Date Initiated by Firm | July 27, 2017 |
Date Posted | January 11, 2018 |
Recall Status1 |
Terminated 3 on November 03, 2020 |
Recall Number | Z-0485-2018 |
Recall Event ID |
78042 |
510(K)Number | K964853 |
Product Classification |
Container, i.V. - Product Code KPE
|
Product | Empty IntraVia Container with Non-DEHP Fluid Path, 500mL
Containers are intended for use in the preparation and administration of drug admixtures. |
Code Information |
Product Code 2J8003, Lot No. DR16J18096. |
Recalling Firm/ Manufacturer |
Baxter Healthcare Corporation 1 Baxter Pkwy Deerfield IL 60015-4625
|
For Additional Information Contact | Center for One Baxter 800-422-9837 |
Manufacturer Reason for Recall | A leak may allow for microbial contamination of the sterile fluid path. |
FDA Determined Cause 2 | Process control |
Action | US consignees were notified via letter on July 27, 2017 and Canadian consignees were notified on July 31, 2017. Instructions included to locate and remove all affected product from the facility, contact Baxter Healthcare to arrange for return of product, complete and return the Reply Form, and to notify customers if the product was further distributed. For further questions, please call (800) 422-9837. |
Quantity in Commerce | 29,088 units |
Distribution | Distribution in the US and Canada. |
Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database | 510(K)s with Product Code = KPE
|