• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall 5.9CM ATTACHMENT, BLACK MAX

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall 5.9CM ATTACHMENT, BLACK MAXsee related information
Date Initiated by FirmOctober 06, 2016
Create DateMay 04, 2017
Recall Status1 Terminated 3 on July 26, 2019
Recall NumberZ-1982-2017
Recall Event ID 76059
510(K)NumberK974025 
Product Classification Table and attachments, operating-room - Product Code BWN
Product5.9CM ATTACHMENT, BLACK MAX Product Usage: Pneumatic system
Code Information All serial numbers shipped to customers prior to 09/27/2016; DFU 18-0062; Part No. B-TURQUOISE.
FEI Number 1045834
Recalling Firm/
Manufacturer
The Anspach Effort, Inc.
4500 Riverside Dr
Palm Beach Gardens FL 33410-4235
For Additional Information ContactAnne Brisson
800-327-6687
Manufacturer Reason
for Recall
Supplied Directions for Use (DFU) did not contain a recommended service interval.
FDA Determined
Cause 2
Under Investigation by firm
ActionConsignees were sent a recall notice via mail, dated 10/10/2016. In the notice, the firm notifies consignees of the Directions For Use (DFU) correction that will include recommended servicing intervals. All DFUs distributed after 09/28/2016 will have the correct DFU. The recalling firm instructs consignees to review the service intervals provided for product received, forward the notification to necessary parties and/or facilities, and complete the attached Verification Form. Complete Verification Forms are to be sent by fax to 877-546-5069 or by email to Anspach4870@stericycle.com. Consignees are also asked to maintain a copy of the notice for their records. Any questions can be directed to Customer Service at 1-800-327-6887, option 1.
Quantity in Commerce49
DistributionWorldwide Distribution - US Nationwide in the states of HI, DC including PR and the countries of Guam and Japan
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
510(K) Database510(K)s with Product Code = BWN
-
-