• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall Zimmer Natural Nail CM Long

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Zimmer Natural Nail CM Longsee related information
Date Initiated by FirmSeptember 03, 2015
Date PostedOctober 14, 2015
Recall Status1 Terminated 3 on March 15, 2016
Recall NumberZ-0119-2016
Recall Event ID 72123
510(K)NumberK091566 
Product Classification Rod, fixation, intramedullary and accessories - Product Code HSB
ProductZimmer Natural Nail CM Long. Intramedullary Fixation Rod. The Zimmer Natural Nail System is intended for temporary fracture fixation and stabilization of the bone.
Code Information All Lots. Item number and name: 47-2493-481-10 ZNN CMN Nail 10mmX48cm 125 L 47-2493-442-10 ZNN CMN Nail 10mmX44cm 130 R 47-2493-480-10 ZNN CMN Nail 10mmX48cm 125 R 47-2493-445-10 ZNN CMN Nail 10mmX44cm 135 L 47-2493-483-10 ZNN CMN Nail 10mmX48cm 130 L 47-2493-444-10 ZNN CMN Nail 10mmX44cm 135 R 47-2493-482-10 ZNN CMN Nail 10mmX48cm 130 R 47-2493-441-11 ZNN CMN Nail 11.5mmX44cm 125 L 47-2493-485-10 ZNN CMN Nail 10mmX48cm 135 L 47-2493-440-11 ZNN CMN Nail 11.5mmX44cm 125 R 47-2493-484-10 ZNN CMN Nail 10mmX48cm 135 R 47-2493-443-11 ZNN CMN Nail 11.5mmX44cm 130 L 47-2493-481-11 ZNN CMN Nail 11.5mmX48cm 125 L 47-2493-442-11 ZNN CMN Nail 11.5mmX44cm 130 R 47-2493-480-11 ZNN CMN Nail 11.5mmX48cm 125 R 47-2493-445-11 ZNN CMN Nail 11.5mmX44cm 135 L 47-2493-483-11 ZNN CMN Nail 11.5mmX48cm 130 L 47-2493-444-11 ZNN CMN Nail 11.5mmX44cm 135 R 47-2493-482-11 ZNN CMN Nail 11.5mmX48cm 130 R 47-2493-441-13 ZNN CMN Nail 13mmX44cm 125 L 47-2493-485-11 ZNN CMN Nail 11.5mmX48cm 135 L 47-2493-440-13 ZNN CMN Nail 13mmX44cm 125 R 47-2493-484-11 ZNN CMN Nail 11.5mmX48cm 135 R 47-2493-443-13 ZNN CMN Nail 13mmX44cm 130 L 47-2493-481-13 ZNN CMN Nail 13mmX48cm 125 L 47-2493-442-13 ZNN CMN Nail 13mmX44cm 130 R 47-2493-480-13 ZNN CMN Nail 13mmX48cm 125 R 47-2493-445-13 ZNN CMN Nail 13mmX44cm 135 L 47-2493-483-13 ZNN CMN Nail 13mmX48cm 130 L 47-2493-444-13 ZNN CMN Nail 13mmX44cm 135 R 47-2493-482-13 ZNN CMN Nail 13mmX48cm 130 R 47-2493-421-10 ZNN CMN Nail 10mmX42cm 125 L 47-2493-485-13 ZNN CMN Nail 13mmX48cm 135 L 47-2493-420-10 ZNN CMN Nail 10mmX42cm 125 R 47-2493-484-13 ZNN CMN Nail 13mmX48cm 135 R 47-2493-423-10 ZNN CMN Nail 10mmX42cm 130 L 47-2493-463-13 ZNN CMN Nail 13mmX46cm 130 L 47-2493-422-10 ZNN CMN Nail 10mmX42cm 130 R 47-2493-461-10 ZNN CMN Nail 10mmX46cm 125 L 47-2493-425-10 ZNN CMN Nail 10mmX42cm 135 L 47-2493-460-10 ZNN CMN Nail 10mmX46cm 125 R 47-2493-424-10 ZNN CMN Nail 10mmX42cm 135 R 47-2493-463-10 ZNN CMN Nail 10mmX46cm 130 L 47-2493-421-11 ZNN CMN Nail 11.5mmX42cm 125 L 47-2493-462-10 ZNN CMN Nail 10mmX46cm 130 R 47-2493-420-11 ZNN CMN Nail 11.5mmX42cm 125 R 47-2493-465-10 ZNN CMN Nail 10mmX46cm 135 L 47-2493-423-11 ZNN CMN Nail 11.5mmX42cm 130 L 47-2493-464-10 ZNN CMN Nail 10mmX46cm 135 R 47-2493-422-11 ZNN CMN Nail 11.5mmX42cm 130 R 47-2493-461-11 ZNN CMN Nail 11.5mmX46cm 125 L 47-2493-425-11 ZNN CMN Nail 11.5mmX42cm 135 L 47-2493-460-11 ZNN CMN Nail 11.5mmX46cm 125 R 47-2493-424-11 ZNN CMN Nail 11.5mmX42cm 135 R 47-2493-463-11 ZNN CMN Nail 11.5mmX46cm 130 L 47-2493-421-13 ZNN CMN Nail 13mmX42cm 125 L 47-2493-462-11 ZNN CMN Nail 11.5mmX46cm 130 R 47-2493-420-13 ZNN CMN Nail 13mmX42cm 125 R 47-2493-465-11 ZNN CMN Nail 11.5mmX46cm 135 L 47-2493-423-13 ZNN CMN Nail 13mmX42cm 130 L 47-2493-464-11 ZNN CMN Nail 11.5mmX46cm 135 R 47-2493-422-13 ZNN CMN Nail 13mmX42cm 130 R 47-2493-461-13 ZNN CMN Nail 13mmX46cm 125 L 47-2493-425-13 ZNN CMN Nail 13mmX42cm 135 L 47-2493-460-13 ZNN CMN Nail 13mmX46cm 125 R 47-2493-424-13 ZNN CMN Nail 13mmX42cm 135 R 47-2493-462-13 ZNN CMN Nail 13mmX46cm 130 R 47-2493-401-10 ZNN CMN Nail 10mmX40cm 125 L 47-2493-465-13 ZNN CMN Nail 13mmX46cm 135 L 47-2493-400-10 ZNN CMN Nail 10mmX40cm 125 R 47-2493-464-13 ZNN CMN Nail 13mmX46cm 135 R 47-2493-403-10 ZNN CMN Nail 10mmX40cm 130 L 47-2493-441-10 ZNN CMN Nail 10mmX44cm 125 L 47-2493-402-10 ZNN CMN Nail 10mmX40cm 130 R 47-2493-440-10 ZNN CMN Nail 10mmX44cm 125 R 47-2493-405-10 ZNN CMN Nail 10mmX40cm 135 L 47-2493-443-10 ZNN CMN Nail 10mmX44cm 130 L 47-2493-404-10 ZNN CMN Nail 10mmX40cm 135 R 47-2493-401-11 ZNN CMN Nail 11.5mmX40cm 125 L 47-2493-340-10 ZNN CMN Nail 10mmX34cm 125 R 47-2493-400-11 ZNN CMN Nail 11.5mmX40cm 125 R 47-2493-343-10 ZNN CMN Nail 10mmX34cm 130 L 47-2493-403-11 ZNN CMN Nail 11.5mmX40cm 130 L 47-2493-342-10 ZNN CMN Nail 10mmX34cm 130 R 47-2493-402-11 ZNN CMN Nail 11.5mmX40cm 130 R 47-2493-341-11 ZNN CMN Nail 11.5mmX34cm 125 L 47-2493-405-11 ZNN CMN Nail 11.5mmX40cm 135 L 47-2493-340-11 ZNN CMN Nail 11.5mmX34cm 125 R 47-2493-404-11 ZNN CMN Nail 11.5mmX40cm 135 R 47-2493-343-11 ZNN CMN Nail 11.5mmX34cm 130 L 47-2493-401-13 ZNN CMN Nail 13mmX40cm 125 L 47-2493-342-11 ZNN CMN Nail 11.5mmX34cm 130 R 47-2493-400-13 ZNN CMN Nail 13mmX40cm 125 R 47-2493-341-13 ZNN CMN Nail 13mmX34cm 125 L 47-2493-403-13 ZNN CMN Nail 13mmX40cm 130 L 47-2493-340-13 ZNN CMN Nail 13mmX34cm 125 R 47-2493-402-13 ZNN CMN Nail 13mmX40cm 130 R 47-2493-343-13 ZNN CMN Nail 13mmX34cm 130 L 47-2493-405-13 ZNN CMN Nail 13mmX40cm 135 L 47-2493-342-13 ZNN CMN Nail 13mmX34cm 130 R 47-2493-404-13 ZNN CMN Nail 13mmX40cm 135 R 47-2493-321-10 ZNN CMN Nail 10mmX32cm 125 L 47-2493-381-10 ZNN CMN Nail 10mmX38cm 125 L 47-2493-320-10 ZNN CMN Nail 10mmX32cm 125 R 47-2493-380-10 ZNN CMN Nail 10mmX38cm 125 R 47-2493-323-10 ZNN CMN Nail 10mmX32cm 130 L 47-2493-383-10 ZNN CMN Nail 10mmX38cm 130 L 47-2493-322-10 ZNN CMN Nail 10mmX32cm 130 R 47-2493-382-10 ZNN CMN Nail 10mmX38cm 130 R 47-2493-321-11 ZNN CMN Nail 11.5mmX32cm 125 L 47-2493-381-11 ZNN CMN Nail 11.5mmX38cm 125 L 47-2493-320-11 ZNN CMN Nail 11.5mmX32cm 125 R 47-2493-380-11 ZNN CMN Nail 11.5mmX38cm 125 R 47-2493-323-11 ZNN CMN Nail 11.5mmX32cm 130 L 47-2493-383-11 ZNN CMN Nail 11.5mmX38cm 130 L 47-2493-322-11 ZNN CMN Nail 11.5mmX32cm 130 R 47-2493-382-11 ZNN CMN Nail 11.5mmX38cm 130 R 47-2493-321-13 ZNN CMN Nail 13mmX32cm 125 L 47-2493-381-13 ZNN CMN Nail 13mmX38cm 125 L 47-2493-320-13 ZNN CMN Nail 13mmX32cm 125 R 47-2493-380-13 ZNN CMN Nail 13mmX38cm 125 R 47-2493-323-13 ZNN CMN Nail 13mmX32cm 130 L 47-2493-383-13 ZNN CMN Nail 13mmX38cm 130 L 47-2493-322-13 ZNN CMN Nail 13mmX32cm 130 R 47-2493-382-13 ZNN CMN Nail 13mmX38cm 130 R 47-2493-301-10 ZNN CMN Nail 10mmX30cm 125 L 47-2493-361-10 ZNN CMN Nail 10mmX36cm 125 L 47-2493-300-10 ZNN CMN Nail 10mmX30cm 125 R 47-2493-360-10 ZNN CMN Nail 10mmX36cm 125 R 47-2493-303-10 ZNN CMN Nail 10mmX30cm 130 L 47-2493-363-10 ZNN CMN Nail 10mmX36cm 130 L 47-2493-302-10 ZNN CMN Nail 10mmX30cm 130 R 47-2493-362-10 ZNN CMN Nail 10mmX36cm 130 R 47-2493-301-11 ZNN CMN Nail 11.5mmX30cm 125 L 47-2493-361-11 ZNN CMN Nail 11.5mmX36cm 125 L 47-2493-300-11 ZNN CMN Nail 11.5mmX30cm 125 R 47-2493-360-11 ZNN CMN Nail 11.5mmX36cm 125 R 47-2493-303-11 ZNN CMN Nail 11.5mmX30cm 130 L 47-2493-363-11 ZNN CMN Nail 11.5mmX36cm 130 L 47-2493-302-11 ZNN CMN Nail 11.5mmX30cm 130 R 47-2493-362-11 ZNN CMN Nail 11.5mmX36cm 130 R 47-2493-301-13 ZNN CMN Nail 13mmX30cm 125 L 47-2493-361-13 ZNN CMN Nail 13mmX36cm 125 L 47-2493-300-13 ZNN CMN Nail 13mmX30cm 125 R 47-2493-360-13 ZNN CMN Nail 13mmX36cm 125 R 47-2493-303-13 ZNN CMN Nail 13mmX30cm 130 L 47-2493-363-13 ZNN CMN Nail 13mmX36cm 130 L 47-2493-302-13 ZNN CMN Nail 13mmX30cm 130 R 47-2493-362-13 ZNN CMN Nail 13mmX36cm 130 R 47-2493-305-10 ZNN CMN Nail 10mmX30cm 135 L 47-2493-341-10 ZNN CMN Nail 10mmX34cm 125 L 47-2493-305-11 ZNN CMN Nail 11.5mmX30cm 135 L 47-2493-325-10 ZNN CMN Nail 10mmX32cm 135 L 47-2493-305-13 ZNN CMN Nail 13mmX30cm 135 L 47-2493-325-11 ZNN CMN Nail 11.5mmX32cm 135 L 47-2493-365-13 ZNN CMN Nail 13mmX36cm 135 L 47-2493-325-13 ZNN CMN Nail 13mmX32cm 135 L 47-2493-344-10 ZNN CMN Nail 10mmX34cm 135 R 47-2493-304-10 ZNN CMN Nail 10mmX30cm 135 R 47-2493-344-11 ZNN CMN Nail 11.5mmX34cm 135 R 47-2493-304-11 ZNN CMN Nail 11.5mmX30cm 135 R 47-2493-344-13 ZNN CMN Nail 13mmX34cm 135 R 47-2493-304-13 ZNN CMN Nail 13mmX30cm 135 R 47-2493-364-10 ZNN CMN Nail 10mmX36cm 135 R 47-2493-324-10 ZNN CMN Nail 10mmX32cm 135 R 47-2493-364-11 ZNN CMN Nail 11.5mmX36cm 135 R 47-2493-324-11 ZNN CMN Nail 11.5mmX32cm 135 R 47-2493-364-13 ZNN CMN Nail 13mmX36cm 135 R 47-2493-324-13 ZNN CMN Nail 13mmX32cm 135 R 47-2493-385-10 ZNN CMN Nail 10mmX38cm 135 L 47-2493-345-10 ZNN CMN Nail 10mmX34cm 135 L 47-2493-385-11 ZNN CMN Nail 11.5mmX38cm 135 L 47-2493-345-11 ZNN CMN Nail 11.5mmX34cm 135 L 47-2493-385-13 ZNN CMN Nail 13mmX38cm 135 L 47-2493-345-13 ZNN CMN Nail 13mmX34cm 135 L 47-2493-384-10 ZNN CMN Nail 10mmX38cm 135 R 47-2493-365-10 ZNN CMN Nail 10mmX36cm 135 L 47-2493-384-11 ZNN CMN Nail 11.5mmX38cm 135 R 47-2493-365-11 ZNN CMN Nail 11.5mmX36cm 135 L 47-2493-384-13 ZNN CMN Nail 13mmX38cm 135 R
FEI Number 3005233524
Recalling Firm/
Manufacturer
Zimmer Gmbh
Sulzer Allee 8
Winterthur Switzerland
For Additional Information ContactConsumer Relations Call Center
800-447-5633
Manufacturer Reason
for Recall
In about 5% of tested samples, the outer tray was found to be compromised. Risks: A damaged outer tray detected during surgery may cause a delay in surgery to obtain another implant. If a compromised packaging is not detected prior or during a surgery by the user, it may lead to periprosthetic infection. This may result in revision or multi-stage revision to treat the infection.
FDA Determined
Cause 2
Package design/selection
ActionOn 9/3/2015, URGENT MEDICAL DEVICE RECALL notifications were sent via courier to the affected distributors and Hospital Risk Managers with instructions for locating and inspecting the affected devices and contacting Zimmer if damage is observed. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. If after reviewing this notification you have questions concerning this notification, please call 1-877-946-2761.
Quantity in Commerce57,603 units
DistributionUS Nationwide distribution.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
510(K) Database510(K)s with Product Code = HSB
-
-