• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall ARROWgard Blue YouBlend TwoLumen Hemodialysis Catheterization Kit for High Volume Infusions

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall ARROWgard Blue YouBlend TwoLumen Hemodialysis Catheterization Kit for High Volume Infusionssee related information
Date Initiated by FirmMay 09, 2017
Create DateJune 12, 2017
Recall Status1 Terminated 3 on September 17, 2018
Recall NumberZ-2334-2017
Recall Event ID 77232
510(K)NumberK993933 
Product Classification Catheter, hemodialysis, non-implanted - Product Code MPB
Product1) ARROWg+ard Blue You-Blend Two-Lumen Hemodialysis Catheterization Kit for High Volume Infusions The large-bore two-lumen catheter permits venous access to the central circulation for rapid fluid administration, temporary or acute hemodialysis, apheresis and hemofiltration. It may be inserted into the jugular, subclavia, or femoral veins Catheter, hemodialysis non-implanted
Code Information Material number: CDA-22122-U1A, CDC-22141-1A CDC-26142-1A  Device Listing D025726
FEI Number 3015859709
Recalling Firm/
Manufacturer
Arrow International Inc
2400 Bernville Rd
Reading PA 19605-9607
For Additional Information Contact
610-378-0131
Manufacturer Reason
for Recall
Arrow International is recalling the affected product because the packaging may not be completely sealed, which may compromise sterility.
FDA Determined
Cause 2
Packaging
ActionArrow International sent an Urgent Medical Device Recall Notification Letter dated May 11, 2017, to affected customers. The firm's notification letter is requesting that customers immediately assess their current inventory and to discontinue and quarantine any product with the specific lot codes listed in the letter. In addition, customers were asked to complete the Recall Acknowledgement form and fax or email it back to Customer Service so they can receive a Returns Good Authorization Number for the product's return. Customers with questions were instructed to contact their local sales representative or Customer Service at 1-866-246-6990.
Quantity in Commerce27,485 units distributed in the U.S., 4,371 units distributed internationally
DistributionWorldwide Distribution - US (nationwide) and Canada
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
510(K) Database510(K)s with Product Code = MPB
-
-