 Class 2 Device Recall American Surgicalsee related information
Date Initiated by FirmMarch 31, 2016
Create DateMay 26, 2016
Recall Status1 Terminated 3 on October 21, 2016
Recall NumberZ-1798-2016
Recall Event ID 73734
510(K)NumberK902921 
Product Classification Neurosurgical paddie - Product Code HBA
ProductAmerican Surgical Uniqcot 1/2" x 2" Ref Number: 67-29 Product Usage: Neurosurgical and Nonneurosurgical Sponges
Code Information Lot numbers:  2015/09 BG 
FEI Number 1221144
Recalling Firm/
Manufacturer
American Surgical Company
45 Congress St
Salem MA 01970-5579
Manufacturer Reason
for Recall
Product packaging defective compromising sterility
FDA Determined
Cause 2
Packaging process control
ActionAmerican Surgical issued an Urgent Recall and Field Safety Notice letter dated March 31, 2016 to affected customers. The letter identified the affected product, problem and actions to be taken. Customers were provided instruction to address affected devices in their inventory. They should address this by either selecting Action A or B: Action A: Users should perform a check of the device before use and, as stated in the IFU, patties should not be used if in an open or damaged pouch. For the affected lots, special attention should be given to lower portion of the pouch where the Tyvek and clear film are sealed. Action B: Place the affected lots in quarantine and contact American Surgical Company for an RMA and replacement. All affected devices are eligible for replacement by ASC. Complete and return confirmation of actions described in Field Safety Notice. Contact:Tel: 1-800-343-0060 Option 2, Fax: 781-595-5460 Email: CustomerService@AmericanSurgical.com
Quantity in Commerce3 boxes
DistributionWorldwide Distribution - US Nationwide and the countries of Australia, Belgium, Canada, France, Germany, Greece, Hong Kong, Hong Kong, Israel, Italy, Italy, Japan, Netherlands, New Zealand, Portugal, Saudi Arabia, South Africa, Spain, Sweden, Switzerland, and United Kingdom.
Total Product Life CycleTPLC Device Report