| | Class 2 Device Recall EXCELSIOR SL10 PRESHAPED J 15OCM |  |
| Date Initiated by Firm | November 04, 2013 |
| Date Posted | December 13, 2013 |
| Recall Status1 |
Terminated 3 on August 27, 2014 |
| Recall Number | Z-0496-2014 |
| Recall Event ID |
66888 |
| 510(K)Number | K042568 |
| Product Classification |
Catheter, percutaneous - Product Code DQY
|
| Product | Stryker brand EXCELSIOR SL-10 PRE-SHAPED C 15OCM; Catalog Number: M0031441940; Product is manufactured by Stryker Neurovascular, Cork, Ireland and distributed by Stryker Neurovascular, Fremont, CA
Excelsior SL-10/Excelsior 1018 Microcatheter: Stryker Neurovascular Microcatheters are intended to assist in the delivery of diagnostic agents, such as contrast media, and therapeutic agents, such as occlusion coils, into the peripheral, coronary, and neurovasculature. GDC 10, 18 Coils: The Gugliemi Detachable Coil is intended for the endovascular embolization of: (1) Intracranial aneurysms, (2) Other neurovascular abnormalities such as arteriovenous malformations and arteriovenous fistula, (3) Arterial and venous embolizations in the peripheral vasculature GDC 18-Fibered VortX Shape Coil: The GDC 18-Fibered VortX Shape Coil is intended for embolization of vascular abnormalities such as (1) Arteriovenous malformations and arteriovenous fistulae of the neurovasculature; (2) GDC 18- Fibered VoliX Shape Coil is also intended for arterial and venous embolizations in the peripheral vasculature. Transform: The Stryker Neurovascular TransForm Occlusion Balloon Catheters are indicated for use in the neurovasculature to temporarily stop or control blood flow and for balloon assisted embolization of aneurysms. |
| Code Information |
17048725 |
| FEI Number |
3008853977
|
Recalling Firm/ Manufacturer |
Stryker Neurovascular 47900 Bayside Pkwy Fremont CA 94538-6515
|
| For Additional Information Contact | Irene Jaworski 510-413-2500 |
Manufacturer Reason for Recall | Products were shipped labeled with "use by" dates that exceed currently available shelf-life data. All products remain within the earliest expiration date of April 2015. |
FDA Determined Cause 2 | Software Manufacturing/Software Deployment |
| Action | Stryker Neurovascular sent an Urgent Medical Device Recall letter dated November 4, 2013, all affected users. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to immediately discontinue use and segregate the affected products. Customers with questions were instructed to contact via NeuroComplaints@stryker.com. Or Fax the response form to 1-800-876-4355.
For questions regarding this recall call 510-413-2500. |
| Distribution | Worldwide Distribution - USa (nationwide) and Internationally to:
Austria, Belgium, Czech Rep, Germany, France, Great Britain, Hungary, Italy, Lithuania, Netherlands, Norway, Poland, Slovakia, Spain, Turkey, Australia, Canada, Chile, Hong Kong, India, Korea, Malaysia, Russia and Japan |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|