Date Initiated by FirmMay 20, 2014
Date PostedAugust 29, 2014
Recall Status1 Terminated 3 on August 14, 2017
Recall NumberZ-2405-2014
Recall Event ID 68536
Product Classification Cesarean section kit - Product Code OHM
ProductCESAREAN PACK - (1) BABY BLANKET IMP. 30" X 30" POLY ABS. (1) NEEDLE & BLADE COUNTER 60c FOAM/STRIP MAG/STRIP LIF (1) DRESSING ISLAND 4" x 10" (1) CESAREAN DRAPE 71" x 125" LIF (2) GOWN SURG REINFORCED LARGE TOWEL/WRAP (10) LAP SPONGES PREWASH XRD LIF (1) BABY BEANNIES (1) SCALPEL #10 WITH HANDLE STAINLESS STEEL (1) DRAPE SHEET 42" X 57" SMS (1) EAR ULCER SYRINGE 2oz. UF (1) RECEIVING BLANKET 100% COTTON (1) SCALPEL #20 WITH HANDLE (1) DURAPREP SURGICAL SOLUTION 26ml UF (2) ABSORBENT TOWEL 15" X 20" L.F (2) LITE GLOVES UF (1) BOWL WITH LID PLASTIC 80oz. (1) TABLE COVER REINFORCED 44" x 78" LIF (1) UTILITY BOWL 32oz (2) PAD OBSTETRICAL X-LARGE ST. (1) WRAPPER 30" X 30" LIF (1) MAYO STAND COVER REINFORCED LIF (1) TUBE SUCTION CONNECT Y." X 12' UF (1) MUCUSTRAP 20cc 1 OFR (1) YANKAUER SUCTION TUBE WITHOUT VENT UF (1) UMBILICAL CORD CLAMP EO sterilized surgical convenient packs packaged into polyester/polyethylene breather bag with Tyvek double vents.
Code Information Product code 900-1689, 20 lots: 111113026 111123373 112051861 112083402 112114645 113015432 113036419 113036729 113047214 113057427 113068155 113088752 113099277 131110194 131210755 140111165 140211884 140312435 140513104 140513464 
FEI Number 2648727
Recalling Firm/
Manufacturer
Customed, Inc
Calle Igualdad Final #7
Fajardo PR 00738
For Additional Information ContactRosemari Melendez
787-622-5151 Ext. 7540
Manufacturer Reason
for Recall
Customed has determined that there is the possibility that packaging integrity may be compromised on different catalog numbers (multiple lots). This could result in an injury to the patient due to product contamination or loss of sterility condition.
FDA Determined
Cause 2
Package design/selection
ActionCustomed Inc. sent a Recall Notification letter dated June 3, 2014 via email and followed by letter to all affected customers. The letter identified the affected products, problem and actions to be taken. The letter informed customers if the affected products are in their inventory to: 1. Remove products from sale (quarantine them) and cease distribution. 2. Complete and return the enclosed acknowledgement form to Customed Inc .Recall Coordinator. Affected products should be returned to Customed Inc for credit. Customers with questions were instructed to contact Customed Inc. Recall Coordinator at 1-787-801-0100 Ext. 7540
Quantity in Commerce20 lots; 1095 units (multiple units per lot)
DistributionWorldwide Distribution - US (nationwide) in the state of Florida including Puerto Rico (USA); & USVI (St John) and the countries of Tortola, Trinidad, Dominican Republic, Costa Rica and Mexico.
Total Product Life CycleTPLC Device Report