Date Initiated by Firm | January 11, 2016 |
Date Posted | March 12, 2016 |
Recall Status1 |
Terminated 3 on April 13, 2017 |
Recall Number | Z-1110-2016 |
Recall Event ID |
72900 |
510(K)Number | K840643 |
Product Classification |
Prosthesis, hip, semi-constrained, metal/polymer, cemented - Product Code JDI
|
Product | Product 12 consists of all product under product code JDI
, and same usage:
Item no:
902602935 6 DEGREE COCR FEM HEAD 28
902602835 6 DEGREE COCR FEM HEAD 28
902603235 6 DEGREE COCR FEM HEAD 32
902603335 6 DEGREE COCR FEM HEAD 32
902603600 6 DEGREE COCR FEM HEAD 36
902603610 6 DEGREE COCR FEM HEAD 36
902603614 6 DEGREE COCR FEM HEAD 36
Product Usage:
For use in total or hemi hip arthroplasty |
Code Information |
lot no62240860 62339639 62395853 62240859 62279765 62339640 62395852 62089508 62094969 62185962 62240862 62273366 62279766 61660149 62279769 62284368 62284368N 62284371 62093453 |
Recalling Firm/ Manufacturer |
Zimmer Manufacturing B.V. Turpeaux Industrial Park Rd 1 km 123.4 Mercedita PR 00715
|
For Additional Information Contact | Consumer Relations Call Center 800-447-5633 |
Manufacturer Reason for Recall | As a result of the insufficient sealer calibration data for product packaged in firm's Building II between August 2010 and April 2013. The affected products are sterile. |
FDA Determined Cause 2 | Packaging |
Action | Zimmer Biomet sent an Urgent Medical Device Recall Letter, dated January 11, 2016, to Distributors, Sales Representatives, Operation Managers, and Risk Managers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to
1.Review the notification and ensure that relevant personnel are aware of the contents.
2. Assist your Zimmer Biomet sales representative with the quarantine of any affected product.
3. Your Zimmer Biomet sales representative will remove the recalled product from your facility.
4. Complete and return the attached Certificate of Acknowledgment form to corporatequality.postmarket@zimmerbiomet.com.
5. If after reviewing this notification you have further questions or concerns please call Customer Service at 1-800-348-2759, or contact your Zimmer Biomet Sales Representative. |
Quantity in Commerce | 209 |
Distribution | US Nationwide Distribution |
Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database | 510(K)s with Product Code = JDI
|