Class 2 Device Recall NexGen Complete Knee Solution

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall NexGen Complete Knee Solutionsee related information
Date Initiated by FirmJune 07, 2013
Date PostedAugust 22, 2013
Recall Status1 Terminated 3 on December 15, 2014
Recall NumberZ-2002-2013
Recall Event ID 65413
Product Classification Prosthesis, knee, patellofemorotibial, semi-constrained, cemented, polymer/metal/polymer - Product Code JWH
Product00597201131 Femur CR POR SURF HDN FEM CO-NID AML Rx, Sterile; 00597201132 Femur CR POR SURF HDN FEM CO-NID AMR Rx, Sterile; 00597201231 Femur CR POR SURF HDN FEM CO-NID BML Rx, Sterile; 00597201232 Femur CR POR SURF HDN FEM CO-NID BMR Rx, Sterile; Used in total knee arthroplasty and indicated for patients with severe pain.
Code Information all codes
FEI Number 1000220733
Recalling Firm/
Manufacturer
Zimmer, Inc.
345 E Main St
Warsaw IN 46580-2746
For Additional Information ContactJaime Weeks
800-613-6131
Manufacturer Reason
for Recall
Surgeons are implanting components of the NexGen Cruciate Retaining (CR) Complete Knee System in combinations that are not approved (micro with standard). Includes CR Micro Femurs, Articular Surface, Patellae, instruments and provisionals combined with standard components. A March 2012 recall reiterated the proper combination labeling, but the mis match problem continues. Zimmer is removing the d
FDA Determined
Cause 2
Labeling design
ActionBeginning June 7, 2013 all distributors will be sent Urgent Device Recall notices via electronic mail, with distributors, hospital risk managers with affected inventory also receiving the notice via courier. The Notice outlined the products recalled; the reason for the recall, possible adverse events related to the recall and directed consignees to locate, quarantine and count all affected product recording the information on the Inventory Return Form acknowledging that recall directions were followed. Further questions should be directed to 1-877-946-2761. Affected product is to be returned to Zimmer Product Service Department , 1777 West Center St. Warsaw, IN 46580 or internation return at Zimmer International Logistics GmbH, Max - Immelmann Allee 12 79427 Eschbach Germany. " Communications outside of the United States will occur approximately two weeks after the United States communications.
Quantity in Commerce192,355 all devices
DistributionWorldwide Distribution-USA (nationwide) and the country of Japan.