Date Initiated by Firm | March 25, 2016 |
Create Date | May 15, 2016 |
Recall Status1 |
Terminated 3 on May 23, 2016 |
Recall Number | Z-1671-2016 |
Recall Event ID |
73786 |
Product Classification |
Mass spectrometer, clinical use - Product Code DOP
|
Product | AB Sciex Triple Quad 4500MD LC/MS/MS System, Part No. 5031257; and QTRAP 4500MD LC/MS/MS System, Part No. 5031231.
Mass spectrometers for in vitro diagnostic use only. |
Code Information |
Serial Numbers: Triple Quad 4500MD LC/MS/MS System BX21081601 BX21101601 BX21091601 BX21111601 BX21121602 BX21131602 BX21141602 BX21151602 QTRAP 4500MD LC/MS/MS System BW20251601 BW20241601 BW20261602 |
Recalling Firm/ Manufacturer |
Ab Sciex 500 Old Connecticut Path Framingham MA 01701-4574
|
Manufacturer Reason for Recall | Under certain conditions, mass spectrometers may report incorrect quantitative results. |
FDA Determined Cause 2 | Under Investigation by firm |
Action | AB SCIEX issued Urgent: Medical Device Correction letter, dated 3/25/16, advising user of the problem and advising not to use the system until corrected. The letter also discussed the risk to health, as well as how to recognize if the issue has occurred. Sciex Field Service Employee (FSE) will be sent to customer site to perform the field corrective action on the affected system. Questions regarding the letter should be directed to SCIEX at 1-289-982-2531. |
Quantity in Commerce | 2 units |
Distribution | Distributed to the state of NC. |
Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|