| | Class 2 Device Recall IMMULITE /IMMULITE 1000 OMMA (CA125) |  |
| Date Initiated by Firm | January 02, 2018 |
| Date Posted | May 04, 2018 |
| Recall Status1 |
Terminated 3 on February 10, 2023 |
| Recall Number | Z-1682-2018 |
| Recall Event ID |
79505 |
| 510(K)Number | K981297 |
| Product Classification |
Test, epithelial ovarian tumor-associated antigen (ca125) - Product Code LTK
|
| Product | IMMULITE /IMMULITE 1000 OM-MA (CA125) |
| Code Information |
OM-MA (CA125), OM-MA (CA125) (IMMULITE/IMMULITE 1000, LKOP1), (IMMULITE 2000/IMMULITE 2000 XPi, LKOP2) Lot codes: CA125 (OM-MA) 0211 0212 0213 |
| FEI Number |
2432235
|
Recalling Firm/ Manufacturer |
Siemens Healthcare Diagnostics, Inc.
511 Benedict Ave
Tarrytown
NY
10591-5005
|
| For Additional Information Contact | Ms. Aarti M. Aziz 914-631-8000 |
Manufacturer Reason for Recall | Concentrations for the level of biotin that does not interfere is not currently listed in the instruction for use for the Dimension Vista CTNI, Dimension Vista MMB, Dimension Sirolimus, ADVIA Centaur FOL, ADVIA Centaur aHBcM and the following IMMULITE assays, 3gAllergy Specific IgE, anti-HBc, BR-MA (CA15-3), CEA, Folic Acid, Gastrin, OM-MA (CA125), Thyroglobulin, and Vitamin B12, and these assays are susceptible to interference from biotin. |
FDA Determined Cause 2 | Error in labeling |
| Action | Siemens Healthcare Diagnostics issued a recall in March 2017 for products with incorrect units for biotin in the Instructions for Use. After further investigation, recalling firm determined that more Siemens products also had missing or incorrect information in the IFUs for which two recalls were initiated, and Amended Corrections and Removals Reports were submitted to the FDA |
| Distribution | Nationwide and Canada, Mexico, and Thailand |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
|