Date Initiated by Firm | January 07, 2016 |
Date Posted | April 14, 2016 |
Recall Status1 |
Terminated 3 on May 03, 2017 |
Recall Number | Z-1450-2016 |
Recall Event ID |
72992 |
Product Classification |
General surgery tray (kit) - Product Code LRO
|
Product | Catalog Number: 9002511
MINOR SURGICAL PACK
Used by medical staff to perform surgical procedures and patient care procedures such as wound healing, suture removals, irrigate wounds, etc. |
Code Information |
LOTS: SL15170 (150317841) 30-APR-16 SL15170 (150317843) 30-APR-16 SL15180 (150619260) 30-JUN-16 SL15184 (150317844) 30-APR-16 SL15199 (150719841) 31-JUL-16 SL15203 (150719841) 31-JUL-16 SL15229 (150820295) 31-AUG-16 SL15245 (150820296) 31-AUG-16 SL15246 (150820296) 31-AUG-16 SL15267 (150920885) 31-AUG-16 SL15268 (150920886) 31-AUG-16 SL15271 (150920887) 31-AUG-16 SL15273 (150920884) 31-AUG-16 SL15273 (150920809) 31-AUG-16 |
Recalling Firm/ Manufacturer |
Customed, Inc Calle Igualdad Final #7 Fajardo PR 00738
|
Manufacturer Reason for Recall | Recall initiated due to inadequate validation/qualification of sterilization processes, and package integrity for products released for distribution as sterile. |
FDA Determined Cause 2 | Packaging |
Action | Recall Notification Letter sent to direct accounts via certified mail with return receipt on January 7, 2016 and January 8, 2016. |
Quantity in Commerce | 344 |
Distribution | Distributed only in Puerto Rico. |
Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|