| Date Initiated by Firm | July 08, 2016 |
| Create Date | August 17, 2016 |
| Recall Status1 |
Terminated 3 on August 08, 2018 |
| Recall Number | Z-2572-2016 |
| Recall Event ID |
74602 |
| 510(K)Number | K926481 |
| Product Classification |
Electrosurgical, cutting & coagulation & accessories - Product Code GEI
|
| Product | Valleylab Laparoscopic Wire J-Hook Electrode Retractable 28CM
Item Code: E2782R28ASP
The Valleylab Laparoscopic Handset and Electrodes are indicated for use in laparoscopic and thoracoscopic surgical procedures where the delivery of electrosurgical current, irrigation and suction via a single handset is desirable |
| Code Information |
Affected Lot Range: 213559X through 60520066X |
| FEI Number |
1219930
|
Recalling Firm/ Manufacturer |
Medtronic 60 Middletown Ave North Haven CT 06473-3908
|
Manufacturer Reason for Recall | Potential for compromise of the package resulting in a breach of the sterile barrier |
FDA Determined Cause 2 | Process control |
| Action | Medtronic/Covidien sent an Urgent Medical Device Recall letter dated July 8, 2016, to all affected customers via Federal Express. The letter identified the product, the problem, and the action to be taken by the customer. The letter instructs customers to discontinue use of the product and return all inventory to Medtronic (Covidien), 195 McDermott Road, North Haven, CT 06473, Attention: Field Returns Department. Customers with questions were instructed to contact their Medtronic representative or Customer Service at 800-882-5878. |
| Quantity in Commerce | 1440 |
| Distribution | Worldwide Distribution - US (nationwide) and Internationally to
Australia, Canada, Japan, Korea, Hong Kong, Thailand, |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
| 510(K) Database | 510(K)s with Product Code = GEI
|