• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall Artis Q floor

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Artis Q floorsee related information
Date Initiated by FirmFebruary 18, 2016
Create DateAugust 29, 2017
Recall Status1 Terminated 3 on November 17, 2017
Recall NumberZ-3088-2017
Recall Event ID 77908
510(K)NumberK123529 
Product Classification Interventional fluoroscopic x-ray system - Product Code OWB
ProductArtis zee/zeego and Artis Q/Q.zen system, x-ray, angiographic AXIOM Artis zee and Q/Q.zen is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to , pediatric and obese patients. Procedures that can be performed with the AXIOM Artis-zee/zeego and Q/Q.zen family include cardiac angiography, neuro angiography, general angiography, rotational angiography, operating room angiography, multipurpose angiography and whole body radiographic/fluoroscopic procedures. AXIOM Artis zee/zeego and Q/Q.zen can also support the acquisition of position triggered imaging for spatial data synthesis
Code Information Artis Q floor - Serial number: 103030 103013 103039 103008 103035 103001 103028 103040 
Recalling Firm/
Manufacturer
Siemens Medical Solutions USA, Inc
40 Liberty Blvd
Malvern PA 19355-1418
For Additional Information ContactMeredith Adams
610-448-6461
Manufacturer Reason
for Recall
Due to a leak in the cooling system, fluids could potentially seep into the equipment cabinet. Loss of coolant can results in a cooling system failure and subsequent system failure. This leek can occur sporadically in affected systems.
FDA Determined
Cause 2
Unknown/Undetermined by firm
ActionSiemens mailed a Customer Safety Advisory Notice (CSAN) dated February 18, 2016, to affected customers informing them that their service organization will contact them to arrange a date to perform the corrective action (replacement of affected plastic couplings to prevent fluid leaks). Customers were advised to contact Siemens service organization for an appointment to schedule the work For questions regarding this recall call 610-.448-6461.
Quantity in Commerce986 units distributed in the USA
DistributionNationwide Distribution
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
510(K) Database510(K)s with Product Code = OWB
-
-