| Date Initiated by Firm | July 15, 2016 |
| Create Date | October 25, 2016 |
| Recall Status1 |
Terminated 3 on June 05, 2017 |
| Recall Number | Z-0098-2017 |
| Recall Event ID |
75186 |
| Product Classification |
Calculator/data processing module, for clinical use - Product Code JQP
|
| Product | Modular Pre Analytics ( EVO/Plus)
Product Usage:
The RSD 800A is a computer-controlled stand-online system for decapping, aliquoting and sorting of primary tubes and for producing and sorting bar coded secondary tubes.. The system operates with an online connection to the host. |
| Code Information |
Not applicable |
| FEI Number |
1823260
|
Recalling Firm/ Manufacturer |
Roche Diagnostics Operations, Inc.
9115 Hague Rd
Indianapolis
IN
46256-1025
|
| For Additional Information Contact | Roche Support Network Customer Support C 800-428-2336 |
Manufacturer Reason for Recall | Roche has received four reports of injury caused by the sharp edge of the center guide rail of an
analyzers sample rack tray. The cleaning procedure for the sample rack tray has been revised to clarify the cleaning process. |
FDA Determined Cause 2 | Device Design |
| Action | On 7/15/2016, URGENT MEDICAL DEVICE CORRECTION notifications were sent to the affected customers via courier. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, an updated cleaning procedure, and instructions for responding to the formal recall notification. Please contact the Roche Support Network Customer Support Center at 1-800-428-2336 if you have questions about the information contained in this UMDC.
On 10/5/2016, URGENT MEDICAL DEVICE CORRECTION UPDATE notifications were sent to the affected customers via courier. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. The notification updated the 7/15/16 notification to include additional affected systems and to provide updated cleaning instructions for all systems via the enclosed Operators Manual addendum. |
| Quantity in Commerce | 9014 in total |
| Distribution | US Nationwide Distribution |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
|