|
Class 2 Device Recall Merge Cardio |
|
Date Initiated by Firm |
April 04, 2016 |
Date Posted |
March 08, 2017 |
Recall Status1 |
Terminated 3 on March 20, 2017 |
Recall Number |
Z-1415-2017 |
Recall Event ID |
76401 |
510(K)Number |
K051649
|
Product Classification |
System, image processing, radiological - Product Code LLZ
|
Product |
Merge Cardio software using EchoIMS. The firm name on the label is Merge Healthcare, Hartland, WI. |
Code Information |
Version 9.0.9 and earlier using EchoIMS. |
Recalling Firm/ Manufacturer |
Merge Healthcare, Inc. 900 Walnut Ridge Dr Hartland WI 53029-8347
|
For Additional Information Contact |
Merge Service and Support 877-741-5369
|
Manufacturer Reason for Recall |
A situation can occur allowing two physicians to access the same study report in EchoIMS when launched from the Cardio Study List without receiving the read-only notification prompt.
|
FDA Determined Cause 2 |
Software design |
Action |
Merge sent an Urgent: Medical Device Recall letter dated March 28, 2016, via e-mail on April 4, 2016. A letter was issued to nonresponders on July 11, 2016.
The letter notifies the customer of the issue, informs them of the workaround, and that no further action was being taken to correct the issue, therefore, the customer needs to be aware of the possible behavior. The notification instructs the customer to pass the information onto all other users, as well as any downstream customers if it was further distributed. The notification also informs the customer their response is required using the enclosed form and the return addressed envelope. The response form asks, in part, if they understand the workaround. The response was required no later than April 18, 2016.
Nonresponders were issued a second letter dated July 11, 2016, asking for a response by August 2, 2016. The letter and response form contained the same information as the first letter. If the second letter results in additional unresponsive customers, a call campaign will be managed by Customer Support and documented via SalesForce. Three call attempts will made before closing the case.
For further information please call (877) 741-5369. |
Quantity in Commerce |
17 sites potentially have the affected versions |
Distribution |
US Distribution was made to medical facilities in CA, FL, IL, MD, MO, OH, OK, TX, VT, and WI. Military distribution was also made. |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database |
510(K)s with Product Code = LLZ and Original Applicant = CAMTRONICS MEDICAL SYSTEMS
|
|
|
|