| | Class 2 Device Recall Adelante Breezway |  |
| Date Initiated by Firm | October 05, 2016 |
| Create Date | December 19, 2016 |
| Recall Status1 |
Terminated 3 on July 26, 2019 |
| Recall Number | Z-0803-2017 |
| Recall Event ID |
75785 |
| 510(K)Number | K101497 K122958 |
| Product Classification |
Introducer, catheter - Product Code DYB
|
| Product | OSCOR ADELANTE BREEZEWAY 10F C55 S79CM
D84CM Catalog # AB101076
Is designed to facilitate the introduction of catheters to any of the heart chambers, including the left atrium via a transseptal puncture. |
| Code Information |
Lot #'s: C1-12572 Exp. June 17, 2019, C1-10896 Exp. July 16, 2018, C1-10896 Exp. July 16, 2018, C1-12139 Exp. November 24, 2018, C1-12572 Exp. June 17, 2019, C1-09133 Exp. August 3, 2017, C1-09133 Exp. August 3, 2017, C1-09943 Exp. March 19, 2018, OR-04641 Exp. March 26, 2018, and C1-10896 Exp. July 16, 2018 |
| FEI Number |
1035166
|
Recalling Firm/ Manufacturer |
Oscor, Inc.
3816 Desoto Blvd
Palm Harbor
FL
34683-1618
|
| For Additional Information Contact | Customer Service 727-937-2511 Ext. 133 |
Manufacturer Reason for Recall | Complaint that during insertion of the dilator through the sheath a fragment of inner liner came out of the sheath prior to use. |
FDA Determined Cause 2 | Nonconforming Material/Component |
| Action | Oscor sent two(2) Product Recall notifications:
-HOSPITALS: Immediately check your inventory to confirm that you do or do not have units from these lot(s) in your possession. Identify and set aside any unit from the identified lot in a manner that ensures the affected product will not be used. Check all storage and usage. Review, complete, sign and return the enclosed Acknowledgement Form attached in the letter, directly to Oscor Inc. at the fax number or e-mail on the form, for immediate replacement (as applicable).
Who to call: Please call Oscor Inc. Customer Service Department at 727-937-2511 or email RGA@Oscor.com Hours of operations: Monday to Friday from 8:30AM to 5:30PM Eastern Time.
- DISTRIBUTORS: Immediately check your inventory to confirm that you do or do not have units from these lot(s) in your possession. Identify and set aside any unit from the identified lot in a manner that ensures the affected product will not be used. Check all storage and usage. Please pull a list of the hospitals (end user) impacted by the affected lots and retrieve inventory for replacement, if available. Review, complete, sign and return the enclosed Acknowledgement Form (Exhibit A) attached in this letter, directly to Oscor Inc. at the fax number or e-mail on the form, for immediate replacement (as applicable). Who to call: Please call Oscor Inc. Customer Service Department at 727-937-2511 or email RGA@Oscor.com Hours of operation: Monday to Friday from 8:30AM to 5:30PM Eastern Standard Time. |
| Quantity in Commerce | 74 |
| Distribution | Worldwide Distribution - US including CA, NJ, MD, TX, FL, NC, MO, and Puerto Rico. Internationally to AUSTRIA, BOLIVIA, CANADA, CHILE, CYPRUS, NETHERLANS, EGYPT, GERMANY, HONG KONG, ISRAEL, ITALY, SWITZERLAND, and VIETNAM |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
|