| | Class 2 Device Recall V60 Ventilator with Proportional Pressure Ventilation and AutoTrak Software Options |  |
| Date Initiated by Firm | February 01, 2017 |
| Create Date | March 27, 2017 |
| Recall Status1 |
Terminated 3 on June 26, 2018 |
| Recall Number | Z-1687-2017 |
| Recall Event ID |
76377 |
| 510(K)Number | K102985 |
| Product Classification |
Ventilator, continuous, minimal ventilatory support,facility use - Product Code MNT
|
| Product | Philips V60 Ventilator with Version 2.20 Software,
Description: V
60 Ventilator,Intl Opt: CFLEX,AVAPS,PPV
V60 Ventilator,Intl Opt: CFLEX,AVAPS
V60 Ventilator,Intl Opt: None
V60 Ventilator,English Opt: None
V60 Ventilator,Engl Opt: CFLEX,AVAPS
V60 Us Demo Unit
V60 USED ENGL OPT:CFLEX, AVAPS,PPV,AT+cl
V60 VENT, JAPAN OPT: CFLEX, AVAPS, AT+
V60 VENT, JAPAN OPT: CFLEX, AVAPS, AT+
Catalog/REF No. 1053613, 1053614, 1053615, 1053616, 1053617, DU1053617, U1053617, 1076709, R1076709 |
| Code Information |
Serial Number 100190253 100195884 201023008 201024196 201025984 201025987 201025988 201025947 201025972 201025996 201026002 201023858 201026449 201026531 201026645 201026658 201026668 201025266 201026252 201026254 201026431 201026433 201026434 201026435 201026436 201026439 201026569 201026577 201026579 201026598 201026601 201026605 201026636 201026639 201026640 201026642 201026646 201026649 201026651 201026652 201026653 201026655 201026659 201026662 201026665 201026667 201026672 201026673 201026680 201026685 201026686 201026687 201026689 201026690 201026691 201026692 201026693 201026694 201026695 201026699 201026700 201026701 201026702 201026705 201026706 201026708 201026709 201026710 201026711 201026712 201026715 201026716 201026717 201026741 201026742 201026744 201026748 201026749 201026753 201026755 201026765 201026782 201024190 201026534 201025309 201025864 201025994 201026438 201024086 201025167 201025173 201025179 201025382 201025402 201025871 201026660 201026671 201026681 201023832 201023850 201023856 201023871 201024010 201024061 201026525 201026568 201026570 201026573 201026574 201026575 201026576 201026602 201026641 201026647 100189742 100189753 100190088 100190090 100190095 100190735 100190738 100190809 100190810 100190811 100190813 100195313 100195319 100195649 100195864 100195896 201022731 201022737 201022738 201022761 201022768 201023092 201023097 201023100 201023101 201023845 201026524 201026567 201026592 201026688 201026704 201026719 201026746 201026752 201026256 201026257 201026532 201026572 201026578 201026644 201026650 201026661 201026714 201026740 201026757 201026780 100190093 100190737 100195641 201023886 201024070 201022727 201023091 201023674 201023675 201024188 201024189 201024193 201024194 201024195 201024198 201024199 201025160 201025166 201025169 201025170 201025171 201025172 201025174 201025175 201025176 201025180 201025185 201025190 201025193 201025194 201025197 201025296 201025316 201025394 201025396 201025397 201025399 201025401 201025407 201025909 201025910 201025921 201025928 201025944 100195312 100195316 100195441 100195442 100195444 100195445 100195446 100195448 100195449 100195451 100195452 100195453 100195455 100195456 100195458 100195465 100195470 100195471 100195472 100195473 100195637 100195640 100195648 100195650 100195653 100195654 100195655 100195656 100195659 100195665 100195668 100195670 100195861 100195870 100195874 100195886 100195893 100195900 100195962 100195963 100195964 100195966 100195973 100195976 100196001 100196014 100196023 100196133 100196138 100196429 100196431 100196436 100196620 100196622 100196623 100196624 100196625 100196627 201022722 201022724 201022725 201022728 201022729 201022730 201022732 201022734 201022739 201022763 201022765 201022766 201022771 201022772 201022967 201022983 201023009 201023052 201023055 201023086 201023087 201023105 201023109 201023110 201023265 201023268 201023270 201023271 201023272 201023276 201023278 201023279 201023668 201023670 201023671 201023672 201023673 201023676 201023677 201023678 201023679 201023841 201023843 201023844 201023846 201023847 201023849 201023861 201023863 201023864 201023868 201023876 201023879 201023883 201023885 201023887 201023888 201023889 201023890 201023892 201023896 201023898 201023899 201023900 201023902 201023905 201023911 201023914 201023922 201023926 201023928 201023929 201023931 201024000 201024003 201024005 201024006 201024008 201024014 201024015 201024030 201024033 201024831 201024832 201024833 201024834 201024835 201024836 201024837 201024838 201024839 201025196 201025198 201025200 201025202 201025280 201025281 201025282 201025283 201025284 201025285 201025286 201025287 201025288 201025289 201025290 201025291 201025294 201025295 201025297 201025298 201025299 201025300 201025301 201025302 201025304 201025306 201025307 201025308 201025310 201025311 201025312 201025313 201025314 201025315 201025317 201025380 201025381 201025383 201025384 201025385 201025386 201025387 201025388 201025389 201025390 201025391 201025392 201025395 201025398 201025403 201025405 201025406 201025855 201025856 201025857 201025858 201025862 201025865 201025866 201025867 201025868 201025870 201025913 201025915 201025920 201025925 201025929 201025931 201025932 201025935 201025937 201025940 201025941 201025942 201025945 201025948 201025949 201025950 201025951 201025952 201025953 201025955 201025956 201025957 201025960 201025961 201025962 201025963 201025964 201025965 201025967 201025968 201025969 201025970 201025971 201025973 201025974 201025975 201025976 201025977 201025978 201025979 201025980 201025981 201025982 201025983 201025985 201025986 201025989 201025990 201025991 201025993 201025997 201025998 201025999 201026000 201026001 201026003 201026350 201026351 201026352 201026353 201026354 201026356 201026357 201026358 201026359 201026380 201026381 201026382 201026383 201026384 201026385 201026386 201026387 201026388 201026389 201026390 201026391 201026392 201026393 201026394 201026395 201026396 201026397 201026398 201026399 201026400 201026401 201026402 201026403 201026404 201026405 201026406 201026407 201026408 201026409 201026410 201026411 201026412 201026414 201026415 201026417 201026418 201026419 201026420 201026421 201026422 201026423 201026424 201026425 201026426 201026427 201026428 201026429 201026440 201026441 201026442 201026443 201026445 201026446 201026450 201026451 201026452 201026453 201026455 201026456 201026457 201026459 201026462 201026463 201026465 201026466 201026468 201026470 201026471 201026472 201026476 201026477 201026478 201026521 201026522 201026523 201025168 201025178 201025182 201025184 201025189 201025191 201025908 201025911 201025912 201025914 201025916 201025917 201025918 201025919 201025922 201025924 201025927 201025930 201025934 201025936 201025938 201025939 201025943 201025946 201025992 100196404 201023059 201025393 201025923 201025926 201025933 201024191 201026432 201026750 100189759 100190736 100190803 201026251 201026255 201026258 201026259 201026355 201026437 201026533 201026600 201026609 201026623 201026624 201026625 201026626 201026627 201026628 201026629 201026630 201026631 201026632 201026633 201026634 201026635 201026637 201026638 201026656 201026669 201026682 201026684 201026697 201026698 201026703 201026718 201026743 201026745 201026747 201026751 201026756 201026758 201026759 201026760 201026761 201026762 201026763 201026764 201026766 201026767 201026768 201026769 201026770 201026771 201026772 201026781 201026783 100190731 100195898 201022985 201023083 201023669 201024016 201024075 201024085 201025161 201025162 201025163 201025164 201025165 201025177 201025181 201025183 201025186 201025187 201025188 201025192 201025195 201025201 201025404 201025863 201025869 201025995 201026004 201023837 201023852 100189758 201022976 201023057 201023099 201023831 201023862 100190730 100195675 201022963 201023090 201025859 201025954 201025966 100190190 100190739 100190801 100190802 100195440 201024004 201024058 201024063 201024073 201024192 201024197 201025199 201025264 201025274 201025293 201025303 201025400 201026475 201026552 201026643 201026648 201026657 201026663 201026707 100136136 100136655 100141903 100146850 100146852 100149714 100153310 100158909 100160567 100160569 100166344 100172321 100173878 100193596
100193599 100193604 100193605 100193607 100193620 100194033 100194034 100194038 100194041 100194307 100194309 100194433 100194434 100194445 100194447 100194451 100194452 100194453 100194455 100194458 100194462 100194466 100194468 100194469 100194472 100194700 100194710 100194713 100194714 100194715 100194717 100194718 100194719 100194720 100194721 100194722 100194724 100194725 100194726 100194727 100194728 100194732 100194869 100194874 100194876 100194879 100194897 100194904 100194905 100194908 100194950 100194951 100194955 100194958 100195042 100195044 100195045 100195050 100195053 100195054 100195056 100195057 100195058 100195059 100195060 100195061 100195062 100195144 100195145 100195146 100195147 100195148 100195149 100195150 100195151 100195152 100195153 100195154 100195155 100195156 100195157 100195158 100195159 100195180 100195181 100195182 100195183 100195184 100195185 100195186 100195187 100195188 100195189 100195190 100195191 100195193 100195194 100195195 100195196 100195197 100195198 100195199 100196481 100196482 201009340 201013041 201013786 201015018 201015057 201015481 201017700 201022430 201022431 201022432 201022433 201022434 201022435 201022436 201022437 201022438 201022439 201022480 201022481 201022482 201022483 201022484 201022485 201022486 201022487 201022488 201022489 201022490 201022491 201022492 201022493 201022494 201022495 201022496 201022497 201022498 201022499 201022500 201022501 201022502 201022503 201022504 201022505 201022506 201022507 201022508 201022509 201022510 201022511 201022512 201022513 201022514 201022515 201022516 201022517 201022518 201022519 201022552 201022553 201022554 201022555 201022556 201022557 201022558 201022559 201022560 201022561 201022562 201022563 201022564 201022565 201022566 201022567 201022568 201022569 201022570 201022571 201022572 201022573 201022574 201022575 201022576 201022637 201022638 201022639 201022640 201022641 201022642 201022643 201022644 201022645 201022646 201022647 201022648 201022649 201022650 201022651 201022652 201022653 201022654 201022740 201022741 201022742 201022743 201022744 201022745 201022746 201022747 201022748 201022749 201022809 201022810 201022811 201022812 201022813 201022814 201022815 201022816 201022817 201022818 201022819 201022820 201022821 201022822 201022823 201022824 201022825 201022826 201022827 201022828 201022829 201022830 201022831 201022832 201022833 201022834 201022835 201022857 201022858 201022859 201022860 201022861 201022862 201022863 201022864 201022865 201022866 201022867 201022868 201022869 201022870 201022871 201022872 201022873 201022874 201022875 201022876 201022877 201022878 201022879 201022880 201022881 201022882 201022883 201022884 201022885 201022886 201023204 201023205 201023206 201023207 201023208 201023209 201023210 201023211 201023212 201023213 201023214 201023215 201023216 201023217 201023218 201023219 201023220 201023221 201023222 201023223 201023224 201023225 201023226 201023227 201023228 201023229 201023230 201023231 201023232 201023233 201023234 201023235 201023236 201023237 201023238 201023239 201023240 201023241 201023242 201023315 201023316 201023317 201023318 201023319 201023320 201023321 201023322 201023323 201023324 201023325 201023326 201023327 201023328 201023329 201023330 201023331 201023332 201023333 201023334 201023335 201023336 201023337 201023338 201023339 201023340 201023341 201023342 201023343 201023344 201023345 201023347 201023348 201023349 201023350 201023351 201023352 201023353 201023354 201023355 201023356 201023357 201023358 201023359 201023360 201023361 201023362 201023363 201023514 201023515 201023516 201023517 201023518 201023519 201023520 201023521 201023522 201023523 201023524 201023525 201023526 201023527 201023528 201023529 201023530 201023531 201023532 201023533 201023534 201023535 201023536 201023537 201023538 201023539 201023540 201023541 201023542 201023543 201023544 201023545 201023546 201023547 201023548 201023549 201023550 201023551 201023552 201023553 201023554 201023555 201023556 201023557 201023558 201023559 201023560 201023561 201023562 201023563 201023572 201023573 201023574 201023575 201023576 201023577 201023578 201023579 201023580 201023581 201023582 201023583 201023584 201023585 201023586 201023587 201023588 201023589 201023590 201023591 201023592 201023593 201023594 201023595 201023596 201023597 201023598 201023599 201023600 201023601 201023602 201023603 201023604 201023605 201023606 201023607 201023608 201023609 201023610 201023611 201023612 201023613 201023619 201023726 201023727 201023728 201023729 201023730 201023731 201023732 201023733 201023734 201023735 201023736 201023737 201023738 201023739 201023740 201023741 201023742 201023743 201023744 201023745 201023746 201023747 201023748 201023749 201023750 201023751 201023752 201023753 201023754 201023755 201023756 201023757 201023758 201023759 201023760 201023761 201023762 201023763 201023764 201023765 201023766 201023767 201023768 201023769 201023770 201023771 201023772 201023773 201023774 201023775 201023780 201024217 201024218 201024219 201024225 201024226 201024227 201024228 201024229 201024230 201024231 201024232 201024233 201024234 201024235 201024236 201024237 201024238 201024239 201024267 201024268 201024269 201024270 201024271 201024272 201024273 201024274 201024275 201024276 201024277 201024278 201024279 201024280 201024281 201024282 201024283 201024284 201024285 201024286 201024287 201024288 201024289 201024290 201024291 201024292 201024293 201024294 201024295 201024296 201024297 201024298 201024299 201024300 201024301 201024302 201024303 201024304 201024305 201024306 201024307 201024308 201024309 201024310 201024311 201024312 201024313 201024314 201024315 201024316 201024335 201024336 201024337 201024338 201024339 201024340 201024341 201024342 201024343 201024344 201024345 201024346 201024347 201024348 201024349 201024351 201024352 201024353 201024354 201024355 201024356 201024357 201024358 201024359 201024360 201024361 201024362 201024363 201024364 201024365 201024366 201024367 201024368 201024369 201024370 201024371 201024372 201024373 201024374 201024375 201024376 201024377 201024378 201024379 201024380 201024381 201024382 201024383 201024384 201024385 201024386 201024400 201024401 201024402 201024403 201024404 201024405 201024406 201024407 201024408 201024409 201024410 201024411 201024412 201024413 201024414 201024415 201024416 201024417 201024418 201024419 201024420 201024421 201024422 201024423 201024424 201024425 201024426 201024427 201024428 201024429 201024430 201024431 201024432 201024433 201024434 201024435 201024436 201024437 201024438 201024439 201024440 201024441 201024442 201024443 201024444 201024445 201024446 201024447 201024448 201024460 201024461 201024462 201024463 201024464 201024465 201024466 201024467 201024468 201024469 201024470 201024471 201024472 201024473 201024474 201024475 201024476 201024477 201024478 201024479 201024480 201024481 201024482 201024484 201024680 201024681 201024682 201024683 201024684 201024685 201024686 201024687 201024688 201024689 201024690 201024691 201024692 201024693 201024694 201024695 201024696 201024697 201024698 201024732 201024733 201024735 201024736 201024737 201024738 201024739 201024740 201024741 201024742 201024743 201024744 201024745 201024746 201024747 201024748 201024749 201024750 201024751 201024752 201024753 201024754 201024755 201024756 201024757 201024758 201024759 201024760 201024761 201024762 201024763 201024764 201024765 201024766 201024767 201024768 201024769 201024770 201024771 201024772 201024773 201024774 201024775 201024776 201024777 201024778 201024779 201024780 201024781 201024860 201024861 201024862 201024863 201024864 201024865 201024866 201024867 201024868 201024869 201024870 201024871 201024872 201024873 201024874 201024875 201024876 201024877 201024878 201024879 201024880 201024881 201024882 201024883 201024884 201024885 201024886 201024887 201024888 201024889 201024890 201024891 201024892 201024893 201024894 201024895 201024896 201024897 201024898 201024899 201024900 201024901 201024902 201024903 201024904 201024905 201024906 201024907 201024908 201024909 201024960 201024961 201024962 201024963 201024964 201024965 201024966 201024967 201024968 201024969 201024970 201024971 201024972 201024973 201024974 201024975 201024976 201024977 201024978 201024979 201024980 201024981 201024982 201024983 201024984 201024985 201024986 201024987 201024988 201024989 201024990 201024991 201024992 201024993 201024994 201024995 201025104 201025105 201025106 201025107 201025108 201025109 201025110 201025111 201025112 201025113 201025114 201025115 201025116 201025117 201025118 201025119 201025120 201025121 201025122 201025123 201025124 201025125 201025126 201025127 201025128 201025129 201025130 201025131 201025132 201025133 201025134 201025135 201025136 201025137 201025138 201025139 201025140 201025141 201025142 201025233 201025234 201025235 201025236 201025237 201025238 201025239 201025368 201025369 201025370 201025371 201025372 201025373 201025374 201025375 201025376 201025377 201025378 201025379 201025480 201025481 201025482 201025483 201025484 201025485 201025486 201025487 201025488 201025489 201025490 201025491 201025492 201025493 201025494 201025495 201025496 201025497 201025498 201025499 201025500 201025501 201025502 201025503 201025504 201025505 201025506 201025507 201025508 201025509 201025510 201025511 201025512 201025513 201025514 201025515 201025516 201025517 201025518 201025519 201025520 201025521 201025522 201025598 201025599 201025640 201025641 201025642 201025643 201025644 201025645 201025646 201025647 201025648 201025649 201025650 201025651 201025652 201025653 201025654 201025655 201025656 201025657 201025658 201025659 201025660 201025661 201025662 201025663 201025664 201025665 201025666 201025668 201025669 201025670 201025671 201025672 201025673 201025674 201025675 201025676 201025677 201025678 201025679 201025680 201025681 201025682 201025683 201025684 201025685 201025686 201025687 201025688 201025689 201025690 201025691 201025692 201025693 201025694 201025695 201025696 201025697 201025698 201025699 201025700 201025701 201025702 201025703 201025704 201025705 201025706 201025707 201025708 201025709 201025710 201025711 201025712 201025713 201025714 201025715 201025716 201025717 201025718 201025719 201025720 201025721 201025722 201025723 201025724 201025725 201025726 201025727 201025781 201025782 201025783 201025784 201025785 201025786 201025787 201025788 201025789 201025790 201025791 201025792 201025793 201025794 201025795 201025796 201025797 201025798 201025799 201025800 201025801 201025802 201025803 201025804 201025805 201025806 201025807 201025808 201025809 201025810 201025811 201025812 201025813 201025814 201025815 201025816 201025817 201025818 201025819 201025820 201025821 201025822 201025823 201025824 201025825 201025826 201025827 201025828 201025829 201025830 201025831 201025832 201025833 201025834 201025835 201025836 201025837 201025860 201025872 201025873 201025874 201025875 201025876 201025877 201025878 201025879 201026020 201026021 201026022 201026023 201026024 201026025 201026026 201026027 201026028 201026029 201026030 201026031 201026032 201026033 201026034 201026035 201026036 201026037 201026038 201026039 201026040 201026041 201026042 201026043 201026044 201026045 201026046 201026047 201026048 201026049 201026050 201026051 201026052 201026053 201026054 201026055 201026056 201026057 201026058 201026059 201026060 201026061 201026109 201026110 201026111 201026112 201026113 201026114 201026115 201026116 201026117 201026118 201026119 201026120 201026121 201026122 201026123 201026124 201026125 201026126 201026127 201026128 201026129 201026130 201026131 201026132 201026133 201026134 201026135 201026136 201026137 201026138 201026139 201026140 201026141 201026142 201026143 201026144 201026145 201026146 201026147 201026148 201026149 201026150 201026151 201026152 201026153 201026154 201026155 201026156 201026157 201026158 201026180 201026181 201026182 201026183 201026184 201026185 201026186 201026187 201026188 201026189 201026190 201026191 201026192 201026193 201026194 201026195 201026196 201026197 201026198 201026199 201026200 201026201 201026202 201026203 201026204 201026205 201026206 201026207 201026208 201026209 201026210 201026211 201026212 201026213 201026214 201026215 201026216 201026217 201026218 201026219 201026220 201026221 201026222 201026223 201026224 201026225 201026226 201026227 201026228 201026229 201026264 201026265 201026266 201026267 201026268 201026269 201026270 201026271 201026272 201026273 201026274 201026275 201026276 201026277 201026278 201026279 201026280 201026281 201026282 201026283 201026284 201026285 201026286 201026287 201026288 201026289 201026290 201026291 201026292 201026293 201026294 201026295 201026296 201026297 201026298 201026299 201026300 201026301 201026302 201026303 201026304 201026305 201026306 201026307 201026308 201026309 201026310 201026311 201026312 201026313 201026314 201026315 201026316 201026317 201026318 201026319 201026320 201026321 201026322 201026323 201026324 201026325 201026326 201026327 201026328 201026329 201026330 201026331 201026332 201026333 201026334 201026335 201026336 201026337 201026338 201026339 201026340 201026341 201026342 201026343 201026344 201026345 201026722 201026723 201026724 201026725 201026726 201026727 201026728 201026729 201026730 201026731 201026732 201026733 201026734 201026735 201026736 201026737 201026738 201026739 201026784 201026785 201026786 201026787 201026788 201026789 201026790 201026791 201026792 201026793 201026794 201026795 201026796 201026797 201026798 201026799 201026800 201026801 201026802 201026803 201026804 201026805 201026806 201026807 201026808 201026809 201026810 201026811 201026812 201026813 201026814 201026815 201026816 201026817 201026818 201026819 201026820 201026821 201026822 201026823 201026824 201026825 201026826 201026827 201026828 201026829 201026830 201026831 201026832 201026833 201026835 201026836 201026837 201026838 201026839 201026849 201026850 201026851 201026852 201026853 201026854 201026856 201026857 201026858 201026859 201026860 201026861 201026862 201026863 201026864 201026865 201026866 201026867 201026868 201026869 201026870 201026871 201026872 201026873 201026874 201026875 201026876 201026877 201026878 201026879 201026880 201026881 201026882 201026883 201026884 201026885 201026886 201026887 201026888 201026889 201026890 201026891 201026904 201026905 201026906 201026907 201026908 201026909 201026910 201026911 201026912 201026913 201026914 201026915 201026916 201026917 201026918 201026919 201026920 201026921 201026922 201026923 201026924 201026925 201026926 201026927 201026928 201026929 201026930 201026931 201026932 201026933 201026934 201026935 201026936 201026937 201026938 201026939 201026940 201026941 201026942 201026943 201026944 201026945 201026946 201026947 201026948 201026949 201026950 201026951 201026952 201026953 201026954 201026955 201026956 201026957 201026958 201026959 201026960 201026961 201026962 201026963 201026964 201026965 201026966 201026967 201026968 201026969 201026970 201026971 201026972 201026973 201026974 201026975 201026976 201026977 201026978 201026979 201026980 201026981 201026982 201026983 201026984 201026985 201026986 201026987 201026988 201026989 201026990 201026991 201026992 201026993 201026994 201026995 201026996 201026997 201026998 201026999 201027000 201027001 201027002 201027003 201027062 201027063 201027064 201027065 201027066 201027067 201027068 201027069 201027070 201027071 201027072 201027073 201027074 201027075 201027076 201027077 201027078 201027079 201027080 201027081 201027082 201027083 201027084 201027085 201027086 201027087 201027088 201027089 201027090 201027091 201027092 201027093 201027094 201027095 201027096 201027097 201027098 201027099 201027100 201027101 201027102 201027103 201027104 201027105 201027106 201027107 201027108 201027109 201027110 201027111 201027112 201027113 201027114 201027115 201027116 201027117 201027118 201027119 201027120 201027121 201027122 201027123 201027124 201027126 201027127 201027128 201027129 201027130 201027132 201027134 201027137 201027139 201027160 201027161 201027162 201027163 201027164 201027165 201027166 201027167 201027168 201027169 201027170 201027171 201027172 201027173 201027174 201027176 201027177 201027178 201027179 201027180 201027181 201027182 201027183 201027184 201027185 201027186 201027187 201027188 201027189 201027190 201027191 201027200 201027201 201027202 201027203 201027204 201027205 201027206 201027207 201027208 201027209 201027210 201027211 201027212 201027213 201027214 201027215 201027216 201027217 201027218 201027219 201027221 201027222 201027223 201027224 201027225 201027226 201027227 201027228 201027229 201027230 201027231 201027232 201027233 201027234 201027235 201027236 201027237 201027238 201027239 201027240 201027241 201027242 201027243 201027244 201027245 201027246 201027248 201027249 201027250 201027251 201027252 201027253 201027255 201027256 201027257 201027258 201027259 201027260 201027261 201027262 201027263 201027265 201027267 201027269 201027270 201027271 201027272 201027275 201027276 201027277 201027278 201027280 201027281 201027282 201027283 201027284 201027286 201027287 201027288 201027320 201027321 201027322 201027323 201027324 201027328 201027329 201027331 201027333 201027334 201027335 201027336 201027338 201027339 201027340 201027341 201027343 201027344 201027345 201027347 201027349 201027350 201027351 201027354 201027355 201027357 201027360 201027362 201027363 201027366 201027368 201027370 201027372 201027374 201027377 201027382 201027383 201027384 201027385 201027386 201027387 201027388 201027389 201027390 201027391 201027392 201027393 201027394 201027395 201026696 201026754 201027307 201027308 201027309 201027310 201027318 201027432 201027433 201027434 201027436 201027440 |
| FEI Number |
2518422
|
Recalling Firm/ Manufacturer |
Respironics California Inc 2271 Cosmos Ct Carlsbad CA 92011-1517
|
| For Additional Information Contact | Christopher D. Barnette 760-918-7300 |
Manufacturer Reason for Recall | The V60 Ventilator with Version 2.20 software installed may falsely detect that the blower motor has stalled. If this condition occurs, the software will cause the ventilator to shut down (Vent Inop) and display Error Code 100E. Ventilatory support will cease. |
FDA Determined Cause 2 | Device Design |
| Action | Philips sent an Urgent- Medical Device Letter dated February 1, 2017, to all affected customers. It will be sent the week of February 6, 2017, to inform customers that Respironics is voluntarily recalling all Philips V60 Non Invasive Ventilators (NIV) that have Version 2.20 Software installed. Customers are informed that the V60 Ventilator with Version 2.20 software installed may falsely detect that the blower motor has stalled. If this condition occurs, the software will cause the ventilator to shut down (Vent Inop) and display Error Code 100E.
Ventilatory support will cease. An audible high-priority alarm will sound continuously for at least 2 minutes when the V60 shuts down for any Vent Inop condition and is operating on battery power. If the V60 is connected to AC power (mains supply), the alarm will continue to sound until an operator intervenes. If the V60 is connected to a remote alarm system, the alarm system will be activated until action is taken by the operator. Error Code 100E only exists in V60 Version 2.20 software. Therefore, V60s running Version 2.10 software are not subject to this particular Vent Inop condition.
E-mail back the complete form attached to the rear of this recall notification to recall. response @ Philips. com or Fax to 1-877 499-7223. For further questions please call (760) 918-7300. |
| Quantity in Commerce | 2512 units |
| Distribution | Worldwide Distribution - US Distribution and to the countries of : Argentina, Austria, Belgium, Canada, Czech Republic, Finland, France, Germany, Hong Kong, Iceland, India, Indonesia, Italy, Japan, Kuwait, Malaysia, Mexico, Netherlands, Norway, Qatar, Saudi Arabia, Spain, Sweden, Switzerland, Turkey, United Arab Emirates, United Kingdom and Vietnam. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
| 510(K) Database | 510(K)s with Product Code = MNT
|
|
|
|