Class 2 Device Recall ADVIA Centaur DHEASO4

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall ADVIA Centaur DHEASO4see related information
Date Initiated by FirmJanuary 02, 2018
Date PostedMay 04, 2018
Recall Status1 Terminated 3 on February 10, 2023
Recall NumberZ-1688-2018
Recall Event ID 79505
Product Classification Radioimmunoassay, dehydroepiandrosterone (free and sulfate) - Product Code JKC
ProductADVIA Centaur DHEA-SO4
Code Information ADVIA Centaur¿ DHEA-SO4; radioimmunoassay, dehydroepiandrosterone (free and sulfate); SMN 10282227  ADVIA Centaur DHEA-S Kit Lot # 00119053 14041053 80449053 90520053 29750054 30423054 39155054 50533054 63267055 68936055 79273055 86427055 06402057 07109057 14187057 37499057 
FEI Number 2432235
Recalling Firm/
Manufacturer
Siemens Healthcare Diagnostics, Inc.
511 Benedict Ave
Tarrytown NY 10591-5005
For Additional Information ContactMs. Aarti M. Aziz
914-631-8000
Manufacturer Reason
for Recall
The concentrations for biotin listed in the non-interfering substances section of the instructions for use for the Dimension TSHL, Dimension LOCI FT4L, Dimension Vista TSH, ADVIA Centaur Cyclosporine, and ADVIA Centaur DHEAS, incorrectly state the level at which biotin does not interfere
FDA Determined
Cause 2
Error in labeling
ActionSiemens Healthcare Diagnostics issued a recall in March 2017 for products with incorrect units for biotin in the Instructions for Use. After further investigation, recalling firm determined that more Siemens products also had missing or incorrect information in the IFUs for which two recalls were initiated, and Amended Corrections and Removals Reports were submitted to the FDA
DistributionNationwide and Canada, Mexico, and Thailand
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.