| | Class 2 Device Recall Composix LP with Echo |  |
| Date Initiated by Firm | April 24, 2014 |
| Date Posted | June 02, 2014 |
| Recall Status1 |
Terminated 3 on December 23, 2015 |
| Recall Number | Z-1695-2014 |
| Recall Event ID |
68124 |
| 510(K)Number | K122436 |
| Product Classification |
Laparoscope, general & plastic surgery - Product Code GCJ
|
| Product | Bard Ventralight" ST Mesh with Echo PS" 8" Circle Number: 5955800
Indicated for use in the reconstruction of soft tissue deficiencies, such as the repair of hernias. |
| Code Information |
Lot Number: HUXK0058 |
| FEI Number |
1213643
|
Recalling Firm/ Manufacturer |
Davol, Inc., Subs. C. R. Bard, Inc. 100 Crossings Blvd Warwick RI 02886-2850
|
| For Additional Information Contact | 800-556-6275 |
Manufacturer Reason for Recall | Pouch holding sterile inflation assembly and inflation adapter may be open or have a weak seal. |
FDA Determined Cause 2 | Packaging process control |
| Action | Davol, Inc., Subs. C. R. Bard, Inc.sent an Urgent Medical Device Recall letter dated April 28, 2014, to all affected consignees. The letter identified the product, the problem, and the action to be taken by the consignee. Consignees were instructed to examine their inventory and quarantine product subject to the recall, contact Davol Customer Service at 1-800-556-6275 or C.R. Bard's Medical Services and Support Department at 1-800-562-0027 for instructions on how to return the product to Davol, Consignees were also instructed to complete and return the recall effectiveness check form, and notify their customers if the product was further distributed.
For questions regarding this recall call 1-800-556-6275. |
| Distribution | Worldwide Distribution - USA (nationwide) and Internationally to AUSTRIA, BELGIUM, CANADA, CYPRUS, ESTONIA, FINLAND, FRANCE,
GERMANY, GREECE, IRELAND, ITALY, LUXEMBOURG, NETHERLANDS,
NORWAY, POLAND, PORTUGAL, SOUTH AFRICA, SPAIN, SWEDEN, SWITZERLAND, and UNITED KINGDOM |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
| 510(K) Database | 510(K)s with Product Code = GCJ
|
|
|
|