| | Class 2 Device Recall GE Healthcare, Advantx, Innova 2000, Innova 2100IQ, Innova 3100, Innova 3100IQ, Innova 4100, Innova |  |
| Date Initiated by Firm | August 16, 2013 |
| Date Posted | October 17, 2013 |
| Recall Status1 |
Terminated 3 on May 30, 2014 |
| Recall Number | Z-0021-2014 |
| Recall Event ID |
66254 |
| 510(K)Number | K022322 K023178 K945375 K960575 K974367 K993037 |
| Product Classification |
Interventional fluoroscopic x-ray system - Product Code OWB
|
| Product | GE Healthcare, Advantx, Innova 2000, Innova 2100IQ, Innova 3100, Innova 3100IQ, Innova 4100, Innova 4100IQ.
Intended for general purpose diagnostic fluoroscopy and radiography studies. |
| Code Information |
Mfg Lot or Serial # System ID 00000026767YY4 201982LCA 00000068001YY7 901384LCA 00000033988YY7 914737LCA 00000034024YY0 954LCA1 00000031679YY4 708216LLCA 00000952037WK0 630856LCN 00000320794BU1 082406010002 00000346260BU3 DR-27820-CO 00000991021WK7 LB1160VA01 00000058879BU8 930047946 00000484712WK5 1453 00000529468WK1 915685LCA 00000528385WK8 082406040003 00000526149WK0 66324TC 00000497670WK0 M4017051 00000512202WK3 1616LCA 00000528587WK9 62134LCV+ 00000533252WK3 00001VAS01 00000548229WK4 506857LCA 00000088970BU9 0910063030 00000233847BU3 082406030001 00000238390BU9 083016598280384 00000245186BU2 RU1488RX20 00000249068BU8 34090VAS01 00000023887YY3 205599LCV 00000055095YY4 207283LC1 00000030139YY0 210428VCATH2 00000023954YY1 212604LC2 00000048123YY4 212604LCA2 00000048175YY4 214376VADV1 00000034312YY9 215576LC2 00000063851YY0 215662LCLP1 00000034351YY7 215955LCEP 00000034379448 216343LCA 00000026282YY4 216363LC 00000023914YY5 216384LCA 00000034355YY8 304256RCL1 00000055081YY4 304487LCA 00000048177YY0 305223LCV 00000023959YY0 305325LCA 00000063853YY6 307577HCL1 00000034352YY5 317818MDI111 00000055057YY4 317818MDI170 00000026283YY2 318462BCL 00000048183YY8 352333ANGIO 00000034327YY7 404991LCA 00000048150YY7 405271UC1 00000063859YY3 405359ELCV 00000023910YY3 407646CL1 00000034390YY5 409532GLCV 00000026316YY0 412578DLX 00000034341YY8 415540ADVLCLP 00000055053YY3 423493PSL 00000048145YY7 440585LCA 00000055092YY1 440835LCA 00000034383YY0 502587SP2 00000022054YY1 502933CL2 00000026276YY6 504380ADV1 00000048128YY3 540596LCV 00000034375YY6 561650LCA 00000030105YY1 601865ADY 00000030168YY9 615544LCA 00000034306YY1 615977SP 00000023934YY3 619340VA6 00000023871YY7 619347LC 00000030130YY9 619457SCL 00000026303YY8 619457SCL2 00000034347YY5 619739LC 00000055077YY2 661282CCLV 00000063864YY3 661326CATH1 00000030134YY1 704456LCV 00000034300YY4 707546ANGIO 00000030126YY7 717782LC1 00000034372YY3 717969LCC 00000030128YY3 718270LC 00000055052YY5 727521LCV 00000034376YY4 806358LCA 00000026272YY5 813462CATH 00000023936YY8 813586ADV1 00000055074YY9 828255MV5 00000030190443 903236LCA 00000034365YY7 904739CV2 00000048167YY1 904MHWCL 00000026290YY7 909387LCV 00000023920YY2 910343NHC1 00000026287YY3 910716BCL2 00000034391YY3 941629LCA 00000055059YY0 941798CATH1 00000048146YY5 972548CATH 00000048139YY0 CSI104 00000048156YY4 CSI106 00000026288YY1 DLPGMH1 00000022064YY0 M27036 00000039134YY1 704456LCV 0000048177YYO0 305223LCV 00000550095YY4 207283LC1 00000022059YY0 609757CCL3 00000026273YY3 813462CATH 00000030111YY9 801350LC2 00000030144YY0 908324CATH 00000022065YY7 1077 00000023879YY0 87373-1 00000023925YY1 082416170001 00000023960YY8 EG1005VA03 00000030181YY2 E4167601 00000048199YY4 M2147228 00000048102YY8 HC1251XR01 00000063868YY4 0835060003 00000063919YY5 IR1495VA01 00000063895YY7 A5111122 00000063910YY4 A5183634 00000048174YY7 A5214921 00000063921YY1 A5365407 00000063917YY9 A5367508 00000063920YY3 A5401513 00000063901YY3 C5120901 00000063877YY5 YX1122 00000063902YY1 YX1140 00000063894YY0 YX1185 00000063913YY8 YX1233 00000063912YY0 YX1314 00000063891YY6 LB1018VA01 00000026296YY4 LB1049VA02 00000030154YY9 LT4000VA02 00000034373YY1 14231LC 00000023958YY2 2235LCV 00000030151YY5 WRT0102 00000023942YY6 SA1000VA01 00000063880YY9 CE229631 00000063882YY5 FEB52037 00000063875YY9 YP629705 00000048103YY6 0853060152 00000048166YY3 0853060155 00000030115YY0 06008VAS02 00000023838YY6 215221CL1 00000063798YY3 303320CL1 00000063779YY3 303320CL2 00000063791YY8 312792RINOV2 00000022078YY0 314768SMLC 00000031479YY9 423493PCL1 00000023794YY1 516351LCA 00000048001YY2 516663LC3 00000031441YY9 603668CLC3 00000048085YY5 706354SLCA 00000031410YY4 708259NWLCA 00000048098YY8 765646CL1 00000048083YY0 770389LCA 00000063721YY5 909865LCLP 00000023830YY3 912633MCL1 00000063805YY6 BPEP 00000063736YY3 MCTHKGMN 00000022073YY1 215537LUA 00000023851YY9 501972SBL2 00000031458YY3 904739CV2 00000023773YY5 1001 00000023795YY8 1021 00000023777YY6 1025 00000023778YY4 1031 00000026180YY0 6227 00000022169YY7 86940 00000023852YY7 87650 00000023841YY0 93162 00000026217YY0 94788 00000023770YY1 1070 00000023855YY0 1021 00000026223YY8 082406120004 00000022095YY4 082406140002 00000026183YY4 082406220002 00000022088YY9 YU4006VA02 00000026260YY0 DEN27227 00000026135YY4 203325LCV 00000047978YY2 205329LCV 00000089619MO1 205592MLC 00000113966M06 219757INVA 00000026090YY1 304348LCA 00000128484M03 312CCHGECATH 00000023746YY1 314577ULCA 00000031342YY9 505848LCA 00000128617MO8 516562INNOVA4 00000128539MO4 563421CATHLAB6 00000063483YY2 601288CV3 00000034255YY0 601288CV4 00000052196YY3 601459LC 00000052199YY7 615264LCV 00000080070YY6 615769SKYCV1 00000071213YY3 615778ESPL2 00000034209YY7 718403LC 00000052138YY5 740446LCA 00000063620YY9 843797LCA 00000052178YY1 914632LCA 00000063533YY4 915649LCV 00000063506YY0 919443NGLCA 00000128531M05 973365LAB1 00000034271YY7 973429CATH 00000118197MO3 973971LAB4 00000113950MD0 973971LAB5 00000022202YY6 314768SMLC 00000023729YY7 910343NHC1 00000023737YY0 212604LC2 00000023761YY0 912633MCL1 00000026160YY2 210428VCATH2 00000034211YY3 603668CLC3 00000034276YY6 215955LCEP 00000047915YY4 415540ADVLCLP 00000047992YY3 212604LCA2 00000089632MO4 205592MLC 00000113966MO6 219757INVA 00000128484MO3 312CCHGECATH 00000063486YY5 X2881402 00000043979YY4 213977LCV 00000043972YY9 215952LCA 00000043989YY3 216778LC2 00000043976YY0 217443ULCV 00000043975YY2 617636LCA 00000043971YY1 619644LC 00000043918YY2 734769SP 00000043915YY8 864455GMLCV 00000043997YY6 920232CATH1 00000043920YY8 010120RX35 00000063978YY1 CS1019VA01 00000043970YY3 EG1023VA02 00000063954YY2 M2080316 00000043954YY7 HC5651XR06 00000043923YY2 HC6499XR01 00000043962YY0 PC6834XR03 00000043931YY5 PM0177XR03 00000063979YY9 H1600VAS05 00000063962YY5 A5189320 00000063969YY0 A5401314 00000043988YY5 YX1131 00000043986YY9 YX1190 00000063956YY7 YX1230 00000043985YY1 YX1308 00000063966YY6 YX1355 00000063960YY9 YX1362 00000063980YY7 YX1369 00000063981YY5 YX1394 00000063985YY6 YX1396 00000063983YY1 YX1438 00000063967YY4 BRD0409 00000043949YY7 PER42000 00000043900YY0 SA1003VA03 00000063971YY6 YP170A12 00000033908YY5 201365LCA 00000026763YY3 201625LCA 00000071104YY4 212604EP1 UNK21320850001 212639LCA 00000071115YY0 228575LCA 00000045551YY9 303450LCA 00000033948YY1 303673LCA 00000047646YY5 303695LCA 00000033964YY8 305682LCA 00000034000YY0 313421LCA 00000034031YY5 314933JMA 00000026716YY1 318388CLCA 00000047612YY7 318675LCA1 00000033980YY4 318681SLCA1 00000047645YY7 501484LC2 00000068006YY6 502587LCV6 00000033918YY4 502875LCA 00000033998YY6 504230LCA 00000047604YY4 504873TSP 00000068036YY3 513870LCA 00000026691YY6 518761GLCA 00000071140YY8 561798LCA 00000033968YY9 601234LCA 00000047686YY1 601268WSP1 00000047676YY2 603432PSP 00000034017YY4 606252LCA 00000047626YY7 606268LCA UNK21320850000 704384LCA2GS 00000026726YY0 706647LCA 00000034029YY9 717346LCA 00000025866YY5 718876LCA 00000033989YY5 718963LCA 00000047684YY6 727341LCA 00000033951YY5 757889RM9 00000033911449 801350LCA 00000025823YY6 804281ANGIO 00000026693YY2 813462ADV1 00000045572YY5 850494WSP2 00000026758YY3 864455GMSR9 00000068003YY3 901661LCA 00000047608YY5 901726M2LCA 00000047621YY8 908527LCA 00000034032YY3 909243AVTX 00000033930YY9 912757LCA 00000047507YY7 CSI105 00000057366YY7 CSI202 00000057385YY7 CSI203 00000068017YY3 CSI204 00000068018YY1 CSI205 00000080217YY3 CSI209 00000080240YY5 CSI210 00000080218YY1 CSI214 00000047687YY9 ECXV1407CLCV 00000034009YY1 336293LCA1 00000071140YYB 561798LCA 00000024987YY0 602258TILT 00000025012YY6 708259NWLCA 00000025828YY5 314577ULCA 00000025853YY3 205877BSPXXX 00000026690YY8 304348LCA 00000026755YY9 505848LCA 00000033911YY9 801350LCA 00000033961YY4 718403LC 00000033975YY4 404991LCA 00000034021YY6 601288CV4 00000034033YY1 973429CATH 00000034049YY7 216343LCA 00000045599YY8 CSI104 00000047647YY3 914632LCA 00000047655YY6 CSI106 00000047669YY7 601459LC 00000047683YY8 919443NGLCA 00000047693YY7 601288CV3 00000057371YY7 740446LCA 00000071116YY8 843797LCA 00000033945YY7 17883ADVLC 00000045573YY3 0364060001 00000080220YY7 35272 00000033997YY8 11597 00000080201YY7 17815 00000047639YY0 23040 00000045561YY8 29306 00000047633YY3 31249 00000045570YY9 40869 00000068007YY4 43035 00000068043YY9 48776 00000045571YY7 49840 00000071139YY0 56719 00000068040YY5 85480 00000068039YY7 85544 00000071123YY4 90690 00000034001YY8 99299 00000068014YY0 74704-1 00000024999YY5 416522LCA 00000071134YY1 082406020001 00000071168YY9 082406020002 00000068046YY2 082406040005 00000071178YY8 082406040020 00000047668YY9 082406050004 00000006800YY9 082406070001 00000071161YY4 082406090002 00000071171YY3 082406090005 00000068026YY4 082406100004 00000071137YY4 082406120001 00000068034YY8 082406120005 00000057382YY4 082406120007 00000071186YY1 082406120009 00000071133YY3 082406140008 00000047662YY2 082406160001 00000047698YY6 082406160007 00000071113YY5 082406160008 00000071183YY8 082406190002 00000047643YY2 082406210002 00000057397YY2 082406210005 00000026771YY6 082406210008 00000047674YY7 082406220003 00000033900YY2 082406250001 00000071148YY1 082406270001 00000071167YY1 082416010001 00000071136YY6 082416160001 00000068041YY3 082416200001 00000068000YY9 082406070001 00000026765YY8 M4486917 00000047653YY1 X2881402 00000033957YY2 GE1000VA01 00000071143YY2 H2077VAS04 00000071157YY2 H2077VAS05 00000068009YY0 H2197VAS09 00000068035YY5 083006592809757 00000034040YY6 IR1063VA02 00000047618YY4 IE1031VA01 00000033937YY4 A5140610 00000071122YY6 YX1128 00000071147YY3 YX1148 00000071159YY8 YX1183 00000071160YY6 YX1184 00000057372YY5 LB1071VA01 00000047610YY1 0834060006 00000047666YY3 MQ1008VA01 00000071173YY9 600123VA01 00000033926YY7 RU1041VA01 00000026769YY0 RU1298VA01 00000047677YY0 RU1556VA01 00000071188YY7 RU1734VA01 00000026768YY2 NE222872 00000034020YY8 NP335806 00000047644YY0 0853060157 00000057374YY1 0853060161 00000071172YY1 0853060164 00000071191YY1 0853060168 00000047601YY0 0853060162 00000057398YY0 0853060157 00000071176YY2 38824624VASOCC 00000080202YY5 38831231VASCTR 00000031708YY1 1751 00000031703YY2 502793G5 00000031740YY4 516968LCV 00000031683YY6 615781LCA 00000031650YY5 203325LCV 00000031655YY4 909387LCV 00000031707YY3 516663LC3 00000031711YY5 212604LCA2 00000031736YY2 615264LCV 00000031738YY8 423493PSL 00000031686YY9 23767 00000031723YY0 604631LCA 00000031728YY9 EG1102VA01 00000031672YY9 NO1000VA01 00000031670YY3 PER18103 00000031676YY0 A020RX06 00000234478BU6 212746L511 00000225366BU4 YX1456 00000000824MC3 303436LCLP 00000120399M01 303778CL1 00000120399MO1 303788SP3 00000120418MO9 412647LABD 00000131446M07 415600LCALP 00000120404MO9 504897ADV3 00000000113MC1 516663EP2 00000000793MC0 609757LCEP 00000131448MO3 615284CV1 00000131456MO6 617525LCN 00000000190MC9 619543UC6 00000000287MC3 717763EP1 00000120440MO3 718780EP 00000120441MO1 806212SPEC 00000120407MO2 815226INN1 00000000132MC1 901726SP1 00000000188MC3 904739EP 00000000565MC2 BPANGIO 00000000288MC1 FHORLCL6 00000000089MC3 563421CATHLAB6 00000000562MC9 BPEP 00000120425MO4 312CCHGECATH 00000000186MC7 010120RX40 00000120403MO1 418656CLCLP 00000131447MO5 A1105902 00000000718MC7 B1607012 00000000156MC0 A5109211 00000000095MC0 A5109212 00000000563MC7 A5267808 00000000794MC8 C5821416 00000120411MO4 YX1449 00000120437MO9 YX1458 00000000434MC1 YX1462 00000120396MO7 YX1515 00000000217MC0 YX1520 00000000719MC5 YX1700 00000120438MO7 LB1040VA01 00000131450MO9 MA1070VA01 00000000203MC0 MA1272VA01 00000120442MO9 0853060175 00000131451MO7 0853060176 00000000388MC9 0853060180 00000000157MC8 0853060182 00000000158MC6 0853060182 00000120431MO2 00243VAS01 00000289997BU9 YX1192 00000305717BU1 SA1039VA01 00000273885BU4 DLP069 00000252139BU1 1322875LCV 00000258324BU3 H2197VAS10 00000265561BU1 A5162231 00000266866BU3 B5820845 00000277196BU2 317818MDI925 00000294681BU2 082406040007 00000305325BU3 082406040010 00000293076BU6 082406090004 00000282496BU9 082406110002 00000294682BU0 082406150001 00000284078BU3 082406190001 00000302110BU2 082406240001 00000272431BU8 EG1914VA01 00000294680BU4 083016824542X 00000302112BU8 083006592018634 00000297990BU4 IR1113VA02 00000272430BU0 IR1379VA02 00000282504BU0 IE1004VA02 00000289984BU7 YX1136 00000297988BU8 YX1186 00000297987BU0 LB1080VA01 000000000PS004 606252CL 00000008679YY1 212604CATH3 00000086338YY1 215955CL1 00000086307TT6 301698LCV 00000086342YY3 304636LCA 00000086365YY4 305549LCVP 00000086081YY7 313593LCA 00000086094YY0 316343L1 00000086318YY3 337494LC3 00000086316YY7 405945CVI 00000086272YY2 409564NLC 00000086337YY3 412232C2 00000086336YY5 607763CL3 00000086327YY4 607763EP 00000086278YY9 610250SC 0000008625BYY1 615769SKYANG2 00000086062YY7 708352LC3 00000086375YY3 732363LCA 00000086343YY1 815741SJLC 00000086259YY9 817255LC 00000086098YY1 865541LCA 000000000PS003 970495HCLA2 00000086092YY4 E21894 00000086323YY3 MCTH003AK 00000086076YY7 617636LCA 00000086079YY1 212604CATH3 00000086264YY9 615769SKYCV1 00000086299YY5 509474LCA 00000086307YY6 301698LCV 00000086329YY0 708938WLCA 00000086344YY9 914632LCA 00000086380YY3 978741LCA1 00000086091YY6 514468BLCA 00000086085YY8 082406050006 00000086070YY0 92539ADV 00000086313YY4 EG1703VA01 00000086352YY2 FI1023VA01 00000086350YY6 FI1060VA01 00000086275YY5 M4217025 000000000PS002 PC1499XR01 00000086306YY8 HU1167RX01 00000086355YY5 083006590088383 00000086281YY3 083006590208491 00000086292YY0 083006590208529 00000086290YY4 083006592018583 00000086314YY2 083006592830240 00000086305YY0 083016590040041 00000086270YY6 IR1379VA01 00000086297YY9 A5401320 00000086073YY4 0850060409 00000086304YY3 LT4000VA03 00000086317YY5 00455VAS01 00000128096MO5 202269CATH 00000118093MO4 205933VACL2 00000118085MO0 301279CATH1 00000109962MO1 303839ANGIO2 00000000592MC6 304843LCA 00000128064MO3 312CCHANGIO1 00000128065MO0 312CCHANGIO2 00000128062MO7 312CCHANGIO3 00000128131MO0 314768M10 00000118046MO2 318363LC 00000118077MO7 330452LCA 00000109982MO9 360456XCL2 00000000711MC2 361788LCPLUS 00000109953MO0 417347FCV3 00000128082MO5 501978LAB1 00000109963MO9 502893LC 00000127515MO5 503681LCV 00000000637MC9 505887CL1 00000109979MO5 513559LCA 00000109961MO3 518580LCV 00000128063MO5 540437LCA 00000118064MO5 561589LCA 00000000219MC6 561863SP1 00000128071MO8 580765VCLCV 00000000145MC3 601445LCV 00000128055M01 601824LCA 00000128066MO8 606678CATH 00000000677MC5 609757EP1 00000128104MO7 614257LCA 00000127528MO8 615222STCL8 00000128127MO8 618241LCA 00000000696MC5 650369SV2 00000000583MC5 713359INA 00000000162MC8 713359INB 00000000679MC1 714670LCA 00000109950MO6 717621LCA 00000128103MO9 724537SP 00000128073MO4 757259LCV 00000109980MO3 757934LCA1 00000128130MO2 773282LCV 00000128126MOO 773564LCA 00000127499MO2 773792REP 00000127494MO3 785354SP1 00000128126MO0 801532LCA 0000000128124M 818884LCA 00000128106M02 828286RHLCA 00000127537MO9 901346CV 00000127511MO4 901346SP 00000127520MO5 901516SP1 00000127527MO0 914666LCA 00000000593MC4 918307SHIN1 00000127497MO6 941473LCA 00000118088MO4 941798LCA 00000118045MO4 954978LCA 00000127532MO0 972969LCA 00000109947MO2 ECXV1407ELCV 00000000135MC4 609757LCEP 00000127535MO3 612863CV80 00000128052MO8 609757LCEP 00000128106MO2 828286RHLCA 00000128124MO5 818884LCA 00000109978MO7 BG4029VA01 00000000906MC8 519663LCP 00000127531MO2 326582ANGI1 00000000684MC1 082406020003 00000127510MO6 082406030005 00000127516MO3 082406030006 00000127540MO3 082406030007 00000128050MO2 082406030008 00000000148MC7 082406030009 00000000683MC3 082406030011 00000128140MO1 082406030012 00000128060MO1 082406040017 00000128086MO6 082406040018 00000000929MC0 082406040024 00000128045MO2 082406050007 00000127500MO7 082406070004 00000127539MO5 082406070005 00000128094MO0 082406070011 00000118052MO0 082406080002 00000127504MO9 082406090010 00000000215MC4 082406090011 00000000729MC4 082406090012 00000000927MC4 082406090015 00000000928MC2 082406090022 00000128048MO6 082406100009 00000128136MO9 082406100013 00000128142MO7 082406100014 00000128144MO3 082406100015 00000000686MC6 082406100016 00000128141MO9 082406110003 00000000636MC1 082406120015 00000128118MO7 082406120017 00000000160MC2 082406130014 00000000143MC8 082406130015 00000000685MC8 082406140019 00000127509MO8 082406150005 00000000146MC1 082406160036 00000128143MO5 082406160037 00000000926MC6 082406160038 00000128100MO5 082406180002 00000128135MO1 082406190009 00000000954MC8 082406190012 00000128056MO9 082406200001 00000000137MC0 082406200002 00000000730MC2 082406200003 00000000147MC9 082406230010 00000128117MO9 082406230011 00000128129MO4 082406270004 00000128093MO2 082406270005 00000000678MC3 082406300001 00000128115MO3 082406310002 00000000801MC1 082406310004 00000128051M00 082416020003 00000109967MO0 082416090001 00000118044MO7 082416100007 00000128085MO8 CS1118VA02 00000128138MO5 EG1021VA03 00000118082MO7 EG1221VA01 00000128092MO4 FI1021VA01 00000127524MO7 A4194813 00000109970MO4 C2195248 00000000214MC7 M2536003 00000118072MO8 M4031077 00000109942MO3 M4181455 00000118089MO2 M4190056 00000118050MO4 M5484925 00000000121MC4 M6414704 00000128049MO4 M9196032 00000000731MC0 M4486917 00000127522MO1 PC4049XR01 00000128059MO3 GP1024VA01 00000000122MC2 08301699537802 00000118051MO2 083006592815007 00000118053MO8 083016598000584 00000109985MO2 083016598067428 00000000681MC7 083016602101193 00000127501MO5 0835060002 00000000220MC4 0835060023 00000128089MO0 IR1632VA02 00000000101MC6 A5251114 00000128113MO8 A5375128 00000000695MC7 A5687002 00000000591MC8 B5381808 00000109969MO6 YX1367 00000127496MO8 YX1448 00000128084MO1 YX1506 00000000149MC5 YX1507 00000000732MC8 YX1525 00000000191MC7 YX1543 00000128077MO5 0850060734 00000128111MO2 KW1001VA02 00000118069MO4 LB1000VA03 00000118083MO5 LB1064VA01 00000128123MO7 LY1013VA01 00000000198MC2 MA1346VA01 00000127541MO1 PK1001VA01 00000127506M04 0843060002 00000127542MO9 PL1465VA01 00000000130MC5 PER43218 00000000150MC3 RU1057VA01 00000128119MO5 RU1379VA01 00000127508MO0 CEE15649 00000128122MO9 NE185034 00000109941MO5 NP083616 00000109987MO8 0853060170 00000128120MO3 0853060178 00000000638MC7 0853060181 00000128081MO7 TN1003VA01 00000000635MC3 01002VAS02 00000000509MC0 34004VAS01 00000127536MO1 48003VAS01 00000000955MC5 65005VAS01 00000128044MO5 78001VAS01 00000127523MO9 38808698VASMET 000000000PS004 610954LCA 00000125275MO8 215361LCA 00000128256MO5 215576LCLP 00000128246M06 256494SP 00000086119YY5 301877LCA 00000089400M06 303839ANGIO1 00000000546MC2 312942EP1 00000000726MC0 314525ACL1A 00000086420YY7 330841LCA2 00000089401MO4 361985LCA 00000086411YY6 406728LCA 00000086416YY5 410535LCA 00000086128YY6 410871LCA 00000086140YY1 509474LCA 00000086135YY1 520622LCA 00000113671MO2 520779LCA 00000086413YY2 610327LCA 00000128244MO1 612863DSAB 00000125286MO5 704384LCA1 00000000183MC4 718670LCA 00000113683MO7 727521LCA 00000113693MO6 732324LCA 00000086422YY3 803775LCA 00000125255MO0 803776LCA 00000113696MO9 808485ANGIO 00000086444YY7 843347LCA 00000000290MC7 864560SPLCA 00000089366MO9 909865INNOVA1 00000089358MO6 909865INNOVA2 00000128262MO3 914366RM5 00000089397MO4 914368CATH 00000125256MO8 970244RM7 00000000209MC7 ECXV1407JLCA 00000113697MO7 LIJ016133 00000125257M06 LVEPINOV5 00000000207MC1 516562INNOVA4 00000000588MC4 212639LCA 00000086123YY7 205329LCV 00000086132YY8 410871LCA 00000086149YY2 205592MLC 00000089398MO2 914368CATH 00000089402MO2 973971LAB4 00000113670MO4 219757INVA 00000113685MO2 504897ADV3 00000128257MO3 973365LAB1 00000128288MO8 601445LCV 00000086112YY0 49638ADVXPF 00000000291MC5 4406 00000086137YY7 23048 00000086148YY4 26009 00000113681MO1 73934 00000125251MO9 76011 00000000703MC9 2524297 00000113682MO9 52207BR 00000086126YY0 514468BLCA 00000113710MO8 519685LCAP 00000089362MO8 082406010003 00000086405YY8 082406010005 00000000120MC6 082406010006 00000125245MO1 082406030003 00000125264MO2 082406030004 00000113717MO3 082406040013 00000125263MO4 082406040015 00000128277MO1 082406040019 00000125274MO1 082406040023 00000000698MC1 082406040026 00000089378MO4 082406060001 00000113719M09 082406070003 00000125272MO5 082406070006 00000000555MC3 082406070009 00000000116MC4 082406080003 00000089370MO1 082406090007 00000113706MO6 082406090008 00000089369MO3 082406090009 00000089359M04 082406100005 00000128245MO8 082406100008 00000125279MO0 082406100010 00000128280MO5 082406110001 00000089361MO0 082406120010 00000086441YY3 082406120011 00000113674MO6 082406120012 00000113678M07 082406120013 00000125244MO4 082406120014 00000128248MO2 082406130011 00000089383MO4 082406140012 00000113684MO5 082406140013 00000113672MO0 082406140014 00000113673MO8 082406140015 00000125260MO0 082406140016 00000086448YY8 082406150002 00000085445YY4 082406150003 00000125266MO7 082406150004 00000125280MO8 082406150006 00000125247MO7 082406150008 00000086418YY1 082406160032 00000113716MO5 082406160033 00000125268M03 082406160034 00000128249MO0 082406160035 00000128281MO3 082406180003 00000113703MO3 082406190004 00000125259MO2 082406190005 00000113704MO1 082406190006 00000125258MO4 082406190007 00000125261MO8 082406190008 00000089363MO6 082406210006 00000125271MO7 082406210009 00000125262MO6 082406220004 00000089393MO3 082406230006 00000125281MO6 082406230008 00000000118MC0 082406230009 00000089379MO2 082406250004 00000089394MO1 082406270002 00000125269M06 082406220004 00000086445YY4 082406150003 00000113678MO7 082406120013 00000113719MO9 082406070003 00000125268MO3 082406160034 00000089357MO8 EG1112VA02 00000086146YY8 M2427732 00000125252MO7 M4196634 00000086407YY4 M2080316 00000086145YY0 HC0335XR08 00000089377MO6 24541491LC 00000086111YY2 93336LCV 00000113695MO1 HU1099VA01 00000086152YY6 083016598000582 00000113707MO4 A5120519 00000125246MO9 A5139622 00000128282MO1 A5158022 00000113713MO2 B5111120 00000128268MO0 C5123609 00000086120YY3 YX1198 00000086147YY6 YX1218 00000086110YY4 YX1312 00000086129YY4 YX1315 00000086415YY7 YX1351 00000089367MO7 YX1359 00000113718MO1 YX1397 00000113689MO4 YX1442 00000125269MO1 YX1504 00000000117MC2 YX1508 00000128291MO2 YX1512 00000000701MC3 YXZ471 00000000091MC9 YX0982 00000000586MC8 KA1048VA01 00000086423YY1 LT4006VA01 00000086412YY4 5272LCPLUS 00000000554MC6 NG1018VA01 00000086124YY5 PL1248VA01 00000128295MO3 RU1047VA01 00000128264MO9 RU1086VA03 00000113699MO3 RU1095VA02 00000128278MO9 RU1111VA01 00000086127YY8 RU1733VA01 00000128255MO7 RU2091VA01 00000000119MC8 RU2400VA01 00000113694MO4 NE222886 00000089365MO1 NP083616 00000125277MO4 TN1003VA01 00000125287MO3 TN1169VA01 00000086443YY9 42002VAS01 00000113711MO6 45001VAS01 00000125250MO1 00584VAS03 00000125253MO5 00584VAS04 00000128296MO1 388555220VASMET 00000086107YY0 38897734VASMET 00000014912VE0 251435ICV1 00000011118VE7 310423INNOVA5 00000322726BU1 336716BCL1 00000015657VE0 414805INNOVA2 00000018525VE6 082416040005 00000013405VE6 A4194812 00000025032VE4 M1754003 00000025025VE8 08301603046202 00000014926VE0 08301628448937 00000013334VE8 083016590020020 00000011622VE8 083016592416686 00000018518VE1 0835160019 00000015655VE4 A5306808 00000010621VE1 YX1344 00000353682BU8 YX1347 00000010969VE4 YX1348 00000018524VE9 YX1468 00000018532VE2 YX1470 00000011128VE6 YX1500 00000018522VE3 LT4001RX01 00000010954VE6 869910INNOVA 00000026276VE6 0835160016 00000041504VE2 YX1482 00000041489VE6 YX1485 00000041484VE7 YX1488 00000039462VE7 207755CL1 00000039446VE0 317818I2K905 00000083327VE7 401274IN3 00000074873VE1 423495CHIRM1 00000065697VE5 610988INNOVA1 00000065691VE8 1157760 00000083383VE5 082416030004 00000074876VE4 082416040007 00000061624VE3 082416040008 00000110181VE5 082416040013 00000079546VE8 082416050001 00000095439VE6 082416160004 00000068468VE8 HC7659XR02 00000074901VE0 083016002100406 00000055080VE6 083016242100883 00000099728VE8 083016242101355 00000104773VE7 083016602200096 00000110174VE0 083016602200116 00000083323VE6 083016602200304 00000087023VE8 083016802100484 00000099719VE7 083016802100854 00000083320VE2 083016802101285 00000110182VE3 083016822100515 00000055070VE7 083016822100563 00000099727VE0 083016842101276 00000110180VE7 083016842101946 00000068484VE5 IR1667VA01 00000052658VE2 YX1483 00000055072VE3 YX1484 00000055071VE5 YX1486 00000039442VE9 0850060674 00000039475VE9 0850060688 00000052650VE9 0850060693 00000068482VE9 0850060709 00000074897VE0 0850060711 00000087011VE3 0850060713 00000099714VE8 0850060715 00000039452VE8 0850060745 00000063253VE9 809840INNOVA 00000074893VE9 0856160001 00000104776VE0 0856160006 00000099713VE0 16016VAS01 00000083324VE4 00285VAS02 00000296175BU3 615769SKYANG1 00000316452BU2 972566ANGIO1 00000313555BU5 416340LCLP 00000317105BU5 083006598280992 00000312523BU4 D5810512 00000313557BU1 0850060437 00000313556BU3 0850060735 00000323058BU8 SA1008VA01 00000307150BU3 0853060165 00000361496BU3 310423LCLP6 00000364921BU7 312CCHANGIO4 00000347810BU4 443481BIPLANE1 00000339403BU8 540776LGVAS 00000354571BU2 757668ANGIO2 00000336748BU9 LIJ015064 00000333315BU0 082406250005 00000330267BU6 HC2563XR06 00000330265BU0 YX1235 00000333314BU3 YX1241 00000345964BU1 YX1243 00000355009BU2 YX1257 00000349682BU5 YX1450 00000364912BU6 YX1452 00000355013BU4 YX1454 00000352680BU3 0853060173 00000439618BU0 082406130016 00000405225BU4 YX1465 00000424091BU7 YX1675 00000405222BU1 SA1114VA01 00000364924BU1 34101VAS02 00000453231BU3 YX1466 00000433768BU9 YX2222 00000433769BU7 0853060186 00000467420BU6 0853060192 00000306073BU8 317818MDI920 00000309133BU7 18891 00000306071BU2 082406030002 00000289987BU0 083006590040140 00000314412BU8 083016598002063 00000324192BU4 083016598289160 00000313561BU3 YX1226 00000320787BU5 YX1353 00000338124BU1 405272SLC2 00000349685BU8 BG4029VA03 00000338115BU9 082406040012 00000329467BU5 082406100006 00000346286BU8 082406100007 00000351639BU0 082406210007 00000352861BU9 082406270003 00000342524BU6 082416200003 00000349691BU6 HR4008VA02 00000376600BU3 08301603023002 00000362381BU6 08301603024062 00000338123BU3 08301606216203 00000361865BU9 08301622006008 00000338116BU7 083016590670230 00000327700BU1 083016598280910 00000352860BU1 IR1569VA01 00000325938BU9 YX1352 00000349686BU6 YX1393 00000349693BU2 YX1395 00000355947BU3 YX1398 00000351636BU6 YX1437 00000354778BU3 YX1440 00000364607BU2 YX1446 00000327701BU9 UK1063VA01 00000331246BU9 IE1560VA01 00000425798BU6 2516621 00000429607BU5 082406090014 00000429608BU3 082406140018 00000394129BU1 08301603125002 00000392100BU4 08301659001325 00000427136BU7 083016602200133 00000414596BU7 083016802100873 00000434604BU5 083016802200044 00000395929BU3 IR1631VA01 00000382284BU8 A5306813 00000380661BU9 YX1501 00000403911BU1 YX1509 00000421900BU2 YX1516 00000383925BU5 0850060609 00000421897BU0 MA1276VA01 00000414598BU3 RU2076VA01 00000385715BU8 34101VAS01 00000465764BU9 2584562 00000445339BU5 082406050008 00000445905BU3 082406090019 00000445903BU8 082406090020 00000441881BU0 082406130017 00000449892BU9 082406190013 00000352550BU8 305585CL2 00000377865BU1 HC2434XR01 00000354848BU4 563516INNOVA 00000383127BU8 95664INNOVA 00000389098BU5 DLP072 00000397404BU5 537527INNOVA 00000419071BU6 SY1007VA01 00000423555BU2 SY2011VA01 00000434301BU8 DZ1027VA01 00000445944BU2 0850060714 00000449636BU0 0856160005 00000008547GP2 50252 00000997077WK3 GP1046VA01 00000004320GQ8 RU1086RX05 00000030435BU2 502587LCV6 00000304358BU5 502875LCA 00000304353BU6 X2881402 00000304372BU6 082406190003 00000109929MO0 281540NECATH 00000109922MO5 316268INNOVAB 00000128826M05 321PBCL1 00000109921M07 404712EI1 00000109891MO2 410337INNOVA3 00000109919MO1 505983CATH 00000114889MO9 513853INNOVA 00000000PROTO7 614293INNOVA1 00000000PROTO8 614293INNOVA2 00000109915M09 617754IN1 00000109938MO1 713794LAB3 000000000PROTO 714229INNOVA 00000128748M01 801299INNOVA 000000PROTO 14 973365CATH 00000000PROTO9 ECXV1406C 00000109896MO1 909865INNOVA2 00000109911MO8 909865INNOVA1 00000109915MO9 617754IN1 00000109921MO7 404712EI1 00000109906MO8 082416040002 00000109907MO6 082416040003 00000114888MO1 083016590670229 00000114891MO5 JO1034VA01 00000109904MO3 00105VAS01 00000146340M05 561548CL2 00000146329MO8 864455GMIN1 00000128823MO2 864455GMIN2 00000128779MO6 205783BCL3 00000128692MO1 207626CL1 00000128649MO1 213726CATH 00000128653MO3 213977INNOVA4 00000146335MO5 214320INNOVA 00000128717MO6 214590INNOVA 00000128829MO9 215448INNOVA 00000120219MO1 215861LC2 00000000517MC3 217258CATH 00000000173MC5 217464INNOVA 00000128741MO6 229259INOV1 00000128795M02 229386INOV1 00000000367MC3 251435ICV4 00000120204MO3 253627INNOVA2 00000146341U03 254519I2K 00000128811MO7 262569INNOVA 00000000175MC0 281338INA 00000128736MO6 281440LAB3 00000128757MO2 303425HCL1 00000146368M06 303715CL1 00000120235MO7 305223IN1 00000128775MO4 305223IN2 00000128687MO1 305325CCL3 00000128648MO3 305585CL3 00000128690MO5 309691PINNOVA 00000128672MO3 310423INNOVA7 00000128743M02 313343LC 00000128694MO7 314569JC1 00000120230MO8 316689INNOVA1 00000128812MO5 317621INNOVA1 00000128816MO6 317621INNOVA2 00000128777MO0 317818I2K110 00000128676MO4 317818I2K120 00000120221MO7 317818MDI915 00000000129WE7 317887INNOVA2 00000128726MO7 321HCL4 00000128732MO5 321HCL5 00000000374MC9 352344CATH1 00000128708MO5 402280CIN1 00000128685MO5 402280CIN2 00000000163MC6 404265AMC2 00000000176MC8 404305SF1 00000128716MO8 404686CRI1 00000128787MO9 404686CRI2 00000146370MO2 410543CATH1 00000128734MO1 412232B 00000000179MC2 417269SCLB 00000000169MC3 417269SCLC 00000128693MO9 417885JCL1 00000128750MO7 417885JCL2 00000128730MO9 417885JCL3 00000128762MO2 417885JCL4 00000128830MO7 419526INNOVA3 00000128825MO7 419526INNOVA4 00000128662MO4 423493PCL3 00000146336MO3 423559BCL1 00000000320MC2 480512IN3 00000000499MC4 508856IN3 00000000429MC1 508856IN4 00000000500MC9 508856IN5 00000000323MC6 508856IN6 00000146339MO7 509473DR7 00000120233MO2 509946CATH1 00000120212M06 512447SI 00000128837MO2 520296CATH1 00000000128MC9 530244XC3 00000128807M05 561548CL1 00000128740MO8 561798CATH1 00000128802MO6 563421CATHLAB2 00000128801MO8 563421CATHLAB4 00000128749MO9 601968INNOVA 00000128737MO4 602588INNOVA 00000146326MO4 603228CCL2 00000128681M04 603650CATH3 00000120205MO0 610431CL1 00000120203MO5 610431CL2 00000000502MC5 610645CL5 00000128673MO1 610648INNOVA 00000152515MO3 610954INNOVA 00000128828MO1 618241GSI 00000000513MC2 626915INNOVA 00000128840MO6 702388INNOVA1 00000128821MO6 706651CCL 00000120200MO1 708333INOVA 00000000501MC7 713794LAB10 00000146337MO1 714533INNOVA 00000128751MO5 714953INNOVA 00000128764MO6 717763INNOVA 00000000504MC1 717763INNOVA2 00000128793MO7 717782INNOVA 00000128833MO1 724284CL 00000128701MO0 732557INNOVA 00000128733MO3 740439CATH 00000128824MO0 740454INNOVA1 00000122739MO0 740454INNOVA2 00000146331MO4 760940INNOVA1 00000128682MO2 770991INNOVA 00000128699MO6 773884HC2000 00000128644MO2 785354HRT1 00000128671MO5 785354HRT2 00000128827M03 801479INNOVA 00000128651MO7 804285CATH4 00000000161MC0 804320CATH1A 00000000126MC3 804320CATH3A 00000128786MO1 804320CATH5 00000152521MO1 804320CATH6 00000128818MO2 817329BGINNOVA 00000128839MO8 817347MPINNOVA 00000000164MC4 817465ARINNOVA 00000000165MC1 818502INNOVA1 00000146342MO1 828262CL 00000000410MC1 828322FIN 00000012023MO4 843339INNOVA 00000128761M04 843577CL1 00000146352M00 850325INNOVA1 00000000201MC4 850325INNOVA2 00000000178MC4 850494INNOVA 00000152518MO7 850494INNOVA2 00000128710M01 860358CV 00000109918M03 864512EP2 00000000202MC2 865218CCL1 00000000180MC0 870508LAB3 00000120243MO1 901384CV3 00000120223MO3 901726R2 00000152525MO2 913317SLSHEART1 00000128760MO6 919934JCL 00000120214MO2 920738INNOVA 00000000140MC4 941483INNOVA1 00000000171MC9 941483INNOVA2 00000146324MO9 956389INNOVA 00000000125MC5 970247INNOVA 00000000128838 972969INNOVA 00000146338MO9 985882INNOVA 00000146345MO4 FHORLCL2 00000000357MC4 FLGRCATH1 00000128731M07 FWCL3772 00000000177MC6 MCTHSC1 00000000514MC0 MCTHSC2 00000128763MO0 MCTHWHS1 00000000099MC2 MOUSE 00000000358MG2 303320CL1 00000128810809 501484LC2 00000152514M06 865541CCL4X 00000428687BU8 618241GSI 00000000324MC4 773878WCL 00000000358MC2 303320CL1 00000128710MI0 860358CV 00000120232MO4 843339INNOVA 00000128731MO7 FWCL3772 00000128739MO0 740454INNOVA2 00000128743MO2 313343LC 00000128784MO6 717763INNOVA2 00000128807MO5 561548CL1 00000128814MO1 303320CL2 00000128827MO3 801479INNOVA 00000128838MO0 972969INNOVA 00000146340MO5 561548CL2 00000152514MO6 865541CCL4X 00000000153MC7 212639LCA 00000128738MO2 0910163002 00000000497MC8 0910162004A 00000146365MO2 BY4015VA01 00000146354MO6 BY4026VA01 00000120208MO4 010120RX39 00000128663MO2 030232RX07 00000128679MO8 BR62349 00000000168MC5 416284INNOVA 00000120210MO0 418656CNOVA1 00000128698MO8 418656CNOVA2 00000128650MO9 519663INNOVA 00000146367MO8 082406140017 00000128720MO0 082416010002 00000128820MO8 082416010003 00000000449MC9 082416020001 00000000085MC1 082416020002 00000128674MO9 082416030001 00000120195MO3 082416040004 00000146344MO7 082416070001 00000128728MO3 082416100001 00000146351MO2 082416100004 00000128798MO6 082416120001 00000120209MO2 082416130005 00000128822MO4 082416130006 00000000448MC1 082416130008 00000120206MO8 082416140001 00000128782MO0 082416210001 00000128752MO3 082416220002 00000003026MC2 082416220005 00000000503MC3 082416220007 00000128797MO8 082416250001 00000000107MC3 082416270001 00000120215MO9 EG1112VA01 00000146358MO7 A4213138 00000146346MO2 M2052042 00000000426MC7 M23864601 00000000123MC0 R4211476 00000128727MO5 HBKR01XR04 00000120238MO1 HC1627XR20 00000128774MO7 PC8109XR02 00000128696MO2 08301621006236 00000128691MO3 0835160013A 00000120202MO7 IE1031VA03 00000120196MO1 A5221706 00000128746MO5 A5381110 00000146359MO5 A5714604 00000128753MO1 B5120716 00000000097MC6 B5346303 00000120237MO3 YX1350 00000120236MO5 YX1443 00000128724MO2 YX1475 00000128772MO1 YX1476 00000128712MO7 YX1478 00000128669MO9 JO1034VA02 00000146361MO1 MQ1000VA01 00000000127MC1 0920169001 00000000460MC6 NO1015VA03 00000128781MO2 600043VA02 00000120600MO6 PL1107VA03 00000128658MO2 PL1494VA01 00000152531MO0 RU1070VA01 00000128654MO1 RU1112VA01 00000128756MO4 RU2096VA01 00000128758MO0 RU2096VA01 00000128706MO9 SA1231VA01 00000152529MO4 DE203229 00000128815MO8 FE416040 00000000447MC3 UP389019 00000120194MO6 0853060171 00000146371MD0 0853060179 00000146371MO0 0853060179 00000146364MO5 TN1012VA01 00000146362MO9 TN1166VA01 00000146363MO7 TN1193VA01 00000146372MO8 TN1194VA01 00000146353MO8 00132VAS01 00000128799MO4 00150VAS01 00000152519MO5 00174VAS01 00000128778MO8 00638VAS01 00000146355MO3 01254VAS01 00000146373M06 10010VAS01 00000128735MO8 00584VAS04 00000128755MO6 00584VAS03 00000120247MO2 216778LABA 00000120290MO2 239939CLB 00000133399MO6 252443NGIN1 00000133408MO5 301891CATH1 00000133410M01 303730CL1 00000120272M00 303778CL2 00000102284MO5 310423INNOVA3 00000120285MO2 310423INNOVA4 00000133417MO6 310540INNOVA 00000133404MO4 318449INNOVA1 00000120280MO3 412367LAB1 00000120287MO8 412367LAB3 00000120245MO6 412647LABC 00000120262MO1 412784C1 00000133401MO0 502587L9 00000120286MO0 504447INNOVA 00000133400MO2 509473DR5 00000120293MO6 612273INNOVA1 00000012025MO6 714999INNOVA 00000133423MO4 719560HCL2 00000120251MO4 727725CATH1 00000120244MO9 727869CATH1 00000120260MO5 804764INNOVA 00000133409MO3 806354INNOVA1 00000120281MO1 816985NKCI2 00000120246MO4 831462CL1 00000120254MO8 936539CL2 00000120273M08 972394INNOVA 00000120263M09 214590INNOVA 00000120267M00 303778CL1 00000120250MO6 714999INNOVA 00000120269MO6 215448INNOVA 00000133410MO1 303730CL1 00000120271MO2 514468BINNOVA 00000133407MO7 819563CINNOVA1 00000133421MO8 905895INNOVA1 00000133422MO6 082416070002 00000120292MO8 082416130007 00000120253MO0 MC2848XR02 00000120252MO2 PC2720XR02 00000120259MO7 A5101425 00000120283MO7 A51088116 00000120257MO1 A5119428 00000120277MO9 A5140730 00000133412MO7 A5196328 00000133416MO8 A5618321 00000133418MO4 A5618322 00000120289MO4 LDN0210 00000120291MO0 LWD0302 00000120278MO7 600062VA01 00000133420MO0 SA1005VA05 00000133419MO2 SA1305VA01 00000421893BU9 706802FINN 00000430040BU6 0910063039 00000424554BU4 082406120016 00000430320BU2 082416100006 00000424195BU6 082416130010 00000434467BU7 082416130012 UNK23351290001 M5480755 00000445906BU1 334286BSP 00000450033BU6 732294INOV 00000438629BU8 0910064007 00000444378BU4 0910064008 00000446690BU0 082416030003 00000437742BU0 082416040010 00000445645BU5 082416100008 00000446837BU7 082416120006 00000446438BU4 082416130015 00000445632BU3 082416160003 00000446285BU9 082416160005 00000442020BU4 082416160006 00000440696BU3 082416210003 00000439906BU9 082416310003 00000446408BU7 H1420VAS04 00000439612BU3 YV0023 00000439905BU1 YV0025 00000460068BU0 5034353100 00000467850BU4 378INV31 00000472308BU6 082406310005 00000468306BU6 082416040015 00000476010BU4 082416040016 00000468309BU0 082416130017 00000477520BU1 082416150001 00000461413BU7 082416190002 00000471686BU6 082416220009 00000467422BU2 IR1026VA02 00000458172BU4 2048000INNOVA 00000509440BU4 602239CLAB2 00000544289BU2 718780CATH3 00000523776BU3 803255CL1 00000490794BU5 0910162010 00000544287BU6 2667313 00000537030BU9 082416010007 00000469583BU9 082416040014 00000512466BU4 082416070006 00000513295BU6 082416100015 00000538573BU7 082416100017 00000511522BU5 082416110002 00000519626BU6 082416130023 00000496756BU8 082416130026 00000555009BU0 M40480167 00000500358BU7 08301600206936 00000544359BU3 08301680520687 00000520963BU0 083006202101017 00000484299BU3 083016202101036 00000540175BU7 083016282101457 00000520405BU2 083016802100866 00000520964BU8 083016802100916 00000531112BU1 083016802101197 00000529519BU1 083016822101187 00000533866BU0 083016862101407 00000488204BU9 083016862101656 00000495988BU8 083016862101916 00000483164BU0 0835160017 00000485317BU2 0835160018 00000552745BU2 0835160021 00000526250BU6 C5294640 00000499364BU8 YV0066 00000546434BU2 0850060728 00000522945BU5 644INNOVA21 00000459368BU7 0920169006 00000533873BU6 2664627INNOVA 00000555781BU4 SK1083VA01 00000478374BU2 SY1001VA02 00000537028BU3 34009VAS06 00000559111BU0 34097VAS01 00000547953BU0 3882643435RXAND 00000488205BU6 850747BR3100 00000511598BU5 0910064009 00000491715BU9 082416040017 00000496319BU5 082416070004 00000485404BU8 082416130018 00000498814BU3 082416130020 00000482712BU7 082416300001 00000499521BU3 2625365INNOVA 00000486023BU5 0850060720 00000486054BU0 0850060721 00000497494BU5 0834160003 00000493876BU7 41018VAS01 00000478506BU9 3882591464VAMET 00000464093BU4 83INNOVA1188379 00000464078BU5 83INNOVA2588441 00000441173BU2 787653INNOVA 00000453913BU6 388262854VASCTR 00000478324BU7 083016002101581 FMI12181DUMM38 619482INNOVA1 FMI12181DUMM62 706291RCL1 00000007C20246 RPINNOVA2100 00000007C20164 2706080 00000489778BU1 2648722INNOV FMI12181DUMM18 B5112507 FMI12181DUMM15 00358VAS01 00000519105BU1 3882643330VACTR 00000000823MC5 3018774100 00000000535MC5 7028804100 00000000211MC2 7278253100 00000000488MC7 213977INNOVA1 00000000644MC5 215728INNOVA 00000000661MC9 251435ICV3 00000000744MC3 256341IN1 00000000694MC0 256386IN1 00000000579MC3 256880IN1 00000000645MC2 281351TINNOVA 00000000646MC0 281964CATH 00000000609MC8 304347INNOVA 00000000647MC8 314768SMCL3 00001205/40447 314768SMOL1 00000000838MC3 317818I2K190 00000000465MC5 352376SP 00000000486MC1 414219INNOVA 00000000733MC6 419227CL3 00000000232MC9 478633INOV 00000000612MC2 480512IN1 00000000705MC4 502765LCA 00000000987MC8 502893CATH 00000000482MC0 508798INC2 00000000466MC3 512341RI 00000000480MC4 516663INNOVA1 00000000692MC4 563421CATHLAB7 00000000654MC4 573761SMC1 00000000691MC6 601268INNOVA1 00000000221MC2 603668CLCB 00000001062MC9 606327ANGIO 00000000229MC5 606833INNOVA 00000000615MC5 610270INNOVA 00000000751MC8 614293INNOVA3 00000000814MC4 614293LIBERTY2 00000000538MC9 615222STCL3 00000000607MC2 615284CV2 00000000657MC7 615327CV1 00000000939MC9 615443LCA 00000000704MC7 619457INNOVA3 00000000664MC3 636916SPCL1 00000000658MC5 702731INNOVA3 00000000750MC0 704487CMIN1 00000000539MC7 708425INNOVA1 00000000622MC1 717270INNOVA1 00000000605MC6 718206LIBERTY1 00000000603MC1 718226INNOVA1 00000000655MC1 718226INNOVA2 00000000809MC4 718918CATH1 00000000737MC7 727521CATH3 00000000606MC4 757736INNOVA 00000000901MC9 760633INNOVA 00000000752MC6 801373CATH1 00000000462MC2 812279INNOVA1 00000000723MC7 815599FHN3100 00000000608MC0 817255NH2000 00000001079MC3 832355SMIN 00000000227MC9 865541CCL2 00000000614MC8 903798WINN 00000000519MC9 918307SHIN2 00000000721MC1 940764INOV1 00000000134MC7 940764INOV2 00000000575MC1 940764INOV3 00000000225MC3 972487ANGIO1 00000000663MC5 CM2100 00000001212MC0 DUCK 00000000377MC2 3178184100920 00000001211MC2 7278253100 0000000535MC50 7028804100 00000000A01386 508798INC2 00000001077MC7 610988CRS3 00000000821MC9 190020RX09 00000000690MC8 514376CARDIO 00000000598MC3 514843AINNOVA 00000000376MC4 514934BANGIO 00000000860MC7 613548INNOVA2 00000000667MC6 082406010011 00000000477MC0 082406040022 00000001080MC1 082406060002 00000000859MC9 082406090018 00000000874MC8 082406090021 00000000843MC3 082406100017 00000000872MC2 082406140020 00000000842MC5 082406190011 00000000989MC4 082406230012 00000001060MC3 082406270006 00000000819MC3 082406290001 00000000877MC1 082416010004 00000000552MC0 082416100005 00000000542MC1 082416120002 00000000734MC4 082416120003 00000000840MC9 082416120004 00000001078MC5 082416130013 00000001367MC2 082416160007 00000000766MC6 082416220004 00000000807MC8 082416220006 00000000844MC1 082416250003 00000000875MC5 EG1003VA01 00000000604MC9 M2030707 00000000688MC2 M2208714 00000000861MC5 M4046541 00000000876MC3 M4488827 00000000551MC2 H2009VAS17 00000000617MC1 0828060057 00000001382MC1 HU1099VA01 00000000649MC4 08301603078802 00000000618MC9 083016842101033 00000000481MC2 0835160015 00000001324MC3 IR1747VA01 00000000832MC6 A5129811 00000000990MC2 A5332104 00000001336MC7 B5603933 00000001258MC3 C5603903 00000000873MC0 A5111122 00000000768MC2 YX1491 00000000228MC7 YV0001 00000000669MC2 YV0003 00000000830MC0 YV0005 00000000765MC8 YV0026 00000000521MC5 LB1050VA02 00000000903MC5 MT1014VA01 00000000648MC6 MA1255VA01 00000000812MC8 2551604LCV 00000000616MC3 PL1008VA01 00000000665MC0 600107VA01 00000000841MC7 RU1018VA03 00000000816MC9 RU1106VA02 00000000902MC7 RU1175VA01 00000000833MC4 RU1463VA01 00000000653MC6 RU1717VA01 00000000839MC1 RU2167VA01 00000000580MC1 RU1016VA01 00000000553MC8 SA1188VA03 00000000577MC7 FE421134 00000000810MC2 FEE08424 00000000815MC1 FEE12418 00000000817MC7 0856160003 00000001173MC4 0856160004 00000000212MC1 00095VAS07 00000000522MC3 00391VAS01 00000000576MC9 01245VAS01 00000000574MC4 10044VAS01 00000000937MC3 77137VAS01 00000000822MC7 388558268VASAND 00000001346MC6 713359INC 00000000746MC8 MCDCATH2 00000000799MC7 8613100 00000001309MC4 6616324100 00000000984MC5 9408984100 00000000631MC2 205783BCL4 00000000675MC9 208523CATH 00000000993MC6 209526ICATH1 00000000748MC4 214324INNOVA2 00000001236MC9 214590PH3100 00000001235MC1 214947MINNOVA 00000001326MC8 214947MINNOVA2 00000001325MC0 215335INNOVA1 00000000742MC7 216844INNOVA 00000000960MC5 251435ICV5 00000000845MC8 303839CL1 00000000761MC7 330837INNOVA 00000000336MC8 337981INNOVA 00000000632MLO 352401CATH 00000001066M00 361980INNOVA2 00000000624MC7 405272SCL4 00000001538MC8 405604D2100 00000001376MC3 405604D3100 00000001383MC9 405840OIC3100 00000000811MC0 407321INCL1 00000000992MC8 409899LAB3 00000000672MC6 410543CATH6 00000000529MC8 412623LAB1 00000000805MC2 512447SI2 00000000470MC5 608324DAL 00000001356MC5 609757CCL5 00000000998MC5 610447CL1 00000001070MC2 610447CL2 00000001075MC1 610891CL1 00000000958MC9 610954INNOVA2 00000000847MC4 615222STCL5 00000000747MC6 615316INOV2 00000000715MC3 615327CV2 00000001064MC5 615769SKYCV2 00000000671MC8 7028803100A 00000000796MC3 707546INNOVA 00000001068MC6 713897CIN 00000000444MC0 727825CATH 00000000443MC2 757889INNOVA 00000000827MC6 773STJINNOVA200 00000000596MC7 817255NH4100 00000001002MC5 817329BG2100 00000000996MC9 817433HSW3100 00000001001MC1 817433HSW4100 00000000762MC5 817922INNOVA 00000001073MC6 856342INNOVA 00000000745MC0 865481CCL1 00000000626MC7 865541CCL1 00000000525MC6 865541CCL5 00000000531MC4 901726SP2 00000000567MC8 918307SHIN4 00000000595MC9 918307SHLC1 00000000983MC7 936266CL1 00000000569MC4 941483INNOVA41 00000001271MC6 956698INN31 00000000797MC1 956994INN1 00000000982MC9 956994INN2 00000000630MC4 972335INN 00000000570MC2 972540INN 00000000144MC6 FWANG3217 00000001390MC4 504903INNOVA31 00000000572MC8 312792RINOV2 00000000634MC6 310423INNOVA7 00000000959MC7 610954INNOVA 0000000827IMO6 773STJINNOVA200 --------992MC8 409899LAB3 00000671MC8000 7028803100A 00000000335MC0 BG4037XR01 00000000764MC1 819563CINNOVA2 00000000623MC9 082416130009 00000000806MC0 082416130011 00000000625MC4 082416210002 00000001574MC3 126430XR03 00000000493MC7 UC1129XR33 00000000492MC9 H2015VAS15 00000000334MC3 A5105151 00000000848MC2 A5108235 00000000995MC1 A5108239 00000000804MC5 A5108714 00000000826MC8 A51152101 00000001003MC3 A5130609 00000000673MC4 A5189713 00000001217MC9 C5343110 00000000626MC2 0850060707 00000000741MC9 0850060708 00000000749MC2 0850060710 00000000956MC3 0850060712 00000000981MC1 477XVIN2553211 00000001072MC8 787844INNOVACL1 00000001076MC9 787844INNOVACL2 00000001321MC9 787844INNOVACL3 00000001539MC6 FP180607 00000001000MC9 TT-2577762-CO 00000000568MC6 TN1059VA01 00000001065MC2 388567772VASMET 00000001218MC7 TT-2586581-CO 00000015106YY8 2135332W40 00000018469YY7 305824AFM 00000014171YY3 507255CC101 00000004423YY0 573882VLUA 00000018479YY6 617340LC 00000016139YY8 702731LUA 00000015099YY5 810664LUA 00000014176YY2 816691NLC2 00000016117YY4 916453ADVTX 00000012900YY7 4221293ADV 00000018484YY6 0961 00000005425YY4 0977 00000012898YY3 60753 00000005428YY8 760598 00000005423YY9 760706 00000015089YY6 22105ANGI1 00000012925YY4 930021926LCV 00000018473YY9 EG1124VA01 00000016131YY5 08300659869002 00000014185YY3 RU1757VA01 00000014207YY5 3178182000935 00000018526YY4 212604LC 00000013003YY9 215785SCA 00000018531YY4 313759LCV 00000015200YY9 317818MDI905 00000016782YY5 408947OC1 00000013007YY0 501484LC1 00000015166YY2 614383LCV 00000013023YY7 615781SHCV1 00000013034YY4 636933JMLC 00000014216YY6 808547QRM8A 00000004460YY2 EG1006VA01 00000014226YY5 45289ADVANTX 00000012333GP1 561548EP 00000001946M33 918682IN41 00000985288WK0 972566INNOVA3 00000994771WK4 4105434100 00000006134M31 7023602100 00000006135M38 7023604100 00000006565M36 8184093131 00000004370GP3 214590LC10 00001003287WK8 215453INNOVA21 00000998276WK0 215955LCLP 00000004431M33 228818VASC 00000505358WK2 302994LCA 00000013745GP5 314966DCL2A 00000516425WK6 317818MDI110 00000006394GP1 318681SLCA2 00000010589GP0 336716BEP3 00000004367GP9 401444EP2 00000009163GP7 404712EP1 00000505536WK3 405272SLCA1 00000002671GP6 412232LCA 00000004628GP4 412623LABI 00000014057GP4 415206CATH 00000004338GP0 415750SCL1 00000004332GP3 415750SCL2 00000012718GP3 502587L2 00000248722WK1 503230CV3 00000497069WK5 504765LSP2 00000004625M30 508941IN1 00000004953M36 509575YM2100 00000001806GP9 570969EP 00000460042WK5 603650LCN 00000001799M36 610250INNOVA1 00000004368GP7 610270LCA 00000988177WK2 614293LIBERTY1 00000276702WK8 618242GSLC 00000993662WK6 619260INNOVA 00000000841GP7 630859LCV 00000010089GP1 651254RM4 00000006158GP0 661949INNOVA 00000001940M36 661PMC3100 00000004329GP9 702731INNOVA 00000984271WK7 706291RINN 00000000997GP7 708216LLCN2 00000003731GP7 713704HLCN 00000002660GP9 713722SBLCN 00000002249GP1 717782EP2 00000001939M38 717782INNOVA31A 00000008813GP8 773702INNOVA 00000009624GP8 775770CATH1 00000994331WK7 775982CATH3 00000004915GP5 775982SNCATH 00000511633WK0 804541ADV 00000012317GP4 806354INNOVA2 00000004334GP9 813615CL2 00000996666WK4 814534C1 00000000959GP7 814877LCA 00000281047WK1 817548ALC 00000524010WK6 817548ALCE 00000006430M33 843674VAS1 00000007949GP1 856582LCA 00000002608M38 856853IN2100 00000997571WK5 918744JB5 00000008934GP2 918744JI3 00000005083GP1 956632INOVA 00000004872GP8 972519NEURO 00000001885GP3 973972LCN 00000013434GP6 FHVIN2000 00000992460WK6 KMCINNOVA 00000005584M38 2646154 00000005119M33 2673668 00000006420GP4 416469INNOVA 00000005581M34 2696283INNOVA 00000002751M36 2655786INNOVA 00001005457WK5 2620943IN2000 00000000237GP8 A4171534 00000276637WK6 IR1062VA01 00001004932WK8 IQ1083VA02 00000002238M34 IQ1126VA01 00000004045M31 IQ1165VA01 00000002850M36 A5102671 00000012700GP1 XF0022 00000001475GP3 YX1187 00000001497GP7 YX1215 00000004725GP8 YX1245 00000007324GP7 YX1453 00000007070GP6 YX1467 00000006896GP5 YX1479 00000011331GP6 YX1487 00000003908GQ1 JO1029VA01 00000007295M39 1257060INNOVA 00000997572WK3 52XVINN2565367 00000006012GP9 54787LC 00000001076GP9 MA1274VA01 00000005568M31 787378INNOVA 00000532889WK3 38820463VASOCC 00000499005WK7 209944DADVLC 00000010180GP8 256331RCL 00000497606WK4 313593LC1 00000537087WK9 314569JRM10 00000002778GP9 314947JCL 00000443486WK4 317818MDI910 00000504181WK9 318473RCATH4 00000000710GP4 352521LCPLUS 00000308920WK8 408236KXSL 00000531725WK0 561468LCLP 00000007197GP7 601288LCN 00000309692WK2 607763CL1 00000451788WK4 619552VLC 00000004993GP2 662841SP1 00000504480WK5 801350LC1 00000457541WK1 803661LC 00000530749WK1 908321LCN 00000497165WK1 916734XR 00000319962WK7 919286LUA 00000000160GP2 CSI107 00000368468WK5 6124ADVCFM 00000418330WK7 0374060028 00000544785WK9 EG1024VA01 00000540850WK5 HC5687XR08 00000002387GP9 YX1200 00000515059WK4 YV0022 00000305721WK3 5578ARCOADVAN 00000420558WK9 RU1045VA01 000000TLK-1909 216476LAB1 000000TLK-2147 301368LUB 000000TLK-1060 415493PACL2 000000TLK-1628 417781JV1 000000TLK-1294 417881HCV1 000000TLK-1382 707544CCA 000000TLK-1736 714883LUA 000000TLK-1713 714887LU 000000TLK-1375 0045 000000TLK-3933 0885 000000TLK-4292 703147 000000TLK-3318 203872LUA 000000TLK-3466 214872NSP 000000TLK-3876 215284LUA 000000TLK-4243 215546AFM 000000TLK-3778 304429V150 000000TLK-1731 305395B44 000000TLK-4376 310673AFM 000000TLK-3588 314569JRM12 000000TLK-1680 314947JCATH 000000TLK-2456 316268SFCL 000000TLK-3900 401274MA1 000000TLK-3045 401274MV2 000000TLK-4305 401729PCL 000000TLK-4257 419381LAB1 000000TLK-2768 502562L1 000000TLK-3070 502587SP 000000TLK-3291 516542LUC 000000TLK-3563 602277D209 000000TLK-3939 619421XSP 000000TLK-3626 704255MMS2 000000TLK-3628 708383OPLC 000000TLK-4355 714827LUA 000000TLK-3402 7186155N47 000000TLK-4012 805SJAFM 000000TLK-3899 812275DLC 000000TLK-3204 904739CV1 000000TLK-4071 908381AFM 000000TLK-3697 913281BSP 000000TLK-2826 915563LUADV 000000TLK-3303 916973KSV 000000TLK-3609 E16327 000000TLK-1136 206456XRM4 000000TLK-4123 316268LB 000000TLK-4059 316268LC 000000TLK-4096 316268LD 000000TLK-2508 316268SFCL 000000TLK-3576 619453VCL 000000TLK-4010 708383OPLC 000000TLK-2716 713697PSP 000000TLK-3807 914378LUA 000000TLK-1488 915653SCL 000000TLK-3636 E16327 000000TLK-3881 51230ADVLCX 000000TLK-4070 512806ADVN 000000TLK-3052 0043 000000TLK-3892 0048 000000TLK-3929 0049 000000TLK-4335 0582 000000TLK-4216 0623 000000TLK-4103 0880 000000TLK-3452 060046 000000TLK-4050 60050 000000TLK-3654 060052 000000TLK-3764 060200 000000TLK-4137 060202 000000TLK-4075 60209 000000TLK-4166 4714REN 000000TLK-4266 4714REN 000000TLK-2726 4165212S16 000000TLK-4057 26814ANGI1 000000TLK-3827 A4181709 000000TLK-3708 0835160032 000000TLK-3732 0835160032 000000TLK-4058 IR1057VA01 000000TLK-3484 LB1047VA01 000000TLK-3551 5588ADVANTX 000000TLK-3959 64147LUADX 000000TLK-3092 NO1005VA03 000000TLK-3902 NO1005VA03 000000TLK-3978 YU4006VA01 0000000TLK3978 YU4006VA02 000000TLK-3887 FEG34312 000000TLK-3505 NE222821 000000TLK-4162 213726RM8 000000TLK-3747 215928C3 000000TLK-3492 704255MMS2 000000TLK-3901 26863ANGI1 000000TLK-4119 01002VAS01 000000TLK-1803 214877UTLU 000000TLK-1360 915653SCL 000000TLK-1441 181VASCULAR 000000TLK-2230 0376060001 00000226588WK2 309655FC1 00000215577WK8 312878WCL 00000000000866 1222 0000000TLK-396 212420LC 0000000TLK-476 212420LUA 00000000000701 213491AVTX 00000000000778 215221CL3 0000000TLK-525 215327LC 0000000TLK-643 215447EP 0000000TLK-484 215662C3 0000000TLK-526 215770C1 00000000000718 215861LC 0000000TLK-709 215928C3 0000000TLK-540 216745LUA 0000000TLK-559 216835CATH 0000000TLK-390 216837LU 0000000TLK-413 301368LUB 0000000TLK-506 301891LC 0000000TLK-651 313284LUA 0000000TLK-630 313421LC 0000000TLK-586 313745CATH2 0000000TLK-618 313967A 0000000TLK-699 316268LB 0000000TLK-710 316268LC 0000000TLK-706 316268LD 0000000TLK-497 317818MDI120 00000000000735 317818MDI150 00000000000837 317818MDI909 00000000TLK211 318448CCATH4 0000000TLK-359 318494LC1 00000000000846 318494LC2 00000000000783 404422NR5 0000000TLK-545 407896CCCATH 0000000TLK-348 414656CCLX 00000000TLK669 415428OCCL2 0000000TLK-200 417885SVA2 00000000000736 480474LC3 0000000TLK-550 502651TJCL 0000000TLK-318 513867LC 00000000000858 602258LC1 00000000000816 615778ECL1 0000000TLK-608 619543UC5 0000000TLK-320 707462LUC 00000000TLK613 714824LUA 0000000TLK-510 714887LU 00000000000724 814371ADV 00000000000677 814765ADV 0000000TLK-528 815963LC 00000000000841 904681TCVX 00000000000835 909824LC2 00000000TLK670 909824LC3 0000000TLK-448 919748BEP 00000000000727 941475CL 0000000TLK-409 M27037 00000000000748 214376VADV1 00000000000789 213491AVTX 00000000000821 561996LC1 00000000TLK660 704371PC2 000000TLK-3694 313421LC 0000000TLK-503 612576ADVX 00000000000728 0207 0000000TLK-702 060206 00000000000695 060208 00000000000694 060418 0000000TLK-565 760200 00000000000830 26863ANGI1 00000000000831 26863ANGI1 00000000000761 0835160012 00000000000741 0119060001 0000000TLK-458 600119RX01 00000000000870 RU1013VA01 00000000000611 06004VAS01 0000000TLK-611 06004VAS01 00000248979WK7 202574LUAC 00000236161WK6 504765LC2 00000249758WK4 713791SCL3X 00000234026WK3 MSIVASLC1 00000224993WK6 318473RCATH2X 00000501908WK8 318798HCATH2 00000454725WK3 402280P1 00000431831WK7 LIJ201178 00000503601WK7 MCTH006YX 00000506323WK5 0910062022 00000418480WK0 0647 00000428844WK5 0792 00000428196WK0 0814 00000437685WK1 0820 00000435118WK5 61993 00000421055WK5 6622LCDLX 00000518759WK6 MA1190VA01 00000316517WK2 0000000 00000333064WK4 417623FCV2 00000273112WK3 804862SR7A 00000336973WK3 MCTH050 00000297074WK7 MSIVASLC2 00000315101WK6 0053 00000330118WK1 060204 00000332783WK0 060205 00000319990WK8 BG4000RX01 00000387204WK1 75650ADVLCV 00000347541WK5 IR1026VA01 00000389587WK7 73959 00000519631WK6 215576LC1 00000459864WK5 RU1158VA03 No susp info 0002853879 No susp info 0002855112 No susp info 2022969CATH2 No susp info 214590CRS No susp info 215662LCLP2 No susp info 303629CL4 No susp info 316689MP No susp info 317818MDI140 No susp info 317818MDI145 No susp info 317818MDI155 No susp info 317818MDI175 No susp info 417837DX1 No susp info 440449EP No susp info 480474LC4 No susp info 507225CC101 No susp info 518387K12S35 No susp info 518387VASC1 No susp info 602258LC2 No susp info 671HVIGLCV No susp info 734426MIR2 No susp info 734426MIRGL No susp info 775770CATH3 No susp info 801566SPC No susp info 830334FCRF1 No susp info 913596GEMNET No susp info 936634WHLC No susp info AIM00902 No susp info CC68820 No susp info ECXV1407I No susp info MCTHWHS2 No susp info MEDIMGLCD No susp info NOPI00285 No susp info NOPI00286 No susp info RADNETRF1238 No susp info 2652ADVXLC No susp info 405949BLCA No susp info 7322AVTXLUC No susp info AH4300XR03 No susp info 6334 No susp info 9864 No susp info 13439 No susp info 39522 No susp info 40140 No susp info 45999 No susp info 060201 No susp info 75155 No susp info 80271 No susp info 99635 No susp info 0002686904 No susp info 0002774953 No susp info 0002830584 No susp info 0043-1 No susp info 3919-1 No susp info BG4549VA01 No susp info 23403ANGI1 No susp info 082406040027 No susp info 082406140006 No susp info 082416ATCINNOVA No susp info 3742513668 No susp info 33863SFX No susp info 07692ADVLC No susp info HR4008VA01 No susp info HR4032XR01 No susp info M2637311 No susp info HC3260XR03 No susp info 2543848ADVLC No susp info 2546301ADLCV No susp info 24516572980IN No susp info 083016592001782 No susp info 0835060001 No susp info IR1164XR01 No susp info IE6000VA01 No susp info A5360421 No susp info A5413834 No susp info BMALTA11 No susp info NXXXXX10 No susp info LY1006VA01 No susp info 33XVLCN544793 No susp info 4915LUC No susp info 64150LUC No susp info 6648LCV No susp info 8183MLCLHU No susp info 9926LCPLUS No susp info LCV+61055 No susp info 0920TESTEQUIP No susp info PK1183VA01 No susp info 21XR59740 No susp info 38223341XR No susp info 38239075LCV No susp info ADOMEDVA01 No susp info 787620IN4100 No susp info 809269LCA No susp info 809852INNOVA No susp info RO4165VA01 No susp info RU1041DM01 No susp info RU1195RX01 No susp info YU4060VA01 No susp info C99999164 No susp info DE539251 No susp info FA49RX02 No susp info FA57RX02 No susp info FA82RX02 No susp info NO003RX01 No susp info 0853060149 No susp info 34107VAS01 No susp info GE001VAS01 No susp info 10238VAS01 No susp info 388200-1XRAND No susp info 3882569347VAMET No susp info 388256938VASMET No susp info 388308XRAND No susp info 38839419VASMET |
Recalling Firm/ Manufacturer |
GE Healthcare, LLC 3000 N Grandview Blvd Waukesha WI 53188-1615
|
| For Additional Information Contact | 262-513-4122 |
Manufacturer Reason for Recall | GE Healthcare has become aware of a potential safety issue involving missing screws in the overhead Video Monitor Suspension of the Advantx, Innova 2000, Innova 2100IQ, Innova 3100, Innova 3100IQ, Innova 4100, Innova 4100IQ Cardiovascular X-ray imaging systems. |
FDA Determined Cause 2 | Process control |
| Action | Consignees were sent on 8/12/2013 a GE Healthcare "Urgent Medical device Correction" letter dated August 12, 2013. The letter was addressed to Hospital Administrators / Risk Mangers, Mangers of Radiology /Cardiology /Surgery Department and Radiologists /Cardiologists /Surgeons.
The letter described the Safety Issue, safety Instructions, Affected Product details, Product Correction and Contact Information (US 800-437-1171, Japan 0120 055 919). |
| Quantity in Commerce | 2184 (1063 US. 1121 OUS) |
| Distribution | Worldwide Distribution-USA (nationwide) including DC and Puerto Rico, and except ND, SD and VT, and the countries of ALGERIA, ARGENTINA, AUSTRALIA, AUSTRIA, BELARUS, BELGIUM, BOLIVIA, BRAZIL, CANADA, CHILE, CHINA, COLOMBIA, COSTA ARICA, CROATIA, CZECH REPUBLIC, DOMINICAN REPUBLIC, ECUADOR, EGYPT, FINLAND, FRANCE, GEORGIA, GERMANY,GREECE, GUADELOUPE, GUATEMALA, HONDURAS, HONG KONG, HUNGARY, INDIA, INDONESIA, IRAN,IRELAND, ITALY, JAMAICA, JAPAN,JORDAN, KAZKHSTAN, REPUBLIC OF KOREA, KUWAIT, LEBANON, LIBYA ARAB JAMHIRIYA, LITHUANIA, MALAYSIA MALTA, MARTINIQUE, MEXICO, MOROCCO, NETHERLANDS, NEW ZEALAND, NICARAGUA, NIGERIA, NORWAY, PAKISTAN, PANAMA, PARAGUAY, PERU, PHILIPPINES, POLAND, PORTUGAL ROMANIA, RUSSIA, SAUDI ARABIA, SERBIA, SLOVAKIA, SPAIN, SWEDEN, SYRIA, TAIWAN, THAILAND, TRINIDAD AND TOBAGO, TUNISIA, TURKEY, UKRAINE, UNITED KINGDOM, and VENEZUELA. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
| 510(K) Database | 510(K)s with Product Code = OWB 510(K)s with Product Code = OWB
|
|
|
|