| | Class 2 Device Recall Invacare TDX |  |
| Date Initiated by Firm | January 15, 2015 |
| Date Posted | February 20, 2015 |
| Recall Status1 |
Terminated 3 on September 23, 2015 |
| Recall Number | Z-1137-2015 |
| Recall Event ID |
69378 |
| 510(K)Number | K023589 |
| Product Classification |
Wheelchair, powered - Product Code ITI
|
| Product | TDX Electric Wheelchair. Invacare FDX, TDX series and Storm Series (3G) power wheelchairs in 20 to 24 inch width configurations with power recline seating system and a conventional style back. |
| Code Information |
Model # TDXSP-CG TDXSP-CG-GT TDXSP-MCG TDXSP-MCG-GT TDXSPSEAT (only for CGTRK & CGR2TK) TDXSR-CG TDXSR-CG-HD TDXSR-MCG TDXSR-MCG-HD TDXSRSEAT (only for CGTRK & CGR2TK) Serial numbers for all models 11KE003137, 11LE003215, 10LE000825, 10JE000783, 08GE001643, 10HE002685, 10HE000921, 12DE000966, 12IE003985, 11KE001216, 07JE005091, 11CE002223, 10HE004991, 11DE002513, 11EE001617, 11AE001071, 09LE002681, 07KE001564, 07GE004941, 08CE002502, 11KE003316, 08HE005195, 08IE002697, 10DE000685, 08IE003493, 10BE002355, 10DE000562, 12IE003063, 12FE004399, 09BE002547, 10DE002382, 08KE000351, 08AE004060, 12JE000192, 12KE001805, 10EE003295, 12KE002002, 11HE002277, 12CE002659, 08EE001584, 09DE003466, 13HE000343, 10HE001917, 11GE003762, 12CE003242, 12JE003316, 08LE004865, 07JE005908, 09HE001839, 08CE001894, 11JE001054, 09FE001441, 08DE000845, 10CE000589, 08BE004611, 10BE002921, 07DE001356, 12DE002032, 09HE004671, 08IE001598, 10KE001144, 08KE004205, 10HE005507, 10HE000564, 12DE003808, 10HE001166, 09DE002092, 11DE004535, 09CE000330, 08HE003225, 12JE001743, 12GE001756, 11FE003542, 07LE004648, 12BE002035, 10BE002352, 09EE003805, 08GE004760, 12GE002328, 12AE005461, 08FE003622, 12DE000616, 07JE000966, 08LE005085, 11EE003780, 12GE003127, 12EE002756, 08FE003317, 13AE001235, 08DE000039, 10AE002104, 08BE002786, 12IE002542, 09CE002495, 08BE004214, 13LE000634, 08HE000567, 07JE004502, 09BE002772, 12JE001455, 13JE000299, 07BR003307, 14FE000447, 08IE002133, 08JE000302, 09BE002767, 12JE001637, 10IE001126, 12BE003254, 07IE004514, 08JE001851, 08EE002568, 09BE003009, 12KE001301, 11BE003967, 07BE003307, 07LE003702, 08JE002532, 12LE000919, 09JE003794, 09DE001338, 08CE003643, 08GE001008, 08BE003889, 08KE001596, 10HE005506, 10FE004526, 11AE003675, 08FE004118, 14DE000609, 11EE002528, 09BE003741, 08LE001401, 11KE001494, 12CE002397, 09FE002368, 08FE005384, 11HE001416, 10BE004558, 10HE000061, 08FE000073, 13EE000415, 09HE001200, 08CE003640, 13AE000611, 10KE000049, 10FE004487, 07LE003939, 09GE001218, 09HE005212, 08EE002092, 09HE000061, 13DE000700, 10LE003114, 10LE002780, 09BE003491, 09IE004153, 08EE002090, 10CE000059, 08KE000352, 11EE003367, 11LE003178, 08GE004089, 09JE004928, 08EE002587, 10LE002030, 07IE001743, 11GE004662, 12FE002247, 08FE000826, 09JE005188, 08EE004178, 10LE002488, 14EE000360, 12DE003959, 06LE005870, 07JE004823, 10BE003527, 08FE001104, 11IE001289, 12FE004402, 12EE002400, 08KE002059, 08HE002624, 10CE000378, 08FE001359, 11JE000004, 14DE000614, 12AE001037, 09FE001429, 08JE003646, 10CE001920, 08FE005130, 11LE002068, 11FE002022, 13AE000413, 11AE003618, 09CE003449, 10EE003289, 09BE001156, 12FE002760, 09FE001438, 11AE000604, 11KE000451, 09DE000608, 10EE003568, 09EE004644, 12GE003409, 08DE002768, 07EE002979, 12BE002274, 10CE004923, 12BE000702, 09KE002236, 10FE000660, 10KE004142, 12HE004086, 12CE002137, 11HE004470, 12DE001511, 10DE000913, 11BE003383, 07FE002211, 09HE005222, 08LE003944, 11AE002642, 13DE000387, 10JE002425, 09LE004673, 10AE002398, 07LE000319, 09BE004157, 08EE004176, 07CE003237, 11BE001143, 10FE005521, 10CE004656, 07CE007696, 11DE001455, 08EE004175, 08BE003251, 11CE003774, 07LE003696, 11AE003813, 11CE003797, 11EE001447, 08FE002779, 09DE000393, 11LE001615, 10HE002966, 09CE001077, 10CE000043, 11EE001821, 09DE001316, 09DE003931, 12EE003784, 08FE004442, 10LE001701, 10AE003313, 07KE002818, 09BE003034, 09HE000553, 12FE001307, 11CE002451, 11BE001658, 08FE002773, 08DE004273, 10BE003226, 09IE001377, 10AE001215, 12CE000052, 12AE004185, 10HE002665, 10JE002112, 10CE005434, 09KE002406, 09JE001828, 08BE003223, 08FE002971, 07GE001983, 12JE002474, 12CE001081, 10KE005253, 10CE000559, 11HE000307, 12BE002646, 09DE000054, 08HE005001, 09KE002424, 11DE001184, 10GE002864, 11AE002806, 12GE000618, 07JE004066, 08JE002537, 10HE005738, 11HE000485, 11BE003943, 11LE004444, 12GE000619, 09FE000968, 08LE004276, 08EE000983, 12HE001837, 11IE002566, 08GE001207, 07GE001139, 09DE000590, 09AE000049, 12FE003798, 06JE005408, 11JE003181, 09DE002692, 07GE003516, 10CE001050, 08JE005151, 08HE002938, 10BE003139, 11KE002768, 09IE001643, 09JE003508, 12HE000488, 09DE001841, 12JE002365, 10DE000055, 12CE002179, 10AE002397, 12DE001985, 10IE004515, 08AE001978, 07JE005564, 10JE003144, 12GE001341, 10GE000588, 12BE004238, 12CE005124, 08EE003064, 11KE003850, 11GE001058, 12HE000442, 10GE003085, 12LE000639, 12DE002142, 08HE004338, 08JE003599, 11GE002340, 10CE002206, 10HE003972, 10BE002108, 12EE003049, 09CE002489, 11DE002845, 11JE002673, 13AE001127, 10JE001489, 11AE000830, 10FE000399, 13GE000757, 07JE005594, 12BE004210, 08HE002902, 11AE001067, 08DE004691, 12AE001229, 09EE004006, 07LE001514, 08CE004481, 09KE002663, 11AE001726, 09KE004269, 12CE000755, 10HE002695, 08EE002306, 09DE002247, 08JE002535, 11BE002876, 08EE002336, 12GE000060, 08IE003963, 08EE002773, 09JE004715, 09AE002739, 11CE005332, 08FE002771, 08GE005259, 09EE001824, 08GE004972, 10EE003738, 10FE001435, 11KE001482, 09BE001633, 11IE002392, 10IE000553, 08HE002554, 08DE001045, 07GE005660, 11KE002809, 09FE005633, 11KE001948, 11CE001083, 08IE000875, 10HE004943, 13FE000057, 11LE001087, 09GE000452, 12BE003041, 11FE005022, 08KE004569, 10JE003934, 09DE002696, 12BE003043, 09GE004571, 12CE004398, 09DE004396, 08KE004567, 09JE005185, 08JE002067, 12CE000955, 10CE005428, 12DE003335, 09GE004294, 09BE003922, 09LE004391, 07EE000010, 12CE005556, 11DE001707, 12DE003511, 09HE004426, 09CE004703, 10EE000442, 12EE002602, 12DE000908, 12FE002553, 12GE002643, 10LE005220, 09DE001103, 10HE005309, 10BE004392, 12FE000730, 11JE000795, 12HE003198, 13BE000035, 09FE000257, 10JE000059, 10FE001652, 12GE000421, 12CE000961, 12IE001536, 12JE001398, 09FE002043, 11DE003678, 11FE002493, 12JE001608, 12DE000914, 12IE003987, 07DE002509, 09FE002805, 11FE004034, 12FE001302, 12JE002159, 12FE002276, 12JE001765, 11BE001695, 09GE002153, 11HE003443, 07EE005198, 12LE000636, 08LE000078, 12JE003311, 11GE004661, 09GE004319, 09HE002104, 07KE004545, 12LE000637, 10CE000055, 12LE000844, 07LE004605, 09GE004947, 10AE003517, 07LE001789, 12LE002553, 10GE001228, 12LE002548, 12CE001716, 09GE005305, 10CE004654, 08HE004989, 12LE002670, 12HE004271, 09CE001082, 07HE002793, 09HE001255, 12JE000989, 08JE003859, 07DE001938, 07JE001172, 07FE005143, 06KE002230, 09HE003310, 10LE002526, 09GE003566, 08HE000568, 10DE002908, 09HE004430, 09AE002722, 10BE000071, 11BE001367, 09IE001827, 08IE004311, 11KE001490, 07EE006274, 09GE002189, 10FE005649, 12EE005008, 08AE000429, 08JE000769, 08DE005132, 10EE000211, 07EE006570, 10HE002469, 12AE002152, 09DE003663, 09BE000264, 10GE003840, 11CE002220, 08LE005088, 10HE005559, 12IE002077, 09GE005551, 10IE002904, 10FE003944, 07BE001559, 13EE000319, 10JE001499, 07GE004937, 09LE003707, 10IE002941, 12EE002148, 10EE003575, 07KE004299, 10KE003495, 07KE000032, 12FE000528, 10KE004134, 09LE001900, 11DE000944, 09KE001436, 10KE003980, 07LE004248, 12IE002287, 12AE004677, 11FE002030, 13JE000053, 07GE003021, 11AE003923, 09HE004421, 12JE002155, 12FE002275, 13DE000869, 11AE002804, 11EE003368, 11CE003855, 10IE003740, 14DE000111, 07BE001779, 08JE002060, 12EE002599, 07DE006057, 11DE000148, 12IE001497, 12KE001592, 07BD001779, 09KE002230, 12IE000828, 08EE000074, 11DE000690, 11EE000688, 08AE003564, 10JE004940, 09LE004121, 09IE002462, 13LE000564, 11DE002840, 11EE004468, 10LE004589, 13BE000613, 09IE004486, 11KE000702, 07LE000323, 12DE000035, 12AE002433, 13FE000132, 09EE003104, 09BE001659, 07JE001430, 08HE001972, 13IE000741, 12EE004269, 10FE002972, 09GE001245, 09JE003514, 09GE003048, 13DE000779, 12AE004447, 14EE000009, 09FE001786, 11BE001694, 10JE001450, 11BE003784, 10FE003949, 09AE005286, 13AE000856, 08AE003820, 11EE004247, 11FE002262, 11IE003479, 10KE004364, 10IE002333, 11HE001044, 13EE000227, 11EE004253, 11LE002025, 08CE000048, 09LE000861, 10IE001348, 12GE001909, 09DE002274, 08LE000706, 08BE000880, 09HE005224, 08GE000232, 09FE003473, 12GE003235, 10AE000444, 11AE000595, 07GE002444, 08HE003524, 07EE002833, 11AE002606, 13AE000480, 09FE000973, 10GE001451, 08FE001841, 11BE001898, 07EE007366, 14BE000121, 13BE000017, 12AE002112, 10KE004104.08FE001903, 07DE003990, 09JE002972, 09EE003958, 12JE003498, 08LE002081, 07JE006064, 09LE001381, 09CE005219, 07FE003558, 09KE002417, 13HE000322, 10DE001349, 08BE000842, 11CE003792, 09IE002599, 07GE003777, 10CE001300, 11JE000494, 12BE000969, 08JE003108, 08AE002670, 09JE004923, 08CE003308, 10KE000857, 11JE001571, 12LE002164, 09HE004978, 10CE003904, 09LE001114, 09BE001398, 08KE003195, 12DE000025, 11BE003779, 08HE003877, 09CE003447, 11LE001089, 10FE001155, 10CE003654, 12EE001527, 12JE002907, 09AE004407, 13LE000164, 12FE001635, 10FE005332, 10LE000582, 13JE000340, 10FE003946, 08BE004610, 07EE005983, 12GE001331, 08FE003041, 11LE000364, 12AE000292, 11LE003613, 08EE001271, 07GE001141, 12HE004275, 08CE002530, 07IE001343, 13DE000957, 07LE002036, 07FE006663, 12KE003323, 12LE001582, 08DE000538, 09HE003084, 12IE001973, 08LE002965, 06LE005873, 09HE004442, 13FE000168, 08GE001831, 11FE001617, 12LE002365, 09FE001439, 07CE005298, 11LE001380, 08GE000560, 08GE002335, 08CE003903, 13JE000194, 11CE001979, 11GE000685, 12CE004391, 11DE004737, 08LE002903, 08EE001818, 07HE006200, 08IE001285, 08BE002785, 08EE005136, 07KE003554, 09AE003356, 08FE002424, 08BE000851, 07LE000852, 08IE003961, 09KE004214, 11AE001723, 08FE004650, 09KE000323, 08DE004055, 09GE005342, 10CE001513, 10AE000446, 10HE005497, 07IE002726, 09KE003154, 08HE004341, 09HE003619, 10FE002973, 07LE003935, 12EE004442, 11HE002725, 10BE000795, 08JE000793, 10FE004527, 10JE000292, 10IE002057, 08AE002871, 09HE004968, 11BE002101, 10CE001248, 11IE000284, 11DE005133, 10JE000751, 09EE002804, 09LE004393, 08BE002236, 07KE003811, 08DE003544, 12CE003042, 09IE002457, 10BE004391, 07KE003814, 11JE002429, 07LE001562, 07EE001297, 11JE003204, 09GE004950, 10FE000395, 08JE000546, 09HE003801, 07LE002035, 12BE004049, 12JE003532, 06LE000137, 09JE002179, 12GE003632, 08DE000554, 08EE005135, 11AE000834, 08DE002297, 07GE005451, 08EE004715, 13JE000142, 08KE002748, 08JE003335, 07EE002835, 08DE002763, 09LE002171, 09IE002592, 09DE003671, 10AE002896, 09DE005079, 09HE003287, 08KE002763, 10JE003933, 10LE004440, 09LE001388, 10IE003938, 09GE000746, 09JE004703, 09BE001163, 12CE005962, 11IE001335, 10BE000571, 14BE000043, 09LE001659, 09KE003780, 09CE001600, 12HE003936, 12BE003481, 11EE001820, 09KE004030, 09LE004678, 12BE001418, 09CE003659, 07CE008711, 07HE005274, 09FE005376, 10EE005191, 10KE000036, 12DE001266, 09DE001845, 09HE000245, 08GE003918, 10KE004775, 11FE002075, 10LE000312, 12IE003097, 09EE002840, 09JE004410, 12GE001516, 11CE005383, 12GE002269, 11BE000296, 07FE002842, 09JE004029, 09HE000037, 09AE003354, 11DE002762, 13HE000044, 11CE004257, 07HE001740, 09LE002161, 11AE001021, 07KE002831, 08EE005128, 07KE000805, 12BE003044, 09EE000037, 09LE002727, 07FE004829, 09IE001915, 08LE000671, 08IE000392, 12BE004050, 13HE000051, 09LE004167, 08GE003572, 08GE005252, 08LE002959, 09FE000728, 12EE000055, 08KE003449, 10CE002247, 09EE002101, 09HE005228, 09DE004574, 09GE000038, 12GE003262, 11JE002400, 10CE004193, 09BE002768, 07HE004756, 08HE000828, 13GE000134, 12IE000039, 09FE002356, 10EE001653, 08GE000564, 09JE003799, 12FE003738, 07HE006203, 12LE000819, 09IE005913, 10FE004260, 06LE005998, 11AE002801, 08AE004495, 08BE004243, 12KE002524, 10DE003639, 10FE005010, 08CE004729, 11GE001566, 08CE004702, 08CE002767, 13HE000472, 13BE000671, 10IE002319, 08FE002427, 09FE001180, 09GE001247, 08HE001463, 08HE000819, 09CE000581, 10IE002320, 10GE004834, 10LE001997, 10GE005273, 08IE003492, 08DE003817, 12LE001799, 10KE003672, 11CE002225, 09GE002191, 14DE000502, 08KE002545, 08EE000046, 12KE001103, 10LE001744, 11JE002665, 10LE000062, 10JE003673, 09BE000484, 09CE000794, 08LE003229, 11BE004595, 11KE002577, 11CE000023, 09GE001897, 09BE003955, 09DE003465, 08JE001043, 11DE003201, 11LE001339, 11DE000577, 08FE004891, 09DE005083, 09JE005173, 07HE005272, 11FE000329, 11LE002706, 13AE001307, 08IE004645, 09HE003284, 10EE005172, 08AE002870, 11FE001862, 12KE001245, 12BE003642, 09KE004266, 09JE003271, 11AE003224, 08BE002552, 11FE004506, 12LE002836, 08FE005385, 11BE001150, 10BE001305, 11DE000603, 09AE002229, 11HE003003, 13GE000360, 11CE001975, 11GE003709, 10BE002554, 12HE000044, 09JE004869, 11KE002209, 09KE004260, 12LE000450, 11GE003708, 10CE002456, 08IE003845, 10EE002451, 11KE002397, 10IE003936, 09DE002255, 13LE000419, 10GE003871, 08KE002997, 10FE005006, 11LE002525, 11CE002031, 09FE004868, 12KE003522, 10HE001920, 09FE001457, 11FE004537, 12AE001469, 11DE003045, 10AE001031, 09BE003501, 10LE005010, 09FE004544, 12KE003513, 12BE003299, 11FE004531, 10AE002406, 10IE003742, 11EE002086, 09LE003912, 13AE000151, 12BE003609, 11IE003254, 12BE003864, 10KE001921, 11IE004534, 10HE004950, 11BE003566, 12CE003867, 11JE002431, 12LE000874, 11BE004121, 12EE004660, 13BE000305, 11DE003052, 12DE002838, 11LE004814, 09HE004169, 07GE005907, 09LE000075, 09HE000051, 11EE002972, 12DE003747, 12DE001982, 10GE001448, 07IE002306, 10AE002866, 10FE003508, 14DE000368, 12HE000884, 12KE003042, 11CE003219, 07KE001392, 10FE000140, 09CE003242, 09LE000864, 12IE002286, 10IE002305, 07HE005225, 08DE004903, 11DE005134, 09GE000947, 09IE004487, 12JE003701, 08GE005013, 10EE003580, 08EE002583, 11GE002142, 10FE003241, 09HE003315, 13IE000320, 08BE002556, 11EE002075, 08CE001891, 11HE001070, 10GE005217, 08LE001866, 07GE005663, 11LE001090, 08EE003932, 08GE000265, 12DE000068, 12CE002129, 07HE001777, 07HE001222, 12HE003153, 09CE002074, 07EE006964, 12DE002197, 13AE000934, 10DE000060, 07IE001490, 09AE002727, 13DE000273, 08BE002997, 12IE002037, 13EE000058, 12FE000042, 07LE000647, 09HE005216, 08EE000035, 07GE005893, 13GE000032, 07IE000924, 08KE003219, 08DE001622, 09HE002362, 10IE004801, 09KE004029, 13IE000015, 08HE001472, 10KE004827, 08DE003825, 09IE004953, 11HE001028, 11EE001643, 08FE003625, 08KE000951, 07BE005320, 09CE002783, 10JE004986, 11KE002806, 12FE001824, 07GE004180, 10AE003115, 09HE001456, 09DE002537, 12EE000777, 13BE000159, 10KE004135, 07LE003022, 10GE004801, 09FE002081, 09JE002053, 09AE000512, 07HE004442, 10FE004524, 08GE001397, 10KE000121, 09GE002194, 09LE001119, 08AE003068, 08CE001400, 11LE002731, 09FE004106, 11HE003212, 11AE003864, 10CE002242, 09KE000886, 08HE004274, 08EE004171, 09FE000039, 11JE004471, 07EE002809, 10GE003597, 10AE002626, 09JE004221, 08FE001303, 10LE000587, 12DE001062, 07FE007020, 10IE003488, 07DE006471, 11JE001311, 09EE001344, 11AE000638, 13AE001295, 07FE007019, 10IE004264, 08BE003220, 13FE000262, 14EE000535, 10HE002691, 14EE000564, 08IE001018, 10JE003474, 08FE000812, 09FE005380, 09GE001268, 09HE003830, 09EE003319, 09CE003677, 10KE000122, 10JE002866, 09IE002316, 08BE004824, 07CE004842, 09LE004165, 09GE004021, 11AE001075, 10KE002926, 11CE005593, 09GE000927, 12FE003511, 09LE004166, 09HE004614, 11AE001449, 11HE001457, 12JE002326, 09LE000865, 12JE000791, 10JE003191, 09JE001836, 11DE002258, 10BE004394, 12LE001804, 10DE002659, 09BE004337, 11DE002260, 09KE000894, 11JE002981, 08AE000809, 10BE002920, 09IE002889, 12AE005013, 12CE005769, 09KE001164, 12AE001698, 08BE001310, 10AE000596, 09AE002001, 10KE004652, 08DE000518, 10CE002728, 12BE004443, 08DE004497, 10AE003946, 09FE000041, 10DE002245, 08DE001050, 10DE002465, 12EE002172, 11DE001669, 14EE000010, 10HE002198, 10BE003533, 08FE002422, 10EE002418, 12KE001604, 11HE003662, 08KE001113, 11LE003610, 09KE000370, 08KE002550, 10EE003101, 13CE000390, 10EE003263, 09GE002467, 07KE003220, 10CE002513, 08KE002546, 10FE001758, 14DE000017, 10GE001623, 10DE000919, 09HE001025, 10CE002747, 10CE003898, 10GE001922, 14FE000285, 11LE004275, 10KE001077, 11DE001182, 11AE001947, 10EE003579, 10IE000797, 11DE003449, 07FE007511, 11GE002337, 12IE002284, 13GE000809, 10FE002476, 10JE001847, 07DE001897, 09IE005308, 07JE002338, 08DE000307, 14DE000687, 08EE003121, 11AE004342, 08AE001466, 10CE003656, 10GE001673, 08HE002905, 08KE002774, 08HE003959, 11GE001579, 11LE004853, 10IE005101, 11HE000490, 08DE004718, 09IE000062, 09JE001259, 11GE001576, 08CE002078, 11GE001901, 10JE000018, 09DE002252, 10IE002071, 10BE002748, 11IE000886, 09EE003576, 10LE003422, 11KE002366, 11DE003681, 09IE000377, 10CE005198, 11LE002127, 10AE001218, 07BE004040, 13EE000135, 08GE003066, 10LE000877, 10LE004231, 11LE002586, 10GE001661, 07GE005068, 08IE000386, 09IE003439, 12JE001372, 11HE000346, 11LE003298, 10JE000057, 08FE001323, 09LE000819, 08HE001691, 08HE002904, 07LE002323, 12AE005073, 10JE001237, 08GE003076, 09LE003877, 06LE003960, 09DE004399, 08AE000095, 12AE005072, 10KE001914, 07JE000255, 08BE005061, 08HE002601, 12AE002625, 09EE002576, 12AE005077, 11IE001084, 08DE004266, 10GE003600, 08HE003224, 11HE002524, 10FE001157, 12AE005078, 11IE002157, 09FE004844, 11DE003200, 10CE003440, 08CE005427, 08DE003063, 12AE005079, 12BE001619, 09HE002625, 11EE004763, 10GE002192, 11KE001687, 08GE003910, 12BE004505, 12BE004687, 09JE001488, 11JE003442, 10DE001363, 14AE000343, 09BE000039, 12DE003312, 12KE001237, 10BE002102, 11KE002813, 10EE001471, 14DE000699, 09KE003781, 12DE003313, 09AE003935, 10CE000377, 12FE000321, 10IE004803, 07HE005227, 10HE003980, 12DE003311, 07IE002727, 10JE003629, 12KE002531, 11CE002944, 07FE002210, 07GE000068, 12DE003309, 13AE001252, 10KE000660, 13DE000557, 08CE002751, 07JE005907, 08KE001864, 12DE003310, 08LE004869, 10LE001481, 13GE000147, 12FE003709, 11DE003480, 10IE000561, 12DE003600, 09BE001186, 11JE000798, 13IE000045, 09BE001154, 08HE005196, 10JE003192, 12DE003601, 08LE000062, 12GE000578, 11HE003663, 09EE002314, 10LE001986, 11EE002937, 12JE002213, 08FE000076, 12KE001999, 11LE001587, 09GE003281, 09CE002994, 11LE002316, 10EE004896, 11FE002496, 07KE000223, 12LE002561, 09HE003608, 11EE002080, 07EE005195, 11CE005590, 08HE000041, 10IE000724, 08JE003638, 09KE001149, 07KE000424, 07HE006197, 12HE003682, 11JE003845, 08IE001081, 09FE002354, 11BE001440, 10FE001153, 07IE001742, 08HE003519, 12KE002537, 11IE002841, 09KE004263, 11FE005032, 10KE005291, 08AE000808, 12AE000057, 08DE001044, 09HE002856, 10KE001143, 11IE001534, 09FE000966, 08AE003563, 12FE003129, 08GE001829, 11LE000850, 12BE004729, 11KE001493, 08KE002766, 08GE005006, 12HE002984, 08LE002676, 08LE004665, 09BE002541, 08HE001121, 07IE004199, 09DE002538, 12IE001011, 13DE000922, 08KE004198, 12BE001134, 09EE001339, 09CE003021, 09EE002100, 07GE003513, 09AE002505, 11IE004311, 12BE004227, 10KE005252, 09CE002995, 09GE003570, 10GE001670, 07EE005416, 08FE000694, 12EE003636, 11FE003756, 10GE000988, 09JE002220, 12EE001227, 08HE004350, 08DE002290, 12GE002462, 12GE001512, 07KE004536, 09KE001439, 11EE000538, 08JE000306, 10EE000220, 12KE003523, 07GE006763, 08GE003067, 09KE003337, 11DE000628, 10CE002484, 11LE000361, 09GE000037, 08LE002667, 09CE000579, 09KE003779, 12FE002948, 10IE000063, 07KE003261, 09AE001728, 11AE003853, 07FE007547, 09KE004023, 12IE000607, 10JE003184, 08GE004133, 09AE002735, 11BE004196, 07FE007553, 09LE001658, 13JE000076, 10KE003922, 10CE004658, 12CE000920, 11HE001449, 08DE000826, 09LE004867, 13KE000059, 11AE003034, 11CE005170, 12EE001032, 12GE001345, 14DE000608, 09LE004679, 07IE004198, 11BE000957, 11DE004132, 12KE001417, 12HE000671, 07EE003824, 10BE003526, 10EE003002, 12GE000614, 11LE001374, 09LE000605, 12IE002291, 08EE003351, 10DE000561, 10HE004673, 13JE000417, 08CE001890, 10DE001885, 12DE003522, 11BE003755, 10EE002431, 08JE004949, 11EE002081, 08DE004902, 10DE001882, 10GE002862, 11IE002996, 10FE002964, 10HE005693, 08HE002897, 08JE003347, 10EE000435, 07FE005872, 07KE004489, 10FE005007, 12HE001241, 09LE001661, 09FE002367, 10FE000406, 07HE001218, 11LE003443, 10KE001653, 09JE000344, 11KE003753, 09GE002843, 10FE003512, 07GE004185, 08CE001642, 11AE003856, 10FE000401, 12KE001246, 09IE004191, 10IE001556, 08JE002896, 08CE003637, 11IE002836, 08FE000888, 13EE000893, 09IE004190, 10HE004948, 09FE005628, 08EE000044, 12LE001956, 08IE003978, 08JE002061, 09IE004160, 11IE002355, 09LE003295, 08EE000704, 07FE002771, 09DE003666, 09DE004395, 11LE002884, 11LE000371, 10DE001356, 07GE006960, 07KE004386, 11GE004348, 09JE003515, 13AE001255, 11KE001951, 10EE000022, 12DE003740, 08GE002534, 11HE001821, 12GE003064, 06JE007067, 12AE002620, 10KE001127, 13EE000352, 11CE004930, 08JE001340, 08AE001076, 09BE003022, 11JE003636, 09CE003230, 09AE003736, 14GE000110, 10FE001143, 09IE004747, 07GE005170, 12GE003063, 10AE002611, 08HE001118, 08HE002232, 11BE004402, 09GE000477, 08EE001585, 11IE000780, 11CE002260, 08JE000505, 09IE002864, 11CE002715, 09FE003743, 08EE001791, 11FE001615, 08HE003523, 10AE001519, 09LE000562, 11EE002525, 09FE003778, 08LE002969, 12IE000463, 10BE003920, 11AE003037, 08DE002714, 12JE000361, 09LE000017, 08LE002957, 10BE000579, 10GE002444, 11DE002251, 11EE000914, 11BE002624, 13HE000402, 09KE000603, 12KE001092, 13DE000746, 11EE001858, 11GE001897, 10BE001504, 09DE005076, 11JE000499, 13AE000314, 08EE004884, 13FE000683, 09JE001275, 11HE003380, 11IE000305, 10LE002312, 12FE001794, 12DE003958, 07EE001104, 11BE000726, 13DE000552, 11IE001757, 10LE004832, 11IE001081, 12JE000975, 09AE004167, 11BE002398, 12BE000037, 12CE000396, 08EE004136, 12FE001627, 11JE001345, 09FE005142, 10IE005096, 12CE005102, 10DE002337, 13CE000161, 12FE000319, 11KE002804, 09JE001832, 08HE004548, 13BE000586, 12FE003103, 07JE001321, 08BE004616, 11LE001797, 12AE005469, 09EE001156, 07FE002715, 08HE002903, 11FE002752, 08KE003404, 09LE001393, 07KE002513, 07DE003993, 09AE002508, 07HE001464, 12HE003439, 09EE001137, 10FE005569, 07IE004197, 08CE004486, 09CE000591, 08JE000771, 06IE008666, 09KE000599, 10JE002167, 08AE000815, 11HE001077, 07CE006316, 11GE002140, 09AE004406, 10HE003939, 12LE001964, 08AE001460, 12FE002533, 07GE003779, 10BE004560, 12EE003798, 12AE004446, 11CE000376, 08CE004461, 09BE000696, 07GE003019, 10GE000065, 12GE000910, 12AE004445, 11BE001902, 08EE005132, 10JE004723, 08FE003037, 10GE001445, 08LE001637, 09JE004448, 07HE007056, 08GE005257, 12KE001796, 08IE003962, 11EE004258, 12HE000440, 08IE001873, 12CE001278, 10HE005310, 07CE006037, 08LE002086, 12AE003252, 08GE003571, 08JE001857, 08CE003689, 10IE002938, 08AE000097, 08LE004699, 12JE000793, 10IE004236, 09DE003470, 07EE005179, 11FE002072, 08KE002570, 09AE004162, 08FE000890, 12KE003044, 09FE002353, 08IE002705, 11IE001721, 09KE002898, 09DE000041, 10EE003753, 11LE001612, 09HE004420, 09BE001671, 12CE001718, 10JE000288, 09DE001693, 08JE004678, 09HE002105, 08CE002545, 10JE004948, 12GE002274, 10LE000831, 09EE001142, 08EE000469, 09HE003295, 08LE000475, 11CE003271, 09JE004169, 11CE004931, 09EE002097, 14BE000256, 10IE003191, 09KE000588, 11IE001325, 10DE004960, 11HE002998, 10GE005204, 14DE000058, 11LE003402, 09LE002451, 11KE001962, 11AE000831, 11CE000796, 08KE001107, 08KE000350, 07IE000965, 10KE002175, 12FE000726, 12GE000013, 07HE003346, 09FE000722, 09EE003796, 10LE000810, 11FE000326, 12HE000045, 08AE001714, 12CE001956, 11IE000042, 07GE000071, 13IE000297, 12KE002694, 12HE004254, 07GE005449, 11CE002264, 12BE001170, 11HE002992, 07AE000992, 13JE000511, 12KE001097, 08HE001971, 12JE003535, 13BE000263, 12BE000038, 09IE003663, 09CE005217, 14AE000181, 08JE001042, 06JE005407, 07JE004503, 09CE005225, 08JE003891, 09FE003732, 14EE000615, 10GE003611, 07GE000004, 06GE005314, 09GE002842, 12CE002400, 09HE004168, 08GE003065, 07DE004005, 07FE002656, 07BE001777, 13LE000542, 13GE000040, 10HE001202, 07JE002081, 09JE003505, 07GE001089, 07DE001600, 14AE000178, 07GE007036, 12EE001915, 12JE002773, 10IE002315, 08DE000304, 07GE004938, 11EE002579, 09BE003316, 12EE003051, 08AE003336, 11BE001161, 07IE001342, 07JE001641, 08LE000930, 12JE001739, 09DE003201, 08HE000291, 11BE002428, 10JE000745, 07JE001821, 10HE004334, 08GE000285, 12JE000983, 08LE000675, 11EE003369, 09IE000822, 09LE000315, 08HE002245, 09DE001699, 09IE000041, 10EE002449, 11JE002188, 08IE003724, 10AE002586, 10AE004224, 09IE000827, 09IE000036, 10HE000374, 12HE004445, 08IE003490, 10AE003278, 12BE001685, 09KE000075, 10BE001514, 10IE003428, 10JE000750, 09FE003044, 10CE001474, 13GE000856, 09FE002625, 10CE002992, 12HE002157, 09DE002253, 10CE001007, 10EE002375, 10LE003140, 13GE000136, 12EE002398, 12KE000087, 09JE001069, 10IE002068, 10EE002387, 11GE002330, 08HE000570, 10GE001002, 12LE002166, 10HE002686, 12FE001634, 10EE002386, 12CE006029, 10CE003697, 11FE001810, 09CE002077, 11IE000274, 07GE006407, 10EE004217, 10CE000062, 12JE002156, 13GE000202, 10BE002743, 12DE000209, 10HE001426, 10EE004218, 10LE003464, 14CE000138, 10FE003506, 10GE005061, 12GE002098, 10BE001505, 10FE000648, 11LE001678, 08FE000463, 10FE005562, 11EE003823, 08FE000072, 09HE004195, 10FE000649, 12FE001011, 12AE004935, 10GE004132, 11FE002080, 10AE003719, 08FE001363, 10FE000650, 12IE002787, 10EE003571, 11BE000060, 11FE002497, 08IE005162, 09GE004560, 10FE001909, 12IE003245, 14CE000495, 11EE004011, 08GE004775, 09BE001634, 10DE003641, 10JE000737, 07IE003629, 10CE001960, 09GE004935, 12GE001696, 09KE001990, 10EE003746, 10JE001784, 07KE001529, 11CE000298, 09KE000557, 07KE004154, 10HE002404, 10IE004512, 10KE004141, 08AE003577, 06LE007409, 10KE001915, 09DE000050, 09DE004576, 11DE003478, 10KE004315, 08HE000060, 09CE003690, 10KE003682, 09BE002555, 09HE002382, 13FE000616, 11AE002805, 08IE004246, 08GE003087, 11EE002327, 09DE005080, 09JE001823, 08BE000280, 11CE001977, 08KE003385, 09EE004657, 11IE002387, 08CE000052, 10JE004716, 09DE001155, 11DE003674, 09AE005327, 09GE000761, 12BE002811, 08HE004349, 10JE004987, 10CE001724, 11JE000043, 09AE004382, 08AE000261, 11CE002505, 09IE002099, 10LE005416, 11HE003390, 12CE000919, 09EE000048, 08KE001607, 11EE004475, 11AE001080, 11AE003858, 09AE004409, 13IE000421, 12LE002731, 09LE002923, 11LE002064, 07IE002722, 11IE003255, 10EE004881, 13IE000612, 07HE001408, 10KE000644, 12KE000090, 07FE007288, 07JE000042, 11HE002276, 07CE005297, 08HE005009, 10KE000643, 14CE000480, 07JE003813, 09GE001681, 08IE001607, 08DE002291, 08IE004660, 10LE003663, 10LE003105, 06KE006044, 13DE000752, 11JE000233, 10FE003207, 09FE005586, 07JE005624, 06JE006242, 08EE000483, 07KE003039, 11DE004352, 10IE002309, 09GE004023, 09JE002938, 08AE003065, 09DE004178, 07GE000072, 07EE005412, 10KE003881, 09HE001043, 09LE002457, 08BE003886, 09DE003194, 07IE000136, 07HE005471, 11AE002638, 09JE002975, 10GE000806, 10GE001870, 12AE001878, 08CE003313, 09DE001652, 11JE001349, 10HE000072, 08CE000588, 10JE002918, 07EE007022, 08DE000045, 10AE001483, 11JE003171, 10HE004004, 09JE004933, 11BE000476, 10DE002108, 08HE002916, 10HE003260, 11LE002529, 10KE004942, 10LE000559, 11CE001778, 11BE003563, 09KE001393, 10IE004796, 13GE000583, 11AE001740, 10LE002533, 12EE004457, 11EE003372, 10DE001843, 10KE002739, 07EE001744, 11DE001327, 11IE002384, 10FE004795, 11LE002733, 10EE001646, 11EE002283, 07CE005265, 11IE001400, 11IE003250, 11GE001290, 12BE002817, 12BE001822, 11HE001667, 07GE006548, 11JE002736, 13AE000983, 11LE001607, 12JE000046, 12BE002307, 12AE003656, 07HE001783, 12FE000604, 08FE005388, 12KE003718, 10FE003777, 12CE002663, 12CE000951, 07IE003978, 11AE000874, 10DE004537, 11JE004482, 10JE001228, 12EE002757, 12FE002286, 07LE002537, 11AE000875, 11HE000517, 12JE000323, 10JE003872, 12KE000877, 14AE000538, 08EE003356, 11FE002753, 09KE001147, 12KE004136, 11HE002513, 07HE005503, 08CE000036, 08FE002202, 09CE002776, 10BE000307, 11GE001896, 11KE000425, 08EE002585, 07LE004212, 08FE003316, 11DE004536, 08IE004650. |
Recalling Firm/ Manufacturer |
Invacare Corporation 1200 Taylor St Elyria OH 44035-6248
|
| For Additional Information Contact | Swasita Saigal 440-329-6595 |
Manufacturer Reason for Recall | If the slack in the wires is not routed and secured correctly, flexing of the back pan under the users weight may cause the headrest knob or clamp bracket to pinch, damage or cut the wire. Smoking, sparking, burning or fire may occur if a wire is repetitively pinched at the same place. |
FDA Determined Cause 2 | Device Design |
| Action | The firm initiated the recall on 01/16/15 by sending a notification letter to customers through the mail. Communication was sent to providers to contact consignees via mail return receipt to inform them of the recall. Recall notification included a Provider letter, Consumer letter, Provider Response card, Unit Disposition form and tracking sheet and the Wiring guide. The notification instructed the provider to visually inspect all potentially impacted chairs. Not all potentially impacted chairs are expected to have improperly routed and secured wires. |
| Quantity in Commerce | 1942 |
| Distribution | Worldwide Distribution-US (nationwide) including DC and the states of AK, AL, AR, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, VT, WA, WI, and WV, and the countries of Canada and Australia. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
| 510(K) Database | 510(K)s with Product Code = ITI
|
|
|
|