| | Class 2 Device Recall 4200 PCA |  |
| Date Initiated by Firm | March 08, 2013 |
| Create Date | July 08, 2016 |
| Recall Status1 |
Terminated 3 on November 07, 2016 |
| Recall Number | Z-2142-2016 |
| Recall Event ID |
68318 |
| Product Classification |
Pump, infusion, pca - Product Code MEA
|
| Product | The LifeCare Pump Model 4200 PCA Infuser makes possible the practical implementation of patient-controlled analgesia. The health care professional sets the prescribed dose, rates, and limits for the drug to be infused. The infuser can be set to operate in one of three modes: PCA Mode, Continuous Mode, and PCA + Continuous Mode. The LifeCare Pump Model 4200 PCA Infuser is designed with a number of mechanical, electrical, and software safeguards that prevent patient injury or drug abuse. |
| Code Information |
List No. 0F782; Serial Numbers: 100 ; 101 ; 102 ; 105 ; 105 ; 106 ; 106 ; 107 ; 108 ; 108 ; 109 ; 111 ; 112 ; 114 ; 116 ; 117 ; 119 ; 120 ; 121 ; 122 ; 123 ; 124 ; 125 ; 125 ; 126 ; 126 ; 127 ; 127 ; 128 ; 128 ; 129 ; 129 ; 130 ; 130 ; 131 ; 131 ; 132 ; 132 ; 133 ; 133 ; 134 ; 134 ; 135 ; 135 ; 136 ; 136 ; 137 ; 137 ; 138 ; 138 ; 139 ; 140 ; 141 ; 142 ; 142 ; 143 ; 143 ; 144 ; 144 ; 145 ; 146 ; 147 ; 147 ; 148 ; 148 ; 149 ; 150 ; 150 ; 151 ; 152 ; 153 ; 154 ; 154 ; 155 ; 155 ; 156 ; 157 ; 158 ; 158 ; 159 ; 160 ; 161 ; 162 ; 162 ; 163 ; 164 ; 164 ; 165 ; 165 ; 166 ; 166 ; 167 ; 167 ; 168 ; 169 ; 170 ; 171 ; 172 ; 172 ; 173 ; 173 ; 174 ; 174 ; 175 ; 175 ; 176 ; 176 ; 177 ; 177 ; 178 ; 178 ; 179 ; 180 ; 181 ; 182 ; 183 ; 183 ; 184 ; 184 ; 185 ; 186 ; 186 ; 187 ; 188 ; 188 ; 189 ; 189 ; 190 ; 191 ; 192 ; 193 ; 194 ; 195 ; 196 ; 197 ; 198 ; 199 ; 200 ; 200 ; 201 ; 201 ; 202 ; 202 ; 203 ; 203 ; 204 ; 204 ; 205 ; 205 ; 206 ; 206 ; 207 ; 207 ; 208 ; 208 ; 209 ; 209 ; 210 ; 210 ; 211 ; 212 ; 212 ; 213 ; 213 ; 214 ; 214 ; 215 ; 216 ; 216 ; 217 ; 217 ; 218 ; 218 ; 219 ; 220 ; 221 ; 222 ; 223 ; 224 ; 224 ; 225 ; 225 ; 226 ; 226 ; 227 ; 227 ; 228 ; 229 ; 229 ; 230 ; 231 ; 232 ; 233 ; 234 ; 235 ; 236 ; 237 ; 238 ; 239 ; 240 ; 241 ; 241 ; 242 ; 242 ; 243 ; 243 ; 244 ; 245 ; 245 ; 246 ; 246 ; 247 ; 247 ; 248 ; 248 ; 249 ; 250 ; 250 ; 251 ; 251 ; 252 ; 252 ; 253 ; 253 ; 254 ; 255 ; 256 ; 256 ; 257 ; 258 ; 259 ; 259 ; 260 ; 260 ; 261 ; 262 ; 263 ; 263 ; 264 ; 265 ; 265 ; 266 ; 266 ; 267 ; 267 ; 268 ; 269 ; 269 ; 270 ; 270 ; 271 ; 271 ; 272 ; 273 ; 273 ; 274 ; 274 ; 275 ; 275 ; 276 ; 277 ; 277 ; 278 ; 279 ; 279 ; 280 ; 280 ; 281 ; 281 ; 282 ; 283 ; 283 ; 284 ; 284 ; 285 ; 286 ; 287 ; 288 ; 288 ; 289 ; 290 ; 290 ; 291 ; 291 ; 292 ; 292 ; 293 ; 293 ; 294 ; 294 ; 295 ; 295 ; 296 ; 296 ; 297 ; 298 ; 299 ; 299 ; 300 ; 301 ; 301 ; 302 ; 303 ; 303 ; 304 ; 305 ; 305 ; 306 ; 307 ; 307 ; 308 ; 309 ; 310 ; 310 ; 311 ; 311 ; 312 ; 312 ; 313 ; 314 ; 315 ; 315 ; 316 ; 317 ; 318 ; 319 ; 319 ; 320 ; 321 ; 323 ; 324 ; 325 ; 326 ; 327 ; 327 ; 328 ; 329 ; 330 ; 330 ; 331 ; 332 ; 332 ; 333 ; 334 ; 335 ; 335 ; 336 ; 336 ; 337 ; 338 ; 339 ; 339 ; 340 ; 340 ; 341 ; 341 ; 342 ; 343 ; 343 ; 344 ; 345 ; 346 ; 346 ; 347 ; 347 ; 348 ; 349 ; 349 ; 350 ; 350 ; 351 ; 352 ; 353 ; 353 ; 354 ; 354 ; 355 ; 356 ; 357 ; 357 ; 358 ; 359 ; 359 ; 360 ; 360 ; 361 ; 361 ; 362 ; 362 ; 363 ; 363 ; 364 ; 364 ; 365 ; 366 ; 366 ; 367 ; 367 ; 368 ; 369 ; 370 ; 370 ; 371 ; 372 ; 372 ; 373 ; 373 ; 374 ; 374 ; 375 ; 376 ; 376 ; 377 ; 377 ; 378 ; 378 ; 379 ; 379 ; 380 ; 381 ; 382 ; 382 ; 383 ; 383 ; 384 ; 384 ; 385 ; 385 ; 386 ; 386 ; 387 ; 387 ; 388 ; 388 ; 389 ; 389 ; 390 ; 390 ; 391 ; 391 ; 392 ; 393 ; 393 ; 394 ; 394 ; 395 ; 395 ; 396 ; 396 ; 397 ; 398 ; 398 ; 399 ; 399 ; 400 ; 400 ; 401 ; 401 ; 402 ; 402 ; 403 ; 403 ; 404 ; 404 ; 405 ; 406 ; 406 ; 407 ; 407 ; 408 ; 408 ; 409 ; 409 ; 410 ; 410 ; 411 ; 411 ; 412 ; 412 ; 413 ; 413 ; 414 ; 414 ; 415 ; 415 ; 416 ; 416 ; 417 ; 417 ; 418 ; 418 ; 419 ; 420 ; 421 ; 422 ; 423 ; 424 ; 425 ; 426 ; 427 ; 428 ; 429 ; 430 ; 431 ; 432 ; 433 ; 434 ; 435 ; 436 ; 437 ; 438 ; 439 ; 440 ; 441 ; 442 ; 443 ; 444 ; 445 ; 446 ; 447 ; 448 ; 449 ; 450 ; 451 ; 452 ; 453 ; 454 ; 455 ; 456 ; 457 ; 458 ; 459 ; 460 ; 461 ; 462 ; 463 ; 464 ; 465 ; 466 ; 467 ; 468 ; 469 ; 470 ; 471 ; 472 ; 473 ; 474 ; 475 ; 476 ; 477 ; 478 ; 479 ; 480 ; 481 ; 482 ; 483 ; 484 ; 485 ; 486 ; 487 ; 488 ; 489 ; 490 ; 491 ; 492 ; 493 ; 494 ;
495 ; 496 ; 497 ; 498 ; 499 ; 500 ; 500 ; 501 ; 501 ; 502 ; 502 ; 503 ; 503 ; 504 ; 504 ; 505 ; 505 ; 506 ; 506 ; 507 ; 507 ; 508 ; 508 ; 509 ; 509 ; 510 ; 510 ; 511 ; 511 ; 512 ; 512 ; 513 ; 513 ; 514 ; 514 ; 515 ; 515 ; 516 ; 516 ; 517 ; 517 ; 518 ; 518 ; 519 ; 519 ; 520 ; 520 ; 521 ; 521 ; 522 ; 522 ; 523 ; 523 ; 524 ; 524 ; 525 ; 525 ; 526 ; 526 ; 527 ; 527 ; 528 ; 528 ; 529 ; 529 ; 530 ; 530 ; 531 ; 531 ; 532 ; 533 ; 533 ; 534 ; 534 ; 535 ; 535 ; 536 ; 536 ; 537 ; 537 ; 538 ; 538 ; 539 ; 540 ; 540 ; 541 ; 541 ; 542 ; 542 ; 543 ; 543 ; 544 ; 544 ; 545 ; 546 ; 546 ; 547 ; 548 ; 548 ; 549 ; 549 ; 550 ; 550 ; 551 ; 551 ; 552 ; 552 ; 553 ; 554 ; 555 ; 555 ; 556 ; 556 ; 557 ; 558 ; 558 ; 559 ; 559 ; 560 ; 560 ; 561 ; 561 ; 562 ; 562 ; 563 ; 563 ; 564 ; 564 ; 565 ; 565 ; 566 ; 566 ; 567 ; 567 ; 568 ; 568 ; 569 ; 570 ; 570 ; 571 ; 571 ; 572 ; 573 ; 573 ; 574 ; 574 ; 575 ; 576 ; 576 ; 577 ; 577 ; 578 ; 578 ; 579 ; 579 ; 580 ; 580 ; 581 ; 581 ; 582 ; 582 ; 583 ; 583 ; 584 ; 584 ; 585 ; 585 ; 586 ; 587 ; 588 ; 588 ; 589 ; 589 ; 590 ; 590 ; 591 ; 591 ; 592 ; 592 ; 593 ; 593 ; 594 ; 594 ; 595 ; 595 ; 596 ; 597 ; 597 ; 598 ; 598 ; 599 ; 599 ; 600 ; 600 ; 601 ; 601 ; 602 ; 602 ; 603 ; 603 ; 604 ; 604 ; 605 ; 605 ; 606 ; 607 ; 607 ; 608 ; 608 ; 609 ; 610 ; 610 ; 611 ; 612 ; 612 ; 613 ; 613 ; 614 ; 614 ; 615 ; 615 ; 616 ; 616 ; 617 ; 617 ; 618 ; 619 ; 619 ; 620 ; 620 ; 621 ; 621 ; 622 ; 622 ; 623 ; 623 ; 624 ; 624 ; 625 ; 625 ; 626 ; 626 ; 627 ; 627 ; 628 ; 628 ; 629 ; 629 ; 630 ; 630 ; 631 ; 631 ; 632 ; 632 ; 634 ; 634 ; 635 ; 636 ; 637 ; 637 ; 638 ; 638 ; 639 ; 640 ; 641 ; 641 ; 642 ; 642 ; 643 ; 643 ; 644 ; 645 ; 645 ; 646 ; 646 ; 647 ; 648 ; 648 ; 649 ; 649 ; 650 ; 650 ; 651 ; 651 ; 652 ; 653 ; 654 ; 654 ; 655 ; 656 ; 656 ; 657 ; 657 ; 658 ; 659 ; 659 ; 660 ; 660 ; 662 ; 663 ; 664 ; 666 ; 667 ; 668 ; 669 ; 670 ; 671 ; 672 ; 673 ; 674 ; 675 ; 676 ; 677 ; 678 ; 679 ; 681 ; 682 ; 683 ; 684 ; 685 ; 686 ; 687 ; 688 ; 689 ; 690 ; 691 ; 692 ; 693 ; 694 ; 696 ; 697 ; 698 ; 699 ; 700 ; 701 ; 702 ; 703 ; 704 ; 705 ; 706 ; 708 ; 709 ; 710 ; 711 ; 712 ; 713 ; 715 ; 716 ; 717 ; 718 ; 719 ; 720 ; 721 ; 723 ; 724 ; 725 ; 728 ; 729 ; 730 ; 731 ; 733 ; 734 ; 735 ; 737 ; 738 ; 741 ; 742 ; 744 ; 745 ; 747 ; 749 ; 751 ; 752 ; 753 ; 754 ; 755 ; 756 ; 757 ; 758 ; 759 ; 760 ; 761 ; 762 ; 763 ; 764 ; 765 ; 766 ; 767 ; 768 ; 769 ; 770 ; 771 ; 772 ; 773 ; 774 ; 775 ; 776 ; 777 ; 778 ; 779 ; 780 ; 781 ; 782 ; 783 ; 784 ; 785 ; 786 ; 787 ; 788 ; 789 ; 791 ; 793 ; 794 ; 795 ; 797 ; 798 ; 799 ; 800 ; 803 ; 804 ; 805 ; 806 ; 807 ; 808 ; 809 ; 810 ; 811 ; 812 ; 813 ; 814 ; 815 ; 816 ; 817 ; 818 ; 819 ; 820 ; 822 ; 823 ; 824 ; 826 ; 827 ; 828 ; 829 ; 830 ; 831 ; 832 ; 833 ; 834 ; 835 ; 836 ; 837 ; 838 ; 839 ; 840 ; 841 ; 842 ; 843 ; 844 ; 845 ; 846 ; 847 ; 848 ; 849 ; 850 ; 851 ; 852 ; 853 ; 854 ; 855 ; 856 ; 857 ; 858 ; 859 ; 860 ; 861 ; 862 ; 863 ; 864 ; 865 ; 866 ; 868 ; 869 ; 870 ; 871 ; 872 ; 873 ; 874 ; 875 ; 876 ; 877 ; 879 ; 880 ; 881 ; 882 ; 883 ; 884 ; 885 ; 886 ; 887 ; 888 ; 889 ; 890 ; 891 ; 892 ; 893 ; 894 ; 895 ; 896 ; 897 ; 898 ; 899 ; 900 ; 901 ; 902 ; 903 ; 904 ; 905 ; 906 ; 907 ; 908 ; 909 ; 910 ; 911 ; 912 ; 913 ; 914 ; 916 ; 917 ; 918 ; 919 ; 920 ; 921 ; 922 ; 923 ; 924 ; 925 ; 926 ; 927 ; 928 ; 929 ; 930 ; 931 ; 932 ; 933 ; 934 ; 935 ; 936 ; 937 ; 939 ; 940 ; 942 ; 943 ; 947 ; 948 ; 950 ; 951 ; 953 ; 955 ; 958 ; 959 ; 960 ; 961 ; 962 ; 963 ; 964 ; 965 ; 966 ; 967 ; 968 ; 970 ; 971 ; 972 ; 973 ; 974 ; 976 ; 977 ; 978 ; 979 ; 980 ; 981 ; 982 ; 983 ; 984 ; 986 ; 987 ; 988 ; 990 ; 991 ; 992 ; 993 ; 994 ; 995 ; 996 ; 997 ; 998 ; 999 ; 1000 ; 1001 ; 1002 ; 1003 ; 1004 ; 1005 ; 1007 ; 1008 ; 1009 ; 1010 ; 1011 ; 1012 ; 1013 ; 1014 ; 1015 ; 1016 ; 1017 ; 1018 ; 1019 ; 1020 ; 1021 ; 1022 ; 1023 ; 1024 ; 1025 ; 1026 ; 1027 ; 1028 ; 1029 ; 1030 ; 1032 ; 1033 ; 1034 ; 1035 ; 1036 ; 1037 ; 1038 ; 1039 ; 1040 ; 1041 ; 1042 ; 1043 ; 1044 ; 1045 ; 1046 ; 1047 ; 1048 ; 1049 ; 1050 ; 1051 ; 1052 ; 1053 ; 1054 ; 1055 ; 1056 ; 1057 ; 1059 ; 1060 ; 1061 ; 1062 ; 1063 ; 1064 ; 1065 ; 1066 ; 1067 ; 1068 ; 1069 ; 1070 ; 1071 ; 1072 ; 1073 ; 1074 ; 1075 ; 1077 ; 1078 ; 1079 ; 1080 ; 1081 ; 1082 ; 1083 ; 1084 ; 1085 ; 1086 ; 1087 ; 1088 ; 1089 ; 1090 ; 1091 ; 1092 ; 1093 ; 1094 ; 1095 ; 1096 ; 1097 ; 1098 ; 1099 ; 1100 ; 1101 ; 1102 ; 1103 ; 1104 ; 1105 ; 1106 ; 1107 ; 1108 ; 1109 ; 1110 ; 1111 ; 1112 ; 1113 ; 1114 ; 1115 ; 1116 ; 1117 ; 1118 ; 1119 ; 1120 ; 1121 ; 1122 ; 1123 ; 1124 ; 1125 ; 1126 ; 1127 ; 1128 ; 1129 ; 1130 ; 1131 ; 1132 ; 1133 ; 1135 ; 1136 ; 1138 ; 1139 ; 1140 ; 1141 ; 1142 ; 1143 ; 1145 ; 1147 ; 1148 ; 1149 ; 1150 ; 1151 ; 1152 ; 1153 ; 1155 ; 1157 ; 1158 ; 1159 ; 1160 ; 1161 ; 1162 ; 1163 ; 1164 ; 1165 ; 1166 ; 1167 ; 1168 ; 1169 ; 1170 ; 1171 ; 1173 ; 1174 ; 1175 ; 1176 ; 1177 ; 1178 ; 1179 ; 1180 ; 1181 ; 1182 ; 1183 ; 1185 ; 1186 ; 1187 ; 1188 ; 1189 ; 1190 ; 1191 ; 1192 ; 1193 ; 1194 ; 1198 ; 1199 ; 1201 ; 1202 ; 1203 ; 1205 ; 1206 ; 1207 ; 1209 ; 1210 ; 1212 ; 1213 ; 1214 ; 1215 ; 1216 ; 1217 ; 1218 ; 1220 ; 1221 ; 1222 ; 1223 ; 1224 ; 1225 ; 1227 ; 1229 ; 1230 ; 1231 ; 1232 ; 1233 ; 1235 ; 1236 ; 1237 ; 1238 ; 1239 ; 1240 ; 1241 ; 1243 ; 1245 ; 1246 ; 1247 ; 1248 ; 1249 ; 1250 ; 1252 ; 1253 ; 1254 ; 1255 ; 1256 ; 1257 ; 1258 ; 1261 ; 1262 ; 1263 ; 1265 ; 1266 ; 1267 ; 1268 ; 1269 ; 1270 ; 1271 ; 1272 ; 1273 ; 1274 ; 1275 ; 1276 ; 1277 ; 1278 ; 6268 ; 10000 ; 10000 ; 10000 ; 10000 ; 10000 ; 10001 ; 10001 ; 10001 ; 10001 ; 10002 ; 10002 ; 10002 ; 10002 ; 10003 ; 10003 ; 10003 ; 10003 ; 10003 ; 10004 ; 10004 ; 10004 ; 10004 ; 10004 ; 10004 ; 10005 ; 10005 ; 10005 ; 10005 ; 10005 ; 10005 ; 10005 ; 10006 ; 10006 ; 10006 ; 10006 ; 10006 ; 10007 ; 10007 ; 10007 ; 10007 ; 10007 ; 10008 ; 10008 ; 10008 ; 10008 ; 10008 ; 10008 ; 10009 ; 10009 ; 10009 ; 10009 ; 10009 ; 10010 ; 10010 ; 10010 ; 10010 ; 10010 ; 10011 ; 10011 ; 10011 ; 10011 ; 10011 ; 10011 ; 10011 ; 10012 ; 10012 ; 10012 ; 10012 ; 10012 ; 10013 ; 10013 ; 10013 ; 10013 ; 10013 ; 10013 ; 10014 ; 10014 ; 10014 ; 10014 ; 10014 ; 10014 ; 10015 ; 10015 ; 10015 ; 10015 ; 10015 ; 10015 ; 10016 ; 10016 ; 10016 ; 10016 ; 10016 ; 10016 ; 10017 ; 10017 ; 10017 ; 10017 ; 10017 ; 10017 ; 10018 ; 10018 ; 10018 ; 10018 ; 10018 ; 10019 ; 10019 ; 10019 ; 10019 ; 10019 ; 10019 ; 10020 ; 10020 ; 10020 ; 10020 ; 10020 ; 10021 ; 10021 ; 10021 ; 10021 ; 10021 ; 10021 ; 10022 ; 10022 ; 10022 ; 10022 ; 10022 ; 10023 ; 10023 ; 10023 ; 10023 ; 10023 ; 10023 ; 10024 ; 10024 ; 10024 ; 10024 ; 10024 ; 10025 ; 10025 ; 10025 ; 10025 ; 10025 ; 10025 ; 10026 ; 10026 ; 10026 ; 10026 ; 10026 ; 10026 ; 10027 ; 10027 ; 10027 ; 10028 ; 10028 ; 10028 ; 10028 ; 10028 ; 10029 ; 10029 ; 10029 ; 10029 ; 10030 ; 10030 ; 10030 ; 10030 ; 10030 ; 10030 ; 10031 ; 10031 ; 10031 ; 10031 ; 10031 ; 10031 ; 10032 ; 10032 ; 10032 ; 10032 ; 10033 ; 10033 ; 10033 ; 10033 ; 10034 ; 10034 ; 10034 ; 10034 ; 10034 ; 10035 ; 10035 ; 10035 ; 10035 ; 10036 ; 10036 ; 10036 ; 10036 ; 10036 ; 10036 ; 10037 ; 10037 ; 10037 ; 10037 ; 10037 ; 10037 ; 10038 ; 10038 ; 10038 ; 10039 ; 10039 ; 10039 ; 10039 ; 10039 ; 10040 ; 10040 ; 10040 ; 10040 ; 10040 ; 10041 ; 10041 ; 10041 ; 10041 ; 10042 ; 10042 ; 10043 ; 10043 ; 10043 ; 10043 ; 10044 ; 10044 ; 10044 ; 10044 ; 10045 ; 10045 ; 10045 ; 10045 ; 10046 ; 10046 ; 10046 ; 10047 ; 10047 ; 10047 ; 10047 ; 10048 ; 10048 ; 10048 ; 10049 ; 10049 ; 10049 ; 10050 ; 10050 ; 10050 ; 10050 ; 10051 ; 10051 ; 10051 ; 10051 ; 10052 ; 10052 ; 10052 ; 10052 ; 10053 ; 10053 ; 10053 ; 10053 ; 10054 ; 10054 ; 10054 ; 10054 ; 10055 ; 10055 ; 10055 ; 10055 ; 10056 ; 10056 ; 10056 ; 10056 ; 10057 ; 10057 ; 10057 ; 10058 ; 10058 ; 10059 ; 10059 ; 10059 ; 10059 ; 10060 ; 10060 ; 10061 ; 10061 ; 10062 ; 10062 ; 10062 ; 10062 ; 10063 ; 10063 ; 10063 ; 10064 ; 10064 ; 10064 ; 10065 ; 10065 ; 10065 ; 10066 ; 10066 ; 10067 ; 10067 ; 10068 ; 10068 ; 10068 ; 10069 ; 10069 ; 10069 ; 10070 ; 10070 ; 10070 ; 10071 ; 10071 ; 10071 ; 10072 ; 10072 ; 10072 ; 10073 ; 10073 ; 10073 ; 10074 ; 10074 ; 10074 ; 10075 ; 10075 ; 10075 ; 10075 ; 10076 ; 10076 ; 10076 ; 10077 ; 10077 ; 10077 ; 10078 ; 10078 ; 10079 ; 10079 ; 10079 ; 10080 ; 10080 ; 10080 ; 10080 ; 10081 ; 10081 ; 10081 ; 10082 ; 10082 ; 10082 ; 10082 ; 10083 ; 10083 ; 10083 ; 10084 ; 10084 ; 10084 ; 10085 ; 10085 ; 10085 ; 10085 ; 10086 ; 10086 ; 10086 ; 10087 ; 10087 ; 10087 ; 10088 ; 10088 ; 10088 ; 10089 ; 10089 ; 10089 ; 10090 ; 10090 ; 10090 ; 10090 ; 10091 ; 10091 ; 10091 ; 10091 ; 10092 ; 10092 ; 10092 ; 10093 ; 10093 ; 10094 ; 10094 ; 10094 ; 10094 ; 10095 ; 10095 ; 10095 ; 10095 ; 10096 ; 10096 ; 10096 ; 10097 ; 10097 ; 10097 ; 10098 ; 10098 ; 10098 ; 10098 ; 10098 ; 10099 ; 10099 ; 10099 ; 10099 ; 10100 ; 10100 ; 10100 ; 10101 ; 10101 ; 10101 ; 10101 ; 10102 ; 10102 ; 10102 ; 10103 ; 10103 ; 10103 ; 10103 ; 10103 ; 10104 ; 10104 ; 10104 ; 10104 ; 10105 ; 10105 ; 10105 ; 10105 ; 10106 ; 10106 ; 10106 ; 10106 ; 10107 ; 10107 ; 10107 ; 10107 ; 10108 ; 10108 ; 10108 ; 10108 ; 10109 ; 10109 ; 10109 ; 10109 ; 10110 ; 10110 ; 10110 ; 10110 ; 10110 ; 10111 ; 10111 ; 10111 ; 10111 ; 10112 ; 10112 ; 10112 ; 10113 ; 10113 ; 10113 ; 10113 ; 10114 ; 10114 ; 10114 ; 10115 ; 10115 ; 10115 ; 10115 ; 10116 ; 10116 ; 10116 ; 10117 ; 10117 ; 10118 ; 10118 ; 10118 ; 10119 ; 10119 ; 10119 ; 10119 ; 10120 ; 10120 ; 10121 ; 10121 ; 10121 ; 10121 ; 10122 ; 10122 ; 10122 ; 10123 ; 10123 ; 10123 ; 10123 ; 10124 ; 10124 ; 10124 ; 10125 ; 10126 ; 10126 ; 10126 ; 10127 ; 10127 ; 10128 ; 10128 ; 10128 ; 10129 ; 10129 ; 10129 ; 10130 ; 10130 ; 10131 ; 10131 ; 10131 ; 10132 ; 10132 ; 10132 ; 10132 ; 10133 ; 10133 ; 10133 ; 10133 ; 10134 ; 10134 ; 10134 ; 10134 ; 10135 ; 10135 ; 10135 ; 10136 ; 10137 ; 10137 ; 10137 ; 10138 ; 10138 ; 10138 ; 10138 ; 10139 ; 10139 ; 10139 ; 10140 ; 10140 ; 10141 ; 10141 ; 10141 ; 10142 ; 10142 ; 10143 ; 10143 ; 10143 ; 10144 ; 10145 ; 10145 ; 10145 ; 10146 ; 10146 ; 10146 ; 10146 ; 10147 ; 10147 ; 10148 ; 10148 ; 10149 ; 10149 ; 10150 ; 10150 ; 10150 ; 10151 ; 10151 ; 10151 ; 10152 ; 10152 ; 10152 ; 10153 ; 10153 ; 10154 ; 10154 ; 10154 ; 10155 ; 10155 ; 10156 ; 10156 ; 10156 ; 10157 ; 10157 ; 10158 ; 10158 ; 10158 ; 10158 ; 10159 ; 10159 ; 10159 ; 10160 ; 10160 ; 10160 ; 10160 ; 10161 ; 10161 ; 10161 ; 10162 ; 10162 ; 10162 ; 10163 ; 10163 ; 10163 ; 10164 ; 10164 ; 10164 ; 10165 ; 10165 ; 10165 ; 10166 ; 10166 ; 10166 ; 10167 ; 10167 ; 10167 ; 10168 ; 10168 ; 10169 ; 10169 ; 10169 ; 10169 ; 10170 ; 10170 ; 10171 ; 10171 ; 10171 ; 10172 ; 10172 ; 10172 ; 10172 ; 10173 ; 10173 ; 10173 ; 10174 ; 10175 ; 10176 ; 10176 ; 10177 ; 10177 ; 10177 ; 10178 ; 10178 ; 10178 ; 10178 ; 10179 ; 10179 ; 10179 ; 10179 ; 10180 ; 10180 ; 10180 ; 10181 ; 10181 ; 10181 ; 10182 ; 10182 ; 10183 ; 10183 ; 10184 ; 10184 ; 10184 ; 10185 ; 10185 ; 10185 ; 10186 ; 10186 ; 10187 ; 10187 ; 10187 ; 10188 ; 10188 ; 10189 ; 10189 ; 10190 ; 10190 ; 10190 ; 10191 ; 10192 ; 10192 ; 10193 ; 10193 ; 10194 ; 10194 ; 10195 ; 10195 ; 10195 ; 10195 ; 10196 ; 10196 ; 10196 ; 10197 ; 10197 ; 10197 ; 10198 ; 10198 ; 10198 ; 10199 ; 10199 ; 10200 ; 10200 ; 10200 ; 10201 ; 10201 ; 10201 ; 10202 ; 10202 ; 10203 ; 10203 ; 10203 ; 10204 ; 10204 ; 10205 ; 10205 ; 10206 ; 10206 ; 10206 ; 10207 ; 10207 ; 10207 ; 10208 ; 10208 ; 10209 ; 10209 ; 10209 ; 10210 ; 10210 ; 10211 ; 10211 ; 10211 ; 10212 ; 10212 ; 10212 ; 10213 ; 10213 ; 10213 ; 10214 ; 10214 ; 10215 ; 10215 ; 10216 ; 10216 ; 10216 ; 10217 ; 10217 ; 10218 ; 10218 ; 10219 ; 10220 ; 10220 ; 10221 ; 10221 ; 10222 ; 10222 ; 10222 ; 10223 ; 10223 ; 10223 ; 10224 ; 10224 ; 10224 ; 10224 ; 10225 ; 10225 ; 10225 ; 10226 ; 10226 ; 10226 ; 10226 ; 10227 ; 10227 ; 10227 ; 10228 ; 10228 ; 10229 ; 10229 ; 10230 ; 10230 ; 10230 ; 10231 ; 10231 ; 10231 ; 10232 ; 10232 ; 10232 ; 10233 ; 10233 ; 10233 ; 10234 ; 10234 ; 10234 ; 10235 ; 10235 ; 10235 ; 10236 ; 10236 ; 10236 ; 10236 ; 10237 ; 10237 ; 10237 ; 10237 ; 10238 ; 10238 ; 10238 ; 10239 ; 10239 ; 10239 ; 10240 ; 10240 ; 10240 ; 10241 ; 10241 ; 10241 ; 10242 ; 10242 ; 10242 ; 10243 ; 10243 ; 10244 ; 10244 ; 10244 ; 10244 ; 10245 ; 10245 ; 10245 ; 10246 ; 10246 ; 10247 ; 10247 ; 10247 ; 10248 ; 10248 ; 10248 ; 10249 ; 10249 ; 10249 ; 10249 ; 10250 ; 10250 ; 10250 ; 10250 ; 10251 ; 10251 ; 10251 ; 10252 ; 10252 ; 10252 ; 10253 ; 10253 ; 10254 ; 10254 ; 10254 ; 10255 ; 10255 ; 10255 ; 10256 ; 10256 ; 10256 ; 10257 ; 10257 ; 10257 ; 10258 ; 10258 ; 10258 ; 10259 ; 10259 ; 10259 ; 10259 ; 10260 ; 10260 ; 10260 ; 10260 ; 10261 ; 10261 ; 10261 ; 10261 ; 10262 ; 10262 ; 10262 ; 10263 ; 10263 ; 10264 ; 10264 ; 10265 ; 10265 ; 10265 ; 10266 ; 10266 ; 10266 ; 10267 ; 10267 ; 10267 ; 10267 ; 10268 ; 10268 ; 10268 ; 10268 ; 10269 ; 10269 ; 10269 ; 10269 ; 10270 ; 10270 ; 10270 ; 10271 ; 10271 ; 10271 ; 10272 ; 10272 ; 10272 ; 10273 ; 10273 ; 10274 ; 10274 ; 10274 ; 10274 ; 10275 ; 10275 ; 10275 ; 10275 ; 10276 ; 10276 ; 10276 ; 10276 ; 10277 ; 10277 ; 10277 ; 10278 ; 10278 ; 10278 ; 10278 ; 10279 ; 10279 ; 10279 ; 10280 ; 10280 ; 10280 ; 10280 ; 10281 ; 10281 ; 10281 ; 10282 ; 10282 ; 10282 ; 10283 ; 10283 ; 10283 ; 10283 ; 10284 ; 10284 ; 10284 ; 10284 ; 10285 ; 10285 ; 10285 ; 10285 ; 10286 ; 10286 ; 10286 ; 10287 ; 10287 ; 10287 ; 10287 ; 10288 ; 10288 ; 10288 ; 10289 ; 10289 ; 10289 ; 10290 ; 10290 ; 10290 ; 10290 ; 10291 ; 10291 ; 10291 ; 10292 ; 10292 ; 10293 ; 10293 ; 10293 ; 10294 ; 10294 ; 10294 ; 10295 ; 10295 ; 10295 ; 10295 ; 10296 ; 10296 ; 10296 ; 10297 ; 10297 ; 10297 ; 10298 ; 10298 ; 10298 ; 10299 ; 10299 ; 10299 ; 10300 ; 10300 ; 10300 ; 10301 ; 10301 ; 10302 ; 10302 ; 10302 ; 10302 ; 10303 ; 10303 ; 10303 ; 10304 ; 10304 ; 10305 ; 10305 ; 10305 ; 10306 ; 10306 ; 10306 ; 10307 ; 10307 ; 10307 ; 10307 ; 10308 ; 10308 ; 10308 ; 10309 ; 10309 ; 10309 ; 10310 ; 10310 ; 10311 ; 10311 ; 10311 ; 10312 ; 10312 ; 10312 ; 10312 ; 10313 ; 10313 ; 10314 ; 10314 ; 10314 ; 10315 ; 10315 ; 10315 ; 10315 ; 10316 ; 10316 ; 10316 ; 10317 ; 10317 ; 10317 ; 10317 ; 10318 ; 10318 ; 10318 ; 10319 ; 10319 ; 10320 ; 10320 ; 10321 ; 10321 ; 10322 ; 10322 ; 10323 ; 10324 ; 10324 ; 10324 ; 10325 ; 10325 ; 10326 ; 10327 ; 10327 ; 10328 ; 10328 ; 10328 ; 10329 ; 10329 ; 10330 ; 10330 ; 10331 ; 10332 ; 10332 ; 10333 ; 10334 ; 10334 ; 10335 ; 10336 ; 10336 ; 10337 ; 10337 ; 10337 ; 10339 ; 10340 ; 10340 ; 10341 ; 10341 ; 10342 ; 10342 ; 10342 ; 10343 ; 10343 ; 10344 ; 10344 ; 10345 ; 10346 ; 10346 ; 10347 ; 10348 ; 10348 ; 10349 ; 10349 ; 10350 ; 10350 ; 10351 ; 10352 ; 10353 ; 10353 ; 10354 ; 10354 ; 10354 ; 10355 ; 10355 ; 10355 ; 10356 ; 10356 ; 10357 ; 10357 ; 10357 ; 10358 ; 10358 ; 10359 ; 10359 ; 10360 ; 10360 ; 10361 ; 10361 ; 10361 ; 10363 ; 10363 ; 10364 ; 10365 ; 10365 ; 10366 ; 10367 ; 10367 ; 10368 ; 10369 ; 10369 ; 10370 ; 10370 ; 10371 ; 10371 ; 10372 ; 10372 ; 10373 ; 10373 ; 10374 ; 10375 ; 10375 ; 10376 ; 10376 ; 10377 ; 10377 ; 10378 ; 10379 ; 10379 ; 10380 ; 10381 ; 10382 ; 10383 ; 10384 ; 10385 ; 10386 ; 10387 ; 10388 ; 10389 ; 10390 ; 10390 ; 10391 ; 10391 ; 10394 ; 10395 ; 10396 ; 10397 ; 10398 ; 10399 ; 10400 ; 10401 ; 10401 ; 10402 ; 10402 ; 10403 ; 10404 ; 10405 ; 10406 ; 10407 ; 10407 ; 10408 ; 10409 ; 10410 ; 10410 ; 10411 ; 10411 ; 10412 ; 10413 ; 10414 ; 10415 ; 10415 ; 10416 ; 10417 ; 10417 ; 10418 ; 10419 ; 10419 ; 10420 ; 10421 ; 10422 ; 10422 ; 10423 ; 10424 ; 10424 ; 10425 ; 10426 ; 10426 ; 10427 ; 10427 ; 10428 ; 10429 ; 10430 ; 10430 ; 10431 ; 10432 ; 10433 ; 10434 ; 10434 ; 10435 ; 10436 ; 10437 ; 10438 ; 10438 ; 10439 ; 10440 ; 10440 ; 10441 ; 10441 ; 10442 ; 10443 ; 10443 ; 10444 ; 10444 ; 10445 ; 10446 ; 10447 ; 10448 ; 10449 ; 10449 ; 10450 ; 10451 ; 10452 ; 10453 ; 10454 ; 10455 ; 10455 ; 10456 ; 10457 ; 10458 ; 10459 ; 10460 ; 10461 ; 10461 ; 10462 ; 10462 ; 10463 ; 10464 ; 10465 ; 10466 ; 10467 ; 10467 ; 10468 ; 10469 ; 10469 ; 10470 ; 10470 ; 10471 ; 10471 ; 10472 ; 10472 ; 10473 ; 10473 ; 10474 ; 10474 ; 10475 ; 10476 ; 10477 ; 10478 ; 10479 ; 10479 ; 10480 ; 10481 ; 10482 ; 10483 ; 10484 ; 10484 ; 10485 ; 10485 ; 10486 ; 10486 ; 10487 ; 10487 ; 10488 ; 10488 ; 10489 ; 10490 ; 10490 ; 10491 ; 10491 ; 10492 ; 10492 ; 10493 ; 10493 ; 10494 ; 10495 ; 10495 ; 10496 ; 10497 ; 10498 ; 10498 ; 10499 ; 10500 ; 10500 ; 10501 ; 10501 ; 10502 ; 10503 ; 10504 ; 10504 ; 10505 ; 10506 ; 10507 ; 10508 ; 10509 ; 10509 ; 10510 ; 10510 ; 10511 ; 10511 ; 10512 ; 10512 ; 10513 ; 10514 ; 10514 ; 10515 ; 10515 ; 10516 ; 10516 ; 10517 ; 10517 ; 10518 ; 10518 ; 10519 ; 10519 ; 10520 ; 10521 ; 10522 ; 10524 ; 10525 ; 10526 ; 10526 ; 10527 ; 10527 ; 10528 ; 10528 ; 10529 ; 10530 ; 10530 ; 10533 ; 10533 ; 10534 ; 10534 ; 10535 ; 10536 ; 10536 ; 10537 ; 10538 ; 10539 ; 10539 ; 10540 ; 10540 ; 10541 ; 10541 ; 10542 ; 10543 ; 10544 ; 10545 ; 10546 ; 10547 ; 10547 ; 10548 ; 10548 ; 10549 ; 10549 ; 10550 ; 10551 ; 10552 ; 10553 ; 10553 ; 10554 ; 10555 ; 10556 ; 10557 ; 10558 ; 10558 ; 10559 ; 10560 ; 10561 ; 10561 ; 10562 ; 10563 ; 10564 ; 10565 ; 10566 ; 10567 ; 10568 ; 10569 ; 10570 ; 10571 ; 10571 ; 10572 ; 10573 ; 10573 ; 10574 ; 10574 ; 10575 ; 10575 ; 10576 ; 10576 ; 10577 ; 10578 ; 10579 ; 10580 ; 10580 ; 10581 ; 10581 ; 10582 ; 10582 ; 10583 ; 10584 ; 10585 ; 10585 ; 10586 ; 10587 ; 10587 ; 10588 ; 10589 ; 10590 ; 10591 ; 10591 ; 10592 ; 10593 ; 10594 ; 10595 ; 10595 ; 10596 ; 10597 ; 10597 ; 10598 ; 10598 ; 10599 ; 10599 ; 10600 ; 10600 ; 10601 ; 10602 ; 10602 ; 10603 ; 10603 ; 10604 ; 10605 ; 10605 ; 10606 ; 10607 ; 10608 ; 10608 ; 10609 ; 10609 ; 10610 ; 10610 ; 10611 ; 10611 ; 10612 ; 10612 ; 10613 ; 10614 ; 10614 ; 10615 ; 10616 ; 10616 ; 10617 ; 10618 ; 10619 ; 10619 ; 10620 ; 10621 ; 10622 ; 10623 ; 10623 ; 10624 ; 10624 ; 10625 ; 10626 ; 10626 ; 10627 ; 10628 ; 10628 ; 10629 ; 10629 ; 10630 ; 10631 ; 10631 ; 10632 ; 10633 ; 10633 ; 10634 ; 10634 ; 10635 ; 10635 ; 10636 ; 10637 ; 10637 ; 10638 ; 10638 ; 10639 ; 10640 ; 10640 ; 10641 ; 10641 ; 10642 ; 10642 ; 10643 ; 10643 ; 10644 ; 10645 ; 10646 ; 10647 ; 10648 ; 10649 ; 10650 ; 10651 ; 10652 ; 10653 ; 10654 ; 10655 ; 10656 ; 10657 ; 10658 ; 10659 ; 10660 ; 10661 ; 10662 ; 10663 ; 10664 ; 10665 ; 10666 ; 10667 ; 10668 ; 10669 ; 10670 ; 10671 ; 10672 ; 10674 ; 10675 ; 10676 ; 10677 ; 10678 ; 10679 ; 10680 ; 10681 ; 10682 ; 10683 ; 10684 ; 10685 ; 10686 ; 10687 ; 10688 ; 10689 ; 10690 ; 10691 ; 10692 ; 10693 ; 10694 ; 10695 ; 10696 ; 10697 ; 10698 ; 10699 ; 10700 ; 10701 ; 10702 ; 10703 ; 10704 ; 10706 ; 10707 ; 10708 ; 10709 ; 10710 ; 10711 ; 10712 ; 10713 ; 10714 ; 10715 ; 10716 ; 10717 ; 10718 ; 10719 ; 10721 ; 10722 ; 10723 ; 10724 ; 10725 ; 10726 ; 10727 ; 10728 ; 10729 ; 10730 ; 10731 ; 10732 ; 10733 ; 10734 ; 10735 ; 10736 ; 10737 ; 10738 ; 10739 ; 10740 ; 10741 ; 10742 ; 10743 ; 10744 ; 10745 ; 10746 ; 10747 ; 10748 ; 10749 ; 10750 ; 10751 ; 10752 ; 10753 ; 10754 ; 10755 ; 10756 ; 10757 ; 10758 ; 10759 ; 10760 ; 10761 ; 10762 ; 10763 ; 10764 ; 10765 ; 10766 ; 10767 ; 10768 ; 10769 ; 10770 ; 10771 ; 10772 ; 10773 ; 10774 ; 10776 ; 10778 ; 10779 ; 10780 ; 10781 ; 10782 ; 10783 ; 10784 ; 10785 ; 10786 ; 10787 ; 10788 ; 10789 ; 10790 ; 10791 ; 10792 ; 10793 ; 10794 ; 10795 ; 10796 ; 10797 ; 10798 ; 10799 ; 10800 ; 10801 ; 10802 ; 10803 ; 10804 ; 10805 ; 10806 ; 10807 ; 10808 ; 10809 ; 10810 ; 10811 ; 10812 ; 10813 ; 10814 ; 10815 ; 10816 ; 10817 ; 10818 ; 10819 ; 10820 ; 10821 ; 10822 ; 10823 ; 10824 ; 10825 ; 10826 ; 10827 ; 10828 ; 10829 ; 10830 ; 10831 ; 10832 ; 10833 ; 10835 ; 10836 ; 10837 ; 10838 ; 10839 ; 10840 ; 10841 ; 10842 ; 10843 ; 10844 ; 10845 ; 10846 ; 10847 ; 10848 ; 10849 ; 10850 ; 10851 ; 10852 ; 10853 ; 10854 ; 10855 ; 10856 ; 10857 ; 10858 ; 10859 ; 10860 ; 10861 ; 10862 ; 10863 ; 10864 ; 10865 ; 10866 ; 10867 ; 10868 ; 10869 ; 10870 ; 10871 ; 10872 ; 10873 ; 10874 ; 10875 ; 10876 ; 10877 ; 10878 ; 10879 ; 10880 ; 10881 ; 10882 ; 10883 ; 10885 ; 10886 ; 10887 ; 10888 ; 10890 ; 10892 ; 10893 ; 10894 ; 10895 ; 10896 ; 10897 ; 10898 ; 10899 ; 10900 ; 10901 ; 10903 ; 10904 ; 10905 ; 10907 ; 10909 ; 10910 ; 10911 ; 10912 ; 10914 ; 10915 ; 10916 ; 10917 ; 10918 ; 10920 ; 10922 ; 10923 ; 10924 ; 10926 ; 10929 ; 10930 ; 10931 ; 10932 ; 10933 ; 10935 ; 10936 ; 10937 ; 10938 ; 10939 ; 10940 ; 10941 ; 10943 ; 10944 ; 10945 ; 10946 ; 10947 ; 10948 ; 10949 ; 10950 ; 10951 ; 10953 ; 10954 ; 10956 ; 10957 ; 10958 ; 10959 ; 10960 ; 10961 ; 10962 ; 10963 ; 10964 ; 10965 ; 10966 ; 10967 ; 10968 ; 10969 ; 10970 ; 10971 ; 10972 ; 10973 ; 10974 ; 10975 ; 10976 ; 10978 ; 10979 ; 10980 ; 10981 ; 10982 ; 10983 ; 10985 ; 10986 ; 10987 ; 10988 ; 10989 ; 10990 ; 10991 ; 10993 ; 10994 ; 10995 ; 10996 ; 10997 ; 10998 ; 10999 ; 11000 ; 11001 ; 11002 ; 11003 ; 11004 ; 11005 ; 11006 ; 11007 ; 11008 ; 11009 ; 11010 ; 11011 ; 11012 ; 11013 ; 11014 ; 11015 ; 11016 ; 11017 ; 11018 ; 11019 ; 11020 ; 11021 ; 11022 ; 11023 ; 11024 ; 11025 ; 11026 ; 11027 ; 11028 ; 11029 ; 11030 ; 11031 ; 11032 ; 11033 ; 11034 ; 11035 ; 11036 ; 11037 ; 11038 ; 11039 ; 11040 ; 11041 ; 11042 ; 11043 ; 11044 ; 11045 ; 11046 ; 11047 ; 11048 ; 11049 ; 11050 ; 11051 ; 11052 ; 11053 ; 11054 ; 11055 ; 11056 ; 11057 ; 11058 ; 11059 ; 11060 ; 11061 ; 11062 ; 11063 ; 11064 ; 11065 ; 11066 ; 11067 ; 11068 ; 11069 ; 11071 ; 11072 ; 11073 ; 11074 ; 11075 ; 11076 ; 11077 ; 11078 ; 11079 ; 11080 ; 11082 ; 11083 ; 11084 ; 11085 ; 11086 ; 11087 ; 11088 ; 11089 ; 11090 ; 18024 ; 40757 ; 40861 ; 102745 ; 102805 ; 85670037 ; 95590001 ; 95590002 ; 95590003 ; 95590004 ; 95590005 ; 95590006 ; 95590007 ; 95590008 ; 95590009 ; 95590010 ; 95590011 ; 95590012 ; 95590013 ; 95590014 ; 95590015 ; 95590016 ; 95590017 ; 95590018 ; 95590019 ; 95590020 ; 95590021 ; 95590022 ; 95590023 ; 95590024 ; 95590025 ; 95590026 ; 95590027 ; 95590028 ; 95590029 ; 95590030 ; 95590031 ; 95590032 ; 95590033 ; 95590034 ; 95590035 ; 95590036 ; 95590037 ; 95590038 ; 95590039 ; 95590040 ; 95590041 ; 95590042 ; 95590043 ; 95590044 ; 95590045 ; 95590046 ; 95590047 ; 95590048 ; 95590049 ; 95590050 ; 95590051 ; 95590052 ; 95590053 ; 95590054 ; 95590055 ; 95590056 ; 95590057 ; 95590058 ; 95590059 ; 95590060 ; 95590061 ; 95590062 ; 95590063 ; 95590064 ; 95590065 ; 95590066 ; 95590067 ; 95590068 ; 95590069 ; 95590070 ; 95590071 ; 95590072 ; 95590073 ; 95590074 ; 95590075 ; 95590076 ; 95590077 ; 95590078 ; 95590079 ; 95590080 ; 95590081 ; 95590082 ; 95590083 ; 95590084 ; 95590085 ; 95590086 ; 95590087 ; 95590088 ; 95590089 ; 95590090 ; 95590091 ; 95590092 ; 95590093 ; 95590094 ; 95590095 ; 95590096 ; 95590097 ; 95590098 ; 95590099 ; 95590100 ; 95590101 ; 95590102 ; 95590103 ; 95590104 ; 95590105 ; 95590106 ; 95590107 ; 95590108 ; 95590109 ; 95590110 ; 95590111 ; 95590112 ; 95590113 ; 95590114 ; 95590115 ; 95590116 ; 95590117 ; 95590118 ; 95590119 ; 95590120 ; 95590121 ; 95590122 ; 95590123 ; 95590124 ; 95590125 ; 95590126 ; 95590127 ; 95590128 ; 95590129 ; 95590130 ; 95590131 ; 95590132 ; 95590133 ; 95590134 ; 95590135 ; 95590136 ; 95590137 ; 95590138 ; 95590139 ; 95590140 ; 95590141 ; 95590142 ; 95590143 ; 95590144 ; 95590145 ; 95590146 ; 95590147 ; 95590148 ; 95590149 ; 95590150 ; 95590151 ; 95590152 ; 95590153 ; 95590154 ; 95590155 ; 95590156 ; 95590157 ; 95590158 ; 95590159 ; 95590160 ; 95590161 ; 95590162 ; 95590163 ; 95590164 ; 95590165 ; 95590166 ; 95590167 ; 95590168 ; 95590169 ; 95590170 ; 95590171 ; 95590172 ; 95590173 ; 95590174 ; 95590175 ; 95590176 ; 95590177 ; 95590178 ; 95590179 ; 95590180 ; 95590181 ; 95590182 ; 95590183 ; 95590184 ; 95590185 ; 95590186 ; 95590187 ; 95590188 ; 95590189 ; 95590190 ; 95590191 ; 95590192 ; 95590193 ; 95590194 ; 95590195 ; 95590196 ; 95590197 ; 95590198 ; 95590199 ; 95590200 ; 95590201 ; 95590202 ; 95590203 ; 95590204 ; 95590205 ; 95590206 ; 95590207 ; 95590208 ; 95590209 ; 95590210 ; 95590211 ; 95590212 ; 95590213 ; 95590214 ; 95590215 ; 95590216 ; 95590217 ; 95590218 ; 95590219 ; 95590220 ; 95590221 ; 95590222 ; 95590223 ; 95590224 ; 95590225 ; 95590226 ; 95590227 ; 95590228 ; 95590229 ; 95590230 ; 95590231 ; 95590232 ; 95590233 ; 95590234 ; 95590235 ; 95590236 ; 95590237 ; 95590238 ; 95590239 ; 95590240 ; 95590241 ; 95590242 ; 95590243 ; 95590244 ; 95590245 ; 95590246 ; 95590247 ; 95590248 ; 95590249 ; 95590250 ; 95590251 ; 95590252 ; 95590253 ; 95590254 ; 95590255 ; 95590256 ; 95590257 ; 95590258 ; 95590259 ; 95590260 ; 95590261 ; 95590262 ; 95590263 ; 95590264 ; 95590265 ; 95590266 ; 95590267 ; 95590268 ; 95590269 ; 95590270 ; 95590271 ; 95590272 ; 95590273 ; 95590274 ; 95590275 ; 95590276 ; 95590277 ; 95590278 ; 95590279 ; 95590280 ; 95590281 ; 95590282 ; 95590283 ; 95590284 ; 95590285 ; 95590286 ; 95590287 ; 95590288 ; 95590289 ; 95590290 ; 95590291 ; 95590292 ; 95590293 ; 95590294 ; 95590295 ; 95590296 ; 95590297 ; 95590298 ; 95590299 ; 95590300 ; 95590301 ; 95590302 ; 95590303 ; 95590304 ; 95590305 ; 95590306 ; 95590307 ; 95590308 ; 95590309 ; 95590310 ; 95590311 ; 95590312 ; 95590313 ; 95590314 ; 95590315 ; 95590316 ; 95590317 ; 95590318 ; 95590319 ; 95590320 ; 95590321 ; 95590322 ; 95590323 ; 95590324 ; 95590325 ; 95590326 ; 95590327 ; 95590328 ; 95590329 ; 95590330 ; 95590331 ; 95590332 ; 95590333 ; 95590334 ; 95590335 ; 95590336 ; 95590337 ; 95590338 ; 95590339 ; 95590340 ; 95590341 ; 95590342 ; 95590343 ; 95590344 ; 95590345 ; 95590346 ; 95590347 ; 95590348 ; 95590349 ; 95590350 ; 95590351 ; 95590352 ; 95590353 ; 95590354 ; 95590355 ; 95590356 ; 95590357 ; 95590358 ; 95590359 ; 95590360 ; 95590361 ; 95590362 ; 95590363 ; 95590364 ; 95590365 ; 95590366 ; 95590367 ; 95590368 ; 95590369 ; 95590370 ; 95590371 ; 95590372 ; 95590373 ; 95590374 ; 95590375 ; 95590376 ; 95590377 ; 95590378 ; 95590379 ; 95590380 ; 95590381 ; 95590382 ; 95590383 ; 95590384 ; 95590385 ; 95590386 ; 95590387 ; 95590388 ; 95590389 ; 95590390 ; 95590391 ; 95590392 ; 95590393 ; 95590394 ; 95590395 ; 95590396 ; 95590397 ; 95590398 ; 95590399 ; 95590400 ; 95590401 ; 95590402 ; 95590403 ; 95590404 ; 95590405 ; 95590406 ; 95590407 ; 95590408 ; 95590409 ; 95590410 ; 95590411 ; 95590412 ; 95590413 ; 95590414 ; 95590415 ; 95590416 ; 95590417 ; 95590418 ; 95590419 ; 95590420 ; 95590421 ; 95590422 ; 95590423 ; 95590424 ; 95590425 ; 95590426 ; 95590427 ; 95590428 ; 95590429 ; 95590430 ; 95590431 ; 95590432 ; 95590433 ; 95590434 ; 95590435 ; 95590436 ; 95590437 ; 95590438 ; 95590439 ; 95590440 ; 95590441 ; 95590442 ; 95590443 ; 95590444 ; 95590445 ; 95590446 ; 95590447 ; 95590448 ; 95590449 ; 95590450 ; 95590451 ; 95590452 ; 95590453 ; 95590454 ; 95590455 ; 95590456 ; 95590457 ; 95590458 ; 95590459 ; 95590460 ; 95590461 ; 95590462 ; 95590463 ; 95590464 ; 95590465 ; 95590466 ; 95590467 ; 95590468 ; 95590469 ; 95590470 ; 95590471 ; 95590472 ; 95590473 ; 95590474 ; 95590475 ; 95590476 ; 95590477 ; 95590478 ; 95590479 ; 95590480 ; 95590481 ; 95590482 ; 95590483 ; 95590484 ; 95590485 ; 95590486 ; 95590487 ; 95590488 ; 95590489 ; 95590490 ; 95590491 ; 95590492 ; 95590493 ; 95590494 ; 95590495 ; 95590496 ; 95590497 ; 95590498 ; 95590499 ; 95590500 ; 95590501 ; 95590502 ; 95590503 ; 95590504 ; 95590505 ; 95590506 ; 95590507 ; 95590508 ; 95590509 ; 95590510 ; 95590511 ; 95590512 ; 95590513 ; 95590514 ; 95590515 ; 95590516 ; 95590517 ; 95590518 ; 95590519 ; 95590520 ; 95590521 ; 95590522 ; 95590523 ; 95590524 ; 95590525 ; 95590526 ; 95590527 ; 95590528 ; 95590529 ; 95590530 ; 95590531 ; 95590532 ; 95590533 ; 95590534 ; 95590535 ; 95590536 ; 95590537 ; 95590538 ; 95590539 ; 95590540 ; 95590541 ; 95590542 ; 95590543 ; 95590544 ; 95590545 ; 95590546 ; 95590547 ; 95590548 ; 95590549 ; 95590550 ; 95590551 ; 95590552 ; 95590553 ; 95590554 ; 95590555 ; 95590556 ; 95590557 ; 95590558 ; 95590559 ; 95590560 ; 95590561 ; 95590562 ; 95590563 ; 95590564 ; 95590566 ; 95590567 ; 95590568 ; 95590569 ; 95590570 ; 95590571 ; 95590572 ; 95590573 ; 95590574 ; 95590575 ; 95590576 ; 95590577 ; 95590578 ; 95590579 ; 95590580 ; 95590581 ; 95590582 ; 95590583 ; 95590584 ; 95590585 ; 95590586 ; 95590587 ; 95590588 ; 95590589 ; 95590590 ; 95590591 ; 95590592 ; 95590593 ; 95590594 ; 95590595 ; 95590596 ; 95590597 ; 95590598 ; 95590599 ; 95590600 ; 95590601 ; 95590602 ; 95590603 ; 95590604 ; 95590605 ; 95590606 ; 95590607 ; 95590608 ; 95590609 ; 95590610 ; 95590611 ; 95590612 ; 95590613 ; 95590614 ; 95590615 ; 95590616 ; 95590617 ; 95590618 ; 95590619 ; 95590620 ; 95590621 ; 95590622 ; 95590623 ; 95590624 ; 95590625 ; 95590626 ; 95590627 ; 95590628 ; 95590629 ; 95590630 ; 95590631 ; 95590632 ; 95590633 ; 95590634 ; 95590635 ; 95590636 ; 95590637 ; 95590638 ; 95590639 ; 95590640 ; 95590641 ; 95590642 ; 95590643 ; 95590644 ; 95590645 ; 95590646 ; 95590647 ; 95590648 ; 95590649 ; 95590650 ; 95590651 ; 95590652 ; 95590653 ; 95590654 ; 95590655 ; 95590656 ; 95590657 ; 95590658 ; 95590659 ; 95590660 ; 95590661 ; 95590662 ; 95590663 ; 95590664 ; 95590665 ; 95590666 ; 95590667 ; 95590668 ; 95590669 ; 95590670 ; 95590671 ; 95590672 ; 95590673 ; 95590674 ; 95590675 ; 95590676 ; 95590677 ; 95590678 ; 95590679 ; 95590680 ; 95590681 ; 95590682 ; 95590683 ; 95590684 ; 95590685 ; 95590686 ; 95590687 ; 95590688 ; 95590689 ; 95590690 ; 95590691 ; 95590692 ; 95590693 ; 95590694 ; 95590695 ; 95590696 ; 95590697 ; 95590698 ; 95590699 ; 95590700 ; 95590701 ; 95590702 ; 95590703 ; 95590704 ; 95590705 ; 95590706 ; 95590707 ; 95590708 ; 95590709 ; 95590710 ; 95590711 ; 95590712 ; 95590713 ; 95590714 ; 95590715 ; 95590716 ; 95590717 ; 95590718 ; 95590719 ; 95590720 ; 95590721 ; 95590722 ; 95590723 ; 95590724 ; 95590725 ; 95590726 ; 95590727 ; 95590728 ; 95590729 ; 95590730 ; 95590731 ; 95590732 ; 95590733 ; 95590734 ; 95590735 ; 95590736 ; 95590737 ; 95590738 ; 95590739 ; 95590740 ; 95590741 ; 95590742 ; 95590743 ; 95590744 ; 95590745 ; 95590746 ; 95590747 ; 95590748 ; 95590749 ; 95590750 ; 95590751 ; 95590752 ; 95590753 ; 95590754 ; 95590755 ; 95590756 ; 95590757 ; 95590758 ; 95590759 ; 95590760 ; 95590761 ; 95590762 ; 95590763 ; 95590764 ; 95590765 ; 95590766 ; 95590767 ; 95590768 ; 95590769 ; 95590770 ; 95590771 ; 95590772 ; 95590773 ; 95590774 ; 95590775 ; 95590776 ; 95590777 ; 95590778 ; 95590779 ; 95590780 ; 95590781 ; 95590782 ; 95590783 ; 95590784 ; 95590785 ; 95590786 ; 95590787 ; 95590788 ; 95590789 ; 95590790 ; 95590791 ; 95590792 ; 95590793 ; 95590794 ; 95590795 ; 95590796 ; 95590797 ; 95590798 ; 95590799 ; 95590803 ; 95590804 ; 95590807 ; 95590965 ; 95610001 ; 95610002 ; 95610003 ; 95610004 ; 95610005 ; 95610006 ; 95610007 ; 95610008 ; 95610009 ; 95610010 ; 95610011 ; 95610012 ; 95610013 ; 95610014 ; 95610015 ; 95610016 ; 95610017 ; 95610018 ; 95610019 ; 95610020 ; 95610021 ; 95610022 ; 95610023 ; 95610024 ; 95610025 ; 95610026 ; 95610027 ; 95610028 ; 95610029 ; 95610030 ; 95610031 ; 95610032 ; 95610033 ; 95610034 ; 95610035 ; 95610036 ; 95610037 ; 95610038 ; 95610039 ; 95610040 ; 95610041 ; 95610042 ; 95610043 ; 95610044 ; 95610045 ; 95610046 ; 95610047 ; 95610048 ; 95610049 ; 95610050 ; 95610051 ; 95610052 ; 95610053 ; 95610054 ; 95610055 ; 95610056 ; 95610057 ; 95610058 ; 95610059 ; 95610060 ; 95610061 ; 95610062 ; 95610063 ; 95610064 ; 95610065 ; 95610066 ; 95610067 ; 95610068 ; 95610069 ; 95610070 ; 95610071 ; 95610072 ; 95610073 ; 95610074 ; 95610075 ; 95610076 ; 95610077 ; 95610078 ; 95610079 ; 95610080 ; 95610081 ; 95610082 ; 95610083 ; 95610084 ; 95610085 ; 95610086 ; 95610087 ; 95610088 ; 95610089 ; 95610090 ; 95610091 ; 95610092 ; 95610093 ; 95610094 ; 95610095 ; 95610096 ; 95610097 ; 95610098 ; 95610099 ; 95610100 ; 95610101 ; 95610102 ; 95610103 ; 95610104 ; 95610105 ; 95610106 ; 95610107 ; 95610108 ; 95610109 ; 95610110 ; 95610111 ; 95610112 ; 95610113 ; 95610114 ; 95610115 ; 95610116 ; 95610117 ; 95610118 ; 95610119 ; 95610120 ; 95610121 ; 95610122 ; 95610123 ; 95610124 ; 95610125 ; 95610126 ; 95610127 ; 95610128 ; 95610129 ; 95610130 ; 95610131 ; 95610132 ; 95610133 ; 95610134 ; 95610135 ; 95610136 ; 95610137 ; 95610138 ; 95610139 ; 95610140 ; 95610141 ; 95610142 ; 95610143 ; 95610144 ; 95610145 ; 95610146 ; 95610147 ; 95610148 ; 95610149 ; 95610150 ; 95610151 ; 95610152 ; 95610153 ; 95610154 ; 95610155 ; 95610156 ; 95610157 ; 95610158 ; 95610159 ; 95610160 ; 95610161 ; 95610162 ; 95610163 ; 95610164 ; 95610165 ; 95610166 ; 95610167 ; 95610168 ; 95610169 ; 95610170 ; 95610171 ; 95610172 ; 95610173 ; 95610174 ; 95610175 ; 95610176 ; 95610177 ; 95610178 ; 95610179 ; 95610180 ; 95610181 ; 95610182 ; 95620001 ; 95620002 ; 95620003 ; 95620004 ; 95620005 ; 95620006 ; 95620007 ; 95620008 ; 95620009 ; 95620010 ; 95620011 ; 95620012 ; 95620013 ; 95620014 ; 95620015 ; 95620016 ; 95620017 ; 95620018 ; 95620019 ; 95620020 ; 95620021 ; 95620022 ; 95620023 ; 95620024 ; 95620025 ; 95620026 ; 95620027 ; 95620028 ; 95620029 ; 95620030 ; 95620031 ; 95620032 ; 95620033 ; 95620034 ; 95620035 ; 95620036 ; 95620037 ; 95620038 ; 95620039 ; 95620040 ; 95620041 ; 95620042 ; 95620043 ; 95620044 ; 95620045 ; 95620046 ; 95620047 ; 95620048 ; 95620049 ; 95620050 ; 95620051 ; 95620052 ; 95620053 ; 95620054 ; 95620055 ; 95620056 ; 95620057 ; 95620058 ; 95620059 ; 95620060 ; 95620061 ; 95620062 ; 95620063 ; 95620064 ; 95620065 ; 95620066 ; 95620067 ; 95620068 ; 95620069 ; 95620070 ; 95620071 ; 95620072 ; 95620073 ; 95620074 ; 95620075 ; 95620076 ; 95620077 ; 95620078 ; 95620079 ; 95620080 ; 95620081 ; 95620082 ; 95620083 ; 95620084 ; 95620085 ; 95620086 ; 95620087 ; 95620088 ; 95620089 ; 95620090 ; 95620091 ; 95620092 ; 95620093 ; 95620094 ; 95620095 ; 95620096 ; 95620097 ; 95620098 ; 95620099 ; 95620100 ; 95620101 ; 95620102 ; 95620103 ; 95620104 ; 95620105 ; 95620106 ; 95620107 ; 95620108 ; 95620109 ; 95620110 ; 95620111 ; 95620112 ; 95620113 ; 95620114 ; 95620115 ; 95620116 ; 95620117 ; 95620118 ; 95620119 ; 95620120 ; 95620121 ; 95620122 ; 95620123 ; 95620124 ; 95620125 ; 95620126 ; 95620127 ; 95620128 ; 95620129 ; 95620130 ; 95620131 ; 95620132 ; 95620133 ; 95620134 ; 95620135 ; 95620136 ; 95620137 ; 95620138 ; 95620139 ; 95620140 ; 95620141 ; 95620142 ; 95620143 ; 95620144 ; 95620145 ; 95620146 ; 95620147 ; 95620148 ; 95620149 ; 95620150 ; 95620151 ; 95620152 ; 95620153 ; 95620154 ; 95620155 ; 95620156 ; 95620157 ; 95620158 ; 95620159 ; 95620160 ; 95620161 ; 95620162 ; 95620163 ; 95620164 ; 95620165 ; 95620166 ; 95620167 ; 95620169 ; 95620170 ; 95620171 ; 95620172 ; 95620173 ; 95620174 ; 95620176 ; 95620177 ; 95620178 ; 95620179 ; 95620180 ; 95620181 ; 95620182 ; 95620183 ; 95620185 ; 95620186 ; 95620187 ; 95620188 ; 95620189 ; 95620190 ; 95620191 ; 95620192 ; 95620193 ; 95620194 ; 95620195 ; 95620196 ; 95620197 ; 95620198 ; 95620199 ; 95620200 ; 95620201 ; 95620202 ; 95620203 ; 95620204 ; 95620205 ; 95620206 ; 95620207 ; 95620208 ; 95620209 ; 95620210 ; 95620211 ; 95620212 ; 95620213 ; 95620214 ; 95620215 ; 95620216 ; 95620217 ; 95620218 ; 95620219 ; 95620220 ; 95620221 ; 95620222 ; 95620223 ; 95620224 ; 95620225 ; 95620226 ; 95620227 ; 95620228 ; 95620229 ; 95620230 ; 95620231 ; 95620232 ; 95620233 ; 95620234 ; 95620235 ; 95620236 ; 95620237 ; 95620238 ; 95620239 ; 95620240 ; 95620241 ; 95620242 ; 95620243 ; 95620244 ; 95620245 ; 95620246 ; 95620247 ; 95620248 ; 95620249 ; 95620250 ; 95620251 ; 95620252 ; 95620288 ; 95620289 ; 95620290 ; 95620291 ; 95620292 ; 95620293 ; 95620294 ; 95620295 ; 95620296 ; 95620297 ; 95620298 ; 95620299 ; 95620300 ; 95620301 ; 95620302 ; 95620303 ; 95620304 ; 95620305 ; 95620306 ; 95620307 ; 95620308 ; 95620309 ; 95620310 ; 95620311 ; 95620312 ; 95620313 ; 95620314 ; 95620315 ; 95620316 ; 95620317 ; 95620318 ; 95620319 ; 95620320 ; 95620321 ; 95620322 ; 95620323 ; 95620324 ; 95620325 ; 95620326 ; 95620327 ; 95620328 ; 95620329 ; 95620330 ; 95620331 ; 95620332 ; 95620333 ; 95620334 ; 95620335 ; 95620336 ; 95620337 ; 95620338 ; 95620339 ; 95620340 ; 95620341 ; 95620342 ; 95620343 ; 95620344 ; 95620345 ; 95620346 ; 95620347 ; 95620348 ; 95620349 ; 95620350 ; 95620351 ; 95620352 ; 95620353 ; 95620354 ; 95620355 ; 95620356 ; 95620357 ; 95620358 ; 95620359 ; 95620360 ; 95620361 ; 95620362 ; 95620363 ; 95620364 ; 95620365 ; 95620366 ; 95620367 ; 95620368 ; 95620369 ; 95620370 ; 95620371 ; 95620372 ; 95620373 ; 95620374 ; 95620375 ; 95620376 ; 95620377 ; 95620378 ; 95620379 ; 95620380 ; 95620381 ; 95620382 ; 95620383 ; 95620384 ; 95620385 ; 95620386 ; 95620387 ; 95620388 ; 95620389 ; 95620390 ; 95620391 ; 95620392 ; 95620393 ; 95620394 ; 95620395 ; 95620396 ; 95620397 ; 95620398 ; 95620399 ; 95620400 ; 95620401 ; 95620402 ; 95620403 ; 95620404 ; 95620405 ; 95620406 ; 95620407 ; 95620408 ; 95620409 ; 95620410 ; 95620411 ; 95620412 ; 95620413 ; 95620414 ; 95620415 ; 95620416 ; 95620417 ; 95620418 ; 95620419 ; 95620420 ; 95620421 ; 95620457 ; 95620458 ; 95620459 ; 95620460 ; 95620461 ; 95620462 ; 95620463 ; 95620464 ; 95620465 ; 95620466 ; 95620467 ; 95620468 ; 95620469 ; 95620470 ; 95620471 ; 95620472 ; 95620473 ; 95620474 ; 95620475 ; 95620476 ; 95620477 ; 95620478 ; 95620479 ; 95620480 ; 95620481 ; 95620482 ; 95620483 ; 95620484 ; 95620485 ; 95620486 ; 95620487 ; 95620488 ; 95620489 ; 95620490 ; 95620491 ; 95620492 ; 95620493 ; 95620494 ; 95620495 ; 95620496 ; 95620497 ; 95620498 ; 95620499 ; 95620500 ; 95620501 ; 95620502 ; 95620503 ; 95620504 ; 95620505 ; 95620506 ; 95620507 ; 95620508 ; 95620509 ; 95620510 ; 95620511 ; 95620512 ; 95620513 ; 95620514 ; 95620515 ; 95620516 ; 95620517 ; 95620518 ; 95620519 ; 95620520 ; 95620521 ; 95620522 ; 95620523 ; 95620524 ; 95620525 ; 95620526 ; 95620527 ; 95620528 ; 95620529 ; 95620530 ; 95620531 ; 95620532 ; 95620533 ; 95620534 ; 95620535 ; 95620536 ; 95620537 ; 95620538 ; 95620539 ; 95620540 ; 95620541 ; 95620542 ; 95620543 ; 95620544 ; 95620545 ; 95620546 ; 95620547 ; 95620548 ; 95620549 ; 95620550 ; 95620551 ; 95620552 ; 95620553 ; 95620554 ; 95620555 ; 95620556 ; 95620557 ; 95620558 ; 95620559 ; 95620560 ; 95620561 ; 95620562 ; 95620563 ; 95620564 ; 95620565 ; 95620566 ; 95620567 ; 95620568 ; 95620569 ; 95620570 ; 95620571 ; 95620572 ; 95620573 ; 95620574 ; 95620575 ; 95620576 ; 95620577 ; 95620578 ; 95620579 ; 95620580 ; 95620581 ; 95620582 ; 95620583 ; 95620584 ; 95620585 ; 95620586 ; 95620587 ; 95620588 ; 95620589 ; 95620590 ; 95620591 ; 95620592 ; 95620593 ; 95620594 ; 95620595 ; 95620596 ; 95620597 ; 95620598 ; 95620599 ; 95620600 ; 95620601 ; 95620602 ; 95620603 ; 95620604 ; 95620605 ; 95620606 ; 95620607 ; 95620608 ; 95620609 ; 95620610 ; 95620611 ; 95620612 ; 95620613 ; 95620614 ; 95620615 ; 95620616 ; 95620617 ; 95620618 ; 95620619 ; 95620620 ; 95620621 ; 95620622 ; 95620623 ; 95620624 ; 95620625 ; 95620626 ; 95620627 ; 95620628 ; 95620629 ; 95620630 ; 95620631 ; 95620632 ; 95620633 ; 95620634 ; 95620635 ; 95620636 ; 95620637 ; 95620638 ; 95620639 ; 95620640 ; 95620641 ; 95620642 ; 95620643 ; 95620644 ; 95620645 ; 95620646 ; 95620647 ; 95620648 ; 95620649 ; 95620650 ; 95620651 ; 95620652 ; 95620653 ; 95620654 ; 95620656 ; 95620657 ; 95620658 ; 95620659 ; 95620660 ; 95620661 ; 95620662 ; 95620663 ; 95620664 ; 95620665 ; 95620666 ; 95620667 ; 95620668 ; 95620669 ; 95620670 ; 95620671 ; 95620672 ; 95620673 ; 95620674 ; 95620675 ; 95620676 ; 95620677 ; 95620678 ; 95620679 ; 95620680 ; 95620681 ; 95620682 ; 95620683 ; 95620684 ; 95620685 ; 95620686 ; 95620687 ; 95620688 ; 95620689 ; 95620690 ; 95620691 ; 95620692 ; 95620693 ; 95620694 ; 95620695 ; 95620696 ; 95620697 ; 95620716 ; 95620717 ; 95620718 ; 95620719 ; 95620720 ; 95620721 ; 95620722 ; 95620723 ; 95620724 ; 95620725 ; 95620726 ; 95620727 ; 95620728 ; 95620729 ; 95620730 ; 95620731 ; 95620732 ; 95620733 ; 95620734 ; 95620735 ; 95620736 ; 95620737 ; 95620738 ; 95620739 ; 95620740 ; 95620741 ; 95620742 ; 95620743 ; 95620744 ; 95620745 ; 95620746 ; 95620747 ; 95620748 ; 95620749 ; 95620750 ; 95620751 ; 95620752 ; 95620753 ; 95620754 ; 95620755 ; 95620756 ; 95620757 ; 95620758 ; 95620759 ; 95620760 ; 95620761 ; 95620762 ; 95620763 ; 95620764 ; 95620765 ; 95620766 ; 95620767 ; 95620768 ; 95620769 ; 95620770 ; 95620771 ; 95620772 ; 95620773 ; 95620774 ; 95620775 ; 95620776 ; 95620777 ; 95620778 ; 95620779 ; 95620780 ; 95620781 ; 95620782 ; 95620783 ; 95620784 ; 95620785 ; 95620786 ; 95620787 ; 95620788 ; 95620789 ; 95620790 ; 95620791 ; 95620792 ; 95620793 ; 95620794 ; 95620795 ; 95620796 ; 95620797 ; 95620798 ; 95620799 ; 95620800 ; 95620801 ; 95620802 ; 95620803 ; 95620804 ; 95620805 ; 95620806 ; 95620807 ; 95620808 ; 95620809 ; 95620810 ; 95620811 ; 95620812 ; 95620813 ; 95620814 ; 95620815 ; 95620816 ; 95620817 ; 95620818 ; 95620819 ; 95620820 ; 95620821 ; 95620822 ; 95620823 ; 95620824 ; 95620825 ; 95620826 ; 95620827 ; 95620828 ; 95620829 ; 95620830 ; 95620831 ; 95620832 ; 95620833 ; 95620834 ; 95620835 ; 95620836 ; 95620837 ; 95620838 ; 95620839 ; 95620840 ; 95620841 ; 95620842 ; 95620843 ; 95620844 ; 95620845 ; 95620846 ; 95620847 ; 95620848 ; 95620849 ; 95620850 ; 95620851 ; 95620852 ; 95620853 ; 95620854 ; 95620855 ; 95620856 ; 95620857 ; 95620858 ; 95620859 ; 95620860 ; 95620861 ; 95620862 ; 95620863 ; 95620864 ; 95620865 ; 95620866 ; 95620867 ; 95620868 ; 95620869 ; 95620870 ; 95620871 ; 95620872 ; 95620873 ; 95620874 ; 95620875 ; 95620876 ; 95620877 ; 95620878 ; 95620879 ; 95620880 ; 95620881 ; 95620882 ; 95620883 ; 95620884 ; 95620885 ; 95620886 ; 95620887 ; 95620888 ; 95620889 ; 95620890 ; 95620891 ; 95620892 ; 95620893 ; 95620894 ; 95620895 ; 95620896 ; 95620897 ; 95620898 ; 95620899 ; 95620900 ; 95620901 ; 95620902 ; 95620903 ; 95620904 ; 95620905 ; 95620906 ; 95620907 ; 95620908 ; 95620909 ; 95620910 ; 95620911 ; 95620912 ; 95620913 ; 95620914 ; 95620915 ; 95620916 ; 95620917 ; 95620918 ; 95620919 ; 95620920 ; 95620921 ; 95620922 ; 95620923 ; 95620924 ; 95620925 ; 95620926 ; 95620927 ; 95620928 ; 95620929 ; 95620930 ; 95620931 ; 95620932 ; 95620933 ; 95620934 ; 95620935 ; 95620936 ; 95620937 ; 95620938 ; 95620939 ; 95620940 ; 95620941 ; 95620942 ; 95620943 ; 95620944 ; 95620945 ; 95620946 ; 95620947 ; 95620948 ; 95620949 ; 95620950 ; 95620951 ; 95620952 ; 95620953 ; 95620954 ; 95620955 ; 95620956 ; 95620957 ; 95620958 ; 95620959 ; 95620960 ; 95620961 ; 95620962 ; 95620970 ; 95620971 ; 95620972 ; 95620973 ; 95620975 ; 95620977 ; 95620978 ; 95620979 ; 95620980 ; 95620981 ; 95620984 ; 95620986 ; 95620987 ; 95620988 ; 95620989 ; 95620990 ; 95620994 ; 95620995 ; 95620996 ; 95620997 ; 95620998 ; 95620999 ; 95621000 ; 95621001 ; 95621002 ; 95621003 ; 95630003 ; 95630029 ; 95630032 ; 95630038 ; 95630041 ; 95630049 ; 95630051 ; 95630052 ; 95630055 ; 95630056 ; 95630063 ; 95630065 ; 95630070 ; 95630075 ; 95630078 ; 95630080 ; 95630081 ; 95630086 ; 95630087 ; 95630088 ; 95630091 ; 95630092 ; 95630094 ; 95630102 ; 95630104 ; 95630106 ; 95630107 ; 95630108 ; 95630110 ; 95630111 ; 95630112 ; 95630116 ; 95630117 ; 95630121 ; 95630122 ; 95630123 ; 95630129 ; 95630130 ; 95630131 ; 95630134 ; 95630137 ; 95630138 ; 95630143 ; 95630148 ; 95630149 ; 95630150 ; 95630151 ; 95630152 ; 95630153 ; 95630154 ; 95630155 ; 95630156 ; 95630159 ; 95630160 ; 95630161 ; 95630162 ; 95630163 ; 95630164 ; 95630165 ; 95630166 ; 95630168 ; 95630169 ; 95630172 ; 95630173 ; 95630174 ; 95630175 ; 95630176 ; 95630178 ; 95630179 ; 95630180 ; 95630183 ; 95630184 ; 95630185 ; 95630187 ; 95630188 ; 95630189 ; 95630190 ; 95630191 ; 95630193 ; 95630196 ; 95630197 ; 95630198 ; 95630199 ; 95630200 ; 95630202 ; 95630203 ; 95630204 ; 95630205 ; 95630206 ; 95630207 ; 95630209 ; 95630210 ; 95630213 ; 95630217 ; 95630218 ; 95630219 ; 95630220 ; 95630221 ; 95630222 ; 95630225 ; 95630226 ; 95630227 ; 95630229 ; 95630230 ; 95630231 ; 95630232 ; 95630233 ; 95630234 ; 95630236 ; 95630238 ; 95630244 ; 95630246 ; 95630248 ; 95630250 ; 95630251 ; 95630252 ; 95630255 ; 95630256 ; 95630257 ; 95630260 ; 95630262 ; 95630270 ; 95630274 ; 95630285 ; 95630286 ; 95630290 ; 95630296 ; 95630300 ; 95630301 ; 95630302 ; 95630303 ; 95630306 ; 95630308 ; 95630309 ; 95630314 ; 95630315 ; 95630316 ; 95630317 ; 95630318 ; 95630319 ; 95630322 ; 95630324 ; 95630326 ; 95630328 ; 95630330 ; 95630331 ; 95630334 ; 95630337 ; 95630339 ; 95630342 ; 95630343 ; 95630349 ; 95630358 ; 95630359 ; 95630360 ; 95630372 ; 95630373 ; 95630374 ; 95630379 ; 95630383 ; 95630386 ; 95630387 ; 95630388 ; 95630389 ; 95630390 ; 95630393 ; 95630394 ; 95630395 ; 95630396 ; 95630397 ; 95630398 ; 95630403 ; 95630405 ; 95630406 ; 95630407 ; 95630408 ; 95630411 ; 95630414 ; 95630429 ; 95630437 ; 95630439 ; 95630440 ; 95630442 ; 95630444 ; 95630447 ; 95630448 ; 95630451 ; 95630452 ; 95630457 ; 95630460 ; 95630461 ; 95630466 ; 95630468 ; 95630469 ; 95630471 ; 95630473 ; 95630474 ; 95630475 ; 95630476 ; 95630481 ; 95630482 ; 95630484 ; 95630487 ; 95630493 ; 95630497 ; 95630498 ; 95630499 ; 95630500 ; 95630502 ; 95630503 ; 95630504 ; 95630505 ; 95630506 ; 95630507 ; 95630508 ; 95630511 ; 95630517 ; 95630518 ; 95630519 ; 95630520 ; 95630523 ; 95630525 ; 95630526 ; 95630527 ; 95630528 ; 95630530 ; 95630531 ; 95630532 ; 95630533 ; 95630534 ; 95630536 ; 95630537 ; 95630538 ; 95630539 ; 95630540 ; 95630541 ; 95630542 ; 95630544 ; 95630546 ; 95630548 ; 95630549 ; 95630550 ; 95630556 ; 95630557 ; 95630558 ; 95630559 ; 95630561 ; 95630588 ; 95630589 ; 95630591 ; 95630592 ; 95630593 ; 95630594 ; 95630596 ; 95630600 ; 95630602 ; 95630605 ; 95630614 ; 95630624 ; 95630625 ; 95630626 ; 95630628 ; 95630629 ; 95630630 ; 95630634 ; 95630635 ; 95630636 ; 95630637 ; 95630638 ; 95630639 ; 95630640 ; 95630641 ; 95630652 ; 95630661 ; 95630662 ; 95630663 ; 95630664 ; 95630668 ; 95630670 ; 95630671 ; 95630672 ; 95630674 ; 95630675 ; 95630681 ; 95630686 ; 95630691 ; 95630692 ; 95630695 ; 95630696 ; 95630698 ; 95630699 ; 95630700 ; 95630701 ; 95630706 ; 95630707 ; 95630708 ; 95630714 ; 95630717 ; 95630719 ; 95630720 ; 95630723 ; 95630724 ; 95630728 ; 95630731 ; 95630739 ; 95630740 ; 95630741 ; 95630744 ; 95630747 ; 95630751 ; 95630754 ; 95630760 ; 95630765 ; 95630766 ; 95630768 ; 95630776 ; 95630785 ; 95630793 ; 95630794 ; 95630800 ; 95630803 ; 95630804 ; 95630806 ; 95630811 ; 95630814 ; 95630832 ; 95630834 ; 95630835 ; 95630837 ; 95630844 ; 95630845 ; 95630846 ; 95630848 ; 95630849 ; 95630850 ; 95630852 ; 95630862 ; 95630863 ; 95630869 ; 95630873 ; 95630879 ; 95630881 ; 95630900 ; 95630901 ; 95630902 ; 95630903 ; 95630906 ; 95630907 ; 95630908 ; 95630912 ; 95630918 ; 95630919 ; 95630926 ; 95630927 ; 95630928 ; 95630929 ; 95630930 ; 95630933 ; 95630934 ; 95630935 ; 95630936 ; 95630938 ; 95630939 ; 95630940 ; 95630946 ; 95630947 ; 95630950 ; 95630952 ; 95630959 ; 95630965 ; 95630970 ; 95630971 ; 95630975 ; 95630976 ; 95630979 ; 95630980 ; 95630981 ; 95630982 ; 95630983 ; 95630984 ; 95630985 ; 95630990 ; 95630991 ; 95630992 ; 95630993 ; 95630995 ; 95631000 ; 95631003 ; 95631004 ; 95631005 ; 95631007 ; 95631009 ; 95631011 ; 95631012 ; 95631014 ; 95631018 ; 95631019 ; 95631020 ; 95631021 ; 95631023 ; 95631024 ; 95631026 ; 95631027 ; 95631028 ; 95631034 ; 95631037 ; 95631038 ; 95631039 ; 95631041 ; 95631044 ; 95631046 ; 95631048 ; 95631049 ; 95631055 ; 95631056 ; 95631058 ; 95631060 ; 95631061 ; 95631076 ; 95631077 ; 95631082 ; 95631083 ; 95631093 ; 95631094 ; 95631095 ; 95631097 ; 95631098 ; 95631099 ; 95631100 ; 95631101 ; 95631102 ; 95631103 ; 95631104 ; 95631105 ; 95631106 ; 95631107 ; 95631108 ; 95631109 ; 95631110 ; 95631111 ; 95631112 ; 95631113 ; 95631114 ; 95631116 ; 95631117 ; 95631119 ; 95631124 ; 95631126 ; 95631128 ; 95631131 ; 95631133 ; 95640013 ; 95640017 ; 95640019 ; 95640020 ; 95640022 ; 95640025 ; 95640032 ; 95640041 ; 95640043 ; 95640044 ; 95640045 ; 95640047 ; 95640049 ; 95640053 ; 95640056 ; 95640058 ; 95640060 ; 95640063 ; 95640064 ; 95640067 ; 95640068 ; 95640069 ; 95640070 ; 95640071 ; 95640072 ; 95640073 ; 95640074 ; 95640076 ; 95640077 ; 95640078 ; 95640080 ; 95640081 ; 95640082 ; 95640083 ; 95640084 ; 95640085 ; 95640086 ; 95640087 ; 95640088 ; 95640089 ; 95640090 ; 95640091 ; 95640092 ; 95640093 ; 95640094 ; 95640095 ; 95640096 ; 95640097 ; 95640098 ; 95640099 ; 95640101 ; 95640102 ; 95640103 ; 95640105 ; 95640106 ; 95640107 ; 95640108 ; 95640109 ; 95640110 ; 95640111 ; 95640112 ; 95640113 ; 95640114 ; 95640117 ; 95640118 ; 95640119 ; 95640122 ; 95640123 ; 95640124 ; 95640125 ; 95640126 ; 95640127 ; 95640129 ; 95640130 ; 95640131 ; 95640132 ; 95640133 ; 95640134 ; 95640135 ; 95640136 ; 95640137 ; 95640138 ; 95640140 ; 95640141 ; 95640142 ; 95640143 ; 95640144 ; 95640145 ; 95640146 ; 95640147 ; 95640148 ; 95640149 ; 95640150 ; 95640151 ; 95640152 ; 95640154 ; 95640155 ; 95640156 ; 95640158 ; 95640159 ; 95640160 ; 95640161 ; 95640162 ; 95640167 ; 95640168 ; 95640170 ; 95640171 ; 95640172 ; 95640173 ; 95640174 ; 95640175 ; 95640176 ; 95640177 ; 95640178 ; 95640179 ; 95640180 ; 95640181 ; 95640182 ; 95640183 ; 95640184 ; 95640190 ; 95640195 ; 95640196 ; 95640197 ; 95640202 ; 95640203 ; 95640204 ; 95640205 ; 95640206 ; 95640209 ; 95640210 ; 95640211 ; 95640212 ; 95640213 ; 95640214 ; 95640215 ; 95640216 ; 95640217 ; 95640218 ; 95640219 ; 95640220 ; 95640221 ; 95640222 ; 95640223 ; 95640224 ; 95640225 ; 95640226 ; 95640227 ; 95640228 ; 95640229 ; 95640230 ; 95640231 ; 95640232 ; 95640234 ; 95640238 ; 95640239 ; 95640240 ; 95640241 ; 95640242 ; 95640243 ; 95640244 ; 95640245 ; 95640246 ; 95640247 ; 95640248 ; 95640249 ; 95640250 ; 95640251 ; 95640253 ; 95640254 ; 95640255 ; 95640256 ; 95640257 ; 95640258 ; 95640259 ; 95640260 ; 95640261 ; 95640262 ; 95640265 ; 95640266 ; 95640268 ; 95640270 ; 95640271 ; 95640272 ; 95640273 ; 95640274 ; 95640275 ; 95640276 ; 95640277 ; 95640278 ; 95640279 ; 95640280 ; 95640281 ; 95640282 ; 95640283 ; 95640284 ; 95640289 ; 95640290 ; 95640291 ; 95640292 ; 95640293 ; 95640294 ; 95640295 ; 95640296 ; 95640297 ; 95640298 ; 95640299 ; 95640300 ; 95640301 ; 95640302 ; 95640303 ; 95640305 ; 95640306 ; 95640311 ; 95640312 ; 95640315 ; 95640316 ; 95640317 ; 95640318 ; 95640319 ; 95640320 ; 95640321 ; 95640349 ; 95640350 ; 95640352 ; 95640353 ; 95650001 ; 95650002 ; 95650003 ; 95650004 ; 95650005 ; 95650006 ; 95650007 ; 95650008 ; 95650009 ; 95650010 ; 95650011 ; 95650012 ; 95650013 ; 95650014 ; 95650015 ; 95650016 ; 95650017 ; 95650018 ; 95650019 ; 95650020 ; 95650021 ; 95650022 ; 95650023 ; 95650024 ; 95650025 ; 95650026 ; 95650027 ; 95650028 ; 95650029 ; 95650030 ; 95650031 ; 95650032 ; 95650033 ; 95650034 ; 95650035 ; 95650036 ; 95650037 ; 95650038 ; 95650039 ; 95650040 ; 95650041 ; 95650042 ; 95650043 ; 95650044 ; 95650045 ; 95650046 ; 95650047 ; 95650048 ; 95650049 ; 95650050 ; 95650051 ; 95650052 ; 95650053 ; 95650054 ; 95650055 ; 95650056 ; 95650057 ; 95650058 ; 95650059 ; 95650060 ; 95650061 ; 95650062 ; 95650063 ; 95650064 ; 95650065 ; 95650066 ; 95650067 ; 95650068 ; 95650069 ; 95650070 ; 95650071 ; 95650072 ; 95650073 ; 95650074 ; 95650075 ; 95650076 ; 95650077 ; 95650078 ; 95650079 ; 95650080 ; 95650081 ; 95650082 ; 95650083 ; 95650084 ; 95650085 ; 95650086 ; 95650087 ; 95650088 ; 95650089 ; 95650090 ; 95650091 ; 95650092 ; 95650093 ; 95650094 ; 95650095 ; 95650096 ; 95650097 ; 95650098 ; 95650099 ; 95650100 ; 95650101 ; 95650102 ; 95650103 ; 95650104 ; 95650105 ; 95650106 ; 95650107 ; 95650108 ; 95650109 ; 95650110 ; 95650111 ; 95650112 ; 95650113 ; 95650114 ; 95650115 ; 95650116 ; 95650117 ; 95650118 ; 95650119 ; 95650120 ; 95650121 ; 95650122 ; 95650123 ; 95650124 ; 95650125 ; 95650126 ; 95650127 ; 95650128 ; 95650129 ; 95650130 ; 95650131 ; 95650132 ; 95650133 ; 95650134 ; 95650135 ; 95650136 ; 95650137 ; 95650138 ; 95650139 ; 95650140 ; 95650141 ; 95650142 ; 95670001 ; 95670002 ; 95670003 ; 95670005 ; 95670007 ; 95670009 ; 95670011 ; 95670012 ; 95670013 ; 95670014 ; 95670015 ; 95670016 ; 95670017 ; 95670019 ; 95670020 ; 95670021 ; 95670022 ; 95670023 ; 95670024 ; 95670026 ; 95670028 ; 95670029 ; 95670030 ; 95670031 ; 95670033 ; 95670034 ; 95670035 ; 95670036 ; 95670037 ; 95670038 ; 95670039 ; 95670040 ; 95670041 ; 95670042 ; 95670043 ; 95670044 ; 95670045 ; 95670046 ; 95670047 ; 95670049 ; 95670050 ; 95670052 ; 95670053 ; 95670054 ; 95670055 ; 95670057 ; 95670058 ; 95670061 ; 95670062 ; 95670063 ; 95670064 ; 95670065 ; 95670067 ; 95670070 ; 95670072 ; 95670073 ; 95670074 ; 95670075 ; 95670076 ; 95670077 ; 95670078 ; 95670079 ; 95670080 ; 95670081 ; 95670082 ; 95670083 ; 95670084 ; 95670085 ; 95670086 ; 95670087 ; 95670088 ; 95670089 ; 95670090 ; 95670091 ; 95670092 ; 95670093 ; 95670094 ; 95670097 ; 95670098 ; 95670099 ; 95670100 ; 95670101 ; 95670102 ; 95670103 ; 95670104 ; 95670105 ; 95670106 ; 95670107 ; 95670108 ; 95670109 ; 95670110 ; 95670111 ; 95670112 ; 95670113 ; 95670114 ; 95670115 ; 95670116 ; 95670117 ; 95670118 ; 95670119 ; 95670121 ; 95670122 ; 95670123 ; 95670124 ; 95670125 ; 95670126 ; 95670127 ; 95670128 ; 95670129 ; 95670130 ; 95670131 ; 95670132 ; 95670133 ; 95670134 ; 95670135 ; 95670136 ; 95670137 ; 95670138 ; 95670139 ; 95670140 ; 95670141 ; 95670142 ; 95670143 ; 95670144 ; 95670145 ; 95670146 ; 95670147 ; 95670148 ; 95670149 ; 95670150 ; 95670151 ; 95670152 ; 95670153 ; 95670154 ; 95670155 ; 95670156 ; 95670157 ; 95670158 ; 95670159 ; 95670160 ; 95670161 ; 95670162 ; 95670163 ; 95670164 ; 95670165 ; 95670166 ; 95670167 ; 95670169 ; 95670171 ; 95670172 ; 95670173 ; 95670174 ; 95670175 ; 95670176 ; 95670177 ; 95670178 ; 95670179 ; 95670180 ; 95670181 ; 95670182 ; 95670183 ; 95670184 ; 95670185 ; 95670186 ; 95670187 ; 95670188 ; 95670189 ; 95670190 ; 95670191 ; 95670192 ; 95670193 ; 95670194 ; 95670195 ; 95670196 ; 95670197 ; 95670198 ; 95670199 ; 95670200 ; 95670201 ; 95670202 ; 95670203 ; 95670204 |
Recalling Firm/ Manufacturer |
Hospira Inc. 275 N Field Dr Lake Forest IL 60045-2579
|
| For Additional Information Contact | Hospira Global Complaint Management 800-441-4100 |
Manufacturer Reason for Recall | Hospira has received reports of PCA pumps not detecting distal occlusions, one of which resulted in a serious injury. The issue is caused by normal wear and tear on the Half Nut (the component/nut that travels up and down the lead screw) which prevents it from properly detecting the pressure build-up associated with a distal occlusion. |
FDA Determined Cause 2 | Component design/selection |
| Action | Hospira sent an IMPORTANT DEVICE INFORMATION letter dated March 8, 2014 to all affected customers.. The letter advised customers to: 1) Perform the Performance Verification Test (PVT) Occlusion Test as defined in the PCA Technical Service Manual (TSM); 2) If the device does not pass the test, remove it from clinical service and contact Hospira at 800-441-4100 (Monday - Friday, 8:00 AM - 5:00 PM CST) to report the issue; and 3) Perform the appropriate troubleshooting and repair activities defined by your facility, which may include returning the device to Hospira for further diagnosing and servicing.
The letter advised the Hospira is in the process of establishing a useful life for the half nut and to determine when it will require replacement. A new Technical Service Manual will be issued in late 2014 that adds a requirement for annual PVT Occlusion testing. The System Operating Manual will include new instructions to reduce the excessive wear on the half nut resulting from incorrectly using the vial information.
Customers who further distributed the devices are requested to notify their customers of the recall and ask them to call Stericycle at 888-965-5801 (Monday - Friday, 8:00 AM - 5:00 PM EST) to receive a reply form.
Questions about this recall should be direct to Hospira at 800-441-4100 (Monday - Friday, 8:00 AM - 5:00 PM CST). |
| Quantity in Commerce | 6,268 infusion pump systems |
| Distribution | Nationwide Distribution to AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY) + District of Columbia + Puerto Rico + Virgin Islands; *** FOREIGN: |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
|
|
|