  Class 2 Device Recall Medtronic Suction Small AxiEM ENT see related information
Date Initiated by Firm June 29, 2017
Create Date August 21, 2017
Recall Status1 Terminated 3 on August 03, 2018
Recall Number Z-3054-2017
Recall Event ID 77739
510(K)Number K153555  
Product Classification Ear, Nose, and Throat Stereotaxic Instrument - Product Code PGW
Product Medtronic Suction 9734308 Small AxiEM ENT, reusable, Rx. The firm name on the label is Medtronic Navigation, Inc., Louisville, CO.
The suction devices are part of a family of reusable ENT instruments that are used with the StealthStationTM AxiEM surgical navigation system. The AxiEM system is an electromagnetic (EM) tracking solution for surgical tools. The system measures an induced voltage created by local magnetic fields from a set of coils in the transmit coil array (TCA). Each surgical tool has a set of receiving coils to sense the magnetic field strength. Using a mathematical algorithm, the position and orientation of the instrument can be computed and overlaid/registered to imagery on a computer monitor, resulting in the ability to track the instrument tip in relation to the patients anatomy. AxiEM ENT Suction devices are used as suction and pointing devices. The suction devices are intended to have an ENT Instrument Tracker mounted on it
Code Information Lot numbers 17022451,17011750, 16120349, 16082989, 16081888, 16071987, 16062713, and 16060712
Recalling Firm/
Manufacturer
Medtronic Navigation, Inc.
826 Coal Creek Cir
Louisville CO 80027-9710
For Additional Information Contact Tom F. Reimann
720-890-3241
Manufacturer Reason
for Recall
The instruments are not able to pass instrument verification which is performed prior to the use of each instrument.
FDA Determined
Cause 2
Device Design
Action Medtronic sent an Urgent Medical Device Recall letter dated June 25, 2017, via overnight mail on June 29, 2017, to all affected customers. Letters to foreign customers are being handled by geography-specific Regulatory Specialists. Customers were instructed examine their inventory for the affected product and immediately quarantine them for return for Medtronic. Customers with affected product were instructed to contact Medtronic Technical Services at 1-800-595-9709 to receive a return material authorization (RMA) and arrange for their return and no charge replacements. Customers were asked to complete and follow instructions on the attached Consignee Response Form. Sign and date the bottom of the form and then return the form to Medtronic at RS NavFCA@medtronic.com or fax it to Medtronic Technical Services at 651-367-7075. Customers with questions were instructed to call 1-800-595-9709.
Quantity in Commerce 60 instruments
Distribution Worldwide Distribution - US (nationwide) and Internationally to Canada.
Total Product Life Cycle TPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
510(K) Database 510(K)s with Product Code = PGW and Original Applicant = MEDTRONIC NAVIGATION, INC.
-
-