Class 2 Device Recall JAMSHIDI (TJ) NEEDLE BM ASP MAC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall JAMSHIDI (TJ) NEEDLE BM ASP MACsee related information
Date Initiated by FirmFebruary 23, 2017
Create DateApril 18, 2018
Recall Status1 Terminated 3 on July 09, 2019
Recall NumberZ-1425-2018
Recall Event ID 76620
510(K)NumberK913306 
Product Classification Tray, surgical, needle - Product Code FSH
ProductJAMSHIDI (TJ) NEEDLE BM ASP MAC [a.) 13G x 3.5, b.) 8G x 4, c.) 11G x 3.5, d.) 8G x 6, e.) 11G x 6]; STERILE;
Code Information Product Code: a.) TJM3513 / LOT 0000836582, 0000838949, 0000843968, 0000849685, 0000852933; b.) TJM4008 / LOT 0000841349, 0000841348, 0000843961, 0000849132, 0000852123, 0000853877; c.) TJM4011 / LOT 0000836575, 0000838950, 0000840195, 0000843075, 0000843076, 0000846708, 0000852932, 0000853878, 0000854861, 0000854983; d.) TJM6008 / LOT 0000842725, 0000853971; e.) TJM6011 / LOT 0000843077, 0000849713, 0000854659
FEI Number 3008002452
Recalling Firm/
Manufacturer
Carefusion 2200 Inc
75 N Fairway Dr
Vernon Hills IL 60061-1845
For Additional Information ContactChristine Cox, RN
847-362-9439
Manufacturer Reason
for Recall
The integrity of the sterile packaging is potentially compromised.
FDA Determined
Cause 2
Package design/selection
ActionOn 2/23/2017, firm notified customers of recall via mailed letter. The letter included a recall response form. Customers were directed to either return the affected product along with the response form or destroy it and return the response form, indicating the product had been destroyed. All affected product will be destroyed.
Quantity in Commerce14,200 units
DistributionNationwide, Australia, Brazil, Canada, China, Japan, Kuwait, Malaysia, Mexico, Netherlands, New Zealand, Philippines, Singapore, South Africa, Taiwan, Thailand, United Arab Emirates, Vietnam
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
510(K) Database510(K)s with Product Code = FSH
-
-