14349776 14349779 14349781 14349785 14349786 14349789 14349790 14349791 14349792 14349794 14349795 14349796 14349798 14349799 14349800 14349801 14349802 14349805 14349808 14349809 14349814 14349816 14349817 14349819 14349820 14349823 14349824 14349825 14349828 14349835 14349836 14349837 14349838 14349839 14349840 14349841 14349845 14349851 14349852 14349853 14349854 14349856 14349857 14349859 14349862 14349863 14349864 14349871 14349872 14349873 14349875 14349877 14349878 14349879 14349881 14349882 14349885 14349886 14349887 14349888 14349889 14349891 14349892 14349895 14349898 14349900 14349901 14349903 14349904 14349905 14349909 14349911 14349912 14349913 14349915 14349916 14349917 14349920 14349924 14349925 14349927 14349928 14349930 14349934 14349936 14349938 14349940 14349943 14349944 14349948 14349949 14349950 14349951 14349955 14349957 14349961 14349962 14349963 14349964 14349965 14349966 14349967 14349968 14349969 14349972 14349977 14349978 14349979 14349983 14349984 14349985 14349986 14349987 14349988 14349989 14349990 14349992 14349993 14349995 14349996 14349997 14350000 14350001 14350002 14350004 14350008 14350010 14350011 14350013 14350014 14350015 14350016 14350017 14350018 14350019 14350020 14350023 14350024 14350025 14350026 14350027 14350029 14350030 14350032 14350034 14350036 14350038 14350041 14350042 14350043 14350044 14350045 14350050 14350052 14350057 14350060 14350061 14350065 14350066 14350068 14350069 14350071 14350073 14350074 14350078 14350079 14350087 14350088 14350090 14350091 14350093 14350094 14350095 14350096 14350097 14350098 14350099 14350100 14350101 14350102 14350104 14350105 14350107 14350109 14350110 14350112 14350115 14350117 14350119 14350121 14350122 14350123 14350125 14350126 14350129 14350131 14350133 14350134 14350135 14350136 14350137 14350138 14350139 14350140 14350141 14350142 14350143 14350144 14350145 14350146 14350150 14350152 14350153 14350159 14350160 14350167 14350169 14350171 14350172 14350174 14350175 14350178 14350179 14350180 14350181 14350182 14350186 14350187 14350188 14350190 14350193 14350195 14350196 14350199 14350200 14350201 14350202 14350203 14350204 14350205 14350206 14350208 14350209 14350210 14350211 14350213 14350214 14350216 14350219 14350220 14350222 14350223 14350224 14350225 14350228 14350231 14350234 14350237 14350239 14350241 14350244 14350246 14350249 14350250 14350252 14350255 14350259 14350260 14350265 14350266 14350268 14350270 14350272 14350278 14350282 14350283 14350285 14350287 14350291 14350303 14350305 14350306 14350311 14350312 14350316 14350317 14350319 14350323 14350327 14350329 14350330 14350336 14350337 14350338 14350341 14350343 14350344 14350345 14350351 14350352 14350355 14350361 14350362 14350368 14350369 14350372 14350384 14350385 14350387 14350389 14350390 14350393 14350398 14350399 14350400 14350404 14350407 14350408 14350410 14350411 14350412 14350416 14350417 14350423 14350425 14350426 14350427 14350428 14350432 14350435 14350436 14350437 14350438 14350441 14350442 14350445 14350446 14350453 14350456 14350458 14350459 14350463 14350465 14350466 14350468 14350469 14350475 14350476 14350477 14350478 14350479 14350480 14350481 14350483 14350485 14350489 14350490 14350492 14350496 14350498 14350499 14350500 14350501 14350502 14350504 14350505 14350510 14350512 14350514 14350518 14350521 14350524 14350525 14350528 14350529 14350532 14350536 14350545 14350547 14350551 14350553 14350558 14350560 14350564 14350565 14350566 14350567 14350568 14350571 14350575 14350577 14350579 14350582 14350586 14350592 14350597 14350598 14350599 14350601 14350602 14350605 14350606 14350610 14350612 14350613 14350615 14350619 14350620 14350621 14350625 14350628 14350631 14350632 14350635 14350637 14350642 14350647 14350649 14350652 14350653 14350663 14350664 14350665 14350669 14350674 14350675 14350676 14350680 14350681 14350682 14350683 14350686 14350688 14350689 14350691 14350693 14350694 14350699 14350700 14350702 14350704 14350706 14350707 14350716 14350717 14350718 14350720 14350730 14350733 14350734 14350736 14350738 14350739 14350740 14350742 14350743 14350751 14350753 14350754 14350756 14350758 14350761 14350763 14350764 14350768 14350769 14350770 14350772 14350773 14350777 14350778 14350779 14350781 14350782 14350783 14350785 14350790 14350792 14350793 14350794 14350795 14350796 14350803 14350805 14350806 14350808 14350809 14350810 14350811 14350813 14350815 14350817 14350819 14350825 14350829 14350830 14350831 14350833 14350838 14350842 14350850 14350852 14350853 14350854 14350859 14350862 14350866 14350867 14350868 14350869 14350870 14350873 14350876 14350878 14350880 14350881 14350883 14350885 14350886 14350887 14350888 14350889 14350890 14350892 14350893 14350894 14350896 14350897 14350899 14350901 14350902 14350903 14350904 14350906 14350907 14350908 14350910 14350911 14350914 14350915 14350917 14350920 14350923 14350924 14350925 14350926 14350927 14350928 14350929 14350931 14350932 14350933 14350934 14350935 14350936 14350940 14350941 14350942 14350944 14350946 14350948 14350949 14350950 14350951 14350952 14350954 14350956 14350957 14350960 14350961 14350964 14350965 14350971 14350974 14350977 14350979 14350980 14350982 14350983 14350984 14350986 14350989 14350990 14350991 14350992 14350994 14350995 14350996 14350997 14351000 14351001 14351003 14351004 14351005 14351007 14351009 14351010 14351015 14351016 14351017 14351018 14351020 14351021 14351022 14351023 14351024 14351025 14351026 14351028 14351030 14351033 14351034 14351038 14351040 14351044 14351045 14351048 14351053 14351054 14351056 14351058 14351061 14351062 14351064 14351065 14351066 14351067 14351070 14351072 14351073 14351076 14351078 14351080 14351081 14351083 14351086 14351088 14351090 14351092 14351093 14351094 14351097 14351099 14351101 14351102 14351105 14351106 14351113 14351116 14351118 14351123 14351124 14351125 14351132 14351133 14351135 14351139 14351141 14351145 14351150 14351155 14351156 14351157 14351161 14351162 14351163 14351164 14351167 14351168 14351172 14351178 14351179 14351180 14351188 14351195 14351202 14351206 14351207 14351209 14351211 14351214 14351215 14351216 14351217 14351219 14351220 14351224 14351225 14351227 14351228 14351230 14351233 14351234 14351235 14351236 14351239 14351240 14351241 14351242 14351246 14351253 14351254 14351255 14351256 14351257 14351258 14351260 14351262 14351264 14351267 14351270 14351272 14351273 14351275 14351280 14351281 14351284 14351285 14351287 14351288 14351289 14351290 14351294 14351296 14351298 14351300 14351301 14351303 14351305 14351306 14351307 14351308 14351309 14351311 14351312 14351315 14351316 14351317 14351318 14351319 14351320 14351323 14351325 14351326 14351329 14351330 14351331 14351332 14351333 14351336 14351339 14351340 14351341 14351344 14351345 14351346 14351347 14351350 14351353 14351354 14351355 14351357 14351358 14351361 14351363 14351364 14351365 14351366 14351367 14351369 14351370 14351372 14351373 14351375 14351380 14351381 14351382 14351386 14351388 14351390 14351393 14351394 14351396 14351399 14351400 14351401 14351404 14351405 14351406 14351409 14351410 14351412 14351414 14351415 14351416 14351417 14351421 14351423 14351424 14351425 14351430 14351432 14351433 14351437 14351441 14351442 14351443 14351444 14351445 14351446 14351447 14351449 14351450 14351451 14351452 14351453 14351455 14351456 14351457 14351459 14351460 14351461 14351464 14351465 14351466 14351467 14351468 14351473 14351476 14351477 14351480 14351481 14351482 14351483 14351485 14351488 14351489 14351491 14351493 14351496 14351498 14351500 14351501 14351502 14351504 14351505 14351506 14351507 14351509 14351512 14351513 14351514 14351517 14351518 14351519 14351522 14351525 14351530 14351537 14351540 14351541 14351542 14351544 14351545 14351547 14351548 14351553 14351554 14351555 14351557 14351558 14351559 14351560 14351562 14351563 14351564 14351565 14351566 14351567 14351568 14351569 14351570 14351571 14351572 14351573 14351578 14351581 14351583 14351584 14351585 14351586 14351588 14351589 14351591 14351592 14351593 14351595 14351596 14351598 14351600 14351602 14351603 14351604 14351605 14351609 14351610 14351614 14351617 14351618 14351619 14351620 14351621 14351622 14351623 14351626 14351627 14351631 14351632 14351633 14351634 14351635 14351637 14351638 14351639 14351640 14351643 14351644 14351645 14351648 14351652 14351653 14351654 14351655 14351656 14351660 14351662 14351665 14351667 14351670 14351671 14351672 14351675 14351682 14351683 14351685 14351689 14351690 14351693 14351696 14351697 14351698 14351699 14351703 14351707 14351711 14351712 14351713 14351715 14351717 14351718 14351719 14351720 14351724 14351725 14351727 14351728 14351731 14351736 14351737 14351739 14351740 14351741 14351742 14351744 14351746 14351747 14351748 14351749 14351750 14351751 14351752 14351753 14351754 14351755 14351756 14351757 14351759 14351760 14351762 14351763 14351766 14351769 14351772 14351774 14351775 14351776 14351777 14351778 14351781 14351783 14351784 14351785 14351787 14351788 14351789 14351790 14351791 14351793 14351796 14351797 14351799 14351804 14351807 14351808 14351810 14351812 14351814 14351816 14351818 14351819 14351820 14351821 14351822 14351825 14351826 14351827 14351828 14351829 14351831 14351833 14351836 14351837 14351838 14351840 14351841 14351842 14351843 14351844 14351845 14351847 14351850 14351851 14351853 14351854 14351856 14351857 14351861 14351862 14351863 14351864 14351865 14351866 14351867 14351868 14351869 14351870 14351872 14351873 14351875 14351877 14351878 14351879 14351880 14351881 14351882 14351883 14351884 14351885 14351886 14351887 14351889 14351890 14351892 14351894 14351897 14351898 14351899 14351901 14351903 14351904 14351905 14351906 14351908 14351909 14351910 14351913 14351915 14351917 14351918 14351919 14351920 14351922 14351923 14351924 14351926 14351928 14351929 14351930 14351931 14351933 14351935 14351936 14351937 14351939 14351941 14351942 14351943 14351946 14351947 14351949 14351953 14351955 14351957 14351964 14351968 14351969 14351971 14351973 14351974 14351976 14351981 14351982 14351984 14351986 14351989 14351991 14351994 14351995 14351996 14351997 14351998 14351999 14352000 14352001 14352003 14352005 14352006 14352008 14352009 14352010 14352011 14352012 14352013 14352015 14352017 14352023 14352028 14352029 14352030 14352033 14352034 14352035 14352036 14352041 14352045 14352047 14352048 14352050 14352051 14352052 14352054 14352055 14352056 14352058 14352059 14352060 14352066 14352067 14352072 14352075 14352076 14352077 14352078 14352079 14352080 14352081 14352082 14352085 14352086 14352088 14352091 14352093 14352095 14352096 14352098 14352101 14352102 14352103 14352104 14352105 14352109 14352110 14352111 14352114 14352115 14352116 14352117 14352118 14352120 14352121 14352122 14352123 14352124 14352125 14352126 14352127 14352128 14352129 14352130 14352131 14352132 14352133 14352135 14352136 14352138 14352139 14352141 14352142 14352143 14352144 14352145 14352146 14352147 14352148 14352149 14352150 14352151 14352152 14352153 14352154 14352155 14352158 14352159 14352161 14352162 14352164 14352165 14352166 14352167 14352168 14352169 14352170 14352171 14352172 14352173 14352174 14352175 14352177 14352179 14352181 14352182 14352183 14352185 14352187 14352188 14352189 14352190 14352191 14352194 14352196 14352197 14352198 14352199 14352201 14352202 14352203 14352204 14352205 14352206 14352210 14352214 14352217 14352218 14352222 14352223 14352225 14352226 14352227 14352230 14352231 14352232 14352234 14352235 14352236 14352237 14352238 14352239 14352240 14352241 14352242 14352243 14352244 14352246 14352247 14352248 14352249 14352255 14352257 14352258 14352259 14352260 14352263 14352264 14352265 14352266 14352267 14352268 14352269 14352270 14352271 14352272 14352273 14352274 14352275 14352276 14352277 14352281 14352282 14352284 14352285 14352286 14352287 14352288 14352291 14352293 14352294 14352295 14352296 14352298 14352299 14352301 14352302 14352305 14352306 14352307 14352309 14352312 14352313 14352314 14352316 14352317 14352318 14352320 14352321 14352322 14352326 14352327 14352330 14352331 14352332 14352336 14352337 14352338 14352339 14352340 14352341 14352342 14352343 14352345 14352346 14352347 14352348 14352349 14352350 14352351 14352352 14352355 14352358 14352359 14352360 14352361 14352362 14352364 14352366 14352367 14352369 14352371 14352373 14352374 14352375 14352376 14352377 14352378 14352380 14352382 14352383 14352385 14352386 14352387 14352389 14352392 14352395 14352397 14352399 14352400 14352403 14352404 14352405 14352406 14352407 14352409 14352410 14352412 14352413 14352414 14352415 14352416 14352417 14352419 14352421 14352424 14352429 14352430 14352432 14352433 14352434 14352437 14352438 14352439 14352440 14352442 14352448 14352449 14352450 14352451 14352453 14352454 14352455 14352456 14352459 14352462 14352464 14352465 14352466 14352467 14352468 14352469 14352470 14352472 14352475 14352476 14352477 14352478 14352481 14352482 14352485 14352486 14352487 14352488 14352489 14352490 14352494 14352495 14352496 14352497 14352499 14352503 14352504 14352505 14352507 14352508 14352509 14352510 14352511 14352512 14352513 14352516 14352517 14352518 14352519 14352521 14352523 14352524 14352525 14352527 14352528 14352530 14352531 14352533 14352534 14352535 14352536 14352537 14352538 14352539 14352541 14352542 14352544 14352546 14352549 14352550 14352551 14352552 14352553 14352556 14352557 14352558 14352559 14352560 14352563 14352564 14352566 14352567 14352568 14352569 14352570 14352571 14352572 14352573 14352574 14352575 14352576 14352577 14352578 14352579 14352580 14352581 14352582 14352583 14352585 14352587 14352588 14352589 14352590 14352591 14352592 14352593 14352594 14352595 14352596 14352600 14352601 14352602 14352603 14352604 14352606 14352607 14352608 14352609 14352610 14352611 14352612 14352613 14352614 14352615 14352617 14352621 14352622 14352623 14352625 14352626 14352629 14352630 14352631 14352633 14352638 14352640 14352641 14352642 14352643 14352645 14352647 14352652 14352653 14352654 14352655 14352656 14352657 14352658 14352659 14352660 14352663 14352665 14352668 14352669 14352670 14352671 14352672 14352673 14352674 14352675 14352676 14352677 14352678 14352680 14352683 14352684 14352685 14352686 14352687 14352688 14352690 14352691 14352693 14352695 14352698 14352699 14352701 14352702 14352703 14352704 14352705 14352706 14352707 14352708 14352710 14352711 14352712 14352713 14352715 14352716 14352717 14352719 14352720 14352725 14352726 14352727 14352728 14352729 14352730 14352731 14352735 14352737 14352738 14352740 14352741 14352745 14352746 14352749 14352750 14352751 14352752 14352755 14352756 14352758 14352761 14352762 14352763 14352764 14352765 14352766 14352768 14352769 14352770 14352771 14352772 14352773 14352776 14352778 14352779 14352781 14352782 14352783 14352784 14352785 14352788 14352789 14352790 14352791 14352792 14352793 14352794 14352795 14352796 14352798 14352799 14352803 14352804 14352805 14352807 14352808 14352809 14352810 14352811 14352813 14352814 14352815 14352816 14352818 14352819 14352821 14352822 14352824 14352826 14352828 14352829 14352830 14352831 14352832 14352833 14352834 14352836 14352838 14352839 14352840 14352841 14352842 14352843 14352844 14352845 14352847 14352848 14352850 14352851 14352852 14352853 14352854 14352855 14352864 14352865 14352866 14352867 14352868 14352870 14352873 14352874 14352875 14352876 14352878 14352880 14352881 14352882 14352883 14352884 14352885 14352886 14352887 14352888 14352889 14352891 14352892 14352893 14352895 14352896 14352897 14352898 14352899 14352900 14352901 14352902 14352903 14352904 14352906 14352907 14352908 14352909 14352910 14352911 14352912 14352914 14352915 14352916 14352917 14352918 14352919 14352920 14352923 14352925 14352926 14352927 14352928 14352929 14352930 14352932 14352933 14352935 14352936 14352938 14352939 14352940 14352941 14352942 14352944 14352947 14352948 14352949 14352950 14352953 14352954 14352956 14352957 14352960 14352961 14352962 14352963 14352964 14352965 14352966 14352967 14352968 14352969 14352970 14352972 14352973 14352974 14352975 14352976 14352977 14352978 14352980 14352982 14352983 14352984 14352986 14352987 14352988 14352989 14352990 14352991 14352992 14352994 14352995 14352998 14353000 14353001 14353002 14353004 14353005 14353006 14353007 14353009 14353012 14353013 14353014 14353015 14353016 14353017 14353018 14353019 14353020 14353021 14353022 14353024 14353025 14353026 14353027 14353028 14353029 14353030 14353031 14353032 14353033 14353034 14353035 14353036 14353037 14353038 14353039 14353040 14353042 14353043 14353044 14353045 14353047 14353048 14353049 14353051 14353052 14353054 14353055 14353057 14353058 14353059 14353060 14353061 14353064 14353066 14353067 14353068 14353069 14353070 14353071 14353072 14353073 14353074 14353075 14353076 14353077 14353078 14353079 14353080 14353081 14353082 14353083 14353084 14353085 14353086 14353087 14353089 14353090 14353091 14353092 14353094 14353095 14353096 14353097 14353098 14353099 14353100 14353101 14353102 14353103 14353104 14353105 14353106 14353107 14353108 14353109 14353110 14353113 14353114 14353115 14353116 14353117 14353118 14353120 14353123 14353124 14353127 14353128 14353129 14353130 14353131 14353136 14353137 14353138 14353139 14353140 14353141 14353142 14353143 14353145 14353146 14353149 14353150 14353151 14353152 14353153 14353155 14353156 14353157 14353158 14353159 14353160 14353161 14353162 14353164 14353165 14353167 14353168 14353169 14353170 14353172 14353173 14353174 14353175 14353176 14353178 14353179 14353180 14353181 14353182 14353183 14353184 14353185 14353186 14353187 14353188 14353189 14353190 14353191 14353192 14353193 14353194 14353195 14353196 14353198 14353199 14353200 14353201 14353202 14353203 14353204 14353205 14353207 14353208 14353210 14353211 14353212 14353213 14353215 14353216 14353217 14353218 14353220 14353221 14353222 14353223 14353225 14353226 14353227 14353228 14353229 14353230 14353231 14353233 14353234 14353235 14353236 14353238 14353239 14353240 14353244 14353245 14353246 14353247 14353248 14353249 14353251 14353252 14353253 14353254 14353255 14353256 14353257 14353258 14353259 14353260 14353262 14353263 14353264 14353265 14353267 14353269 14353272 14353273 14353274 14353275 14353276 14353278 14353279 14353280 14353282 14353283 14353284 14353285 14353287 14353289 14353292 14353294 14353295 14353296 14353297 14353298 14353300 14353303 14353304 14353305 14353306 14353307 14353309 14353311 14353312 14353313 14353314 14353316 14353317 14353318 14353319 14353321 14353322 14353323 14353325 14353326 14353327 14353328 14353329 14353330 14353331 14353332 14353334 14353336 14353337 14353339 14353340 14353341 14353344 14353345 14353346 14353347 14353348 14353349 14353350 14353352 14353353 14353354 14353357 14353359 14353360 14353361 14353362 14353363 14353364 14353365 14353366 14353367 14353369 14353370 14353373 14353374 14353375 14353376 14353377 14353379 14353381 14353382 14353383 14353384 14353386 14353387 14353388 14353389 14353390 14353391 14353393 14353394 14353395 14353396 14353397 14353398 14353399 14353400 14353401 14353402 14353403 14353404 14353405 14353406 14353408 14353409 14353410 14353411 14353413 14353414 14353415 14353418 14353421 14353423 14353424 14353425 14353426 14353427 14353428 14353429 14353430 14353431 14353432 14353433 14353434 14353435 14353436 14353437 14353439 14353440 14353441 14353444 14353445 14353448 14353449 14353450 14353452 14353453 14353455 14353456 14353457 14353458 14353460 14353463 14353464 14353465 14353466 14353467 14353469 14353470 14353471 14353472 14353473 14353474 14353475 14353477 14353481 14353482 14353483 14353486 14353487 14353489 14353491 14353492 14353493 14353494 14353495 14353497 14353498 14353499 14353500 14353501 14353502 14353503 14353504 14353505 14353506 14353507 14353510 14353511 14353516 14353517 14353518 14353519 14353522 14353523 14353525 14353526 14353528 14353529 14353530 14353531 14353534 14353535 14353536 14353537 14353539 14353540 14353541 14353543 14353544 14353546 14353548 14353550 14353552 14353553 14353554 14353555 14353558 14353559 14353561 14353563 14353564 14353566 14353567 14353568 14353569 14353570 14353571 14353572 14353573 14353574 14353575 14353576 14353577 14353578 14353579 14353582 14353583 14353585 14353586 14353587 14353588 14353592 14353593 14353595 14353596 14353597 14353599 14353600 14353601 14353603 14353605 14353607 14353609 14353610 14353611 14353612 14353613 14353614 14353615 14353617 14353620 14353622 14353623 14353624 14353625 14353630 14353631 14353632 14353633 14353635 14353636 14353637 14353638 14353639 14353640 14353641 14353642 14353643 14353645 14353646 14353647 14353648 14353649 14353650 14353652 14353653 14353654 14353656 14353657 14353658 14353659 14353660 14353661 14353665 14353666 14353667 14353669 14353670 14353671 14353673 14353675 14353676 14353678 14353680 14353681 14353682 14353684 14353685 14353687 14353689 14353692 14353693 14353694 14353697 14353698 14353699 14353700 14353701 14353702 14353707 14353708 14353709 14353715 14353716 14353717 14353718 14353720 14353721 14353722 14353723 14353725 14353726 14353728 14353730 14353731 14353732 14353735 14353736 14353737 14353738 14353741 14353742 14353744 14353745 14353747 14353748 14353749 14353752 14353753 14353754 14353757 14353758 14353759 14353760 14353761 14353762 14353763 14353764 14353765 14353766 14353767 14353768 14353769 14353770 14353772 14353773 14353774 14353775 14353778 14353782 14353783 14353784 14353785 14353788 14353789 14353792 14353793 14353794 14353799 14353800 14353801 14353802 14353803 14353805 14353809 14353810 14353811 14353812 14353814 14353816 14353817 14353819 14353821 14353823 14353824 14353825 14353826 14353828 14353831 14353833 14353835 14353837 14353840 14353841 14353847 14353848 14353849 14353850 14353851 14353852 14353853 14353854 14353856 14353857 14353858 14353860 14353861 14353862 14353864 14353866 14353868 14353871 14353874 14353875 14353876 14353879 14353880 14353881 14353883 14353884 14353885 14353888 14353891 14353892 14353893 14353895 14353896 14353897 14353898 14353901 14353902 14353904 14353905 14353906 14353907 14353908 14353909 14353912 14353913 14353914 14353915 14353916 14353917 14353918 14353919 14353921 14353922 14353923 14353924 14353925 14353926 14353927 14353928 14353929 14353930 14353931 14353933 14353934 14353935 14353936 14353937 14353939 14353940 14353941 14353942 14353943 14353948 14353949 14353950 14353954 14353956 14353958 14353959 14353960 14353961 14353963 14353965 14353967 14353969 14353970 14353971 14353972 14353974 14353975 14353977 14353979 14353980 14353981 14353983 14353985 14353988 14353989 14353990 14353991 14353992 14353993 14353996 14353997 14353998 14354000 14354001 14354003 14354004 14354005 14354006 14354007 14354008 14354009 14354010 14354011 14354012 14354013 14354015 14354018 14354019 14354021 14354022 14354023 14354024 14354026 14354028 14354030 14354032 14354033 14354034 14354035 14354036 14354037 14354038 14354040 14354043 14354044 14354045 14354046 14354047 14354051 14354054 14354055 14354057 14354058 14354059 14354060 14354061 14354065 14354066 14354067 14354068 14354069 14354070 14354072 14354073 14354075 14354076 14354077 14354078 14354079 14354081 14354082 14354083 14354084 14354089 14354090 14354091 14354092 14354093 14354094 14354095 14354096 14354097 14354098 14354099 14354100 14354101 14354102 14354104 14354105 14354106 14354107 14354112 14354114 14354116 14354118 14354120 14354121 14354127 14354129 14354130 14354132 14354133 14354134 14354135 14354136 14354137 14354140 14354143 14354146 14354148 14354150 14354151 14354153 14354154 14354155 14354158 14354159 14354160 14354161 14354164 14354165 14354168 14354170 14354173 14354174 14354177 14354178 14354179 14354180 14354181 14354182 14354190 14354192 14354193 14354194 14354195 14354198 14354199 14354200 14354202 14354206 14354210 14354212 14354213 14354215 14354216 14354217 14354219 14354220 14354222 14354223 14354224 14354230 14354231 14354232 14354233 14354234 14354235 14354236 14354238 14354239 14354240 14354241 14354242 14354243 14354244 14354245 14354247 14354248 14354249 14354251 14354253 14354256 14354257 14354259 14354260 14354261 14354262 14354264 14354265 14354267 14354268 14354269 14354271 14354274 14354275 14354276 14354277 14354278 14354279 14354280 14354281 14354283 14354285 14354287 14354289 14354291 14354292 14354294 14354300 14354304 14354510 14354511 14354512 14354513 14354514 14354515 14354516 14354517 14354518 14354519 14354521 14354523 14354524 14354525 14354526 14354528 14354529 14354530 14354531 14354532 14354533 14354534 14354537 14354538 14354539 14354540 14354541 14354542 14354543 14354544 14354545 14354546 14354547 14354548 14354549 14354550 14354551 14354552 14354553 14354554 14354555 14354556 14354557 14354558 14354559 14354560 14354561 14354563 14354565 14354566 14354567 14354568 14354569 14354571 14354573 14354574 14354575 14354576 14354578 14354579 14354580 14354581 14354583 14354584 14354585 14354587 14354588 14354589 14354590 14354591 14354593 14354594 14354595 14354598 14354599 14354601 14354602 14354603 14354605 14354606 14354607 14354609 14354610 14354612 14354613 14354614 14354615 14354616 14354617 14354618 14354619 14354620 14354621 14354622 14354623 14354625 14354626 14354627 14354628 14354630 14354631 14354632 14354633 14354634 14354635 14354639 14354640 14354643 14354645 14354647 14354648 14354649 14354650 14354652 14354653 14354654 14354655 14354656 14354657 14354658 14354659 14354660 14354661 14354662 14354663 14354664 14354665 14354666 14354667 14354668 14354669 14354670 14354671 14354672 14354674 14354676 14354678 14354680 14354681 14354682 14354683 14354684 14354685 14354687 14354688 14354689 14354690 14354691 14354692 14354693 14354694 14354695 14354696 14354697 14354698 14354699 14354700 14354701 14354702 14354704 14354705 14354707 14354710 14354712 14354714 14354717 14354720 14354721 14354722 14354724 14354726 14354728 14354729 14354730 14354731 14354732 14354733 14354734 14354735 14354737 14354738 14354739 14354740 14354741 14354742 14354743 14354746 14354747 14354749 14354750 14354751 14354752 14354753 14354754 14354755 14354756 14354757 14354758 14354761 14354762 14354763 14354764 14354765 14354766 14354767 14354768 14354769 14354771 14354772 14354773 14354774 14354775 14354776 14354777 14354778 14354779 14354780 14354782 14354783 14354784 14354785 14354786 14354787 14354788 14354789 14354791 14354792 14354796 14354798 14354799 14354800 14354801 14354802 14354804 14354805 14354806 14354807 14354808 14354810 14354811 14354812 14354813 14354814 14354815 14354817 14354818 14354819 14354820 14354821 14354822 14354825 14354826 14354827 14354829 14354832 14354833 14354835 14354836 14354837 14354838 14354840 14354841 14354843 14354844 14354845 14354846 14354847 14354848 14354849 14354852 14354853 14354854 14354855 14354856 14354857 14354858 14354859 14354860 14354861 14354862 14354863 14354864 14354867 14354868 14354869 14354870 14354871 14354874 14354875 14354876 14354877 14354878 14354879 14354880 14354882 14354883 14354885 14354886 14354887 14354888 14354889 14354890 14354892 14354893 14354894 14354895 14354896 14354897 14354898 14354900 14354901 14354902 14354903 14354904 14354905 14354907 14354908 14354909 14354910 14354911 14354912 14354913 14354914 14354915 14354916 14354917 14354918 14354921 14354922 14354923 14354924 14354925 14354927 14354928 14354929 14354930 14354931 14354932 14354933 14354934 14354936 14354938 14354939 14354940 14354943 14354944 14354945 14354947 14354948 14354949 14354951 14354952 14354954 14354955 14354956 14354957 14354959 14354960 14354962 14354963 14354964 14354965 14354966 14354968 14354969 14354970 14354971 14354972 14354973 14354974 14354976 14354978 14354979 14354981 14354982 14354983 14354985 14354986 14354987 14354988 14354989 14354990 14354991 14354992 14354993 14354995 14354996 14354997 14354998 14354999 14355000 14356421 14356422 14356424 14356425 14356426 14356429 14356430 14356432 14356433 14356434 14356435 14356440 14356444 14356445 14356447 14356448 14356450 14356451 14356453 14356455 14356456 14356458 14356459 14356460 14356461 14356462 14356463 14356465 14356467 14356469 14356471 14356472 14356474 14356475 14356476 14356478 14356480 14356484 14356485 14356486 14356487 14356489 14356490 14356491 14356492 14356493 14356495 14356496 14356497 14356499 14356500 14356501 14356502 14356503 14356506 14356507 14356508 14356509 14356510 14356511 14356513 14356514 14356515 14356517 14356518 14356519 14356522 14356525 14356526 14356527 14356528 14356530 14356531 14356532 14356533 14356535 14356536 14356537 14356538 14356540 14356541 14356542 14356544 14356545 14356547 14356548 14356549 14356550 14356551 14356553 14356554 14356555 14356556 14356559 14356560 14356561 14356562 14356563 14356564 14356565 14356566 14356567 14356568 14356569 14356572 14356573 14356574 14356579 14356580 14356581 14356582 14356583 14356584 14356585 14356587 14356588 14356590 14356591 14356592 14356593 14356594 14356595 14356596 14356597 14356598 14356599 14356600 14356602 14356603 14356605 14356606 14356607 14356608 14356609 14356610 14356612 14356614 14356617 14356618 14356619 14356620 14356621 14356622 14356623 14356624 14356625 14356626 14356627 14356628 14356629 14356630 14356632 14356634 14356635 14356636 14356637 14356638 14356639 14356640 14356641 14356642 14356643 14356644 14356645 14356646 14356647 14356648 14356649 14356650 14356651 14356652 14356653 14356654 14356655 14356656 14356657 14356658 14356659 14356660 14356661 14356662 14356663 14356664 14356665 14356666 14356667 14356668 14356669 14356671 14356672 14356673 14356674 14356675 14356676 14356677 14356678 14356679 14356680 14356681 14356682 14356683 14356684 14356685 14356686 14356687 14356688 14356689 14356690 14356691 14356692 14356693 14356694 14356695 14356696 14356697 14356698 14356699 14356700 14356701 14356702 14356703 14356704 14356705 14356706 14356707 14356708 14356709 14356710 14356711 14356712 14356713 14356714 14356715 14356716 14356717 14356718 14356719 14356720 14356721 14356722 14356723 14356724 14356725 14356727 14356728 14356729 14356730 14356731 14356732 14356733 14356734 14356735 14356736 14356737 14356738 14356739 14356740 14356741 14356742 14356743 14356744 14356745 14356746 14356747 14356748 14356749 14356750 14356751 14356752 14356753 14356754 14356755 14356756 14356757 14356758 14356759 14356760 14356761 14356762 14356763 14356764 14356765 14356766 14356767 14356768 14356769 14356770 14356771 14356772 14356773 14356774 14356775 14356776 14356777 14356778 14356779 14356780 14356781 14356782 14356783 14356784 14356785 14356786 14356787 14356788 14356789 14356790 14356791 14356792 14356793 14356794 14356795 14356796 14356797 14356798 14356799 14356800 14356801 14356802 14356803 14356804 14356805 14356806 14356807 14356808 14356809 14356810 14356812 14356813 14356814 14356815 14356816 14356817 14356818 14356819 14356820 14356821 14356822 14356823 14356824 14356825 14356826 14356827 14356828 14356829 14356830 14356831 14356832 14356833 14356834 14356835 14356836 14356837 14356838 14356839 14356840 14356841 14356842 14356843 14356845 14356846 14356847 14356848 14356849 14356850 14356851 14356853 14356854 14356855 14356856 14356857 14356858 14356859 14356860 14356861 14356862 14356863 14356864 14356865 14356866 14356867 14356868 14356869 14356870 14356871 14356872 14356873 14356874 14356875 14356876 14356877 14356878 14356879 14356880 14356881 14356882 14356883 14356885 14356886 14356887 14356888 14356889 14356890 14356891 14356892 14356893 14356894 14356895 14356896 14356897 14356898 14356899 14356900 14356901 14356903 14356904 14356905 14356906 14356907 14356908 14356909 14356910 14356911 14356912 14356913 14356914 14356915 14356916 14356917 14356918 14356920 14356921 14356923 14356924 14356925 14356926 14356927 14356928 14356930 14356932 14356933 14356935 14356937 14356938 14356940 14356941 14356942 14356943 14356944 14356945 14356946 14356947 14356948 14356949 14356953 14356954 14356955 14356957 14356958 14356959 14356960 14356961 14356962 14356963 14356964 14356965 14356967 14356968 14356969 14356971 14356972 14356974 14356975 14356976 14356977 14356978 14356979 14356980 14356981 14356982 14356983 14356985 14356987 14356988 14356991 14356992 14356993 14356994 14356995 14356996 14356997 14356999 14357000 14357001 14357002 14357003 14357005 14357006 14357007 14357008 14357009 14357013 14357014 14357018 14357019 14357020 14357021 14357023 14357024 14357025 14357027 14357028 14357029 14357030 14357032 14357034 14357036 14357037 14357038 14357039 14357040 14357041 14357042 14357043 14357044 14357045 14357046 14357048 14357049 14357050 14357052 14357054 14357055 14357056 14357058 14357060 14357061 14357062 14357063 14357066 14357067 14357068 14357069 14357070 14357071 14357074 14357077 14357079 14357082 14357084 14357085 14357086 14357087 14357089 14357090 14357091 14357093 14357094 14357095 14357096 14357097 14357098 14357099 14357100 14357102 14357103 14357104 14357105 14357107 14357108 14357110 14357111 14357112 14357113 14357114 14357115 14357116 14357118 14357120 14357121 14357122 14357124 14357126 14357128 14357129 14357130 14357131 14357132 14357134 14357135 14357136 14357137 14357138 14357139 14357141 14357142 14357143 14357144 14357145 14357146 14357147 14357148 14357149 14357151 14357152 14357154 14357156 14357157 14357158 14357159 14357160 14357162 14357163 14357164 14357166 14357167 14357168 14357169 14357171 14357172 14357175 14357176 14357178 14357179 14357182 14357183 14357184 14357186 14357187 14357189 14357190 14357191 14357192 14357193 14357194 14357195 14357196 14357198 14357199 14357200 14357201 14357202 14357203 14357205 14357206 14357207 14357209 14357210 14357211 14357212 14357213 14357214 14357215 14357216 14357217 14357218 14357220 14357222 14357223 14357224 14357225 14357228 14357230 14357232 14357234 14357235 14357237 14357240 14357241 14357242 14357243 14357244 14357245 14357246 14357249 14357250 14357251 14357252 14357253 14357254 14357256 14357257 14357258 14357259 14357260 14357261 14357264 14357265 14357266 14357267 14357269 14357270 14357271 14357274 14357275 14357276 14357277 14357279 14357281 14357282 14357283 14357284 14357285 14357287 14357288 14357289 14357290 14357291 14357292 14357294 14357295 14357296 14357297 14357298 14357299 14357300 14357302 14357305 14357306 14357307 14357308 14357309 14357311 14357312 14357313 14357314 14357316 14357319 14357320 14357321 14357322 14357324 14357327 14357330 14357331 14357334 14357335 14357337 14357341 14357343 14357345 14357346 14357347 14357349 14357350 14357352 14357353 14357354 14357355 14357356 14357357 14357359 14357360 14357361 14357363 14357364 14357365 14357366 14357368 14357372 14357373 14357374 14357377 14357378 14357380 14357381 14357384 14357385 14357386 14357389 14357391 14357393 14357394 14357395 14357396 14357397 14357398 14357400 14357402 14357404 14357406 14357407 14357410 14357417 14357418 14357421 14357426 14357436 14357440 14357449 14357450 14357452 14357456 14357458 14357462 14357463 14357468 14357472 14357474 14357483 14357484 14357487 14357490 14357494 14357496 14357497 14357501 14357506 14357508 14357510 14357511 14357523 14357524 14357526 14357530 14357533 14357537 14357538 14357539 14357549 14357551 14357559 14357561 14357564 14357565 14357566 14357567 14357568 14357569 14357570 14357571 14357573 14357580 14357581 14357583 14357584 14357585 14357588 14357591 14357597 14357598 14357600 14357601 14357602 14357604 14357606 14357608 14357609 14357611 14357612 14357613 14357614 14357615 14357617 14357620 14357621 14357622 14357623 14357624 14357625 14357626 14357627 14357628 14357629 14357630 14357631 14357632 14357633 14357634 14357635 14357636 14357637 14357638 14357639 14357640 14357641 14357642 14357643 14357644 14357645 14357646 14357647 14357648 14357649 14357650 14357651 14357652 14357653 14357655 14357656 14357657 14357659 14357660 14357662 14357664 14357665 14357666 14357667 14357668 14357669 14357670 14357671 14357672 14357674 14357675 14357676 14357677 14357678 14357679 14357680 14357681 14357682 14357683 14357684 14357685 14357686 14357687 14357688 14357690 14357691 14357692 14357693 14357695 14357697 14357698 14357699 14357701 14357702 14357703 14357704 14357706 14357707 14357710 14357711 14357714 14357715 14357717 14357718 14357719 14357721 14357722 14357723 14357724 14357725 14357726 14357727 14357728 14357730 14357731 14357733 14357734 14357735 14357736 14357737 14357738 14357741 14357742 14357743 14357744 14357745 14357746 14357749 14357750 14357751 14357752 14357753 14357754 14357755 14357756 14357757 14357758 14357760 14357761 14357762 14357763 14357764 14357766 14357767 14357768 14357769 14357770 14357771 14357774 14357775 14357776 14357778 14357779 14357781 14357785 14357786 14357787 14357788 14357789 14357790 14357791 14357792 14357793 14357794 14357795 14357796 14357797 14357798 14357799 14357800 14357801 14357802 14357803 14357804 14357806 14357807 14357808 14357809 14357810 14357812 14357814 14357815 14357816 14357817 14357818 14357819 14357821 14357822 14357823 14357824 14357825 14357826 14357827 14357828 14357829 14357832 14357834 14357835 14357839 14357840 14357842 14357843 14357847 14357849 14357851 14357853 14357857 14357858 14357859 14357861 14357862 14357863 14357864 14357865 14357866 14357871 14357872 14357873 14357874 14357875 14357876 14357877 14357878 14357879 14357880 14357882 14357884 14357886 14357887 14357888 14357889 14357890 14357893 14357894 14357895 14357898 14357899 14357900 14357903 14357905 14357907 14357911 14357912 14357924 14357926 14357927 14357929 14357931 14357932 14357933 14357934 14357935 14357936 14357937 14357939 14357941 14357942 14357943 14357944 14357947 14357950 14357952 14357953 14357954 14357955 14357956 14357957 14357959 14357960 14357962 14357963 14357965 14357966 14357969 14357970 14357971 14357973 14357976 14357978 14357979 14357983 14357985 14357989 14357990 14357991 14357992 14357994 14357996 14357997 14357998 14358001 14358002 14358005 14358006 14358007 14358008 14358010 14358011 14358012 14358013 14358015 14358019 14358020 14358021 14358027 14358031 14358032 14358033 14358036 14358044 14358045 14358047 14358054 14358058 14358061 14358064 14358068 14358070 14358074 14358075 14358077 14358079 14358080 14358081 14358083 14358084 14358085 14358087 14358088 14358089 14358090 14358092 14358094 14358096 14358097 14358099 14358102 14358103 14358104 14358107 14358108 14358110 14358112 14358115 14358116 14358117 14358118 14358119 14358120 14358121 14358125 14358127 14358129 14358130 14358131 14358132 14358133 14358134 14358135 14358137 14358138 14358139 14358140 14358142 14358143 14358144 14358145 14358148 14358151 14358154 14358158 14358159 14358161 14358163 14358164 14358165 14358166 14358168 14358169 14358170 14358171 14358172 14358173 14358174 14358176 14358177 14358178 14358180 14358181 14358183 14358184 14358185 14358186 14358187 14358188 14358189 14358191 14358192 14358194 14358195 14358196 14358197 14358198 14358199 14358200 14358203 14358204 14358205 14358206 14358207 14358208 14358209 14358210 14358211 14358212 14358214 14358216 14358217 14358218 14358222 14358223 14358224 14358226 14358228 14358229 14358230 14358231 14358232 14358233 14358235 14358236 14358237 14358242 14358243 14358244 14358245 14358246 14358247 14358249 14358250 14358252 14358254 14358257 14358258 14358259 14358260 14358262 14358264 14358266 14358267 14358268 14358269 14358271 14358276 14358277 14358278 14358279 14358280 14358281 14358282 14358283 14358284 14358285 14358286 14358287 14358288 14358289 14358290 14358291 14358292 14358293 14358294 14358295 14358296 14358297 14358298 14358299 14358300 14358301 14358302 14358303 14358304 14358305 14358306 14358307 14358308 14358309 14358310 14358311 14358312 14358313 14358314 14358315 14358316 14358317 14358318 14358319 14358321 14358322 14358323 14358324 14358325 14358326 14358327 14358328 14358329 14358330 14358331 14358332 14358333 14358334 14358335 14358336 14358337 14358338 14358339 14358340 14358341 14358342 14358343 14358344 14358345 14358346 14358347 14358349 14358350 14358351 14358352 14358354 14358355 14358356 14358358 14358359 14358360 14358361 14358362 14358363 14358364 14358365 14358366 14358367 14358368 14358369 14358370 14358371 14358372 14358373 14358374 14358375 14358376 14358377 14358379 14358380 14358381 14358382 14358383 14358384 14358385 14358386 14358387 14358388 14358389 14358390 14358391 14358392 14358393 14358394 14358395 14358396 14358397 14358398 14358399 14358400 14358401 14358402 14358403 14358404 14358405 14358406 14358407 14358408 14358409 14358410 14358411 14358412 14358413 14358414 14358415 14358416 14358417 14358418 14358420 14358421 14358422 14358423 14358424 14358425 14358426 14358428 14358429 14358430 14358431 14358432 14358433 14358434 14358435 14358436 14358437 14358438 14358439 14358440 14358441 14358442 14358443 14358444 14358445 14358446 14358447 14358448 14358449 14358450 14358451 14358452 14358453 14358454 14358455 14358456 14358457 14358458 14358459 14358460 14358461 14358462 14358463 14358464 14358465 14358466 14358467 14358468 16452240 16452242 16452317 16452371 16452403 16452441 16452471 16452502 16452561 16452562 16452579 16452618 16452693 16452699 16452756 16452826 16452832 16452844 16452896 16452934 16452963 16452966 16452984 16452987 16452997 16453005 16453006 16453007 16453008 16453013 16453016 16453019 16453022 16453023 16453026 16453028 16453030 16453035 16453039 16453041 16453049 16453050 16453051 16453055 16453058 16453060 16453070 16453079 16453085 16453086 16453089 16453091 16453092 16453101 16453104 16453108 16453109 16453111 16453114 16453115 16453123 16453126 16453127 16453136 16453141 16453144 16453151 16453154 16453156 16453158 16453160 16453164 16453165 16453172 16453174 16453176 16453177 16453178 16453179 16453180 16453181 16453182 16453183 16453184 16453187 16453189 16453191 16453192 16453195 16453196 16453199 16453202 16453205 16453207 16453208 16453209 16453211 16453212 16453213 16453214 16453215 16453216 16453218 16453220 16453222 16453223 16453226 16453227 16453229 16453230 16453231 16453232 16453233 16453236 16453238 16453239 16453241 16453242 16453243 16453246 16453247 16453248 16453250 16453251 16453252 16453253 16453254 16453256 16453257 16453258 16453260 16453261 16453263 16453264 16453266 16453269 16453270 16453272 16453273 16453274 16453275 16453276 16453277 16453280 16453281 16453282 16453284 16453285 16453286 16453287 16453291 16453292 16453293 16453297 16453298 16453299 16453303 16453316 16453317 16453318 16453321 16453323 16453325 16453329 16453331 16453335 16453337 16453339 16453342 16453344 16453345 16453346 16453347 16453353 16453354 16453360 16453361 16453367 16453372 16453374 16453375 16453376 16453377 16453381 16453382 16453386 16453387 16453388 16453390 16453391 16453393 16453394 16453398 16453400 16453401 16453402 16453405 16453408 16453410 16453414 16453415 16453417 16453422 16453423 16453427 16453428 16453430 16453432 16453433 16453435 16453438 16453445 16453446 16453448 16453449 16453453 16453455 16453456 16453457 16453458 16453461 16453463 16453464 16453466 16453468 16453474 16453475 16453480 16453481 16453484 16453485 16453488 16453492 16453493 16453496 16453497 16453499 16453502 16453503 16453505 16453508 16453510 16453514 16453516 16453518 16453519 16453520 16453522 16453526 16453528 16453530 16453531 16453533 16453534 16453540 16453541 16453542 16453547 16453549 16453550 16453554 16453556 16453559 16453564 16453568 16453571 16453573 16453576 16453577 16453582 16453584 16453586 16453590 16453593 16453597 16453599 16453602 16453603 16453605 16453606 16453608 16453612 16453615 16453619 16453620 16453627 16453630 16453632 16453635 16453638 16453642 16453646 16453648 16453649 16453650 16453651 16453658 16453660 16453661 16453662 16453663 16453664 16453667 16453668 16453672 16453673 16453674 16453675 16453676 16453677 16453679 16453681 16453682 16453683 16453684 16453686 16453688 16453689 16453690 16453691 16453692 16453693 16453694 16453697 16453699 16453701 16453705 16453707 16453709 16453711 16453714 16453720 16453725 16453726 16453728 16453731 16453733 16453736 16453739 16453740 16453741 16453744 16453745 16453750 16453753 16453754 16453755 16453756 16453757 16453758 16453760 16453762 16453764 16453765 16453768 16453770 16453774 16453776 16453780 16453782 16453783 16453784 16453786 16453787 16453792 16453795 16453796 16453797 16453801 16453802 16453803 16453804 16453807 16453810 16453812 16453813 16453815 16453817 16453819 16453820 16453821 16453826 16453828 16453832 16453833 16453836 16453837 16453838 16453841 16453843 16453844 16453845 16453848 16453849 16453851 16453852 16453855 16453856 16453859 16453862 16453864 16453865 16453867 16453868 16453869 16453870 16453871 16453872 16453873 16453874 16453875 16453877 16453878 16453882 16453885 16453887 16453888 16453889 16453891 16453893 16453894 16453897 16453900 16453903 16453904 16453905 16453907 16453908 16453911 16453914 16453916 16453919 16453920 16453921 16453922 16453924 16453926 16453928 16453929 16453931 16453932 16453939 16453940 16453945 16453946 16453948 16453949 16453950 16453951 16453952 16453953 16453954 16453956 16453959 16453962 16453964 16453965 16453966 16453968 16453975 16453976 16453977 16453978 16453979 16453984 16453988 16453989 16453991 16453993 16453996 16453997 16453999 16454004 16454005 16454006 16454007 16454010 16454011 16454013 16454014 16454015 16454016 16454022 16454024 16454026 16454032 16454033 16454035 16454036 16454038 16454039 16454040 16454044 16454047 16454048 16454052 16454054 16454056 16454057 16454058 16454059 16454060 16454061 16454062 16454064 16454065 16454066 16454069 16454074 16454079 16454082 16454086 16454087 16454093 16454095 16454096 16454100 16454101 16454103 16454109 16454112 16454116 16454117 16454118 16454119 16454122 16454125 16454129 16454130 16454131 16454132 16454135 16454137 16454138 16454139 16454140 16454141 16454143 16454144 16454145 16454147 16454148 16454149 16454151 16454152 16454153 16454154 16454155 16454157 16454159 16454161 16454163 16454164 16454165 16454168 16454169 16454170 16454172 16454173 16454175 16454176 16454178 16454180 16454181 16454183 16454184 16454185 16454186 16454188 16454189 16454190 16454192 16454194 16454195 16454196 16454197 16454198 16454199 16454200 16454201 16454204 16454206 16454207 16454209 16454210 16454211 16454213 16454214 16454215 16454217 16454218 16454219 16454220 16454221 16454222 16454223 16454224 16454226 16454227 16454229 16454230 16454231 16454233 16454235 16454236 16454237 16454238 16454241 16454243 16454244 16454247 16454248 16454249 16454250 16454251 16454252 16454253 16454254 16454255 16454256 16454257 16454259 16454260 16454261 16454262 16454263 16454264 16454267 16454268 16454269 16454270 16454271 16454272 16454273 16454274 16454275 16454276 16454277 16454278 16454279 16454280 16454281 16454282 16454283 16454284 16454285 16454287 16454288 16454289 16454291 16454292 16454293 16454294 16454295 16454296 16454297 16454299 16454300 16454301 16454302 16454303 16454304 16454306 16454309 16454311 16454314 16454316 16454317 16454318 16454319 16454320 16454321 16454323 16454324 16454325 16454329 16454330 16454332 16454333 16454334 16454335 16454336 16454339 16454340 16454342 16454343 16454345 16454346 16454348 16454349 16454351 16454352 16454353 16454354 16454355 16454358 16454360 16454361 16454363 16454364 16454365 16454367 16454368 16454369 16454370 16454371 16454373 16454374 16454376 16454377 16454378 16454379 16454380 16454381 16454382 16454384 16454386 16454387 16454388 16454389 16454390 16454393 16454394 16454396 16454397 16454398 16454399 16454400 16454401 16454403 16454409 16454410 16454412 16454414 16454417 16454418 16454419 16454421 16454422 16454423 16454425 16454427 16454428 16454429 16454431 16454432 16454433 16454434 16454435 16454436 16454437 16454438 16454439 16454441 16454443 16454444 16454445 16454446 16454447 16454448 16454451 16454452 16454453 16454454 16454455 16454456 16454457 16454458 16454459 16454461 16454462 16454463 16454464 16454465 16454466 16454468 16454469 16454470 16454471 16454472 16454473 16454475 16454476 16454477 16454478 16454480 16454481 16454482 16454483 16454484 16454485 16454486 16454487 16454488 16454489 16454490 16454491 16454492 16454495 16454496 16454497 16454499 16454500 16454501 16454502 16454503 16454504 16454505 16454506 16454507 16454509 16454510 16454511 16454512 16454513 16454515 16454516 16454517 16454518 16454519 16454521 16454522 16454523 16454524 16454525 16454526 16454527 16454529 16454530 16454531 16454532 16454534 16454535 16454536 16454538 16454540 16454541 16454543 16454545 16454546 16454547 16454548 16454549 16454550 16454551 16454552 16454553 16454554 16454555 16454556 16454557 16454558 16454560 16454561 16454562 16454563 16454564 16454565 16454566 16454568 16454569 16454571 16454572 16454573 16454575 16454576 16454577 16454578 16454579 16454580 16454581 16454583 16454584 16454587 16454588 16454590 16454591 16454592 16454593 16454595 16454596 16454597 16454598 16454599 16454601 16454602 16454603 16454605 16454606 16454607 16454608 16454610 16454612 16454614 16454615 16454616 16454619 16454620 16454621 16454623 16454624 16454625 16454626 16454627 16454628 16454629 16454630 16454633 16454634 16454635 16454636 16454638 16454640 16454641 16454643 16454645 16454646 16454647 16454648 16454649 16454651 16454654 16454655 16454656 16454657 16454658 16454660 16454661 16454663 16454667 16454668 16454669 16454670 16454671 16454672 16454673 16454674 16454676 16454677 16454678 16454679 16454681 16454683 16454684 16454685 16454686 16454687 16454688 16454690 16454691 16454692 16454693 16454695 16454698 16454699 16454700 16454701 16454702 16454703 16454707 16454708 16454710 16454712 16454713 16454715 16454717 16454721 16454722 16454723 16454724 16454727 16454728 16454729 16454730 16454731 16454732 16454733 16454734 16454735 16454736 16454737 16454738 16454739 16454740 16454741 16454742 16454743 16454744 16454745 16454746 16454747 16454748 16454749 16454752 16454753 16454755 16454758 16454759 16454760 16454762 16454763 16454764 16454766 16454767 16454772 16454773 16454774 16454775 16454776 16454777 16454778 16454779 16454783 16454784 16454785 16454786 16454787 16454788 16454789 16454791 16454792 16454793 16454794 16454798 16454799 16454800 16454803 16454805 16454807 16454809 16454811 16454813 16454817 16454818 16454820 16454825 16454827 16454831 16454833 16454835 16454836 16454838 16454839 16454841 16454843 16454845 16454846 16454847 16454852 16454854 16454857 16454858 16454859 16454860 16454861 16454862 16454864 16454865 16454867 16454868 16454871 16454873 16454874 16454875 16454876 16454878 16454879 16454880 16454883 16454884 16454885 16454886 16454888 16454890 16454892 16454894 16454895 16454897 16454899 16454901 16454903 16454904 16454905 16454907 16454909 16454910 16454911 16454912 16454913 16454914 16454915 16454916 16454918 16454919 16454920 16454921 16454924 16454925 16454926 16454929 16454930 16454933 16454934 16454935 16454937 16454938 16454939 16454940 16454941 16454943 16454944 16454945 16454946 16454950 16454952 16454953 16454954 16454955 16454956 16454957 16454958 16454961 16454962 16454963 16454964 16454965 16454967 16454968 16454970 16454971 16454973 16454975 16454976 16454977 16454978 16454980 16454981 16454982 16454983 16454984 16454985 16454986 16454987 16454990 16454991 16454994 16454996 16454997 16454998 16455000 16455003 16455005 16455008 16455012 16455015 16455016 16455017 16455018 16455020 16455021 16455022 16455026 16455028 16455029 16455030 16455032 16455033 16455034 16455036 16455037 16455040 16455041 16455044 16455045 16455046 16455047 16455049 16455050 16455051 16455053 16455056 16455057 16455058 16455059 16455060 16455066 16455067 16455068 16455073 16455078 16455080 16455081 16455083 16455084 16455086 16455088 16455089 16455090 16455091 16455093 16455094 16455095 16455098 16455102 16455103 16455104 16455105 16455108 16455111 16455113 16455114 16455116 16455118 16455119 16455120 16455121 16455122 16455123 16455124 16455128 16455130 16455131 16455134 16455135 16455136 16455138 16455139 16455142 16455145 16455146 16455147 16455149 16455150 16455153 16455154 16455157 16455161 16455163 16455164 16455165 16455168 16455171 16455172 16455173 16455175 16455176 16455178 16455179 16455181 16455182 16455183 16455184 16455185 16455186 16455187 16455188 16455189 16455190 16455191 16455192 16455194 16455195 16455196 16455197 16455198 16455199 16455200 16455201 16455202 16455205 16455206 16455207 16455208 16455209 16455211 16455212 16455213 16455216 16455219 16455220 16455221 16455223 16455224 16455225 16455226 16455228 16455229 16455230 16455231 16455233 16455234 16455236 16455237 16455238 16455239 16455241 16455243 16455244 16455245 16455246 16455247 16455248 16455250 16455251 16455252 16455254 16455256 16455257 16455259 16455260 16455261 16455262 16455263 16455264 16455265 16455267 16455268 16455270 16455271 16455274 16455275 16455278 16455279 16455282 16455283 16455286 16455287 16455289 16455291 16455294 16455296 16455298 16455304 16455305 16455306 16455308 16455310 16455313 16455315 16455316 16455317 16455318 16455319 16455320 16455321 16455322 16455323 16455324 16455325 16455327 16455328 16455330 16455332 16455338 16455340 16455343 16455344 16455345 16455346 16455347 16455351 16455352 16455355 16455357 16455358 16455359 16455361 16455362 16455365 16455370 16455372 16455376 16455378 16455379 16455382 16455384 16455385 16455386 16455388 16455389 16455390 16455391 16455392 16455393 16455395 16455396 16455397 16455398 16455399 16455400 16455402 16455403 16455406 16455407 16455409 16455410 16455412 16455413 16455415 16455417 16455418 16455420 16455422 16455424 16455425 16455426 16455429 16455430 16455432 16455433 16455434 16455435 16455436 16455437 16455438 16455439 16455440 16455441 16455442 16455443 16455444 16455445 16455446 16455450 16455452 16455453 16455454 16455455 16455456 16455457 16455459 16455461 16455462 16455463 16455464 16455465 16455466 16455468 16455469 16455470 16455472 16455474 16455475 16455477 16455478 16455479 16455481 16455482 16455483 16455484 16455485 16455486 16455487 16455489 16455492 16455494 16455495 16455496 16455498 16455499 16455500 16455502 16455503 16455505 16455506 16455507 16455508 16455509 16455512 16455513 16455514 16455515 16455520 16455523 16455530 16455532 16455535 16455536 16455537 16455538 16455539 16455540 16455541 16455542 16455545 16455546 16455547 16455548 16455550 16455551 16455552 16455554 16455555 16455556 16455557 16455558 16455559 16455561 16455563 16455564 16455567 16455569 16455570 16455571 16455572 16455573 16455574 16455575 16455577 16455580 16455584 16455585 16455586 16455587 16455588 16455589 16455590 16455591 16455593 16455594 16455595 16455597 16455599 16455601 16455603 16455604 16455605 16455606 16455607 16455610 16455613 16455615 16455616 16455625 16455626 16455628 16455630 16455632 16455633 16455634 16455638 16455639 16455640 16455641 16455642 16455644 16455646 16455649 16455651 16455652 16455653 16455654 16455655 16455656 16455659 16455662 16455664 16455665 16455666 16455669 16455670 16455671 16455673 16455676 16455680 16455681 16455685 16455687 16455690 16455693 16455697 16455699 16455700 16455703 16455708 16455710 16455711 16455712 16455713 16455716 16455718 16455719 16455722 16455724 16455726 16455727 16455728 16455730 16455733 16455734 16455740 16455741 16455742 16455744 16455745 16455747 16455748 16455751 16455753 16455754 16455756 16455759 16455760 16455761 16455762 16455764 16455768 16455769 16455770 16455771 16455774 16455775 16455779 16455780 16455783 16455784 16455786 16455787 16455790 16455791 16455792 16455793 16455796 16455798 16455800 16455803 16455877 16455878 16456657 16456658 16456659 16456660 16456661 16456662 16456663 16456664 16456665 16456667 16456668 16456669 16456670 16456671 16456672 16456673 16456674 16456675 16456676 16456677 16456678 16456679 16456680 16456681 16456682 16456683 16456684 16456685 16456756 16456757 16456760 16456769 16456770 16456771 16456772 16456775 16456796 16456797 16456798 16456802 16456806 16456807 16456818 16456821 16456822 16456825 16456833 16456834 16456836 16456899 16456924 16456930 16456968 16457069 16457070 16457071 16457072 16457073 16457074 16457075 16457076 16457077 16457078 16457079 16457081 16457083 16457084 16457085 16457086 16457087 16457088 16457089 16457090 16457092 16457093 16457097 16457098 16457102 16457105 16457106 16457107 16457110 16457111 16457114 16457115 16457116 17227411 17227417 17227425 17227456 17227557 17227649 17227650 17227652 17227653 17227654 17227655 17227656 17227658 17227659 17227660 17227661 17227662 17227663 17227664 17227666 17227667 17227668 17227669 17227670 17227671 17227672 17227673 17227674 17227675 17227677 17227678 17227679 17227681 17227682 17227683 17227684 17227685 17227686 17227687 17227688 17227689 17227690 17227691 17227692 17227693 17227694 17227695 17227696 17227697 17227698 17227699 17227700 17227701 17227702 17227703 17227704 17227705 17227706 17227707 17227708 17227709 17227710 17227711 17227712 17227713 17227714 17227715 17227716 17227717 17227718 17227719 17227720 17227721 17227722 17227723 17227724 17227725 17227726 17227727 17227728 17227729 17227730 17227731 17227732 17227733 17227734 17227735 17227736 17227737 17227738 17227739 17227740 17227741 17227742 17227743 17227744 17227746 17227747 17227748 17227749 17227750 17227751 17227752 17227753 17227754 17227755 17227756 17227757 17227758 17227759 17227760 17227761 17227762 17227763 17227764 17227765 17227766 17227768 17227769 17227770 17227771 17227772 17227773 17227774 17227775 17227776 17227777 17227778 17227779 17227780 17227781 17227782 17227783 17227784 17227786 17227787 17227788 17227789 17227790 17227791 17227792 17227793 17227794 17227795 17227796 17227797 17227798 17227799 17227800 17227801 17227803 17227804 17227805 17227806 17227807 17227809 17227810 17227811 17227812 17227813 17227814 17227816 17227817 17227818 17227819 17227820 17227821 17227822 17227823 17227824 17227825 17227826 17227827 17227828 17227830 17227831 17227832 17227833 17227834 17227835 17227836 17227837 17227838 17227839 17227840 17227841 17227842 17227843 17227844 17227845 17227846 17227847 17227848 17227849 17227850 17227851 17227852 17227853 17227854 17227855 17227856 17227858 17227859 17227861 17227862 17227863 17227864 17227865 17227866 17227867 17227868 17227869 17227870 17227871 17227872 17227873 17227874 17227881 17227882 17227889 17227890 17227893 17227898 17227900 17227901 17227903 17227912 17227913 17227918 17227927 17227932 17227938 17227940 17227950 17227957 17227975 17227976 17227978 17227979 17227980 17227981 17227983 17228000 17228011 17228012 17228020 17228028 17228035 17228036 17228037 17228043 17228044 17228049 17228054 17228055 17228057 17228058 17228059 17228060 17228062 17228069 17228074 17228078 17228088 17228090 17228091 17228100 17228103 17228104 17228105 17228106 17228107 17228108 17228110 17228113 17228114 17228116 17228127 17228129 17228130 17228131 17228132 17228134 17228137 17228139 17228140 17228141 17228142 17228143 17228147 17228148 17228151 17228154 17228155 17228156 17228157 17228158 17228159 17228162 17228163 17228164 17228166 17228167 17228168 17228170 17228171 17228172 17228174 17228176 17228177 17228178 17228179 17228180 17228181 17228183 17228184 17228187 17228189 17228190 17228192 17228193 17228196 17228197 17228198 17228199 17228200 17228202 17228204 17228205 17228208 17228209 17228210 17228211 17228213 17228214 17228215 17228216 17228217 17228218 17228219 17228220 17228221 17228223 17228224 17228225 17228226 17228227 17228228 17228229 17228230 17228233 17228234 17228235 17228236 17228238 17228241 17228242 17228246 17228247 17228248 17228251 17228252 17228253 17228254 17228255 17228256 17228259 17228260 17228262 17228263 17228264 17228266 17228267 17228269 17228272 17228273 17228274 17228275 17228276 17228280 17228281 17228282 17228283 17228284 17228285 17228286 17228288 17228289 17228291 17228292 17228293 17228294 17228295 17228296 17228297 17228298 17228299 17228301 17228304 17228306 17228307 17228311 17228312 17228313 17228315 17228317 17228318 17228319 17228320 17228323 17228324 17228325 17228326 17228329 17228331 17228333 17228334 17228335 17228336 17228337 17228338 17228339 17228342 17228343 17228344 17228345 17228347 17228350 17228351 17228352 17228354 17228355 17228357 17228358 17228359 17228361 17228364 17228365 17228366 17228367 17228371 17228372 17228373 17228374 17228375 17228377 17228378 17228380 17228382 17228384 17228386 17228388 17228389 17228390 17228391 17228392 17228393 17228394 17228395 17228396 17228399 17228401 17228403 17228405 17228406 17228407 17228413 17228421 17228423 17228424 17228425 17228426 17228427 17228428 17228429 17228430 17228432 17228434 17228435 17228436 17228438 17228439 17228440 17228441 17228443 17228446 17228447 17228448 17228449 17228450 17228451 17228452 17228453 17228454 17228455 17228457 17228458 17228459 17228460 17228461 17228462 17228463 17228464 17228465 17228466 17228467 17228468 17228469 17228470 17228471 17228472 17228473 17228474 17228475 17228476 17228477 17228478 17228479 17228480 17228481 17228482 17228483 17228484 17228485 17228486 17228487 17228488 17228489 17228490 17228491 17228492 17228493 17228494 17228495 17228496 17228497 17228498 17228499 17228500 17228501 17228502 17228503 17228504 17228505 17228506 17228507 17228508 17228509 17228510 17228511 17228512 17228513 17228514 17228515 17228516 17228517 17228518 17228519 17228520 17228521 17228522 17228523 17228524 17228525 17228527 17228528 17228529 17228530 17228531 17228532 17228533 17228534 17228535 17228536 17228537 17228538 17228539 17228540 17228541 17228542 17228543 17228544 17228545 17228546 17228547 17228548 17228549 17228550 17228551 17228552 17228553 17228554 17228555 17228556 17228557 17228558 17228559 17228560 17228561 17228562 17228563 17228564 17228565 17228566 17228567 17228568 17228569 17228570 17228571 17228572 17228573 17228574 17228575 17228577 17228578 17228579 17228580 17228581 17228583 17228584 17228585 17228586 17228587 17228588 17228589 17228590 17228591 17228592 17228593 17228594 17228595 17228596 17228597 17228598 17228599 17228600 17228601 17228602 17228603 17228604 17228605 17228606 17228607 17228608 17228609 17228610 17228611 17228612 17228613 17228614 17228615 17228616 17228617 17228618 17228619 17228620 17228621 17228622 17228623 17228624 17228625 17228626 17228627 17228628 17228629 17228630 17228631 17228632 17228633 17228634 17228635 17228636 17228637 17228638 17228639 17228640 17228641 17228642 17228643 17228644 17228645 17228646 17228647 17228648 17228649 17228650 17228651 17228652 17228653 17228654 17228655 17228656 17228657 17228658 17228659 17228660 17228661 17228662 17228663 17228664 17228665 17228666 17228667 17228668 17228669 17228670 17228671 17228672 17228673 17228674 17228675 17228676 17228677 17228678 17228679 17228680 17228681 17228682 17228683 17228685 17228686 17228687 17228688 17228689 17228690 17228692 17228693 17228694 17228695 17228696 17228697 17228698 17228699 17228700 17228701 17228702 17228703 17228704 17228705 17228706 17228707 17228708 17228709 17228710 17228711 17228712 17228713 17228714 17228715 17228716 17228717 17228718 17228719 17228720 17228721 17228722 17228723 17228724 17228725 17228726 17228727 17228728 17228729 17228730 17228731 17228732 17228733 17228734 17228736 17228737 17228738 17228739 17228740 17228741 17228742 17228743 17228744 17228745 17228746 17228747 17228748 17228749 17228750 17228751 17228752 17228753 17228754 17228755 17228756 17228757 17228758 17228759 17228760 17228761 17228762 17228763 17228764 17228765 17228766 17228767 17228768 17228769 17228770 17228771 17228772 17228773 17228774 17228775 17228776 17228777 17228778 17228779 17228780 17228781 17228782 17228783 17228784 17228785 17228786 17228787 17228788 17228789 17228790 17228791 17228792 17228793 17228794 17228795 17228796 17228797 17228798 17228799 17228800 17228801 17228802 17228803 17228804 17228805 17228806 17228807 17228808 17228809 17228810 17228811 17228812 17228813 17228814 17228815 17228816 17228817 17228818 17228819 17228820 17228821 17228822 17228823 17228824 17228825 17228826 17228827 17228828 17228829 17228830 17228831 17228832 17228833 17228834 17228835 17228836 17228837 17228838 17228839 17228840 17228841 17228842 17228844 17228845 17228846 17228847 17228848 17228849 17228850 17228851 17228852 17228853 17228854 17228855 17228856 17228857 17228858 17228859 17228860 17228861 17228862 17228863 17228864 17228865 17228866 17228867 17228868 17228869 17228870 17228871 17228872 17228873 17228874 17228876 17228877 17228878 17228879 17228880 17228881 17228882 17228883 17228884 17228885 17228886 17228887 17228888 17228889 17228890 17228891 17228892 17228893 17228894 17228895 17228896 17228897 17228898 17228899 17228900 17228901 17228902 17228903 17228904 17228905 17228906 17228907 17228908 17228909 17228910 17228911 17228912 17228913 17228914 17228915 17228916 17228917 17228918 17228919 17228921 17228922 17228924 17228925 17228926 17228927 17228928 17228929 17228930 17228931 17228932 17228933 17228934 17228936 17228937 17228938 17228939 17228940 17228941 17228942 17228943 17228944 17228945 17228946 17228947 17228948 17228949 17228950 17228951 17228952 17228953 17228954 17228955 17228956 17228957 17228958 17228959 17228960 17228961 17228962 17228963 17228965 17228966 17228967 17228968 17228969 17228970 17228971 17228972 17228973 17228974 17228975 17228976 17228978 17228979 17228980 17228981 17228982 17228983 17228984 17228985 17228986 17228987 17228989 17228990 17228991 17228992 17228995 17228996 17228997 17228998 17228999 17229000 17229001 17229002 17229003 17229005 17229006 17229007 17229008 17229009 17229011 17229013 17229014 17229015 17229016 17229017 17229018 17229019 17229020 17229021 17229022 17229023 17229024 17229025 17229026 17229027 17229028 17229029 17229030 17229031 17229032 17229033 17229034 17229035 17229036 17229037 17229038 17229039 17229040 17229042 17229043 17229044 17229045 17229046 17229047 17229048 17229049 17229050 17229051 17229052 17229053 17229054 17229055 17229056 17229057 17229058 17229059 17229060 17229062 17229063 17229064 17229065 17229066 17229067 17229068 17229069 17229071 17229072 17229073 17229074 17229075 17229076 17229079 17229080 17229081 17229082 17229084 17229085 17229087 17229089 17229090 17229091 17229092 17229093 17229094 17229095 17229096 17229097 17229099 17229100 17229102 17229103 17229104 17229105 17229106 17229107 17229108 17229109 17229110 17229111 17229112 17229113 17229114 17229115 17229116 17229117 17229118 17229119 17229120 17229122 17229124 17229125 17229126 17229127 17229129 17229130 17229131 17229133 17229135 17229136 17229137 17229138 17229139 17229140 17229141 17229142 17229143 17229144 17229145 17229146 17229147 17229148 17229149 17229150 17229151 17229152 17229153 17229154 17229155 17229156 17229157 17229158 17229159 17229160 17229161 17229162 17229163 17229164 17229165 17229166 17229167 17229168 17229169 17229170 17229171 17229172 17229173 17229174 17229175 17229176 17229177 17229178 17229179 17229180 17229181 17229182 17229183 17229184 17229185 17229186 17229187 17229188 17229189 17229190 17229191 17229192 17229193 17229194 17229195 17229197 17229199 17229201 17229202 17229203 17229204 17229205 17229206 17229207 17229208 17229209 17229210 17229211 17229213 17229214 17229215 17229216 17229217 17229218 17229219 17229221 17229222 17229223 17229224 17229225 17229226 17229227 17229228 17229229 17229230 17229231 17229232 17229233 17229234 17229235 17229236 17229237 17229238 17229239 17229240 17229241 17229242 17229243 17229245 17229246 17229247 17229248 17229249 17229250 17229251 17229252 17229253 17229254 17229255 17229256 17229257 17229258 17229259 17229260 17229261 17229262 17229263 17229264 17229266 17229267 17229268 17229269 17229270 17229271 17229272 17229273 17229274 17229275 17229276 17229277 17229278 17229279 17229280 17229281 17229282 17229283 17229284 17229285 17229286 17229287 17229288 17229289 17229290 17229291 17229292 17229293 17229294 17229295 17229296 17229297 17229298 17229299 17229300 17229301 17229302 17229304 17229305 17229306 17229307 17229308 17229309 17229310 17229311 17229312 17229313 17229314 17229315 17229316 17229317 17229318 17229319 17229320 17229321 17229322 17229323 17229324 17229325 17229326 17229327 17229328 17229329 17229330 17229331 17229332 17229333 17229334 17229335 17229336 17229337 17229338 17229339 17229340 17229341 17229342 17229343 17229344 17229345 17229347 17229348 17229349 17229350 17229351 17229353 17229354 17229355 17229356 17229357 17229358 17229359 17229360 17229361 17229362 17229363 17229364 17229366 17229367 17229368 17229369 17229370 17229371 17229372 17229373 17229374 17229375 17229376 17229377 17229378 17229379 17229380 17229381 17229382 17229383 17229384 17229385 17229386 17229387 17229388 17229389 17229390 17229391 17229392 17229393 17229394 17229395 17229396 17229397 17229398 17229399 17229400 17229401 17229402 17229403 17229404 17229405 17229406 17229407 17229408 17229409 17229410 17229411 17229412 17229413 17229414 17229415 17229416 17229417 17229418 17229421 17229422 17229423 17229424 17229425 17229426 17229427 17229428 17229429 17229430 17229431 17229432 17229433 17229434 17229435 17229436 17229437 17229439 17229440 17229441 17229442 17229443 17229444 17229446 17229447 17229448 17229449 17229450 17229451 17229452 17229453 17229454 17229455 17229456 17229457 17229458 17229459 17229460 17229462 17229463 17229464 17229465 17229466 17229467 17229468 17229469 17229470 17229471 17229472 17229473 17229474 17229475 17229476 17229477 17229478 17229479 17229480 17229481 17229482 17229483 17229484 17229485 17229486 17229487 17229488 17229489 17229490 17229491 17229492 17229493 17229494 17229495 17229496 17229497 17229498 17229499 17229500 17229501 17229502 17229503 17229504 17229505 17229506 17229507 17229508 17229509 17229510 17229511 17229512 17229513 17229514 17229515 17229516 17229517 17229518 17229519 17229520 17229521 17229522 17229523 17229524 17229525 17229526 17229527 17229528 17229529 17229530 17229531 17229532 17229533 17229536 17229537 17229538 17229539 17229540 17229541 17229542 17229543 17229544 17229545 17229546 17229547 17229548 17229549 17229550 17229551 17229552 17229553 17229554 17229555 17229557 17229558 17229559 17229560 17229561 17229562 17229563 17229564 17229565 17229566 17229567 17229568 17229569 17229570 17229571 17229572 17229573 17229574 17229575 17229576 17229577 17229578 17229579 17229580 17229581 17229582 17229584 17229585 17229586 17229587 17229589 17229590 17229592 17229594 17229595 17229596 17229597 17229599 17229602 17229604 17229605 17229606 17229608 17229610 17229611 17229612 17229614 17229616 17229617 17229619 17229621 17229622 17229623 17229624 17229627 17229628 17229629 17229630 17229631 17229632 17229634 17229635 17229636 17229637 17229638 17229639 17229640 17229641 17229642 17229643 17229645 17229646 17229647 17229648 17230664 17230665 17230666 17230667 17230668 17230669 17230670 17230671 17230674 17230675 17230676 17230677 17230678 17230679 17230680 17230681 17230682 17230683 17230684 17230685 17230686 17230687 17230688 17230689 17230690 17230691 17230692 17230694 17230695 17230696 17230697 17230698 17230699 17230700 17230701 17230702 17230703 17230704 17230705 17230706 17230707 17230708 17230709 17230710 17230711 17230712 17230713 17230714 17230716 17230717 17230718 17230719 17230720 17230722 17230723 17230724 17230725 17230726 17230727 17230728 17230729 17230731 17230732 17230733 17230734 17230735 17230736 17230737 17230738 17230739 17230740 17230741 17230742 17230743 17230744 17230745 17230746 17230747 17230748 17230749 17230750 17235110 17235651 17235652 17235653 17235655 17235656 17235657 17235658 17235659 17235660 17235661 17235663 17235664 17235665 17235666 17235667 17235668 17235669 17235670 17235671 17235672 17235674 17235675 17235676 17235677 17235678 17235679 17235680 17235681 17235682 17235683 17235684 17235686 17235687 17235688 17235689 17235690 17235691 17235692 17235693 17235694 17235695 17235696 17235697 17235698 17235700 17235701 17235702 17235703 17235707 17235708 17235709 17235711 17235713 17235715 17235716 17235717 17235718 17235719 17235720 17235721 17235723 17235724 17235726 17235727 17235728 17235729 17235730 17235731 17235733 17235734 17235735 17235736 17235737 17235738 17235739 17235740 17235741 17235742 17235744 17235745 17235746 17235747 17235749 17235750 17235751 17235752 17235753 17235754 17235755 17235756 17235757 17235758 17235759 17235761 17235762 17235763 17235764 17235765 17235766 17235768 17235769 17235770 17235771 17235772 17235773 17235774 17235777 17235778 17235779 17235780 17235781 17235782 17235784 17235785 17235786 17235787 17235788 17235789 17235790 17235792 17235793 17235794 17235795 17235796 17235797 17235798 17235801 17235802 17235803 17235804 17235805 17235806 17235808 17235809 17235820 17235826 17235827 17235828 17235830 17235831 17235833 17235834 17235835 17235836 17235837 17235838 17235839 17235840 17235841 17235842 17235843 17235844 17235846 17235848 17235849 17235850 17235851 17235852 17235853 17235854 17235855 17235856 17235857 17235858 17235859 17235860 17235861 17235863 17235864 17235866 17235868 17235869 17235870 17235872 17235873 17235880 17235881 17235882 17235884 17235885 17235886 17235887 17235888 17235929 17235936 17235937 17235938 17235939 17235940 17235944 17235945 17235946 17235947 17235949 17235950 17235951 17235952 17235954 17235955 17235956 17235957 17235958 17235959 17235960 17235961 17235962 17235963 17235964 17235965 17235966 17235967 17235968 17235969 17235970 17235971 17235972 17235973 17235974 17235975 17235976 17235977 17235978 17235979 17235980 17235981 17235983 17235984 17235985 17235986 17235987 17235988 17235991 17235993 17235997 17235998 17235999 17236002 17236003 17236006 17236007 17236008 17236009 17236010 17236015 17236016 17236018 17236019 17236020 17236021 17236025 17236044 17236555 17237970 17237973 17237977 17237978 17237980 17237993 17237999 17238001 17238002 17238006 17238009 17238010 17238011 17238013 17238016 17238017 |
Recalling Firm/ Manufacturer |
Hospira Inc. 275 N Field Dr Lake Forest IL 60045-2579
|
For Additional Information Contact |
Mr. Chris Eustace 224-212-4892
|
Manufacturer Reason for Recall |
May go to a white screen due to a problem with the SOM2 module. A white screen is a diagnostic screen displayed to the user as part of the safety shutdown procedure of the device. The infusion and the processor stop, sound a 3 minute audible alarm from the backup buzzer and deactivate the couch screen and the automated loading/unloading of cassettes. Delay or interruption of therapy.
|
FDA Determined Cause 2 |
Device Design |
Action |
Hospira sent an Urgent Device Recall Letters, dated February 5, 2013, to all affected customers. Customers were directed to notify their consignees if they further distributed the product and confirm they have done so by return of the reply form. Healthcare professionals were to contact Stericycle. Hospira discontinued support of these devices as of June 30, 2015. Customers with questions were instructed to call 1-800-201-9064.
For questions regarding this recall call 224-212-4892. |
Quantity in Commerce |
10,337 |
Distribution |
Worldwide Distribution - US (nationwide) and Canada |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database |
510(K)s with Product Code = FRN and Original Applicant = ABBOTT LABORATORIES 510(K)s with Product Code = FRN and Original Applicant = HOSPIRA, INC.
|