• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall LOCATOR Implant Attachment System

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall LOCATOR Implant Attachment Systemsee related information
Date Initiated by FirmJanuary 16, 2014
Date PostedFebruary 06, 2014
Recall Status1 Terminated 3 on May 15, 2014
Recall NumberZ-0935-2014
Recall Event ID 67315
510(K)NumberK072878 K994257 
Product Classification Implant, endosseous, root-form - Product Code DZE
ProductLOCATOR Implant Attachment System. The LOCATOR Implant Attachment System is a universal hinge, resilient attachment for endosseous implants. The LOCATOR Implant Attachment System is designed for use with overdentures or partial dentures, retained in whole or in part, by endosseous implants in the mandible or maxilla. FDA 510(k): K994257, K072878
Code Information Part Numbers: 8875-SPI, 8875-SPI-TMS, 8876-SPI, 8876-SPI-TMS, 08877-SPI, 08877-SPI-TMS, 08878-SPI, 08878-SPI-TMS, 08879-SPI, 08880-SPI; Lot Numbers: 11052, 15306, 20665, 21351, 12025,12259, 22440, I0K7X, I0K7Y, 11963,15307,17676,18934, 25294, 25567, I0JQ7, 12258,12941,15000, 19735, 20666, I0K7Z, I0KY7, 12703, 14568, 19832, 22170, I0LCG, 11093, 13364, 14151, 15624, 20043, 25026, 13530, I0LE4, 10055, 14152, 15001, 25027, 10892, 10893.   Since the LOCATOR¿ Implant Attachment System (including the restorative components) is not sold by the Zest Anchors as a sterilized product, it does not have a designated shelf-life.
FEI Number 2023950
Recalling Firm/
Manufacturer
Zest Anchors LLC
2061 Wineridge Pl
Escondido CA 92029-1947
For Additional Information Contact
760-743-7744
Manufacturer Reason
for Recall
Zest Anchors, LLC is voluntarily recalling specific lots of LOCATOR Implant Attachment Systems because it is currently provided with a label that indicates that the final torque applied to the abutment should be 30 N-cm. It was recently noted by Zest that in order to provide an optimal performance by increasing the factor of safety between the failure load of the abutment screw and the recommend
FDA Determined
Cause 2
Unknown/Undetermined by firm
ActionZest Anchors, LLC initiated this recall by sending customers recall notification letter on 01/16/2014 via FedEx overnight. The customer notification letter, titled, "IMPORTANT: MEDICAL DEVICE VOLUNTARY RECALL NOTICE", informed customers of the recall, reason for voluntary recall, product information with part numbers and lot numbers, actions to be taken, and contact information. The customer notification letter was accompanied by a response form titled "MEDICAL DEVICE RECALL RETURN RESPONSE Acknowledgement & Receipt Form".
Quantity in Commerce3,130
DistributionDistributed in Ohio and Switzerland.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
510(K) Database510(K)s with Product Code = DZE
-
-