| | Class 2 Device Recall TUBING 7/8" X 10', W/WAND & TIP |  |
| Date Initiated by Firm | March 10, 2015 |
| Date Posted | March 27, 2015 |
| Recall Status1 |
Terminated 3 on January 21, 2017 |
| Recall Number | Z-1344-2015 |
| Recall Event ID |
70721 |
| 510(K)Number | K001082 K910515 |
| Product Classification |
Apparatus, exhaust, surgical - Product Code FYD
|
| Product | VACUUM TUBING 7/8" X 6' (2.22cm X 182.88cm), REF 24009, QTY 24\CS, NON-STERILE
Intended use: The smoke evacuator and accessories are intended to evacuate the smoke created by electrosurgery, laser surgery, or power tool surgical procedures. Locations include Operating Rooms, Trauma, Endoscopy and Laparoscopy Sites. |
| Code Information |
Lot Numbers: C9344, C10091, C10158, C10214, C10221, C10257, C10264, C10298, C10342, C11060, C11104, C11118, C11136, C11180, C11216, C11237, C11290, C11313, C11334, C12060, C12096, C12180, C12206, C12235, C12243, C12334, C12355, C13008, C13032, C13042, C13053, C13067, C13128, C13179, C13212, C13248, C13281, C13303, C13317, C13338, C14036, C14057, C14121, C14140, C14169, C14183, C14197, C14233 |
Recalling Firm/ Manufacturer |
Microtek Medical Inc 602 N Lehmberg Rd Columbus MS 39702-4406
|
| For Additional Information Contact | Roman Blahoski 651-250-4385 |
Manufacturer Reason for Recall | This recall is being conducted because a cohesive band used to package the devices contains latex and the product is not labeled as required by 21 CFR 801.437; it does not contain Caution: The Packaging of This Product Contains Natural Rubber Latex Which May Cause Allergic Reactions. |
FDA Determined Cause 2 | Error in labeling |
| Action | The firm, Ecolab - Microtek Medical, sent an "Urgent: Recall - Mislabeled Product Immediate Action Required" letter dated March 2015 to its US customers on 03/10/2015 via letter and email. The letter describes product, problem and actions to be taken. The customers were instructed to immediately examine your inventory and promptly quarantine the lot(s) listed; complete and return the enclosed Recall Response Form via fax to Quality Assurance Manager, Ecolab-Microtek Medical at (662) 244-3011 or email to: recall@ecolab.com RE: Recall - Mislabeled Product ASAP, and if you have further distributed the products subject to this recall, please notify your customers at once.
Upon receipt of the completed response forms, Ecolab-Microtek Medical will issue a return material authorization (RMA) and ask that you return any unopened product listed above to Ecolab-Microtek Medical at the address listed in the attached form.
If you have any questions in regard to this recall, please contact Customer Service by phone at 1-800-824-3027 or email; customerservice@microtekmed.com. |
| Quantity in Commerce | 30796 units |
| Distribution | US (nationwide) Distribution. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
| 510(K) Database | 510(K)s with Product Code = FYD 510(K)s with Product Code = FYD
|
|
|
|