|
Class 2 Device Recall Cobas b 123 POC system |
 |
Date Initiated by Firm |
September 03, 2015 |
Date Posted |
January 23, 2017 |
Recall Status1 |
Terminated 3 on June 05, 2017 |
Recall Number |
Z-1077-2017 |
Recall Event ID |
75709 |
510(K)Number |
K111188
|
Product Classification |
Electrode measurement, blood-gases (pco2, po2) and blood ph - Product Code CHL
|
Product |
Cobas b 123 POC system The cobas b 123 POC system is a fully automated POC system for whole blood in vitro measurement of pH, blood gases (B3G), electrolytes Nat, K+, iCaWt (ISE), hematocrit (THct), metabolites (Glu, Lac), total hemoglobin (tHb), hemoglobin derivatives (O2Hb, HHb, COT~b, MetHb), and oxygen saturation (SO2). In addition, the cobas b 123 POC system calculates derived parameters. It is dedicated for use in a Point-of-Care environment and laboratory. The integrated AutoQC module and the oximeter module are available as an option |
Code Information |
Models: cobas b 123 <2> POC system: 05122252001 cobas b 123 < 4 > POC system: 05122287001 Serial numbers: 11074 2274 2550 2302 2304 2309 1253 2420 2337 2147 2389 2418 2425 1749 2462 2338 2339 1850 2328 2500 1374 2301 1349 2400 2344 2350 1860 1941 2006 2128 |
Recalling Firm/ Manufacturer |
Roche Diagnostics Operations, Inc. 9115 Hague Rd Indianapolis IN 46256-1025
|
For Additional Information Contact |
800-526-2272
|
Manufacturer Reason for Recall |
under specific settings, an issue may occur during simultaneous Sensor Cartridge and Fluid Pack change on the cobas b 123 <2> POC system and cobas b 123 <4> POC system.
The issue occurs when the software function [AutoQC as follow-up] is configured to run all three levels of AutoQC only after a Fluid Pack change, but not after a Sensor Cartridge change. When both are changed simultaneously, starting with the Sensor Cartridge and followed by the Fluid Pack, the analyzer carries out only the follow-up actions associated with the Sensor Cartridge change after completing the change workflow. As a result, no follow-up AutoQC is performed and the three expected AutoQC measurements for the Fluid Pack change are not carried out.
Without running quality control, there is a remote possibility that system issues would not be detected and wrong results would not be excluded on all parameters: pH, PO2, PCO2, Na+, K+, Ca++, Cl-, Glu, Lac, Hct, SO2, O2Hb, COHb, MetHb, HHb, and Bili.
|
FDA Determined Cause 2 |
Software design |
Action |
Roche issued an Analyzer Bulletin on September 3, 2015, to all affected Consignees. The recall notifications included a description of the reason for the recall, affected product, and consignee responsibilities; the notification did not include instructions for responding to the notification.
Actions Required
" Follow the workaround outlined in this Analyzer Bulletin to avoid the described issue.
" File this Analyzer Bulletin for future reference.
Questions
Please contact the Roche Support Network Customer Support Center at 1-800-526-2272 if you have questions about the information contained in this Analyzer Bulletin. |
Quantity in Commerce |
30 |
Distribution |
Nationwide Distribution including CA, LA, NE, PR, SC, TX, VA, and WV |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database |
510(K)s with Product Code = CHL and Original Applicant = Roche Diagnostics
|
|
|
|