New Search Back to Search Results
  Class 2 Device Recall Customed see related information
Date Initiated by Firm January 07, 2016
Date Posted April 14, 2016
Recall Status1 Terminated 3 on May 03, 2017
Recall Number Z-1447-2016
Recall Event ID 72992
Product Classification General surgery tray (kit) - Product Code LRO
Product Catalog Number: 9002259A
LAPAROSCOPY-COLORECTAL SURGICAL PACK
Used by medical staff to perform surgical procedures and patient care procedures such as wound healing, suture removals, irrigate wounds, etc.
Code Information LOTS: SL15173 (150418356) 31-MAY-16 SL15195-01 (150619256) 30-JUN-16 SL15199 (150619256) 30-JUN-16 SL15208 (150719758) 31-JUL-16 SL15212 (150719758) 31-JUL-16 SL15224 (150719758) 31-JUL-16 SL15243 (150820288) 31-AUG-16 SL15260 (150920289) 30-SEP-16  
Recalling Firm/
Manufacturer
Customed, Inc
Calle Igualdad Final #7
Fajardo PR 00738
Manufacturer Reason
for Recall
Recall initiated due to inadequate validation/qualification of sterilization processes, and package integrity for products released for distribution as sterile.
FDA Determined
Cause 2
Packaging
Action Recall Notification Letter sent to direct accounts via certified mail with return receipt on January 7, 2016 and January 8, 2016.
Quantity in Commerce 84
Distribution Distributed only in Puerto Rico.
Total Product Life Cycle TPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
-
-