 Class 2 Device Recall Radial Artery Catheterization Kitsee related information
Date Initiated by FirmApril 11, 2018
Create DateMay 30, 2018
Recall Status1 Terminated 3 on May 15, 2020
Recall NumberZ-1970-2018
Recall Event ID 79890
510(K)NumberK810675 
Product Classification Wire, guide, catheter - Product Code DQX
ProductRadial Artery Catheterization Kit; Product Codes: ASK-04020-AH, ASK-04020-MIHS, ASK-04120-HF1
Code Information Lot Numbers:  Product code ASK-04020-AH: 13F17B0102 13F17C0283 13F17F0221 13F17G0480 13F17J0045 13F17L0127   Product code ASK-04020-MIHS: 13F17A0174 13F17F0139   Product code ASK-04120-HF1: 13F17C0360 13F17D0034 13F17E0093 13F17F0468 13F17G0269 13F17J0101 13F17K0311 13F17L0489           
FEI Number 3015859709
Recalling Firm/
Manufacturer
Arrow International Inc
2400 Bernville Rd
Reading PA 19605-9607
Manufacturer Reason
for Recall
Product sterility may be compromised due to unsealed packaging.
FDA Determined
Cause 2
Under Investigation by firm
ActionOn April 11, 2018 Arrow International mailed Urgent Medical Device Recall notifications to affected customers. Customers were instructed to take the following actions: 1. If you have affected stock, immediately discontinue use and quarantine any products with the product codes and lot numbers listed above, so that the affected products can be returned to Arrow. 2. To return product, complete the enclosed Recall Acknowledgement Form and fax it to 1-855-419-8507, Attn: Customer Service or email to recalls@teleflex.com. This will allow us to document the amount of product you have on hand for return. A customer service representative will contact you with a Return Goods Authorization (RGA) Number and will provide instructions for the return of products to Arrow International. 3. If you have no affected stock, please complete the enclosed Recall Acknowledgment Form and fax it to 1-855-419-8507, Attn: Customer Service or email to recalls@teleflex.com. This will allow us to document your receipt of this letter. Customers with questions may contact your local sales representative or Customer Service at (866)246-6990.
Quantity in Commerce416,055 total products
DistributionUS Nationwide, AUSTRALIA, AUSTRIA, BELGIUM, CANADA, CZECH REPUBLIC, DENMARK, ECUADOR, FRANCE, GERMANY, GREECE, HONG KONG, INDONESIA, IRELAND, ITALY, JAPAN, KOREA, NETHERLANDS, NEW ZEALAND, PANAMA, PHILIPPINES, POLAND, PUERTO RICO, ROMANIA, SAUDI ARABIA, SINGAPORE, SLOVENIA, SOUTH AFRICA, SPAIN, SWEDEN, SWITZERLAND, THAILAND, TRINIDAD AND TOBAGO, UNITED KINGDOM
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
510(K) Database510(K)s with Product Code = DQX
-
-