| | Class 2 Device Recall QuickFlap |  |
| Date Initiated by Firm | June 22, 2018 |
| Create Date | August 29, 2018 |
| Recall Status1 |
Terminated 3 on July 08, 2019 |
| Recall Number | Z-2900-2018 |
| Recall Event ID |
80371 |
| 510(K)Number | K120352 |
| Product Classification |
Plate, cranioplasty, preformed, alterable - Product Code GWO
|
| Product | QUIKFLAP, 3 X 2-HOLE PLATES WITH TAB, 1.5X4mm SELF DRILLING SCREWS, Part Number 12-01522S, for use on the cranium. |
| Code Information |
1000096159, 1000096935, 1000096936, 1000097334, 1000097915, 1000101820, 1000109424, 1000115216, 1000119029, 1000125079, 1000125080, 1000130030, 1000136426, 1000141584, 1000147073, 1000151497, 1000156030, 1000157137, 1000162227, 1000163908, 1000167016, 1000169751, 1000176390, 1000178196, 1000181747, 1000184371, 1000186595, 1000187362, 1000191062, 1000193585, 1000194476, 1000199414, 1000205660, 1000210937, 1000211803, 1000212626, 1000214335, 1000222495, 1000233193, 1000239650, 1000247182, 1000256556, 1000266164, 1000270836, 1000277104, 1000283590, 1000287393, 1000294080, 1000086193, 1000298088 |
Recalling Firm/ Manufacturer |
Stryker Leibinger GmbH & Co. KG Botzinger Str. 41 Freiburg Im Breisgau Germany
|
| For Additional Information Contact | Julie Brown 269-389-3946 |
Manufacturer Reason for Recall | After a three year term, there is a potential for the external packaging (sterile barrier) to become compromised. |
FDA Determined Cause 2 | Package design/selection |
| Action | URGENT MEDICAL DEVICE RECALL letters dated 6/22/18 were sent to customers with the following requested actions to be taken:
1. Please inform users of this Medical Device Removal and pass this notice to all appropriate
individuals within your organization.
2. Return all affected products available at your location to:
Stryker CMF
Attn: Recall Coordinator
6300 Sprinkle Rd
Portage, MI 49002
REF: PFA #1739180
Or, Contact Stryker Customer Service and refer to PFA #1739180 with any questions about returning the product to us.
3. Complete and sign the enclosed Business Reply Form and fax a copy to: (877) 648-7114 or email a copy to CMF-custserv@stryker.com.
4. Keep a copy of the completed and executed Customer response Form for your records. |
| Quantity in Commerce | 22,831 total |
| Distribution | US distribution in the states: AZ, FL, NH, and OH, Australia, Canada, France, New Zealand, and UK. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
| 510(K) Database | 510(K)s with Product Code = GWO
|
|
|
|